Most tumours are heterogeneous in oxygen tension (pO2) and show regions with severely hypoxic cells (Vaupel et al, 1989). Two main types of hypoxia have been recognised: chronic hypoxia, arising from limitations in oxygen diffusion, and acute hypoxia, resulting from transient cessations in microregional blood flow (Horsman, 1995; Brown, 1999). Tumour hypoxia causes resistance to radiation therapy and some forms of chemotherapy (Brown, 1999; Wouters et al, 2002) and promotes malignant progression and the development of metastatic disease (Rofstad, 2000; Höckel and Vaupel, 2001). Clinical investigations have shown that extensive hypoxia in the primary tumour is associated with locoregional treatment failure and poor disease-free and overall survival in several histological types of cancer (Rofstad, 2000; Höckel and Vaupel, 2001; Wouters et al, 2002).

The spatial heterogeneity in pO2 in tumours has been studied in great detail (Vaupel et al, 1989; Vaupel, 1990). Steep gradients in pO2 and pO2 values ranging from 0 mmHg to those found in well-oxygenated normal tissues are characteristic features of tumours, implying that tumour tissues may show extensive spatial heterogeneity in cellular radiation sensitivity (Wouters and Brown, 1997) and hypoxia-induced gene expression (Acker and Plate, 2002; Harris, 2002). In contrast, studies of the temporal heterogeneity in pO2 in tumours are sparse (Dewhirst et al, 2000). However, examinations of fluctuations in microvessel blood flow in window chamber tumours and measurements of the kinetics of red blood cell flux in experimental and human solid tumours by laser Doppler flowmetry have suggested that temporal heterogeneity in pO2 and acute hypoxia are common events in tumour tissues (Endrich et al, 1979; Brizel et al, 1993; Chaplin and Hill, 1995; Hill et al, 1996). Moreover, radiobiological studies of murine tumours and human tumour xenografts have given evidence that acute hypoxia may be a significant cause of tumour radiation resistance (Yamaura and Matsuzawa, 1979; Chaplin et al, 1987; Rofstad and Måseide, 1999). Measurements of temporal heterogeneity in pO2, however, have been reported for tumours of one line only, the rat R3230Ac mammary adenocarcinoma. In these experiments, recessed-tip oxygen microelectrodes (Linsenmeier and Yancey, 1987) were used to record fluctuations in pO2 in tumours implanted subcutaneously (Dewhirst et al, 1998; Braun et al, 1999) or in window chambers (Dewhirst et al, 1996; Kimura et al, 1996). The studies suggested that tissue pO2 and red blood cell flux fluctuated at low frequencies and were temporally coordinated, and demonstrated that acute hypoxia is a frequently occurring phenomenon in R3230Ac tumours (Dewhirst et al, 1996, 1998; Kimura et al, 1996; Braun et al, 1999).

The low number of investigations of the temporal heterogeneity in pO2 in tumours most likely reflects that adequate equipment for measurement of fluctuations in pO2 in tissues has not been commercially available (Dewhirst et al, 2000). A new oxygen-sensing device, the OxyLite system, was marketed recently (Young et al, 1996; Griffiths and Robinson, 1999). This device measures pO2 by using a fluorescence quenching technique. Light pulses induce fluorescence in ruthenium chloride incorporated into a silicone rubber polymer at the tip of 220-μm-diameter fibre-optic probes. The lifetime of the fluorescence is inversely proportional to the oxygen tension at the probe tip. Comparative studies have shown that tumour pO2 distributions and hypoxic fractions measured with the OxyLite system are similar to those obtained with Eppendorf polarographic electrodes (Collingridge et al, 1997; Seddon et al, 2001), recessed-tip microelectrodes (Braun et al, 2001) and radiobiological assays (Urano et al, 2002). The OxyLite system has been used successfully to measure pO2 changes in experimental tumours following different types of physiological intervention (Bussink et al, 2000; Braun et al, 2001;Demeure et al, 2002; Jarm et al, 2002; Jordan et al, 2002).

In the work reported here, the OxyLite system was used to measure pO2 fluctuations in unperturbed A-07 human melanoma xenografts transplanted orthotopically in BALB/c-nu/nu mice. Xenografted A-07 tumours have been shown to retain many biological characteristics of the donor patient's tumour, including cell cycle distribution, angiogenic potential, vascular density, metastatic pattern and several histological and pathophysiological parameters (Rofstad, 1994). The main purpose of the work was to establish the usefulness of the OxyLite system in studying the temporal heterogeneity in pO2 in tissues and to characterise the pO2 fluctuations in a clinically relevant tumour model. Significant evidence was found that the pO2 in many tumour regions fluctuated around the threshold values for hypoxia-induced radiation resistance and hypoxia-induced gene expression.

