Introduction

Increasing demands for high living qualities raise the requirements for improved healthcare in modern society. As a supplementary approach to disease treatment, wearable and implanted biomedical devices arouse intense interest in clinical medicine1,2,3. To develop versatile and comfortable electronic biomedical devices, tremendous research efforts have been devoted to exploiting various kinds of functional materials, such as metal foils4, carbon-based nanomaterials5,6,7, metal-based nanomaterials8, hybrid nanocomposite9,10, conductive polymers11,12, Si-based semiconductors13, etc. 14 These materials have been widely used in bone repairment, material package, sensor fabrication and other fields15,16,17. Additionally, more innovative methods were introduced to modify the materials and devices18. However, traditional materials have several limitations in stretchable biomedical devices. Most metal foils and their alloys exhibit high elastic modulus19, making them inflexible and cannot fit the movement of the human body. Most of the conductive polymers and carbon-based nanotubes have the potential toxicity, poor ductility and incompatible mechanical strength with humans20,21,22,23. Synthesis of metal-based nanoparticles is costly, and it is also difficult to control the size and shape of the particles. Moreover, metal-based nanoparticles may also have biotoxicity, and the high degradation resistance of metal-based nanoparticles is also detrimental to the environment23,24.

To satisfy the requirements of good stretchability and good electrical conductivity simultaneously, liquid metals (LMs) are regarded as the candidates for the design of high-performance electrical devices. The LMs with low melting points generally consist of only a certain range of elements. For instance, Hg, which serves as a part of the traditional thermometer, is the most common LM in daily life. However, the toxicity of Hg results in poor biocompatibility25. Besides, alkali metal elements like Rb have very low chemical stability26. Recently, Ga- and Bi-based LMs have attracted extensive research interest due to their excellent mechanical and chemical properties.

Unlike Hg and Rb, the Ga- and Bi- alloys show near zero vapor pressures at room temperature27,28, so there is no worry about Ga or Bi evaporating into humans, and destroying the nervous system as does Hg vapor. Moreover, the Ga and Bi-based LM devices exhibit huge potential in biomedical applications, owing to their marvelous biocompatibility29,30,31,32,33,34, stretchability35,36,37, high thermal38,39,40,41,42,43,44,45 and electrical36,46 conductivity. Thus, they have been intensely investigated recently. For instance, scientists have fabricated LM electrodes used in many basic electronic devices47, such as capacitors48,49, inductors49,50,51, memristors52,53,54, and optoelectronic devices55,56,57, achieving good performances in electrical biomedical facilities. Moreover, LM-based devices, such as bone cement, e-vessels, and exoskeletons, can excellently conform with human movement, due to their superior stretchability performance. Besides, LM-based materials can also be used in non-electrical equipment, like drug carriers58. Recently, LM-based materials have been widely applied in biomedical fields, such as digestive tract contrast agents59, nerve repairment60,61,62, and sensors63,64,65,66,67,68. (Fig. 1). Although several reviews focus on the properties69, the biomaterial application70, manufacture and modification methods71,72 of LMs were proposed these days, there is a lack of comprehensive review on the stretchable LM-based biomedical devices with an emphasis on human therapy and function augmenting.

Fig. 1: The biomedical applications and challenges of LMs.
figure 1

The biomedical applications include therapy and human function augmenting, but the high cost and corrosivity limit the wide spread of LMs.

In this review, we focus on the applications of LM-based materials in stretchable biomedical devices. The basic properties of LMs will be first introduced. Then, we will discuss the recent development of LM-based materials applied in biomedical, including human therapy and human function augmenting. Finally, the existing challenges and future prospects of LM-based biomedical applications will be outlined.

Overview Of Lm

Before the elaboration on the application of LMs in biomedical devices, a brief introduction of the LMs and their material systems assists the understanding of the limitations in device manufacturing and the existing challenges. Herein, we emphasize the architecture, manufacturing and modification of LMs as follows.

Architecture

Various architectures of LMs have different application scenarios. For example, the LM-based microfluid can be utilized in nerve repairment, while LM-composites can be used to manufacture bone-cements and exoskeletons73,74. Herein, we focus on the LM-based composites, microfluids and printed particles to introduce the structure of the LMs.

LM composites, in collaboration with other materials, present promising solutions with vast potential for both fundamental research and practical applications. As illustrated in Fig. 2a–c, the basic composite strategies are categorized into three groups: LM composites with a core-shell structure75, LM-polymer composites76, and LM-particle composites77,78. The functions of these composite strategies are exemplified through typical applications in several key fields. The controlled integration of base LM with functional materials, such as metal nanoparticles, polymers, and drug molecules, allows for precise tuning of LM’s intrinsic properties. This tunability opens up significant opportunities to address challenging issues across various sectors, including thermal management, biomedicine, chemical catalysis, flexible electronics, and soft robotics79.

Fig. 2: The architecture of LMs.
figure 2

a The sketch of LM composites with a core-shell structure75. Copyright 2017, Science. b The sketch of LM-polymer composites76. Copyright 2011, John Wiley and Sons. c The schematic graphs of LM particles78. Copyright 2019, John Wiley and Sons. d The LM- microfluid83. Copyright 2014, National Academies Press. e The process of printing LMs95. Copyright 2013, Springer Nature.

Microfluids, which involve the integration of functional fluids into a chip-like system with a compact size and low energy consumption for portable applications, hold great promise in areas such as drug/cell screening, sensing, health monitoring, and microfabrication. The substitution of traditional fluids with LMs introduces additional capabilities to the system, shifting its applications beyond bio-related fields to electronics and related domains (Fig. 2d)70,80,81,82,83. LMs offer distinct advantages in LM-based microfluidics. Firstly, LM can maintain electrical continuity even under significant deformations. Secondly, LM readily forms alloys with various metals, facilitating direct electrical connections between LM-based electronics and substrate materials, including other electronic components. Thirdly, LM wets and fills channels despite its high surface tension, owing to the rapid oxidation of the gallium-based LM surface, forming a self-limiting gallium oxide skin that is a few nanometers thick. This skin possesses low surface tension while exhibiting high affinity and adhesion to various substrates. The combination of conductivity and deformability in LM opens up the possibilities in electronics, enabling soft, flexible, stretchable, reconfigurable, and even self-healable features84,85,86. To date, LM-based microfluidics have demonstrated excellent performance in soft, flexible, and wearable electronics87, soft robotics88, biomedical sensors89, and transient circuits90. Additionally, the generation and manipulation of microdroplets play a crucial role in microfluidics research. The fabrication of LM droplets and their dispersion in elastomers has advanced in recent years. Microdroplet composites, derived from these advancements, find significant applications in the fields of electricity, heat, mechanics, and supercooling. Therefore, beyond filling the LM continuous phase in microfluid chips, the study of LM microdroplets/composites is closely interconnected with the field of LM microfluidics.

