Abstract
N95 filtering facepiece respirators (FFRs) are essential for the protection of healthcare professionals and other high-risk groups against Coronavirus Disease of 2019 (COVID-19). In response to shortages in FFRs during the ongoing COVID-19 pandemic, the Food and Drug Administration issued an Emergency Use Authorization permitting FFR decontamination and reuse. However, although industrial decontamination services are available at some large institutions, FFR decontamination is not widely accessible. To be effective, FFR decontamination must (1) inactivate the virus; (2) preserve FFR integrity, specifically fit and filtering capability; and (3) be non-toxic and safe. Here we identify and test at-home heat-based methods for FFR decontamination that meet these requirements using common household appliances. Our results identify potential protocols for simple and accessible FFR decontamination, while also highlighting unsuitable methods that may jeopardize FFR integrity.
Similar content being viewed by others
Introduction
Coronavirus Disease of 2019 (COVID-19), caused by the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), is predominantly transmitted through exposure to short-range droplets and aerosols1. To protect against infectious airborne particles2, filtering facepiece respirators (FFRs) such as N95 respirators3 are recommended by the United States Center for Disease Control and Prevention (CDC) for healthcare professionals4. With appropriate donning and doffing procedures, combined with proper fit, N95 FFRs can effectively limit cross-contamination and the transmission of COVID-195.
The global increase in demand for personal protective equipment (PPE) during the COVID-19 pandemic has resulted in FFR shortages3. This led the Food and Drug Administration (FDA) to issue an emergency use authorization for N95 FFR reuse and decontamination during the pandemic6. To decontaminate FFRs, a minimum 3-log (99.9%) reduction in SARS-CoV-2 viral activity7 needs to be achieved. In addition to inactivating the virus, decontamination strategies must maintain overall FFR integrity, as measured by fit (ability to form a tight seal) and filtration (ability to limit penetration of particulates)8.
Several decontamination methods have been shown to satisfy these criteria, including, ultraviolet (UV) germicidal irradiation (radiation-based); hydrogen peroxide vapor (HPV; chemical-based); and heat-humidity treatments9. Unfortunately, these procedures can be difficult to implement at a large scale, and/or present safety risks at a small scale, especially in resource-limited environments. HPV, for example, is a severe irritant that can be hazardous to the lungs and eyes. Most UV-C light sources have spectral variation, and emit wavelengths that can damage the skin or eyes. Proper decontamination by UV-C light requires every surface to receive a sufficient dose of radiant energy, which is challenging to achieve for FFRs with three-dimensional shapes and metal inserts10.
Another significant obstacle is that methods that provide full sterilization using heat, such as autoclaving, are likely to damage FFR integrity. As a result, existing FFR decontamination methods, even though effective against SARS-CoV-2, do not typically eliminate all pathogens. To prevent cross-contamination from pathogens that haven’t been eliminated, decontaminated FFRs need to be returned to the original user, posing a major logistical challenge on an institutional scale, particularly during a healthcare crisis. Therefore, establishing safe and economical decontamination methods that can be implemented by the end user is both necessary and urgent.
In this study, we recognized the potential of widely available, commercial household heat-generating appliances for FFR decontamination. We evaluated the ability of several appliances to maintain decontamination conditions, and tested the effects of repeated treatments on FFR fit and material filtration performance. Of the devices tested, we identify rice cookers with thermostatically-controlled warming functions as the most reliable at maintaining conditions for heat-based decontamination while retaining FFR fit and filtration integrity for up to five treatment cycles. We confirmed that boiling respirators is damaging to FFR fit.
Results
Establishing decontamination parameters
Previous studies suggest that temperatures of 70–85 °C for 30 min at 50% humidity11,12 or 60 min with dry heat (increased temperature alone without added moisture)13 are sufficient to inactivate SARS-CoV-2 from FFR fabric, while also preserving fit and filtration efficiency for up to five rounds of treatment, in line with CDC’s recommendations9. In contrast, treatment with steam autoclaves or immersion boiling should inactivate SARS-CoV-2 from FFRs, but is likely to reduce FFR filtration efficiency and impact fit after as few as one round of treatment14,15,16 (Supplementary Table S1).
