Abstract
Enzo Life Sciences Nuclear-ID™ Red DNA stain is useful for staging cells on the basis of their DNA content. The stain can be applied to live, detergent-permeabilized or fixed cells for quantification of their distribution among the three main phases of the cell cycle (G0/G1, S and G2/M). The far-red emission of the dye facilitates multiplexing with GFP and other fluorescent probes. With the dye, cells may be analyzed by fluorescence microscopy, flow cytometry and laser-scanning cytometry.
Main
The ability to identify cell cycle position through DNA content analysis is fundamental to investigations of cell cycle–regulated protein expression and assessment of perturbations in cell cycle progression. Hoechst 33342 is one of the few live-cell DNA stains used in cell cycle analysis. However, this dye requires UV-light excitation, and many laboratory flow cytometers are equipped only with a 488 nm and/or 633 nm laser. Few cell-permeable DNA dyes are excitable at these wavelengths, forcing reliance on time-consuming methods that require cell fixation and permeabilization. Propidium iodide staining of DNA in fixed cells has been widely adopted in cell cycle analysis, but it stains all double-stranded nucleic acids, so cells must be treated with RNase before staining1,2.
Nuclear-ID Red DNA stain is a novel DNA-intercalating dye that freely enters live cells, allowing cell cycle analysis without laborious permeabilization and fixation steps. Since the dye intercalates exclusively into double-stranded DNA, no RNase treatment is required and the fluorescence intensity of stained cells is directly proportional to their DNA content.
Cell cycle analysis by flow cytometry
With respect to live-cell analysis of cell cycle, Nuclear-ID Red DNA stain has similar performance characteristics to Hoechst 33342 stain. In untreated cells, the G0/G1 peak coefficient of variation is generally less than 10%, and the G2/G1 mean fluorescence ratio is roughly 1.8. Unlike Hoechst, the Nuclear-ID Red DNA stain is excited by either 488-nm or 633-nm laser sources. Staining is easy to perform: simply add the dye, incubate and analyze. The assay has been validated using a wide range of cell types, cell concentrations, dye concentrations, staining times and temperatures. Nuclear-ID Red DNA stain is especially suitable for monitoring drug-induced changes in cell cycle dynamics (Fig. 1). We exposed Jurkat cells to 0.1 μg ml−1 nocodazole, 3 μM etoposide or 30 μg ml−1 aphidicolin. Then we incubated the cells with a 1:250 dilution of Nuclear-ID Red DNA stain in medium and assayed the cultures using the FL3 channel of a FACSCalibur flow cytometer (BD Biosciences), equipped with a 488-nm laser. Flow cytometric analysis revealed the dye's ability to accurately profile DNA content and cell cycle distribution in cells subjected to these treatments.
Cell cycle analysis by laser-scanning microplate cytometry
The inability of flow cytometry to analyze adherent cells in situ, its requirement for relatively high densities of cells (105–106 cells per milliliter) and its inherently low-throughput nature has provided an impetus for development of alternative instrument platforms, such as laser-scanning microplate-based cytometry3. Using Nuclear-ID Red DNA stain with a microplate cytometer allows all cell processing steps to be performed in microplates. Scientists at TTP Labtech grew cells in culture medium alone or in culture medium with 0.2 μg ml−1 nocodazole (Fig. 2). A 1:1,000 dilution of Nuclear-ID Red DNA stain in 0.2% Triton X-100–containing phosphate-buffered saline was added after ∼20-hour drug incubation, and plates were scanned on an Acumen eX3 fluorescence microplate cytometer using the instrument's 633-nm laser. Cell cycle analysis of adherent HeLa cells using this instrument showed excellent correlation with results achieved by flow cytometry.
Multiplexing with GFP
Nuclear-ID Red DNA stain is a far-red–emitting dye (emission maximum, 640 nm) that provides minimal spectral overlap with GFP or with antibody conjugates of fluorescein (FITC), DyLight 488 (Thermo-Fisher Scientific) and Alexa Fluor 488 (Invitrogen Corporation). As the dye is highly resistant to photobleaching, it is suitable for wide-field and confocal fluorescence microscopy. We detected the nucleus with Nuclear-ID Red dye in GFP-expressing HeLa cells, containing a TurboGreen-tagged mitochondrial targeting sequence of human cytochrome c oxidase subunit VIII precursor (Marinpharm GmbH) (Fig. 3). Cells were stained with a 1:1,000 dilution of Nuclear-ID Red DNA stain. The Nuclear-ID Red dye was imaged using a Texas Red filter set, and the green fluorescence was imaged using a FITC filter set, on an Axio Observer Z1 wide-field fluorescence microscope equipped with an Apotome slider module (Zeiss).
Conclusions
The principal advantage of Nuclear-ID Red DNA stain in live cells is that the assay protocol is simple, as there is no need for fixation or RNase treatment. Nuclear-ID Red DNA stain is compatible with a broad range of instruments, including conventional and confocal fluorescence microscopes, as well as flow cytometers and microplate-based cytometers. Its far-red emission makes the stain ideal for multiplexing with common fluorescent probes, such as GFP.
References
Darzynkiewicz, Z., Crissman H.A. & Robinson, J.P. Flow cytometry In Methods in Cell Biology (2nd edn.) vol. II, part B (Academic Press, Inc., San Diego, 1994).
Nunez, R. DNA measurement and cell cycle analysis by flow cytometry. Curr. Issues Mol. Biol. 3, 67–70 (2001).
Bowen, W.P. & Wylie, P.G. Application of laser-scanning fluorescence microplate cytometry in high content screening. Assay Drug Dev. Technol. 4, 209–221 (2006).
Acknowledgements
The authors thank D. Onley and P. Wylie of TTP Labtech for performing the experiments on the Acumen eX3 instrument.
Author information
Authors and Affiliations
Corresponding author
Additional information
Disclaimer
This article was submitted to Nature Methods by a commercial organization and has not been peer reviewed. Nature Methods takes no responsibility for the accuracy or otherwise of the information provided.
Rights and permissions
About this article
Cite this article
Xiang, Y., Cox, H., Lebedeva, I. et al. A cell-permeant dye for cell cycle analysis by flow and laser-scanning microplate cytometry. Nat Methods 6, an2–an3 (2009). https://doi.org/10.1038/nmeth.f.275
Issue Date:
DOI: https://doi.org/10.1038/nmeth.f.275