Abstract
Transcatheter embolization is an important treatment method in clinical therapy, and vascular embolization material plays a key role in embolization. The temperature-sensitive p(N-isopropylacrylamide-co-butyl methylacrylate) (PIB) nanogel is a novel embolic agent. To evaluate the feasibility of the nanogel as a blood vessel embolization agent, we aimed to assess the effect of embolization with PIB nanogels in the rabbit renal artery by non-invasive computed tomography (CT) perfusion, macroscopic and histological examination. Ten healthy adult Japanese rabbits were used to implement RAE of PIB nanogels in their right kidneys. CT perfusion scans were performed pre- and post-treatment at various time-points (1, 4, 8, and 12 weeks). Two rabbits were euthanized and histologically examined at each time-point, and the remaining rabbits were euthanized at 12 weeks after embolization. The RAE efficacy of the nanogels was further confirmed by macroscopic and histological examination. The renal volume and renal blood flow (BF) of the right kidney were significantly decreased post-treatment compared with those pre-treatment (volume: pre, 9278 ± 1736 mm3; post 1 week, 5155 ± 979 mm3, P < 0.0001; post 4 weeks, 3952 ± 846 mm3, P < 0.0001; post 8 weeks, 3226 ± 556 mm3, P < 0.0001; post 12 weeks, 2064 ± 507 mm3, P < 0.0001. BF: pre, 530.81 ± 51.50 ml/min/100 ml; post 1 week, 0 ml/min/100 ml, P < 0.0001; post 4 weeks, 0 ml/min/100 ml, P < 0.0001; post 8 weeks, 0 ml/min/100 ml, P < 0.0001; post 12 weeks, 0 ml/min/100 ml, P < 0.0001). No revascularization or collateral circulation was observed on histological examination during this period, and PIB nanogels were dispersed in all levels of the renal arteries. Twelve weeks after embolization, CT perfusion showed no BF in the right renal artery and renal tissue, a finding that was consistent with histological examination showing complete embolization of the right renal artery with a lack of formation of collateral vessels. The effect of embolization on PIB was adequate, with good dispersion and permanency, and could be evaluated by non-invasive and quantitative CT perfusion.
Similar content being viewed by others
Introduction
With the rapid development of interventional radiological technology, blood vessel embolism materials (including metal coils, anhydrous ethanol, Lipiodol, onyx, and polyvinyl alcohol (PVA)) have expanded the clinical indications, such as the treatment of postpartum hemorrhage, gastrointestinal bleeding1,2,3, preoperative embolotherapy of tumors4,5, and embolization of vascular malformations6. However, the clinical application of these materials has some complications, including poor flowability of the solid embolic agent that makes it difficult to uniformly embolize the peripheral vessel. Additionally, solid embolic agents such as spring coils cause obstructive nephropathy7. Although liquid embolic agents have good flowability, Lipiodol embolization is often incomplete or totally eradicated by tissue, pulmonary hypertension is often secondary to ethanol embolism8, and high concentrations of onyx can result in neurotoxicity9. The ideal vascular embolization material has the following characteristics: good dispersibility, radiopacity, biocompatibility and embolism permanency10,11. Therefore, it remains a considerable challenge to develop safe and effective liquid embolic materials with a peripheral embolization effect.
Temperature-sensitive nanogels have attracted increasing attention during the last decades because they undergo reversible and rapid volume phase transitions in response to ambient temperature. First, these nanogels have good fluidity in fine catheters because they remain in the liquid state at a low critical solution temperature (LCST). Second, their small size enables them to overcome some biological barriers. Furthermore, a gel forms in situ, conforming to the shape and size of the blood vessels, minimizing trauma to non-target vessels and reducing the use of nanogels. In addition to good dispersibility and permanent embolism, the p(N-isopropylacrylamide-co-butyl methylacrylate) (PIB) nanogels developed by us exhibited a lower inflammatory vascular response, better fluidity, and better operability than PVA-embolized particles and Lipiodol12,13. Temperature-sensitive PIB nanogels have been used as novel blood vessel embolic materials to solve the dilemma of peripheral artery embolization and permanent embolization12,13,14. With the progress of preparation technology, the resultant PIB nanogels have become more biocompatible, and the convenience of operation has been markedly improved. Since PIB nanogels have different mechanisms of temperature sensitive artery embolization from other clinical embolic agents, it is difficult to precisely evaluate their performance in renal artery embolization (RAE).