Identifying home appliances capable of accurately attaining decontamination conditions
After considering a wide range of commercial heat-generating appliances we excluded those with (1) minimum temperatures above 70 °C, such as standard gas ovens21, (2) maximum temperatures below 70 °C, such as commercial tumble dryers and hair dryers22 and (3) appliances that pose potential safety hazards, including microwaves which17,18 may cause FFRs with metallic components to spark or the plastic in the straps to melt19,20. Based on these temperature and safety criteria, we selected rice cookers with Keep Warm settings and sous vide (immersion cooker) machines for further evaluation21,22. Two types of rice cookers with Keep Warm settings—the Cuckoo (CR-0655F, retail ~ $110), and the Aroma (ARC-743-1NGB, retail ~ $20)—and an AuAg Immersion Circulator sous vide machine (A808, 950 W, retail ~ $50) were selected for testing.
Assessing basic device performance
Initial testing was performed to confirm whether each device could reach and maintain the target temperature of 70 °C for one hour (Supplementary Table S2). The Keep Warm setting of the Cuckoo rice cooker (71.7 °C ± 0.3 °C) and the AuAg Immersion Circulator sous vide machine (70.1 °C ± 0.7 °C) were confirmed to have precise temperature control. In contrast, the Keep Warm setting of the Aroma rice cooker does not allow for precise temperature control, with recorded temperatures of 67.5 °C ± 7.8 °C across the hour-long treatment, this appliance was excluded from all subsequent tests. This result highlights the importance of confirming each device’s ability to maintain the specified temperature range before use.
We also tested for spatial variations in temperature in the Cuckoo rice cooker by measuring the temperature at one central and two peripheral locations of an untreated test FFR (Supplementary Table S2, Supplementary Fig. S1). Confirming an even distribution across the FFR during the treatment cycle, the temperature at all three FFR locations stabilized close to 70 °C after the initial heating period (70.5 °C ± 0.4 °C at both lateral locations and 71.7 °C ± 0.3 °C at the central location).
Evaluating the effects of appliance-based heat treatment on quantitative FFR fit
We tested the effects of three different decontamination protocols on FFR fit and function: (1) boiling at 100 °C for 10 min, (2) warming to 70 °C for 60 min using the Cuckoo rice cooker and (3) warming to 70 °C for 60 min using the sous vide machine (Fig. 1). Even though boiling was expected to impair FFR function, it was tested as it would be particularly simple to perform at home8,15.
Experiments were performed on two models of donated NIOSH-approved FFRs: the Halyard Fluidshield 3 (TC-84A-7521) (Supplementary Fig. S2a) and the Uniair San Huei SH3500 (TC-84A-4313) (Supplementary Fig. S2b). The Halyard Fluidshield 3 filter facepiece respirator was included because it is a commonly used respirator at major hospitals. The Halyard Fluidshield 3 is a surgical N95 duck-billed respirator (SN95 FFR) with enhanced splash protection. The Uniair San Huei was among the few FFRs that could be procured when there were severe FFR shortages, and was chosen for comparison to the Halyard FFR. The Uniair San Huei is a foldable, cup-shaped respirator without additional splash protection (N95 FFR). This research was conducted when PPE supplies were limited so care was taken to minimize the total number of FFRs used in these experiments (Fig. 1 and Supplementary Methods for full details of treatments and fit testing protocols).
Individual FFRs underwent one to five rounds of heat treatment (in a folded configuration) followed by quantitative fit testing and material filtration assessment (Fig. 1b). Control FFRs were not exposed to any treatment (Fig. 2a). FFRs assigned to the rice cooker were placed in breathable paper bags to mimic how a home user would be advised to handle contaminated FFRs (Fig. 2f)9. For heat treatment with the sous vide machine, FFRs were sealed in a polypropylene plastic bag before being placed in water and two 140 g weights were placed on top of the plastic bag to ensure full submersion (Fig. 2g). For boiling, the fabric was immersed directly in water while the elastic straps were kept above the water line to prevent altering their elasticity (Fig. 2b) in accordance with a previous study8. For all three conditions, extra care was taken to avoid deforming the FFR.