With the advent of the information age, electronic products are playing an increasingly crucial role. In recent decades, various electronic manufacturing methods have been extensively explored91,92,93. LM printed electronic (LMPE) (Fig. 2e) is one of the most cutting-edge materials for solving the issue of low conductivity of the traditional inorganic nanomaterials94,95. From both theoretical and technical perspectives, LM is a highly desirable choice for preparing electronic inks due to its high charge mobility (up to 106 S m−1) and adjustable physical and chemical properties90,96, such as low melting point, high thermal conductivity, and electrical conductivity. The fundamental principles and practical technologies for molding LM electronic inks and printers have been established. A group of LMPE equipment, inks, and application products like electronic skins has even transitioned into the industry, witnessing rapid industrialization in the field. LMPE has garnered widespread attention from academia to industry, indicating a bright future in the biomedical field97.

Manufacture and modification

LMs demonstrate their outstanding properties in various fields, while the processes of manufacture and modification affect the performances of LM-based facilities, such as chemical and electric stabilities, thermal sensitivity, anti-bacterial ability, viscosity, biocompatibility, and magnetism72,98,99.

Magnetic LMs (MLM) are introduced to have a better simulation of the dynamic stiffness change. For example, Li et al.74 proposed a methodology to synthesize the MLM which could be applied in bone repair. The magnetic silicon particles (Fe@SiO2) were added into the prepared Galinstan (special weight percent: Ga 68.5%, In 21.5%, and Sn 10%), and the mixture will be stirred to guarantee homogeneous distribution of the Fe@SiO2. The as-fabricated MLM has a good ability to respond to the changes in the external magnetic field by simulating the regeneration of bones.

Moreover, several fabrication methods of non-wettable and nonsticky LM marbles (LMMs) were designed to address the stickiness problem of oxidized LM alloys exposed to the air. In 2018, Chen et al.100 proposed a nonsticky LM droplet with coated graphene shells. As illustrated in Fig. 3a, the LM droplets were dropped and rolled on the graphene sheets. Due to the adhesion between the oxidized LM layer and the graphene sheet, the LM droplet was packed by a nest-like graphene cover (Fig. 3b).

Fig. 3: The fabrication and modification of LMs.
figure 3

a The coating procedure of graphene sheets100. Copyright 2018, John Wiley and Sons. b A photograph of the graphene-coated LM droplets100. Copyright 2018, John Wiley and Sons. c The preparation and application of the galvanic substitution method modified EGaIn-Au NPs101. Copyright 2021, Elsevier. d The synthesis process of LM@pZrO2 NPs and the flexibility under scheme103. Copyright 2009, Royal Society of Chemistry. e Preparation of LM@MSN105. Copyright 2019, Elsevier. f Fabrication of LMs based on sono-chemical assembly modification106. Copyright 2009, Royal Society of Chemistry.

The galvanic substitution surface modification occurs, owing to the difference of the electrochemical potential gradient between the two metals/metal ion pairs, which not only assists in enhancing the photothermal conversion efficiency of LM nanoparticles (NPs)101 but also improves the catalytic activity in certain areas102. For example, Guo et al.101 proposed an in-situ interfacial galvanic substitution reaction to prepare the heterogeneous eutectic gallium indium-Au (EGaIn-Au) nanoparticles. As shown in Fig. 3c, the HAuCl4 solution was added into the EGaIn-Au nanoparticles suspension, triggering the in-situ growth of the Au-coated shell. The photothermal performance test demonstrated its excellent behavior in thermotherapy on cancer treatment.

Post-synthetic strategies were also adopted for core-shell structural modification of LMs. Figure 3d illustrates a synthesis method for ZrO2-covered EGaIn nanoparticles, which exhibits enhanced efficiency in internalizing into cells103. Zheng. et al.104 elucidated a route for the synthesis of a conductive LM-based core-shell particle that could be used to fabricate self-repairing soft circuits. The silver encapsulates the LM particles via screen printing, and the whole material shows remarkable resistance stability (ΔR/R0 < 1.65% after 10,000 bending cycles). Furthermore, the silica coating strategy also enhances the thermal stability and chemical stability of LM NPs105 (Fig. 3e).

Polymers, which have strong intra- and inter-molecular interactions, are utilized for encapsulating LMs which raises scientists’ interest these days. Figures 3f and 4a show the common methodologies for polymer encapsulation, including sonochemical assembly106, chemical-initiated crosslinking polymerization stragedy107, and atom transfer radical polymerization (ATRP)108. For example, Li et al.109 utilized aqueous ultrasonication to prepare the marine polysaccharide-wrapped LM, and the alginate assisted in the downsizing and formation of microgel shells for EGaIn droplets (Fig. 4b). Thrasher and his team proposed a photo-initiated polymersation for encapsulation110. The poly-LM networks exhibited an outstanding performance in resistance stability over large strains. Moreover, the reversible addition-fragmentation chain-transfer (RAFT) polymersation111 and the ring-opening polymersation112 also serve as the LM modification strategy.

Fig. 4: The fabrication and modification of LMs.
figure 4

a The structure and modification process of ATRP108. Copyright 2019, Springer Nature. b The sonicating and encapsulating procedure of EGaIn109. Copyright 2018, John Wiley and Sons. c The synthesis process of enzymatic LM nanobots114. Copyright 2021, American Chemical Society. d The manufacture of LMNP@CM115. Copyright 2020, John Wiley and Sons.

