Introduction

Diffusion-weighted imaging (DWI) is an important component of multi-parameter magnetic resonance imaging (MRI). The second edition of the Prostate Imaging Reporting and Data System version 2 (PI-RADS v2) indicates that DWI is the most important sequence for the diagnosis of peripheral prostate cancer1. The apparent diffusion coefficient (ADC) derived from the mono-exponential model is the most widely used parameter in clinical trials for prostate cancer diagnosis, tumour grading, early monitoring of prostate cancer responses to clinical treatment, and relapse assessment2,3,4. However, the diffusion decay curve in the tissue is often observed to deviate from the mono-exponential decay. Intravoxel incoherent motion (IVIM) is a diffusion model that aims to separate the perfusion effect from the diffusion effect. Some studies have applied IVIM in prostate imaging and showed promising results for the evaluation of prostate cancer5,6.

However, the stability of DWI-derived quantitative parameters is a prerequisite for their use as imaging biomarkers in clinical studies. In addition, the variety of imaging parameters may lead to conflicting results in quantitative analysis. The two-compartment IVIM theory assumes that the transverse relaxation time is the same in both compartments7. However, blood and tissue have different T2 values, and echo time (TE) may thus have an effect on the quantitative parameters of IVIM. Lemke et al.8 showed that in normal pancreas IVIM imaging, the f values was significantly affected by the TE value, while D and D* were not significantly affected. Jerome et al.9 reported that DWI scanning using the shortest TE value overestimated the f value of liver tissue. To our knowledge, no report has examined the effect of TE value on prostate IVIM quantitative parameters, although the 2012 PI-RADS v1 prostate MR guideline suggests that the TE should be as short as achievable (typically <90 ms) in prostate DWI10. However, IVIM was not included in the PI-RADS v1 and v2, and this recommendation did not sufficiently assess the TE effect on the quantification parameters. Relevant data and experience are also needed1,10. Therefore, in this study, we fixed all imaging parameters except for TE to investigate the effect of TE values on the quantitative parameters of IVIM in the peripheral zone (PZ) and central zone (CZ) of the normal prostate and to explore the reproducibility of the quantitative parameters of IVIM.

Results

The test-retest reproducibility of ADC, D, D*, and f in the PZ and CZ of the normal prostate

The coefficient of variation (CV) values for the ADC, D, D* and f are shown in Table 1 and Supplementary Fig. S1. The Bland-Altman analysis of the ADC, D, D*, and f values are shown in Table 2 and Supplementary Fig. S2. With TE values in the range of 60–120 ms, the CV of the ADC and D values in the PZ were in the range of 1.641–2.635%, and the ranges of the Bland-Altman limits of agreement (BA-LA) were −4.965% to 4.263% and −6.889% to 7.453%, respectively. The CV range of the ADC and D with different TE values in the CZ was 2.632–6.661%, and the BA-LA values ranged from −8.052% to 6.270% and from −24.413% to 21.284%, which indicated that ADC and D exhibited good repeatability; f had moderate reproducibility with a CV of 12.337–25.108% and BA-LA values of −43.961% to 30.994% and −52.185% to 62.516%; D* had moderate to poor reproducibility with a CV between 11.640% and 33.644% and BA-LA of −29.917% to 35.054% and −96.248% to 81.146%.

Table 1 Coefficient of variation (CV, in %) for the ADC, D, D* and f in the peripheral zone and central zone.
Table 2 Bland-Altman limits of agreement(BA-LA, in %) for the ADC, D, D* and f in the peripheral zone and central zone.

Correlations between ADC, D, D*, and f and the TE value in the PZ

Table 3 summarizes the ADC and IVIM-derived parameters in the PZ of the normal prostate for different TE values. The Spearman’s rho analysis of the TE values and the peripheral and central ADC, D, D*, and f values is shown in Table 4. Analysis of variance (ANOVA) showed a significant difference in the ADC, D and D* values obtained during the same examination (p < 0.05). Spearman’s rho analysis showed that ADC, D and D* were correlated with TE (p < 0.05). ANOVA showed no significant difference in f with different TEs (p > 0.05). Spearman’s rho analysis showed no significant correlation between f and TE (p > 0.05). The dependences of the ADC, D, D*, and f values in the PZ on different TE values are shown in Fig. 1.