Treatment-exposed and control FFRs were fit tested using the built-in N95 setting of a TSI PortaCount Pro + Respirator Fit Tester 8038. Quantitative fit of FFRs is tested for each individual user in seven standardized test conditions: normal breathing, deep breathing, head side to side, head up and down, talking, bending over, and normal breathing. These results are commonly combined into an overall fit factor computed as the geometric mean of the individual conditions. FFRs with a mean fit factor of 100 or greater, pass the fit test. Three technical replicates of all measures were taken for each FFR (Fig. 2a–d). To assess re-sealing of the mask after doffing and re-donning, other standard tests often additionally include a grimace exercise designed to break the face seal23. As FFRs may not re-seal in the same way as during initial donning for subsequent exercises and, consequently, impact fit unpredictably23, this test condition was excluded from the current protocol. Instead, re-sealing capabilities of the FFR was assessed through repeated fit testing after doffing and re-donning.
As shown in Fig. 2c, exposing Halyard Fluidshield 3 FFRs to 70 °C of dry heat for 60 min in the Cuckoo rice cooker up to five times did not decrease their overall fit quality significantly (F(1, 4) = 2.57, p = 0.18). Treatment with the sous vide machine and boiling, on the other hand, significantly decreased fit quality of Halyard Fluidshield 3 FFRs, leading to failed fit testing after five (F(1, 4) = 36.76, p < 0.01; Fig. 2d) and one (F(1, 4) = 12.54, p < 0.05; Fig. 2b) cycles of treatment, respectively (Supplementary Table S3). We conclude that, among the three tested methods, the Cuckoo rice cooker, and by extension, dry heat, is best at preserving Halyard Fluidshield 3 FFR fit.
Even untreated, the Uniair FFR fit our test subjects poorly. It repeatedly failed preliminary user seal tests and produced variable quantitative fit test results, often with the same mask failing two out of three replications of the control fit test (Supplementary Figs. S3, S4, Supplementary Tables S4 and S5). This contrasts with the Halyard Fluidshield 3 FFR which, when correctly sized to our testers, consistently passed preliminary user seal tests and produced more consistent quantitative fit test scores. Based on the poor fit performance and lack of trends in fit quality with treatment conditions (Supplementary Fig. S4), we cannot draw meaningful conclusions on how the number of treatment repetitions affected the Uniair FFR fit. These results underscore the importance of performing regular quantitative fit testing of FFRs to each individual user to ensure proper fit under a respiratory protection program24. While it is possible that trends in fit following heat-treatment hold for similarly manufactured FFRs, the conclusions of this study regarding FFR fit after heat treatment pertain primarily to the Halyard Fluidshield 3 (TC-84A-7521).
Comparison of individual fit test categories gives insights into how to improve fit
Quantitative fit testing of FFRs involves a series of seven test conditions (normal breathing, deep breathing, head side to side, head up and down, talking, bending over, and normal breathing), but results are often only examined using the mean fit factor. However, a more detailed examination of results for each part of the test can provide valuable insights (Fig. 3).
FFRs treated in the Cuckoo rice cooker passed the overall fit test after five treatment cycles, however, there was a progressive decrease in the goodness of fit when the subject was bending over (Fig. 3, Supplementary Table S6). Boiling and heat treatment using a sous vide machine progressively decreased fit scores in all conditions. For sous vide machine treatment, while FFRs still passed the overall fit test after four rounds of treatment (Fit Factor (FF) = 123), fit factors while bending over were significantly lower than in untreated FFRs (p < 0.05) (for further details see Fig. 3, Supplementary Tables S6, S7, S8).
Our results suggest that rice cookers and sous vide machines are viable options for the decontamination of N95 FFRs. Moreover, they highlight the (1) importance to make technical replicate measurements, which is currently not required for fit testing, and (2) record each category’s fit factor. While dry heat is the method that preserves fit best, we propose that paying special attention to movement-specific fit testing results when fitting a respirator could aid the end user to adjust their behavior while wearing FFRs (e.g. kneeling vs. bending).
Evaluating the effects of appliance-based heat treatment on FFR filtration performance
The effect of appliance-based heat treatment on FFR filtration performance was evaluated for treatment and control FFRs that had passed at least one round of treatment (FF ≥ 100, Uniair FFRs were excluded). We developed a modified version of the standard NIOSH test (TEB-APR-STP-0059) to assess the penetration of non-neutralized saline aerosols as the difference in particle counts before and after passing an aerosol through the FFR material (Fig. 4a). Filtration efficiency (FE) was measured for particles in six different size classes (0.3 μm, 0.5 μm, 1 μm, 2.5 μm, 5 μm, and 10 μm) by a calibrated particle counter. Like similar custom-built devices22, our apparatus was designed to normalize for constant flow rates regardless of filter impedance (Fig. 4a, Supplementary Table S9). Furthermore, our experiments were conducted with a steady flow rate per unit area of the filter that exceeds the rates expected under standard NIOSH testing conditions, which generally decreases filtering efficiency compared to lower flow rates25,26,27, suggesting a conservative estimate of FE. Testing was performed on two different central sites on each heat-treated and control FFR.
