Case report

A 52-year-old male was referred to the Department of Restorative Dentistry in 1999 requesting the restoration of the missing left maxillary first and second molars (26 and 27). These teeth had been extracted several years previously possibly due to endodontic or structural problems. The patient had no relevant medical history, was a non-smoker with no history of parafunction but was employed in a high stress occupation.

After discussions with the patient and undertaking a full evaluation including radiographic assessment and study models, it was decided to restore the edentulous space using an implant-retained bridge. Subsequently, two 15 mm Mark II Brånemark regular platform fixtures were inserted in 1999 under a local anaesthetic using a traditional two-stage procedure. In January 2000 two 17º angulated abutments (Fig. 1) (Nobel Biocare UK Ltd, Nobel House, The Grand Union Office, Packet Boat Lane, Uxbridge UB8 2GH, UK) were placed and a two-unit porcelain fused to metal restoration inserted a month later. The bridge conformed to the existing canine-guided occlusion with posterior disocclusion in protrusion. It was retained with gold slotted screws inserted at 10 Ncm torque.

Figure 1
figure 1

Peri-apical radiograph following abutment placement in January 2000

Regular standard reviews were uneventful until June 2005, when prosthetic screw loosening was noted but no problems were detected with the abutments. At this time the original prosthetic screws were replaced with titanium 'Torquetite' screws (Nobel Biocare UK Ltd, Nobel House, The Grand Union Office, Packet Boat Lane, Uxbridge UB8 2GH, UK) at the recommended increased insertion torque of 15 Ncm. Unfortunately, the patient presented four months later complaining of bridge loosening.

The prosthesis was removed again when it was observed on this occasion that the distal fixture (27) had fractured at the alveolar bone margin with the abutment screw still intact within the abutment (Fig. 2).

Figure 2
figure 2

Removed coronal portion of fractured fixture still attached to abutment

On inspection, the cross-sectional surface of the fractured fixture appeared homogenous with no obvious gross voids or other signs of manufacturing imperfections. There were however a number of obvious cracks across its diameter.

As an immediate measure, an acrylic bridge retained by temporary cylinders was provided, with canine guidance, posterior disocclusion in protrusion and light contacts in intercuspal position (ICP); the latter being coincidental with retruded contact position (RCP). No attempt was made to engage the temporary cylinder on the fractured implant surface at this stage.

An intra-oral peri-apical radiograph (Fig. 3) obtained at this time showed no change in bone levels at the non-fractured fixture (26). However, there appeared to be further bone loss associated with the fractured implant (27).

Figure 3
figure 3

Peri-apical radiographs of implants following removal of coronal portion of fractured fixture and the placement of a temporary bridge

At this stage the following management options were considered and discussed with the patient:

  1. 1

    Accept the missing second molar, and provide a new single unit restoration to replace the first molar using the intact implant

  2. 2

    Provide a replacement implant retained bridge supported by two fixtures, using the intact fixture and either:

    • Remove and replace the fractured fixture

    • Salvage the fractured fixture

    • Accept the fractured fixture and insert a new implant more mesially, resulting in a cantilever two implant supported bridge design

  3. 3

    Provide an implant retained bridge using the intact fixture together with an additional two implants to provide a three fixture supported restoration. This could be achieved by placing an additional fixture together with either:

    • Removal and replacement of the fractured fixture

    • Salvaging the fractured fixture.

A decision was made to salvage the fractured fixture and provide a three implant retained bridge by placing an additional implant between the existing two fixtures.

As a result, under a local anaesthetic, the fractured fixture was exposed by a punch excision of the overlying mucosa. The fractured surface was then refaced to provide a flattened area at right angles to the long axis of the fixture so a connecting abutment would more easily locate. This recontouring was accomplished by use of the Endmill Bur (Nobel Biocare UK Ltd, Nobel House, The Grand Union Office, Packet Boat Lane, Uxbridge UB8 2GH, UK), which consists of cutting flutes and a central guide pin for positioning in the fixture thread to ensure correct location of the cutting flutes. The bur was used with a slow speed of 20 revs. per minute and a 40 Ncm torque, with copious irrigation and lubrication (Fig. 4). An additional Brånemark, Mark IV, 15 mm long, wide platform fixture with an enhanced titanium oxide surface (Tiunite Nobel Biocare UK Ltd, Nobel House, The Grand Union Office, Packet Boat Lane, Uxbridge UB8 2GH, UK) was placed off-centre, between the existing implants.

Figure 4
figure 4

Refaced, smooth surface of retained portion of fractured fixture

A customised abutment was also placed on the fractured fixture at this time. This latter component incorporated a 17º Multi-Unit abutment with a customisable gold prosthetic screw (Fig. 5), which was shortened to an appropriate length. Beneath the abutment a customised cylindrical titanium connecting abutment was inserted which replaced the fractured portion of the fixture. This connecting abutment also had a TiUnite surface.

Figure 5
figure 5

Customised abutment incorporating connecting abutment and gold screw

A 17º Multi-Unit abutment was placed on the implant in the first molar site and a straight 1 mm Multi-Unit abutment on the more central wide platform fixture (Figs 6 and 7).

