Perinatal acute care is associated with specific clinical demands and challenges. The relatively small number of acute care cases occurring in many institutions, together with the presence of senior staff assuming the responsibility for their management, may hinder the establishment of adequate training for less experienced individuals. In other areas of acute care medicine, realistic simulators contribute to training of healthcare professionals and to patient safety.

Clinically relevant learning objectives for medical staff in Neonatology include the recognition of cardiovascular signs and symptoms occurring in pathologies that manifest shortly after birth, such as persistent pulmonary hypertension, left ventricular overload by a patent ductus arteriosus (DA), myocardial depression, and hemorrhage.

This article introduces a fundamental component of a perinatal acute care simulator, namely a simulation engine, for normal hemodynamic transitions from the fetal to the early neonatal period. A simulation engine can contribute to a life-like training environment by providing real-time, automatic evolution of clinical signs and monitored signals and of their response to therapeutic interventions. To be able to simulate normal physiology and provide a platform for future simulation of the above-mentioned incidents and pathologies, it is necessary that the engine reflect essential aspects of fetal hemodynamics and its subsequent transition to the neonate. Among others, structures such as lungs, placenta, heart, systemic arterial circulation, and the various fetal shunts need to be included. The simulation engine should be able to simulate realistic blood pressures and flow rates in these structures during fetal, transitional, and neonatal periods.

METHODS

The proposed simulation engine for normal hemodynamic transitions from the fetal to the early neonatal period consists of a fetal hemodynamic model controlled by a time- and event-based script (1). The script triggers events occurring at birth such as onset of breathing, cord clamping, and the transitory phase up to 24 h after birth, via changes to selected model parameters. The Instituto de Engenharia Biomédica, Porto, Portugal, and the Máxima Medical Center, Veldhoven, The Netherlands, approved this study.

Fetal hemodynamic model.

In this study, we used an adaptation of a previously described model for educational simulation of the neonatal circulation (2,3). The latter was based on the linearized, improved model of cardiovascular physiology, originally presented by Beneken (1965, PhD thesis), which is also the basis for the model of cardiovascular physiology of the human patient simulator (HPS; developed at the University of Florida and commercially available from Medical Education Technologies, Inc., Sarasota, FL) and for a recently published model of infant cardiovascular physiology (4). In the discussion section, we will come back to other fetal hemodynamic models.

To reflect the unique aspects of fetal circulation, the neonatal model was adapted as follows (Fig. 1):

Figure 1
figure 1

Hydraulic analog for the fetal cardiovascular model. Fetal shunts: 1—ductus arteriosus, 2—foramen ovale, 3—ductus venosus. Fetal circulation represented by 11 compliant compartments. Heart chambers compartments are represented by time-variant elastances to mimic heart contractility. Connections between compartments represent blood flow and include vessel resistance to the passage of blood. Heart valves are included in the atrial and ventricular outflows. Inertial behavior of the blood is included in the DA and in the intrathoracic arteries outflow.

  • the DA was included between the pulmonary arteries and the aorta (part of the intrathoracic arteries compartment);

  • the foramen ovale (FO) was included between the atria with a valve-like flap, and;

  • the placental circulation (including the ductus venosus) was placed in parallel with the systemic circulation, returning all flow to the intrathoracic veins compartment.

The proposed fetal hemodynamic model consists of two major parts: The heart and seven vascular compartments. We refer to previous studies by our group for a detailed description of the mathematics associated with implementation of the model and similar extensions (24).

Numerical values for most model parameters were obtained directly from the scientific literature (514). Numerical values for a few parameters that are specific to our model needed to be derived (Supplemental Methods, Supplemental Digital Content, http://links.lww.com/PDR/A57). Table 1 presents the complete parameter set for a full-term fetus with an assumed weight of 3 kg.

Table 1 Fetal cardiovascular model parameters (assumed weight: 3kg)

Scripted transitions at birth.

Scripted transitions can be automatic or controlled by the simulation instructor. Twenty-four hours after birth, transitions are complete and the model structure (but not all parameter values) matches the previously published model for a 1-week-old neonate. The following subsections describe the scripted model parameter changes in detail.

Intrathoracic pressure.