Digital subtraction angiography (DSA) has long been the imaging gold standard in the evaluation, treatment, and follow-up of vascular disorders15. Some researchers have attempted to use DSA to assess the long-term effect of embolization12,13,14,16. However, a major drawback of DSA is the invasiveness of vascular puncture. Computed tomography (CT) perfusion imaging is a new, accurate, non-invasive, functional imaging technology that can provide information on hemodynamic changes in organs and tissues17,18. One study showed that CT perfusion imaging could provide data concerning renal morphological changes and quantitative renal hemodynamic data19. Thus, it is reasonable to use CT perfusion to assess the embolization effect of PIB in the renal artery.
In our previous study13, PIB was first developed as a novel temperature-sensitive blood vessel embolic agent to address the dilemma of flowability and embolization in transarterial chemoembolization of hepatocellular carcinoma. This study aimed to observe the feasibility of PIB as a blood vessel embolization agent using CT perfusion imaging, macroscopic and histological examination, thus providing an experimental basis for the clinical application of a novel embolic agent.
Materials and methods
Experimental materials
PIB nanogels, as an embolic agent in this experiment, were provided by the National Engineering Research Center for Nanomedicine, College of Life Science and Technology, Huazhong University of Science and Technology. The PIB nanogel is a new type of temperature-sensitive nanogel that remains in a liquid state at a low temperature and turns into a gelatinous solid at temperatures above its LCST (35 °C). Because it undergoes a transition from a flowable sol phase to a solidified gel phase at its LCST, the PIB nanogel is in a liquid state at a lower temperature (25 °C, compared with 35 °C) and turns into a solid white gel at a higher temperature (37 °C, compared with 35 °C) (Fig. 1).
Study design
The experiment was designed to assess PIB nanogels and monitor their residue time. The right renal artery of rabbits was embolized with PIB nanogels, and CT perfusion was performed pre- and post-treatment at various time points (1, 4, 8, and 12 weeks). Efficacy was confirmed by macroscopic and histological examination after CT perfusion.
Animal model preparation
Ten healthy adult Japanese long-eared white rabbits were provided by the experimental animal breeding plant of Tongji Medical College, Huazhong University of Science and Technology. The rabbits were aged 3–5 months, had body weights between 2.5 and 3 kg, and were of either sex. All animal procedures were performed in accordance with the ARRIVE guidelines and National Institutes of Health Guidelines for the Care and Use of Laboratory Animals, and the study was approved by the Institutional Animal Care and Use Committee at Tongji Medical College, Huazhong University of Science and Technology (Certificate no. IACUC-2018–2274).
Vascular embolization protocol
First, the animals were fasted for 12 h before each procedure. Sodium pentobarbital (2.0 wt%, 30 mg/kg) was injected intravenously for anaesthesia. Next, interventional procedures were performed using a DSA unit with strict sterile technique. The right femoral artery was isolated and accessed through an open puncture using an 18-gauge sheath needle. After successful puncture, the needle core was withdrawn and a short guidewire was inserted. Then, the sheath of the puncture needle was withdrawn and a 4F arterial sheath was placed into the femoral artery. Sodium heparin (100 IU/kg) was then administered using a three-way switch. A 4F Cobra catheter was inserted into the right renal artery using a 4F arterial sheath. After injecting a contrast agent (Iopamiron, Bayer HealthCare, 350 mg I/mL; 3 ml/s; a dose of 6 ml per rabbit) into the rabbits, renal arterial angiography was performed. After right renal arterial angiography, the PIB nanogels were injected into the rabbit with an average dose of 2 ml per rabbit via the 4F Cobra catheter (Fig. 2). The distal branches of the renal artery were embolized. Next, the catheter was slightly retracted to embolize the proximal vessels of the renal artery. After injection of the PIB nanogels, the catheter was withdrawn, and the distal femoral artery near the puncture point was ligated.
