Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Paper
  • Published:

Axial penile buckling forces vs Rigiscan™ radial rigidity as a function of intracavernosal pressure: why Rigiscan does not predict functional erections in individual patients

An Erratum to this article was published on 19 July 2000

Abstract

Aim: An improved understanding of the relationship between radial and axial rigdity values would enable better appreciation of the clinical usefulness of RigiScan™, the most widely utilized determination of erectile rigidity testing. Previous studies have shown that axial rigidity (measured by buckling forces) correlated well with radial rigidity (measured by RigiScan™) for radial rigidity values below 60%. For radial rigidity exceeding 60%, there was poor correlation. Heretofore, there has been no physiologic explanation of this phenomenon.

Methods: During dynamic pharmacocavernosometry in 36 impotent patients, we investigated the relationship between axial buckling forces and RigiScan™ radial rigidity and, for the first time, how they both vary with pressure, (which we varied over over a wide functional range). In addition, we recorded multiple penile length and diameter values enabling us to relate, also for the first time, axial and radial rigidity to individual mechanical erectile tissue and penile geometric properties.

Results: Marked differences were found in the manner RigiScan™ radial rigidity units and axial buckling force magnitudes increased with increases in intracavernosal pressure values in each individual. The former asymptotically approached a maximum finite value while the latter increased continuously towards infinity. Based on data in this study, RigiScan™ radial rigidity values greater than 55% may be considered a necessary criteria for vaginal intromission capability in all partners but it is not a sufficient one.

Conclusions: Axial and radial rigidity share a common dependency upon intracavernosal pressure, however, they are also dependent upon other unique physical determinants. For axial rigidity, additional dependent variables include cavernosal erectile tissue properties and penile geometry, while for radial rigidity, this may include tunical surface wall tension properties. Clinical devices which assess functional penile rigidity should utilize axial and not radial rigidity testing.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Udelson, D., Park, K., Sadeghi-Najed, H. et al. Axial penile buckling forces vs Rigiscan™ radial rigidity as a function of intracavernosal pressure: why Rigiscan does not predict functional erections in individual patients. Int J Impot Res 11, 327–337 (1999). https://doi.org/10.1038/sj.ijir.3900443

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/sj.ijir.3900443

Keywords

This article is cited by

Search

Quick links