Main

Noninvasive and detailed assessment of cardiopulmonary hemodynamics in the sick preterm infant is now possible using Doppler echocardiography. However, conventional techniques only allow intermittent observations resulting in repeated disturbance of the infant, which can itself change hemodynamic parameters. To date, no continuous method of measuring cardiac output noninvasively is available for use in the newborn. In contrast, in adult practice, cardiac output measurements have been made continuously using the partially invasive technique of transesophageal Doppler echocardiography (1, 2); routine use is recommended by some units to optimize management in adults with circulatory failure. Miniature transesophageal probes have been developed for use in infants and children (3, 4). However, sizes suitable for the newborn are not yet available.

Continuous Doppler recordings of neonatal cerebral blood flow (57) have been possible and, when combined with continuous monitoring of intra-arterial blood pressure (8), arterial oxygen saturation (9), and content, have provided valuable insight into the circulation not obtained through intermittent measurements.

Precordial Doppler echocardiographic signals are easier to obtain in the newborn than in the adult but, although Doppler signals from the aorta and pulmonary artery have been used to derive cardiac output in the newborn (1015), no method of obtaining continuous recordings has been described. Because the use of such continuous data has an important role during neonatal intensive care, we have developed a new steerable and fixable precordial Doppler probe.

We aimed to evaluate the performance of this newly designed ultrasound probe in the measurement of precordial Doppler echocardiographic signals in healthy term infants and to record these signals continuously for up to 1 h and, thereby, to derive values for cardiac output.

METHODS

Doppler Machinery and Probe

A 4-MHz pulsed-wave Doppler crystal was inserted into a specially designed steerable joystick device that was made in such a way that it could screw into a commercially available disposable adhesive transcutaneous fixation ring (Radiometer, Copenhagen) (Figs. 1 and 2). The signals are then transmitted to and processed by a nonimaging pulsed-wave Doppler system originally designed for use in peripheral vascular Doppler sonography (Dopstation, Scimed Ltd.).

Figure 1
figure 1

Steerable Doppler probe with 4-MHz pulsed-wave Doppler crystal (11-mm diameter) consisting of a joystick device and a screwable base (total height, 2.7 cm).

Figure 2
figure 2

Attachment of probe to the infant's chest with use of a commercially available disposable adhesive transcutaneous fixation ring (Radiometer, Copenhagen). Written parental consent was obtained.

The Doppler frequency-shift signals are displayed in real time and trend along with the power of the signal. Signal power gives an indication of the volume of moving blood within the pulsed-wave Doppler sample gate independently of the speed of blood sampled. As it is related to the emitted ultrasound frequency, the signal power is also known as “relative backscatter power” (16). A decrease in the power of the signal indicates that the sample volume has decreased and implies that the sample gate has moved out of the blood vessel that is being interrogated.

Safety of the 4-MHz Doppler transducer had been examined with respect to potential adverse biologic effects by use of a hydrophone and head amplifier. The ultrasound beam was visualized using a Schlieren system, which showed that the beam was tightly focused at 1.5 cm from the face of the transducer where the energy output of the transducer was highest. Energy output measured 2.26 mW/cm2. The American Institute of Ultrasound in Medicine recommends an acceptable safety level of 100 mW/cm2 (17). Ethical approval for the study of human infants with use of this system was obtained from the local Research Ethics Committee.

Study Population

Twenty-five healthy term infants were recruited from the postnatal wards after one or both parents gave informed consent. Infants were included in the study only if they were in a calm and resting state. Six of these 25 infants became unsettled either at the beginning or during the recording and were excluded. Each infant underwent a clinical cardiovascular examination together with a cross-sectional echocardiogram to exclude structural cardiac abnormality by use of the Hewlett-Packard Sonos 2500 ultrasound machine.

The study was divided into two phases: the first in which different echocardiographic approaches were investigated in six infants, and the second in which continuous measurement of Doppler signals from the pulmonary artery was obtained in 13 infants.

