Abstract
Highly precise micromanipulation tools that can manipulate and interrogate cell organelles and components must be developed to support the rapid development of new cell-based medical therapies, thereby facilitating in-depth understanding of cell dynamics, cell component functions, and disease mechanisms. This paper presents a literature review on micro/nanomanipulation tools and their control methods for single-cell surgery. Micromanipulation methods specifically based on laser, microneedle, and untethered micro/nanotools are presented in detail. The limitations of these techniques are also discussed. The biological significance and clinical applications of single-cell surgery are also addressed in this paper.
Similar content being viewed by others
Introduction
The molecular biology of individual cells must be explored to understand the normal physiology and functional mechanism of cells in response to disease or injury. Cellular functions, such as metabolism, cell motility, and gene expression, are greatly affected by the properties of intracellular structures and organelles. Technologies that can explore the genetic factors and molecules that contribute to disease evolution at a single-cell level have become increasingly important to elucidate the molecular basis underlying the proliferation, transformation and metastasis of cancer cells1,2,3,4. Research on whole tissues delivers only a statistical average of numerous activities occurring in different cells. There has been a paradigm shift in modern surgery and medicine, with technologies becoming smaller and more efficient than their larger predecessors, increasing their effectiveness5. This has led to major improvements in areas such as minimally invasive surgery, drug delivery, diagnostics, and cell manipulation6. Studies of single-cell manipulation and surgery may suggest that genetic changes activated by tumorigenesis-related signaling pathways can cause healthy cells to mutate, becoming cancer cells7. Therefore, micro/nanomanipulation methods and tools that can perform single-cell manipulation and surgery have attracted worldwide attention in recent years, as they provide information about individual cells and their organelles8,9,10,11,12,13.
There are numerous key applications of single-cell surgery, such as cloning12, preimplantation and diagnosis14, gene editing15, cellular therapy16, and understanding the functions and activities of subcellular organelles and components7. Several micro/nanomanipulation tools and systems for single-cell manipulation and surgery, such as glass microneedles17,18,19, optical tweezers (OTs)20, microfluidics21, atomic force microscopy (AFM)22, dielectrophoresis23, and magnetic tweezers24, have been used for manipulating and interrogating single cells. Glass microneedles can be fabricated by pulling glass capillaries on heating filaments. The resultant capillaries feature a conical shape with one end from a few micrometers to a few nanometers in diameter. Several programmable micropipette pulling instruments can create micropipette tips with different sizes and shapes depending on their application, such as force probes for investigating muscle physiology. Kishino and Yanagida25 fabricated a micropipette 0.3 μm in diameter at the tip and 70–100 μm in length from a 1 mm glass capillary by using an electrode micropipette puller. Optical tweezers are strongly focused laser beams that can be used as end-effector tools to trap and manipulate dielectric particles with sizes ranging between a few nanometers and a few micrometers26. Given their noninvasiveness, flexibility, and precise manipulation of objects, OTs have been applied to numerous biological tasks, such as the rotation, transportation, stretching, and assembly of cells. Microfluidics is an interdisciplinary methodology at the intersection of many fields, such as nanotechnology, micromechanics, analytical chemistry, bioengineering, and microelectronics. Microfluidics was invented in the early 1990s by Manz et al.27. The goal of microfluidic technologies is to create a platform with a size of a few square centimeters or less to simplify operations involved in biology and chemistry, such as sample separation, preparation, detection, sorting reaction, and lysis. Improvements in on-chip integration and soft lithography have enabled us to create a network of microchannels that can be used to understand cell biology at a single-cell level by controlling fluid at the microliter and picoliter scales.
The above techniques and tools can be integrated to achieve precise, versatile, and advanced cellular and subcellular manipulations. Compared to traditional chemical and biological methods for cell manipulation, these methodologies offer numerous advantages, such as control of single or multiple cells, accurate simultaneous manipulation, the ability to approach subcellular structures and organelles, and flexible and repeatable cell manipulation. Cell micromanipulation methods also play important roles in precise cell surgery. In general, micro/nanomanipulation tools for single-cell surgery can be divided into three types: laser-based, microneedle-based, and untethered-micro/nanotool-based methods. The following sections describe the details of tools and methods developed for single-cell surgery using lasers, microneedles, and untethered micro/nanorobots.
Micro/nanomanipulation methods for single-cell surgery
In recent decades, several micro/nanomanipulation tools have been developed for micromanipulation tasks (Fig. 1)28,29,30,31,32,33,34,35,36,37,38. Specific mRNAs were extracted from live cells with AFM tips39, and the cells were punctured by nanorobots39. Carbon nanotubes attached at one end of a glass micropipette were used to transfer fluids in cells, interrogate cells, and perform electrochemical and optical diagnostics40. A number of laser systems have been used to perform surgery on cell membranes, including continuous wave and pulsed picosecond and nanosecond lasers41. Many other approaches use microscale devices, such as sampling probes42,43, semiclosed microchips44 and computer-assisted patch clamping45, to analyze cellular contents. The following subsections describe the details of the tools and methods developed for single-cell surgery using lasers, microneedles, and untethered microdevices.
Laser-based single-cell surgery methods
Optical transfection, which uses light in the form of a focused laser to deliver drugs into cells, has also garnered scientific interest. Optics-based technology is also useful for manipulating biological materials at the micron and submicron scales46. Various laser types may generate small temporary holes that facilitate the transport of plasmid DNA and other macromolecules. Laser-based single-cell surgery methods provide a noncontact, rapid, and sterile way of introducing membrane-impermeable compounds into cells. High-speed laser pulses for single-cell surgery have been widely studied in the past few years47,48. A noninvasive argon fluoride excimer laser with a wavelength of 193 nm was used to drill a hole in the outer membrane (zona pellucida) of mouse oocytes, as shown in Fig. 2a49. Figure 2b shows the application of high-intensity femtosecond (fs) laser pulses to live mammalian cells to perform nanosurgical cell isolation and membrane surgery50. As a high-resolution ablation tool, fs lasers were also used to disrupt single organelles of cells, such as single mitochondria (Fig. 2c)51.
Chen et al. developed a method of cell fusion by cutting the cell membrane52, as shown in Fig. 2d. In their method, two cells were placed near each other, and the membranes of both cells at the common junction were carefully cut by a solid-state laser, which resulted in cell fusion. However, the efficiency of cell fusion by this laser ablation was very low.
The use of an ultrafast pulsed laser for optical transfection allows precise targeting and excellent viability53,54. Even though different optical technologies are continuously being developed, several issues remain with the intracellular delivery of genes/transcripts, such as low efficiency and repeatability. Combining optical transfection with microfluidic technology has widely been utilized to improve transfection efficiency and throughput. To accomplish optical reprogramming of large cell populations, Uchugonova et al. developed an ultrashort fs laser-microfluidic cell transfection device55. Ultrashort laser pulses cause temporary membrane permeabilization in a microfluidic tube containing several genes, allowing the generation of contamination-free induced pluripotent stem cells55.
Schomaker et al. presented an optical transfection technique based on photosensitive materials56. They cultured cells with gold nanoparticles (AuNPs) and subjected them to fs laser pulses. Laser stimulation can activate photosensitive molecules localized to the membranes of endocytic vesicles, resulting in localized membrane permeabilization of the cell57. Cell processing for gene and cell treatments typically employs numerous distinct methods for gene transfer and cell separation or removal. Lukianova-Hleb et al. pioneered the use of plasmonic nanobubbles (PNBs) to simultaneously transfect target cells while eliminating undesired subsets of other cells58. After brief laser irradiation, transient PNBs were produced around AuNPs, resulting in a nanoscale explosion that enabled molecular cargo transmembrane injection.
OTs have been effectively used to investigate numerous biological applications in recent years21,59,60. As demonstrated in Fig. 2e, OTs utilize a Gaussian laser beam, which can create a 3D light gradient when focused by an objective lens, to manipulate microscopic particles that face two types of forces: the gradient force introduced by a scattering force and the gradient of field intensity produced by the photons hitting the particle along their transmission direction26. This scattering force must be overcome by the gradient force along the optical axis to form a stable trap. Trapping of micrometer-sized transparent particles was introduced by Arthur Ashkin in 197061 and then used to manipulate and trap bacteria and viruses62. OTs have also been exploited to perform numerous biological tasks, such as cell stretching and transportation63,64, mechanical property calibration65, and cell assembly57, because they can operate microparticles noninvasively, flexibly, and precisely. As the power of the OTs is in the range of a few hundred milliwatts in the sample plane, it is not sufficient to counteract the internal pressure of the cell or break the cell membrane to force external particles or internal organelles to move in or out of the cell, respectively. Therefore, OT has been used in conjunction with other micromanipulation tools to perform cell surgery. Two OTs were pointed on a single cell, and their position was controlled to alter the 3D position of the cell, which eventually made biopsies easier by bringing the nucleus opposite to the micropipette, as shown in Fig. 2f. Inna V. Il’ina et al. established a laser-based polar body biopsy method for preimplantation genetic diagnosis57. They used a fs laser to cut a hole in the membrane of an embryo and then utilized OTs to extract the polar body from the embryo, as shown in Fig. 2g. The removal of cell organelles from a yeast cell by OTs was presented in66. First, a fs laser was used to disrupt the cell membrane, and OTs were used to extract subcellular parts by changing the functioning mode of the laser. Similarly, we utilized OT to control the 3D orientation of a single cell prior to cell surgery, which greatly improved the efficiency of the surgical process21.
A laser-induced cavitation bubbled methodology (BLAST) was developed to increase the throughput of delivery of the particles in mammalian cells67. The abrupt generation of bubbles induces high stress on the cell membrane, which eventually creates pores in the membrane for intracellular delivery, as shown in Fig. 2h. Under rapid laser scanning, BLAST methodology can inject micron-sized particles into 100,000 cells in 1 min. The existing laser scissors methods, such as the fs laser methods described above, enable precise surgery in cells at the organelle level, but they are limited by throughput. High-throughput surgery methods such as BLAST can rapidly perform surgery on thousands of cells in parallel, but they fail to achieve organelle-level precision simultaneously. Achieving high precision and throughput of cell surgery at the same time is challenging but needed. Robotic control of laser scissors may enhance the speed of processed cells while maintaining precision levels.