Materials and methods

Mice and tumours

Adult (8−10 weeks of age) female BALB/c-nu/nu mice were used as host animals for xenografted tumours. The mice were maintained under specific pathogen-free conditions at constant temperature (37.0±0.5°C) and humidity (50−60%). Sterilised food and water were given ad libitum. Experiments were performed with tumours of the A-07 human melanoma line, established as described previously (Rofstad, 1994). Tumours were initiated from exponentially growing cell cultures verified to be free from Mycoplasma contamination. The cells were cultured in RPMI-1640 medium (25 mM HEPES and L-glutamine) supplemented with 13% bovine calf serum, 250 mg l−1 penicillin and 50 mg l−1 streptomycin. Approximately 3.5 × 105 cells in 10 μl of Ca2+ and Mg2+-free Hanks' balanced salt solution were inoculated intradermally into the hindmost part of the left mouse flank by using a 100-μl Hamilton syringe. The tumours were subjected to measurements of pO2 when having volumes within the range of 300−700 mm3. Tumour volume (V) was calculated as V=(π/6)ab2, where a is the longer and b is the shorter of two perpendicular diameters, measured with calipers. Animal experiments were approved by the Institutional Committee on Research Animal Care and were performed in accordance with the ethical standards of the UKCCCR ‘Guidelines for the Welfare of Animals in Experimental Neoplasia’ (Workman et al, 1998).

Measurement of pO2

Tumour pO2 was measured with a two-channel fibre-optic oxygen-sensing device (OxyLite™ 2000, Oxford Optronix, Oxford, UK). The probes were supplied precalibrated by the manufacturer. The calibration data were scanned into a computer by means of a barcode wand. The calibration specification was ±0.7 mmHg or <±10% of actual pO2, whichever was greater. To ensure correct pO2 readings, the accuracy of the precalibration data was controlled for every probe by measuring pO2 in Ringers solutions flushed with N2 gas containing 0, 0.5, 1.0, 2.5 or 5.0% O2. This procedure was repeated at the end of each experiment to control for drift in the system. Atmospheric pressure was measured with a DPI 705 barometer (Newport Electronics, Deckenpfronn, Germany).

The mice were kept under anaesthesia during the pO2 measurements. A mixture of fentanyl citrate/fluanisone (Janssen Pharmaceutika, Beerse, Belgium) and midazolam (Roche, Basel, Switzerland) was administered intraperitoneally in doses of 0.63/20 and 10 mg kg−1, respectively. The body core temperature of the mice, measured with a rectal probe, was kept at 37.5–38.0°C by using a heating lamp and a heating pad. The tumour temperature under these conditions, measured with an OxyLite probe having a fine thermocouple wound around the optic fibre, was within 37.0–38.0°C, that is, it was similar to the tumour temperature during the growth period prior to the pO2 measurements.

Most tumours had an ellipsoidic shape, and tissue pO2 was measured simultaneously in two positions in each tumour, that is, centrally in each of the two halves of the ellipsoid. The probes were inserted into the tumours through skin punctures made with 21- or 23-gauge needles. After insertion, the probes were retracted by approximately 1 mm to minimise the pressure on the sensors before they were fixed firmly and pO2 measurements were initiated. The pO2 readings were recorded every 5, 10 or 30 s for at least 60 min and were stored to disc and displayed as time traces by using a data-acquisition system (LabView, National Instruments, Austin, TX, USA). The mice were watched continuously during the experiments and reflex movements were recorded. At the end of the experiments, the mice were killed without withdrawing the probes from the tumour tissue, and it was ensured that the pO2 dropped rapidly to 0 mmHg. Data were discarded if (a) changes in tumour pO2 were associated with reflex movements of the host mouse, (b) tumour pO2 did not drop to 0 mmHg after killing of the host mouse or (c) the probe calibration at the end of an experiment gave data that differed significantly from the precalibration data.