Evaluation of Different Echocardiographic Approaches

Four echocardiographic windows, all of which have been used previously to assess cardiac output in adults (1821), were evaluated to determine the most useful site for continuous recording of Doppler signals:1) the main pulmonary artery, insonated from the midhigh left parasternum;2) the ascending aorta, insonated from the suprasternal notch or high right parasternum;3) the tricuspid valve, insonated from the lower left sternal edge or apex; and 4) the mitral valve, insonated from the apex.

Main pulmonary artery.

Recordings from the main pulmonary artery were found to be the most stable of all sites investigated. However, during movement, the signal quality was reduced along with a decrease in the power of the signal but returned to the resting value after the infant had settled, provided the joystick position had not changed.

Ascending aorta.

Signals were difficult to obtain continuously from the ascending aorta because of poor adhesion of the mount, particularly in the suprasternal notch. The probe would require redesign to solve this problem. Signals from the high parasternum were obtained for short periods, but small movements of the infant's head deflected the probe, preventing longer-term recordings.

Tricuspid valve.

Recordings of flow across the tricuspid valve were sometimes obtainable for a few minutes with the probe at the lower left sternal edge, but, at the apex, signals were lost with respiration due to excessive ribcage movements during normal resting respiration.

Mitral valve.

Recordings of flow across the mitral valve from the probe positioned at the apex were only intermittently possible for the same reason as for tricuspid flow.

Continuous Recordings

Because recordings from the pulmonary artery were most promising, continuous recording of pulmonary artery flow for a period up to 1 h was attempted in the remaining 13 babies. The study was stopped earlier if the infant became too restless.

Pulmonary artery diameter.

Pulmonary artery diameter was measured in the standard parasternal short-axis view to allow derivation of right ventricular output from the Doppler flow integral. Measurements were taken on five occasions at the hinge points of the valve leaflets seen on the frozen image, using the scroll facility. The average of these five measurements was stored in the Dopstation and used as a constant for the continuous calculation of right ventricular output.

Positioning of the probe.

Initially, the best Doppler signal was obtained manually by using fine manipulation guided by the Doppler signal only, positioning the probe over the pulmonary artery, and searching for a strong stable flow pattern with flow away from the probe. Figure 3 shows the output from the Dopstation: the optimal signal had minimal spectral dispersion, a smooth outline, and the highest velocity within the pulmonary artery. The signals are displayed as frequency shift measured in kHz (top trace). Gain and filter settings were adjusted to minimize artifact and valve clicks. Power of the signals, or relative backscatter power, is displayed at the bottom half of the screen. The range was arbitrarily set between 0 and 64.

Figure 3
figure 3

Printer output of the Dopstation showing Doppler flow signals during insonation of the pulmonary artery (top trace) with simultaneous display of signal power (bottom trace). Flow away from the probe is displayed above the zero line.

The adhesive transcutaneous fixation ring was then positioned at that particular site and the Doppler probe was screwed into it. Further fine adjustment of the direction of the ultrasound beam was achieved by manipulating the joystick within the adhesive ring and adjusting the depth of the pulse Doppler sample gate to give an optimal signal. The depths selected ranged between 2.3 and 2.8 cm. The length of the Doppler sample gate was fixed at 0.5 cm.

Analysis of signals.

The signals were processed by the Dopstation to display the data in real time and trend. The flow velocity integral was calculated by the Dopstation from the mean velocity, averaging the signals over 6-s epochs. This was automatically multiplied by the cross-sectional area of the pulmonary artery (derived from the previously measured pulmonary artery diameter) to give a 6-s flow volume. This was multiplied by 10 and adjusted for birth weight to give values in mL·kg−1·min−1.