The possible biological applications of laser-based cell surgery devices are broad. High-throughput laser-based surgical methods, such as BLAST, are now able to deliver ultralarge cargoes into relatively cells in a minimally invasive manner that was not previously possible with other methods41,67. For example, the delivery of mitochondria for the study of diseases caused by mutated mitochondrial DNA, the delivery of whole chromosomes for cell engineering, and the delivery of intracellular pathogens for the study of pathogenesis all become possible. Due to the massively parallel and near-simultaneous nature of delivery achieved by methods such as BLAST67, a single chip can be used to conduct experiments and generate enough data for statistical analysis. Researchers will be able to observe large numbers of infected cells over time to examine phenomena such as bacterial localization and intracellular proliferation due to the ability to transfer bacteria into 100,000 host cells at a time. Such studies are practically difficult to perform using standard pipette-based delivery systems, as they do not provide the throughput required for accurate statistical analysis, and rapid events do not benefit from synchronization due to coinfection.
Micro/nanoneedle-based single-cell surgery methods
Intracellular surgery using microneedles dates back to the work of Barber et al., who injected living cells with substances such as bacteria by using a very tiny glass needle loaded with injection solutions68. Glass microneedles, which can be fabricated by pulling glass capillaries on heating filaments, have been used for single-cell manipulation and surgery in recent decades69. Given their low cost and easy fabrication, microneedles have been widely used in single-cell surgical methodologies. Microneedles attached to a micromanipulator have been used for several cell surgical tasks, such as cell biopsy70, cell injection71, cell cutting72, patch clamping73 and organelle removal and transfer74. Cell injection is the procedure of injecting external materials into a cell by a microneedle-based microinjection system75. A microneedle controlled by a micromanipulator perforates the cell membrane with its small and sharp tip and then injects materials into the cells, as illustrated in Fig. 3a. To restrict the transmission of inherited mtDNA diseases, the pronucleus of diseased women is injected by a glass microneedle to enucleate donor eggs, which may lead to offspring with healthy mtDNA76. Although this assisted reproduction strategy offers a healthy embryo, it cannot be utilized after birth or on somatic cells77. Another use of glass microneedles is patch clamping. Patch clamping is a method of studying the ionic currents and electrical behavior of individual isolated living cells and cell membranes. Micropipette tips coated with electrodes are placed on the cell surface, and a part of the cell membrane is aspirated to create a small opening for measuring the current levels.
To minimize the physical damage to cells due to penetration, several nanoneedle-based systems have been developed. Recently, a scanning ion conductance microscopy-based nanobiopsy system was developed to extract femtoliter samples of intracellular content by glass micropipette to analyze mRNA and mitochondrial DNA. This technique utilizes electrowetting to take up samples of cytoplasm into a glass nanopipette and then applies high-throughput sequencing technology to analyze the biopsied cellular material78, as shown in Fig. 3b. Similarly, researchers developed DEP-based nanotweezers79 made of two closely spaced electrodes with gaps as small as 10–20 nm to trap DNA and proteins in a dielectrophoretic manner, as shown in Fig. 3c.
Researchers have used this technique to extract nucleic acids from living cells for gene expression research without affecting cell survival, in addition to trapping single molecules. They also demonstrated the extraction of a single mitochondrion by using nanotweezers. To minimize cell surgery damage, researchers have developed a carbon nanotube-based endoscope that can interrogate cells, transfer fluids, and conduct optical and electrochemical diagnostics at the single-organelle level40 (Fig. 3d). The endoscope, which is created by inserting a multiwalled carbon nanotube (length, 50–60 µm) into the tip of a glass pipette, has a spatial resolution of ~100 nm and can reach organelles without disturbing the cell. When the nanotube is loaded with magnetic nanoparticles, the endoscope may be moved remotely to transfer nanoparticles and attoliter quantities of fluids to and from specific places. Given that the endoscopes are placed on ordinary glass micropipettes, they easily fit common equipment, opening up a wide variety of possibilities for minimally invasive intracellular probing, medication administration, and single-cell surgery.
The abovementioned micromanipulation procedures also necessitate skillful operation by a technician, which is time-consuming. The technician can inject only one cell at a time, restricting throughput. The development of technologies such as automation equipment and robotic systems in recent years has significantly increased the therapeutic efficiency of microinjection technology, although significant throughput and scalability need to be achieved (Fig. 3e, f)22. Automation of these complex single-cell surgical tasks is in high demand to reduce the probability of contamination resulting from human errors, labor-intensive work, process uncertainty, and variable outcomes. We developed several robot-aided single-cell surgery systems to automate organelle-level complex surgical procedures17,19,21,60,63,64,80,81. These automated systems were developed to reduce the human error involved in the precision surgery of single cells. Recently, we developed a robotic single-cell biopsy system for mitochondria and nuclei, as shown in Fig. 3g19. A microfluidic chip was used to trap single cells, and a glass micropipette was used to biopsy mitochondria from single cells in an automated way with high precision. Similarly, mitochondria extracted from a single cell can be further transferred to another cell using a microneedle-based mitochondrial transfer system, as shown in Fig. 3h80. Precise cell injections were performed with an initial 3D reconstruction of a cell to determine the specific target location, such as the nucleus inside the cell82.
Biomedical applications of microneedle-based single-cell surgery methods are limited by the throughput and physical damage to cells caused by microneedle penetration into the cell as the cell size decreases relative to the microneedle tip. For large cells, such as oocytes (100 µm diameter), a large number of mitochondria can be injected relatively easily by using a micropipette with a tip size of approximately 5–10 µm. However, for small cells, such as somatic cells ~20 µm in diameter, the size of the micropipette tip cannot be larger than 1 µm during mitochondrial injection to keep the recipient cell alive83. When injection pressure is applied, injected materials may become stuck at the tip and clog the micropipette tip, which can lead to low efficiency of microneedle-based mitochondrial injection. Therefore, the glass microneedle-based mitochondrial injection method has several limitations, including clogging of the glass micropipette tip84, physical damage to the cell, and limited opportunity for the repeated injection into the same cell.
To address the above problems, the nanothermal blade method was developed85, which uses a micropipette with a diameter of >3 µm to prevent clogging, as shown in Fig. 3i. This method uses pulsed laser-induced bubble cavitation to open holes in the cell membrane and then employs a synchronized fluid to pump mitochondria. Moreover, in this method, the injection efficiency of mitochondria is only 2–3% because the micropipette is only placed on the surface of the cell membrane and is not tightly wrapped inside the cell membrane. Therefore, during the injection process, mitochondria may be unable to enter the cell.
Integrating microneedles onto traditional patches has shown the ability to effectively deliver a range of drugs, including proteins, antibodies and vaccines86. Most microneedles have tips tens of microns in size and cannot accurately target individual cells, resulting in uneven and invasive delivery of plasmids or other macromolecules. To address this problem, nanoneedles with sharp smaller ends (<100 nm) were fabricated. They provide high precision and minimally invasive manipulation at the single-cell level. Kim et al. demonstrated the use of nanowire (NW) arrays to penetrate cells, allowing gene transfer into mammalian stem cells87. When NWs enter cells or cells are placed on NWs, molecules attached to the NW walls may dissociate and enter the cytoplasm (Fig. 3j)88. In this way, biomacromolecules can be transferred into cells without chemical modification or viral packaging. Shalek et al. used chemical vapor deposition techniques to create vertical NWs89. In Shalek’s study, HeLa cells with 1,1-dioctadecyl-3,3,3,3-tetramethylindodicarbocyanine (DID)-tagged membranes were placed on green fluorescently labeled NWs, and the permeation process was observed86. This technology has the potential to deliver siRNA, peptides, DNA, proteins and impermeable inhibitors into sensitive cell types such as neurons and immune cells. However, the NWs were unable to enter the cell immediately after contact90. Although the NWs developed above allow high-throughput cell injection, this approach may not enable precise dose injection. Furthermore, active injection or control of the penetration time of NWs still needs to be addressed, as direct penetration of cell membranes by mechanical structures often leads to permanent cell damage. This situation requires continuous innovation and improvement of micro/nanoneedle preparation methods and materials to provide high delivery efficiency while minimizing damage to cells.
Untethered microtools for single-cell surgery
On-chip microtools
Several microfluidic-based micromanipulation platforms for untethered control of surgery tools have been developed to perform contamination-free automatic single-cell surgery91,92,93,94,95,96. Animal cells were first softened by chemical treatment, and then two orthogonal channels controlled by an external magnetic field were used to cut the cell in half. The half of the cell containing the nucleus was used for cloning97. A microfluidic device integrated with thermopneumatic actuators was used for controlled cell lysis98. Upper channels were used for the introduction of cells, and lower lysed cells could be collected at lower channels. An oocyte was cut by a magnetically manipulated cutter to remove its nucleus, as shown in Fig. 4a99. On-chip nuclear removal (enucleation) was performed using a magnetic field-controlled microrobot to increase the speed of enucleation100 (Fig. 4b). Similarly, after grasping an oocyte with a microgripper, the cytoplasm of the cell was extruded with a micro knife, as shown in Fig. 4c101. Several other contamination-free, automated and high-throughput microfluidic-based enucleation procedures of oocytes were performed92,95. Although the above microrobotic-based platforms have been shown to be efficient surgical techniques, they have not been used for surgery of somatic cells because of the severe damage caused during cell dissection.
