Abstract
While fluorescence microscopes and atomic force microscopes are widely used to visualize, track, and manipulate single biomolecules, the resolution of these methods is limited by sample drift. To minimize drift, active feedback methods have recently been used to stabilize single molecule microscopes on the sub-nanometer scale. However, these methods require high intensity lasers which limits their application in single molecule fluorescence measurements. Furthermore, these feedback methods do not track user-defined regions of the sample, but rather monitor the relative displacement of an unknown point on a fiducial marker, which limits their use in biological force measurements. To overcome these limitations, we have developed a novel method to image, track and stabilize a sample using low laser intensities. We demonstrate the capabilities of our approach by tracking a user-chosen point on a fiducial marker at 8.6 kHz and stabilizing it with sub-nanometer resolution. We further showcase the application of our method in single molecule fluorescence microscopy by imaging and stabilizing individual fluorescently-tagged streptavidin proteins under biologically relevant conditions. We anticipate that our method can be easily used to improve the resolution of a wide range of single molecule fluorescence microscopy and integrated force-fluorescence applications.
Similar content being viewed by others
Introduction
Imaging, tracking and manipulation of single biological molecules are widely used to obtain key mechanistic insights into biological processes. Single molecule fluorescence microscopes are routinely employed to localize and track proteins with nanometer spatial and millisecond temporal resolution1,2,3,4,5. Interactions between single biological molecules are also commonly measured using mechanical probes such as atomic force microscopes (AFMs)6,7,8,9. In fact there have been several recent attempts at combining single molecule AFM and fluorescence microscopy in order to measure force-induced changes in the conformation and activity of single proteins10,11 and to ‘cut and paste’ single fluorescent biomolecules for the bottom-up assembly of nanoscale structures12,13,14,15.
Despite their widespread use, the resolution of single molecule AFM and fluorescence methods is severely limited by thermal and mechanical drifts of the microscope and sample. This drift is particularly pronounced in fluid environments and at ambient temperatures: conditions which are essential for biologically relevant measurements. For instance, while it is theoretically possible to localize a single stationary fluorescent molecule with sub-nanometer accuracy, drift of both the microscope stage and the molecule itself blurs the fluorescence image and reduces localization precision. Similarly, when probing a single biomolecule with AFM, undesired drift of the sample and the AFM tip makes it impossible to repeatedly address the same molecule. Drift has also made it particularly difficult to develop instrumentation that integrates single molecule AFM and fluorescence-based measurements in a high-throughput fashion.
Since cryogenic temperature measurements, which have traditionally been used to combat undesired drift16,17, are not compatible with biological experiments, several active stabilization methods for microscopes have recently been developed18,19,20,21,22,23. Of these techniques, back-scatter based active stabilization provides the high spatial and temporal resolution required to track molecular dynamics. Back-scatter based active stabilization platforms, which are commonly called ultrastable microscopes, typically reflect a laser off a sample fiducial and detect the backscattered light using a position sensitive detector. To map detector signals to drift, the fiducial is initially scanned through a known area and the detector response is correlated to fiducial displacement via a polynomial fit for each spatial dimension18. With the detector signal correlated to fiducial displacement, sample movement is monitored in real-time and the sample is stabilized by moving the microscope stage. While back-scatter based active stabilization achieves impressive angstrom-level stability over a significant bandwidth (0.1–10 Hz), it requires high laser intensities which quickly and irreversibly bleach fluorescent dyes. This precludes the use of this active-feedback method in single molecule fluorescence microscopy, which requires significantly lower laser intensities. Furthermore, while the mapping technique used in back-scatter based active stabilization provides high displacement resolution, it does not lend itself to pinpointing a specific feature on the fiducial. For instance, when this method has been used to stabilize an AFM18, the AFM tip is not directly imaged, and as such the precise location of the tip apex is unknown. Consequently, the location of the AFM tip apex is determined by scanning the tip over a hard feature on the sample. This limits the use of this feedback method in AFM based single molecule force measurements, where the tip is functionalized with soft biological molecules that can be easily destroyed by compressive forces during scanning.
