Introduction

Coronavirus disease 2019 (COVID-19) is caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), wreaking havoc worldwide, with over 33 million confirmed cases and in excess of 998,000 reported deaths by 28 September 2020.1 SARS-CoV-2 is transmitted primarily through direct or indirect contact with infected droplets. However, transmission of SARS-CoV-2 through aerosol has been implicated, particularly in the presence of highly concentrated aerosols in enclosed environments.2

It is accepted that aerosols are produced during a range of dental procedures, posing potential risks to both dental practitioners and patients. This concern, which prompted decisions to cease elective dental procedures during the sustained human-to-human transmission phase of the COVID-19 pandemic, has led to implementing transmission-based infection prevention and control precautions, including changes to personal protective equipment (PPE) requirements and intensified international debate on the generation of aerosol within dentistry and orthodontics. Routine use of fallow periods to allow for settling of suspended aerosol has also been variously instituted following selected aerosol generating procedures (AGPs), limiting the capacity for the provision of dental care. Notwithstanding this, there has been little consistency in the definition of an AGP, recommendation concerning PPE levels or the necessity for fallow periods following procedures.

The question of aerosol transmission has been contested within orthodontics, with British national guidance implying that orthodontic treatment is essentially non-aerosol generating.3 Furthermore, the NHS defines a dental AGP as one 'using high-speed devices such as ultrasonic scalers and high-speed drills',3 with the use of slow-speed handpieces (SHPs) not specifically mentioned. Orthodontic debonding and bonding procedures typically involve use of SHPs and conventional acid etch-based techniques, respectively, although alternatives do exist.4

Previous research into orthodontic-specific aerosol has focused on procedures involving enamel clean-up associated with debonding of fixed appliances.5,6,7,8,9 Particulate generation has been attributed to the use of handpieces, burs and water irrigants with a mass median aerodynamic diameter (MMAD) of up to 50 μm,6 although particles with MMAD of less than 0.75 μm have also been noted.9 Smaller particles are of more concern, as they are expected to take longer to settle10 and are more likely to reach the terminal alveoli of the lungs.7 Particles with an MMAD of 0.5-10 μm are considered to have the highest risk for transmitting infection.11 Respirable particulates can be produced during orthodontic debonding, irrespective of whether a fast handpiece (FHP) or SHP is used with or without water irrigation.9 However, FHPs used with water irrigation are considered likely to induce the most aerosol production.8 It has also been suggested that pre-procedural mouth rinses may offer no measurable benefit.6 The make-up of released particles incorporating composite resin, filler and metallic material has also been illustrated.5,12,13

In previous studies, qualitative sampling of particle emission allowing assessment of potential lung penetration and the collection of respirable particulate fractions has been undertaken with a Marple Cascade Impactor (Thermo Fisher Scientific, Franklin, USA).7,8,9 This monitor has also allowed for scanning electron microscopy and energy-dispersive x-ray spectroscopic (EDX) analysis.7 Other cascade impactors have also been used, such as the Six-Stage Viable Anderson Cascade Impactor (Thermo Fisher Scientific, Franklin, USA), which has been used to represent six stages of the respiratory tract.6 With regards to quantitative measurements, a Personal DataRAM pDR-1200 Real-Time Monitor (Thermo Fisher Scientific, Franklin, USA), which is a form of photometric monitor, has been used to measure the concentration of respirable particles.7,8 Markers such as fluorescence and citric acid used in waterlines and the spread of splatter have been used as surrogate measures of dental aerosol release.14,15,16 Alternatives have included attempts to identify environmental contamination using microbiological markers by cultured air sampling techniques.14,17 However, these approaches are designed to identify the characteristics and behaviour of aerosol produced in a dental operatory. Fluid dynamic methods involving more sensitive particle analysis offer the potential to identify the characteristics and behaviour of an aerosol, including the size, spread and longevity of particles following dental procedures.12,18,19 Few studies in dentistry have used particle analysis for assessment of aerosol.13,18