RESULTS

Continuous measurements were obtained in 13 infants (aged 13 h to 4.5 d with birth weights ranging from 2890 to 3880 g). The median total recording time was 40 min (range, 24–60 min). During these recordings, movement interfered briefly with the recording to the extent that the signal was either decreased or lost transiently, typically during a wriggle or stretch. However, once the infant settled down again to its original position, the signal returned. Figure 4 shows a typical trace: the top graph represents right ventricular output measured as mL/min recorded over 1 h. The power of the signal (relative backscatter power) is recorded simultaneously over the same period and is displayed at the bottom of the screen. The vertical lines marked the times when the infant moved. If the signal did not restabilize, this was clearly indicated by an altered wave form with reduced velocity and a reduction in the power of the signal, indicating that the decrease in cardiac output was, in fact, artifactual. Signal power decreased sharply during exaggerated movement. This was clearly identifiable from considering the trace. Across the study population, periods during which the signal decreased because of movement were associated with an average decrease in power signal of 28%. This usually occurred when the probe was knocked or the wire pulled during a brief episode of restlessness. The probe was then readjusted to obtain the optimal signal and to return the power of the signal to starting values. The arrows on the trace demarcate the periods of artifactual decrease in cardiac output (Fig. 4).

Figure 4
figure 4

Printer output of Dopstation showing trending of right ventricular (RV) output (mL/min;top trace) and power of the signal (arbitrary units;bottom trace) over time (62-min screen). Cardiac output expressed in mL·kg−1·min−1 was derived after the recording. The vertical lines are markers denoting infant movement and are associated with a decrease in signal power. Data between arrows were excluded.

Median length of recording was 40 min, varying from 24 to 60 min. Episodes of movement identified by observation in which there was a simultaneous decrease in signal power have been excluded. The percentage of the recording that was excluded together with the length of recording achieved for each infant is given in Table 1. Between 15 and 53% of the total recording time was lost because of movement artifact.

Table 1 Median right ventricular output, length of recording, and percentage of record excluded due to movement artifact for each infant

Median right ventricular output was 180 mL·kg−1·min−1 (range, 148–246 mL·kg−1·min−1). Figure 5 shows the median and interquartile ranges for individual children (baby A–M). Whereas the values for median right ventricular output varied widely, for each individual child, the 6-s epochs showed much less variation with interquartile ranges varying only between 13.2 and 29.9 mL·kg−1·min−1, or between 8.3 and 14.5% of the median value.

Figure 5
figure 5

RV output (RVO) in study infants. Median, interquartile ranges, and minimum/maximum values displayed as box-and-whisker plot to show distribution of individual 6-s epochs.

DISCUSSION

Blood pressure measurement alone is a poor indicator of adequacy of cardiac output. It is well recognized that newborn infants are often able to maintain systemic blood pressure close to normal until cardiovascular collapse approaches. Further, in the preterm infant, there is wide variation in normal blood pressure values (22, 23) associated with adequate peripheral perfusion. Thus, the potential clinical value of a system for recording cardiac output continuously in the newborn is great, particularly if this can be achieved noninvasively. For example, there is increasing evidence that low left and right ventricular output measured with Doppler echocardiography is associated with poor outcome in asphyxiated infants or those with persistent hypoxemia (24, 25).

The system evaluated in this article represents, to our knowledge, the first attempt to record precordial Doppler echocardiographic signals continuously at any age. Newborn infants, especially those with hyaline membrane disease, have good ultrasonic windows because of the relative lack of inflated lung anterior to the heart (air being largely impervious to ultrasound), making this age group potentially the most suitable for such a system. However, we found it impracticable to record flow patterns continuously from the ascending aorta or either inlet valve in unsedated infants. We have the impression that the difficulty with obtaining ascending aortic flow was related mostly to the design of the probe, particularly to its size and shape, which may be possible to overcome in the future. However, we suspect that continuous recordings from the apex will never be possible because of ribcage movement.

