Introduction

The number of intravitreal injections has increased significantly, outnumbering cataract surgery as the most common intraocular procedure. [1] There are well-established indications for intraocular antiangiogenic use: AMD, diabetic macular edema, retinal vein occlusions, myopic choroidal neovascularization, and proliferative diabetic retinopathy [2].

Virtually all brands of syringes used for intraocular injection have silicone oil (SO) on their inner walls, which is added during their manufacturing, a process known as “siliconization”. Among other purposes, it facilitates gliding of the plunger during injection.

Since most, if not all, syringes have not been developed specifically for intraocular use, insulin/tuberculin syringes are commonly used. The SO present in their interior can be released into the vitreous as droplets during intravitreal injection [3,4,5,6,7]. It might lead to an annoying complaint of floaters that, in some cases, requires vitrectomy. Although vitrectomy is considered a safe procedure, there are inevitable complications possibly associated with any surgery. The American Society of Retina Specialist 2018 Preferences and Trends Membership Survey showed that the 60.4% of US respondents have seen SO in the vitreous, while 5.2% already have performed vitrectomy to remove the oil. Moreover, 1.8% reported that patients have sought legal action because of the presence of the floaters [8].

Another important issue concerning the SO released by the syringes refers to a possible association to inflammation. A case-control study from our group showed a chronological link between noninfectious inflammatory reaction after intravitreal injection of aflibercept in six patients and the use of a specific brand of syringe subsequently shown to release a significant amount of SO [3, 9].

All these findings and the relevance of this topic led us to study whether agitation promotes the release of SO by different brands of syringes.

Materials and methods

Eight syringe models were analyzed for the release of SO. Detailed information about each one is presented in Table 1. All syringes were tested with their own needles. Solidor (B), Injex (C), BD Ultra fine II (D), and SR (E) have fixed needles, while Descarpack (A), BD Plastipak (F), SR (G), and SR (H) have attachable needles.

Table 1 Syringes detailed description, including the needle used, manufacturer, distributor, and lot number.

Thirty syringes of each model were tested under three different handling conditions (ten syringes of each model for each condition): positive control (fluid with deliberate addition of SO) with agitation by flicking (group 1); fluid with agitation by flicking (group 2) and fluid without agitation (group 3). The fluid used for all syringes was distilled water (Samtec Biotecnologia Ltda, São Paulo, Brazil) and the SO used for the positive control was the same used in vitrectomy surgeries (RS-Oil ECS, Alchimia, Ponte San Nicolo, Italy, batch: P180020016, Ref.: RSO 010–00).

Preparation of syringes

In order to simulate the clinical conditions of intravitreal injection, 0.06 mL of distilled water was aspirated into all the syringes of the three groups (filled by drawing with their own needle). Only the syringes of the positive control group had addition of a drop of SO to the tip of the plunger rubber (RS-Oil ECS, Alchimia, Ponte San Nicolo, Italy, lot: P180020016, Ref.: RSO 010–00). In groups one and two, after aspiration of distilled water, five flicks were given with the syringe needle facing upward and, then, ten more flicks with the needle of the syringe facing downward. No agitation was performed in Group 3 after aspiration of the liquid. Immediately after this procedure (carried out in one syringe at a time), the fluid was ejected onto a glass slide (Yancheng Huida Medical Instruments Co., Jiangsu, China) and analyzed with a light microscope (Nikon Eclipse 200, Shanghai, China). A single investigator (CSDJr) performed all steps in all syringes in order to avoid inter-examiner variability.

Microscopic analysis of syringe fluid

For each syringe, a different slide was used and was observed using an optical microscope under ×100 optical magnification. An additional 50% digital magnification and photographic recording were used with an iPhone 6S (Apple Inc., California, USA) attached to a microscope adapter (Eye2Mobile LLC, California, USA). An image was taken capturing the largest number of SO droplets per solution drop (ten drops onto each glass slide) for each syringe. The number of droplets of SO, when present, was analysed by the investigator who prepared the syringes (CSDJr) and another blinded examiner (GBM).

Fourier-transform infrared spectroscopy (FTIR)

A sample of each of the eight syringe models was studied by FTIR. After the plunger was removed, the rubber tip was gently rubbed onto a cesium iodide crystal to transfer the possible content of SO present in the syringe. Identification was performed using the Nicolet iS10 FTIR (Thermo Electron Scientific Instruments LLC, Madison, Wisconsin, USA) with bands ranging from 4000/cm−1 to 400/cm−1, resolution of 4/cm−1 and 32 scans.