Comparisons between filtration efficiencies in the control and heat-treated FFRs indicate that none of the tested decontamination strategies reduced FFR FE below the 95% required of ‘N-95’ FFRs (Fig. 4b, Supplementary Table S9). To both validate the sensitivity of our custom-built filtration test rig and to demonstrate how even small leaks can significantly affect effective filter performance, we used an 18-gauge needle to introduce small holes in the filter after normal testing. FE decreased as a function of the number of pinholes introduced into the FFR material (Supplementary Fig. S5), and pressure across the FFR material decreased at a flow rate of 10 lpm (Supplementary Table S10). The validity of these results is further supported by passing scores during the quantitative fit tests, as any damage to the N95 FFR material would compromise the overall filterability of the respirator and result in the FFR failing the fit test. On the other hand, FE testing also demonstrates that flow and filtration properties of the FFR material are not the only important consideration, as proper fit is critical to maintain a high level of protection.
Discussion
It has been previously shown that SARS-CoV-2 can be inactivated in FFRs by 1-h treatments with dry 70 °C dry heat13. Under these conditions, some FFRs have been shown to preserve fit and filtration efficiency for up to ten decontamination cycles13,28. Here we assessed three heat-based methods that are likely to inactivate SARS-CoV-2 using common household appliances and simple methods: a rice cooker, a sous vide machine and boiling in water.
We compared the fit and filtration properties of two FFR models, the Halyard Fluidshield 3 and the Uniair San Huei, after up to five sequential rounds of heat treatment to those of untreated FFRs. Even before treatment, the Uniair San Huei FFR exhibited widely varying fit test results in our pool of testers, and was thus excluded from further analyses. This result emphasizes the importance of fitting FFRs to each individual user, as proper fit is paramount for appropriate protection.
We show that the Halyard Fluidshield 3 fit and filtration efficiency are preserved for up to five rounds of treatment using the Keep warm function of a rice cooker with precise temperature control. As the CDC does not recommend re-using FFRs more than five times, we did not subject the respirators to more than five rounds of treatment9. Sous vide machine treated FFRs preserved fit only up to four rounds, possibly because the respirator was slightly deformed by the water and weights. FFRs boiled in water failed the fit test after just one round of treatment.
By examining fit test results in more detail, we found that for all decontamination strategies the fit factor, and therefore protection, was reduced the most and after the fewest treatment cycles while bending over (Fig. 3). This indicates that a poor fit under this condition may serve as an early warning sign for a decrease in overall fit.
Flow and filtration analysis of treated FFRs demonstrated that none of the heat treatment methods lowered filtration efficiency below the required 95%. Interestingly, even boiling for up to three treatment cycles did not lower the FE of the Halyard Fluidshield 3 FFRs below this threshold. The effects of boiling on FE have been conflicting, with some studies reporting > 92% efficiency after boiling8, and others reporting < 90%29. Variability in the FFR models used and the challenge particles tested may contribute to these differences. Here we used a FFR that was designed to protect against fluid splashes in the operating room and is considered to be water-resistant. Moreover, we also used non-neutralized challenge particles, which can interact with the FFR’s surface charge and affect absolute FE measurements. Further research is warranted to determine whether splash-protective respirators, in general, are more resistant to changes in FE during humid or water-immersive treatment conditions than conventional FFRs8,28, although it should be highlighted that we saw significant degradation in fit and thus efficacy with boiling, making it an unsuitable approach for decontamination overall.
While not recommended for clinical or routine use, the decontamination methods evaluated here demonstrate that household appliances can constitute a low-cost strategy for N95 decontamination that could be performed in resource-constrained settings or situations where no viable alternative exists. Before attempting decontamination, however, it is essential for the user to verify that the selected appliance can stably maintain a temperature of 70 °C for 60 min with spatial homogeneity across the device. Additionally, after each decontamination cycle, FFRs need to be examined for damage, burns or tears. As highlighted by differences in the goodness of fit between FFR models and between individual FFRs exposed to the same decontamination treatment, it is imperative to carefully examine FFR fit before the first use and after each treatment cycle to ensure tight fit of the FFR all around the face by performing a user seal check20.