Perfusion CT
The parameters of perfusion CT scanning and contrast agent injection are summarized in Table 1. A 320-detector row CT scanner (Aquilion One; Toshiba Medical Systems) was used in the volumetric scan mode. A 22-gauge needle was placed in the ear vein of the rabbit, and sodium pentobarbital (2.0 wt%, 30 mg/kg) was injected intravenously for anesthesia. Next, a nonionic contrast material (Iopamiron; 350 mg I/mL; Bayer HealthCare) was administered via a power injector (Ulrich CT Plus 150; Ulrich Medical). The scan area of the perfusion CT was set to cover at least both kidneys. During the scan, the rabbits were placed in the supine position, and straps were fixed to the rabbit's subcostal abdominal wall to minimize image artifacts caused by respiratory movements. Perfusion was calculated using the maximum slope model (Body Perfusion; Toshiba Medical System), and the results were shown in ml/min per 100 ml. Additionally, color perfusion maps of renal blood flow (BF) were obtained.
Pathology
Macroscopic and microscopic pre- and post-RAE treatment
At various time points (1, 4, and 8 weeks), two rabbits were euthanized after CT perfusion was completed, the remaining rabbits were euthanized at 12 weeks, and the kidneys were harvested for general pathological observation. Next, a coronal incision of the kidneys was performed, and a dividing ruler was placed next to the kidneys. The macroscopic presentation at each time point was observed, noting the size, shape, texture, color, cortex, and medullary infarction of the right kidney after the RAE, in addition to observing the presence of any surviving kidney tissue. The post-RAE renal tissues were fixed in 10% formalin for 48 h, dehydrated in graded alcohol, and embedded in paraffin. Then, 4- to 5-μm sections were cut. The tissue samples were then stained with Masson’s trichrome at various time points to evaluate ultrastructural changes and estimate the treatment effects.
Statistical analysis
Statistical analysis was performed using SPSS 20.0 and Microsoft Excel. All the data were expressed as mean ± standard deviation. Dunnett’s tests were used to assess the differences in the renal volume and BF in the right kidney pre- and post-treatment during the 12 weeks.
Results
Dispersion and permanency
During the arterial embolization procedure, all levels of the renal arteries (large, small and peripheral) were embolized with PIB nanogels, indicating the PIB nanogels were well dispersed in the blood vessels. A perfusion CT examination was performed to obtain the correlation quantitative index of the paired renal volume and renal BF. These data showed that the right renal volume decreased (pre, 9278 ± 1736 mm3; post 1 week, 5155 ± 979 mm3; post 4 weeks, 3952 ± 846 mm3; post 8 weeks, 3226 ± 556 mm3; post 12 weeks, 2064 ± 507 mm3) and the left side increased in a compensatory manner (Fig. 3a). The right renal BF decreased from 530.81 ± 51.50 ml/min/100 ml to zero after right RAE (Fig. 3b). Significant differences were found in the renal volume (P < 0.0001) and BF (P < 0.0001) of the right kidney pre- and post-treatment during the 12 weeks. Long-term follow-up CT examination showed no BF reperfusion and revascularization after right kidney embolization (Fig. 4). Additionally, the macro-pathological pictures (Fig. 5) of RAE at four various time points (1, 4, 8, and 12 weeks) showed that the right kidney gradually shrank after embolization, and the left kidney enlarged in a compensatory manner. At the first week, the embolized kidney was yellowish white, and the renal medulla was ischemic. At the fourth week, the embolized kidney was yellow, the volume shrank, and the texture became hard, but the renal edge was yellowish white. At the eighth and twelfth weeks, the color of the right kidney was pale, the volume was smaller, and the texture was harder. Additionally, calcification of the right renal tissue was observed. Distinguishing between the cortex and medulla was difficult. The results of macroscopic examination to assess right renal volume changes were similar to those of CT perfusion imaging. These results indicated that PIB nanogels contributed to effective embolization within 12 weeks. The histological results of the right kidney indicated that the nanogels were immobilized in the renal arteries and branches throughout the entire experimental period (Fig. 6). The renal cells and tissues maintained their intact structure and morphology at 1-week post-embolization, while edema and coagulative necrosis occurred (Fig. 6a) and began to disintegrate at 4 weeks post-embolization (Fig. 6b). After embolization for 8 weeks, the pyknosis, rupture, dissolution and disappearance of renal cells indicated that PIB nanogel embolization resulted in obvious embolic gangrene, and the gel remained in the blood vessels (Fig. 6c). The structure and morphology of the renal tissues were entirely destroyed, and significant calcifications were found 12 weeks after embolization (Fig. 6d).
