Main

Lung disease, especially hyaline membrane disease, causes significant morbidity in preterm newborns (born before 37 wk after the beginning of the last menstrual period) (1). It is currently treated invasively by intratracheal instillation of surfactant (1). Aerosol therapy offers a noninvasive alternative and has the advantage of administering drugs locally to the site of action, thus avoiding systemic effects. Therefore, aerosol therapy may offer a potential therapeutic regimen for lung disease in preterm newborns, as well as for other conditions in which the application of aerosolized drugs such as steroids, β-mimetics, or prostacyclines is a treatment option. However, aerosol therapy in this age group has so far been proven to be inefficient and inefficacious (2). This may be due to the fact that current delivery devices do not meet the requirements of the specific anatomical and physiologic characteristics of the airways and breathing patterns in this special group of patients (3).

As in vivo aerosol deposition studies in preterm infants are difficult to perform, there is a need for an in vitro model of the upper airways in this age group. Airway models have proven to be a valuable tool in assessing in vitro aerosol delivery (46) and help define the optimal combination of device and drug formulation for aerosol delivery in different age groups.

Several human upper airway casts have been developed to study aerosol deposition (7,8) from infancy to adulthood. Both mathematical and anatomical models have been developed for adults (912). They show that the upper airways of adults (nose, mouth, pharynx, and larynx) are an important trap for inhaled aerosols (10,1316).

Infant airway anatomy and physiology differ considerably from older children and adults. These differences are likely to influence lung deposition, shown by only <1% of nominal dose being delivered to the lungs of ventilated infants as opposed to 8–22% in adults (17). Airway morphometrics change dramatically during the first months of life, as well as breathing patterns, airway resistance, and lung volume (18). Lung deposition of drugs is influenced by all of these parameters (19), altering the trajectory of an aerosol particle through the upper airways into the lungs. However, so far no model is available for preterm infants. Therefore, we developed and characterized the premature infant nose throat (PrINT)-upper airway model of a premature newborn to study aerosol delivery in this age group.

MATERIALS AND METHODS

Development of airway model.

The PrINT airway model was developed in three phases: i) first, a three-planar magnetic resonance imaging (MRI) scan of a preterm infant was performed at the University Hospital of Geneva, Switzerland (imaging phase); ii) second, the airway model was constructed using three-dimensional (3D)-printing and a photopolymer termed FullCure 720 (construction phase); iii) third, a high-resolution computed tomography (CT) scan was performed on the airway model, providing data to validate the anatomical accuracy (validation phase). Furthermore, flow-pressure patterns were obtained using an infant lung function testing device (Exhalyzer D, Ecomedics, Dürnten, Switzerland) (20) and aerosol deposition experiments were performed. The study was approved by the local medical ethical review board of the University Children's Hospital of Bern.

Imaging.

The model was constructed from an MRI (Avanto, 1.5 Tesla, Siemens Medical Solutions, Erlangen, Germany) data set acquired from a healthy male premature infant, with a gestational age of 32 wk (length: 34 cm; weight: 1750 g), who underwent the above mentioned investigation in the context of a brain development imaging study (21). The child was sedated, lying in supine position and breathing spontaneously. The duration for the MRI data acquisition was 40 min.

Acquisition of the primary MRI data were achieved using two different imaging modes: a 3D Fourier-transforming fast gradient recalled sequence (1.5-mm coronal slices, flip angle 25°C, repetition time 15 ms, echo time 4.4 ms, field of view 18 cm, matrix 256 × 256) and a double-echo (proton density and T2-weighted) spin-echo sequence (1.5-mm coronal slices, repetition time 3500 ms, echo times 30 and 150 ms, field of view 18 cm, matrix 256 × 256, interleaved acquisition). The volume of pixel (voxel) dimensions for both acquisitions was 0.7 × 0.7 × 1.5 mm. The slices were then transferred to an imaging workstation for additional 3D and multiplanar reconstructions using Mimics, and Magics Software (Materialise, Munich, Germany).

Reconstruction.