Previous research has failed to elucidate the quantity and nature of particles produced following orthodontic procedures and the duration required for particulates to reduce to baseline levels. This information would be helpful in determining the potential risk of SARS-CoV-2 from aerosol production during orthodontic procedures, including handpiece-based debonding and the use of 3-in-1 air-water syringes. The aim of this pilot study was therefore to identify whether aerosol is produced during orthodontic procedures and, if so, to estimate the duration of this effect. Secondary aims were to clarify the intensity of the aerosol produced, including the range of particle sizes, and to explore whether the use of water during debonding enhances or reduces emitted aerosol.

Methodology

Orthodontic treatment procedures were simulated in a closed side-surgery in a dental hospital setting (Institute of Dentistry, Barts and The London School of Medicine and Dentistry) using a dental manikin. The dental hospital has an air exchange system (with six air changes per hour) which was left functioning throughout the procedures.

A 1 mm layer of composite-based adhesive (3M Transbond XT Light Cure Paste Adhesive, Minnesota, USA) was light-cured on the buccal surface of all upper and lower acrylic teeth, including second molars (28 teeth in total). Adhesive was placed using orthodontic attachments with the base coated in petroleum jelly to prevent composite from adhering to the attachment surface, with 300 g of pressure applied for five seconds. Excess was removed and the adhesive light-cured for 20 seconds per tooth with a light-emitting diode (LED) curing light. A permanent black marker pen was used to identify the location of the adhesive and demarcate the surface of the manikin (Fig. 1).

Fig. 1
figure 1

a) Experimental model to simulate debonding with the location of adhesive demarcated. b) Adhesive was placed on the tooth surface with pressure to expel the excess adhesive using molar tubes and brackets as appropriate covered in a petroleum jelly layer. c) An even layer (approx. 1 mm) of composite-based adhesive remained on each tooth surface. d) Models of each arch were secured in place using silicone transfer jigs

Debond procedures were simulated with adhesive removal using a 1:1 contra-angle handpiece (W&H Synea Vision WK-56 LT, Bürmoos, Austria); handpieces were operated at maximum speed (up to 25,000 rpm) with and without the use of internal water coolant, with standard air coolant used but limited by regulating additional external air coolant for some experiments. Experiments were also undertaken using 37% phosphoric acid etch placed on all 28 teeth with use of combined 3-in-1 air-water syringe to remove etch. A two-bracket repair was also simulated with adhesive removal from two teeth (premolar and incisor). All experiments were undertaken using four-handed dentistry technique using high-volume suction (HVS). The operator and assistant only were present in the room during the pre-procedural period and left on completion of the procedure. Both operator and assistant wore fluid-resistant surgical masks (FRSMs) during the procedures. All procedures were undertaken by the same operator (PSF) in the same closed surgery. Procedures were repeated on separate days over a period of three weeks to ensure that variation in external factors pertaining to air exchange and air-conditioning systems were accounted for.

An optical particle scanner (OPS 3330, TSI Inc., Minnesota, USA) and a spectrometer particle scanner (NanoScan SMPS 3910, TSI Inc., Minnesota, USA; Figure 2) were used to assess particulate matter. The sampling inlets were located at a fixed position at 7 o'clock and located 8 cm away from the maxillary left central incisor (Aerosol Instrument Manager AIM v10.3.1.0 and NanoScan Manager v1.0.0.19, TSI Inc., Minnesota, USA), with a sampling rate of one minute. Measurement commenced ten minutes before the procedure to assess the baseline concentration of particulate matter and continued for 30 minutes following the procedure. Using both counters, it was possible to measure particles in 26 bins, ranging from 10 nm to 10 μm in diameter. As a single SAR-CoV-2 viron is approximately 80-100 nm in diameter,20 information on particles smaller than 80 nm was deemed to be irrelevant to virus transmission and was discarded.' In order to reduce the remaining data, the particle counts were combined into four categories: 'very small' (0.08-0.26 μm), 'small' (0.26-0.9 μm), 'medium' (0.9-2.7 μm) and 'large' (2.7-10 μm). In most cases, no liquid coolant was used and any increase in aerosol concentration can be attributed solely to the solid matter removed; however, when water is used, the measurements may reflect a combination of the solid matter and liquid aerosols, which will evaporate within a few seconds.21