It is also important to note that exposure to dry heat at 70 °C does not result in sterilization, the complete elimination of all forms of microbial life. Pathogens such as Clostridium difficile will not be eliminated by treatment methods like the ones described in this study, and thus, treated FFRs should only be re-used by the same individual30.
Materials and methods
The Columbia University Institutional Review Board reviewed this study’s protocol and determined that ethical approval could be waived (IRB reference AAAT5503). All necessary participant consent was obtained.
Assessing basic device performance
Initial assessment of whether each device could reach and maintain the target temperature of 70 °C for the treatment duration of one hour was performed. For each rice cooker, a temperature sensor was placed in an elevated plastic tub containing 120 ml of water within the inner pot while the rice cookers were heated using the Keep Warm setting (Supplementary Table S2). For the sous vide machine, a temperature sensor was immersed in a polystyrene box with 3.2 L of water and warmed using a sous-vide temperature setting of 70 °C. All device temperatures were recorded during initial heating and then for one hour after each appliance’s temperature had stabilized. The total process time, including time to reach the target temperature, were also recorded (Supplementary Table S2, Supplementary Fig. S1).
To confirm that the temperatures were distributed evenly across the FFR during the treatment cycle (70 °C for 60 min), we repeated these measures in the Cuckoo rice cooker at one central and two peripheral locations of an untreated test FFR (Supplementary Table S2, Supplementary Fig. S1). For the Cuckoo rice cooker, the temperature at all three FFR locations stabilized close to 70 °C after the initial heating period (70.5 °C ± 0.4 °C at both lateral locations and 71.7 °C ± 0.3 °C at the central location).
FFR treatment
We tested the effects of three different decontamination protocols on FFR fit and function: (1) boiling at 100 °C for 10 min, (2) 70 °C for 60 min using a Cuckoo rice cooker and (3) 70 °C for 60 min using a sous vide machine (Fig. 1). Tests were performed on two NIOSH-approved N95 FFR models, the standard Halyard Fluidshield 3 (TC-84A-7521) and the Uniair San Huei SH3500 (TC-84A-4313) (Supplementary Fig. S2). FFRs were randomly assigned to either one of the three heat-treatment or a control group. Both the experimenter and test subject were blind to this assignment. FFRs were subjected to up to three cycles of decontamination by boiling or up to five cycles of dry heat decontamination (Fig. 1).
Temperature within each appliance used for dry heat treatment was recorded throughout the treatment with a wireless sensor (Inkbird Thermometer IBS-TH1) for the rice cooker and MeatStick Wireless Thermometer (4335995327) for the sous vide machine. Both thermometers were benchmarked against a calibrated thermometer (Oakton Acorn Series pH 5 Meter and Thermometer). For dry heat treatment in the rice cooker, FFRs were placed in breathable paper bags inside the preheated rice cooker (Fig. 2e) after which the temperature was maintained at 70 °C for 60 min. For sous vide treatment, FFRs were sealed in Double Protection Ziploc© Freezer Bags before being placed in water preheated to 70 °C. In order to completely submerge the FFRs, two 140 g weights were placed on top of the plastic bag (Fig. 2g). The temperature of the sous vide machine was set to 70 °C for 60 min. FFRs subjected to boiling were submerged directly in 1.3 L of boiling water at 100 °C within a 1.4 L pot (Fig. 2c) and were drained and allowed to air-dry overnight post-treatment.
FFR fit test
Quantitative fit tests were performed using the manufacturer provided “N95” setting of a calibrated TSI PortaCount Pro + Respirator Fit Tester 8038 with ambient particulates and aqueous aerosols produced by an ultrasonic humidifier (New York city tap water). Adherent to the Occupational Safety and Health Administration’s (OSHA’s) Respiratory Protection Standard (29CFR 1910.134), the Condensation Nuclei Counter (CNC) Quantitative Fit Test for Filtering Facepiece Respirators was used24. This test reports a quantitative fit factor score for sequential test conditions (normal breathing, deep breathing, head side to side, head up and down, talking, bending over, and normal breathing) as well as an overall fit factor computed as the mean of the scores achieved in the individual conditions. Based on the OSHA Respiratory Protection Standard, an overall fit factor of ≥ 100 is required for passing, regardless of performance on each test component24. It should be noted that the system gives a maximum possible score of 200 + reported as 20131.