Reconstruction took place at the Institute for Rapid Product Development at the University of Applied Sciences in St. Gallen, Switzerland. Optimal negative and positive gray scale thresholds were selected for each slice of the original MRI (coronal slices), stored as a Digital Imaging and Communications in Medicine-Format (DICOM), to provide maximal accuracy of air-mucosa distinction (Fig. 1). Subsequently, the slices were joined together to compose a 3D reconstruction image, which was stored in Standard Transformation Language (STL) format, and transferred over local area network (LAN) to the fabrication devices. The generated STL file also allowed us to define a choice of sagittal, coronal, and axial cross-sections of the model.

Figure 1
figure 1

Two-dimensional visual comparison of a selected (AC) sagittal and (DF) transversal cross-section through the upper airway; (A) the original MRI data from the infant, (C) the control CT scan of the constructed upper airway model. (B) Shows the corresponding STL file at the section position through the left nostril down to the larynx. (DF) show an axial cross section through the nose, (D) the original MRI data file, (F) the control CT scan, and (E) the corresponding STL file. The highlighted line in A corresponds to the air-mucosa border.

We constructed two models using a photopolymer resin (PolyJet FullCure 720, Objet Geometries Ltd., Sint-Stevens-Woluwe, Belgium), a transparent, stiff material widely used in rapid product development techniques on an Objet Eden 330 3D Printer (Objet Geometries, Rehovot, Israel). Printing layer thickness was 0.016 mm. Of this material, we built one model showing the entire head's surface (Fig. 2A) and a reduced laboratory model (Fig. 2BD), consisting of the face and air conducting parts only to conduct all further tests. Anatomical structures included in both models were all air conducting parts from the nostrils to 4 mm below the glottis.

Figure 2
figure 2

(A) Side view of the complete upper airway model; B, C, and D show a lateral, up-front and side view of the reduced laboratory model (face and air conducting parts), all built in FullCure 720.

Validation.

A CT scan (Philips Brilliance 16P, Philips Medical Systems, Cleveland, OH) was performed on the airway model (voltage: 90 kV, current: 95 mA) with a slice thickness of 0.8 mm. Both the original MRI and the model CT scan data sets underwent comparison using Materialise Magics, and Mimics software, achieved by two methods. First, a slice per slice comparison was done optically followed by measurements of landmarks. Then a calculation of airway volumes and cross-sectional areas was performed. To facilitate the comparison, limiting cross-section planes were defined behind the nostrils and below the glottis providing specific anatomical landmarks. Deviations in volume of less than 1% were defined as highly accurate.

A flow-resistance profile through the air conducting part was done using the constructed preterm upper airway model and an Infant Pulmonary Lung function Testing System Exhalyzer D Infant (Ecomedics, Dürnten, Switzerland), which was calibrated according to the procedure described by the European Respiratory and American Thoracic Society Task Force (20). As the intention was to assess the model's physical properties and not in vivo physiologic conditions, we carried out measurements that included pressure and flow values exceeding a preterm infant's range. To record the flow-resistance profile in inspiration, a face mask (Laerdal Infant Silicon Mask 0/1, Laerdal Medical, Germany) was attached to the airway model using therapeutic putty to ensure an airtight fit on the nonflexible surface of the airway model. Attached to the facemask were a pressure transducer and an ultrasonic flowmeter (Spiroson, Ecomedics, Dürnten, Switzerland), followed by an air-supply tube and a variable air valve. The air flow regulator was subsequently set to randomly chosen, increasing values, beginning at a flow of zero l/min eventually reaching a maximum value of 18.8 L/min corresponding to pressure values of 0 and 11.1 kPa, respectively.

Second, the set-up was changed using the subglottic opening at the air entrance, hence simulating the expiratory pathway. Therefore, the measurement and air-supply setup was placed at the caudal end of the model connected by a 22F-T-piece-connector (No. 1983, Intersurgical, Wokingham, UK) to ensure an airtight link to the model. Again, flow-pressure measurements were done as previously described. The airflow ranged from 0 up to 18.6 L/min corresponding to a pressure of 0 and 11.5 kPa, respectively.