Fourier analysis

Characteristic frequencies of pO2 fluctuations were searched for by subjecting the pO2 traces (oxygen tension vs time) to Fourier analysis. The Fourier analysis was performed by using IDL software (Research Systems, Boulder, CO, USA). The program calculated frequency spectra from pO2 traces by accomplishing a discrete Fourier transformation. The highest detectable frequency was given by the sampling frequency and was 100 mHz at a sampling frequency of 5 s. The lowest detectable frequency was given by the length of the observation period and was approximately 0.4 mHz at an observation period of 40 min.

Results

Measurement of pO2 in A-07 tumours with OxyLite probes gave values that varied substantially with time. The variations were more pronounced shortly after the probe insertion than towards the end of the observation period. Two characteristic pO2 traces are illustrated in Figure 1A. The initial pO2 was above 5 mmHg in both traces. In trace #1, the pO2 decreased for several minutes before it increased gradually and reached a high and fairly stable level. In trace #2, the pO2 decreased rapidly with time before a low and fairly stable level was reached. To investigate whether the initial changes in pO2 were artefacts caused by the probe insertion, similar pO2 measurements were performed in muscle tissue in BALB/c-nu/nu mice. Two typical pO2 traces from muscle tissue are presented in Figure 1B. The muscle pO2 traces were qualitatively similar to pO2 trace #1 in Figure 1A, strongly suggesting that the initial pO2 changes were artefacts.

Figure 1
figure 1

Examples of pO2 traces recorded with OxyLite probes in A-07 tumours (A) and muscle tissue (B). The traces show that the changes in pO2 were more pronounced shortly after the probe insertion than towards the end of the observation period.

To determine the length of the time period in which the pO2 readings in A-07 tumours obviously were influenced by the probe insertion, 50 randomly selected pO2 traces were normalised and summed. The normalisation was performed by dividing all pO2 values in a trace by the highest pO2 value recorded in that trace, excluding pO2 values recorded during the first 10 min when determining the normalisation factor. The pO2 trace representing the sum of the normalised pO2 traces is plotted in Figure 2. This plot suggests that the changes in pO2 recorded within the first 20 min after the probe insertion were systematic and hence were artefacts caused by the probe, whereas the pO2 changes recorded beyond 20 min were random and hence most likely represented true variations in tissue pO2. Consequently, the temporal heterogeneity in pO2 in A-07 tumours was studied by only considering pO2 values recorded beyond the first 20 min after the insertion of a probe, that is, the first 20 min of each pO2 trace was excluded in the analysis presented below.

Figure 2
figure 2

Sum of 50 normalised pO2 traces recorded with OxyLite probes in A-07 tumours. The composed trace shows that the changes in pO2 recorded within the first 20 min after the probe insertion were systematic, whereas those recorded beyond the first 20 min were random.

A total of 38 A-07 tumours were subjected to pO2 measurements and a total of 70 reliable pO2 traces were obtained. The pO2 traces showed substantial differences, regardless of whether they were recorded in the same tumour or in different tumours. Mean pO2 differed among the traces from 0 to 38 mmHg. Examples of characteristic pO2 traces are illustrated in Figure 3. Both well-oxygenated and poorly oxygenated tumour regions could show pO2 traces without significant fluctuations (Figure 3A). However, significant pO2 fluctuations were detected in most tumour regions (Figure 3B). The pO2 traces were analysed by using two threshold values for hypoxia, that is, pO2=5 and 10 mmHg. Some tumour regions were not hypoxic at all during the observation period, whereas others were hypoxic during the entire period (Table 1). Acute hypoxia, that is, pO2 fluctuations around the threshold values, was detected in 29% (10 mmHg) and 39% (5 mmHg) of the tumour regions. To characterise the kinetics of the acute hypoxia, the number of times per hour the pO2 decreased below the threshold values and the fractional time the pO2 was below the threshold values were calculated for each of the tumour regions showing acute hypoxia. The durations of the hypoxic periods were also determined. The median values and the ranges of these parameters are presented in Table 1. A similar analysis has been performed previously for R3230Ac rat tumours (Dewhirst et al, 1998), and the results of this analysis were included in Table 1 for comparison.

Figure 3
figure 3

Examples of pO2 traces recorded with OxyLite probes in A-07 tumours. The traces refer to tumour regions without significant fluctuations in pO2 (A) and tumour regions showing significant fluctuations in pO2 around pO2 values of 5 and 10 mmHg (B).

