Introduction

Silicone oil is a valuable tool in the management of various forms of complex vitreoretinal disease. Raised intraocular pressure (IOP) remains, however, a significant complication of intraocular silicone oil with reported incidences of between 5.9 and 48% of eyes.1, 2, 3, 4, 5 When topical medication fails to control the IOP these eyes present a challenging management problem. Silicone oil removal may result in pressure control although recent reports indicate that additional antiglaucomatous medication will still be necessary in many eyes3, 4 and that even with such additional medication over 50% of eyes will have uncontrolled intraocular pressure.3, 6 Furthermore, removal of silicone oil may be contraindicated in eyes where there is a perceived high risk of retinal redetachment or postoperative hypotony.1

Reduction of aqueous production by ciliary body destruction is a potential method of IOP control. Cyclocryotherapy, while often effective in pressure reduction, is commonly associated with marked intraocular inflammation and is difficult to titrate with a high incidence of subsequent hypotony.7 Cyclodiode laser cyclodestruction offers a more controlled method of pressure reduction. This report documents the long-term follow-up (13–113 months, median 68 months) of 38 eyes which underwent trans-scleral diode laser ciliary body destruction to control raised IOP associated with intraocular silicone oil.

Patients and methods

Using a computerised database patients were identified who had undergone cycloablation by diode laser at Moorfields Eye Hospital during the period 1993–2000 and were followed up till January 2003. In total, 41 patients were identified all of whom had adequate data for retrospective analysis. Three patients who had less than 12 months follow-up were excluded and the remaining 38 records were analysed further.

Data were recorded retrospectively from case notes and included primary vitreoretinal diagnosis, number and type of vitreoretinal procedures, previous ophthalmic history, lens status, and topical and systemic medications used to control IOP. The number of cyclodiode treatments together with laser power, duration, and number of pulses were documented. IOP was recorded prior to cyclodiode, 1 day postlaser and at 12 months and final follow-up visit. Statistical analysis was carried out using paired t- and McNemar's tests. Raised IOP was defined as IOP greater than 21 mmHg.

Results

In all, 38 patients in total were identified who fulfilled the criteria for this study. The follow-up period ranged from 13 to 113 months (median 68 months).

Totally, 24 patients (63.2%) were males and the mean age in all patients was 53 years (range 21–85). Overall, 23 right eyes and 15 left eyes were treated.

The primary diagnosis included retinal detachment (RD) (17 patients), RD and proliferative vitreoretinopathy (PVR) grade C or greater (nine patients), RD secondary to trauma (seven patients), RD related to proliferative diabetic retinopathy (three patients), RD and macular hole (one patient) and RD due to giant retinal tear (one patient). The total number of vitreoretinal procedures patients underwent is shown in Figure 1, which demonstrates that overall 30 patients (78.9%) had one or two vitreoretinal (VR) procedures. All patients whose IOP was not controlled at 1 year (10) and final follow-up (5) had had one or two previous VR procedures.

Figure 1
figure 1

Number of vitreoretinal procedures.

At the time of first cyclodiode treatment, four (10.5%) patients were phakic, 16 (42.1%) patients were pseudophakic and 18 (47.4%) patients were aphakic. In total five patients had had a retinectomy: in two this extended to 90° and the remaining three it was between 90 and 270°. At the time of first cyclodiode 17 patients (44.7%) had silicone oil in situ. There was previous history of glaucoma in four patients (10.5%).

Cyclodiode laser was performed at a power of 1.5–2.0 W, for a duration of 1.5–2.0 s and with 20–60 applications. Overall, 18 patients (47.4%) required only one cyclodiode treatment and the remaining patients received up to four treatments to control their raised IOP (Figure 2). There was no statistically significant correlation between the laser power and the success in IOP reduction in this group of patients.

Figure 2
figure 2

Number of cyclodiode procedures.

Changes in IOP following cyclodiode, at 1 year and final follow-up are shown in Table 1. It shows that 28 patients (73.8%) at 1 year and 33 (86.8%) at their final visit had controlled IOP. Among those with uncontrolled IOP, seven out of 10 at 1 year and four out of five at last visit had IOPs of 30 mmHg or less. The mean IOP was 18.6 mmHg (SD 8.6) at 1 year and 13.9 mmHg (SD 8.4) at the final visit compared to the pre-op mean value of 31.4 mmHg (SD 10.9) (P<0.001, paired t-test). Hypotony (IOP of less than 5 mmHg) was found in four patients (10.5%) at 1 year and in five patients (13.2%) at the final follow-up one of whom was a trauma case who underwent enucleation 95 months after initial silicone oil injection and 75 months after cyclodiode treatment.The IOP at 1 year follow-up in this eye was 2 mmHg.

Table 1 IOP of the study eye, pretreatment, immediately post-treatment, and at 1 year and final visit

There were significant reductions in the numbers of patients using acetazolamide and topical hypertensive medications. In total, 28 patients were on two or more topical antihypertensive medications prior to their cyclodiode treatment. This was reduced to 18 patients at 1 year and 13 patients at the final follow-up visit (McNemar's P=0.013 and 0.0007, respectively). With regard to the systemic medications, preoperatively 16 patients were taking oral acetazolamide, whereas at 1 year five patients were on oral acetazolamide and at the final visit only three patients required this systemic treatment (Table 2). (McNemar's P=0.0034 and 0.001, respectively).