Untethered micro/nanotools
To perform untethered surgeries on single cells, several untethered micro/nanotools and systems have been established recently102. Bacteria-driven microswimmers have been created that possess active locomotive and bacterial sensing capabilities with the desirable enclosure and viscoelastic characteristics of a soft double-micelle microswimmer for active transport and transfer of cargo to living cells (e.g., genes)103 (Fig. 4d). This in vitro model demonstrates that soft microswimmers offer a promising opportunity for biomedical applications for active and/or targeted transport and delivery in in vitro models (e.g., organ-on-a-chip devices) and stagnating or low-flow liquid areas in the body. Researchers have also reported “dual-action microdaggers,” which are plant-derived biogenic micromotors that can produce a cellular incision followed by drug release, enabling extremely targeted drug delivery104 (Fig. 4e). The biogenic hybrid micromotor with “dual action” (e.g., cell microdrilling and drug release) enables noninvasive surgery with the single-cell targeting accuracy, as well as the additional benefit of drug release. Schmidt and colleagues created biohybrid sperm micromotors (Spermbots) as a targeted drug delivery system to treat female reproductive system disorders105 (Fig. 4f). 2PP 3D printing was utilized to create magnetic tubular microstructures for transporting motile sperm that serve as a propeller or transporter and anticancer medication carrier. Drugs such as doxorubicin (DOX) can be loaded onto sperm, giving good encapsulation, transport, and transfer stability. These spermbots can be magnetically directed to swim within the in vitro tumor model and release sperm to distribute DOX locally via fusion of cell and sperm membranes. Spermbots can also be potentially used in the human physiological environment because of their biohybrid design. Spermbots may have promising applications in gynecological treatments, for example, the treatment of ovarian cancer, cervical cancer, and other gynecological diseases, while avoiding harmful side effects of drugs on healthy tissues and/or organs. Researchers have developed an ultrasound-driven and near-infrared (NIR)-operated nanoswimmer that can puncture the membrane of a cancer cell106 (Fig. 4g). These nanoswimmers were able to move efficiently and controllably toward target cells under manipulation by an acoustic field. According to the experimental and theoretical findings, the immediate photothermal action generates sufficient photomechanical force to perforate the membrane of the cell. Such NIR-assisted nanoswimmer-enabled cell membrane poration has several advantages over conventional chemical and physical cell poration techniques, including active, fast, and precise targeting for single-cell surgery. Thus, this technique has tremendous potential for a number of biological applications, including gene transfer and fertilization. Artificial motorized sperm cells were created—a novel kind of hybrid micromotor in which customized microhelices serve as motors to help sperm cells with motion deficiencies in completing their usual function107 (Fig. 4h). This robot can collect, transport, and release single immotile living sperm cells in fluidic tubes. Although certain difficulties remain on the path to successful fertilization using artificially motorized sperm, the promise of this new method for assisted reproduction is already evident.
Although the above untethered micro/nanomanipulation platforms present several advantages, such as contamination-free and precise surgery without the use of expensive micromanipulators, the precision, efficiency, and throughput of these micromanipulation systems are low. Cell manipulation systems such as the pressure-driven injection system (Mitopunch) can produce hundreds of modified cells simultaneously108. By applying force on a mechanical plunger, thousands of cells can be injected with foreign substances at once, which can greatly improve the throughput of cell surgery.
Advances in medical robotics have the potential to improve current medicine and overall quality of life5. As these robotic platforms become more compact, they open the door to new applications of precision medicine. The unconstrained micro- and nanorobots used for precision medicine continue to face technical, regulatory, and commercial difficulties that must be overcome before they can be widely used in clinical settings. Nonetheless, recent translation from proof-of-concept to in vivo studies suggests their promise for precision medicine and personalized therapy109. The shrinking of robotic platforms provides the potential to improve patient care and diagnosis. These small robotic doctors may offer us access to hard-to-reach areas of the body, as well as the ability to perform a variety of medical treatments. Despite advances in medical micro/nanorobotics over the past decade, one of the unmet requirements and important hurdles in the field is the translation of these instruments into broad clinical applications, which is still a long way off.
Microrobots have a wide range of applications in precision medicine, including drug delivery, biologic delivery, gene delivery, and live cell delivery; surgical tools for biopsy, tissue penetration, intracellular delivery, or biofilm degradation; diagnostic tools such as physical and chemical biosensors or isolation tools; and optical, ultrasonic, magnetic and radionuclide imaging tools. Targeted delivery is the most established use, and current efforts are largely focused on animal testing. To understand the dynamics of micro/nanorobots and maximize their efficiency and capabilities, imaging must be used in conjunction with delivery, manipulation, or diagnostics. Current nano/microtool research shows that the gap between precision medicine and micro/nanorobotics has been narrowing. Nonetheless, each application presents unique barriers to clinical translation. Micro/nanorobots must operate in hard-to-reach areas of the body for delivery and surgical applications; therefore, recovery/degradation procedures are critical to ensure that they do not endanger the health of patients110. Furthermore, before micro/nanorobots can be used in therapeutic settings, they must overcome safety, technical, regulatory, financial and commercial hurdles. Although there is still a long way to go, the use of microrobots in precision medicine can definitely enhance diagnosis and treatment, leading to improved quality of life for patients. Microrobots may aid in precision medicine while also reducing the cost and pain of major surgical procedures.
Control methods
Cells and their intracellular components are small111,112 and fragile113,114, and they are distributed randomly in the medium112,113 with irregular shapes115,116. Operation failure and damage are likely to occur under inappropriate manipulation control. In addition to developing cell surgery tools, current studies face the following two challenges: how to achieve precise position localization of targeted intracellular specimens for successful cell surgery micromanipulation and how to regulate cell damage for a high survival rate. However, localizing operation positions can meet unignorable difficulties resulting from the complexity of intracellular environments and culture media, such as counterpart cells and auxiliary manipulation tools117,118,119. Random distribution of biological targets requires tedious height adjustment of optical lenses to ensure precise detection of targets120. Moreover, the imaging quality is deteriorated by imaging blur and noise during such optical observation, making the boundary of the targeted object unclear121. Therefore, visual servoing should be carried out based on the position feedback from advanced imaging processing. For the second challenge, cell morphology is usually adopted to estimate cell survivability, where dead cells often exhibit broken surfaces and lose adhesion to the substrate (for adherent cells). Fluorescence is another reflection of cell viability because cells undergoing damage exhibit low fluorescence intensity. For example, calcium flux78 and membrane potential probes122 are widely used for cell and mitochondrial function tests, respectively. However, these observation methods are qualitative and not accessible for cell damage regulation control. Force provides a quantitative parameter for cell viability measurement; cell damage from the interaction of manipulation tools can be monitored during micromanipulation; and the measured force signal can be adopted for the robotic manipulation control of such tools123. In practice, precise force sensing is difficult because of the small amplitude of the response signals of biological targets124,125,126 and high-level environmental noise7,127,128. With the help of state-of-the-art sensing instruments and postcomputation procedures, force-based control during manipulation is achievable. Thus, the success rates of cell surgery tasks can be improved, and operation damage can be reduced.
Visual servoing and controls
Successful robotic surgery manipulation control heavily relies on the localization of biological targets and manipulation tools. Targeted objects can be imaged and sampled with optical microscopes, such as wide-field fluorescence microscopes and confocal fluorescence microscopes. Out-of-focus blurs and imaging noise are inevitable in original sampled images and may produce inaccurate contours and localization of targeted specimens. Therefore, computational image processing methods are required to restore the morphology, position, orientation, and other operational details of targets.
Threshold binary is a widely adopted segmentation method to obtain distinct boundaries of targeted objects, where sampled features are binarized into bright spots or black backgrounds in terms of a threshold value. To smooth the sampled images, Gauss80, Median129, Wiener130, and Hilbert filters131 were adopted before the segmentation process. Thereafter, cell contours can be highlighted with edge detectors, such as the Canny edge detector80 and the Sobel operator132. Cells or other biological targets are normally covered by other devices and similar objects. Direct segmentation with a constant threshold value can hardly distinguish them from one another and may erase target boundaries near similar targets. Therefore, some adaptive thresholding processes were proposed that utilize identification factors to adjust the threshold value of each image. The hue, saturation and value (HSV) range value is proposed to divide the region of interest into four ranges, and the specific threshold value can be obtained by applying these values on the HSV plane image19, as shown in Fig. 5a. Figure 5b shows an Otsu thresholding process whose threshold value is derived from the area and roundness of the targets133. The segmented boundary can be used to locate the target object and facilitate manipulation, and the success rate of automated mitochondrial extraction and oocyte enucleation can reach 60%19 and 93.3%133. Notably, these segmentation approaches can extract features only at a fixed imaging depth, and the targeted specimen is assumed to be a regular sphere, whose center is usually localized as an operating position. For targets with irregular morphology, the appropriate operation position should be selected through spatial information134.
Recently, some template-based depth acquisition methods have been developed, where a focal image of the biological targets or surgery tools is sampled and set as a template for comparison with sampled images at different optical depths. The focal imaging depth can be estimated when the pixel differences between the sampled images are minimized135,136. However, external disturbances, noises, and imaging errors can produce inaccurate depth when the imaging field moves greatly. To alleviate this problem, some optimized template-matching criteria have been proposed based on template similarity score137,138, three frames18, and centroid coordinates139. With the depth information obtained, the microinjection efficiency is greatly improved, with experimental results showing that more than 1500 cells can be processed automatically within 1 hour18. However, template selection is manual and labor intensive, and a new template is required when targets change. Imaging practices have shown that the light intensity of sampled targets follows an organized distribution model. Many attempts called depth from focus (DFF) methods140,141 have been applied to locate the focal depth where a rapid change in light intensity occurs, as shown in Fig. 5c. Depth from defocus (DFD) methods142,143 are efficient because they can estimate current depth by comparing neighboring defocused images, as shown in Fig. 5d. However, DFF and DFD estimations suffer from low depth resolution and become inaccurate if the target’s height is smaller than the sampling depth intervals.