Feasibility of aerosol deposition experiments.

As an example, we measured aerosol delivery of a typically used drug, Budesonide solution (Budeflow, PARI GmbH Starnberg, Germany), from a perforated vibrating membrane nebulizer (e-Flow, PARI GmbH Starnberg, Germany) attached to our airway model with a facemask (Laerdal Neo, model 0, Laerdal Medical, Germany). The Budesonide solution was specifically developed for use with perforated membrane devices. Particle size as measured by the manufacturer shows a mass median diameter (MMD) of 3.12 μm with a geometric SD (GSD) of 1.9.

The nebulizer was connected directly to the face mask to reduce the dead space to a maximum of 4 mL and then attached to the upper airway model using Parafilm (Pechiney, Chicago) to ensure a tight fit. Air tightness was tested before and after each run by using a water column allowing ± 1 cm H2O per minute. Ambient air was kept constant at 23°C and 50% humidity. The air in the bypass-flow-system supplying the impactor was also kept constant at 23°C and a humidity of 50 ± 5%.

The model was connected to a Next Generation Pharmaceutical Impactor (NGI, MSP Corp. Shoreview, MN) whose temperature was constantly measured and maintained at 23°C, with a driving flow of 15 L/min.

Constant inspiration was simulated with flows of 1, 5 and 10 L/min through our model to evaluate lung dose. Constant nebulization was performed until depletion of nebulizer. Each experiment was repeated six times.

Budesonide concentration was measured using HPLC. Emitted dose was defined as total amount of drug emitted by the nebulizer. Lung dose was defined as amount of drug recovered in the impactor and expressed as percentage of emitted dose. Nominal dose was defined as total amount of drug filled in to the nebulizer.

Statistical analysis.

Statistics were carried out using analysis of variance (StatView 512+; Abacus Concepts Inc, Berkeley, USA) with a significance level of 95% (p < 0.05).

RESULTS

Modeling and validation.

The final upper airway model is shown photographically in Figure 2BD. Figure 1 shows a visual comparison between the CT scan (C) of the model and the original MRI scan (A), for illustration the STL file (B) is provided at the corresponding section position. For real volume comparison, 3D reconstruction was done (Fig. 3).

Figure 3
figure 3

3D-Validation of cast geometry: Extracted air conducting parts of (A) the original MRI data file and (B) control CT scan of the cast. For volume calculation, limiting section planes were introduced at the nostrils (n) and below the glottis (g). Further shown are the pharynx (p), the piriform sinus (s), and the trachea (t). The difference in volume was <1%.

The airway volume within the original MRI was 1277 mm3 and that of the CT scan of the model built in FullCure 720 was 1265 mm3 (Fig. 3). These values showed an absolute difference in volume of 12 mm3, which was equivalent to a relative deviation of 0.94%. Their discrepancy in diameter with a mean aberration of approximately 0.03 mm is not shown.

Flow characteristics and resistance of cast.

Fig. 4 shows corresponding pressure values to each applied flow. The pressure slopes prove to be nonlinear, representing square functions (Inspiration: y = 143.08x2 − 12.16x + 0.67; Expiration: y = (137.64x2 − 7.1721x + 0.26)−1, coefficients of determination (r2) are 0.9988 and 0.9994, respectively). Mean upper airway resistance was 20.5 cm H2O · l−1 · s (range: 5.8–36.7) for the inspiratory, and 23.1 cm H2O · l−1 · s (range: 4.2–37.9) for the expiratory pathway.

Figure 4
figure 4

Resistance profile within the PrINT model for inspiratory and expiratory flows.

Feasibility of aerosol deposition measurements.

Increasing constant inspiratory flow resulted in a decrease in lung dose (Fig. 5).

Figure 5
figure 5

Lung dose (LD) for budesonide, expressed as percentage of nominal dose (ND). Displayed are mean values and error bars in SD, ‡ and * p < 0.001 (1 vs 5 L/min, 5 vs 10 L/min, respectively).