Fig. 2
figure 2

Working principle of the instruments. a) NanoScan 3910. b) OPS 3330. Images are courtesy of TSI Incorporated, Shoreview, MN (USA) and protected by copyright

Tests were first inspected to ensure that aerosol levels were stable during the initial fallow period and cases in which large fluctuations were observed were disregarded. To quantify any changes in the particle concentration, the median of the concentration of particles in each size range during each procedure was calculated and was compared to the corresponding value in the initial ten-minute fallow period before each procedure. The Mann-Whitney U test was applied to estimate the probability that the median of the measurements acquired during the procedure was different to that of the fallow period (that is, that the concentration of aerosols had increased). In this context, a high P value indicates a high likelihood that the procedure has led to a change in the median aerosol concentration. This statistical test was not applied to the 3-in-1 air-water syringe used with etch or the breakage repair, as these procedures took less than one minute and therefore represented a single data point.

Results

Most procedures ranged in duration from 4-11 minutes for debonding using either air, no air, water or a FHP (Table 1), with the exception of washing of enamel with 3-in-1 air-water syringe and the repair of two teeth, which both took less than 60 seconds. Data showing the percentage change in the median concentration of aerosols during each procedure (relative to the initial fallow time) and the probability that this change is statistically significant (based on the Mann-Whitney U test) are presented in Tables 1 and 2, respectively. Most procedures led to a statistically significant increase in aerosol concentrations in most size ranges (Tables 1 and 2).

Table 1 Percentage change in the median concentration of particles in various size ranges during the procedures, with respect to the baseline levels (measured in the ten-minute period prior to the start of the procedure). For the final two procedures (repair and the use of 3-in-1 air-water syringe used to remove etch), the values show the percentage change in the single data point recorded during the procedure, relative to the baseline
Table 2 Probability that there has been a change in the median particle concentration measured during a procedure relative to baseline. Probabilities were calculated using the Mann-Whitney U test, with values close to 1 indicating a high probability that the dental procedure led to an increase in aerosols within that size range

The variations in the measured concentrations of aerosols during each experiment are demonstrated in Figure 3. The use of handpiece-based debonding resulted in an increase in particulate matter (Fig. 3a and Table 1). A standard debonding procedure (involving air but no water) was associated with a clear increase in the 'very small' and 'small' particles but only for a short period, with no significant increase for other sizes. Debonding procedures without supplementary air coolant appeared to produce similar levels of aerosol to standard debonding. A sharp increase in particles at the start of one procedure with air was observed, which we believe is most likely caused by the suboptimal position of the HVS with respect to the handpiece as the procedure commenced, and appears to be an outlier. However, all three repeats show a consistent increase in the concentration of 'small' and 'medium' particles (Fig. 3b[ii], Fig. 3b[iii] and Table 2). Debonding with water tended to result in large peaks, producing particles of all sizes throughout the experiment. The aerosol increase during debonding with water appeared to be of a higher magnitude than for standard debonding procedures due to the production of both solid particles and droplets of water. However, it is significant that this increase in aerosol levels remains for several minutes after the end of the procedure, which is significantly longer than the predicted evaporation time of liquid aerosols.21 This therefore reflects an increase in potentially infectious solid matter, rather than simply droplets of water from the coolant system.