To enhance the behavior in biocompatibility, several bioengineering modification methods were proposed ingeniously113. For instance, Xu et al.114 designed enzyme-coated LM-based nanobots utilized in bio-imaging (Fig. 4c). Firstly, the LMs were sonicated in the dopamine solution to synthesize the polydopamine-based LM NPs. Then urease and cefixime trihydrate were loaded and the enzyme was finally coated at the surface of LM NPs with the assistance of a glutaraldehyde linker through electrostatic adsorption. Similarly, Fig. 4d illustrates a cell membrane-coating strategy for surface engineering115, which improved the anti-tumor ability of the LM NPs.

Generally, various surface engineering methods have been exploited to construct protective layers for LM improving the performances of LM-based nanomaterials. This modification enhances the chemical stability and biocompatibility of LM-based materials, enabling their wider applications in the fields of biology, energy, the aviation industry, and other scenarios. However, several LM-based materials are difficult to apply on a large scale due to the complexity and high cost of the preparation methods. Therefore, there is an urgent need for scientists to further invent more low-cost production processes for LM materials.

Properties Of Lm

To understand why LMs are qualified for biomedical devices, being acquainted with their basic properties is of great significance. The biocompatibility of the materials ranks first when being used for the manufacturing of biomedical devices. Moreover, good chemical stability is critical to avoid generating potentially toxic substances. Table 1 compares the different properties between LMs and other biomedical materials, indicating the potential advantages of LMs in the field of biomedical devices. Due to high fluidity35,116,117, low viscosity118,119,120, low melting point121, and special crystallinity122, LMs and LM composites show good stretchability, which endows biomedical devices with good fit for human movements. Along with the properties mentioned above, the excellent electrical conductivity of LM ensures its applications in electrical equipment68,123. Herein, we elucidate the characteristics of biocompatibility, chemical stability, stretchability, and conductivity of LMs before overviewing their biomedical application.

Table 1 Chemical, mechanical, and electrical properties of different biomedical materials

Biocompatibility

Ga-based and Bi-based alloys are two kinds of common LMs in the healthcare field. In the past few years, bio-safeties of LM-based materials have been widely tested through both in vitro and in vivo experiments. Various categories of LMs and LM alloys have different toxicity. Sano et al.124 carried out a study of the direct toxicity of Bi. They selected enough male and female Crj:CD (SD) IGS rats (SPF) and then divided them into two experiments. One group of rats was studied for acute oral toxicity with a dose of 0 and 2000 mg kg−1, and the other group was used to unveil the potential toxic effect of Bi, where the rats were administrated orally at a dose of 0, 40, 200, and 1000 mg kg−1 daily for 4 weeks. The relevant parameters were detected. As illustrated in Fig. 5a–d, there was no significant difference among each group. Even high Bi dose (1000 mg kg−1) orally administrated rats shared similar necropsy and histopathological findings with those at the dose of 0 mg kg−1 (Table 2). The results showed that neither the acute nor long-term oral administration led the adverse reactions in rats. The lethal dose 50% mortality rate dosage of Bi turned out to be higher than 2000 mg kg−1 for the rats, so the toxicity of bismuth is much lower than Hg. However, the inorganic Bi compounds have the potential to be adsorbed and diffused in human organisms125, and administrating the Bi compounds will cause acute renal failure126. Therefore, a longer in vivo toxicity investigation of bismuth is of great significance to further determine the relationship between the safe concentration threshold of bismuth.

Fig. 5: Biotoxicity test of Bi with different doses124.
figure 5

a Minerals and Electrolytes of rats under different Bi Dose concentrations. b Liver function of rats under different Bi Dose concentrations. c Renal function of rats under different Bi Dose concentrations. d Protein markers of rats under different Bi Dose concentrations. ad The error bars show the standard deviation.

Table 2 Histological findings within the 28-d perennial oral dose toxicity study124

Ga-based materials are another mainstream LM. Recently, many researchers have focused on the cytotoxicity of Ga. For example, Liu et al.59 found that Ga was a potential digestive tract radiography agent and conducted the safety test simultaneously. They separated 10 rats into two groups randomly, where the rats in the experimental group were given 0.2 mL Ga while the control group received deionized water by gavage. Then the rats were utilized for histological and basic indexes observations. The data of the two groups can be regarded as the same, verifying the biocompatibility of Ga. Moreover, not only do Ga and Bi have low cytotoxicity, but piles of experiments also demonstrated the safety of Ga-based and Bi-based alloys127,128,129,130. Nonetheless, the Ga-based LMs will get oxidation into gallium oxide hydroxide (GaOOH) in an oxygenated high-temperature atmosphere, and the toxicity of GaOOH limits the practical application of Ga-based biological devices131.

Additionally, to investigate the cytotoxicity of the ions released from eutectic gallium-indium alloy (EGaIn) on human cells. Kim et al.132 conducted a toxicity test of EGaIn in an aqueous environment. The results revealed that although mechanical agitation would augment the cytotoxicity, EGaIn could be generally considered safe in a normal aqueous environment. Furthermore, although organometallic compounds are summarized as more toxic generally133, many experiments indicate their good biocompatibility116,134,135,136,137. However, it should be noted that Sn and many Sn salts have low cytotoxicity, while many organotin compounds harm the nervous system133.

With the development bio-application of LM NPs138, the toxicity aroused scientists’ attention. For example, Narayanasamy et al.139 reported that the prolonged-acting, muti-targeting Ga-based NPs helped inhibit human immunodeficiency virus (HIV) infection. And there was no observation about macrophage cytotoxicity caused by the Ga-NPs in 24 h. Moreover, Lin et al.140 prepared poly (1-octadecene-alt-maleic anhydride) (POMA) grafted LM NPs and investigated their toxicity. The Chinese hamster ovary (CHO) cells were exposed to LM NPs. The results showed the low cytotoxicity of LM NPs under the concentration of 62.5 μg mL−1.