For small biological objects with irregular morphology and random distribution, 3D image reconstruction of biological targets can be carried out to provide a comprehensive geometric analysis. Most out-of-focus emissions can be physically filtered by a pinhole aperture in a confocal microscope; the 3D morphology of endothelial cells144 and bladder cancer cell nuclei145 have been reconstructed for applications of cell microinjection and biophysics measurement, as shown in Fig. 5e. Wide-field microscopes are extensively used. The in-plane features of sampled images contain out-of-focus blurs projected from adjacent depths and noise. Noise-regulated maximum likelihood estimation deconvolution was proposed to eliminate such deteriorations82. After filtered segmentation, the deconvoluted images of the neonatal human dermal fibroblast (HDFn) nucleus can be reconstructed into a 3D model, and an optimal position can be obtained under volume-based geometric analysis to improve nuclear delivery accuracy and efficiency, as shown in Fig. 5f17. The 3D reconstruction information can greatly improve the operation accuracy while reducing the damage to biological objects. Specifically, the success rate of mitochondrial extraction experiments is 20% higher than that of traditional 2D methods, preventing the cell destruction that usually occurs in traditional methods. The 3D reconstruction information was also applied to automated intracellular delivery. The success rate of nucleus delivery increased to 71%, well above the value of 48% obtained in 2D experiments. More than 85% of processed cells showed strong biological activity when incubated 24 h after delivery. However, the computational complexity and image sampling of 3D reconstruction require a relatively long processing duration, and real-time reconstruction is difficult to realize. Thus, a tradeoff exists between control accuracy and manipulation throughput.
With such visual localization feedback, the robotic manipulation controller manipulates surgery tools or biological objects to minimize the error between their origin and destination positions is minimized. Differentiation terms of position error derived from the velocity of the manipulated objects are essential for motion control. However, direct measurement of such velocity not only requires complicated hardware schemes but also suffers from system disturbances and external disturbances. Some velocity observers and estimators were proposed to compensate for the error caused by inaccurate information of the camera position. An uncertainty and disturbance estimation-based observer was proposed to estimate cell angular velocity for cell orientation control146. A model-free disturbance observer147 and high-gain observer148 were developed to estimate the velocity and state errors for the control of microrobots in blood vessels. With such observers, a simple saturated proportional–integral–derivative (PID) controller was proposed for the asymptotic regulation of cell rotation149. A feedforward plus PD feedback controller was proposed for high-throughput cell microinjection132. In addition, several sliding mode controllers were developed to manipulate microneedles for intracellular delivery17 and biopsy150. By using such advanced motion controllers with precise visual feedback, the operation accuracy and success rate can be improved.
Force sensing and controls
Current surgical tools manipulate cells and other biological objects by either physical interaction or field-driven forces; the interaction forces between them unavoidably do great harm to the biological viability of the cells if a certain value is exceeded. Therefore, interaction force control is significant for damage regulation in robotic cell surgery manipulation. Nevertheless, the amplitude of the force signal during manipulation is small (less than a few hundreds of micronewtons), and the signal accompanied by strong background noise from the complicated environment of living cells, which puts force sensing techniques with high sensitivity and signal-to-noise ratio (SNR) into urgent demand.
AFM has a very high force-sensing resolution at the piconewton level. The forces between the probing tip and a sample surface cause cantilever deflection, which is detected by an optical system composed of a laser and a photodetector. Customized magnetic microgrippers151, cantilevered micropipette probes152, and nanofluidic channel probes152 have been integrated into AFM systems for applications in microassembly, microinjection, and intracellular delivery, as shown in Fig. 5g. In addition to the labor-intensive optical alignment and adjustment, signals from AFM are measured by the outer reflection of the laser beam, which can result in inaccurate force measurements when reflected light transmits through the culture medium153,154. MEMS sensors, which take advantage of task-oriented design, high sensing performance, and easy mass production, have been widely used for force sensing in robotic cell surgery manipulation. On the basis of the resistance change resulting from exerted forces, the piezoresistive force sensor normally produces stable force signals with relatively low resolution at the micronewton level. A soft flexure beam cell microinjector embedded with a piezoresistive force sensor has been developed, which has a force resolution of hundreds of micronewtons155. Piezoresistive probes with a resolution of tens of nano-Newtons have been proposed, which typically have flexible beams156 or springs157 with widths less than 5 µm and lengths greater than 1000 µm to achieve high stress concentrations, but the cost is significantly difficult to fabricate, as shown in Fig. 5h. MEMS capacitive force sensors can measure the forces when external forces produce transverse or longitudinal movement of parallel electrodes. With comb-structured capacitive electrode plates, they can incorporate multiple sensing arrays within one unit, thereby achieving high sensing sensitivity. Although a narrow air gap between the comb plates is preferable for higher sensitivity, capacitive sensors are much easier to fabricate than piezoresistive sensors because the comb plates are typically less than a few hundred micrometers long. Several capacitive force sensors using differential comb capacitors have been proposed and can measure nanonewton-level forces158,159,160,161, as shown in Fig. 5i. However, the fabrication consistency deteriorates as the number of combs increases. Furthermore, capacitive sensors are susceptible to high cross-coupling and low linearity owing to unavoidable parasitic capacitances162,163; thus, a complicated readout circuit is needed. Piezoresistive and capacitive force sensors are mainly integrated with surgery tools for measuring the forces exerted by such tools. Piezoelectrical sensors can be mounted at the targeted cell side, and loaded forces can compress or stretch the film and produce corresponding electrical voltage. Polyvinylidene fluoride (PVDF) film can be adhesively bonded to a cell bracket for cell indentation force measurement164,165, as shown in Fig. 5j. Considering the macroscopic film size, PVDF piezoelectric sensors are easy to fabricate, while their force resolution is relatively low at the micronewton level166,167. In addition, they are not applicable for force measurement in intracellular manipulation because they can hardly measure forces inside the cytomembrane; thus, their spatial resolution is greatly limited. The optical sensor has large sensing ranges for observing the displacement of the sensing apparatus with known stiffness, and its resolution is comparable with the pixel size of the imaging microscope. Micromachined springs168, flexible pillars169, microrobots170, and other types of optical sensors have been developed to measure intracellular forces under microscopic fields, and they can be easily integrated into untethered microtools, as shown in Fig. 5k. However, optical observation suffers from imaging blurs and noises, which can lead to difficulty in distinguishing the sensing structure from backgrounds, especially for cases that are not at the focal plane171. To improve measurement accuracy, optical sensors usually use flexible materials or low-rigidity structures. In addition to increased fabrication difficulty, geometric deviations can create stiffness inconsistencies between fabricated sensors, so optical sensors require careful performance calibration172. In general, force sensors should be selected in terms of micromanipulation tasks and their own characteristics. Sensing resolution and signal conditioning should be evaluated first, especially for intracellular organelle micromanipulation tasks that require sensing resolution higher than the nanonewton level. Furthermore, the convenience of integration with manipulation tools should be considered.
Force feedback control can be simultaneously initiated once force signals are obtained from appropriate sensors. Direct force control and indirect force control are the two main categories of manipulation force control. For the indirect type, the forces exerted on cells are typically converted to the position of manipulation tools depending on optical observations, guaranteeing a smooth switch. An impedance controller173,174 and a PID controller with optimal control functions175 have been proposed for maintaining the desired force and displacement amplitude. A two-loop control framework controller improves the force regulation performance, where a force tracking nonlinear controller with measured feedback from PVDF piezoelectrical sensors is an external loop for the internal impedance control loop176. The penetration force of zebrafish embryos can be adjusted to 216 µN, and the relative root mean square error (RMSE) of force tracking is approximately 0.37. Given that the force is estimated from optical observations, indirect force control suffers from inaccurate control parameters. It may be unreliable for complicated manipulation tasks such that target deformation is undetectable. Direct force controllers move manipulation tools with straightforward force feedback from integrated sensors once they arrive at the target position under position control. Some PID/PD controllers based on incremental function177 and model-compensated prediction schemes178,179 have been proposed to regulate cell loads, where the maximum force tracking error is limited to 90 µN177 and the regulation resolution reaches 50 µN178,179. The force controller takes charge when the measured force exceeds a threshold value; thus, a stable switch between position and force control is strongly needed. A weight-based fusion approach with introduced weight coefficients in the control variable has been proposed to fuse the two position control variables and force in an overlap interval of embryo microinjection manipulation180. A revised noise-insensitive extended high gain-observer181 and an event-based switch criterion182 have been proposed to minimalize the switch overshoot for contact force regulation. Compared with the traditional PID controller, the force overshoot is greatly improved by 49%179, and the force tracking RMSE is reduced by 34.2%. From the aforementioned illustration, it is clear that force regulation controllers can greatly decrease the cell deformation resulting from tool interaction; cell breakage, bursting, and other major damage can be prevented as well.
Discussion and conclusion
Micro/nanoscale biological cell operations, such as single-cell surgery, have recently become important due to their key biological applications in precision medicine. Single-cell surgical approaches, such as cell injection, cell biopsy, and the extraction and transfer of cell organelles from a single cell, have key biological applications, such as studying diseases and their causes, in greater depth. The development of tools and methods for single-cell surgery can play an important role in the treatment of diseases, such as aging and neurodegenerative disorders, epilepsy, and type 2 diabetes. In this article, the tools and systems for single-cell organelle surgery are summarized in detail. The current challenges and drawbacks of existing methodologies are also highlighted.
As mentioned above, several single-cell surgery systems, such as single-cell injection and organelle manipulation systems, have been developed in recent years. Single-cell organelle-level surgery or manipulation is more challenging than cell injection for the following reasons. First, in the case of cell injection, a precise and specific location is generally not mandatory inside the cell; in cell organelle manipulation or surgery, the tooltip must be positioned at a precise location of the organelle in the 3D space of the cell. Second, unlike organelle biopsy, cell fluidic injection can be completed with a tip size of 0.1–1 µm, which is an acceptable range to keep the cell alive after injection. However, organelle biopsy or transfer requires a relatively large tooltip size depending on the organelle size, which can easily damage the cell. Third, in organelle manipulation, the cell-holding power must be readjusted during tool withdrawal. Therefore, introduction of the tooltip at a precise position inside the cell, use of a relatively larger tip, and suitable control of the cell during organelle biopsy or removal cause single-cell organelle manipulation or removal to be more complex than cell injection.