Fig. 3
figure 3

Variation in the measured concentrations of aerosols during each experiment. The rows correspond to the different procedures: a) standard debond; b) debond without air coolant; c) debond with water; d) repair of two teeth using standard debond approach (with air only); e) repair of two teeth; and f) debond using the fast handpiece. The columns show the concentrations for various size ranges of aerosols: i) very small; ii) small; iii) medium; and iv) large. The grey shaded regions indicate the time during which the procedure took place and the colours indicate the different repeated measurements, as shown in Table 2. Note that the y-axis scale is different for the fast handpiece procedures (row f)

The use of an FHP with water led to the most significant increase in particles. However, in this case, the increase was associated with short spikes that were very large, rather than a continuous increase, as was seen in previous cases. An additional test was performed to assess the aerosol produced during a repair procedure, in which the standard debond (with air but no water) was applied to only two teeth (Fig. 3d). As in the case of the complete debond (Fig. 3a), this was associated with an increase in aerosol concentration in the 'very small' and 'small' size ranges, although the relatively short duration meant the total increase in concentration is much smaller. Finally, the combined use of 3-in-1 air-water syringe to remove etch did not result in any detectable increase in the aerosol levels.

Discussion

The aerosol dilemma has been thrown into sharp focus since the onset of the COVID-19 pandemic, with binary discussion around the presence or otherwise of aerosol associated with various dental procedures.3 There has also been consideration of the potential risk attached to various procedures. This risk-based approach appears to be relevant in relation to orthodontics, in particular. Specifically, the majority of orthodontic procedures may lead to splatter while not producing aerosol. However, on the basis of the present study, it is apparent that SHP-based debonding procedures do generate aerosol, although the magnitude of aerosol release is markedly lower than that emitted by FHP use.

The nature of the aerosol released during debonding has not been fully quantified in this analysis. However, given that no salivary substitute was used, it is likely to constitute dry debris in the absence of the use of water. The viability of envelope viruses such as SARS-CoV-2 is known to be reduced when virons are not in a liquid environment;22 however, in a clinical setting, viral particles are likely to reside in or be mixed with saliva, increasing the potential for transmission. Furthermore, a significant volume of 'small' and 'very small' particles were measured during orthodontic treatment procedures, in keeping with previous research.8 These are within inhalable and respirable fractions and it is believed that such droplets can conceivably lead to transmission of SARS-CoV-2.23

The release of fine particulate matter during full debonding procedures was observed consistently. Debonding under irrigation was also evaluated as it was hypothesised that adjunctive use of water might increase the mass and volume of any generated aerosol, making it more amenable to removal with HVS. A beneficial effect of irrigation, however, was not observed. Moreover, the performance of debonding with reduced air pressure on the SHP did not eliminate aerosol release. It therefore appears that debonding procedures involving the use of SHPs at conventional rotational speed will inevitably lead to some aerosol release. Notwithstanding this, the concentration and mass of released particles using the SHP for debonding was markedly lower than that observed with the use of FHPs. Moreover, particulate levels reduced to background levels within a period of five to ten minutes. It is important to note that the air filtration system within the environment assessed is likely to have facilitated this rapid dispersal. As such, given the time necessary to complete an orthodontic debond (including scanning or impression-taking), a post-procedural surgery fallow period is unlikely to be necessary. This may not, however, be true of all closed surgical environments where air filtration facilities are lacking.

Aerosol release was not noted during removal of acid etch, suggesting that this procedure can continue to be undertaken without concerns relating to aerosol generation and potential associated risk.3 Alternatives including the use of self-etch primer have been promoted. However, given the concerns associated with mechanical preparation of the enamel, a reduction in bond strength may be risked. While this may be offset with pre-etching of the enamel for up to ten seconds as a form of enamel pre-conditioning,24 our results indicate that these modifications may not be required.