Figure 6
figure 6

Abutments secured onto resultant three fixtures

Figure 7
figure 7

Peri-apical radiograph after salvaging the fixture, the placement of additional implant and abutment connection

Impressions were recorded with abutment copings and an acrylic screw retained interim bridge inserted using temporary cylinders. The bridge conformed with the canine guided lateral excursions and had light contacts in ICP with posterior disocclusion in protrusion.

A definitive bridge was subsequently fabricated from a milled titanium framework and inserted three months later (Fig. 8). This restoration copied the occlusal features of the temporary restoration. During the period between insertion of the temporary and definitive restorations, no further complications were observed.

Figure 8
figure 8

Definitive bridge in place

The patient has been followed-up regularly and no problems have developed.

Discussion

The fracture of a dental implant is an uncommon occurrence with most studies reporting an incidence of between 0-1% in Brånemark fixtures (Table 1).2,3,4,5,6,7,8,9,10 An early study reported an incidence of 3.5%1 but this may have been due to the inclusion of implants inserted whilst the technique was being developed and the longer maximum follow up period of 15 years.1 Implant fracture occurs at all levels of the fixture, usually at around five years after insertion, and with the majority in the maxilla.1 The literature also reports a 1.9 % fracture rate in 157 ITI implants (Straumann AG, Waldenburg, Switzerland) that were followed for up to two years after loading.11 Case reports and case series of fractured ITI,12,13 and Core-Vent (Zimmer Dental, 1900 Aston Avenue, Carlsbad, CA 92008-7308.USA) fixtures also exist.14,15

Table 1 Incidence of dental implant fracture

Potential causes of implant fracture

The main factors which may lead to a risk of implant fracture appear to be:

  • Bending overload16

  • Manufacturing imperfections8

  • Restoration design17

  • Accuracy of fit of restoration1

  • Implant numbers, dimensions and positioning17,18

  • Marginal bone loss1

  • Occlusion and parafunctional habits8,12,17

  • Chemical factors.19

Bending overload

Bending overload is defined as the load on an implant-supported prosthesis that exerts a bending moment on the fixture cross-section at the crestal bone level, leading to marginal bone loss and/or implant fatigue fracture.16 Previous case series and reports, involving scanning electron micrograph (SEM) investigations of fractured implant surfaces, revealed fatigue striations10,20,21 consistent with those obtained in vitro when implants were cyclically loaded to fracture.20 This would suggest that bending overload and resultant fatigue failure is the usual mode of fracture of an implant. Some of the SEMs of the fractured implant in this case had a similar appearance to those in previously reported cases (Fig. 9).

Figure 9
figure 9

Scanning electron micrograph of fractured section of removed coronal portion of fractured fixture × 15

Manufacturing imperfections

There are no reports of implant fractures due to manufacturing imperfections, with previously reported SEMs of fractured implant surfaces failing to show any characteristic defects. 8,10,12,20,21

Similarly, the SEM of the fractured implant surface in this case revealed uniformity of the microstructure and a homogenous grain structure which would suggest an absence of any manufacturing error (Fig. 9).

Restoration design

Cantilever design bridges increase the stress upon an implant, and have been found to be associated with fractured implants (Table 2).8,10,17,20,22Fractured fixtures also appear to occur more frequently in single implant supported, single unit restorations (Table 2).10,17,12,13,14,22 Implant fractures associated with a combined dento-implant supported restoration have also been reported (Table 2).5,15,17,21The reports relating fractured implants to restoration design either are retrospective or this information has been missing, or are case reports. It is therefore unknown if the distribution of implant fractures, with regard to restoration type, is simply a reflection of the cohorts investigated. However, from these reports it does appear fractured implants in partially edentulous cases are often associated with cantilever, single tooth, or dento-implant supported restorations.

Table 2 Factors associated with fractured implants

Although the bridge in this case was not of cantilever design, molar units are larger than the supporting fixture platforms so some cantilever forces may have developed within the restoration. For this reason some clinicians advocate the placement of two fixtures for each molar unit replaced.23 A decision was made to reduce the potential for a cantilever effect in the new restoration by placing an additional fixture.

Accuracy of fit of restoration

Previous studies have shown implant fracture in partially edentulous fixed prostheses occurs with older, less passively fitting prostheses.1,24,25,26 A non-passive fit can also lead to screw loosening (Table 2).4,8,10,17,21

Although this patient's original bridge appeared clinically acceptable, it was constructed using a cast metal framework which may have increased the risk of a less than optimum fit. In addition screw loosening had been observed immediately prior to the implant fracturing. However screw loosening has been reported in 5-25% of patients during routine implant review.4,27As such this is not, by itself, necessarily a sign of impending implant fracture. However, if it does occur the prostheses should be re-evaluated as well as other design and occlusal features.28

It is anticipated that the new bridge, which was constructed from a milled titanium framework, provides a more accurately fitting prosthesis by eliminating errors that may arise in the casting process.29The bridge was completed by veneering with composite rather than porcelain to avoid possible firing deformation and also to make the restoration more serviceable in the future.