Table 3 ADC, D, D* and f in the peripheral zone with different TE values, and ANOVA results.
Table 4 The Spearman’s rho analysis between ADC, D, D*, and f and the TE value.
Figure 1
figure 1

The scatter plot distribution of ADC, D, D*, and f according to the TE value. In the peripheral zone, ADC, D, and D* increased as TE increased, while f was stable. In the central zone, the ADC increased, but not obviously, with increases in TE, while D was stable as TE increased. The D* and f values increased as TE increased. Jittering (width = 1 ms, height = 0 ms) was applied to reduce overplotting.

Correlations between ADC, D, D*, and f and the TE values in the CZ

The ADC and IVIM-derived parameters for the CZ of the normal prostate for different TE values are summarized in Table 5. The Spearman’s rho analysis between TE value and the peripheral and central ADC, D, D*, and f values is shown in Table 4. ANOVA showed a significant difference for D* and f values obtained within the same examination (p < 0.05). Spearman’s rho analysis showed that D* and f had significantly positive correlations with TE (p < 0.05). ANOVA showed no significant differences (p > 0.05) in the ADC or D values for different TEs in the CZ of the prostate. Spearman’s rho analysis showed that ADC and TE exhibited a positive correlation (p < 0.05), while D and TE had no significant correlation (p > 0.05). The dependence of ADC, D, D*, and f values in the CZ on different TE values is shown in Fig. 1.

Table 5 Results of ADC, D, D* and f in the central zone with different TE value and ANOVA results.

Image signal-to-noise ratio (SNR)

Table 6 shows the SNR when TE was changed from 60 to 120 ms at b = 1000 s/mm2. The SNR was 13.509–20.374 using two measurements of different TEs. The SNR decreased with increases in TE values.

Table 6 Signal-to-noise ratio at b = 1000 s/mm2.

Discussion

Our study showed that ADC and D measurements had good reproducibility in both the PZ and CZ of the prostate with variable TE values from 60–120 ms. Thus, the ADC and D values can be used as reliable parameters, especially for clinical longitudinal studies. Relative to the ADC and D values, the reproducibility of f and D* was poor, suggesting that the stability of ADC and D measurements is higher than that of D* and f measurements. This result is consistent with other prostate IVIM findings11,12. Jambor et al.13 showed that the bi-exponential model parameters are sensitive to noise and are less reproducible than the mono-exponential model. Kakite et al.14 studied the short-term reproducibility of hepatocellular carcinoma IVIM-DWI, and they also concluded that the perfusion parameters D* and f were less reproducible and that ADC and D were highly reproducible. Clinically, IVIM-DWI, as a biomarker of imaging, is required to have good reproducibility, especially for follow-up of prostate cancer patients, to accurately reflect the dynamic changes in the lesion, thus helping to develop the correct treatment plan.