Table 1 Parameters describing the kinetics of acute hypoxia in A-07 human melanoma xenografts and R3230Ac rat mammary adenocarcinomas

Two pO2 traces were recorded simultaneously in most tumours. None of the traces, one of the traces, or both traces could show significant pO2 fluctuations. The pO2 values of concurrent traces were subjected to correlation analysis to investigate whether the pO2 fluctuations in different regions of the same tumour were temporally coordinated. Positive correlations were found in some tumours (Figure 4A), whereas inverse correlations were seen in others (Figure 4B). However, there was no correlation between the two series of pO2 values in the majority of the tumours, implying that the pO2 fluctuations in different regions of a tumour in general were temporally independent.

Figure 4
figure 4

Examples of pO2 traces recorded simultaneously with OxyLite probes in two regions of the same A-07 tumours. The traces refer to a tumour where the pO2 values were correlated (A) and a tumour where the pO2 values were inversely correlated (B).

Moreover, the pO2 traces were subjected to Fourier analysis to investigate whether the pO2 fluctuated at characteristic frequencies. The Fourier analysis resulted in frequency spectra that were not qualitatively different, regardless of whether they were derived from the same tumour or from different tumours. The frequency spectra indicated that the pO2 fluctuated at very low frequencies, that is, at frequencies lower than 1.5–2.0 mHz, corresponding to less than 0.1 cycle min−1. Significant fluctuations at higher frequencies could not be detected. Data from a characteristic tumour region are presented in Figure 5, showing the pO2 trace (Figure 5A) and the corresponding frequency spectrum (Figure 5B).

Figure 5
figure 5

Example of a pO2 trace recorded with an OxyLite probe in an A-07 tumour (A) and the corresponding frequency spectrum (B). The frequency spectrum, which was obtained by subjecting the pO2 data to Fourier analysis, suggests that the pO2 fluctuated at frequencies lower than 1.5–2.0 mHz.

Discussion

Tissue pO2 in A-07 human melanoma xenografts was monitored continuously over periods of at least 60 min by using OxyLite fibre-optic probes. The study showed that fluctuations in pO2 at the microregional level occur frequently in A-07 tumours. Moreover, acute hypoxia was found to be a common phenomenon in these tumours, which is in agreement with the conclusions from a previous study in which radiobiological and immunohistochemical assays were used to detect hypoxia in A-07 and other human melanoma xenografts (Rofstad and Måseide, 1999). The OxyLite system has been used previously to study changes in tumour pO2 following treatment with blood flow and tissue oxygenation modifying agents (Bussink et al, 2000; Braun et al, 2001;Demeure et al, 2002; Jarm et al, 2002; Jordan et al, 2002). The present study is the first in which the OxyLite system has been used successfully to study temporal heterogeneity in pO2 in unperturbed tumours.

Studies of temporal heterogeneity in pO2 in tumour tissue with the OxyLite system, however, require precautions against potential methodical pitfalls, as revealed by the work reported here. First, it was observed that the pO2 values recorded shortly after the probe was inserted into tumour tissue varied systematically with time and hence were influenced signifcantly by the probe insertion. A similar artefact was seen when tissue pO2 was measured in P22 rat carcinosarcomas with OxyLite probes (Seddon et al, 2001). The pO2 readings stabilised after approximately 10 min in P22 tumours, whereas in A-07 tumours, reliable pO2 readings could not be obtained until 20 min after the probe insertion. The artificial pO2 readings obtained during the first 20 min of measurement in A-07 tumours were most likely a result of vasoconstrictive reactions to tissue trauma induced by the probe. However, other factors may also have contributed, as discussed in detail for P22 tumours (Seddon et al, 2001). Second, it was observed that abrupt changes in the pO2 readings could occur simultaneously with reflex movements of the host mouse. These changes did probably not reflect temporal heterogeneity in pO2, but were rather a consequence of minor changes in probe position and hence the spatial heterogeneity in pO2, as it is well known that tumour tissues can show steep pO2 gradients (Vaupel, 1990; Horsman, 1995; Braun et al, 2001; Urano et al, 2002). Third, it was observed that some correctly precalibrated probes after a few measurements in tissue suddenly began recording erroneous absolute values of pO2. Therefore, it is essential to kill the host animals after each experiment and ensure that the pO2 drops rapidly to 0 mmHg, and to control the probe calibration regularly in Ringers solutions, as was followed in the work reported here.