Typically, enucleation methods are performed on cells 50–1000 µm in diameter. The removal and transfer of subcellular components from small biological cells (<25 µm in diameter), such as human mesenchymal stem cells (MSCs), are more complicated than those from large biological cells, such as zygotes. The core challenges in small-cell surgery are precise micropipette or cell positioning and the prevention of irreversible physical damage to these sensitive cells during surgery.
Notably, the direct penetration of cell membranes through mechanical structures often results in irreversible cellular damage. Solving this problem requires continuing innovation and the development of micro/nanoneedle fabrication techniques and materials to achieve high delivery efficiency while preventing cellular damage. Furthermore, current single-cell surgery or modification methods are primarily limited by throughput and efficiency. To further expand the use of modified cells for in vivo or clinical applications, which require millions of engineered cells, the throughput of single-cell surgical approaches needs to be significantly increased. Less invasive micromanipulation systems, such as pressure-driven BLAST (Fig. 2h)67 and Mitopunch108, can generate hundreds of modified cells simultaneously.
With the rapid development of precision engineering and medical technology, the use of engineering methods for surgical intervention on single biological cells is an emerging technology in the field of medical applications. This development is in line with the trend of expanding modern therapies from the organ level to the cellular level to prolong patient survival. Although existing cell modification engineering methods exhibit many advantages, such as high precision and selectivity in generating functionally altered cells, these methods have relatively low throughput compared to traditional chemical or biological methods in cell therapies. There is a high demand for the development of automated table-top microfabrication systems capable of efficiently and rapidly processing tiny biological samples. Automated micromanipulation systems can incorporate manipulators, such as vision-guided and robot-driven micropipettes and OTs, and microfluidic chip devices for automated cell processing. Dynamic image processing and robust control techniques can be used to realize the large-scale cell modification process required for clinical therapy. The success of these systems will validate the feasibility of the large-scale production of functionally altered cells by integrating robotics and fabrication techniques into cell manipulation. The successful production of therapeutic quantities of characteristically modified cells will have broad and diverse applications in biopharmaceuticals, gene and cell therapy, and tissue engineering.
Table 1 summarizes and compares different technologies for single-cell surgery.
References
Ramachandran P. et al. Resolving the fibrotic niche of human liver cirrhosis at single-cell level. Vol 575. Springer US (2019).
Yuan, G. C. et al. Challenges and emerging directions in single-cell analysis. Genome Biol. 18, 1–8 (2017).
Füger, P. et al. Microglia turnover with aging and in an Alzheimer’s model via long-term in vivo single-cell imaging. Nat. Neurosci. 20, 1371–1376 (2017).
Pan, X. W. U. et al. Identification of a potential mechanism of acute kidney injury during the COVID-19 outbreak: a study based on single-cell transcriptome analysis. Intensive Care Med. 46, 1114–1116 (2020).
Nelson, B. J., Kaliakatsos, I. K. & Abbott, J. J. Microrobots for minimally invasive medicine. Annu. Rev. Biomed. Eng. 12, 55–85 (2010).
Patra, D. et al. Intelligent, self-powered, drug delivery systems. Nanoscale 5, 1273–1283 (2013).
Liu, J., Wen, J., Zhang, Z., Liu, H. & Sun, Y. Voyage inside the cell: Microsystems and nanoengineering for intracellular measurement and manipulation. Microsyst. Nanoeng. 1, 1–15 (2015).
Pertoft H., Laurent T. C. Isopycnic separation of cells and cell organelles by centrifugation in modified colloidal silica gradients. Methods Cell Sep. 25–65 (1977).
Schubert, C. Single-cell analysis: The deepest differences. Nature 480, 133–137 (2011).
Farbehi, N. et al. Single-cell expression profiling reveals dynamic flux of cardiac stromal, vascular and immune cells in health and injury. Elife 8, 1–39 (2019).
Chung, W. et al. Single-cell RNA-seq enables comprehensive tumour and immune cell profiling in primary breast cancer. Nat. Commun. 8, 1–12 (2017).
Ji, Q. et al. Single-cell RNA-seq analysis reveals the progression of human osteoarthritis. Ann. Rheum. Dis. 78, 100–110 (2019).
Han, Y. et al. The logic of single-cell projections from visual cortex. Nature 556, 51–56 (2018).
Sermon, K. Preimplantation genetic diagnosis. J. Bras. Reprod. Assist 8, 7–20 (2004).
Li, H. et al. Generation of biallelic knock-out sheep via gene-editing and somatic cell nuclear transfer. Sci. Rep. 6, 1–12 (2016).
Patananan, A. N., Wu, T. H., Chiou, P. Y. & Teitell, M. A. Modifying the mitochondrial genome. Cell Metab. 23, 785–796 (2016).
Gao, W. et al. Precise automated intracellular delivery using a robotic cell microscope system with three-dimensional image reconstruction information. IEEE/ASME Trans. Mechatron. 25, 2870–2881 (2020).
Pan, F. et al. Automated high-productivity microinjection system for adherent cells. IEEE Robot Autom. Lett. 5, 1167–1174 (2020).
Shakoor, A. et al. Achieving automated organelle biopsy on small single cells using a cell surgery robotic system. IEEE Trans. Biomed. Eng. 66, 2210–2222 (2019).
Kim, S. H. & Ishiyama, K. Magnetic robot and manipulation for active-locomotion with targeted drug release. IEEE/ASME Trans. Mechatron. 19, 1651–1659 (2014).
Xie, M., Shakoor, A., Shen, Y., Mills, J. K. & Sun, D. Out-of-Plane rotation control of biological cells with a robot-tweezers manipulation system for orientation-based cell surgery. IEEE Trans. Biomed. Eng. 66, 199–207 (2019).
Zhukovskiy, Y. & Koteleva, N. Development of augmented reality system for servicing electromechanical equipment. J. Phys. Conf. Ser. 1015, 358–365 (2018).
Kremser, L., Blaas, D. & Kenndler, E. Capillary electrophoresis of biological particles: Viruses, bacteria, and eukaryotic cells. Electrophoresis 25, 2282–2291 (2004).
Cheah, C. C., Li, X., Yan, X. & Sun, D. Observer-based optical manipulation of biological cells with robotic tweezers. IEEE Trans. Robot 30, 68–80 (2014).
Kishino, A. & Yanagida, T. Force measurements by micromanipulation of a single actin filament by glass needles. Nature 334, 74–76 (1988).
Zhang, H. & Liu, K. K. Optical tweezers for single cells. J. R. Soc. Interface 5, 671–690 (2008).
Manz, A., Graber, N. & Widmer, H. M. Miniaturized total chemical analysis systems: A novel concept for chemical sensing. Sens. Actuators B Chem. 1, 244–248 (1990).
Pierini, F., Zembrzycki, K., Nakielski, P., Pawłowska, S. & Kowalewski, T. A. Atomic force microscopy combined with optical tweezers (AFM/OT). Meas. Sci. Technol. 27, 025904 (2016).
Ramadan, A., Inoue, K., Arai, T. & Takubo, T. New architecture of a hybrid two-fingered micro-nano manipulator hand: Optimization and design. Adv. Robot 22, 235–260 (2008).
Yamanishi Y., Kuriki H., Sakuma S., Onda K., Arai F. Local ablation by micro-electric knife. 2012 7th IEEE Int Conf Nano/Micro Eng Mol Syst NEMS 2012., 129–132 (2012).
Chung, K. & Lu, H. Automated high-throughput cell microsurgery on-chip. Lab Chip 9, 2764–2766 (2009).
Actis, P. et al. Electrochemical nanoprobes for single-cell analysis. ACS Nano 8, 875–884 (2014).
Nikitin, V. A. Microtools for Microsurgery of a Single Cell in Field of Cellular Engineering (IntechOpen, 2012).
Song, B. et al. Cellular-level surgery using nano robots. J. Lab Autom. 17, 425–434 (2012).
Satava, R. M. How the future of surgery is changing: Robotics, telesurgery, surgical simulators and other advanced technologies. Jurnalul Chir. 5, 1–21 (2006).
van Bergeijk, P., Hoogenraad, C. C. & Kapitein, L. C. Right time, right place: probing the functions of organelle positioning. Trends Cell Biol. 26, 121–134 (2016).
Jeong, H., Li, T., Gianchandani, Y. B. & Park, J. High precision cell slicing by harmonically actuated ultra-sharp SixNy blades. J. Micromech. Microeng. 25, 025007 (2015).
Yoshioka M., Miyazako H., Wagastuma A., Mabuchi K., Hoshino T. Single-cell membrane surgery by virtual electrodes using an inverted electron beam lithography system. 2014 IEEE/SICE Int Symp Syst Integr SII 2014. Published online, 68–71 (2014).
Osada, T., Uehara, H., Kim, H. & Ikai, A. mRNA analysis of single living cells. J. Nanobiotechnol. 1, 2 (2003).
Singhal, R. et al. Multifunctional carbon-nanotube cellular endoscopes. Nat. Nanotechnol. 6, 57–64 (2011).
Stevenson, D. J., Gunn-Moore, F. J., Campbell, P. & Dholakia, K. Single cell optical transfection. J. R. Soc. Interface 7, 863–871 (2010).
Horn, P. J. et al. Visualization of lipid droplet composition by direct organelle mass spectrometry. J. Biol. Chem. 286, 3298–3306 (2011).
Pan, N. et al. The Single-probe: A miniaturized multifunctional device for single cell mass spectrometry analysis. Anal. Chem. 86, 9376–9380 (2014).
Takeuchi, M., Nakajima, M. & Fukuda, T. In 2009 IEEE International Conference on Robotics and Automation. Published online, 1838–1843 (2009).
Azizian M., Patel R., Gavrilovici C., Poulter M. Computer-assisted patch clamping. Proc - IEEE Int Conf Robot Autom. Published online, 4131–4136 (2010).
Maragò, O. M., Jones, P. H., Gucciardi, P. G., Volpe, G. & Ferrari, A. C. Optical trapping and manipulation of nanostructures. Nat. Nanotechnol. 8, 807–819 (2013).