Our study demonstrated the significant influence of TE value on the ADC and IVIM parameters in different locations in the normal prostate. Our results showed that the f value of the CZ increased with the extension of TE value, and that the f value of the PZ was not significantly correlated with the TE value, which was mainly due to the T2 relaxation time difference between the PZ and CZ. Lemke et al.8 showed that the f value of normal pancreatic tissue increased with an increase in the TE value, and they suggested that TE values had a greater effect on tissue with a short T2 relaxation time. According to the theory of the IVIM double compartment model, the signal attenuation of tissue is attributed to pure molecular diffusion and microcirculation or blood perfusion in the capillaries15. Bojorquez et al. have reported a calculated T2 value for the prostate of 80 ± 34 ms16, and Gibbs et al. found a normal PZ T2 value of 136 ± 40 ms17. Stanisz et al. have reported the blood T2 relaxation time of 275 ± 50 ms18. The T2 relaxation time of blood is much longer than that of central gland tissue. A longer TE results in more signal attenuation of tissues with short T2 relaxation time, which increases the signal fraction of the capillary component8 and results in an increase in the value of f. Different from CZ, the change in the TE value did not significantly increase the perfusion signal in the PZ. This may have been because the T2 relaxation time of PZ tissue is not much different from the T2 relaxation time of blood, making the bias less pronounced in the PZ. In our study, the D value was calculated using a b value greater than the 200 s/mm2 DWI signal, at which the effect of the capillary perfusion component is assumed to be negligible19,20. Theoretically, the D value should be stable as the TE changes. Our results showed that no statistically significant TE dependence of D was observed in the CZ (Spearman’s rho 0.179 and 0.172; p > 0.05). This is in agreement with the result of Lemke et al.8, who explored the effect of the TE value on the value of f and D in the healthy pancreas. The relaxation times of both the pancreas and TZ are much shorter than that of blood, which may explain the similar variations of D with TE. They also applied a corrected IVIM model to correct for different relaxation times of the two compartments (and hence minimize the effect of TE and TR), and with the corrected model, the bias in f and D values vanished. Therefore, they suggested that TE has a definite effect on f in the standard IVIM model but that this can be accounted for with a corrected model. While the D values in the PZ of our study had a significant correlation with the TE, one possible reason for this bias may be that the threshold to fit D (200 s/mm2 in this study) is not appropriate and relatively low. To validate this possibility, we increased the threshold up to 500 s/mm2. However, when we set the threshold to 500 s/mm2, the bias was also found in the parameter D (for different TEs) in the PZ. Therefore, the variations of D with TE were not due to an inappropriate b-threshold. We speculate that the special structure of the PZ causes this phenomenon. The peripheral tissue has a relatively loose microstructural organization that contains many glands with large lumina and fluid in the glandular lumina21. The fluid and glandular tissue have different T2 relaxation times; therefore, increasing the TE value would cause different signal decay. This may explain in part why the D value changes with the TE in the PZ. However, further studies are needed to confirm this phenomenon.

The ADC in this study was a mono-exponential signal fitted with all b values of 0–1000 s/mm2. The ADC represents the common contribution of both diffusion and perfusion components. Thus, the variation in both diffusion and perfusion would lead the variation in ADC values with the change in TE values. In our study, the ADC values of both PZ and CZ showed significant correlations with TE. Previous work also explored the effect of TE value on the measurement of ADC, obtaining similar result22. Of all the calculated parameters, D* showed the worst reproducibility. Accurate measurement of D* is still challengeable. Jerome et al.9 studied the effect of the TE value on the liver IVIM parameter values, taking into account the uncertainty of D*, which did not explain the changes in D*. The accurate measurement of the value of D* and evaluation of the effect of TE on D* require further studies.

It should be noted that image quality is important for accurate estimations of diffusion parameters. To evaluate the image quality, we computed the SNR of DW images with a b value of 1000 s/mm2. In our study, the SNRs ranged from 13.509–20.374 using two measurements of different TEs (60–120 ms). Although the longer TE for DWI caused impairment of the SNR, the SNR in our study still reached a reasonably high level and was close to that of a previous study23. We used a 3.0 T MRI scanner, with 4 averages and a 5-mm slice thickness, which played positive roles in increasing SNR.

This study had some limitations. First, only a limited number of cases were included in the present study, and large studies will be needed in the future to confirm these results. Second, the volunteers had a wide age range (20–70 years old), which increased the variance caused by age. Third, this study only included volunteers with a healthy prostate. The next step should evaluate prostate pathologies such as prostate cancer. Finally, we used a relatively short TR of DWI in this study in order to save examination time, which had an effect on pixel intensity. However, the acquisition time needed to obtain accurate measurements was still longer, and some subjects were intolerant. In a future study, we plan to examine a large sample of prostate cancer patients by optimizing the existing program and reducing the scan time.