Statistical analysis

Data analysis was performed qualitatively and quantitatively using the STATA 14.0 software (StataCorp LP, College Station, TX, USA). The presence/absence of SO droplets was analysed in terms of percentage in each type of syringe (A, B, C, D, E, F, G, and H) and under different conditions (positive control and with or without flicking). The comparison between the different levels of the variables of interest was performed using the Fisher's exact test. Firth's logistic regression was used to evaluate the presence/absence of SO droplets (dependent variable) in relation to the syringe types and evaluation conditions (predictive variables), presented as odds ratio. The number of oil droplets in each syringe was evaluated and compared between different syringe types and evaluation conditions from the Kruskal–Wallis test. The Dunn test was used as a post hoc analysis. For all the tests, a significant P value was considered when lower or equal to 0.05.

Results

A total of 240 syringes were analysed: 30 of each 8 models (A, B, C, D, E, F, G, and H). Ten syringes were evaluated in three different conditions (positive control, with agitation, and without agitation).

Qualitative analysis

The presence of SO droplets in each type of syringe and condition is shown in Table 2.

Table 2 Qualitative data for each condition and type of syringe.

The presence of SO droplets was observed in all positive controls. When agitated by flicking, 100% of the samples of the syringes disclosed SO, except for type F (BD Plastipak), which presented 40% of positivity. Without agitation, a smaller percentage of samples with SO were observed, except for type B (Solidor), which presented 100% of positivity. The Fisher’s exact test showed a significant difference for all syringe models, except for type B (100% of samples with SO).

Firth's logistic regression for the presence of oil droplets as response variable and syringe model and condition as predictive variables is presented in Table 3. The F syringe was used as the reference because it presented a lower percentage of oil.

Table 3 Firth’s logistic regression for presence of silicone oil droplets adjusted for syringe type and condition.

According to the syringe model, it was observed that all of them had a greater chance of presenting SO droplets when compared with type F (p < 0.05), when the condition was controlled. Regarding the condition, the logistic regression adjusted for the syringe type indicated that the syringes subjected to agitation by flicking had a 265-fold greater chance of presenting oil droplets when compared with the syringes without flicking (OR = 263.18; 95% CI: 15.43–4490.00; P < 0.001). As expected, the positive controls had a much higher likelihood of presenting oil droplets when compared with the syringe without flicking (OR = 7838.06; 95% CI: 125.42–489835.90; P < 0.001).

Quantitative analysis

Quantitative data on the number of SO droplets in each syringe (A, B, C, D, E, F, G, and H) under different conditions (positive control, agitation, and without agitation) are shown in Table 4. The Kruskal–Wallis’ test was used to evaluate the differences between the number of SO droplets between the types of syringe and the handling conditions.

Table 4 Quantitative analysis of the number of silicone oil droplets in each syringe under different handling conditions.

Statistically significant differences were observed between at least two conditions (p < 0.05) that were assessed by the Dunn’s test for all types of syringes. For syringes A, B, C, D, F, G, and H, there was a significant difference between the three conditions (P < 0.05). Syringe E presented a smaller number of droplets in the no-agitation condition when compared with the control (P < 0.001) and to the agitation one (P < 0.001). Syringe F showed a greater number of droplets in the positive controls in comparison with both agitation (p < 0.001) and no-agitation (P < 0.001). However, there was no statistically significant difference between agitation and no-agitation (P = 0.135).

Fourier-transform infrared spectroscopy (FTIR)

Analysis of the plunger tip by FTIR indicated the presence of polysiloxane (SO) in all models of syringe (Fig. 1). Although polysiloxane was the predominant compound, other compounds were found in small proportions: ester carbonyl (syringes A and F) and carboxylic acid salt carbonyl (syringes E, F, and G).

Fig. 1
figure 1

Fourier-transform infrared spectroscopy graph shows the characteristic bands of polysiloxane (silicone oil) in all syringes (A, B, C, D, E, F, G, H).

Discussion

Previous publications have reported the presence of SO droplets in the vitreous of individuals previously treated with intravitreal injections, assuming that oil comes from the interior of the syringes used in this procedure [4,5,6,7]. The present study identified the presence and release of SO in various models of syringe used for intravitreal injections under different handling conditions, especially after agitation by flicking.

In the positive controls, all ten syringe samples of each brand released SO droplets. When they were subjected to agitation by flicking in the study group, all syringes, except for six samples of the BD Plastipak, released droplets of SO. Without agitation, the syringes that released the smallest number of SO droplets were, respectively, the BD Plastipak (F), SR (G), and Descarpack (A). Notably, we did not verify whether these differences were consistent across multiple lots. Our previous studies have also shown similar results [9, 10].

The choice of syringes for this study was based on the models most commonly used in some European countries, North America, and Brazil. BD Ultra-Fine, although manufactured in the USA, is widely found and used in Brazil, and also is frequently sold with prefilled bevacizumab by compounding pharmacies in the USA. BD Plastipak is widely available in Europe, in Canada, and in Brazil. The others are more easily found in Brazil.