Implant numbers, dimensions and positioning

Theoretical models suggest the effects of loading on implant supported restorations can be significantly reduced by the placement of additional implants, the use of wider platform fixtures and the avoidance of implants being positioned in a straight line.17,18 In previous reports, fixtures have fractured when a straight-line arrangement of the implants has been adopted (Table 2).8,17,20Therefore, in this case, a decision was made to support the new bridge with three implants rather than two, with the additional central implant being a wide platform and positioned off centre in an attempt to improve load distribution (Fig. 6).

Marginal bone loss

Recently, figures for acceptable bone loss associated with fixtures of various designs and loading protocols have been proposed.30 However, in edentulous arches, the mean acceptable maximum value for marginal bone loss around Brånemark Mark II implants restored following the original two stage protocol is 1.2 mm during healing and the first year after bridge connection, and 0.1 mm annually thereafter.1,31 Lekholm and co-workers reported a mean bone loss of 0.7 mm over a ten-year period in partially edentulous jaws.3

In the case of this patient, the marginal bone level around the fractured fixture was eventually up to the fifth thread, if both the removed and retained portions of the fractured fixture are considered. At the intact fixture (26) site, there was no change in marginal bone level since initial loading. At the latter site this was acceptable, whilst at the fracture site the amount of bone loss was excessive.

Increased marginal bone loss has also been associated with implant fractures (Table 2).8,10,12,14,17,20,21 In the 15-year study by Adell et al. in 1981, fixtures that had rapid bone loss of approximately 3 mm a year all presented with eventual mechanical complications such as screw, fixture and bridge fractures.1 This would suggest that any rapid marginal bone loss should be investigated for possible mechanical complications, including fixture fracture.

Increased marginal bone loss has also been associated with excessive loading and this may occur within a year of the problem being initiated.4,32 Typically, this is seen on radiographs, as a crater-like pattern of bone loss around the fixture.33 In this case, until the time of fracture, there had been no evidence of excessive bone loss at any of the implant sites.

However, it is unclear whether marginal bone loss is a cause or an effect of implant fracture, or if they are both consequences of unfavourable loading. There is also the possibility that when an implant begins to fracture, it becomes secondarily infected which causes the accelerated marginal bone loss.

Whatever the explanation, any significant bone loss at the crestal level increases the unsupported coronal length of the implant, increasing the crown to implant ratio, resulting in the fixture becoming more at risk to bending forces.

Where excessive bone loss has occurred, the weaker hollow portion of the implant, apical to the abutment screw may become the fulcrum for the loading. This area is obviously weaker and as such, at this stage the implant may become particularly vulnerable to fracture.20,22This may particularly be the case in fixtures whose design features incorporate a significant void in this region.12,13,14

Histological analysis of fractured implant portions retained in bone reveal a high degree of osseointegration in the retained portion, apical to marginal bone levels.12,15In this case, radiographic examination of the retained portion showed close apposition of bone around the implant periphery, apical to the marginal bone level, indicating a high degree of osseointegration.

Occlusion and parafunctional habits

Implant fracture has been associated with parafunction but there is no significant scientific evidence to support this increased risk (Table 2).8,12,17 Although the patient reported no history of parafunction, and there were no clinical signs suggesting any habit, parafunction can not be eliminated as a contributory factor. The patient reported that his natural upper left molars had been extracted since they were heavily broken down and this may have been a consequence of parafunction. The patient was also in a high stress profession which may also have contributed to the problem.

During the new reconstruction, however, care was taken to avoid excessive occlusal forces by ensuring that all restorations achieved light contacts in ICP with canine guidance and disocclusion on protrusion and excursions. The staggered placement of three implants, with the additional implant being wide platform, also potentially distributed the occlusal loads more favourably. An acrylic occlusal splint has also been provided for night time wear.

Chemical factors

Titanium implant components adsorb hydrogen in the biological environment and it has been suggested that this makes them more brittle and prone to fatigue.19 A further case report of a fractured titanium implant supporting a nickel-chromium-molybdenum alloy restoration, suggested cytotoxic nickel ions may have accelerated bone resorption. The authors suggested that this resulted in the implant becoming more prone to bending and fatigue fracture.13

However, there is currently little additional evidence to support these suggestions as major contributors to implant fracture.

Planning considerations

A decision was made to retain the fractured implant as a significant length of fixture had maintained osseointegration and its removal would have resulted in bone loss. Furthermore, there is a possibility that any replacement fixture may not have integrated.

Salvaging a fractured dental implant is not always feasible. In this case the fracture level was favourable, with sufficient screw threads available in the retained portion to locate the Endmill bur's guide pin and to secure an abutment screw. If this had not been the case then retaining the fractured fixture as a viable functioning unit would not have been possible.

Conclusion

Dental implants are a predictable method of replacing missing teeth. When complications occur, consideration should be given to potential causes and how they can be overcome.

Implant fracture is a rare cause of failure but when it does occur it can present significant treatment planning and technical challenges. When considering remedial treatment following a fracture consideration should be given to the implant numbers, their design and features as well as their position. In addition the design of any restoration should aim to reduce unfavourable loading, particularly cantilever forces and ensuring a passively fitting framework.