Abstract
Microbeam radiation therapy is a novel pre-clinical external beam therapy that uses high-brilliance synchrotron X-rays to deliver the necessary high dose rates. The unique conditions of high dose rate and high spatial fractionation demand a new class of detector to experimentally measure important beam quality parameters. Here we demonstrate the highest spatial resolution plastic scintillator fibre-optic dosimeter found in the literature to date and tested it on the Imaging and Medical Beam-Line at the Australian Synchrotron in a X-ray beam where the irradiation dose rate was 4435 Gy/s. With a one-dimensional spatial resolution of 10 μm the detector is able to resolve the individual microbeams (53.7 ± 0.4 μm wide), and measure the peak-to-valley dose ratio to be 55 ± 17. We also investigate the role of radioluminescence in the optical fibre used to transport the scintillation photons, and conclude that it creates a significant contribution to the total light detected.
Similar content being viewed by others
Introduction
Synchrotron microbeam radiation therapy (MRT) is a novel external beam cancer therapy currently in the pre-clinical research stage. The highly collimated and high brilliance X-rays required for MRT can only be produced in a synchrotron. Despite the relatively small number of synchrotron facilities world-wide, there is emerging research into compact synchrotron technologies1,2,3 which can allow MRT to become a widespread treatment of currently untreatable cancers4,5. The current hypothesis for the high resistance to damage of the healthy tissue is that the vasculature is unaffected by microbeams ranging from 25 μm to 100 μm width and 200 μm to 400 μm spacing, while the tumour vasculature is destroyed6,7,8,9,10. Synchrotron x-rays are required for MRT due to their high collimation, to ensure the microbeam quality, and the high brilliance, to ensure the dose can be delivered quickly enough to minimise dose-blurring from patient movement. Independent detector technologies must be developed to provide essential quality assurance of the X-ray beam and treatment plan. An overview of devices capable of MRT dosimetry has been presented previously11, hence here we will only present a summary high spatial resolution real time dosimeters that are capable of MRT dosimetry.
Radiochromic film is considered a dosimetric standard, but is limited in its application to MRT. The high range of doses between the peaks and valleys means that two exposures of different duration on separate film are required to measure the peak and valley doses. Further, it has been shown that film measurements can be inaccurate to up to 15%12. Metal oxide semiconductor field-effect transistors (MOSFET) have the required spatial resolution for MRT dosimetry, but lack the radiation hardness to withstand the high dose rate13. A number of silicon strip detectors (SSD) have been developed for MRT beam quality monitoring14,15,16. These detectors have reached an estimated resolution of 15 μm defined by the electrical field distribution under biasing.
A commercial single crystal diamond detector, the PTW microDiamond, as also been applied to MRT fields due to its very high spatial resolution. It claims a 1 μm one-dimensional sensitive volume (with 2.2 mm diameter in the other dimensions)17.
Scintillator dosimeters have been applied to a variety of contexts. LINAC photon dosimetry has been done with a variety of detectors, including plastic plastic scintillators (which are advantageous due to their water equivalence18,19,20) and inorganic dosimeters (which are advantageous due to their superior light output and radiation resistance, as well as characteristics that allow easy stem-effect filtering21,22,23,24). A preliminary study of Europium and Lithium doped yttrium oxide crystal scintillator has been applied to x-ray micro-fields, with limitations due to the non-water equivalence of the crystalline material25. In this work we will be focusing on plastic scintillator fibre-optic dosimeters (FODs). Plastic scintillator FODs are desirable due to their water equivalence, radiation hardness and energy independence18,26. Specifically, the radiation hardness will be an important factor for MRT dosimetry. Beddar et al. found that their plastic scintillator FOD saw a 2.8% decrease in sensitivity after exposure to 10 kGy of Cs-137 gamma rays18. This level of radiation hardness is appropriate for exposures to MRT x-rays with minimal detector sensitivity deviations. While an inorganic solution (such as a crystal detector21 or doped-silica detector24) would provide a smaller sensitivity to radiation damage, the primary focus of this work is to develop a dosimeter that is as water equivalent as possible for MRT. Typically, scintillator dosimeters are applied when high spatial resolution is not required, because plastic scintillators have a low light yield (typically 11,000 photons/MeV27) and so the dosimeter sensitivity is limited by the small scintillator volumes required for high resolution dosimetry. However, the high brilliance of a synchrotron source provides a measurable response in scintillator dosimeters. Optical detectors have been applied to imaging microbeam X-rays in the past25,28, however these detectors have not been demonstrated and tested at highly brilliant synchrotron light source facilities.