Emerson (2017) reported that some BD insulin syringes also released SO droplets [11]. The experiments were performed under steady-state conditions of manipulation of the syringes. It was speculated that the absence of a dead space in that specific syringe model was the cause. Theoretically, the dead space serves as a trap for high-resistance components, such as SO droplets.

Other studies have shown that insulin syringes release more SO droplets when subjected to more vigorous agitation, especially after agitation by flicking, compared with the syringes without prior agitation [9, 12, 13]. The practice of flicking the syringe during preparation of the medication is quite common, as an attempt to remove air bubbles from the syringe prior to injection. In addition, priming the plunger, different temperatures, and the presence of air bubbles (without agitation) had little impact on the release of SO by the syringes [12]. In the agitation groups, the presence of air also showed an incremental effect on the release rate [12]. Another study from our group also demonstrated the presence of SO obtained from the outer surface of 26- and 30-gauge needles commonly used to be attached to the syringes for injection [10]. A recent study from our group has shown no significant amount of SO released by the inner surface of most commercially available needles (unpublished data). This finding, added to the fact that the total area of the needle inner surface is much smaller than that of any syringe, makes us believe that the needles did not have any significant impact on the main findings of the study.

Other studies unrelated to ophthalmology [14,15,16] corroborate the fact that there is a greater release of SO from the inner walls of the syringes when they are agitated. In addition, the formation of protein aggregates is described from the interaction of certain medications with SO, air bubbles, and agitation. A similar process can occur with intravitreal injections, causing serious and previously unanticipated damage to eye.

FTIR confirmed the presence of polysiloxane (SO) at the tip of the plunger of all models plus two other compounds in smaller amounts: ester carbonyl (A and F) and carboxylic acid salt carbonyl (E, F, and G). Neither do they come from the production of silicone nor represent possible impurities. Further studies are required to understand it.

With the increasing number of intravitreal injections, the number of individuals with SO droplets in the vitreous and relevant complaint also increases. Further complications may occur when vitrectomy is required to treat floaters secondary to SO droplets in the vitreous. In addition, retinal specialists worldwide are concerned about the occurrence of ocular inflammation without clearly defined etiology after intravitreal injections of antiangiogenic drugs. Noninfectious vitritis was reported in 0.10% of cases after 66,356 injections of bevacizumab, 0.02% after 26 injections of ranibizumab (Lucentis, Genentech), and 0.16% after 8,071 injections of aflibercept (Eylea, Regeneron Pharmaceuticals, Tarrytown, NY) [17]. The American Society of Retinal Specialists Therapeutic Surveillance Committee also reported cases of sterile inflammation related to aflibercept without a clear explanation suggested by the authors [18]. Our group also identified six cases of ocular inflammation after aflibercept injections with the use of the SR syringe [3], which was later proven to release SO [9, 10, 12]. Since then, our group recommends that the syringes not be agitated during its handling for intravitreal injection, as it is speculated that the release of SO droplets may contribute to the inflammatory reaction. In agreement with this hypothesis, some studies have reported a more intense protein aggregation and the formation of insoluble molecules resulting from the agitation of the syringes in the presence of SO from the syringes [14,15,16].

Finally, it has also been reported that SO droplets can act as immunological adjuvants in dermatological procedures performed in the subcutaneous space in mice [19, 20]. In addition, a recently published study showed undesirable effects of SO on therapeutic proteins, including adsorption to the oil droplets and increased secretion of several innate cytokines from human peripheral blood mononuclear cells [21]. The authors concluded that the SO droplets form complexes with pharmaceutical proteins that can potentially invoke early- and late-stage immune responses [21].

The BD Plastipak syringe was shown to release some SO after agitation, which differs from one previous study of our group [9]. Not only was the size sample of the current study larger but also examiners were more trained than in the initial research period. In addition, FTIR showed confirmed that the BD Plastipak syringe really has SO on the tip of its plunger [13].

A limitation of this study was the quantitative analysis of the SO droplets, since the variation in size of these droplets was not taken into consideration and only the field with the highest number of SO droplets was chosen for quantification. A more accurate quantitative analysis test would be needed to draw more definite conclusions about the amount of SO. However, it does not invalidate the current findings because the qualitative analysis of the syringe samples with the presence of SO was reliable and reproducible.

In conclusion, syringes commonly used for intravitreal injections frequently release SO droplets, especially when agitated. In order to prevent potential hazards to the eye, we recommend that the syringes not be agitated at the time of the injection.

Summary

What was known before

  • SO might be seen in the vitreous—Syringes seem to be the source

What this study adds

  • Most syringes have and release SO—Agitation of the syringes causes additional release of SO