Furthermore, the environmental impact of LMs cannot be ignored. Yu et al.141 reported the environmental pollution effects of Ga. As illustrated in Fig. 6a, due to low solubility in water, Ga has a low concentration in the daily aquatic environment. However, the Ga levels in inhalable gas samples are quite higher in manufacturing areas (Fig. 6b). Similarly, the inhalable gas levels of Bi in the workplace are also high, which has caused a few bismuth poisoning events142. Besides, Ga may be oxidized to GaOOH in an aqueous environment131, which decreases the biocompatibility of the LM-based devices. Despite the direct pollution caused by the bulk LMs, the environmental impact of Ga and Bi-based LMs is much lower than lead and Hg142. Moreover, the scientists also devoted themselves to investigating the modification methods that decrease the pollution of LM-based materials. For example, Mao et al.143 proposed a nanocellulose-based LM circuit, which is soluble in water and the LM ink could be recycled conveniently.

Fig. 6: Average concentrations of Ga in different area141.
figure 6

a The average concentration of Ga in different water resources. The error bars show the standard deviation. b The average concentration of Ga in control and manufacturing areas. The error bars show the standard deviation.

In summary, with excellent biocompatibility, many Ga-based and Bi-based LMs are relatively safe to be utilized for biomedical devices. Nevertheless, further investigations are needed to determine the toxicity and pollution of LM-based materials.

Chemical stability

Owing to the high reactivity of Ga, LMs containing Ga are easily oxidized in humid air, which can improve the stability of the Ga-based alloy by forming an inactive oxidation layer. Scharmann et al.144 found that regardless of low or high relative humidity for a short or long period, the oxidized Ga (Ga2O3 mostly, which is stable) mainly scattered at the surface, and the Ga atoms dispersed in the deeper inside, while Sn and In were predominantly stable. However, in a humid environment, the liquid Ga tends to be over-oxidized and transform to a crystalline structure (GaOOH, which can serve as a catalyst) when the Ga-based LMs are heated to 343 K145,146. Inversely, the Ga oxide layers could also be reduced under negative potential or in an acid or base environment147. Moreover, Ga would react with other metals (like Cu) in the LM-based devices due to its high reactivity. For example, HerØ et al.148 studied the long-term corrosion of Ga alloys in vivo. They tested GF (65% Ga, 19% In, 16% Sn), Galloy (62% Ga, 25% In, 13% Sn, 0.05% Bi) and EGaIn (75.5% Ga, 24.5% In). The samples were exposed to the oral cavity for 2–9 months. The scanning electron microscopy results showed that the CuGa2 phase corroded the most substantially in the three alloys. Compared with other alloys, Galloy was less corrosive, indicating that modifying the microstructure of LM-based alloys helps improve the corrosion resistance.

On the other hand, scientists have developed various LM composites to enhance stability. For example, Li et al.109 introduced a method to fabricate a biocompatible aqueous ink, which was manufactured by introducing the EGaIn into marine polysaccharides microgels, reinforcing the stability of LM composites. Furthermore, Wu et al.149 proposed a Ga-Mg alloy which has the potential to act as a biodegradable material, where Ga was modified and utilized to decrease the degradation rate of Mg in body fluid. Therefore, the corrosivity of gallium could be reduced.

Moreover, increasing investigations focus on the rupture of LM-based NPs. Liu et al.150 prepared a series of polyurethane-based LM NP materials (Fig. 7a) and investigated the rupture stress by numerical calculations together with experiments. Researchers stretched the polyurethane-based LM NP materials cyclically, and Fig. 7b–e illustrated that the smaller NPs had higher activation strain and were less likely to rupture. When the oxide layer breaks, the LM exudation will cause the insulation surfaces to be electrically conductive, which may have potential harm to human bodies and limit the application of LM-based biomedical devices.

Fig. 7: The rupture of LM-based NPs.
figure 7

a The design of the polyurethane-based LM NP material150. b The morphology changes of different polyurethane-based LM NP materials150. c The stress-strain curves of different polyurethanes coupled with LM NPs activation stress, where the NPs break in the orange area and remain intact in the purple area150. d The experimental and theoretical activation strain of 50%LMNPs-0.3/85PU (50% vol% LMNPs-0.3 dispersed in 85PU matrix) and 50%LMNPs-1.4/85PU. The error bars show the standard deviation150. e The experimental and theoretical activation strain of 50%LMNPs-0.3/65PU and 50%LMNPs-1.4/65PU. The error bars show the standard deviation150. ae Copyright 2021, Springer Nature.

Overall, although the performance of LMs might degrade in high temperatures or the acid or base environment, various modification methods assist in enhancing their chemical stability naturally or artificially, both of which enable LM-based devices with long working life. Nonetheless, further investigations about how to optimize the performance of LM-based devices are required. Moreover, more research is needed to understand the effect of the effusion of LM when the gallium oxide shell is ruptured because of the device’s deformation.

Stretchability and electrical conductivity

The good stretchability of a biomedical device enables a long operating life and improved adaptability to the human body. It is nonsensical to only discuss flexibility while ignoring changes in electrical conductivity for materials. Therefore, researchers often analyze stretchability, conductivity, and resistance stability simultaneously.

The LMs with high flexibility are suitable for basic biomedical device application151,152,153,154,155. Tang et al.116 introduced a printable metal-polymer conductor (MPC). As illustrated in Fig. 8a, the LM particles were embedded in the polymer matrix. Not only did the MPC maintain approximately 30% of its initial conductivity (8 × 103 S cm−1) when subjected to a large strain of 500%, but also a few significant changes in its resistance (ΔR/R < 3%) were observed after 50% strain for 10,000 cycles (Fig. 8b). Similarly, Wang et al.117 presented another type of superelastic conductor with an original conductivity of 8331 S cm−1. These highly flexible conductors could be stretched up to 800% for 10,000 cycles, and the resistance increased 8 times compared with the initial one.

Fig. 8: The stretchability and electrical conductivity of LMs.
figure 8

a The schematic component of MPC116. Copyright 2018, Elsevier. b ΔR/R changes with a strain of 50% for 10,000 cycles of MPC. The error bars show the standard deviation116. Copyright 2018, Elsevier. c Fractional resistance changes of the LM-based elastic conductor with annular-shaped branch channels79. Copyright 2023, Elsevier. d The scheme of the LM-based elastomer with annular-shaped branch channel79. Copyright 2023, Elsevier. e The design of the annular-shaped branch channels with continuous performance79. Copyright 2023, Elsevier. f Formation and activation of polymerized LM networks110. Copyright 2019, John Wiley and Sons.

