Introduction

Although radioactivity was discovered more than 120 years ago1, decay properties of a considerable number of radioactive isotopes close to the line of stability are still not precisely known. Intriguingly, this knowledge gap also includes isotopes with medium or long half-lives that are readily produced within nuclear facilities or due to interactions of matter with cosmic radiation like 32Si or 53Mn2,3,4. The majority of decay data for radioisotopes are based on measurements that were conducted at the beginning of the second half of the last century. However, the development of more precise measurement tools like high resolution inductively coupled plasma—mass spectrometry (ICP-MS), accelerator mass spectrometry (AMS) as well as the improvements in the liquid scintillation counting (LSC) together with new powerful chemical separation methods have opened the possibility to improve the statistical uncertainty and accurateness of such data.

The half-life of a radionuclide can be determined by various methods. Often, when the half-life is not too long, it can be determined by means of repeat measurements. Such a method, however, is limited to radionuclides of half-lives up to ca. 100 years and crucially depends on the stability of the sample and the measurement setup. Moreover, it may require repeated measurements for a long time (e.g. several years). In such a case, a more suitable alternative is a half-life determination via a combination of precise determinations of the activity and the number of atoms of the respective radionuclide.

However, two requirements need to be met for the second method. Firstly, if the number of atoms is determined by mass spectrometry, the sample needs to be free from isobaric and polyatomic interferences. Secondly, for the activity measurements, the efficiency of the measurement setup needs to be precisely known and the sample must not contain radioisotopes that interfere with the signal of the target nuclide.

Despite the seeming simplicity of the involved techniques, the half-lives of a significant number of not so rare radioisotopes are still insufficiently constrained; sometimes because isobaric impurities hamper quantifying the number of atoms or because the pure radioisotope is hard to obtain in sufficient amounts for precise measurements. Another major complication during radionuclide production for half-life measurements is the simultaneous production of other radioisotopes of the target element which may require an additional separation using a mass separator.

93Mo is one of the last four radionuclides (the other three being 91Nb, 205Pb and 236Np5,6,7,8,9) with a half-life longer than 1 year and less than 100 million years whose half-life was never measured directly. 93Mo decays purely via electron capture (EC) to the metastable 93mNb (T1/2 = 16.12(12) y) and the ground state of 93Nb. The decay scheme of 93Mo is presented in Fig. 1.

Figure 1
figure 1

Evaluated decay scheme of 93Mo from5. Reprinted from Nuclear Data Sheets, 112/5, Coral M. Baglin, Nuclear Data Sheets for A = 93, 1163–1389, Copyright (2011), with permissionn from Elsevier.

The currently accepted evaluated half-life of 4000(800) years is marked as tentative and carrying a significant uncertainty5. This is very surprising because it is the only isotope across the chart of nuclei produced by neutron activation of a stable isotope, where the decay properties are so vaguely known. The main source of 93Mo is the neutron capture of 92Mo in molybdenum rich materials, typically steel, Iconel alloys or cables used for constructions within nuclear power plants10,11,12. Molybdenum is also a potential material for nuclear fusion devices where high neutron fluxes unavoidably lead to the production of 93Mo as one of the main long-lived radioactive products13. Other sources of 93Mo in nuclear waste are particle accelerator facilities where the neutrons are emitted due to nuclear reactions. The proton accelerator at Paul Scherrer Institute where the molybdenum containing alloys are used for window materials to separate different part of the accelerator complex14 can serve as an example. Due to its long half-life 93Mo becomes the dominant radioactive dose contributor to humans in low and intermediate level radioactive waste when the shorter-lived radionuclides have decayed in the time period from 5000 to 30,000 AD15. Highlighting the requirement of a better-constrained half-life value of 93Mo, preliminary calculations show 93Mo to become the major radionuclide in the final repository of low-level nuclear waste over the time16.