Uchugonova, A. Nanosurgery of cells and chromosomes using near-infrared twelve-femtosecond laser pulses. J. Biomed. Opt. 17, 101502 (2012).
Sumer, B. D., Goyal, V., Truelson, J. M. & Myers, L. L. Transoral robotic surgery and transoral laser microsurgery for oropharyngeal squamous cell cancer. J. Robot Surg. 7, 377–383 (2013).
Palanker, D. et al. Technique for cellular microsurgery using the 193???nm excimer laser. Lasers Surg. Med. 11, 580–586 (1991).
Kohli, V., Elezzabi, A. Y. & Acker, J. P. Cell nanosurgery using ultrashort (femtosecond) laser pulses: Applications to membrane surgery and cell isolation. Lasers Surg. Med. 37, 227–230 (2005).
Shen, N. et al. Ablation of cytoskeletal filaments and mitochondria in live cells using a femtosecond laser nanoscissor. MCB Mech. Chem. Biosyst. 2, 17–25 (2005).
Chen S., et al. Laser-induced fusion of human embryonic stem cells with optical tweezers. Appl Phys Lett. 103 (2013).
Uchugonova, A., König, K., Bueckle, R., Isemann, A. & Tempea, G. Targeted transfection of stem cells with sub-20 femtosecond laser pulses. Opt. Express 16, 9357 (2008).
Kohli, V. & Elezzabi, A. Y. Laser surgery of zebrafish (Danio rerio) embryos using femtosecond laser pulses: Optimal parameters for exogenous material delivery, and the laser’s effect on short- and long-term development. BMC Biotechnol. 8, 1–20 (2008).
Uchugonova, A., Breunig, H. G., Batista, A. & König, K. Optical reprogramming of human cells in an ultrashort femtosecond laser microfluidic transfection platform. J. Biophotonics 9, 942–947 (2016).
Schomaker, M. et al. Characterization of nanoparticle mediated laser transfection by femtosecond laser pulses for applications in molecular medicine. J. Nanobiotechnol. 13, 1–15. (2015).
Il’ina, I. V. et al. Noncontact microsurgery and micromanipulation of living cells with combined system femtosecond laser scalpel-optical tweezers. Biophotonics Photonic Solut. Better Heal Care III 8427, 84270S (2012).
Lukianova-Hleb, E. Y., Mutonga, M. B. G. & Lapotko, D. O. Cell-specific multifunctional processing of heterogeneous cell systems in a single laser pulse treatment. ACS Nano 6, 10973–10981 (2012).
Shakoor, A. et al. Automated optical tweezers manipulation to transfer mitochondria from fetal to adult mscs to improve antiaging gene expressions. Small 2103086, 1–17 (2021).
Xie, M., Shakoor, A. & Wu, C. Manipulation of biological cells using a robot-aided optical tweezers system. Micromachines 9, 1–13 (2018).
Ashkin, A., Dziedzic, J. M. & Yamane, T. Optical trapping and manipulation of single cells using infrared laser beams. Nature 330, 769–771 (1987).
Ashkin, A. & Dziedzic, J. M. Optical trapping and manipulation of viruses and bacteria. Science (80-) 235, 1517–1520 (1987).
Xie M., Shakoor A. Finite-time control for biological cells manipulation with robot-tweezers system: Theory and experiments. IEEE/ASME Int Conf Adv Intell Mechatronics, AIM. 2019-July, 1646–1651 (2019).
Xie, M., Shakoor, A., Wu, Z. & Jiang, B. Optical manipulation of biological cells with a Robot-Tweezers system: a stochastic control approach. IEEE Trans. Circuits Syst. II Express Briefs 67, 3232–3236 (2020).
Gou, X., Yang, H., Fahmy, T. M., Wang, Y. & Sun, D. Direct measurement of cell protrusion force utilizing a robot-aided cell manipulation system with optical tweezers for cell migration control. Int J. Rob. Res. 33, 1782–1792 (2014).
Ando, J., Bautista, G., Smith, N., Fujita, K. & Daria, V. R. Optical trapping and surgery of living yeast cells using a single laser. Rev. Sci. Instrum. 79, 2–6 (2008).
Wu, Y. C. et al. Massively parallel delivery of large cargo into mammalian cells with light pulses. Nat. Methods 12, 439–444 (2015).
Barber, M. A. A technic for the inoculation of bacteria and other substances into living cells. J. Infect. Dis. 8, 348–360 (1911).
Chen, B. K. et al. Nano-dissection and sequencing of DNA at single sub-nuclear structures. Small 10, 3267–3274 (2014).
Actis, P., Kan, S. Y., Penfold, C., Vilozny, B. & Pourmand, N. Single-Cell biopsy using nanopipettes. Biophys. J. 102, 188a (2012).
Xu, Q. Review of Microinjection Systems (Springer International Publishing, 2018).
Yang, R. et al. Cellular level robotic surgery: Nanodissection of intermediate filaments in live keratinocytes. Nanomed. Nanotechnol., Biol. Med. 11, 137–145 (2015).
Aerts, J. T. et al. Patch clamp electrophysiology and capillary electrophoresis-mass spectrometry metabolomics for single cell characterization. Anal. Chem. 86, 3203–3208 (2014).
Yu J., Zhao Q., Cui M., Sun M., Zhao X. Robotic donor cell injection in Somatic Cell Nuclear Transfer (SCNT). Proc World Congr Intell Control Autom. 2015-March, 2821–2825 (2015).
Sun, Y. & Nelson, B. J. Biological cell injection using an autonomous microrobotic system. Int J. Rob. Res. 21, 861–868 (2002).
Richardson, J. et al. Concise reviews: assisted reproductive technologies to prevent transmission of mitochondrial DNA disease. Stem Cells 33, 639–645 (2015).
Vogel, G. FDA considers trials of “three-parent embryos. Science (80-) 343, 827–828 (2014).
Actis, P. et al. Compartmental genomics in living cells revealed by single-cell nanobiopsy. ACS Nano 8, 546–553 (2014).
Nadappuram, B. P. et al. Nanoscale tweezers for single-cell biopsies. Nat. Nanotechnol. 14, 80–88 (2019).
Shakoor, A. et al. In 2019 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) 659–664 (IEEE, 2019).
Shakoor A., et al. A high-precision robot-aided single-cell biopsy system. 8, 5397–5402 (2017).
Gao, W., Shakoor, A., Zhao, L., Jiang, Z. & Sun, D. 3-D image reconstruction of biological organelles with a robot-aided microscopy system for intracellular surgery. IEEE Robot Autom. Lett. 4, 231–238 (2019).
Chow, Y. T. et al. Single cell transfection through precise microinjection with quantitatively controlled injection volumes. Sci. Rep. 6, 24127 (2016).
King, M. P. & Attardi, G. Injection of mitochondria into human cells leads to a rapid replacement of the endogenous mitochondrial DNA. Cell 52, 811–819 (1988).
Wu, T. H. H. et al. Mitochondrial transfer by photothermal nanoblade restores metabolite profile in mammalian cells. Cell Metab. 23, 921–929 (2016).
Chang, H. et al. Cryomicroneedles for transdermal cell delivery. Nature Biomedical Engineering 5, 1008–1018 (2021).
Kim, W., Ng, J. K., Kunitake, M. E., Conklin, B. R. & Yang, P. Interfacing silicon nanowires with mammalian cells. J. Am. Chem. Soc. 129, 7228–7229 (2007).
Qi, S., Yi, C., Ji, S., Fong, C. C. & Yang, M. Cell adhesion and spreading behavior on vertically aligned silicon nanowire arrays. ACS Appl Mater. Interfaces 1, 30–34 (2009).
Shalek, A. K. et al. Vertical silicon nanowires as a universal platform for delivering biomolecules into living cells. Proc. Natl Acad. Sci. USA 107, 1870–1875 (2010).
Xu, A. M. et al. Quantification of nanowire penetration into living cells. Nat. Commun. 5, 3613 (2014).
Huang, S. H., Hung, L. Y. & Lee, G. B. Continuous nucleus extraction by optically-induced cell lysis on a batch-type microfluidic platform. Lab. Chip 16, 1447–1456 (2016).
Hagiwara M., Ichikawa A., Kawahara T., Arai F. High speed enucleation of oocyte using magnetically actuated microrobot on a chip. 2012 7th IEEE Int Conf Nano/Micro Eng Mol Syst NEMS 2012. Published online, 364–367 (2012).
Ichikawa, A., Sakuma, S., Shoda, T., Arai, F. & Akagi, S. On-chip enucleation of oocyte using untetherd micro-robot with gripping mechanism. 17th Int Conf. Miniaturized Syst. Chem. Life Sci. MicroTAS 2013 3, 1635–1637 (2013).
Park I. S., Lee Y. R., Hong S. J., Lee K. Y., Chung S. K. On-chip enucleation using an untethered microrobot incorporated with an acoustically oscillating bubble. Proc IEEE Int Conf Micro Electro Mech Syst. 2015(February), 1059–1062 (2015).
Sakuam S., Arai F. ON-chip oocyte enucleation by robot integrated microfluidic chip. Proc IEEE Int Conf Micro Electro Mech Syst. Published online, 225–228 (2013).
Campuzano, S., Esteban-Fernández De Ávila, B., Yáñez-Sedeño, P., Pingarrón, J. M. & Wang, J. Nano/microvehicles for efficient delivery and (bio)sensing at the cellular level. Chem. Sci. 8, 6750–6763 (2017).
Ichikawa, A., Tanikawa, T., Matsukawa, K., Takahashi, S. & Ohba, K. Automated Cell-Cutting for Cell Cloning. SICE J. Control Meas. Syst. Integr. 3, 75–80 (2010).
Irimia, D., Tompkins, R. G. & Toner, M. Single-cell chemical lysis in picoliter-scale closed volumes using a microfabricated device. Anal. Chem. 76, 6137–6143 (2004).