Apart from the low Young’s modulus of the materials, the structure of the circuit also helps improve the stretchability. For instance, Li and his team proposed annular-shaped branch channels to reinforce the stretchability of LM-embedded elastomeric conductors. As shown in Fig. 8c, the change of resistance stayed below 2% within 1000 cycles under 60% strain. Figure 8d, e show the construction scheme of the LM-based component. Additionally, another structural design strategy was introduced by Thrasher110. As demonstrated in Fig. 8f, core-shell shaped LM particles were synthesized with the acrylate ligands, forming polymerized LM networks (Poly-LMNs). The resistance of Poly-LMNs changed about 10% under 100% strain over 10,000 cycles.

Generally, LM-based devices are equipped with both excellent stretchability and good conductivity as well as resistant stability. Their outstanding performances make them to have wide application prospects. However, reliable stability of resistance can only be achieved within a small range of deformation. Therefore, methods to enhance the electrical stability under large-scale strain are required to be investigated in the future.

Therefore, many LM-based conductors are highly durable, exhibit excellent stretchability, and maintain good conductivity after continuous stretching. These advantages make LMs promising candidates for manufacturing flexible electronic biomedical devices.

Applications oriented toward human therapy

Nerve repairment and neural interface

Nerve injury, a challenging condition to cure, affects thousands of patients annually. Unfortunately, current treatment protocols are ineffective in repairing damage to the central nervous system156. Additionally, when the nerves are injured, the axons are often stretched over long distances, leading to a lack of contact between tissues and prolonged atrophy of target tissues157. Therefore, researchers are actively seeking a method to repair injured nerves efficiently. Fortunately, LMs offer a potential solution to this problem.

Since the 20th century, scientists have been conducting research on neural implants. In 1999, Williams et al.158 introduced a protocol for recording chronic and multi-site unit data in the cerebral cortex of awake animals over an extended period. This protocol served as a model system for studying the biocompatibility of neural implants and laid the groundwork for future neural implant studies.

Over the past decade, many neural experiments utilizing LMs have been carried out. For instance, Jin et al.135 proposed a method for directly fabricating 3D medical electronic devices in vivo in 2013. They utilized gelatin as the primary packing material with GaInSn alloy inside. Gelatin was injected into porcine tissue, then a mold was filled into the gelatin and removed after shaping. Finally, the LM was injected into the shaped gelatin to form medical electronics (Fig. 9a). The as-formed electrode demonstrated the potential of LMs as an efficient electrocardiograph and stimulator electrode. In the same year, Hallfors et al.73 depicted a protocol to stimulate neurons together with the subcellular compartments moderately. As illustrated in Fig. 9b–d, LM electrodes were combined with microfluid culture platforms, and these LM electrodes were used to target and depolarize axons, emerging as a useful tool for treating multiple neurological diseases.

Fig. 9: LM electrodes for nerve repairing.
figure 9

a Fabrication processes of a 3D medical electronic device135. Copyright 2013, Springer Nature. b The schematic diagram of the LM-based electrode73. c The profile view of the red line in figure b73. d The DIM image of the interface region73. bd Copyright 2013, Royal Society of Chemistry. e Images of an LM-based fluidic cuff electrode161. f The photographs and schematic of the LM-based cuff electrode in the stretchability test161. e, f Copyright 2022, Elsevier.

Following the initial research, an in vivo experiment was conducted further to explore the potential of LM in the neural application. In 2014, Zhang et al.159 testified that LM could serve as a medium for transmitting neural signals and stimulating the muscles of frogs. In 2016, Liu et al.160 conducted experiments on mice, utilizing a combination of liquid gallium (Ga) and silicone rubber tube shells to connect two injured nerves. The electrode connected to the injured nerve exhibited similar outcomes to the uninjured nerves. As a result, the relevant muscles were able to avoid the issue of atrophy. Additionally, a microelectrode array based on LM was introduced in the subsequent year60. This flexible array addressed the problem of potential mechanical incompatibility, which is often encountered with traditional rigid stimulating electrodes. Remarkably, even deceased frogs were able to move their legs when implanted with the microelectrode array rhythmically. This advancement in therapy opened up the possibilities for restoring nerve functions.

However, these initial experiments discussed earlier did not involve the complete implantation of peripheral nerve interfaces or the recording of peripheral neural signals. In 2022, Tang et al.161 fabricated LM-based fluidic cuff electrodes (Fig. 9e) and implanted them into rats for two weeks (Fig. 9f). The results demonstrated that these stretchable nerve cuff electrodes were able to transmit neural stimuli to the peripheral nerve and activate the somatosensory cortex. Importantly, the long-term implantation did not adversely affect the well-being of the rats.

Based on the previous research, it can be concluded that LM-based neural electrodes exhibit excellent biocompatibility and flexibility. Their high conductivity makes them an outstanding signal transmission characteristic. Therefore, LMs are expected to play a significant role in nerve repair, such as in neural prostheses, neural stimulator systems, nerve connection agents, etc.

Furthermore, equipped with good biocompatibility, high conductivity, and good stretchability, LMs can also be used to fabricate neural interfaces. For instance, Dong et al.162 provided a fabrication strategy to manufacture a stretchable electrode to monitor real-time epileptiform activities. The electrode was compatible with the brain surface and could record the signals. Other researchers have also reported plentiful LM-based neural interface materials fabrication methods163,164,165,166,167 and devoted themselves to the relevant biological experiments, promoting the development of neural interfaces. In the future, neural interface systems may serve as monitors for human health recording and have the potential to realize the interaction between human beings and electronics. Although a series of experiments are needed to achieve the goal of nerve therapy and neural interface fabrication, LM is undoubtedly a promising material for neurosurgery.

Liquid-solid phase transition alloy cement

On a daily basis, our bodies endure the weight and strain that can cause small cracks in our bones168. Remarkably, our bones possess the ability to self-repair, promptly refilling these small cracks before they can cause irreversible damage. However, as we age, the self-repair capability of bones diminishes, making orthopedic diseases such as osteoporosis more common among the elderly. Additionally, the violent impact also leads to fracture. Therefore, using some methods to cover the shortage of self-repairing bones is of great significance due to their structural material function in human bodies.