The CNC quantitative fit test was performed three times for each FFR. All tests were performed on the same test subject to control for variability in facial features. As per CDC recommendations, a user seal check was performed prior to each fit test measurement and adjustments were made until the user deemed that there were no detectable air leaks during forceful breathing9.
FFR filtration efficiency test
Due to the limited availability of NIOSH test apparatus, a modified version of the standard procedure (TEB-APR-STP-0059) was developed to determine the particulate filter efficiency of N95 FFRs24. FFR fabric was challenged by a NaCl aerosol generated using a humidifier (Vicks© Filter-Free Cool Mist Humidifier) filled with 2% saline solution. Non-neutralized particles were released into a polycarbonate test chamber (~ 150 L) with a minimum 0.3-μm particle count of 5000 particles per ft3. Each FFR was manually secured over an iso-KF 25 fitting. After ~ 10 cm of ~ 25 mm ID stainless steel tubing, another KF-25 to barbed fitting joined via 3/8″ ID silicone tubing connected to the both the calibrated particle counter (Extech Instruments VCP300) and a high-performance surgical smoke removal device used as an evacuator (Buffalo Filter VisiClear) (Fig. 4a). Pressure drop across the filter was monitored with a Sensirion SDP8-500 differential pressure sensor, and the air flow volume was monitored by a Sensirion Mass Flow Meter (part number SFM3300-D). Both devices were read through their I2C communications channel through a custom program running on an Arduino Uno and then recorded on a PC at rates over 10 Hz.
Tests were conducted at a flow rate of 10 lpm with a 4.54 cm2 area of effective filtration, which exceeds the rate of 85 lpm used during NIOSH testing when our filter patch is scaled proportionally to the area of the entire FFR32. We compensated for any variation in the impedance of the FFR material by changing the strength of the suction of the evacuator to obtain a flow rate of 10 lpm through each FFR or blank prior to each filtration test.
Before measuring the filtration efficiency of each treated or control FFR, the baseline ambient particle count within the aerosol shield chamber was measured as the particle count downstream of an open nozzle. First, particles were suctioned through either a blank or FFR at a flow rate of 10 lpm for 20 s, then the suction was switched off. Following a 5 s pre-counting sample draw, the cumulative number of same-sized particles downstream of each FFR or blank were counted over a 20 s sampling period at a 2.83 lpm sampling rate controlled by the particle counter’s internal pump. The total volume of the internal tubing, sensors, and stainless port was ~ 1.35 L, and designed such that the bulk of the air sampled by the particle counter during its total 25 s counter operation was air drawn through the filter unit at the 10 lpm volume flow rate, not air drawn through the filter by the pump action of the particle counter. Particles binned into 0.3 μm, 0.5 μm, 1 μm, 2.5 μm, 5 μm, and 10 μm size classes were measured. The filtration efficiency (FE) was calculated using the equation,
where \({C}_{mask}\) and \({C}_{chamber}\) denote the concentrations of particles downstream of the FFR and within the aerosol chamber, respectively. For each FFR, flow test measurements of two different central N95 material areas were obtained to sample the variance.
To assess the sensitivity of our modified flow test, two FFR samples were punctured with an 18-gauge needle up to three times (Supplementary Fig. S5, Supplementary Table S10). Each puncture was at most 0.3% of the filter area, yet reduced filtering efficiency of the 0.3 μm particles by roughly 25% (e.g., from > 95 to 75%). This large reduction in efficiency from such a small leak highlights the importance of proper fit, with no voids, for effective protection.
Statistical analysis
All statistical analyses were run in the R software environment (v4.0.1). The decrease of fit factors over repeated treatment was evaluated using linear regressions (stat_smooth function, method = lm, ggplot2 v3.3.0). Variance component analyses were calculated with nested one-Way ANOVA (aov(score ~ treatment + Error(FFR)). F-values and significance levels for the F-statistics reported33.
Data availability
All data is available in the main text or the Supplementary materials.
References
Cook, T. M. Personal protective equipment during the coronavirus disease (COVID) 2019 pandemic—A narrative review. Anaesthesia 75, 920–927 (2020).