During fetal life, the bronchoalveolar tract is filled with fetal lung fluid, maintaining the lungs in a distended state and with an intrathoracic pressure of around 0 mm Hg (13,14). At birth, the fluid in the bronchoalveolar tract is replaced by air. In spontaneously breathing newborns, the diaphragm contracts during inspiration, thereby lowering the intrathoracic pressure toward negative values of approximately −3 mm Hg (14,15).

Pulmonary vascular resistance (PVR).

Immediately after birth, concurrent with the onset of breathing, PVR decreases abruptly (16). After this period, there is a more gradual decline in resistance over the following days and weeks (16,17). From the studies of Dawes et al. (18) in fetal lambs, it was possible to estimate a reduction in PVR of 74% after the start of ventilation. A neonatal PVR value between 0.75 and 0.9 mm Hg · mL−1 · s was reported by Brook et al. (19), and it was assumed that this is reached by the end of the first week of life. In our model, PVR is distributed over model parameters RPP and RLAIN (see Table 1 for abbreviations). An exponential was used to mimic the initial rapid decrease immediately after birth, and a linear function to reproduce the slower decrease thereafter, reported here only for the first 24 h of life:

Equation

where RPP0 is the resistance in pulmonary peripheral vessels for a normal-term fetus. Similarly, for RLAIN:

Equation

Systemic vascular resistance.

With cord clamping, total systemic resistance (TSR) approximately doubles (20). An initial rapid rise in TSR is followed by a slower, more sustained increase over time. The mechanisms behind this considerable increase are not fully understood, but it is thought that there are at least three contributing factors: removal of the low-resistance placental bed, catecholamine release (21,22), and a high sympathetic tone in the immediate postnatal period (23). Measurements of TSR in full-term human neonates during the first 24 h of life are available from several sources (2426). Based on these data, the following piecewise linear approximation was established:

Equation

Cord clamping is assumed to take place at 1 min. Only the peripheral vessel resistance [RSP(t)] is increased, whereas the resistances of extrathoracic arteries and veins (RETHA and RETHV) remain constant. TSR(t) corresponds to the sum of these three parameters.

DA resistance.

It is well known that the DA constricts rapidly after birth, with functional closure occurring within 12 to 36 h in the majority of term infants. Experimental data for DA diameter (DAD) changes in humans can be found in Rasanen et al. (11), for the full-term fetus, and in several other publications (2528) for term neonates, during the first 24 h of life. To model DA closure, we used a dynamic resistance that increases with neonatal age. The following curve of the DAD in centimeters, as a function of age t in hours, was fitted to these data (Sá Couto, unpublished data):

Equation

Migliavacca et al. (29) present an application of Poiseuille's law to calculate ductus arteriosus resistance (DAR) as a function of its diameter:

Equation

where K is a proportionality constant in mm Hg · s · cm.

In our model, K was set to 0.0145 mm Hg · s · cm to match fetal DAR and DAD (0).

Software implementation.

Model equations were numerically integrated using the Euler forward method with a step size of 0.1 ms. All simulations were implemented in Matlab V7.5(R2007b; MathWorks, Inc.) on a personal computer with an 2.33-GHz Intel Core Duo processor.

RESULTS

To evaluate the realism of the described simulation engine, we compared simulation results with target data. This comparison was performed at three physiologically and educationally relevant stages: Term fetus, ventilation and cord clamping, and the evolution during the first 24 h of life.

Term fetus.

Table 2 presents target data and simulation results for the term fetus. Note that to facilitate comparison across species, all flow rates are normalized to fetal weight. All but one of the simulated flow rates fall within one SD of the mean target data obtained in the term fetal lamb (30,31). Pulmonary blood flow rate is the exception, but like the other flow rates, matches human target data (11). Combined ventricular stroke volume (CVSV) and output (CVO) also match human target data (8). Simulated arterial blood pressures are within one SD of animal target data (30,31). This comparison demonstrates that the fetal model has clinically appropriate ventricular outputs, flow rates, and pressures at baseline.

Table 2 Target data (8,11,30,31) and simulation results for the term fetus

Ventilation and cord clamping.

Table 3 presents target data (30) and simulation results for newborn ventilation and ventilation with cord clamping. Comments concerning the target data are included in the discussion section. The ventilation column shows that all simulated output variables are within one SD of the target data, with the exception of arterial pressures that are within two SDs. Ventilation with cord occlusion results in a correct reversal of DA flow, from a predominantly right-to-left to a predominantly left-to-right shunt, and in increased left ventricular output. All simulated output variables are within or at one SD of target data. In the discussion section, we will analyze the single exception, DA right-to-left flow.