Lung dose decreased with increasing flow from 61.8 (±5.3) %, 26.0 (±1.5) % to 9.0 (±0.8) % of nominal dose at 1, 5, and 10 L/min, respectively. The differences were statistically significant (p < 0.001 for each measurement condition). Drug-recovery was 86.7, 77.0, and 79.9% of the nominal dose for 1, 5, and 10 L/min, respectively.

DISCUSSION

Based on MRI data obtained from a premature neonate of 32-wk gestational age, we developed an anatomically correct model of the upper airways to study in vitro aerosol delivery. The cast's highly nonlinear flow characteristics and the feasibility of measuring flow-dependent drug delivery into the airways could be demonstrated. To our knowledge, this is the first description of a premature infant upper airway model and its suitability for aerosol studies so far.

The overall goal in inhalation therapy is to deliver medication to the bronchopulmonary tree and lower airways, avoiding deposition in the upper airway tract. Generally, for simulation of aerosol distribution in the respiratory tract, a geometric model of the upper airways is used, the United State Pharmacopoeia (USP) throat, and for the lower airways, the Andersen cascade impactor. Both are designed for simulation of aerosol deposition patterns in adults. The USP throat is a 90° bent tube with a smooth inner surface, assuming that the adult patient inhales via a mouthpiece. As infants inhale via the nose, and upper airways geometry differs not only in dimensions but also in proportion from those of adults, this model is not suitable for studying aerosol deposition in newborn infants. Furthermore, in a number of in vitro studies (46,22,23) it has been pointed out, that the use of a realistic human upper airway replica would be preferable to the USP throat. The main reason, why anatomically correct upper airway models are of interest is that the upper airways act as an important trap for inhaled aerosols (10,14,16,24,25). Aerosol deposition occurs mostly in the larynx and trachea (26,27), and is determined by pharyngeal and glottal morphologic constrictions and consequent flow jets (28). It is therefore imperative to have realistic in vitro models of the upper airways including the larynx and tracheal opening.

The SAINT model, an upper airway model of a 9-mo-old infant, described by Janssens et al. (4), served as master for our work. The model is derived from a CT scan and reconstructed using stereolithography. In contrast to the already existing upper airway model of Phalen et al. (14), which was an extrapolated simplification of an adult nose model by reducing adult airway dimensions by 50%, this model was the first realistic upper airway model of a 9-mo-old infant to correctly represent the anatomy of the upper airways, including the laryngeal part and specifically taking into account nose breathing. The model includes the face, nasal cavity, and pharynx extending to the subglottic region. The nasal airway is open for air passage whereas the oral airway is closed. It has been used in many different studies and is well established (6,23,2931). In contrast to the above mentioned model, we derived our model from a MRI dataset of a preterm infant, used a different material, and performed aerosol experiments, which can be used for quality assessment.

As model reconstruction is based on an imaging dataset, the quality of data-transposition is crucial. Because of ethical considerations (high radiation), we did not perform CT on a preterm infant, although image resolution and acquisition time would have been preferable. Our dataset consisted of an MRI with limited resolution due to voxel size and comparatively long acquisition time. There are studies in adults (22,23) describing successful evaluation of the upper airways using respiratory-gated MRI to evaluate airway dynamics during tidal breathing, allowing a better distinction of the air mucosa border. However, this technique is not yet established in preterm neonates. Therefore, the manual definition of the optimal threshold for air mucosa distinction is the most critical point in the reconstruction process in our model. The accuracy of this step can only be judged by indirect measurements of physiologic data. Our measurement of an upper airway volume in our model of 1.27 mL is consistent with the anatomic and physiologic dead spaces of 1–2 mL/kg reported in other studies (32,33), which are between. In our infant, with a birth weight of 1750 g, the estimated dead space would be expected to be between 1.75 and 3.5 mL. We only measured the upper airways from the nostrils down to 4 mm below the glottis, excluding the trachea and bronchi, accounting for the difference between measured and expected volume. The airway resistance measured within our model was at the upper end of the reported range for nasal resistance in Caucasian children weighing between 1.5 and 10.2 kg (34), supporting the hypothesis that our geometrically identical in vitro model is also comparable in flow-resistance behavior with in vivo measurements. The relatively high resistance can be explained by the additional resistance of pharynx and larynx as well as the rigid, nonexpandable geometry of our model. In vivo, the upper airway geometry changes during tidal breathing, but simulation of dynamically changing geometry is technically difficult. These findings underline the assumption that the reconstruction process was anatomically correct and within the physiologic range.