Santarpia, J. L. et al. The size and culturability of patient-generated SARS-CoV-2 aerosol. J. Expo. Sci. Environ. Epidemiology. 1–6. https://doi.org/10.1038/s41370-021-00376-8 (2021).
Tabah, A. et al. Personal protective equipment and intensive care unit healthcare worker safety in the COVID-19 era (PPE-SAFE): An international survey. J. Crit. Care 59, 70–75 (2020).
CDC. Infection Control Guidance for Healthcare Professionals about Coronavirus (COVID-19). https://www.cdc.gov/coronavirus/2019-ncov/hcp/infection-control.html. Accessed 30 Aug 2020 (2020).
Wang, X., Pan, Z. & Cheng, Z. Association between 2019-nCoV transmission and N95 respirator use. J. Hosp. Infect. 105, 104–105 (2020).
CDC. Recommended Guidance for Extended Use and Limited Reuse of N95 Filtering Facepiece Respirators in Healthcare Settings. https://www.cdc.gov/niosh/topics/hcwcontrols/RecommendedGuidanceExtUse.html. Accessed 30 Aug 2020 (2020).
FDA. Enforcement Policy for Face Masks and Respirators During the Coronavirus Disease (COVID-19) Public Health Emergency (Revised). https://www.fda.gov/media/136449/download. Accessed 30 Aug 2020 (2020).
Juang, P. S. C. & Tsai, P. N95 respirator cleaning and reuse methods proposed by the inventor of the N95 mask material. J. Emerg. Med. 58, 817–820 (2020).
CDC. COVID-19 Decontamination and Reuse of Filtering Facepiece Respirators. https://www.cdc.gov/coronavirus/2019-ncov/hcp/ppe-strategy/archived-decontamination-reuse-respirators.html. Accessed 30 Aug 2020 (2020).
Andersen, B. M., Bånrud, H., Bøe, E., Bjordal, O. & Drangsholt, F. Comparison of UV C light and chemicals for disinfection of surfaces in hospital isolation units. Infect. Control Hosp. Epidemiol. 27, 729–734 (2006).
Massey, T. L. et al. Quantitative Fit Evaluation of N95 Filtering Facepiece Respirators and Coronavirus Inactivation Following Heat Treatment. Ann Work Expo Heal. https://doi.org/10.1093/annweh/wxab020 (2021).
Anderegg, L. et al. A scalable method of applying heat and humidity for decontamination of N95 respirators during the COVID-19 crisis. PLoS One 15, e0234851 (2020).
Fischer, R. J. et al. Effectiveness of N95 Respirator Decontamination and Reuse Against SARS-CoV-2 Virus. Emerg. Infect. Dis. 26, 2253–2255 (2020).
van Straten, B. et al. Sterilization of disposable face masks by means of standardized dry and steam sterilization processes; an alternative in the fight against mask shortages due to COVID-19. J. Hosp. Infect. 105, 356–357 (2020).
Kumar, A. et al. Decontamination of N95 masks for re-use employing 7 widely available sterilization methods. Plos One. 15, e0243965 (2020).
N95Decon. Heat and Humidity for Bioburden Reduction of N95 Filtering Facepiece Respirators. http://www.n95decon.org/files/heat-humidity-technicalreport. Accessed 30 Aug 2020 (2020).
Appliances, G. E. Wall ovens and ranges. https://products.geappliances.com/appliance/gea-support-search-content?contentId=18501. Accessed 15 Jun 2020.
Gluesenkamp, K. Residential Clothes Dryer Performance Under Timed and Automatic Cycle Termination Test Procedures. https://web.ornl.gov/sci/buildings/docs/2014-10-09-ORNL-DryerFinalReport-TM-2014-431.pdf Accessed 30 Sep 2020 (2014).
Pascoe, M. J. et al. Dry heat and microwave generated steam protocols for the rapid decontamination of respiratory personal protective equipment in response to COVID-19-related shortages. J. Hosp. Infect. 106, 10–19 (2020).
Viscusi, D. J., Bergman, M. S., Eimer, B. C. & Shaffer, R. E. Evaluation of five decontamination methods for filtering facepiece respirators. Ann. Occup. Hyg. 53, 815–827 (2009).