The stability of the signals from the pulmonary artery relates to a number of factors. Pulmonary arterial blood flow is directed from the front of the chest toward the back. Thus, the probe can be placed on the infant's chest so it is away from the head and neck but, at the same time, still in line with blood flow. Respiratory movement did not noticeably affect alignment with flow. In contrast, respiration alters alignment with flow through the inlet valves when they are insonated from a fixed part of the chest near the apex of the heart.

The demonstration that continuous recording of right ventricular output was possible in healthy infants for up to 1 h suggests that the technique has considerable potential use in intensive care. Ultimately, we aim to use this system in unwell infants, particularly hypoxemic infants in the first days of life, most of whom will be receiving muscle relaxants or sedation and will be less likely to move the probe themselves. Thus, interference and movement artifacts should be even less of a problem.

The steerable joystick system was very effective in maintaining the recording of Doppler signals from the pulmonary artery in healthy term infants studied when asleep or resting quietly. Even with this method, it was difficult to maintain a stable probe position in the infant who was restless, feeding, or in any other position than prone. This may be somewhat improved by subtle refinements in manufacture of the probe, including making the wire attachment to it lighter than at present, although we doubt recordings will be feasible in the very active infant.

An important future development that will allow this or similar systems to be used routinely will be the automatic detection of changes in position of the Doppler sampling gate. The continuous recording of signal power in this study confirms that decreases in power occur with excessive movement and correlate with decreases in the Doppler-shift signal. True decreases in the mean Doppler-shift frequency (i.e. due to decreasing stroke volume) would be associated with an unchanged power of the signal, as the volume of blood within the sample gate remains constant. The best cutoff point for such an automated system is not yet clear and requires further study.

The absolute values for right ventricular output presented here are much less important than the serial continuous data. Variation in the baseline value can be expected due to error in the measurement of the pulmonary artery diameter. Indeed, the data may better be presented as “minute distance” derived from the Doppler signal [minute distance (cm/min) = flow velocity integral (cm) × heart rate (beats per minute)] (26, 27). This parameter is directly proportional to cardiac output and does not include measurement of pulmonary artery diameter. In future, refinement of the software will allow continuous display of minute distance.

Our data show remarkable stability in right ventricular output in the resting infant. In newborns undergoing intensive care, it is known that cardiac output can change up to 3-fold in response to changes in pulmonary blood flow due to fluctuating pulmonary vascular resistance. It, therefore, seems likely that the technique will be sensitive enough to detect significant change in cardiac output.

From the data presented here, it is not possible to differentiate the separate influences of stroke volume and heart rate on the variability of right ventricular output because the system currently analyzes 6-s epochs, and beat-to-beat counting by the system is unreliable. This additional information could be of important physiologic and clinical interest. We are in the process of developing an ECG gated beat-to-beat counting system.

It is, of course, interesting to speculate whether this system will be of value in neonatal intensive care. Potential future problems with the system in intensive care include interference from ventilator noise and disturbance of the system during handling. Further, turbulence in the pulmonary artery from left-to-right shunting may disturb the Doppler wave form in preterm infants recovering from respiratory failure. The technique would also be limited by factors affecting the echocardiographic window, such as chronic lung disease or pulmonary interstitial emphysema. One may, however, anticipate that it will find a place in the hemodynamic monitoring of larger newborns with hypoxemia or birth asphyxia who typically have low cardiac output (25). Such infants do not have turbulent flow in the pulmonary artery from left-to-right shunting and would clearly benefit from a system to monitor cardiac output.

The present study has shown that continuous recording of right ventricular output is possible in resting healthy infants for up to 1 h. It is, therefore, the first to record cardiac output continuously by a noninvasive means in the human infant. Therefore, it has considerable potential in the study of physiology of the healthy newborn circulation, particularly if further refinements to the probe and its lead attachments allow it to be more stable in the more active infant. The study also provides normative data for comparison with infants undergoing intensive care, where the system is currently being evaluated and further developed.