In summary, the ADC and D values of a normal prostate exhibit good reproducibility, whereas D* and f have moderate or poor repeatability. The ADC and quantitative parameter values of IVIM are clearly affected by the TE value. In clinical applications of IVIM, especially for dynamic assessment of prostate disease and follow-up after treatment, changes in D* and f values may be due to poor reproducibility rather than the disease itself. Considering the effects of the TE value on the parameter values, TE values should be consistent before and after dynamically assessing prostate disease and during follow-up after treatment. The minimum TE is typically used in clinical studies; however, studies have showed that even using the smallest TE may overestimate the IVIM parameters of the tissue9. It is necessary to pay attention to the effects of TE value on parameter values. A corrected IVIM model may reduce or eliminate the effect of the TE value on quantitative parameters and may improve the clinical application value of IVIM.

Materials and Methods

Subjects

This prospective study was approved by the Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology institutional review board, and informed content was obtained from each participant. All methods were performed in accordance with the relevant guidelines and regulations. From September 2016 to December 2016, 17 healthy volunteers (mean age, 42 years; age range, 20–67 years) were eventually enrolled in this study. The inclusion criteria were as follows: (a) no obvious symptoms and signs of the urinary system and (b) no MRI contraindications (such as metal implants, pacemakers, claustrophobia, etc.). The exclusion criteria consisted of (a) failure to complete the examination for any reason; (b) motion artefacts that prevented image analysis; and (c) MRI results showed prostate disease (such as prostatitis and obvious benign prostatic hyperplasia).

MR Imaging Technique

All volunteers were examined with a 3.0 T system (Magnetom Skyra, Siemens, Erlangen, Germany) using an 18-channel phased-array coil above and a spine coil underneath the pelvis. One hour before the examination, the volunteers were asked to defecate to reduce the magnetic susceptibility artefacts produced by gas in the rectum. All examinations included triplane half-Fourier acquisition single-shot turbo spin-echo (HASTE), axial T2-W images, and axial DW MR images. The DWI was acquired using a monopolar single-shot echo planar imaging sequence with 9 b values (0, 10, 20, 50, 100, 200, 500, 800, and 1000 s/mm2). The HASTE image was acquired with a repetition time (TR) of 1510 ms, echo time (TE) of 87 ms, field of view (FOV) of 300 × 300, matrix of 320 × 256, and slice thickness/gap of 3/0 mm. The axial T2W images were acquired using the following parameters: TR = 3700 ms, TE = 104 ms, FOV = 180 × 180 ms, matrix = 384 × 346, slice thickness/gap = 5/0 mm. The DWI data were scanned 6 times with variable TE values of 60, 70, 80, 90, 100, and 120 ms. The other parameters were consistent for all scans (TR = 2000 ms, FOV = 225 × 180 mm, matrix = 90 × 90, slice thickness/gap = 5/0 mm, number of slices = 12 and the number of averages = 4 for each b value and the geometrical averaged data were inline calculated, acquisition time = 3.26 min). The pulse gradient duration δ and the separation between leading edges of the pulse gradient Δ were (13.9, 29.1), (18.9, 34.1), (23.9, 39.1), (28.9, 44.1), (33.9, 49.1) and (43.9, 59.1) ms with variable TE values of 60, 70, 80, 90, 100, and 120 ms. A duplicate examination was performed, and the volunteers were asked to rest for 15–20 minutes and return to the scanner before the second examination. Thus, each volunteer underwent a 12 DWI image series during this study.

Image Analysis

A segmented algorithm was applied using in-house-developed software (MATLAB R2012b, MathWorks Software) to fit the IVIM model24. It assumed the perfusion component could be neglected for diffusion data at a b-value threshold of 200 s/mm2, and then the diffusion coefficient (D) and S0′ were calculated using the mono-exponential equation Sb = S0′ × exp (−b × D). Sb is the signal intensity with the gradient, and S0′ is the b = 0 s/mm2 intercept of the mono-exponential fit of high b value data (≥200 s/mm2). Then, the perfusion fraction (f) was calculated by f = (1 − S0′)/S0. At last, the calculated D and f were applied in the IVIM equation fit of all b values (0–200 s/mm2) to measure D*. The ADC was calculated from all the b values (0–1000 s/mm2) using the mono-exponential equation Sb = S0 × exp (−b × ADC). Plots of representative subject IVIM data and corresponding IVIM fits are shown in Supplementary Fig. S3.