First efforts in search for a long-lived Mo isotope have been undertaken at Oak Ridge National Laboratory more than 80 years ago. In these early investigations of neutron irradiated molybdenum, the 93Mo half-life has been estimated to be at least 70 years17. The first indirect determination of the half-life18 is based on the comparison of isotope production yields for (d,2n) reactions with a deuteron beam energy of 21 MeV in the atomic mass region of A = 85–109 with 93Mo produced by chemical separation from a bombarded niobium foil. The 93Mo activity was determined by measuring the relaxation radiation with X-ray spectrometry, while the number of produced atoms was estimated by comparing it to (d,2n) reactions on 85Rb, 103Rh and 109Ag, assuming a constant production yield within this atomic mass interval. The obtained value for the 93Mo half-life was later adjusted by the evaluator to the currently adopted value taking into account updated nuclear decay properties used in the work of5,18.

In this work we demonstrate the power of state of art chemical separations combined with modern instrumentation to determine the nuclear properties of radionuclides with very high precision. The isobaric impurities present the biggest obstacle in the precise determination of the number of atoms by mass spectrometry. Therefore, chemical separations play the crucial role in half-life determination by an absolute activity measurement. We communicate here chemical separation yielding significantly high decontamination factors of 93Mo from the interfering 93Nb isobar leading to precise determination of the 93Mo half-life by absolute activity determination.

Experimental techniques and methodology

Separation and purification of 93Mo sample

In this work, we produced a sample of around 5 × 1015 atoms of 93Mo from a 2.1 mm thick high purity niobium disc which was irradiated with 72 MeV protons. The cross section of the 93Nb(p,n)93Mo reaction is approximately 15 mb19. The content of radionuclides in the irradiated disc was investigated by the means of gamma-ray and X-ray spectroscopy. Three parts of the irradiated niobium disc with high radioactivity content identified by means of autoradiography were cut out. A more detailed description of the used Nb degrader, its processing and characterization is given in the Supplementary materials. These were dissolved in a mixture of 2 mL 48% HF and a 100 µL 70% HNO3. The separation of 93Mo from niobium and radioactive contaminants consisted of two separate chemical systems. In the first one the acid composition was adjusted to 6.5 mol/L HCl/0.1 mol/L HF and applied to the column filled with 2 mL of tributylphosphate (TBP) resin. Under this chemical composition molybdenum is fully retained on resin whereas niobium together with 22Na, 60Co, 93Zr and 54Mn contaminants pass through the chromatographic column. The fraction containing 93Mo was thereafter eluted from the resin by distilled water from a MilliQ system. In order to determine the decontamination factor (DF) from niobium, the sample was spiked before the separation with carrier free 95Nb obtained from the inhouse produced 95Zr/95Nb generator (for details see Supplementary material). The decontamination factor is defined as:

$${D}_{\mathrm{F}}=\frac{{C}_{{\mathrm{Nb}}_{\mathrm{L}}}}{{C}_{{\mathrm{Nb}}_{\mathrm{E}}}}\equiv \frac{{A}_{{95}_{{\mathrm{Nb}}_{\mathrm{L}}}}}{{A}_{{95}_{{\mathrm{Nb}}_{\mathrm{E}}}}}$$

where CNb represents the concentration of niobium in the sample in the loading (L) and elution (E) fractions of the separation, that is corresponding to the measured specific radioactivity of 95Nb (\({A}_{{95}_{\mathrm{Nb}}}\)) in the loading and elution fractions corrected for the decay during the separation procedure and measurement. Hence, the decontamination factor from niobium was determined by γ-spectrometric measurements using the 765.8 keV line of 95Nb measured by an n-type HPGe detector. The measurements were performed with a live time of ca. 24 h and corrected for the decay of 95Nb during this period.

The second cycle of purification was performed on alumina substrate in a chromatographic column according to20. The eluate from the TBP resin was evaporated to dryness, re-dissolved in 0.1 mol/L HF and spiked again with 95Nb tracer. The solution was thereafter pushed through a chromatographic column filled with alumina substrate. During this procedure Zr, Nb, Mo, Ta and Fe were retained on the alumina substrate. The fraction containing 93Mo was thereafter selectively eluted from the substrate by 20% NH4OH solution. During this step, a nearly complete separation from remaining niobium as well as from traces of iron and 179Ta was achieved. The simplified schematics of the separation procedure is shown in Fig. 2.

Figure 2
figure 2

Schematics of the separation procedure for 93Mo master solution.

Each of the three sub-samples from the irradiated Nb disc underwent a separation procedure on both columns separately before they were merged together and underwent one more separation on TBP resin resulting in the “93Mo master solution”.

