Introduction

Cataracts frequently form after the use of silicone oil as a tamponade agent in vitreoretinal surgery.1,2,3,4,5,6,7 In some patients, removal of silicone oil can be combined with phacoemulsification of the cataract and intraocular lens (IOL) implantation.1,2 thus avoiding subsequent surgery. Accurate determination of IOL power can present a dilemma since the true axial length (AL) of the eye is often unknown, and the echographically measured AL (MAL) of the silicone oil-filled eye is greater than the true AL.8,9,10 We have previously established a conversion factor of 0.71 which allows estimation of the true AL from the MAL so that desired IOL power can be accurately estimated despite the presence of silicone oil.8 Whilst most biometry machines allow a conversion of the speed of sound through vitreous to 987 metres/second, the absorption of sound waves through the silicone oil and the resultant loss of sensitivity often makes it impossible to get an accurate A-scan. The object of this study was to test the accuracy of our conversion factor in clinical practice.

Materials and methods

Eleven phakic eyes with silicone oil (Oxane®, Chauvin, France; viscosity 1300 centistokes) in the posterior segment were studied. Nine eyes had undergone successful retinal reattachment surgery (five for rhegmatogenous retinal detachment and four with traction retinal detachment complicating proliferative diabetic retinopathy), and two eyes had undergone macular hole surgery. All eyes had subsequently developed visually significant lens opacity. MALs were obtained, using A-mode echography (10 MHz central frequency, Quantel BVI AXIS biometry apparatus). With the patient in the upright position, the transducer was oriented so that the ultrasound beam was perpendicular to the globe. B-mode echography was also performed to ensure that the MAL measurement obtained with the A-scan was accurate and that the retinal spike seen was not an artefact. The MAL was multiplied by the conversion factor of 0.71 in order to estimate the CAL of the eye. The CAL was then inserted into the SRK/T biometry formula, and IOL power for the desired postoperative refraction determined. Refraction was carried out using manual retinoscopy in all patients.

Surgical technique

All patients had combined removal of silicone oil, phacoemulsification and intraocular lens implant. The IOL was then implanted into either the capsular bag—Group 1 (seven patients) or the ciliary sulcus—Group 2 (four patients). A variety of foldable IOLs were used (Table 1), including Rayner rigid polymethylmethacrylate (PMMA) (one patient), Alcon Acrysof soft acrylic (four patients), Storz Hydroview (five patients), and Allergan SI40 silicone (one patient). The corresponding A constant for each IOL was employed in the SRK/T biometry formula.11

Table 1 Predicted and actual refraction after combined removal of silicone oil, phacoemulsification and intraocular lens implantation

Results

Measured axial length (MAL) and calculated axial length (CAL) (Table 2)

Table 2 Measured axial length (MAL) and calculated axial length (CAL) in 11 phakic silicone oil-filleda eyes

MAL ranged from 29.76 mm to 40.49 mm (mean 34.02 mm). CAL (MAL × 0.71) ranged from 21.13 mm to 28.75 mm (mean 24.15 mm).

Time between combined cataract surgery and removal of silicone oil and refraction

The time between removal of silicone oil and refraction ranged from 2 weeks to 25 weeks (mean 8.27 weeks).

Predicted and actual postoperative refraction (spherical equivalent)—(Table 1)

Desired postoperative refraction ranged from + 0.31 D to −5.84 D and actual postoperative refraction ranged from +1.00D to −9.0 D. The difference between the suggested target and actual postoperative refraction ranged from 0 dioptres to 1.83 D (mean 0.74 D, SD 0.5 D) in Group 1, and 0.10 D to 3.16 D (mean 1.31 D, SD 0.98 D) in Group 2.

Discussion

Long-term complications including cataract, raised intraocular pressure, and band keratopathy, frequently occur after the use of silicone oil.1,2,3,4,5,6,7 One study showed that after one year, 30 out of 61 eyes (49%) exhibited some degree of clinically recognisable lens changes, which were not present before surgery.6 The incidence of this finding rose to 74% of 35 eyes assessed at 2 years. Another study showed that eyes that had undergone vitrectomy with silicone oil for severe proliferative diabetic retinopathy (PDR) had the least progression of PDR and the best prognosis for visual acuity when the silicone oil was removed, the cataract extracted and an IOL implanted at the same operation.12

Removal of silicone oil is therefore the preferred option, since the incidence of complications, causing sometimes irreversible deterioration of vision is reduced. Using the combined technique of cataract extraction, silicone oil removal and lens implantation, patients are often spared the burden of more than one operation.

At present, deciding the power of IOL to be used in eyes undergoing this combined technique poses a dilemma. Some advocate measuring the AL prior to every vitrectomy, but the AL may be different after surgery especially if encircling explants are used and measuring the AL accurately in eyes with extensive RD may be difficult. If pre-operative biometry is not done, then IOL power calculation is dependent on the usual parameters, namely the desired postoperative refractive error, the anterior chamber depth factor, the average corneal power, and most significantly, the AL of the eye.13 The presence of silicone oil renders the determination of AL difficult. AL measurements obtained from the contra-lateral eye or the AL of an average eye are therefore often used, but there are also limitations with these methods. In eyes that have silicone oil as a tamponade agent, there is an apparent increase in AL, as measured by echography.8,9,10 This is because the velocity of sound in silicone oil is slower than in vitreous humour, being 987 m/s in silicone oil of viscosity 1000 centistokes, compared to 1532 m/s in aqueous and vitreous.13 When the vitreous humour is replaced with silicone oil, it takes approximately 1.5 times as long for an ultrasound pulse to traverse the vitreous cavity.10 The depth of the vitreous cavity is therefore overestimated if allowance is not made for the accompanying reduction in the velocity of sound (Distance = velocity × time).

Whilst most biometry machines allow a conversion of the speed of sound through vitreous to 987 m/s, the absorption of sound waves through the silicone oil and the resultant loss of sensitivity often makes it impossible to get an accurate A-scan. In addition, the velocity of sound varies with the viscosity of the silicone oil, becoming increasingly slower as the viscosity increases. The viscosity of the silicone oil used in this study was 1300 centistokes. Sound velocity through this more viscous oil would therefore be less than through oil of viscosity 1000 centistokes, but the exact velocity is unknown. The conversion factor of 0.71 was determined in our previous study using silicone oil of viscosity 1300 centistokes.8 It was here used to estimate true AL from the MAL of 11 silicone oil-filled eyes.

This study has shown that it is possible to calculate the true AL in eyes with silicone oil by using a conversion factor of 0.71. This can minimise postoperative refractive errors resulting from inaccurate IOL power calculation (Table 1). In group 1, where IOL was placed in the bag, 5/7 eyes (71.4%) were within 0.56 D of predicted refraction. The median difference between predicted and actual refraction was only 0.5 D for this group. As expected this study also shows that sulcus placement of the IOL gives a less predictable result than placement in the capsular bag: the median difference between predicted and actual refraction in group 2 was 0.98 D. Patient 3, in group 2, had the largest difference between the predicted and actual refraction (3.16 D). The patient is a myope with an axial length of 28.75 mm and this factor may have contributed to the inaccurate biometry in this patient. Some authorities say that if an eye requires the use of silicone oil, the accuracy of biometry does not matter due to poor visual prognosis. This may be true in eyes with advanced PVR, but often silicone oil is used because of problems with patients posturing following procedures like macular hole repair. Therefore this is not a completely valid criticism. This conversion factor can only be used for silicone oil of viscosity 1300 centistokes. Oils of different viscosity will have a different constant, which can be determined as previously described.8