Polymethyl methacrylate (PMMA) cement and calcium phosphate cement (CPC) are two traditional bone cement used to repair bone defects169,170. Nevertheless, these traditional materials have many flaws171. For example, during the polymerization process, these materials generate excessive heat and then impair the surrounding tissue, leading to the decreased mechanical properties of bone cement172. Moreover, traditional materials do not return to a liquid state after solidification, making their removal challenging. Additionally, the leakage of cement can potentially cause severe cardiac injuries173. Therefore, there is an urgent need for new types of bone cement to address these issues. LMs are anticipated to provide solutions to these challenges.

In 2014, Liu et al.129 proposed a Bi-based liquid-solid transition alloy with a low melting point of 57.5 °C. Unlike traditional PMMA bone cement, which consists of separate powder and liquid systems, the liquid-solid transition alloy only has one system, where the liquid and solid phases can be reversibly transformed. This simplified the complex preparation process. Distinctively, while the solidification of PMMA cement involved a chemical reaction, LM cement relied solely on a physical process. Furthermore, the temperature decreases rapidly from its peak during solidification (Fig. 10a), reducing trauma to surrounding tissues. Surgeons can also preheat the materials at a lower temperature or prechill the prosthesis and the bone cavity before filling to minimize the time spent at high temperatures. This mild thermal effect not only holds the potential to improve repair effectiveness but also lowers the likelihood of premature cement failure. Rapid cement molding can also reduce the incidence of infection during surgery. However, although LM cement has a higher modulus, it behaves less effectively than acrylic cement in compression and bending strength tests. Therefore, utilizing the LM cement in non-bearing positions might be a more suitable choice. Nevertheless, LM cement performs well in fatigue, crack resistance, and safety tests. Furthermore, with appropriate modifications, LM cement may perform better174. Further research on the Bi-based LM demonstrated excellent characteristics in corrosion resistance (Fig. 10b) and a capacity for osseointegration. Moreover, the injected alloy remained immobilized for 210 days128.

Fig. 10: LM-based bone cement for bone repairment.
figure 10

a The temperature-time curve of the cement injecting process129. Copyright 2014, Elsevier. b Mass change of Bi alloy immersed in HBSS solution128. Copyright 2021, John Wiley and Sons. c The sketch map of the fabrication process74. Copyright 2022, Elsevier. d The temperature change produced by the liquid-solid phase transition of the LM-based bone cement175. e The 3D numerical knee model for the temperature simulation175. fh The temperature distribution of diverse melting point/volume/amount successively175. dh Copyright 2019, Springer Nature.

Another LM-based scaffold for bone tissue engineering has been developed to provide a changeable material for bone regeneration. Li et al.74 introduced magnetic liquid metal (MLM) and porous magnetic liquid metal (PMLM), which could change their shapes by changing the external magnetic field. The MLM scaffold was fabricated by introducing magnetic silicon dioxide particles into Galinstan, while adding polyethylene glycol as a template resulted in the fabrication of the PMLM (Fig. 10c). The result demonstrated that the external magnetic field could change the stiffness of MLM by rearranging the distribution of magnetic particles, which can stimulate the osteogenic differentiation of mesenchymal stem cells. Moreover, as shown in Fig. 10d, other research also reported the rapid cooling property of LM-based bone cement during solidification175.

Nevertheless, several investigations indicated that the supercooling effect of LM during the solid-liquid phase transition may result in potential side effects on human tissue. For example, Zhang et al.175 evaluated the negative effects of latent heat from LMs on biological tissues. BiInSn alloy, one of the LM-based bone cement materials, was selected for the test of thermophysical properties. They injected different amounts of bone cement with various melting points into the bone cavities and then measured the temperature of the bone-cement interface together with the peripheral bone tissue by using thermocouples to evaluate the potential thermal necrosis. Besides, a 3D human knee model was constructed to assist in evaluating the thermal parameters numerically (Fig. 10e). The data indicated that not only did the larger amount of bone cement cause a higher latent heat, but the higher melting point of LM alloy also resulted in the peak temperature soaring. Moreover, the LM bone cement with higher latent heat had an extended solidification period, causing irreversible damage to the bone tissue.

In conclusion, LM-based cement has good biocompatibility, low melting point, and can be introduced into bodies easily128. However, the high latent heat associated with LMs requires more time for solidification and may lead to an irreversible impairment of the surrounding tissues175. In general, the LM with a higher melting point (Fig. 10f), higher diameter (Fig. 10g), or a larger amount (Fig. 10h) generates a higher temperament in bone tissues. Therefore, various orthopedic diseases could be treated with different strategies. For example, the LMs with a higher melting point could serve as bone tumor treatment devices, while the lower melting point LMs can be used to treat bone defects. Nonetheless, deeper investigations are required to unveil the miracle of LM bone cements.

E-vessels

Despite the advancements in medical technologies, cardiovascular disease continues to pose a threat to individuals’ health. Coronary artery bypass grafting surgery is one of the main treatments to cure the disease.

For instance, Quint et al.176 introduced an arterial tissue engineering method to fabricate short-term-transplantation tissue-engineered vessels (TEVs). The results indicated that the endothelial progenitor cells and endothelial cells planted in TEVs did not show a tendency for graft occlusion. Besides, another research reported a TEV that shared a similar burst strength with human vessels (>2000 mmHg)177. However, lots of the tissue engineering blood vessels (TEBVs), which are fabricated by conventional methods and materials like polyvinyl alcohol-silica (PVA-SiO2)178, merely function as stents, and hardly help the new blood vessels to regenerate176,179,180,181. Besides, the complex interactions between the TEBV and blood might induce inflammations, leading to other issues like thrombosis or neointimal hyperplasia180,182. Fortunately, Jiang et al.183 introduced a kind of TEBV, which was conductive, stretchable and biocompatible, meeting the requirements of clinical surgeries.