One of the challenges with optical dosimetry is the so-called stem effect: light generated in the optical fibre that is not related to the dose deposited in the scintillator sensitive volume18. There are two sources of this stem light: Cherenkov radiation and radioluminescence. Due to the refractive index of the optical fibre core, there is no Cherenkov radiation generated under synchrotron X-ray irradiation26, instead only radioluminescence is present. It dominates at lower wavelengths29, and occurs less in PMMA-core optical fibres than silica-core optical fibres30.
We have previously tested FODs with MRT beams that have a spatial resolution of 50 μm31 and 20 μm11. In this work we demonstrate a FOD with a spatial resolution of 10 μm, the highest of a plastic scintillator detector found in the literature.
Materials and Methods
The design of the scintillator fibre-optic dosimeter is summarised in Fig. 1. A 10 ± 2 μm thick film of BC-400 plastic scintillator is optically coupled to a 1 mm diameter core Eska CK-40 optical fibre. The scintillator thickness defines the one-dimensional spatial resolution of the FOD probe as 10 μm in the axial direction, and 1 mm in the radial direction (determined by the optical fibre core diameter). The scintillation light (peaking at 423 nm) is transported via optical fibre to a SensL MiniSM Silicon Photomultiplier 10035 (SiPM)32. The scintillator volume is 0.0380 mm3, while the collection volume (encompassed by the optical fibre acceptance cone) is 0.00785 mm3 Table 1.
The scintillation and SiPM collection efficiency spectra can be found in Archer et al.31. As the scintillator is not covered, experiments must be done in darkness to minimise the amount of ambient light entering the optical fibre. Measuring the background signal allows a simple subtraction to be done, but excess light will limit the dynamic range of the detector.
The measurements were performed on the Imaging and Medical Beam-Line (IMBL) at the Australian Synchrotron. The 3.032 GeV, 200.2 mA electron beam is subject to a 3.0 T wiggler magnetic field to produce synchrotron X-ray radiation (Fig. 2). These X-rays are filtered to control the spectrum and intensity, with filter combination F4 (as defined in Table 3, Stevenson et al.33). The spectrum is shown in Fig. 3. The beam was shaped with a beam defining aperture (BDA) to a width of 30 mm and a height of 2.014 mm, 1.052 mm or 0.532 mm, then was further shaped with a conformal mask to 20 mm × 20 mm. A tungsten multi-slit collimator (MSC) with gaps of 50 μm and pitch 400 μm is used to spatially fractionate the beam into microbeams. The dose rates of the various fields without the MSC in place are presented in Table 2.
The FOD was mounted in edge-on mode (fibre axis perpendicular to the beam direction and parallel to the direction of fractionation) in a specially constructed PMMA holder, allowing the microbeam fractionation to be measured with a spatial resolution of 10 μm, the thickness of the scintillator. By scanning the probe continuously across the beam at 1 mm/s, and sampling the SiPM response with an analogue front end at a high frequency (2000 Hz) a profile of the microbeams can be measured. The PMMA holder minimised alignment issues experienced with thin detectors such as this; the cross section exposed to the beam has thickness t given by:
where t0 is the one-dimensional spatial resolution (10 μm), d is the diameter of the optical fibre core (1000 μm) and θ is the misalignment angle in radians. For small angles, the increase in effective thickness is proportional to the core diameter, and hence is sensitive to misalignment. The PMMA holder will ensure that the detector is aligned within 1° in both rotational axis (about the x and z axes defined in Fig. 2). The z axis alignment can be further refined using the DynMRT rotation stage (by minimising the measured microbeam width), but there is currently no procedure possible with the current setup to align about the x axis.
Results and Discussion
Microbeam scans
Figure 4 shows the intrinsic microbeam array scan (with 10-point moving average smoothing applied – this moving average covered a 5 μm width). The inset shows the average and 95% confidence interval of all the (un-smoothed) microbeams. Figure 5 also shows the peak and valley values across the microbeam array. The peak height is very accurate and so no uncertainties are presented, while the valley values were averaged over a 200 μm region between the peaks, and the standard deviation is shown. The full-width at half-maximum (FWHM) was calculated to be 53.7 ± 0.4 μm. The peak-to-valley dose ratio (PVDR) over the entire microbeam array is 55 ± 17. Over the central 15 microbeams, the PVDR is 18.2 ± 1.5. This scan was repeated from 6 mm depth in water to 70 mm depth, which can be seen in Fig. 6. The step size is 2 mm up to 20 mm depth, and 5 mm for all deeper depths. For comparison, a microbeam scan was also done with a PTW microDiamond dosimeter, which has a spatial resolution of 1 μm17. The PMMA holders used for both detectors were identical, except for the opening to hold the detectors due to their differing geometry.