Discussion
In this study, artery embolization efficacy of PIB nanogels was successfully evaluated by CT perfusion scan and histological examination before and after embolization. PIB nanogels exhibited thermo-sensitive sol–gel transition, that is, remained a liquid state at LCST and turned into a gelatinous solid as the ambient temperature increased above LCST. If their LCSTs were tailored close to physiological temperature (37 °C), ‘smart’ PIB nanogels could achieve the sol–gel transition when they entered into blood vessels and are widely used in blood vessel embolic materials11,14, drug delivery20,21,22, precise cancer therapy23,24, tissue engineering25, and cell sheet engineering26,27. Most of the studies focused on their phase transition behavior theory, drug delivery and promising biomedical applications. However, little has been reported on the clinical transition of the smart polymeric materials.
The ideal vascular embolic material has the following characteristics: favorable dispersibility, radiopacity, biocompatibility and embolism permanency10,11. In previous studies, PIB nanogels have been confirmed to be non-adhesive, controllable, and low inflammatory reactions of the blood vessels12. Transcatheter arterial embolization (TAE) treatment is highly dependent on the dispersion and integration of the embolization material. Temperature-sensitive PIB nanogels remain liquid at a temperature lower than their LCST (35 °C) and have good flowability in fine catheters. When they enter in vivo, the body temperature is higher than their LCST (35 °C), PIB nanogels change from a liquid state to a gelatinous solid in response to ambient temperature, and in situ gelate in blood vessels, matching well the shape and size of the targeted blood vessels. PIB nanogels achieved effective RAE that was uniform and persistent because the embolic agent was stable in the embolized vessel throughout the entire 12 weeks, with no microvascular formation in the embolized kidney (Fig. 6). This phenomenon led to a significant drop in the right kidney volume and BF after embolization (Fig. 3). Atrophy and calcification were observed in the right kidneys (Fig. 4). Because of the different proficiencies of the operators, the segment of the renal artery embolism might be slightly different. After embolization of the peripheral vessels, the catheter was slightly retracted to embolize the proximal renal artery. The segment of renal artery embolization varied with the distance of catheter retraction. When the catheter was retracted a shorter distance, the distal renal artery was embolized, resulting in segmental embolization of the renal artery and making the renal artery appear relatively intact (Fig. 5b); however, the kidney was obviously affected. Additionally, the PIB nanogels remained in the blood vessel during the 12-week follow-up based on pathologic examination (Fig. 6), confirming the embolization effect of the PIB nanogels. Regarding the nanogel dose, our previous study had shown that various levels of renal arteries could be embolized by adjusting the nanogel injection dosages, and a dose of 2 ml could embolize all levels of renal arteries (including peripheral, small and large arteries)16. Zhao12 studied the dispersion of the gel according to different injection rates; however, the long-term effects of the treatment were not observed by the imaging methods and remain unknown. Additionally, DSA is a traumatic examination that may cause various complications15,28,29,30, increasing the risk of death in experimental animals. In this study, we observed the dispersion of PIB in the right kidney and evaluated both the long-term effect of treatment by CT, macroscopic and histological examination in a 12-week follow-up examination.