Table 3 Target data (30) and simulation results for transitions shortly after birth

Evolution during the first 24 h of life.

Table 4 presents target data (32,33) and simulation results for left ventricular stroke volume and output from the term fetus until the 24-h old neonate. With one exception, all simulation results are within one SD of target data. Left ventricular output at 12 h is one SD removed from the target data.

Table 4 Target data (32,33) (mean ± SD) and simulation results of the left ventricle performance from before birth to 24 hours after birth

Table 5 presents target data (34,35) and simulation results for arterial and atrial pressures 24 h after birth. All but one of the simulation results are within one SD of target data. Pulmonary arterial diastolic pressure is at one SD.

Table 5 Target data (34,35) (mean ± SD) and simulation results of blood pressures at 24 hours after birth

Figure 2 shows target data (36) and simulation results for the evolution of the mean aortic-pulmonary pressure difference during the first 24 h of life. These results demonstrate a plausible evolution of the simulated variable. The pulsatile nature of the model is illustrated in Figure 3, depicting the evolution of systemic and pulmonary arterial pressures in the fetus and at 2, 12, and 24 h after birth, respectively.

Figure 2
figure 2

Mean aortic-pulmonary pressure difference in the first 24 h after birth. Open circles: Target data (36) from 39 normal-term human infants; Closed dots and line: Simulation results.

Figure 3
figure 3

Simulated systemic (heavy line) and pulmonary (light line) arterial blood pressure before birth and at 2, 12, and 24 h after birth.

Figure 4 provides an overview of all simulated phenomena by presenting the scripted model parameters and the evolution of modeled flow rates and pressures from the fetus until the 24-h old neonate. With the onset of breathing, intrathoracic pressure decreases abruptly to negative values, as observed in spontaneously breathing newborns. In addition, PVR decreases considerably in the first minutes after birth (bottom panel). Consequently, arterial pressures increase and atrial pressures decrease (middle panel), but no relevant changes are observed in flow rates (top panel). An immediate result of cord clamping is the sudden increase in systemic resistance (bottom panel), causing an increase in systemic arterial blood pressure (middle panel) and the cessation of placental blood flow (top panel). Flow rate in the DA and pulmonary blood flow (top panel) are deeply affected by cord clamping. The continuous decrease of pulmonary resistance (bottom panel) further increases pulmonary blood flow (top panel). The increase in systemic resistance (bottom panel) and subsequent increase in left atrial pressure (middle panel) cause—at near constant right atrial pressure (middle panel)—a decrease in flow across the FO (top panel). The combination of increased left atrial pressure, because of higher pulmonary venous return, and diminished inferior vena cava return to the right atria results in a reversal of the left and right atrial pressures (middle panel), functionally closing the FO, as can be observed around minute 30 (top panel). From this time on, pulmonary inflow matches left ventricle flow rate (top panel). As PVR falls below systemic vascular resistance (bottom panel), the bidirectional DA flow decreases significantly, changing to a continuous left-to-right flow at the third hour of life (top panel). These changes in DA flow cause the left ventricle to pump more blood than the right ventricle. As DA resistance increases over time (bottom panel), flow through the ventricles and lungs converge (top panel). Another consequence of changes in DA flow is the increasing difference between mean systemic and pulmonary arterial pressures, which can be observed between the second and 24th h after birth (middle panel).

Figure 4
figure 4

Simulated model parameter changes and output variables in the first 24 h after birth. RV, right ventricle; DA (R-L), ductus arteriosus, right to left flow; LV, left ventricle; PL, placenta; FO, foramen ovale; PA, pulmonary arterial; DA (L-R), ductus arteriosus, left to right flow; SA, systemic arterial; RA, right atrium; LA, left atrium; PVR, pulmonary vascular resistance; TSR, total systemic resistance; DAR, ductus arteriosus resistance; PTH, mean intrathoracic pressure.