The width of the microbeams is expected to be 50 μm, which agrees well with the measured average FWHM of 53.7 ± 0.4 μm. While this is greater than the intrinsic microbeam width, it is expected due to the dose blurring over the 10 μm sensitive volume of the scintillator. This also validates the alignment of the detector to the incident X-rays. Measurements with the microDiamond detector gave a FWHM of 52.5 ± 5 μm (which can be seen in Fig. 4). This agrees within uncertainty with the FOD. The large uncertainty here is due to the detector being stepped by 5 μm between measurements. Interpolation between dose measurements allowed a more refined FWHM to be acquired. The microDiamond profile matches well with the FOD profile, with the exception of between −50 and −25 μm positions, where the FOD response is slightly higher than the microDiamond. We believe this is due to the higher spatial resolution of the microDiamond. The agreement between the shape of both profiles suggests that the FOD is measuring an accurate beam profile with minimal dose-blurring. Alternatively, it is possible that the microDiamond effective sensitive volume is larger than the theoretical case, and so both detectors may be experiencing similar levels of dose-blurring. This may be due to their being an effect from the packaging around the diamond sensitive volume to keV x-rays. The microDiamond has a 2.2 mm diameter of sensitive volume, and so is more sensitive to misalignment then the FOD.
To evaluate the PVDR, the central microbeams are considered for consistency. There is a roll-off of dose in the valleys towards the edge of the microbeam array, giving a large difference in PVDR across the profile. This can be seen in both the peaks and valleys in Fig. 5. The asymmetry in this figure indicates a slight rotational misalignment in the MSC. The average PVDR of all microbeams at 6 mm depth is 55 ± 17. However, over the central 15 microbeams, where the valleys are much more consistent, the PVDR is 18.2 ± 1.5. This agrees well with the central microbeam being measured with the microDiamond giving 17.1 ± 0.8.
Depth dose scans
With the FOD mounted in edge-on mode, the probe was scanned vertically in the z direction (defined in Fig. 2) through the HDR broadbeam (with no MSC fractionating the field) to measure how the broadbeam dose changes with depth (depth dose). This was repeated at smaller field heights of 1.052 mm and 0.532 mm. This methodology is consistent with other dosimetry devices (see Fournier et al. 3.134). The SiPM charge was measured with a commercial electrometer, PTW UNIDOSwebline, used for dosimetry. The charge was integrated over the duration of the scan, allowing relative dose to be measured. The dark current was too high to allow the electrometer to zero the readings (due to signal saturation on the highest sensitivity) so a “dark scan” with no X-rays allowed this to be characterised and subtracted in analysis. Figure 7 shows the depth dose measured with the FOD, along with the response from just the optical fibre. The response in the optical fibre is due to radioluminescence in the fibre.
It can be seen in Fig. 7(b) that the response with only optical fibre in the field has a higher relative response at low depths than the scintillator and fibre together, Fig. 7(a). This over-response is around 40% higher than the FOD signal at the same depth. Further, this effect is more significant with larger field heights. It is not due to high dose rate due to the effect also being seen at much lower dose rates11,31. The reduced discrepancy between the FOD signal and ionisation chamber, compared to the fibre-only signal, suggests that the discrepancy is minimal in the scintillator itself. As the optical fibre core is PMMA, the relative generation of radioluminescence is low, compared to silica core optical fibres30. However, it is still a significant part of the total light signal collected. We estimate this to be around 52% of the signal with the scintillator in the centre of the field, however this value is dependent on the exact fraction of fibre inside the field for this measurement. To adequately deal with this effect in future measurements, and any quality assurance methods, a secondary probe in parallel with the FOD measuring only the radioluminescence in the optical fibre can be used18,19,26. Other methods such as filtration or spectral separation are not efficient due to the strong overlap between the radioluminescence and scintillation spectra.