The diffusion data were processed with a prototype tool called Body Diffusion Toolbox. The ADC map of the mono-exponential model and the D, D*, and f maps of the IVIM model were calculated. One radiologist (ZYF, with 5 years of clinical experience in the interpretation of prostate MR images) performed the quantitative image analysis of two examinations for each volunteer. For each of the diffusion examinations, regions of interest (ROIs) were drawn on the b = 0 s/mm2 images with TE = 60 ms. On the three largest axial images at the mid-gland level, ROIs were drawn within the outer border of the PZ and CZ, respectively. Then, the ROI was copied to the other series (TE = 70, 80, 90, 100, and 120 ms) obtained during the same examination. The mean ADC, D, D*, and, f values were recorded for each ROI. Our data analysis was based on ROI levels, and co-registration was not performed.

We computed the SNR to evaluate the image quality. To calculate the SNR, elliptical ROIs were delineated in the bilateral internal obturator muscle on the TE = 60 ms, b = 0 s/mm2 image, and the ROIs were automatically copied to other non-zero-value images of the same DWI sequence. The ROIs were then copied to the other DWI series (TE = 70, 80, 90, 100, and 120 ms) obtained during the same examination. The signal intensities of the PZ, CZ, and bilateral internal obturator muscle were recorded. The SNR was calculated by the formula: SNR = SIROI/SDmuscle, where SIROI is the signal intensity of the prostate of on the DW image with a b value of 1000 s/mm2 and SDmuscle is the standard deviation of the signal intensity of the bilateral internal obturator muscle. Although this is not a perfect method given use of parallel imaging, other researchers have used this same method to measure SNR25. For the ROIs drawing method, see Fig. 2.

Figure 2
figure 2

ROI drawing method. (a) Axial T2WI, (b) Axial DWI with TE = 60 ms and a b value of 0 s/mm2. The ROI was drawn within the outer border of the central zone (yellow solid line) and peripheral zone (red solid line). Then, the ROI was automatically copied to each parameter map. The oval ROI was drawn on the bilateral obturator muscle and automatically copied to an image with a b value of 1000 mm2/s (c) to measure the signal-to-noise ratio. (dg) Parameter maps of D, D*, f and ADC, respectively, calculated by DWI with TE = 60 ms.

Statistical Analysis

The reproducibility of ADC, D, D*, and f values calculated using DWI data was evaluated by the coefficient of variation (CV) and 95% Bland-Altman limits of agreements (BA-LA). CV values ≤10% indicated good reproducibility, 10–25% indicated moderate reproducibility, and values ≥25% indicated poor reproducibility.

The correlations between ADC, D, D*, and f and the TE value were analysed using Spearman’s rho analysis. The correlation was low (if any correlation) when 0 ≤ r < 0.25; the correlation was weak when 0.25 ≤ r < 0.5; the correlation was moderate when 0.5 ≤ r < 0.75; and the correlation was strong when r ≥ 0.75. The ADC and IVIM-derived quantitative parameters for different TE values were compared using one-way analysis of variance (ANOVA).

The statistical software included SPSS software (SPSS for Windows 19.0, Chicago, IL, USA) and MedCalc 13.0.0.0 (MedCalc Software, Mariakerke, Belgium). MedCalc was used to calculate the CV and 95% BA-LA. A value of p < 0.05 indicated a statistically significant difference.

Ethical approval and informed consent

This prospective study was approved by Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology Institutional Review Board, and informed content was obtained from each participant.

Data availability statement

The datasets generated and/or analysed during the current study are available from the corresponding author on reasonable request.