Abstract
We present a rapid (<10 s), cost-effective, unique single-step method for fabricating paper-based devices without necessitating any expensive instrumentation, simply by deploying correction pens that are otherwise commonly used for masking typos in printed or written matters. The marked regions formed by deposits from the correction pen demonstrate ubiquitous flow resistances to typical aqueous solutions and organic solvents in the transverse direction, resulting in a preferential bulk flow along the axial direction of the paper channels ‘fabricated’ in the process. Considering the simplicity and cost-effectiveness of this platform, it is deemed to be ideal for (bio) chemical sensing and point-of-care diagnostics in resource-limited settings.
Similar content being viewed by others
Introduction
Paper, as a versatile substrate, has garnered attention in the miniaturization and microfluidics community due to its low cost, passive fluid transport via capillary action, porosity, biodegradability, biocompatibility and random network structure. The pioneering efforts of Whiteside’s group1,2,3 have transformed a simple paper into paper-based analytical devices in association with fluorescence, colorimetric, and electrochemical detection systems4,5. Afterwards, many other interesting phenomena like electrically modulated flow control6, micromixing7,8, separation9, energy generation10,11,12 etc. have been explored and studied using paper-based device. The contribution of this device is significant in various fields including clinical, food industries, smart phone based sensing, environmental applications and point-of-care tests13,14,15. Of late, scientists and researchers have found a deep interest in studying the fundamentals of fluid transport through the random porous structure of the paper matrix16,17,18.
The hydrophilic nature of the paper pores, coupled with patterning of hydrophobic barriers across the direction of preferential transport, is a quintessential consideration in constructing paper-based analytical devices. Plethora of expensive techniques has been used for fabricating hydrophobic barriers on paper, including photolithography1,19,20, flexography printing21, plasma treatment22,23,24, cutting25,26 and vapour phase deposition27,28,29. Various low-cost techniques like PDMS printing30, wax printing31,32, ink-jet printing33,34,35,36, wax dipping37,38, screen printing39,40 and stamping41,42 have also been developed. The common drawbacks with the aforementioned techniques are high cost, long fabrication time, expertise, requirement of external armamentarium like laser, oven, printer and stamps. Access to such instruments still remains a challenge in resource-limited laboratories, and consequently, this may end up in delayed diagnosis or detection of ailments using body-fluid based procedures43,44. One way to overcome this problem is by free-hand drawing, where a novice can fabricate paper-based devices45,46.
In this work, we describe a rapid, frugal and facile proof-of-concept of fabricating paper-based analytical devices using correction pen. Traditionally, correction pen is used as an agent to mask errors in printed or written text. Exact composition of correction pen ink is a trade secret. However, literature suggests that it is a mixture of titanium dioxide, solvents, resins and colorant47. To the best of our knowledge, this combination has never been leveraged as a barrier patterning agent on paper. Our one-step fabrication involves direct manual deposition (i.e, free-hand drawing) of the correction fluid on filter paper. Strikingly, this method requires no heating, no complex instrumentation, and no trained personnel.
Results and Discussions
Fabrication of Correction pen-based barrier on paper
First, we have constructed circular devices through direct patterning of correction liquid from the pen on filter paper. After solvent evaporation (~15 min), we observed that Titanium dioxide gets embedded on paper and blocks the perforated pores, consequently forming a visible hydrophobic barrier (Fig. 1a). This has been confirmed by wetting the device with water (Fig. 1b). When coloured water was introduced to the fabricated device, the hydrophobic barriers have confined the same without any disruption or leakage (Fig. 1c,d). Through free hand-drawing, we have designed our respective institute’s acronyms and further checked their water confinement ability (Fig. 2). We have also designed a channel and assessed the wicking ability of paper device (Supplementary Fig. 1). This method offers advantages like less fabrication time (10 s for circular device), one-time deposition of correction fluid, 1 mm resolution and no dependence of heating step.