Our analysis was based on the assumption that the pO2 readings recorded beyond the first 20 min after the probe insertion were not influenced significantly by the tissue trauma caused by the probe. The OxyLite probes have a tip diameter of 220 μm and may therefore now and then cause severe tissue damage during the insertion, for example by destroying or compressing larger vessels. Consequently, it cannot be excluded that the pO2 readings in some of the tumour regions studied here were influenced by probe-induced tissue damage also beyond the first 20 min. However, several observations suggest that this potential problem, if present, was of minor significance. First, pO2 was measured in normal tissues also, and significant pO2 fluctuations were never observed beyond 20 min, as illustrated for muscle tissue in Figure 1B. Second, we have shown that the mean pO2 measured beyond 20 min is inversely correlated to the fraction of radiobiologically hypoxic cells in A-07 and R-18 human melanoma xenografts (Brurberg et al, unpublished data). Moreover, comparative studies performed in other laboratories have demonstrated that the pO2 distributions measured in experimental tumours with OxyLite probes are similar to those obtained with Eppendorf polarographic electrodes (Collingridge et al, 1997; Seddon et al, 2001) and recessed-tip microelectrodes (Braun et al, 2001).

Previous studies of the rat R3230Ac mammary adenocarcinoma have led to the suggestion that fluctuations in tissue pO2 and acute hypoxia may be commonly occurring phenomena in tumours (Dewhirst et al, 1996, 1998; Kimura et al, 1996; Braun et al, 1999). The present study of A-07 human melanoma xenografts, which have been shown to retain several characteristic biological features of the donor patient's tumour (Rofstad, 1994) and hence most likely are more relevant models of tumours in man than are R3230Ac tumours, confirmed this suggestion. A direct comparison of the temporal heterogeneity in pO2 in A-07 and R3230Ac tumours is difficult, however, because OxyLite fibre-optic probes were used to measure pO2 in the present work and recessed-tip microelectrodes were used to measure pO2 in the R3230Ac tumours. The tip diameter of the OxyLite probes is 220 μm, and the sampling volume has been estimated to be approximately 1000 cells (Griffiths and Robinson, 1999; Seddon et al, 2001). In contrast, the recessed-tip microelectrodes had a diameter of only 10–12 μm (Dewhirst et al, 1998), and therefore, they had a sampling volume that was substantially smaller than that of the OxyLite probes (Braun et al, 2001). Since tumours are spatially heterogeneous in pO2 at the microregional level (Vaupel, 1990; Horsman, 1995), microelectrodes are expected to measure larger fluctuations in pO2 in tumours than are OxyLite probes (Braun et al, 2001). Nevertheless, the temporal heterogeneity in pO2 measured in A-07 tumours was remarkably similar to that measured in R3230Ac tumours, as can be seen from the comparison of A-07 and R3230Ac tumours presented in Table 1.

Temporal heterogeneity in pO2 in tumour tissue has to be caused by temporal heterogeneity in either oxygen delivery, that is, blood supply, or oxygen consumption, that is, cell respiration, or both. There is no experimental evidence that the rate of respiration may fluctuate synchronously in cells within tumour microregions. On the other hand, there is ample evidence that the blood supply may fluctuate significantly at the microregional level in both experimental and human tumours (Endrich et al, 1979; Brizel et al, 1993; Chaplin and Hill, 1995; Hill et al, 1996). Studies of R3230Ac tumours transplanted to window chambers have suggested that the fluctuations in pO2 in these tumours were temporally coordinated with fluctuations in red blood cell flux (Dewhirst et al, 1996; Kimura et al, 1996). Moreover, Fourier analysis of pO2 traces recorded with recessed-tip microelectrodes and red blood cell flux traces recorded by laser Doppler flowmetry revealed that both pO2 and blood flow fluctuated at very low frequencies in subcutaneous R3230Ac tumours (Braun et al, 1999). The data reported here for A-07 tumours are in good agreement with those of the R3230Ac tumours. Thus, the pO2 frequency spectra of the A-07 tumours suggested that the pO2 fluctuated at low frequencies also in these tumours, that is, at frequencies lower than 1.5–2.0 mHz, corresponding to less than 0.1 cycle min−1. Fluctuations in pO2 in this frequency range could result from fluctuations in blood flow caused by vasomotion in supplying arterioles, haemodynamic mechnisms and/or microvascular remodelling via intussusceptive vascular growth (Braun et al, 1999). Haemodynamic mechanisms may be particularly significant because the blood viscosity is elevated in tumour tissues and the tumour microvascular network is irregular and chaotic (Vaupel et al, 1989).