Hagiwara, M., Kawahara, T., Feng, L., Yamanish, Y. & Arai, F. High precision magnetically driven microtools with ultrasonic vibration for enucleation of oocytes. 2010 Int. Symp. Micro-NanoMechatronics Hum. Sci. From Micro Nano Scale Syst. to Robot. Mechatronics Syst. MHS 2010, Micro-Nano GCOE 2010, Bio-Manipulation 2010, 47–52 (2010).
Inomata N., Mizunuma T., Yamanishi Y., Kudo S., Arai F. On-chip magnetically driven micro-robot for enucleation of oocyte. 20th Anniv MHS 2009 Micro-Nano Glob COE - 2009 Int Symp Micro-NanoMechatronics Hum Sci. Published online, 493–498 (2009).
Ichikawa, A. et al. On-chip enucleation of an oocyte by untethered microrobots. J. Micromech. Microeng. 24, 095004 (2014).
Farooq, U. et al. Cell lysis induced by nanowire collision based on acoustic streaming using surface acoustic waves. Sens. Actuators B Chem. 345, 130335 (2021).
Singh, A. V., Hosseinidoust, Z., Park, B. W., Yasa, O. & Sitti, M. Microemulsion-based soft bacteria-driven microswimmers for active cargo delivery. ACS Nano 11, 9759–9769 (2017).
Srivastava, S. K., Medina-Sánchez, M., Koch, B. & Schmidt, O. G. Medibots: Dual-Action biogenic microdaggers for single-cell surgery and drug release. Adv. Mater. 28, 832–837 (2016).
Xu, H. et al. Sperm-Hybrid micromotor for targeted drug delivery. ACS Nano 12, 327–337 (2018).
Wang, W., Wu, Z., Lin, X., Si, T. & He, Q. Gold-Nanoshell-Functionalized polymer nanoswimmer for photomechanical poration of single-cell membrane. J. Am. Chem. Soc. 141, 6601–6608 (2019).
Medina-Sánchez, M., Schwarz, L., Meyer, A. K., Hebenstreit, F. & Schmidt, O. G. Cellular cargo delivery: toward assisted fertilization by sperm-carrying micromotors. Nano Lett. 16, 555–561 (2016).
Patananan, A. N. et al. Pressure-Driven mitochondrial transfer pipeline generates mammalian cells of desired genetic combinations and fates. Cell Rep. 33, 108562 (2020).
Soto, F., Wang, J., Ahmed, R. & Demirci, U. Medical micro/nanorobots in precision medicine. Adv. Sci. 7, 1–34 (2020).
Wei, T. et al. Development of Magnet‐Driven and Image‐Guided Degradable Microrobots for the Precise Delivery of Engineered Stem Cells for Cancer Therapy. Small 16, 1906908 (2020).
Saha R., Bajger M., Lee G. Segmentation of cervical nuclei using SLIC and pairwise regional contrast. In: Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS. 3422–3425 (2018).
Singh, L., Nag, T. C. & Kashyap, S. Ultrastructural changes of mitochondria in human retinoblastoma: correlation with tumor differentiation and invasiveness. Tumor Biol. 37, 5797–5803 (2016).
Wang W., Hou J., Zhu Z., Fang H. Is mitochondrial cell fragility a cell weakness? In: Mitochondrial DNA and Diseases. Springer Singapore 107–116 (2017).
Pfeifer, C. R., Vashisth, M., Xia, Y. & Discher, D. E. Nuclear failure, DNA damage, and cell cycle disruption after migration through small pores: A brief review. Essays Biochem. 63, 569–577 (2019).
Uversky, V. N. Intrinsically disordered proteins in overcrowded milieu: Membrane-less organelles, phase separation, and intrinsic disorder. Curr. Opin. Struct. Biol. 44, 18–30 (2017).
Fermie, J. et al. Single organelle dynamics linked to 3D structure by correlative live-cell imaging and 3D electron microscopy. Traffic 19, 354–369 (2018).
Kuznetsov, A. V. & Margreiter, R. Heterogeneity of mitochondria and mitochondrial function within cells as another level of mitochondrial complexity. Int J. Mol. Sci. 10, 1911–1929 (2009).
Kang, N., Guo, Q., Islamzada, E., Ma, H. & Scott, M. D. Microfluidic determination of lymphocyte vascular deformability: Effects of intracellular complexity and early immune activation. Integr. Biol. (U. Kingd.) 10, 207–217 (2018).
Martinez Torres, C. et al. Enlightening intracellular complexity of living cells with quantitative phase microscopy. Quant. Phase Imaging II 9718, 97182C (2016). In:Vol.
Milani, M., Ballerini, M. & Batani, D. High resolution microscopy techniques for the analysis of biological samples: a comparison. Eur. Phys. J. Appl. Phys. 26, 123–131 (2004).
Sibarita, J. B. Deconvolution microscopy. Adv. Biochem. Eng. Biotechnol. 95, 201–243 (2005).
Pasini, E. M., Van Den Ierssel, D., Vial, H. J. & Kocken, C. H. A novel live-dead staining methodology to study malaria parasite viability. Malar. J. 12, 190 (2013).
Wei, Y. & Xu, Q. A survey of force-assisted robotic cell microinjection technologies. IEEE Trans. Autom. Sci. Eng. 16, 931–945 (2019).
Läubli, N. F. et al. 3D mechanical characterization of single cells and small organisms using acoustic manipulation and force microscopy. Nat. Commun. 12, 1–11 (2021).
Zhang, H., Wei, X., Gao, Y. & Cretu, E. Frequency characteristics and thermal compensation of MEMS devices based on geometric anti-spring. Journal of Micromechanics and Microengineering 30, 085014 (2020).
Krenger, R., Burri, J. T., Lehnert, T., Nelson, B. J. & Gijs, M. A. Force microscopy of the Caenorhabditis elegans embryonic eggshell. Microsystems & nanoengineering 6, 1–11 (2020).
O’Bryan, C. S., Schulze, K. D. & Angelini, T. E. Low force, high noise: Isolating indentation forces through autocorrelation analysis. Biotribology 20, 100110 (2019).
Offroy, M., Razafitianamaharavo, A., Beaussart, A., Pagnout, C. & Duval, J. F. L. Fast automated processing of AFM PeakForce curves to evaluate spatially resolved Young modulus and stiffness of turgescent cells. RSC Adv. 10, 19258–19275 (2020).
Reddy D., Dheeraj, Kiran, Bhavana V., Krishnappa H. K. Brain tumor detection using image segmentation techniques. Proc 2018 IEEE Int Conf Commun Signal Process ICCSP 2018. Published online, 18–22 (2018).
Devi T. G., Patil N. Analysis & evaluation of Image filtering Noise reduction technique for Microscopic Images. 2020 Int Conf Innov Trends Inf Technol ICITIIT 2020. Published online (2020).
Obara, B., Roberts, M. A. J., Armitage, J. P. & Grau, V. Bacterial cell identification in differential interference contrast microscopy images. BMC Bioinforma. 14, 134 (2013).
Chow, Y. T. et al. A high-throughput automated microinjection system for human cells with small size. IEEE/ASME Trans. Mechatron. 21, 838–850 (2016).
Liu Y., Wang X., Zhao Q., Zhao X., Sun M. Robotic batch somatic cell nuclear transfer based on microfluidic groove. IEEE Trans Autom Sci Eng. Published online, 1–10 (2020).
Ma, P. et al. Molecular detection of cordycepin-induced HeLa cell apoptosis with surface-enhanced raman spectroscopy. Applied Sciences 9, 3990 (2019).
Zhang, Y., Tan, K. K. & Huang, S. Vision-Servo system for automated cell injection. IEEE Trans. Ind. Electron 56, 231–238 (2009).
Lofroth, M. & Avci, E. Auto-focusing approach on multiple micro objects using the prewitt operator. Int. J. Intell. Robot Appl. 2, 413–424 (2018).
Yang, L., Paranawithana, I., Youcef-Toumi, K. & Tan, U. X. Automatic vision-guided micromanipulation for versatile deployment and portable setup. IEEE Trans. Autom. Sci. Eng. 15, 1609–1620 (2018).
Yang L., Youcef-Toumi K., Tan U. X. Towards automatic robot-assisted microscopy: An uncalibrated approach for robotic vision-guided micromanipulation. IEEE Int Conf Intell Robot Syst. 2016-Novem(October), 5527–5532 (2016).
Yamato, K., Chiba, H. & Oku, H. High speed three dimensional tracking of swimming cell by synchronous modulation between TeCE Camera and TAG Lens. IEEE Robot Autom. Lett. 5, 1907–1914 (2020).
Avci, E. et al. High-speed automated manipulation of microobjects using a two-fingered microhand. IEEE Trans. Ind. Electron 62, 1070–1079 (2015).
Avci E., et al. Towards high-speed automated micromanipulation. Proc - IEEE Int Conf Robot Autom. Published online, 1718–1723 (2013).
Wang, H. et al. Three-Dimensional autofocusing visual feedback for automated rare cells sorting in fluorescence microscopy. Micromachines 10, 567 (2019).
Bai K., et al. Automated sorting of rare cells based on autofocusing visual feedback in fluorescence microscopy. IEEE Int Conf Intell Robot Syst. Published online, 1567–1572 (2019).
Wang W. H., et al. A microrobotic adherent cell injection system for investigating intracellular behavior of quantum dots. In: 2008 IEEE International Conference on Robotics and Automation. 407–412 (2008).
Wang, X. et al. Intracellular manipulation and measurement with multipole magnetic tweezers. Science robotics 4, eaav6180 (2019).
Xie, M., Shakoor, A., Li, C. & Sun, D. Robust orientation control of multi-DOF cell based on uncertainty and disturbance estimation. Int. J. Robust. Nonlinear Control 29, 4859–4871 (2019).
Dong, D., Lam, W. S. & Sun, D. Electromagnetic actuation of microrobots in a simulated vascular structure with a position estimator based motion controller. IEEE Robot Autom. Lett. 5, 6255–6261 (2020).
