Abstract
BACKGROUND:
The aim of this cohort, prospective study was to compare the diagnostic value of intrapartum fetal pulse oximetry (FPO) with that of fetal scalp blood gas (FSBG) for an abnormal neonatal outcome in cases with abnormal fetal heart rate (FHR) tracings.
METHODS:
Fetal oxygen saturation was continuously monitored with Nellcor N-400 FPO during labor. Simultaneous FSBG determinations were obtained. The results were analyzed in relation to umbilical arterial cord blood pH and neonatal outcome. Studied FPO cutoff levels were 30 and 40% hemoglobin saturation and that of FSBG pH was 7.2.
RESULTS:
During the study, there were 9825 deliveries; 415 had abnormal FHR. Only 150 fulfilled the whole screening panel. When the outcome variable was umbilical arterial pH, the positive predictive values of the three methods (FPO30, FPO40, FSBG) were 57, 61 and 65% and the negative predictive values were 43, 39 and 35% respectively. The sensitivity of FPO30 was highest (75%). Considering abnormal neonatal outcome, again the sensitivity was also highest for FPO30 (89%). The sensitivity of FSBG was 82%. The specificity of the three methods were 53, 49 and 38% respectively.
CONCLUSION:
The diagnostic value of intrapartum FPO compares favorably with FSBG. FPO seems to be a reliable and less invasive tool and may decrease unnecessary interventions and unnecessary fetal scalp blood sampling in cases of suspected fetal distress. The FPO cutoff of 30% saturation defined by previous studies appears to be appropriate.
This is a preview of subscription content, access via your institution
Access options
Subscribe to this journal
Receive 12 print issues and online access
$259.00 per year
only $21.58 per issue
Buy this article
- Purchase on Springer Link
- Instant access to full article PDF
Prices may be subject to local taxes which are calculated during checkout
Similar content being viewed by others
References
Prentice A, Lind T . Fetal heart rate monitoring during labor. Lancet 1987;2:1375–1377.
MacDonald D, Grant A, Sheridan-Pereira M, Boylan P, Chalmers I . The Dublin controlled randomized trial of intrapartum fetal heart rate monitoring. Am J Obstet Gynecol 1985;152:524–539.
Boog G . Acute fetal distress. J Gynecol Obstet Biol Reprod 2001;30(5):393–432.
Saling E . Fetal scalp blood analysis. Am J Obstet Gynecol 1985;153:717–720.
Clark SL, Paul RH . Intrapartum fetal surveillance: the role of fetal scalp blood sampling. Am J Obstet Gynecol 1985;153:717–720.
Van den Berg P, Schmidt S, Gesche J, Saling E . Fetal distress and the condition of the newborn using cardiotocography and fetal blood analysis during labor. Br J Obstet Gynecol 1998;94:72–75.
Chua S, Yam J, Razvi K, Yeong SM, Arulkumaran S . Intrapartum fetal oxygen saturation monitoring in a busy labour ward. Eur J Obstet Gynecol Reprod Biol 1999;82(2):185–189.
Gardosi JO, Schram CM, Symonds EM . Adaptation of pulse oximetry for fetal monitoring during labor. Lancet 1991;1:1265–1267.
Johnson N, Johnson VA, Fisher J . Fetal monitoring with pulse oximetry. Br J Obstet Gynaecol 1991;98:36–41.
Dildy CA, Clark SL, Loucks CA . Preliminary experience with intrapartum fetal pulse oximetry in humans. Obstet Gynecol 1993;81:630–635.
MacNamara H, Chung DC, Lilford R, Johnson M . Do fetal pulse oximetry readings correlate with cord blood oxygenation and acidemia? Br J Obstet Gynaecol 1992;99:735–738.
Carbonne B, Audibert F, Segard L, Sebban E . Fetal pulse oximetry: correlation between changes in oxygen saturation and neonatal outcome. Eur J Obstet Gynecol Reprod Biol 1994;57:73–77.
Langer B, Boudier E, Haddad J . Fetal pulse oximetry during labor of 62 patients. Fetal Diagn Ther 1996;11:37–45.
Dildy GA, Katz M, Clark SL, Jongsma SW . Intrapartum fetal pulse oximetry, fetal oxygen saturation trends during labor and relation to delivery outcome. Am J Obstet Gynecol 1994;171:179–184.
Skoczylas M, Laudanski T . Usefulness of the examination of fetal blood oxygen saturation (FSpO2) and fetal heart rate (FHR) as a prognostic factor of the newborn outcome. Ginekol Pol 2003;74(10):1284–1289.
Reuss JL . Factors influencing fetal pulse oximetry performance. J Clin Monit Comput 2004;18(1):13–24.
Carbonne B, Langer B, Goffinet F, et al. Multicenter study on the clinical value of fetal pulse oximetry. Am J Obstet Gynecol 1998;178(5):1100–1101.
Garite TJ, Dildy GA, McNamara H, et al. A multicenter controlled trial of fetal pulse oximetry in the intrapartum management of nonreassuring fetal heart rate patterns. Am J Obstet Gynecol 2000;183(5):1049–1058.
Luttkus AK, Dudenhausen JW . Fetal pulse oximetry. Clin Obstet Gynecol 1996;10:295–306.
Schmidt S, Koslowski S, Sierra F, Meyer-Wittkopf M, Heller G . Clinical usefulness of pulse oximetry in the fetus with non-reassuring heart rate pattern? J Perinat Med 2000;28(4):298–305.
Seelbach-Gobel B, Heupel M, Kuhnert M, Butterwegge M . The prediction of fetal acidosis by means of intrapartum fetal pulse oximetry. Am J Obstet Gynecol 1999;180(1 Part 1):73–81.
Westgate J, Garibaldi JM, Greene KR . Umbilical cord blood gas analysis at delivery: a time for quality data. BJOG 1994;101(12):1054–1063.
Van den Berg P, Schmidt S, Gesche J, Saling E . Fetal distress and the condition of the newborn using cardiotocography and fetal blood analysis during labor. Br J Obstet Gynecol 1998;94:72–75.
Kuhnert M, Schmidt S . Intrapartum management of nonreassuring fetal heart rate patterns: a randomized controlled trial of fetal pulse oximetry. Am J Obstet Gynecol 2004;191(6):1989–1995.
Author information
Authors and Affiliations
Additional information
Source of Support: None, institutional resources.
Rights and permissions
About this article
Cite this article
Bakr, A., Al-Abd, M. & Karkour, T. Fetal Pulse Oximetry and Neonatal Outcome: A Study in a Developing Country. J Perinatol 25, 759–762 (2005). https://doi.org/10.1038/sj.jp.7211406
Published:
Issue Date:
DOI: https://doi.org/10.1038/sj.jp.7211406