Barrier’s chemical compatibility
A wide gamut of research has already been reported on chemical compatibility of microfluidic devices to solvents and other reagents. Lee et al. have demonstrated the barrier compatibility of PDMS-based microfluidic devices with a comprehensive list of organic solvents48. Later, Dornelas et al. have proposed a simple rubber stamp based contact printing of PDMS on paper49. The authors have examined the barrier compatibility of PDMS and Wax with solvents and surfactants. Wang et al. used a different approach, where they have used sol-gel method and inkjet-printing to fabricate barriers on paper50. Furthermore, they compared chemical compatibility of Methylsilsesquioxane (MSQ) based hydrophobic barriers with wax and alkyl ketene dimer and concluded that MSQ barriers were not disrupted by surfactants. In order to use our method for bio-sensing or point-of-care testing, it is of paramount importance to study the fabricated barrier’s resistance or inertness to different reagents. For this, we used 10 devices per reagent and monitored the barrier disruption. Interestingly, the results revealed that 70% or higher devices have successfully confined the reagents without any disruption. Device’s compatibility with different chemicals is depicted in Fig. 3 and Supplementary Table 1. Through barrier compatibility test, we infer that our fabrication method holds great promise in confining most of the commonly used aqueous solutions and organic solvents.
Surface Characterization
Figure 4 shows the SEM micrographs of coated and uncoated part of the Whatman(R) filter paper (Grade 1). The difference in the surface morphology corroborates to pore blockage by particles present in correction pen. To analyse the elemental composition of the coated and uncoated surface, we performed Energy Dispersive Spectroscopy and found that coated surface possess abundance of Titanium (Supplementary Fig. 2 and Table 2). To strengthen our statement, we have also carried out AFM studies on the barriers and plain paper surface (Supplementary Fig. 3). We have found that Root Mean Square height value of coated and uncoated is 0.145 µm and 0.758 µm respectively (data not shown). This result clearly attributes to the difference between the morphology of two surfaces. Overall, it is clear from the results that the components present in correction pen block the pores and mask the paper surface, which results in sample confinement.
We have further studied the wetting property of coated and uncoated filter paper by water contact angle measurements. Figure 5 clearly indicates the difference in the surface wettability. The results reveal that the initial contact angle of coated and uncoated paper is 75° and 22° respectively. High contact angle of coated paper attributes to the hydrophobic nature of the surface (due to deposition of TiO2 particles), whereas the less contact angle of uncoated paper attributes to the hydrophilic nature of the surface. Time lapse images of contact angle measurements of both the surfaces are provided (Supplementary Fig. 4). Songok et al. have used liquid flame spray technique and coated titanium dioxide nanoparticles on paper. The authors reported that the coated TiO2 nanoparticles impart hydrophobicity on the paper surface. Moreover, the authors have observed variations in the wettability of paper surface by measuring the contact angle. They reported that uncoated paper surface is hydrophilic (CA~80°), whereas the TiO2 coated paper surface is super hydrophobic (CA~150°)51. The above finding is in good agreement with our proposed approach and supports the utility of simple titanium dioxide based correction pen for fabricating hydrophobic barriers.
We also have characterized the surface by studying the water penetration rates of correction pen based hydrophilic channel and the plain filter paper. Interestingly, the water penetration profile in both fabricated device and the plain paper follow similar trends (Fig. 6). In our opinion, correction pen or Titanium-based composites, inarguably, can be added to the list of low-cost materials like PDMS and Wax.
Biochemical assays
To validate the functionality of the correction pen-based devices, we have performed basic colorimetric assays. Owing to the simplicity, we have chosen glucose and protein estimation using glucose oxidase enzymatic mixture and Bradford reagent respectively. We have studied the change in colour intensity by varying the glucose and protein concentration. For this, we have selected a wide range of concentration (0–300 mg/dl) and (0–2000 mg/dl) for standard glucose and protein respectively. Figure 7 suggests variation in colour intensity is proportional to the concentration of the analyte.
Moreover, upon addition of blood (20 µl) to the circular device, we observed no cross reactivity of blood sample with the barrier (Fig. 8a). This implies that the barrier’s composition limits surface interaction, confirming its suitability for any enzymatic or biochemical assays. Shelf-life of paper-based devices is always a crucial parameter for long-term storage in resource-limited settings. We observed no leakage or barrier disruption from the stored device (1 month) device (Fig. 8b). The main limitation of our fabrication method is the inability to confine alcohols and surfactants similar to wax barriers; however, this issue can be addressed49. Rajendra et al. have described wax barrier disruption due to surfactants and their importance in biological assays52. The authors developed silicone resins to circumvent such barrier disruption.
Further, we have explored the applicability of the platform towards developing a quantitative assay for a very important blood parameter, namely, the blood glucose level. For this, we used blood samples of known glucose concentration (estimated through EM 360 automated clinical analyser). Later, we assessed the glucose levels in blood plasma and their relative colour intensities. 12 blood samples of varying glucose concentrations were collected. Each sample was deposited in 3 different devices and the average colour intensities ± standard deviations were measured. As it can be witnessed from Fig. 9, a linear increase in the grey scale value attributes to increase in the glucose concentration. All the reported data points agreed within 10% of the calibration curve (essentially, the straight line). During this study, no barrier leakage or disruption was observed. This suggests that our method can be easily employed for colorimetric detection of blood-based parameters through point-of-care testing.