Mass spectrometric analysis

The homogeneous 93Mo master solution was split to use one part for LSC measurements and the other part for MC-ICP-MS determinations of the 93Mo content using a Nu Instruments Plasma 3 MC-ICP-MS. Given the large first ionization potential of Argon21 the inductively coupled Ar-plasma ion source of this mass spectrometer efficiently ionizes almost all elements of the periodic table. This efficient ionization capability and the ability of the state-of-the-art Plasma 3 MC-ICP-MS to detect up to 22 ion beams simultaneously allow highly precise determinations of intensity-ratios of mass-specific ion beams of the same or multiple elements. Exploiting this ability two different methods were employed to define the 93Mo content of the master solution: gravimetric standard addition with an extra addition of Zr as internal standard22 as well as reverse isotope dilution23. The former method makes use of the increase in the intensities of 95Mo and 97Mo ion beams per gravimetrically defined addition of natural Mo (LGC Molybdenum Standard 976(3) µg/g, k = 2, LOT # 987189-15, traceable to NIST SRM 314, lot #130418), cancelling intensity variations due to plasma fluctuations by referencing to ion beams of 90Zr or 91Zr from a gravimetrically defined addition of natural Zr (LGC Zirconium Standard 993(2) µg/g, k = 2, LOT # 150521R-15, traceable to NIST SRM 3169, lot #130920). The latter method—in the following referred to as “reverse isotope dilution”—utilizes the change in 93Mo-to-95,97Mo ion beam intensity ratios per gravimetrically defined addition of natural Mo.

All solutions were introduced into the system via a self-aspiring “PFA-ST MicroFlow” consuming ca. 50 μL/min and an “Apex HF” desolvating nebulizer (both from Elemental Scientific). Ion beams from mass 90 (Zr) through to 97 (Mo) were collected simultaneously in Faraday cups connected to amplifiers with a 1011 Ω resistor in their feedback loop. Instrumental background signal contributions to each ion beam were removed by subtracting the signals obtained during 120 s of measurement prior to each analysis on the pure 0.28 M HNO3–0.01 M HF solution that was used for standard and sample preparation. Each sample and standard measurement comprises 100 ten-second-long integrations of the ion beam intensities. All gravimetric additions were done on a Mettler Toledo XP56 scale (10–6 g scale interval). In detail, four gravimetric mixtures of the 93Mo master solution with different amounts of the Mo standard solution and approximately the same amount of a Zr standard solution as well as one Mo standard-free mixture of the 93Mo master solution and the Zr standard solution were prepared. Note that if the same solution is used for all mixtures, the exact concentration of the Zr standard solution is irrelevant for the results of the standard addition with an internal standard.

The content of a stable isotope of Mo in the master solution (here: interference-free 95Mo and 97Mo) is then calculated from the linear change in.

  1. a)

    “Standard addition with internal standard”: the 95Mo and 97Mo signal intensity referenced (i.e., “divided by”) to the 90Zr or 91Zr ion beam intensity and multiplied by the weight ratio of the added internal standard and the 93Mo master solution in the mixture (see22 for general details and formulae) or

  2. b)

    “Reverse isotope dilution”: the measured ratios 95Mo/93Mo and 97Mo/93Mo of the mixture compared to the respective value of the Mo standard-free mixture (see23 for general details and formulae).

with the gravimetrically defined weight ratio of the added amount of the Mo standard and the master solution in the mixture.

Final concentration uncertainties are derived by Monte-Carlo propagation and include:

  1. (i)

    the certified concentration uncertainty (k = 1) of the Mo reference solutions,

  2. (ii)

    weighing uncertainties as one standard deviation obtained from repeatedly (N = 10) weighing each dilution and addition step during and before preparation of all mixtures containing the master solution,

  3. (iii)

    the uncertainties of repeated (N = 5) mass spectrometric analyses of each mixtures containing the master solution as one standard deviation,

  4. (iv)

    and the overall scatter in concentration results from combining all abovementioned uncertainties in

    1. a.

      Monte-Carlo models of repeated linear regressions (N = 1 × 106) for standard addition calculations and

    2. b.