The aerosol deposition measurements were conducted with the intention to provide a quality assessment tool to guarantee consistency of the aerosol filtering behavior of the model, to be applied as pre- and postexperiment calibration. Therefore, the applied flows were chosen arbitrarily within the range of published maximal expiratory flows during forced expiration in newborns of about 6 L/min in (35), and from unpublished observations (U. Frey, personal communication) in preterm infants during clinical routine lung function testing, quiet sleep, and tidal breathing with peak tidal flows of 3.9 L/min (range 2.5–6.5 L/min). Although, dead space is not relevant during constant inspiratory flow experiments, we took great care to design experiments in a manner usable for further aerosol deposition studies and according to the experiments previously described by Janssens et al. (4). Our data showed the expected reciprocal dependency between inspiratory flow and lung dose due to higher inertial impaction of aerosol particles in the upper airways, resulting in decreasing aerosol penetration to the lower airways. This may be due to secondary flow structures that arise in areas of higher inertial forces (transitional-to-turbulent flow), complicating flow structures as well as aerosol transport and deposition. Computation fluid dynamic simulations of the upper airways in adults have shown that secondary flow structures arise as early as in the nostrils, in-part contributing to the development of turbulent flow structures in upstream airway sections, such as the glottis, leading to deposition of aerosols in the airway wall in these regions (36,37). Increasing inspiratory flows resulted in a lower lung dose.

This finding is in good agreement with other studies using upper airway cast models (4,6,30). In an in vitro study by Schuepp et al. (30), using the SAINT model and the same nebulizer–drug combination we used, lung doses, calculated as percentage of emitted dose, were in the range of 7.9–42.3%. In contrast to the sinusoidal breathing patterns they applied, we used constant inspiratory flows (no expiration). Assuming that half the time of a breathing cycle is expiration, a comparison of their lung doses with our results was only possible if we divided our results by two (corrected range 4.5–30.8%). Therefore, we concluded that lung dose in preterm infants is less than in a 9-mo-old child.

As the authors state themselves, the material used to build the SAINT model (Stereocol) interfered with the HPLC signal during wash out using ethanol, impeding direct determination of upper airway deposition from direct wash out of the cast (4). We have tested FullCure 720 extensively for solubility behavior with ethanol and methanol. By washing out the airways five times with 10 mL of methanol, we could detect extractable material from the final cast that could potentially interfere with the budesonide analysis. The error introduced by this was estimated to be around 3% and was considered acceptable for the purpose of the measurements conducted here. By soaking the model in methanol over 24 h, we detected unacceptably high levels of extracted material, leading us to the conclusion to limit rinsing to a maximum of five times with 10 mL of methanol. Furthermore, the minimal aberrations in the geometry from the original data, proven to be highly accurate by the validation process comparing anatomical landmarks, distances, and relative deviation in airway volume of 0.94%, endorsed our material selection.

In conclusion, we described an anatomically correct preterm infant nose throat-model and showed that this model may be used for realistic in vitro aerosol experiments; allowing obtainment of in vitro information, which is difficult or impossible to obtain in vivo. Inhaled therapy in preterm infants is difficult and may be improved by designing new inhalation devices and patient-interfaces. Further work on the influence of different breathing patterns and applied aerosol particle sizes on lung deposition would be an important step in improving aerosol therapy in the highly vulnerable group of premature infants with drugs such as topical steroids, prostacycline, or lung surfactant.