Conclusions
To summarize, in this study, we have described a quick, reliable and cost-effective one-step method to fabricate paper-based devices. We have used simple free-hand drawing to deposit Titanium dioxide based correction fluid and have blocked the pores of filter paper. The deposit exhibits unique resistance to different aqueous solutions and solvents. We have confirmed the workability or functionality of the device through standard glucose, protein and plasma glucose assays. The fabrication cost of one device is 0.0037 US $ and time is <10 s respectively. We are aware that our method has two limitations. The first one is minor toxicity. The second is inability to confine surfactants as well as alcohols. It is well known that the role of surfactants is restricted only to cell lysis and alcohols have negligible role in medical diagnostics. Henceforth, this method can safely be used as a simple substitute for paper-based analytical devices that are fabricated by other techniques, for common purposes. In addition, this may pave a way for developing frugal Titanium-based composites or inks for constructing paper-based device for bio-sensing applications compatible with the requirements of the economically deprived community at large.
Methods and Materials
Materials
Whatman(R) cellulose filter paper (Grade 1) was purchased from GE Life sciences, India. Correction pens with tip size of 0.8 mm (Camlin, Faber Castle & Cello brands) and Fountain pen ink were obtained from local stationery. Glucose assay kit was supplied by Arkray Health Care, India. Protein Estimation kit was supplied by Himedia Laboratories, India. Triton-X was purchased from Loba Chemie. Ethanol, Isopropanol, Acetone, Acetonitrile, Dimethyl formamide, Sodium dodecyl sulphate were purchased from Merck, India. All other reagents and chemicals of analytical grade were purchased from Sigma-Aldrich, India.
Fabrication of paper-based analytical devices
Correction pen was gently pressed in order to dispense fluid onto the filter paper (deposition on both the sides of the paper is advisable). The fabricated device was cured for 30 min at room temperature (25 °C). One side of the filter paper was sealed using 3 M adhesive tape. The thickness of plain and coated filter paper was measured using Mitutoyo Digital Micrometer (293–831).
Image acquisition
Device images were captured using Nikon D5200 camera at a fixed distance. After acquisition, images were processed using FIJI and MS PowerPoint. Greyscale measurement for plasma glucose was performed using FIJI. Firstly, the images were converted to 8-bit grey scale. Secondly, the images were inverted to have a positive slope. Finally, mean grey value was measured.
Surface characterization
SEM analysis was performed using Zeiss EVO 18, smart SEM software. Energy Dispersive Spectroscopy for Elemental analysis was carried out using EDAX (Ametek), TEAM software. Surface morphology of coated and uncoated filter paper was performed using an Agilent Technologies, AFM 5100 silicon nitride cantilever (PPP-NCL, Nanosensors Inc., USA) in tapping mode. Water penetration rate was assessed using a ruler, a camera and a timer. Water contact angle was recorded using Phantom V641 high-speed camera.
Ethical approval and blood sample collection
An approval of ethical clearance (Approval No: IIT/SRIC/DR/2017 Dated: April 27, 2017) was obtained from Institute Ethical Committee (IEC) for experimenting with blood sample. The diabetic and normal blood samples were collected from B C Roy Technology Hospital, Indian Institute of Technology Kharagpur. An informed consent was received from all participants. Only one sample was collected from each participant after receiving the approval of informed consent from the patient or Guardians. All the experiments using human blood samples were performed according to the guidelines and regulations of Institute Ethical Committee (Indian Institute of Technology Kharagpur, India), Department of Biotechnology (Ministry of Science and Technology, Govt. of India) and Indian Council of Medical Research, India.
Biochemical assays
Glucose assay was performed in this order. 10 µl of Glucose Oxidase enzyme mix was added to the fabricated device followed by the addition of 10 µl standard glucose of varying concentrations. For protein assay, 15 µl of Bradford reagent was added to the device. Later, 5 µl Bovine Serum Albumin of different concentrations were added. Conventional blood glucose level was measured using Erba Mannheim 360 Automated clinical chemistry analyser. Plasma was obtained by centrifuging at 3500 rpm for 5 min. In paper devices, 10 µl of Glucose Oxidase enzyme mix was added initially, followed by 10 µl of Plasma sample.
