Abstract
Purpose Responses to topically applied ocular drugs vary between patients. The volume of drug instilled is of particular importance as one of many sources of response variation, but the reliability of drop volume from eye drop bottles is unknown. Hence, the repeatability of drop volume and factors affecting this for a variety of drug manufacturers were considered in this study.
Methods Nineteen bottles, one from each primary manufacturer in the UK, were examined. The mass of all drops expelled from each bottle was measured with respect to the bottle type, handling angle, drop number, drug and concentration. The accuracy (repeatability and trueness) of drops from each bottle was also evaluated.
Results Drop volume varied significantly between drug manufacturers, ranging from 33.8 µl to 63.4 µl. The handling angle of the bottle also influenced drop volume, with angles less than 60° giving smaller drops. Drop number exhibited no significant effect upon drop volume. However, the drug type and its concentration did significantly affect the volume of the drop expelled from the bottle, with higher concentrations giving rise to larger drops. Repeatability coefficients across the range of bottles varied between ± 2.24 µl and ± 10.76 µl (mean ± 5.07 µl).
Conclusions It is well reported that drug volume instilled has a significant effect on the degree of response. However, there are currently no official regulations concerning eye drop volume in either the UK or the USA. Since drop volume has been shown to vary significantly depending upon a variety of factors, it may be appropriate that the regulatory bodies consider the consequences of variable drop size.
Similar content being viewed by others
Article PDF
References
Lund W . Pharmaceutical codex. 12th ed. London: Pharmaceutical Press, 1994:161–2.
File RR, Patton TF . Topically applied pilocarpine: human pupillary response as a function of drop size. Arch Ophthalmol 1980;98:112–5.
Gray RH . The influence of drop size on pupil dilatation. Eye 1991;5:615–9.
Patton TF . Pharmacokinetic evidence for improved ophthalmic drug delivery by reduction of instilled volume. J Pharm Sci 1977;66:1058–9.
Kessler C, Bleckmann H, Kleintges G . Influence of the strength, drop size and viscosity of metipranolol eye drops on the concentration of the substance in human aqueous humour. Graefes Arch Clin Exp Ophthalmol 1991;229:452–6.
Sugaya M, Nagataki S . Kinetics of topical pilocarpine in the human eye. Jpn J Ophthalmol 1978;22:127–41.
Brown RH, Hotchkiss ML, Davis EB . Creating smaller eyedrops by reducing eydropper tip dimensions. Am J Ophthalmol 1985;99:460–4.
Craig EW, Griffiths PG . Effect on mydriasis of modifying the volume of phenylephrine drops. Br J Ophthalmol 1991;75:222–3.
Fell JT . Surface and interfacial phenomena. In: Aulton ME, editor. Pharmaceutics: the science of dosage form design. Edinburgh: Churchill Livingstone, 1988:53–5.
German EJ, Hurst MA, Wood D . Eye drop container: a source of response variation? Ophthalmic Physiol Opt 1997;17:196–204.
Green K . The role of surfactants as bactericides in topical drug delivery. STP Pharm Sci 1992;2:34–8.
British Standard ISO-5725-1 Accuracy (trueness and precision) of measurement methods and results. Part 1. General principles and definitions, 1994.
Bland JM, Altman DG . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;l:307–10.
Bland IM . An introduction to medical statistics. 1st ed. Oxford: Oxford University Press, 1993:118.
Personal communication, Technical Services, Merck Sharp Dohme Ltd.
Author information
Authors and Affiliations
Corresponding author
Additional information
The study was funded by a grant to E.J.G. from the College of Optometrists
Extracts of this work were presented at the Second International Symposium on Experimental and Clinical Ocular Pharmacology and Pharmaceutics, Munich, 1997
Rights and permissions
About this article
Cite this article
German, E., Hurst, M. & Wood, D. Reliability of drop size from multi-dose eye drop bottles: is it cause for concern?. Eye 13, 93–100 (1999). https://doi.org/10.1038/eye.1999.17
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1038/eye.1999.17
Keywords
This article is cited by
-
Design and simulation of a piezoelectric micropump for drug delivery systems
Microsystem Technologies (2023)
-
Efficacy and safety of mydriatic microdrops for retinopathy of prematurity screening: an external pilot crossover randomized controlled trial
Journal of Perinatology (2022)
-
Glaukom – Erkrankung der Augenoberfläche („ocular surface disease“) – Augentropfen – Konservierungsmittel: Ein Überblick
Spektrum der Augenheilkunde (2021)
-
A study on the change of droplet volume by the increase of flow rate
SN Applied Sciences (2020)
-
A novel TRPM8 agonist relieves dry eye discomfort
BMC Ophthalmology (2017)