DISCUSSION

We present a combination of a mathematical model and a time- and event-based script for educational simulation of fetal to early neonatal hemodynamic transitions. The pulsatile cardiovascular model presented by Beneken (1965, PhD thesis) was selected as a basis for this work because of its proven track record in simulating adult, infant, and neonatal cardiovascular physiology for educational simulation (14). Inclusion of the DA, FO, and placental flow was relatively straightforward. Other models of fetal cardiovascular physiology have been described in the past (6,10), but because of their complexity and large parameter sets were judged to be difficult to apply in our envisioned educational application, which involves repeated parameter estimation and manipulation to represent different patients, pathologies, and critical situations. The relatively simple circulation model presented by Huikeshoven et al. (7) has a complex heart model. The structure of the presented model for a 24-h-old neonate is identical to that of a previously described 1-week-old neonate (2), but parameters differ considerably. This should in principle be exclusively because of the age difference, but an evaluation of the evolution from 24 h to 1 week is needed to confirm this statement.

Simulations results demonstrate that, in general, the model is capable of correctly representing fetal hemodynamics and its evolution during the first 24 h of life. The validation of events occurring shortly after birth (ventilation and cord clamping) is limited by the available experimental data, which do not completely represent normal physiologic conditions. Target data used for this purpose includes three sequential events: ventilation, oxygenation, and umbilical cord occlusion. Each event was continued for 15 min, as described in the footnote of Table 3. To reproduce these conditions, a few adaptations in scripted parameters were needed: 1) matching to reported timeline; 2) representation of artificial ventilation with a positive average intrathoracic pressure of +2 mm Hg (instead of −3 mm Hg). Intrathoracic pressure is positive throughout inspiration and remains slightly positive, whenever positive end-expiratory pressure is applied (15,37); and 3) delayed rise in DA resistance, corresponding to the initiation of ventilation with 100% oxygen. In these conditions, simulation results match the target data. The only exception is right-to-left DA flow during ventilation with cord clamping, which is higher than in target data. This may be explained by differences between species. Flow through the DA after birth may not be the same in the human neonate and in the newborn lamb. In the latter, the PVR apparently falls sharply within minutes after the onset of respiration, and the fetal direction of flow suddenly becomes reversed (38).

Figures 3 and 4 illustrate potential educational applications of the model and show its integrated response. Evolution of aortic and pulmonary artery pressures can be simulated in accelerated time, providing students with a unique experience of these important yet complex phenomena.

Potential limitations may result from the lack of sufficient experimental data and heterogeneity of available data to provide model parameters and compare simulation results with independent target data. However, despite different flow measurements techniques in animal and in human subjects, the simulation results provided fairly consistent values of the relative flow rates and pressures in the different parts of the circulation.

Manipulation of specific model parameters or inclusion of new input variables is straightforward and allows for simulation of relevant clinical acute scenarios. An increase in pulmonary peripheral resistance can be used to mimic persistent pulmonary hypertension and simulate a right-to-left ductal shunt. Conversely, combining a decrease of this parameter with a persistent DA leads to a significant left-to-right shunt with volume overload of the left heart and systemic hypotension as clinically relevant consequences. Another illustrative example is the use of the ventricular time-variant elastances to simulate myocardial depression. We refer to a previous article from our group, where several congenital heart defects are modeled using the same methodology (2).

Expansion of this model with transport of oxygen and carbon dioxide seems to be the most pressing future work, because it would allow for the simulation of other clinically relevant variables, such as oxygen saturations or blood pH. As no baroreflex control mechanism has been incorporated yet, the current model precludes simulations of dynamic responses to blood loss or fluid expansion. Incorporation of natural control mechanisms (baroreflex and chemoreflex), congenital heart disease, or pathologies that may occur during the transitional period will further enhance its potential educational impact.

Envisioned educational applications for this model include interactive screen-based or full-body simulators. The specific interventions available in those simulators will be identified based on a more detailed training needs analysis and training program design. Such analyses are beyond the scope of this study. However, the selected model structure, variables, and parameters already anticipate a broad range of interventions. For instance, the inclusion of a mean intrathoracic pressure already anticipates the use of this model to train resuscitation procedures in a full-body simulator.

In conclusion, we present an original model for educational simulation of fetal to early neonatal hemodynamic transitions. Fetal and neonatal endpoints show realistic simulated hemodynamic variables. The transitions at birth and during the first 24 h of life are also represented realistically. This model constitutes an essential component of educational simulators for perinatal acute care. It can also be used to demonstrate the complex hemodynamic transitions at birth.