References
Martinez, A. W., Phillips, S. T., Butte, M. J. & Whitesides, G. M. Patterned Paper as a Platform for Inexpensive, Low-Volume, Portable Bioassays. Angew. Chemie Int. Ed. 46, 1318–1320 (2007).
Martinez, A. W. et al. Simple Telemedicine for Developing Regions: Camera Phones and Paper-Based Microfluidic Devices for Real-Time, Off-Site Diagnosis. Anal. Chem. 80, 3699–3707 (2008).
Martinez, A. W., Phillips, S. T., Whitesides, G. M. & Carrilho, E. Diagnostics for the Developing World: Microfluidic Paper-Based Analytical Devices. Anal. Chem. 82, 3–10 (2010).
Xia, Y., Si, J. & Li, Z. Fabrication techniques for microfluidic paper-based analytical devices and their applications for biological testing: A review. Biosens. Bioelectron. 77, 774–789 (2016).
Akyazi, T., Basabe-Desmonts, L. & Benito-Lopez, F. Review on microfluidic paper-based analytical devices towards commercialisation. Anal. Chim. Acta 1001, 1–17 (2018).
Mandal, P., Dey, R. & Chakraborty, S. Electrokinetics with “paper-and-pencil” devices. Lab Chip 12, 4026 (2012).
Dey, R., Kar, S., Joshi, S., Maiti, T. K. & Chakraborty, S. Ultra-low-cost ‘paper-and-pencil’ device for electrically controlled micromixing of analytes. Microfluid. Nanofluidics 19, 375–383 (2015).
Arun, R. K. et al. Paper-PDMS hybrid microchannel: a platform for rapid fluid-transport and mixing. J. Micromechanics Microengineering 26, 105008 (2016).
Kar, S., Maiti, T. K. & Chakraborty, S. Capillarity-driven blood plasma separation on paper-based devices. Analyst 140, 6473–6476 (2015).
Arun, R. K., Halder, S., Chanda, N. & Chakraborty, S. A paper based self-pumping and self-breathing fuel cell using pencil stroked graphite electrodes. Lab Chip 14, 1661–1664 (2014).
Arun, R. K., Singh, P., Biswas, G., Chanda, N. & Chakraborty, S. Energy generation from water flow over a reduced graphene oxide surface in a paper–pencil device. Lab Chip 16, 3589–3596 (2016).
Das, S. S., Kar, S., Anwar, T., Saha, P. & Chakraborty, S. Hydroelectric power plant on a paper strip. Lab Chip 18, 1560–1568 (2018).
Cate, D. M., Adkins, J. A., Mettakoonpitak, J. & Henry, C. S. Recent developments in paper-based microfluidic devices. Anal. Chem. 87, 19–41 (2015).
Altundemir, S., Uguz, A. K. & Ulgen, K. A review on wax printed microfluidic paper-based devices for international health. Biomicrofluidics 11, 041501 (2017).
Sher, M., Zhuang, R., Demirci, U. & Asghar, W. Paper-based analytical devices for clinical diagnosis: recent advances in the fabrication techniques and sensing mechanisms. Expert Rev. Mol. Diagn. 17, 351–366 (2017).
de Azevedo, E. N., Alme, L. R., Engelsberg, M., Fossum, J. O. & Dommersnes, P. Fluid imbibition in paper fibers: Precursor front. Phys. Rev. E 78, 066317 (2008).
Masoodi, R. & Pillai, K. M. Darcy’s law-based model for wicking in paper-like swelling porous media. AIChE J. 56, 2257–2267 (2010).
Chaudhury, K., Kar, S. & Chakraborty, S. Diffusive dynamics on paper matrix. Appl. Phys. Lett. 109, 224101 (2016).
OuYang, L., Wang, C., Du, F., Zheng, T. & Liang, H. Electrochromatographic separations of multi-component metal complexes on a microfluidic paper-based device with a simplified photolithography. RSC Adv. 4, 1093–1101 (2014).
He, Q., Ma, C., Hu, X. & Chen, H. Method for Fabrication of Paper-Based Microfluidic Devices by Alkylsilane Self-Assembling and UV/O3 -Patterning. Anal. Chem. 85, 1327–1331 (2013).
Olkkonen, J., Lehtinen, K. & Erho, T. Flexographically Printed Fluidic Structures in Paper. Anal. Chem. 82, 10246–10250 (2010).