The e-vessel (Fig. 11a) was manufactured by rolling up a poly(L-lactide-co-ε-caprolactone) (PLC) based MPC membrane (Fig. 11b) with the assistance of a polytetrafluorethylene mandrel. The MPC had a conductivity of about 8 × 103 S cm−1, and its resistance remained stable, even after being bent and rubbed approximately 1000 times. This remarkable conductivity of the MPC enables the e-vessel to hold a promising application in monitoring and treating diseased regions. Moreover, in-situ monitoring and in vivo implantation in rabbits demonstrated good biocompatibility and stability of the e-vessel. Figure 11c illustrates the diameter of the e-vessel during the implantation, while Fig. 11d shows the velocity distribution of blood flow at the e-vessels from numerous rabbits (Fig. 11e, f) at different time points. Both of which indicated the marvelous biocompatibility of the e-vessel. Furthermore, the e-vessel also performs marvelously in mechanical testing (Fig. 12a–f) and in vivo (Fig. 12g). MPC has a much greater elastic modulus (130 MPa) and maximum tensile strength (27 MPa) than the native vessels. Though the compliance of MPC is lower than the native vessels, MPC shares a similar burst pressure (about 2800 mmHg) with the native carotid artery. Generally, the robust mechanical properties like maximum elongation, elastic modulus, and maximum tensile strength guarantee the MPC a potentially competent in replacing the pathological vessel.

Fig. 11: LM-based e-vessels in tissue treatment.
figure 11

a Photographs of the electronic blood vessel183. b Photographs of the MPC-PLC membrane183. c The diameter changes in the electronic blood vessel after different times post-implantation. The error bars show the standard deviation183. d The scatter diagram of the velocity of blood flow at the operational site, where the middle velocity was about 0.47 m·s−1. The error bars show the standard deviation183. e, f Image before injecting the contrast medium. The red box indicates the position of the implanted electronic blood vessel. The scale bar is 1 cm183. af Copyright 2020, Elsevier.

Fig. 12: The different behaviors between e-vessels and native vessels.
figure 12

a The curve of stress-strain comparison between the electronic vessel and native vessel183. b The compliance test comparison between the electronic vessel and the native vessel183. c The burst pressure test comparison between the electronic vessel and the native vessel183. d The elongation at break comparison between the electronic vessel and native vessel183. e The modulus of elasticity comparison between the electronic vessel and the native vessel183. f The ultimate tensile strength comparison between the electronic vessel and the native vessel183. bf The error bars show the standard deviation. g A schematic of the e-vessel of the rabbit183. ag Copyright 2020, Elsevier.

In summary, the electron vessel shows the promise in treating cardiovascular disorders. In the future, by combining the e-vessel with other stretchable electronics, such as sensors, the multifunctional electron vessels could serve as replacements for the original vessels and enable the monitoring and treatment of cardiovascular disease. Nonetheless, further research and in vivo testing still hold immense importance before widespread application can be realized.

Heater

Moderate calefaction on the skin can accelerate blood circulation, thus relieving the pain and the cramp184. Thermotherapy is a highly effective clinical therapy for a wide range of diseases, known for its low side effects. However, traditional heaters often face challenges when maintaining their heating temperature during stretching or movement, reducing their functionality185. For example, the resistance of conventional conductive fibers changes significantly after stretching, which may result in the uncontrollability of temperature when using heaters, leading to inconvenience and even hurt186. On the other hand, although the addition of conductive particles can stabilize the stretching conductivity of the conductive fibers, it also increases the stiffness of the conductive fibers187. In contrast, LM-based heaters provide an ingenious solution97,188,189,190.

Wang et al.185, proposed a stretchable LM-based heater that showcases the immense potential of LM in wearable thermotherapy devices. The heater is made of Ga-based LM and polydimethylsiloxane (PDMS). The LM@PDMS heater demonstrated outstanding conductivity (1.81 × 103 S cm−1) and stretchability (over 100% strain). Figure 13a illustrates the relationship between saturation temperature and stretching strain under 2.0 V for heating 240 s, and the decay of saturation temperature is less than 8%. Moreover, Wang and the team captured photos of the heaters during various exercise states, accompanied by corresponding infrared thermal images (Fig. 13b), thus illustrating the heater’s ability to maintain a stable heating temperature during physical activity. Furthermore, owing to its high stretchability and simple fabrication procedures, the LM@PDMS heater can be easily molded into various shapes to ensure compatibility with different parts of the body.

Fig. 13: Electrical and thermal behavior tests of LM-based heaters.
figure 13

a The strain-temperature diagram of the stretchable LM-heater under 2.0 V for heating 240 s. The error bars show the standard deviation185. b Photographs and IR thermal images of exercise at different states185. a, b Copyright 2018, John Wiley and Sons.

Another LM-based heater was introduced in 2022. Tan et al.189 proposed a strategy for manufacturing liquid metal/carbon nanotube@Ecoflex (LCEF) coaxial conductive filaments. As shown in Fig. 14a–g, the LCEF textiles showed excellent electro-thermal performance under diverse voltages, which indicated the LCEF fabrics have a promising application in wearable heaters.

Fig. 14: Electro-thermal performance of LCEF fabrics.
figure 14

a Infrared images of the woven fabric at different voltages189. b The temperature sensing result of the LCEF is from 30 °C to 110 °C. The error bars show the standard deviation189. c The heating results of the woven fabric at different voltages189. d The heating results of the woven fabric at 12 V189. e Stepped heating results of the woven fabric at different voltages189. f Heating cyclic test results of the woven fabric at 6 V and 12 V189. g Long-term heating stability test of the woven fabric at 6 V and 12 V189. ag Copyright 2023, Elsevier.

Besides, there are multitudes of potential applications for LM-based heaters, such as wireless heater-based thermotherapy platforms188, electronic tattoos97, and the LM-filled magnetorheological elastomer190. Nonetheless, the LM heaters still cannot adjust the temperature automatically with the tissue temperature changing, which might lead to potential scald. Therefore, more research is significantly required to cope with this issue.

