Skip to main content

Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Original Article
  • Published:

Evaluation of the A&D UA-767 and Welch Allyn Spot Vital Signs noninvasive blood pressure monitors using a blood pressure simulator

Abstract

The performance of five units of the A&D UA-767 NIBP monitor and five units of the Welch Allyn Spot Vital Signs noninvasive blood pressure monitor was evaluated with the Biotek BP Pump blood pressure simulator under a variety of conditions. Using the simulator to provide a normal blood pressure waveform at 80 bpm over a range of pressures, it was found that the mean bias for the combined results from the A&D monitors was 1.9±2.8 mmHg and from the Welch Allyn monitors was 0.7±2.4 mmHg. No individual measurement showed a bias greater than 10 mmHg. A bias of greater than 5 mmHg was present in 28 out of 150 measurements for the A&D monitor and 10 out of 150 measurements for the Welch Allyn monitor. These results are comparable with ratings achieved by the instruments when tested previously according to the British Hypertension Society protocol, but testing with a simulator allowed assessment of aspects of performance which were not included in the British Hypertension Society protocol.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Figure 1
Figure 2

Similar content being viewed by others

References

  1. Langford NJ, Ferner RE . Toxicity of mercury. J Hum Hypertens 1999; 13: 651–656.

    Article  CAS  Google Scholar 

  2. O'Brien E et al. Working group on blood pressure monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults. Blood Pressure Monit 2002; 7: 3–17.

    Article  Google Scholar 

  3. O'Brien E et al on behalf of the European Society of Hypertension Working Group on Blood Pressure Monitoring. Blood pressure measuring devices: validated instruments. BMJ 2001; 322: 531–536.

    Article  CAS  Google Scholar 

  4. Association for the Advancement of Medical Instrumentation. The National Standard of Electronic or Automated Sphygmomanometers. AAMI: Arlington, VA, 1987.

  5. O'Brien E et al. The British Hypertension Society protocol for the evaluation of automated and semiautomated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens 1990; 8: 607–619.

    Article  CAS  Google Scholar 

  6. O'Brien E et al. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993; 11(Suppl 2): S43–S63.

    Google Scholar 

  7. Association for the Advancement of Medical Instrumentation. American National Standard. Electronic or Automated Sphygmomanometers. AAMI: Arlington, VA, 1993.

  8. O'Brien E, Atkins N . A comparison of the BHS and AAMI protocols for validating blood pressure measuring devices: can the two be reconciled? J Hypertens 1994; 12: 1089–1094.

    CAS  PubMed  Google Scholar 

  9. Amoore JN, Geake WB . Evaluation of the Critikon 8100 and Spacelabs 90207 non-invasive blood pressure monitors using a test simulator. J Hum Hypertens 1997; 11: 163–169.

    Article  CAS  Google Scholar 

  10. Rogoza AN, Pavlova TS, Sergeeva MV . Validation of A&D UA-767 device for the self easurement of blood pressure. Blood Pressure Monit 2000; 5: 227–231.

    Article  CAS  Google Scholar 

  11. Jones CR, Taylor K, Poston L, Shennan AH . Validation of the Welch Allyn ‘Vital Signs’ oscillometric blood pressure monitor. J Hum Hypertens 2001; 15: 191–195.

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to P D Davis.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Davis, P., Dennis, J. & Railton, R. Evaluation of the A&D UA-767 and Welch Allyn Spot Vital Signs noninvasive blood pressure monitors using a blood pressure simulator. J Hum Hypertens 19, 197–203 (2005). https://doi.org/10.1038/sj.jhh.1001804

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/sj.jhh.1001804

Keywords

This article is cited by

Search

Quick links