      Monte-Carlo models (N = 1 × 106) of inverse isotope dilution.

Liquid scintillation counting

The liquid scintillation (LS) measurements were carried out in two custom-built TDCR (i.e., triple-to-double coincidence ratio) counters at PTB. The first system is referred to as M2724 and it is equipped with a MAC3 coincidence unit25. The second TDCR counter, M29, makes use of an FPGA-based coincidence module referred to as the “4KAM”26. In addition, two commercial LS counters were used to apply the CIEMAT/NIST efficiency tracing (CNET) method: a Wallac 1414 Guardian™ liquid scintillation spectrometer and a TriCarb2800 TR from PerkinElmer. For details see Supplementary materials.

Two LS sample series were prepared. The first series was made using 20 mL glass vials, while 20 mL “low diffusion” polyethylene vials (with a “PTFE®-type coating on the inside surface”) were used for the second series. It has to be noted that all LS samples including the 3H samples (see below) were prepared from the same Ultima Gold (UG) batch. Small amounts of nitromethane were added to some of the samples in glass vials in order to reduce the counting efficiency by means of chemical quenching. The calibration curves, i.e. the counting efficiency of 3H as a function of the quenching indicator (called SQP(E) in case of the Wallac and tSIE in case of the TriCarb, respectively), were measured with the aid of a PTB standard solution of 3H. The LS samples containing 3H have the same sample composition and geometry as the 93Mo LS samples. All weighing procedures were carried out using a Mettler Toledo XP26, which is calibrated once a year ensuring traceability to the German national mass standard. LS spectra as well as an estimation of potential adsorption effects are given in the Supplementary materials.

Results

The half-life of 93Mo was determined through the following steps:

  1. 1.

    Preparation of the 93Mo master solution free of radioactive impurities and isobaric interferences.

  2. 2.

    Determination of the 93Mo content by Multiple-Collector Inductively Coupled Plasma Mass Spectrometry (MC-ICP-MS).

    • The sample should be free of isobars and molecules that could lead to interferences in the ICP-MS measurement.

  3. 3.

    Activity measurement of 93Mo by Liquid Scintillation (LS) counting with the following basic conditions to be fulfilled:

    • The sample must not contain significant radioactive impurities influencing the activity determination and should be chemically stable over the period of measurement.

The uncertainties of the presented values in the manuscript are given for k = 1 if not stated otherwise.

Production of 93Mo master solution

The element niobium is monoisotopic with mass 93. Therefore, the decontamination factor from the macro amount of Nb needed to be at least 1 × 1012 to avoid isobaric interference with 93Mo to make reliable MC-ICP-MS measurements possible. The achieved overall decontamination factor from niobium for the final purified 93Mo sample is larger than 1.6 × 1014.

The determined decontamination factors from Nb in different separation steps are presented in Table 1.

Table 1 Decontamination factors of 93Mo from Nb for each separation step.

Taking the mass of the niobium disc material that was used (1.8 g, corresponding to 1.65·1022 atoms) for molybdenum extraction into account, the amount of stable Nb in the master solution has been reduced to 4 × 107 atoms or less. This amount of Nb contributes to the overall budget of mass-93 nuclides by less than 10–7 and is thus insignificant. Before preparation of the final master solution, the sample was measured for 5 × 105 s live-time on an n-type HPGe detector (26% relative efficiency, Canberra) for detection of potential γ-ray emitting radioactive impurities. No gamma emissions were found apart from the natural background. The absence of radioactive impurities was also confirmed by LS counting measurements. After the chemical separation, the sample was evaporated to dryness and subsequently dissolved in 0.28 mol/L HNO3–0.01 mol/L HF mixture creating the “93Mo master solution” for both MC-ICP-MS and LS counting.

Mass spectrometry analysis

A scan across the 93Mo signal peak utilizing the high mass resolution capabilities of the Nu Instruments Plasma 3 MC-ICP-MS indicates that the signal at mass 93 comes only from 93Mo atoms and potential molecular interferences can be excluded (Fig. 3).