Applications oriented towards the human function augmenting

Exoskeleton device and artificial muscle

Exoskeleton devices have the potential to assist elderly and disabled patients in movement, increase human strength, and seize human’s sights. Scientists have already proposed designs for such devices191. With the soar of technology, plentiful novel materials are being explored for exoskeleton devices, and LM is considered as a potential material in this field192. The artificial joint is a key focus in the exoskeleton and human function augmenting, and in 2010, scientists developed a smart hydraulic joint for robotic structures193, making the manufacturing of artificial joints for mankind a promising endeavor194.

For instance, Liu et al.195 reported a flexible mechanical exoskeleton (Fig. 15a) with a low melting point and easily underwent liquid-solid phase transition. This exoskeleton had two working states. When the wearer needed to carry loads, the LM-based exoskeleton was cooled into a solid state with the assistance of the cooling device, which then supported the arm handling the loads. On the contrary, the exoskeleton would be heated into a liquid state after working to improve the comfort level. The mechanical test, compared with the paraffin phase change material (Fig. 15b), demonstrated the extraordinary mechanical strength of the LM-based exoskeleton.

Fig. 15: LM-based exoskeleton devices and e-skins.
figure 15

a The top view of the flexible mechanical exoskeleton195. b Comparison of mechanical strength between conventional paraffin-based mechanical joints and that based on low melting point alloy195. a, b Copyright 2014, American Society of Mechanical Engineers. c The temperature changes of TE-skin under various airflows, where the temperature of the film heater is 40 °C205. d, e A bracket was charged by TE-skin on the human arm205. f, g A cellphone was dialed by the TE-skin205. cg Copyright 2024, John Wiley and Sons.

Similarly, the LM-based exoskeleton can also be applied to artificial muscles. Shu et al.196 proposed a liquid metal artificial muscle (LMAM) that could work in various solutions with a pH range from 0 to 14. The LMAM had a maximum expanding speed of 15 mm s−1 and generated actuation strains of up to 87%. Moreover, the low driving voltage (0.5 V) indicated that a small energy supply system satisfied the need for LMAM. Furthermore, LM consisted of the actuating component, averting mechanical fatigue and movement limitation. Not only could the LMAM mimic the tail fin of the fish, but it also had a promising future in biomedical devices.

The intrinsic fluidity and the electrical conductivity of Ga-based liquid metal (GbLM) enable the GbLM to play an irreplaceable role in stretchable electron devices. A method of manufacturing the high-aspect-ratio LM was presented in 2022, and the high-aspect-ratio LM pattern could be applied to the area of wearable flexible devices, orienting towards human function augmenting and therapy197.

E-skin

Electronic skins have also attracted wide scientific interest. One of the most important functions of e-skins is monitoring human health. Several e-skin monitors have been proposed in recent years by researchers198,199, and LMs have already been regarded as a potential material for e-skins200,201.

For instance, Gao et al.202 proposed an LM-based electronic skin based on embedded Galinstan microchannels and could distinguish the pressure variation less than 50 Pa under the initial pressure of fewer than 100 Pa within 90 ms. Another research reported a wearable touch sensor that could operate from 4 kPa to 100 kPa with a high sensitivity of 0.05 kPa−1 203. The e-skins not only served as sensors, but they could also be used for temperature-conditioned devices204. In 2021, Xiang et al.205 introduced a self-powered LM-based electronic skin, which helped with thermoregulation. The team adjusted the proportion of Ga, In, and Sn to modify the melting point of the alloy, so that thermoregulating electronic skin (TE-skin) could transform its phase according to the surrounding temperature. As shown in Fig. 15c, the temperature of TE-skin kept stable at around 30 °C when melted, which was a comfortable temperature for human skin206. Moreover, LM-based electronic skins could also be utilized as anti-microbial screens and other electronic devices (Fig. 15d–g). Xu et al.207 proposed an EGaIn-based soft conductor with high anti-leakage and anti-microbial characteristics. The experiment showed that the soft conductor could inactivate more than 97% of the SARS-CoV-2 virus in 10 s. Thus, it could serve as a long-term biomedical electronic skin.

Challenges and prospects

Equipped with properties such as good biocompatibility29,30,31,32,33,34, electrical conductivity36,46, and stretchability37, LMs are promising materials for biomedical devices. Nevertheless, several challenges must be addressed before their widespread use.

First and foremost, the corrosion of LMs in a biological environment is an issue that needs to be solved. Many LMs are susceptible to corrosion in the presence of moisture208, which can damage LM-based devices and impair their performance. Additionally, LMs are highly reactive and can release ions60, potentially causing harm to local tissue if they react with human organs133. Moreover, there is still a lack of long-term biosafety experiments and clinical practice, limiting the clinical results60,209. Another challenge is the high cost of LMs compared to other materials used in biomedical devices210. Furthermore, the fabrication of LM-based biomedical devices requires specialized equipment164, adding to the manufacturing costs. Additionally, LMs are not ideal for fabricating disposable electronic devices211. The stability of LM-based devices is also a significant issue that needs to be addressed. The active nature of Ga, Bi, In, and Sn atoms can result in incorrect connections under electric fields209. Besides, the connection between rigid electron devices and the LM circuit is unstable, leading to the inhomogeneous current distribution and accelerating the failure of LM-based devices.

LMs also show their marvelous effect in other biomedical fields. For instance, LMs can serve as raw materials for fabricating nano-targeted medicine for cancer treatment212. However, as a relatively new material, LMs require deeper investigation before being applied to biomedical devices. Further research is needed to optimize the application of LMs in biomedical devices, such as modifying LM-based nano-targeted medicine using proteins213. Despite these challenges, the prospects for LMs in biomedical devices remain promising. With continued research and development, more effective, efficient, economical, and easily processed facilities can be developed. LM-based biomedical devices can revolutionize medical technology.

Conclusions

As biocompatible, electrical, and thermally conductive, stretchable, and chemically stable materials, LMs exhibit promising prospects in biomedical applications. LMs can serve as neural interfaces, alloy cements, e-vessels, and heaters, facilitating disease treatment and improving human health. LMs also hold great potential for human function augmenting, such as exoskeletons and e-skins. Nevertheless, there are challenges to address in the biomedical use of LMs, including high costs and corrosive issues. With deeper investigations and further development of LMs, the revolution of medical technology can be realized, satisfying the public’s demand for an enhanced quality of life.