Li, X., Tian, J., Nguyen, T. & Shen, W. Paper-Based Microfluidic Devices by Plasma Treatment. Anal. Chem. 80, 9131–9134 (2008).
Li, X., Tian, J. & Shen, W. Progress in patterned paper sizing for fabrication of paper-based microfluidic sensors. Cellulose 17, 649–659 (2010).
Kao, P.-K. & Hsu, C.-C. One-step rapid fabrication of paper-based microfluidic devices using fluorocarbon plasma polymerization. Microfluid. Nanofluidics 16, 811–818 (2014).
Spicar-Mihalic, P. et al. CO2 laser cutting and ablative etching for the fabrication of paper-based devices. J. Micromechanics Microengineering 23, 067003 (2013).
Fang, X., Wei, S. & Kong, J. Paper-based microfluidics with high resolution, cut on a glass fiber membrane for bioassays. Lab Chip 14, 911 (2014).
Haller, P. D., Flowers, C. A. & Gupta, M. Three-dimensional patterning of porous materials using vapor phase polymerization. Soft Matter 7, 2428 (2011).
Kwong, P. & Gupta, M. Vapor Phase Deposition of Functional Polymers onto Paper-Based Microfluidic Devices for Advanced Unit Operations. Anal. Chem. 84, 10129–10135 (2012).
Demirel, G. & Babur, E. Vapor-phase deposition of polymers as a simple and versatile technique to generate paper-based microfluidic platforms for bioassay applications. Analyst 139, 2326–2331 (2014).
Bruzewicz, D. A., Reches, M. & Whitesides, G. M. Low-Cost Printing of Poly(dimethylsiloxane) Barriers To Define Microchannels in Paper. Anal. Chem. 80, 3387–3392 (2008).
Lu, Y., Shi, W., Jiang, L., Qin, J. & Lin, B. Rapid prototyping of paper-based microfluidics with wax for low-cost, portable bioassay. Electrophoresis 30, 1497–1500 (2009).
Koo, Y., Sankar, J. & Yun, Y. High performance magnesium anode in paper-based microfluidic battery, powering on-chip fluorescence assay. Biomicrofluidics 8, 054104 (2014).
Abe, K., Suzuki, K. & Citterio, D. Inkjet-Printed Microfluidic Multianalyte Chemical Sensing Paper. Anal. Chem. 80, 6928–6934 (2008).
Su, W., Cook, B. S., Fang, Y. & Tentzeris, M. M. Fully inkjet-printed microfluidics: a solution to low-cost rapid three-dimensional microfluidics fabrication with numerous electrical and sensing applications. Sci. Rep. 6, 35111 (2016).
Yamada, K., Henares, T. G., Suzuki, K. & Citterio, D. Paper-Based Inkjet-Printed Microfluidic Analytical Devices. Angew. Chemie Int. Ed. 54, 5294–5310 (2015).
Xu, C., Cai, L., Zhong, M. & Zheng, S. Low-cost and rapid prototyping of microfluidic paper-based analytical devices by inkjet printing of permanent marker ink. RSC Adv. 5, 4770–4773 (2015).
Songjaroen, T., Dungchai, W., Chailapakul, O. & Laiwattanapaisal, W. Novel, simple and low-cost alternative method for fabrication of paper-based microfluidics by wax dipping. Talanta 85, 2587–2593 (2011).
Songjaroen, T., Dungchai, W., Chailapakul, O., Henry, C. S. & Laiwattanapaisal, W. Blood separation on microfluidic paper-based analytical devices. Lab Chip 12, 3392 (2012).
Dungchai, W., Chailapakul, O. & Henry, C. S. A low-cost, simple, and rapid fabrication method for paper-based microfluidics using wax screen-printing. Analyst 136, 77–82 (2011).
Sameenoi, Y., Nongkai, P. N., Nouanthavong, S., Henry, C. S. & Nacapricha, D. One-step polymer screen-printing for microfluidic paper-based analytical device (μPAD) fabrication. Analyst 139, 6580–6588 (2014).
de Tarso Garcia, P., Garcia Cardoso, T. M., Garcia, C. D., Carrilho, E. & Tomazelli Coltro, W. K. A handheld stamping process to fabricate microfluidic paper-based analytical devices with chemically modified surface for clinical assays. RSC Adv. 4, 37637–37644 (2014).
Akyazi, T., Saez, J., Elizalde, J. & Benito-Lopez, F. Fluidic flow delay by ionogel passive pumps in microfluidic paper-based analytical devices. Sensors Actuators B Chem. 233, 402–408 (2016).