CT perfusion is a functional imaging technique17,31,32 that can noninvasively evaluate the blood supply and kinetics, reflecting physiological function changes. An important use of CT perfusion is the evaluation of cerebral ischemia, which can provide important evidence to support the treatment choice of patients with ischemic stroke33,34. Recently, studies have indicated the potential of CT perfusion to predict angiogenesis and treatment efficacy in multiple cancers, such as hepatocellular carcinoma17, renal cell carcinoma35, lung cancer31, and gastric cancer32, among others. In this study, we evaluated the embolization effect of the right renal artery using CT perfusion, a major advantage of which is its non-invasive nature. Additionally, this method can depict the morphologic characteristics of the kidney and provide additional quantitative factors to reflect the embolization effect. During follow-up, the right kidney volume decreased gradually after embolization, while the volume of the left kidney increased in a compensatory manner after the operation and then decreased slightly. After embolizing the right renal artery, the right renal BF decreased significantly to 0, and the left renal BF increased in a compensatory manner. The lack of vascular enhancement reflects necrosis and is regarded as successful treatment36, demonstrating that the gelling agent has excellent permanency. Macroscopic examination of renal tissue at post-mortem verified the results obtained using CT perfusion; thus, evaluating the embolic effect is feasible by CT perfusion.
The current study has some limitations. First, because of the different proficiency of operators, the position of renal artery embolism was slightly different. Second, the number of rabbit samples was limited, and further studies are warranted. Studies regarding these properties of our new gel are now ongoing.
Conclusion
The effect of embolization on PIB was adequate, with good dispersion and permanency. The nanogel can reach peripheral blood vessels and remain in them for a long time. Thus, this gel is a promising blood vessel embolic material and can be used in vivo. CT perfusion can be used to non-invasively and quantitatively evaluate the embolization effect of temperature-sensitive nanogels.
References
Kim, T. H. et al. Uterine artery embolization for primary postpartum hemorrhage. Iran. J. Reprod. Med. 11, 511–518 (2013).
Park, K. J. et al. Postpartum hemorrhage from extravasation or pseudoaneurysm: efficacy of transcatheter arterial embolization using N-butyl cyanoacrylate and comparison with gelatin sponge particle. J. Vasc. Interv. Radiol. 26, 154–161. https://doi.org/10.1016/j.jvir.2014.10.001 (2015).
Urbano, J., Manuel Cabrera, J., Franco, A. & Alonso-Burgos, A. Selective arterial embolization with ethylene-vinyl alcohol copolymer for control of massive lower gastrointestinal bleeding: feasibility and initial experience. J. Vasc. Interv. Radiol. 25, 839–846, https://doi.org/10.1016/j.jvir.2014.02.024 (2014).
Li, Q. et al. Incidence and therapeutic frequency of extrahepatic collateral arteries in transcatheter arterial chemoembolization of hepatocellular carcinoma: experience from 182 patients with survival time more than 3 years. Eur. J. Radiol. 84, 2555–2563. https://doi.org/10.1016/j.ejrad.2015.10.006 (2015).
Sun, C. J. et al. Transcatheter arterial embolization of acute gastrointestinal tumor hemorrhage with Onyx. Indian J. Cancer. 51(Suppl 2), e56–e59. https://doi.org/10.4103/0019-509x.151988 (2015).
Wang, B., Gao, B. L., Xu, G. P., Xiang, C. & Liu, X. S. Endovascular embolization is applicable for large and giant intracranial aneurysms: experience in one center with long-term angiographic follow-up. Acta Radiol. 56, 105–113. https://doi.org/10.1177/0284185113520312 (2015).
Poyet, C., Grubhofer, F., Zimmermann, M., Sulser, T. & Hermanns, T. Therapy-resistant nephrolithiasis following renal artery coil embolization. BMC Urol. 13, 29. https://doi.org/10.1186/1471-2490-13-29 (2013).
Saba, R. et al. Pulmonary arterial hypertension secondary to ethanol sclerotherapy for renal artery embolization. Case Rep. Crit. Care. 2014, 452452. https://doi.org/10.1155/2014/452452 (2014).
Bakar, B. et al. Evaluation of the toxicity of onyx compared with n-butyl 2-cyanoacrylate in the subarachnoid space of a rabbit model: an experimental research. Neuroradiology 52, 125–134. https://doi.org/10.1007/s00234-009-0594-8 (2010).
Mawad, D. et al. Synthesis and characterization of novel radiopaque poly(allyl amine) nanoparticles. Nanotechnology 21, 335603. https://doi.org/10.1088/0957-4484/21/33/335603 (2010).