One of the primary challenges with using this dosimeter is the low light signal measured at the SiPM, which limits the applicability to high dose rates. We estimate that, using the ionisation chamber results at 75 mm depth and the low FOD response at this depth, that the minimum dose rate that can be confidently measured is 200 Gy/s. One method for increasing the light signal is to use an inorganic scintillator, which typically have a much higher yield. However, this will increase the dose perturbations due to the higher atomic mass elements used. Increasing the optical fibre core diameter will also increase the collected light, but will make the detector more sensitive to misalignment.
Conclusion
In this work we have presented the highest spatial resolution plastic scintillator fibre-optic dosimeter found in the literature with a collection volume of 0.00785 mm3 and a one-dimensional spatial resolution of 10 μm. Synchrotron X-ray microbeams have been resolved with this detector and measured the microbeam FWHM to be 53.7 ± 0.4 μm. The detector is limited by radioluminescence in the optical fibre, with solutions to this being investigated. This detector has the potential to be applied to characterising highly brilliant synchrotron X-rays and quality assurance in microbeam radiation therapy.
References
Hadsell, M. et al. A first generation compact microbeam radiation therapy system based on carbon nanotube X-ray technology. Applied Physics Letters 103, 183505 (2013).
Eggl, E. et al. The Munich Compact Light Source: initial performance measures. Journal of Synchrotron Radiation 23, 1137–1142 (2016).
Chen, M. et al. Tunable synchrotron-like radiation from centimeter scale plasma channels. Light-Science & Applications 5, e16015 (2016).
Bravin, A., Olko, P., Schueltke, E. & Wilkens, J. J. SYRA3 COST Action - Microbeam radiation therapy: Roots and prospects. Physica Medica 31, 561–563 (2015).
Grotzer, M. A., Schueltke, E., Braeuer-Krisch, E. & Laissue, J. A. Microbeam radiation therapy: Clinical perspectives. Physica Medica 31, 564–567 (2015).
Bouchet, A. et al. Preferential effect of synchrotron microbeam radiation therapy on intracerebral 9l gliosarcoma vascular networks. International Journal of Radiation Oncology Biology Physics 78, 1503–1512 (2010).
Bouchet, A. et al. Synchrotron X-Ray Boost in the Microbeam Radiation Therapy Mode Improves Glioma Control After Conventional X-Ray Fractions. International Journal of Radiation Oncology Biology Physics 96, E94–E95 (2016).
Brauer-Krisch, E. et al. Effects of pulsed, spatially fractionated, microscopic synchrotron x-ray beams on normal and tumoral brain tissue. Mutation Research-Reviews in Mutation Research 704, 160–166 (2010).
Bronnimann, D. et al. Synchrotron microbeam irradiation induces neutrophil infiltration, thrombocyte attachment and selective vascular damage in vivo. Scientific Reports 6, 33601 (2016).
Crosbie, J. C. et al. Tumor cell response to synchrotron microbeam radiation therapy differs markedly from cells in normal tissues. International Journal Of Radiation Oncology Biology Physics 77, 886–894 (2010).
Archer, J. et al. Synchrotron x-ray microbeam dosimetry with a 20 micrometer resolution scintillator fibre-optic dosimeter. Journal of Synchrotron Radiation 25, 826–832 (2018).
Bartzsch, S., Lott, J., Welsch, K., Brauer-Krisch, E. & Oelfke, U. Micrometer-resolved film dosimetry using a microscope in microbeam radiation therapy. Medical Physics 42, 4069–4079 (2015).
Rosenfeld, A. et al. MOSFET dosimetry of an X-ray microbeam. IEEE Transactions on Nuclear Science 46, 1774–1780 (1999).
Petasecca, M. et al. X-tream: a novel dosimetry system for synchrotron microbeam radiation therapy. Journal of Instrumentation 7, P07022 (2012).
Fournier, P. et al. X-tream quality assurance in synchrotron x-ray microbeam radiation therapy. Journal of Synchrotron Radiation 23, 1180–1190 (2016).
Davis, J. A. et al. Characterisation and evaluation of a pnp strip detector for synchrotron microbeam radiation therapy. Biomedical Physics & Engineering Express 4, 044002 (2018).
Livingstone, J., Stevenson, A. W., Butler, D. J., Hausermann, D. & Adam, J.-F. Characterization of a synthetic single crystal diamond detector for dosimetry in spatially fractionated synchrotron x-ray fields. Medical Physics 43, 4283–4293 (2016).