The findings from the present study are consistent with previous research identifying aerosol and splatter, which may be a possible route for microbial transmission within a clinical environment.11,18 Two primary clinical considerations exist: how to protect the treating clinician and assistant from the aerosol released, and whether aerosol emitted poses a risk to subsequent patients. The magnitude of the risk associated with orthodontic debonding is difficult to quantify; notwithstanding, given the low levels of aerosol produced with the SHP relative to FHP use, it would be reasonable to infer a commensurately lower risk profile. However, the routine use of enhanced PPE and, in particular, masks would be sensible whether the FHP or SHP is used for adhesive removal during debonding. FRSMs have been shown to have limited value in relation to dust resistance, with a quoted 96% reduction pertaining to fluid resistance, in isolation.8 Furthermore, Type I and Type IR face masks have a bacterial filtration efficiency (BFE) of 95%, whereas Type II and Type IIR face masks have a slightly higher BFE of 98%.25 These medical masks are tested in the direction of exhalation (inside to outside) and take into account the efficiency of bacterial filtration. Surgical masks of this type protect from droplet transmission of the infection but will not offer adequate protection against close-range aerosol transmission for viruses such as coronavirus.25 As such, further refinement and guidance in relation to the appropriate mask protection is required.

In terms of the risk attached to orthodontic debonding, it is important to note that differences exist between removal of superficial composite material and operative dentistry. The latter may involve excavation of infected dentine and enamel, and pooled microbial communities; however, debonding is based on removal of a surface composite layer where infected dental tissue is not encountered. The associated risk on the production of harmful bioaerosol is, however, uncertain; SARS-CoV-2 receptors are present in salivary glands and the pharynx, meaning that salivary and pharyngeal secretion could therefore potentially be infected. It is worth noting that excess adhesive flash surrounding orthodontic attachments may be exposed to the oral environment for some time. This could theoretically be reduced by limiting adhesive flash; although this has failed to reduce the production of particulates in the inhalable, respirable and thoracic fractions that have been demonstrated.7

The present study has highlighted the potential for relatively short-lived and low concentration of aerosol release during orthodontic debonding. While quantification of risk is fraught, it would be reasonable to assume that this is likely to be limited, given the lower particulate volumes and concentration for orthodontic debonding using SHPs relative to procedures involving FHPs. In certain jurisdictions, decisions were made during the first wave of the COVID-19 pandemic to cease elective appointments and indeed orthodontic emergencies, presumably on the premise of the 'precautionary principle', to mitigate against viral spread given the non-zero probability of risk. It is accepted, however, that orthodontics involves an ongoing course of treatment with recognised risks associated with non-compliance and unsupervised care, including unwanted tooth movement and iatrogenic damage.26 As such, while remote monitoring offers some promise in mitigating potential associated problems, the risks attached to in-person attendance should be balanced with the undeniable need for progression of treatment and physical attendance on a regular basis during active courses of treatment.27

A significant limitation of the present study was that the particulate model did not measure viral load generated from these procedures. Dental aerosols are distinct from respiratory aerosols as they produce their own external aerosol, with the potential for infectivity residing in the patient's mucus, nasal and oral secretions. The air and water jets produced by FHPs and SHPs in the oral cavity may become contaminated; the associated microbial concentration and risk of infectivity has not yet been elucidated fully. While quantification of the precise microbial contamination and infectious dose of the particles assessed was not possible, one can estimate the risk based on the known ability of the smaller particles to persist longer in the air and penetrate deeper in the respiratory tract. In general, smaller particles are more likely to be airborne; as such, the risk of airborne spread was considered.28

Conclusions

Based on this simulated study undertaken within a closed surgery, particulate matter is released during orthodontic debonding. The concentration and volume of matter produced using the SHP for adhesive removal during debonding is markedly less than that associated with FHP use. No increase in particulates was associated with prolonged use of a 3-in-1 air-water syringe. Particulate levels reduced to baseline levels over a short period (approximately five minutes) in this environment. Therefore, in the current SARS-CoV-2 risk environment, we recommend that orthodontic debonding be carried out using the SHP rather than the FHP and without the use of water as an irrigant, in order to mitigate the risk of infection to staff and patients. While it is difficult to extrapolate to other air environments, in this setting, the fallow periods between patients can be reduced to approximately five minutes and use of the 3-in-1 air-water syringe can be considered to be a low-risk dental procedure.