Ma, Y. et al. The studies on highly concentrated complex dispersions of gold nanoparticles and temperature-sensitive nanogels and their application as new blood-vessel-embolic materials with high-resolution angiography. J. Mater. Chem. B. 2, 6044–6053. https://doi.org/10.1039/C4TB00748D (2014).
Zhao, H. et al. Temperature-sensitive poly(N-isopropylacrylamide-co-butyl methylacrylate) nanogel as an embolic agent: distribution, durability of vascular occlusion, and inflammatory reactions in the renal artery of rabbits. AJNR Am. J. Neuroradiol. 34, 169–176. https://doi.org/10.3174/ajnr.A3177 (2013).
Zhao, Y. et al. Permanent and peripheral embolization: temperature-sensitive p(N-isopropylacrylamide-co-butyl methylacrylate) nanogel as a novel blood-vessel-embolic material in the interventional therapy of liver tumors. Adv. Funct. Mater. 21, 2035–2042. https://doi.org/10.1002/adfm.201002510 (2011).
Li, L. et al. Rational design of temperature-sensitive blood-vessel-embolic nanogels for improving hypoxic tumor microenvironment after transcatheter arterial embolization. Theranostics 8, 6291–6306. https://doi.org/10.7150/thno.28845 (2018).
Grossberg, J. A., Howard, B. M. & Saindane, A. M. The use of contrast-enhanced, time-resolved magnetic resonance angiography in cerebrovascular pathology. Neurosurg. Focus FOC. 47, E3. https://doi.org/10.3171/2019.9.focus19627 (2019).
Qian, K. et al. The studies about doxorubicin-loaded p(N-isopropyl-acrylamide-co-butyl methylacrylate) temperature-sensitive nanogel dispersions on the application in TACE therapies for rabbit VX2 liver tumor. J. Control Release. 212, 41–49. https://doi.org/10.1016/j.jconrel.2015.06.013 (2015).
Kim, S. H., Kamaya, A. & Willmann, J. K. CT perfusion of the liver: principles and applications in oncology. Radiology 272, 322–344. https://doi.org/10.1148/radiol.14130091 (2014).
Sun, Z. et al. Predicting perigastric lymph node metastasis in gastric cancer with CT perfusion imaging: a prospective analysis. Eur. J. Radiol. 122, 108753. https://doi.org/10.1016/j.ejrad.2019.108753 (2020).
Cai, X. R. et al. Assessment of renal function in patients with unilateral ureteral obstruction using whole-organ perfusion imaging with 320-detector row computed tomography. PLoS ONE 10, e0122454. https://doi.org/10.1371/journal.pone.0122454 (2015).
Wan, J. et al. Doxorubicin-induced co-assembling nanomedicines with temperature-sensitive acidic polymer and their in-situ-forming hydrogels for intratumoral administration. J. Control. Release 235, 328–336. https://doi.org/10.1016/j.jconrel.2016.06.009 (2016).
Zhao, H., Xu, J., Huang, W., Zhao, Y. & Yang, X. Thermosensitive nanogels with cross-linked Pd(II) ions for improving therapeutic effects on platinum-resistant cancers via intratumoral formation of hydrogels. Chem. Mater. 31, 5089–5103. https://doi.org/10.1021/acs.chemmater.9b00986 (2019).
Xiong, W. et al. Dual temperature/pH-sensitive drug delivery of poly(N-isopropylacrylamide-co-acrylic acid) nanogels conjugated with doxorubicin for potential application in tumor hyperthermia therapy. Colloids Surf. B Biointerfaces. 84, 447–453. https://doi.org/10.1016/j.colsurfb.2011.01.040 (2011).
Zhao, H. et al. Spatiotemporally light-activatable platinum nanocomplexes for selective and cooperative cancer therapy. ACS Nano 13, 6647–6661. https://doi.org/10.1021/acsnano.9b00972 (2019).
Wan, J. et al. Precise synchronization of hyperthermia–chemotherapy: photothermally induced on-demand release from injectable hydrogels of gold nanocages. Nanoscale. 10, 20020–20032. https://doi.org/10.1039/C8NR06851H (2018).