Figure 3
figure 3

Signal profiles at high mass resolving power illustrating the lack of molecular interferences on mass 93 in the analyte. Molecular interferences whose mass difference to that of 93Mo is smaller than the mass range of the scan would show (e.g., from low to high mass: 93Mo-only, 93Mo+  + 53Cr40Ar+, 53Cr40Ar+-only) in the signal profile as prominent steps given the logarithmic scale. Further, note the lack of significant signal contributions from Metal Argides (MAr+) at metal (Cr, Fe) impurities on the ppm level.

No significant impurities in the sample analyte were found in a separate mass scan with the only discernable impurity being Fe at a concentration of ca. 1.5 ppm (see Supplementary materials). Note that this level of impurity does not produce detectable Metal-Argide signals (i.e., 40 mass units above the impurity’s mass) as attested by a scan across mass 93 on a 1 ppm Fe–Cr solution (Fig. 3). The recorded ion beam at mass 93 is thus free of bias from nuclides or molecules other than 93Mo.

The overall 93Mo-amount determination further is free from significant bias resulting from molecular interferences on the remaining involved masses as attested by the excellent agreement among 93Mo-results using either reverse isotope dilution or the standard addition method whose results agree within less than 0.16% (Fig. 4). Moreover, 93Mo-results from isotope dilution data based on 93Mo/95Mo and 93Mo/97Mo agree within less than 0.12%; results from standard addition using 95Mo or 97Mo and referencing to either 91Zr or 90Zr agree within less than 0.13% (Fig. 4).

Figure 4
figure 4

Left: 93Mo-amount results using either the isotope dilution or the standard addition method. Right: 93Mo-amount results from standard addition using either 95Mo or 97Mo, internally normalizing to either 90Zr or 91Zr and averaging results using a mass discrimination correction based on Mo and Zr.

The total number of 93Mo atoms per gram of the 93Mo master solution was determined to be 4.415(23) × 1015.

Liquid scintillation counting

The measurements of the 93Mo LS samples and background samples yield experimental net count rates. In order to translate these into absolute activities, the corresponding counting efficiencies must be known. In this work the counting efficiencies were determined by means of the triple-to-double coincidence ratio (TDCR) method and the CIEMAT/NIST efficiency tracing (CNET) method27. These methods can be considered as being well established in radionuclide metrology. However, they have never been applied to 93Mo before. In the case of 93Mo, a major challenge is to compute the required LS counting efficiencies of the weak electron-capture (EC) decays. The computations included a stochastic atomic rearrangement model which proved its worth for several other EC radionuclides28,29. The calculations, however, require accurate nuclear and atomic data as input. In particular the fractional electron-capture probabilities must be well-known (see Supplementary material). The calculations also require knowledge on the decay probabilities of the two involved EC branches of 93Mo. The determination of these probabilities was achieved with a complex numerical procedure using long-term measurement data and is explained in the Supplementary materials, too. The probability PEC,isomer is also required to estimate the growing contribution of 93mNb to the overall counting rate.

Once, when the probabilities of the two EC branches are known, one can calculate the counting efficiencies for both methods. The ionization quenching function was computed using a Birks parameter kB = 0.0075 cm/MeV as described in30. Figure 5 shows the computed double coincidence counting efficiency εD as a function of the TDCR parameter. As for many complex decay schemes, the curve shows an ambiguity, i.e. parts where one TDCR value could lead to more than one value for εD. In order to find the correct result, it is important to apply an efficiency variation (here realized by means of chemical quenching with nitromethane). When looking at the parts with high TDCR values, one can also identify a region where small changes of the TDCR parameter result in huge changes of the corresponding counting efficiency. This may lead to high uncertainties and a somewhat larger model dependence. Hence, all data with double counting efficiencies εD > 52.2% (in particular those with TDCR > 85%) were not used in the analysis of the activity concentration. This applies only to one measurement of a glass vial in the TDCR system M27 and all TDCR measurements with the PE vial.

Figure 5
figure 5

Computed double counting efficiency εD as a function of the TDCR parameter. The ambiguity is a well-known effect which occurs for several radionuclides with complex decay schemes (see, e.g.27). The right figure shows the same data in the ROI only.

The efficiency curve for the CNET method is shown in Fig. 6. This curve does not show any ambiguity and no experimental data was discarded.

Figure 6
figure 6

Computed 93Mo counting efficiency vs. 3H counting efficiency to apply the CNET method.