Vieira, F. R. & Pecchia, J. A. An exploration into the bacterial community under different pasteurization conditions during substrate preparation (Composting–Phase II) for Agaricus bisporus cultivation. Microbial Ecol. 75, 318–330 (2018).
Konda, A. et al. Aerosol filtration efficiency of common fabrics used in respiratory cloth masks. ACS Nano 14, 6339–6347 (2020).
Richardson, A. W., Hofacre, K. C., Weed, J., Holm, R. & Remiarz, R. Evaluation of a faster fit testing method for elastomeric half-mask respirators based on the TSI PortaCount. J. Int. Soc. Respir. Protect. 31, 43–56 (2014).
OSHA. 1910.134—Respiratory Protection. https://www.osha.gov/laws-regs/regulations/standardnumber/1910/1910.134. Accessed 30 Sept 2020 (2011).
Gao, S. et al. Penetration of combustion aerosol particles through filters of NIOSH-certified filtering facepiece respirators (FFRs). J. Occup. Environ. Hyg. 12, 678–685 (2015).
Eninger, R. M. et al. Filter performance of N99 and N95 facepiece respirators against viruses and ultrafine particles. Ann. Occup. Hyg. 52, 385–396 (2008).
Huang, S.-H. et al. Factors affecting filter penetration and quality factor of particulate respirators. Aerosol. Air Qual. Res. 13, 162–171 (2013).
Liao, L. et al. Can N95 respirators be reused after disinfection? How many times?. ACS Nano 14, 6348–6356 (2020).
Ou, Q., Pei, C., Kim, S. C., Abell, E. & Pui, D. Y. H. Evaluation of decontamination methods for commercial and alternative respirator and mask materials—View from filtration aspect. J. Aerosol. Sci. 150, 105609 (2020).
Kazanowski, M., Smolarek, S., Kinnarney, F. & Grzebieniak, Z. Clostridium difficile: Epidemiology, diagnostic and therapeutic possibilities—A systematic review. Tech. Coloproctol. 18, 223–232 (2014).
CDC. Determination of Particulate Filter Penetration Test Powered Air-Purifying Respirator Filters Standard Testing Procedure (STP). https://www.cdc.gov/niosh/npptl/stps/pdfs/TEB-APR-STP-0001-508.pdf Accessed 30 Aug 2021 (2009).
Brosseau, L. M., McCullough, N. V. & Vesley, D. Mycobacterial aerosol collection efficiency of respirator and surgical mask filters under varying conditions of flow and humidity. Appl. Occup. Environ. Hyg. 12, 435–445 (1997).
Crawley, M. J. Statistics: An Introduction Using R (Wiley, 2015).
Acknowledgements
We thank Richard Hormigo for circuit design for the filtration efficiency testing rig; David Brenner for helpful discussions; VisiClear for the Buffalo Filter VisiClear; the Zuckerman Mind Brain Behavior Institute and Columbia Environmental Health and Safety for the space to perform experiments; Columbia Researchers Against COVID-19 for assembling this team; Wearing is Caring; Columbia University Institutional Review Board for assessing these experiments; Peter Andolfatto, Rui Costa, Ivaylo Ivanov, and Andrés Bendesky for support throughout the project; Anil Lalwani for providing the FFRs used in this study. Figure 1 was designed and drawn by co-author Elaine Lee.
Funding
The Zuckerman Mind Brain Behavior Institute, Columbia University provided funding to support this project.
Author information
Authors and Affiliations
Contributions
T.X.C., A.P., D.S.P. and E.M.C.H. conceived the study and designed the experiments. T.X.C. and A.P. performed the experiments, analyzed the data and wrote the manuscript. N.A.S. wrote the manuscript. K.Y-M. performed the literature survey. T.X.C., A.P., E.M.C.H. and L.E. made the figures. D.S.P. designed and assembled the FFR filtration efficiency test and provided training. N.A.S., J.H. and L.W. provided specialist support. S.F. provided training on the PortaCountPro+. T.X.C., A.P., N.A.S. and E.M.C.H. wrote and applied for the IRB. All the authors commented and edited the manuscript.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Chen, T.X., Pinharanda, A., Steinemann, N.A. et al. Evaluation of at-home methods for N95 filtering facepiece respirator decontamination. Sci Rep 11, 19750 (2021). https://doi.org/10.1038/s41598-021-99129-8
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-99129-8
This article is cited by
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.