Klouda, L. et al. Thermoresponsive, in situ cross-linkable hydrogels based on N-isopropylacrylamide: fabrication, characterization and mesenchymal stem cell encapsulation. Acta Biomater. 7, 1460–1467. https://doi.org/10.1016/j.actbio.2010.12.027 (2011).
Akimoto, J. et al. Facile cell sheet manipulation and transplantation by using in situ gelation method. J. Biomed. Mater. Res. B Appl. Biomater. 102, 1659–1668. https://doi.org/10.1002/jbm.b.33148 (2014).
Shimizu, K., Fujita, H. & Nagamori, E. Oxygen plasma-treated thermoresponsive polymer surfaces for cell sheet engineering. Biotechnol. Bioeng. 106, 303–310. https://doi.org/10.1002/bit.22677 (2010).
Chen, J. & Gailloud, P. Safety of spinal angiography: complication rate analysis in 302 diagnostic angiograms. Neurology. 77, 1235–1240. https://doi.org/10.1212/WNL.0b013e3182302068 (2011).
Willinsky, R. A. et al. Neurologic complications of cerebral angiography: prospective analysis of 2,899 procedures and review of the literature. Radiology 227, 522–528. https://doi.org/10.1148/radiol.2272012071 (2003).
Schaafsma, J. et al. Cost-effectiveness of magnetic resonance angiography versus intra-arterial digital subtraction angiography to follow-up patients with coiled intracranial aneurysms. Stroke 41, 1736–1742. https://doi.org/10.1161/strokeaha.110.585083 (2010).
Fraioli, F. et al. Whole-tumor perfusion CT in patients with advanced lung adenocarcinoma treated with conventional and antiangiogenetic chemotherapy: initial experience. Radiology 259, 574–582. https://doi.org/10.1148/radiol.11100600 (2011).
Sun, Z. et al. An application study of low-dose computed tomography perfusion imaging for evaluation of the efficacy of neoadjuvant chemotherapy for advanced gastric adenocarcinoma. Gastric Cancer Off. J. Int. Gastric Cancer Assoc. Jpn. Gastric Cancer Assoc. 21, 413–420. https://doi.org/10.1007/s10120-017-0763-0 (2018).
Campbell, B. C. V. & Khatri, P. Stroke. Lancet 396, 129–142. https://doi.org/10.1016/S0140-6736(20)31179-X (2020).
Campbell, B. et al. Comparison of computed tomography perfusion and magnetic resonance imaging perfusion-diffusion mismatch in ischemic stroke. Stroke 43, 2648–2653. https://doi.org/10.1161/strokeaha.112.660548 (2012).
Reiner, C. S. et al. Computed tomography perfusion imaging of renal cell carcinoma: systematic comparison with histopathological angiogenic and prognostic markers. Invest. Radiol. 48, 183–191. https://doi.org/10.1097/RLI.0b013e31827c63a3 (2013).
Du, F., Jiang, R., Gu, M., He, C. & Guan, J. The clinical application of 320-detector row CT in transcatheter arterial chemoembolization (TACE) for hepatocellular carcinoma. Radiol. Med. 120, 690–694. https://doi.org/10.1007/s11547-015-0523-3 (2015).
Acknowledgements
We would like to thank all the participants. We acknowledge financial support from the National Natural Science Foundation of China (Grant No. 81371661).
Author information
Authors and Affiliations
Contributions
All the authors have contributed significantly to the content of the manuscript. P.H. and Y.Z. designed the study. Z.Z. and C.C. performed the experiments and wrote the manuscript. K.Q., H.L., X.Z. and H.Z. performed the experiments. G.M., Y.C., N.J. and C.Z. analyzed the data.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Zhang, Z., Cen, C., Qian, K. et al. Assessment of the embolization effect of temperature-sensitive p(N-isopropylacrylamide-co-butyl methylacrylate) nanogels in the rabbit renal artery by CT perfusion and confirmed by macroscopic examination. Sci Rep 11, 4826 (2021). https://doi.org/10.1038/s41598-021-84372-w
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-84372-w
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.