Peeling, R. W., Holmes, K. K., Mabey, D. & Ronald, A. Rapid tests for sexually transmitted infections (STIs): the way forward. Sex. Transm. Infect. 82, v1–v6 (2006).
Mabey, D., Peeling, R. W., Ustianowski, A. & Perkins, M. D. Diagnostics for the developing world. Nat. Rev. Microbiol. 2, 231–240 (2004).
Chagas, C. L. S. et al. Hand drawing of pencil electrodes on paper platforms for contactless conductivity detection of inorganic cations in human tear samples using electrophoresis chips. Electrophoresis 36, 1837–1844 (2015).
Qin, Y. et al. Paper-Based, Hand-Drawn Free Chlorine Sensor with Poly(3,4-ethylenedioxythiophene):Poly(styrenesulfonate). Anal. Chem. 88, 10384–10389 (2016).
Toedt, J., Koza, D. & Van Cleef-Toedt, K. Chemical composition of everyday products. (Greenwood Press, 2005).
Lee, J. N., Park, C. & Whitesides, G. M. Solvent Compatibility of Poly(dimethylsiloxane)-Based Microfluidic Devices. Anal. Chem. 75, 6544–6554 (2003).
Dornelas, K. L., Dossi, N. & Piccin, E. A simple method for patterning poly(dimethylsiloxane) barriers in paper using contact-printing with low-cost rubber stamps. Anal. Chim. Acta 858, 82–90 (2015).
Wang, J. et al. Hydrophobic sol-gel channel patterning strategies for paper-based microfluidics. Lab Chip 14, 691–695 (2014).
Songok, J. et al. Paper-based micro fluidics: Fabrication technique and dynamics of capillary-driven surface flow. ACS Appl. Mater. Interfaces 6, 20060–20066 (2014).
Rajendra, V., Sicard, C., Brennan, J. D. & Brook, M. A. Printing silicone-based hydrophobic barriers on paper for microfluidic assays using low-cost ink jet printers. Analyst 139, 6361–6365 (2014).
Acknowledgements
We gratefully acknowledge the financial support provided by the Indian Institute of Technology Kharagpur [Sanction Letter no.: IIT/SRIC/ATDC/CEM/2013-14/118, dated 19.12.2013] and the Royal Academy of Engineering, United Kingdom [Newton-Bhaba Fund]. AP & NKM acknowledge the financial support from Vision Group on Science and Technology, Government of Karnataka under SMYSR and RGS/F Scheme [Sanction Letter no.: KSTePS/VGST/SMYSR-2016-17/GRD-595/2017-18, KSTePS/ VGST-RGS/F/GRD No.711/2017-18]. We greatly acknowledge Indian Science Academies for NKM’s fellowship. We thank Prof. Rabibrata Mukherjee & Mr. Palash Dhara for providing AFM facility. We acknowledge the help of Mr. Sumit Kumar for the contact angle measurements. We are also grateful to Dr. Bidhan Pramanick, Mr. Sankha Shuvra Das for their fruitful discussions. We thank Mr. Rajiv Das, BC Roy Technology Hospital for blood samples. We also thank Dr. Vijendra Prabhu, Dr. P. Y. Prakash & Prof. A. Senthil Kumar for proof reading the manuscript.
Author information
Authors and Affiliations
Contributions
S.C. directed and designed the research. N.K.M. designed the research, performed the experiments and analysed the data. A.P. performed preliminary experiments on fabrication, chemical compatibility, standard glucose and water penetration rate. S.K.B. was involved in preliminary studies using ink and data analysis. S.C., N.K.M. and S.K.B. wrote the manuscript with the inputs from the other author.
Corresponding author
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Mani, N.K., Prabhu, A., Biswas, S.K. et al. Fabricating Paper Based Devices Using Correction Pens. Sci Rep 9, 1752 (2019). https://doi.org/10.1038/s41598-018-38308-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-018-38308-6
This article is cited by
-
Smart Bags and Smartphone for On-The-Spot Detection of Bleached Garlic
Food Analytical Methods (2024)
-
Tuning Hydrophobicity of Paper Substrates for Effective Colorimetric detection of Glucose and Nucleic acids
Analytical and Bioanalytical Chemistry (2023)
-
3D Paper-based milk adulteration detection device
Scientific Reports (2022)
-
Latex-Based Paper Devices with Super Solvent Resistance for On-the-Spot Detection of Metanil Yellow in Food Samples