Ma, W., Li, J., Niu, F., Ji, H. & Sun, D. Robust control to manipulate a microparticle with electromagnetic coil system. IEEE Trans. Ind. Electron 64, 8566–8577 (2017).
Xie, M., Li, X., Wang, Y., Liu, Y. & Sun, D. Saturated PID control for the optical manipulation of biological cells. IEEE Trans. Control Syst. Technol. 26, 1909–1916 (2018).
Shakoor A., et al. A robotic surgery approach to mitochondrial transfer amongst single cells. In: IEEE International Conference on Intelligent Robots and Systems. (2019).
Xie, H., Meng, X., Zhang, H. & Sun, L. Development of a magnetically driven microgripper for piconewton force-controlled microscale manipulation and characterization. IEEE Trans. Ind. Electron 67, 2065–2075 (2020).
Xie, H. et al. Living cell manipulation and in situ nanoinjection based on frequency shift feedback using cantilevered micropipette probes. IEEE Trans. Autom. Sci. Eng. 17, 142–150 (2020).
Müller, D. J. & Dufrêne, Y. F. Atomic force microscopy: A nanoscopic window on the cell surface. Trends Cell Biol. 21, 461–469 (2011).
Shi, X., Zhang, X., Xia, T. & Fang, X. Living cell study at the single-molecule and single-cell levels by atomic force microscopy. Nanomedicine 7, 1625–1637 (2012).
Wei, Y. & Xu, Q. Design and testing of a new force-sensing cell microinjector based on small-stiffness compliant mechanism. IEEE/ASME Trans. Mechatron. 4435, 1–1 (2020).
Stavrov, V. T., Shulev, A. A., Hardalov, C. M., Todorov, V. M. & Roussev, I. R. All-silicon microforce sensor for bio applications. Smart Sens., Actuators, MEMS VI 8763, 87630Y (2013).
Grech, D. et al. A Quasi-Concertina force-displacement MEMS probe for measuring biomechanical properties. Sens. Actuators, A Phys. 275, 67–74 (2018).
Kim, K., Sun, Y., Voyles, R. M. & Nelson, B. J. Calibration of multi-axis MEMS force sensors using the shape-from-motion method. IEEE Sens J. 7, 344–351 (2007).
Kim, K., Cheng, J., Liu, Q., Wu, X. Y. & Sun, Y. Investigation of mechanical properties of soft hydrogel microcapsules in relation to protein delivery using a MEMS force sensor. J. Biomed. Mater. Res - Part A 92, 103–113 (2010).
Zhang, Y., Oh, Y., Stauffer, D. & Polycarpou, A. A. A microelectromechanical systems (MEMS) force-displacement transducer for sub-5 nm nanoindentation and adhesion measurements. Review of Scientific Instruments 89, 045109 (2018).
Ruppert, M. G., Fowler, A. G., Maroufi, M. & Moheimani, S. O. R. On-chip dynamic mode atomic force microscopy: A silicon-on-insulator MEMS approach. J. Microelectromechanical Syst. 26, 215–225 (2017).
Kim, K., Cheng, J., Liu, Q., Wu, X. Y. & Sun, Y. Investigation of mechanical properties of soft hydrogel microcapsules in relation to protein delivery using a MEMS force sensor. J. Biomed. Mater. Res. - Part A 92, 103–113 (2010).
Li, Z., Gao, S., Brand, U., Hiller, K. & Wolff, H. A MEMS nanoindenter with an integrated AFM cantilever gripper for nanomechanical characterization of compliant materials. Nanotechnology 31, 305502 (2020).
Xie, Y., Sun, D., Tse, H. Y. G., Liu, C. & Cheng, S. H. Force sensing and manipulation strategy in robot-assisted microinjection on zebrafish embryos. IEEE/ASME Trans. Mechatron. 16, 1002–1010 (2011).
Xie Y., Sun D., Liu C. Penetration force measurement and control in robotic cell microinjection. 2009 IEEE/RSJ Int Conf Intell Robot Syst IROS 2009. Published online, 4701–4706 (2009).
Xie, Y., Zhou, Y., Lin, Y., Wang, L. & Xi, W. Development of a microforce sensor and its array platform for robotic cell microinjection force measurement. Sens. (Switz.) 16, 1–13 (2016).
Zhou Y., Liu S., Yang C., Chong L., Yu X. Design of a microforce sensor based on fixed-simply supported beam: Towards realtime cell microinjection. 2015 IEEE Int Conf Cyber Technol Autom Control Intell Syst IEEE-CYBER 2015. 1080–1084 (2015).
Rajagopalan, J., Tofangchi, A. & Saif, M. T. A. Linear high-resolution BioMEMS force sensors with large measurement range. J. Microelectromechanical Syst. 19, 1380–1389 (2010).
Dostanić, M. et al. A Miniaturized EHT platform for accurate measurements of tissue contractile properties. J. Microelectromechanical Syst. 29, 881–887 (2020).
Adam, G. & Cappelleri, D. J. Towards a real-time 3D vision-based micro-force sensing probe. J. Micro-Bio Robot 16, 23–32 (2020).
Gao, W., Zhao, L., Jiang, Z. & Sun, D. Advanced biological imaging for intracellular micromanipulation: methods and applications. Appl Sci. 10, 7308 (2020).
Polacheck, W. J. & Chen, C. S. Measuring cell-generated forces: a guide to the available tools. Nat. Methods 13, 415–423 (2016).
Huang, H., Sun, D., Mills, J. K., Li, W. J. & Cheng, S. H. Visual-based impedance control of out-of-plane cell injection systems. IEEE Trans. Autom. Sci. Eng. 6, 565–571 (2009).
Huang, H. B., Sun, D., Mills, J. K. & Cheng, S. H. Robotic cell injection system with position and force control: Toward automatic batch biomanipulation. IEEE Trans. Robot 25, 727–737 (2009).
Dai, C. et al. Robotic manipulation of deformable cells for orientation control. IEEE Trans. Robot 36, 271–283 (2020).
Sun, D. et al. A force control approach to a robot-assisted cell microinjection system. Int J. Rob. Res 29, 1222–1232 (2010).
Wang, G. & Xu, Q. Design and precision position/force control of a piezo-driven microinjection system. IEEE/ASME Trans. Mechatron. 22, 1744–1754 (2017).
Zhang, W. et al. Force-Controlled mechanical stimulation and single-neuron fluorescence imaging of drosophila larvae. IEEE Robot Autom. Lett. 6, 3736–3743 (2021).
Zhang, W., Qu, J., Zhang, X. & Liu, X. A model compensation-prediction scheme for control of micromanipulation systems with a single feedback loop. IEEE/ASME Trans. Mechatron. 22, 1973–1982 (2017).
Xu Q. Micromachines for Biological Micromanipulation. Springer International Publishing (2018).
Zhang, W., Dong, X. & Liu, X. Switched Fuzzy-PD control of contact forces in robotic microbiomanipulation. IEEE Trans. Biomed. Eng. 64, 1169–1177 (2017).
Xu, Q. Design and smooth position/force switching control of a miniature gripper for automated microhandling. IEEE Trans. Ind. Inform. 10, 1023–1032 (2014).
Palanker, D. et al. Technique for cellular microsurgery using the 193‐nm excimer laser. Lasers Surg. Med. 11, 580–586 (1991).
Shalek, A. K. et al. Nanowire-mediated delivery enables functional interrogation of primary immune cells: Application to the analysis of chronic lymphocytic leukemia. Nano Lett. 12, 6498–6504 (2012).
Wang, X. et al. A microengineered cell fusion approach with combined optical tweezers and microwell array technologies. RSC Adv. 3, 23589–23595 (2013).
Ammi, M. & Ferreira, A. Realistic visual and haptic rendering for biological-cell injection. Proc. - IEEE Int Conf. Robot Autom. 2005, 918–923 (2005).
Li, J. & Pumera, M. 3D printing of functional microrobots. Chem. Soc. Rev. 50, 2794–2838 (2021).
Dong, M. et al. 3D‐printed soft magnetoelectric microswimmers for delivery and differentiation of neuron‐like cells. Advanced Functional Materials 30, 1910323 (2020).
Guillaume-gentil, O. et al. Force-controlled fluidic injection into single cell nuclei. Small 9, 1904–1907 (2013).
Acknowledgements
Figure 1 was created with BioRender.com. This work was partly supported by grants from the Hong Kong Research Grants Council (Ref. nos. C1134-20G and 11211421), the Shenzhen Science and Technology Project, China (Ref. no. SGDX20201103093 00502), the National Science Foundation of China (Ref. no 52105589), and the RGC Postdoctoral Fellowship Scheme (PDFS) (Ref. no. PDFS2021-1S05).
Author information
Authors and Affiliations
Contributions
D.S. proposed the idea and conceived the project with Z.J. A.S., W.G., D.S., and Z.J. wrote the manuscript. L.Z. provided important pieces of advice on this work. All authors reviewed and approved the manuscript.
Corresponding author
Ethics declarations
Conflict of interest
The authors declare no competing interests.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Shakoor, A., Gao, W., Zhao, L. et al. Advanced tools and methods for single-cell surgery. Microsyst Nanoeng 8, 47 (2022). https://doi.org/10.1038/s41378-022-00376-0
Received:
Revised:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41378-022-00376-0
This article is cited by
-
In vivo femtosecond laser nanosurgery of the cell wall enabling patch-clamp measurements on filamentous fungi
Microsystems & Nanoengineering (2024)
-
Design of an automated robotic microinjection system for batch injection of zebrafish embryos and larvae
Microsystems & Nanoengineering (2024)
-
3D-printed PEDOT:PSS for soft robotics
Nature Reviews Materials (2023)
-
Single-Cell Sequencing Technology and Its Application in the Study of Central Nervous System Diseases
Cell Biochemistry and Biophysics (2023)
-
A bionic approach for the mechanical and electrical decoupling of an MEMS capacitive sensor in ultralow force measurement
Frontiers of Mechanical Engineering (2023)