Beddar, A. S., Mackie, T. R. & Attix, F. H. Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: 1. Physical characteristics and theoretical considerations. Physics in Medicine & Biology 37, 1883–1900 (1992).
Archer, J. et al. High-resolution fiber-optic dosimeters for microbeam radiation therapy. Medical Physics 44, 1965–1968 (2017).
Beaulieu, L. & Beddar, S. Review of plastic and liquid scintillation dosimetry for photon, electron, and proton therapy. Physics in Medicine & Biology 61, R305 (2016).
Justus, B. L. et al. Gated fiber-optic-coupled detector for in vivo real-time radiation dosimetry. Appl. Opt. 43, 1663–1668 (2004).
Pittet, P. et al. Fiber background rejection and crystal over-response compensation for gan based in vivo dosimetry. Physica Medica: European Journal of Medical Physics 29, 487–492 (2013).
Veronese, I. et al. Infrared luminescence for real time ionizing radiation detection. Applied Physics Letters 105, 061103 (2014).
Veronese, I. et al. Real-time dosimetry with Yb-doped silica optical fibres. Physics in Medicine & Biology 62, 4218 (2017).
Belley, M. D. et al. Fiber-optic detector for real time dosimetry of a micro-planar x-ray beam. Medical Physics 42, 1966–1972 (2015).
Beddar, A. S., Mackie, T. R. & Attix, F. H. Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: 2. Properties and measurements. Physics in Medicine & Biology 37, 1901–1913 (1992).
Saint-Gobain Crystals. BC-400, BC-404, BC-408, BC-412, BC-416 Premium Plastic Scintillators, https://www.crystals.saint-gobain.com/sites/imdf.crystals.com/files/documents/bc400-404-408-412-416-data-sheet.pdf (visited on 12/05/2018).
Okada, G. et al. Spatially resolved measurement of high doses in microbeam radiation therapy using samarium doped fluorophosphate glasses. Applied Physics Letters 99, 121105 (2011).
Therriault-Proulx, F., Beaulieu, L., Archambault, L. & Beddar, S. On the nature of the light produced within pmma optical light guides in scintillation fiber-optic dosimetry. Physics in Medicine & Biology 58, 2073–2084 (2013).
Nowotny, R. Radioluminescence of some optical fibres. Physics in Medicine & Biology 52, N67 (2007).
Archer, J. et al. X-ray microbeam measurements with a high-resolution scintillator fibre-optic dosimeter. Scientific Reports 7, 12450 (2017).
sensL. Minism photomultiplier modules, http://www.sensl.com/downloads/ds/DS-MiniSM.pdf.
Stevenson, A. W. et al. Quantitative characterization of the X-ray beam at the Australian Synchrotron Imaging and Medical Beamline (IMBL). Journal of Synchrotron Radiation 24, 110–141 (2017).
Fournier, P. et al. X-tream dosimetry of highly brilliant x-ray microbeams in the MRT hutch of the australian synchrotron. Radiation Measurements 106, 405–411 (2017).
Acknowledgements
This research was undertaken on the Imaging and Medical Beam Line at the Australian Synchrotron, part of ANSTO (AS181/IMBL/12956, AS183/IMBL/13671). Authors JA, JD, and ML acknowledge the support of the Australian NH&MRC (APP1093256). This research has been conducted with the support of the Australian Government Research Training Program Scholarship.
Author information
Authors and Affiliations
Contributions
J.A. wrote the main manuscript text and prepared figures. J.A., J.D. and M.L. collected the data. M.C. produced simulation data. E.L. designed and fabricated the fiber-optic probes and constructed the signal detection system. E.L., A.R. and M.L. jointly conceptualised and supervised the work. All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Archer, J., Li, E., Davis, J. et al. High spatial resolution scintillator dosimetry of synchrotron microbeams. Sci Rep 9, 6873 (2019). https://doi.org/10.1038/s41598-019-43349-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-019-43349-6
This article is cited by
-
One-dimentional periodic structure infiltrated by (PVA/CV + CF)-polymer for high-performance sensitivity
Optical and Quantum Electronics (2022)
-
Megavolt bremsstrahlung measurements from linear induction accelerators demonstrate possible use as a FLASH radiotherapy source to reduce acute toxicity
Scientific Reports (2021)
-
Highly sensitive photonic crystal gamma ray dosimeter
Optical and Quantum Electronics (2021)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.