The counting efficiencies of 93mNb were also calculated with the stochastic model but a detailed description is omitted here. It should, however, be noted, that the LS counting efficiency of 93mNb is very high (e.g. the double counting efficiency in TDCR is about 99.7%) which is due to the fact that 93mNb ejects mainly conversion electrons. Hence, both the variability of the counting efficiency and the related uncertainty are very small.

With the known value of PEC,isomer, it is straight forward to compute the counting efficiencies. All required corrections were applied (background subtraction, consideration of ingrowing 93mNb, decay corrections). The quenching indicator tSIE measured in the TriCarb was found to be slightly biased. The bias was identified by comparing the 93Mo sample without nitromethane and the background sample which have basically the same composition. Such a bias was also observed for a few other radionuclides and may be explained by an overlap of the LS spectrum created by the radionuclide under study and the spectrum created due to the external 133Ba standard in the counter. After application of the correction, the TriCarb results (activity concentrations) are found to be about 0.45% larger than the uncorrected data. The Wallac data did not show any bias for the quenching indicator SQP(E), and consequently, no correction was needed.

A figure that shows the determined activity concentration from all samples as determined using the TDCR and CNET method, respectively, is given in the Supplementary material. As expected, the TDCR data with (double) counting efficiency of more than 52% show a larger spread which can be explained from the TDCR efficiency curve (Fig. 5). Hence, these data are discarded. The overall spread of remaining results (Fig. 7) is much lower (standard deviation 0.22%).

Figure 7
figure 7

Activity concentration as obtained for the CNET and the TDCR method without data that were discarded. The uncertainty bars represent only a statistical component which was calculated as a standard deviation of the mean of several repetition measurements.

The unweighted mean of the CNET results is aCNET = 20.06(24) kBq/g with a relative standard uncertainty of 1.16%. The unweighted mean of the TDCR results is aTDCR = 20.03(22) kBq/g with a relative standard uncertainty of 1.11%. The results agree very well. Since the uncertainties are similar too, the final result is obtained as the unweighted mean, which is a = 20.04(24) kBq/g. This result is also shown as dashed line in Fig. 7. The uncertainty was taken from the CNET result as a conservative approach. The probability of EC decay to the isomer state 93mNb was determined to be PEC,isomer = 95.7(16)% and, consequently, the probability for the EC branch to the ground state is 1 − PEC,isomer = 4.3(16)%. Details for this determination are given in the Supplementary materials. All activity concentrations were stated for a reference date 1 July 2019, 0:00 CET. Full uncertainty budgets are shown in Supplementary material. For more detailed information on the uncertainty assessment in LS counting see31.

The half-life of 93Mo and discussion

The half-life of 93Mo was calculated based on the measured number of 93Mo (\({N}_{{93}_{\mathrm{Mo}}}\)) atoms in the master solution and its measured specific activity (\({A}_{{93}_{\mathrm{Mo}}}\)) using the formula:

$$T_{1/2} = \frac{{\ln (2) \cdot N_{{93_{{{\text{Mo}}}} }} }}{{A_{{93_{{{\text{Mo}}}} }} }}$$

where \({N}_{{93}_{\mathrm{Mo}}}\) is the number of 93Mo atoms per gram of the master solution 4.415(23)∙1015 atoms/g and \({A}_{{93}_{\mathrm{Mo}}}\) its activity concentration 20.04(24) × 103 Bq/g. For the half-life calculation one year is assumed to have 365.242198 days. Using these numbers, the half-life value of 93Mo was determined to be T1/2 = 4839(63) years. At the same time the probabilities of electron-capture decay to the radioactive isomer 93mNb was re-determined yielding the value of 95.7(16)%. The obtained half-life value is compatible with the currently adopted value of 4000(800) years while the uncertainty has been decreased by a factor of 15. The new value for the probability of the decay to the radioactive isomer 93mNb 95.7(16)% is by factor of 8 more precise than the currently accepted value of 88(12)%. The precise determination of the 93Mo half-life and its decay scheme provides now the possibility towards more accurate radionuclide inventory calculations together with better estimates of the radioactive dose emission from low and intermediate nuclear wastes over long time periods.