Table 2 Systemic medication pretreatment, 1 year and last visit

The relation between the lens and the silicone oil status and the IOP at 1 year and at final follow-up is shown in Table 3. There were 18 (47.5%) aphakic, 16 (42.1%), and four (10.5%) phakic patients at 1 year. These numbers did not change at the final follow-up.

Table 3 Lens and silicone oil status at 1 year and final follow-up

At 1 year, 15 patients (39.5%) had silicone oil in situ 12 of whom had normal IOP. The remaining three had IOPs of 22, 28, and 45 mmHg. By the final follow-up 13 (34.2%) patients had silicone oil in situ with 10 of them having normal IOPs. The remaining three had IOPs of 22, 24, and 44 mmHg.

There was no correlation between the duration of silicone oil in situ and the final outcome of controlled IOP following cyclodiode treatment.

Discussion

Secondary IOP elevation after silicone oil injection has proved to be a challenging problem to control medically. The incidence of silicone oil related glaucoma is reported to be between 5.9 and 48%.1, 2, 3, 4, 5 Known risk factors for silicone oil-induced raised IOP include pre-existing glaucoma, diabetes, and aphakia.2, 3 The mechanism of silicone oil related glaucoma may be due to migration of emulsified oil to the anterior chamber and into the trabecular meshwork, synaechial angle closure, rubeosis iridis, pupillary block, and inflammation, or to a combination of these. Medical treatment aims at reducing the aqueous production and inflammatory response using aqueous suppressants and steroids as well as cycloplegics. Prophylactic surgical inferior iridectomy reduces the risk of pupillary block glaucoma.

Although there are reports of normalisation of IOP following silicone oil removal8 this issue remains controversial and several previous reports suggest that removal of silicone oil alone may be insufficient to alleviate the glaucoma.6, 9, 10, 11, 12, 13 It is, however, generally accepted that where possible prompt removal of silicone oil will minimise its potential ocular side effects. This is not always desirable because of the risk of redetachment which has been reported in up to one in three patients9 particularly those with complex previous vitreoretinal pathology.14

Glaucoma surgery on these difficult cases is often unsuccessful due to pre-existing conjunctival scarring following vitreoretinal procedures.4 Glaucoma drainage implants offer an alternative surgical option but oil migration can occur through the tube into the subconjunctival space inciting an inflammatory reaction.15, 16 Cyclodiode treatment can provide good IOP control as reported by Bloom et al17 and Han et al18 with success rates of 66 and 82% respectively. Our data indicate a long-term success rate of 74% at 1 year and 86% at final follow-up of up to 113 months (median 68 months). Notably, our overall results show that of the five cases which had ‘uncontrolled’ IOP at final visit four had IOP of 30 mmHg or less. This may represent an adequate level of IOP control in eyes with complex multiple pathologies (and a poor visual prognosis). To our knowledge these are the longest follow-up data reported on silicone oil-related glaucoma patients.

In the present study, 17 patients (45%) had silicone oil in situ at the time of first cyclodiode treatment, 15 (39.5%) patients had oil in situ at 1 year and 13 (34%) patients at the final follow-up visit. At 1 year and final follow-up visits only three patients with oil in situ had IOPs greater than 21 mmHg of whom two had IOPs of less than 30 mmHg. Removal of silicone oil was performed in three patients following cyclodiode (two patients within the first year of treatment). Our long-term-follow up results (of up to 113 months) are similar to those in a previous report by Han et al18 where good IOP control was achieved with the silicone oil in situ up to 68 weeks after diode laser photocoagulation. Overall, these findings suggest that initial IOP control in these eyes is maintained long term.

Systemic antiglaucoma medication was necessary in only 13% of patients at 1 year and 8% at the final follow-up compared to 42% prior to cyclodiode. In addition to IOP control, cyclodiode treatment has an important role in reducing the need for the long-term use of systemic medication and the inevitable side-effects associated with this.

Chronic hypotony is a common problem following cyclodestructive procedures. Cyclodiode treatment does however carry a lower incidence of complications compared to cyclocryotherapy. Suzuki et al19 compared 36 eyes which had undergone cyclodiode with 80 eyes treated with cyclocryo. Three patients in the cyclodiode group (8%) had complications one of whom had hypotony whereas the rate of complication in the cyclocryo-treated group was 30% (24 patients, two with hypotony).19 Schuman et al20 reported a 10% incidence of hypotony and phthisis after Nd:Yag cyclophotocoagulation, which is similar to the 10.5% hypotony rate (including one case of enucleation) at the final follow-up (5% at 1 year) in this study. All patients in our study with hypotony had the silicone oil in situ at the final follow-up.

Cyclodiode treatment has the advantage of being repeatable if required and 52% of the patients in this study had more than one treatment session. In our study group, nearly 20% of patients had more than two VR procedures, however, none of these had uncontrolled IOP at 1 year or at final follow-up visit. The long-term effect of cyclodiode treatment in lowering IOP in a controlled manner in silicone oil-related glaucoma is encouraging and this treatment modality should be considered as an effective method of treatment in these often challenging eyes. It would seem reasonable to consider such treatment at an early stage in the management of these cases.