Studies of intratumour heterogeneity in temporal variation in pO2 have not been reported so far. The pO2 measurements performed in R3230Ac tumours were all restricted to a single point at a time in each tumour (Dewhirst et al, 1996, 1998; Kimura et al, 1996; Braun et al, 1999). Therefore, these studies did not provide information on the fractional tumour volume showing pO2 fluctuations or on the temporal coordination of the pO2 fluctuations in different tumour regions. Attempts to obtain information of this type were made in the present work by measuring pO2 simultaneously in two distinctly different regions of the same A-07 tumours. In some tumours, significant fluctuations in pO2 could not be detected in any of the regions, and in others, pO2 fluctuated significantly in one of the regions only, suggesting that the pO2 fluctuations usually involved only a fraction of the tumour volume. Many A-07 tumours, however, showed significant pO2 fluctuations in both regions. These fluctuations were in general not temporally coordinated, suggesting that they were caused primarily by redistribution of the perfusion within the tumours rather than changes in global perfusion. Thus, in some tumours, the pO2 values in the two regions analysed simultaneously were inversely correlated. However, our experiments can by no means exclude the possibility that also the global perfusion and hence the fraction of acutely hypoxic cells varied significantly with time. In fact, this possibility is very likely, considering the irregular and heterogeneous nature of the microvasculature of tumours, and is supported by the observation that the two pO2 series recorded simultaneously were strongly correlated in some tumours. Studies of temporal heterogeneity in pO2 involving mapping of the pO2 distribution in whole tumours are needed, however, before this question can be settled.

The study reported here may have significant implications for the radiation therapy of cancer. First, clinical studies attempting to eliminate the chronic hypoxia in tumours during radiation therapy are being performed, but so far, the therapeutic results have been unsatisfactory (Overgaard and Horsman, 1996). The present observations suggest that a significant fraction of the hypoxic cells in tumours are acutely hypoxic, and acutely hypoxic cells may be more resistant to radiation therapy than chronically hypoxic cells (Pettersen and Wang, 1996; Zölzer and Streffer, 2002). Consequently, treatment strategies aiming at reducing the fraction of acutely hypoxic cells may prove more successful in improving the outcome of radiation therapy than those aiming at reducing the fraction of chronically hypoxic cells. Second, serious attempts have been initiated to improve the local control of radiation-resistant tumours by using intensity modulated radiation therapy for selective boosting of hypoxic subvolumes (Tome and Fowler, 2000; Chao et al, 2001; Popple et al, 2002). Our observations suggest that the spatial distribution of the acutely hypoxic regions in tumours may change rapidly with time. Consequently, efficient selective boosting of hypoxic subvolumes may require novel technology for imaging of tumour hypoxia and guiding of intensity-modulated radiation therapy.

The present observations are also relevant for our understanding of the malignant progression of tumours and the development of metastatic disease. It is well established that tumour hypoxia activates DNA transcription factors, for example, HIF-1, and leads to increased expression of a large number of genes, and a high expression of these genes has been shown to be associated with poor prognosis in several histological types of cancer (Rofstad, 2000; Höckel and Vaupel, 2001; Acker and Plate, 2002; Harris, 2002; Wouters et al, 2002). Some of the genes that are activated under hypoxic conditions encode proteins involved in the metastatic process, for example, angiogenesis factors and proteolytic enzymes. Studies of human melanoma xenografts have shown that tumour hypoxia may promote metastasis by upregulating the expression of interleukin-8 (Rofstad and Halsør, 2002) and urokinase-type plasminogen activator receptor (Rofstad et al, 2002). The study reported here showed that a significant fraction of the hypoxic volume of tumours may be acutely hypoxic and that the spatial distribution of the acutely hypoxic regions may change rapidly with time. It is possible that the dynamic nature of the acute hypoxia in tumours may lead to hypoxia-induced gene expression without loss of viability in a substantial fraction of the malignant cells and hence may promote aggressiveness and metastatic spread. Consistent with this suggestion is the observation that experimentally imposed acute hypoxia but not chronic hypoxia-enhanced spontaneous metastasis in KHT murine tumours (Cairns et al, 2001).

In summary, the present study showed that significant fluctuations in tissue pO2 and acute hypoxia are commonly occurring phenomena in A-07 human melanoma xenografts. If A-07 tumours are relevant models of tumours in man, acute hypoxia may be an important cause of resistance to radiation therapy and malignant progression in human cancer.