To extend active feedback to single molecule fluorescence applications, we have developed an optical feedback method that uses laser intensities which are several orders of magnitude lower than existing ultrastable methods. Instead of using a fitting approach to map fiducial drift, we utilize a simple discrete map to track fiducial motion with high temporal resolution. Since we directly image the fiducial, our method allows us to stabilize a user-selected point, thereby adding high accuracy to the precision of ultrastable microscopy. We demonstrate the capabilities of our method by stabilizing a user-defined point on a fiducial with sub-nanometer resolution and at a tracking rate of 8.6 kHz. To showcase its application in single molecule fluorescence microscopy, we demonstrate this method by stabilizing individual fluorescently-tagged streptavidin proteins under biologically relevant conditions.
Methods
Imaging a user-defined point on a fiducial
Since previous ultrastable techniques have focused on active stabilization of an AFM tip18,19, we directly benchmarked our method using an AFM tip as a fiducial for sample stabilization. An AFM cantilever (Oxford Instruments, TR400PSA) was glued to a coverslip and mounted on a custom confocal fluorescence microscope24, with the apex of the AFM tip pointed towards the objective (Fig. 1(a)). For reference, an SEM image of an identical tip is shown in Fig. 1(b). A 532 nm laser, coupled to a single mode fiber, was circularly polarized and focused on the AFM tip with a high numerical aperture (1.42) oil immersion objective. The low intensity laser light (4-5 µW) used in our experiments is compatible with single molecule fluorescence microscopy. The piezoelectric sample stage (Physik Instrumente, P-733.3, E-712 controller) raster-scanned the fiducial through the focused laser, and light backscattered off the tip was collected by the objective and directed through a quarter waveplate and polarizing cube beam splitter, splitting the backscattered and incident light18. Backscattered light was then focused by a 4X objective through a 50 μm pinhole onto the detection face of a quadrant avalanche photodiode (QAPD) (First Sensor, QA4000-10) operated in proportional mode. Using a QAPD rather than a conventional quadrant photodiode (QPD) allows for detection of reflected laser light that is several orders of magnitude lower in intensity, making this detection scheme ideal for low light fluorescence applications.
The QAPD was mounted on a custom printed circuit board which converted the APD quadrant currents into voltages, and output three voltage values: total current from the sensor, the difference between the left and right halves of the sensor, and the difference between the bottom and top halves of the sensor. A field programmable gate array (FPGA, National Instruments, USB 7856-R utilizing a Xilinx Kintex-7 K160T chip) rapidly digitized each value with an onboard multichannel successive approximation register ADC and produced two difference signals. The difference signals were normalized by dividing each by the total (sum) signal to eliminate sensitivity to intensity fluctuations. These normalized values yield the 2D position of the reflected laser on the face of the detector, where the laser position on the detector face varies deterministically with the 2D position of the sample fiducial.
Transmitted backscattered light was reflected by a second dichroic and focused by a 4X objective on the detection face of a single photon-counting avalanche photodiode (APD, Micro Photon Devices). The photon counts were timestamped with the FPGA’s onboard clock and stored in onboard registers. This allowed direct imaging of the fiducial produced by the sum channel of the QAPD and the APD (Fig. 1(c) and (d)). Furthermore, because the AFM tip is made of silicon nitride, which is known to luminesce at a longer wavelength when excited with 532 nm light25,26, an image of the AFM tip was simultaneously generated by using a 4X objective to focus the luminescence on a second APD (Fig. 1(e)). As described in the results section, simultaneously acquired back-scattered (Fig. 1(c) and (d)) and luminescence images of the AFM tip (Fig. 1(e)) permitted us to unambiguously identify the tip apex. In subsequent measurements, we used the tip apex as a fiducial, generating high resolution maps for tracking the tip.
Tracking fiducial position
Upon locating the AFM tip apex, we scanned a small area (100 nm × 100 nm) around this point with the piezo stage. The difference channels from the QAPD were digitized, normalized, and stored on the host computer for processing. The data was formatted to produce a 2D scan image, with the horizontal and vertical difference channels shown in Fig. 2(a) and (b) respectively. The full range of the scan was evenly divided into contour lines (Fig. 2(c) and (d)). The 2D position data associated with the intersections of these contour lines were stored on the FPGA to provide a fast conversion from normalized measurement to tip position (Supplementary Methods). The number of contours was optimized to provide high resolution within the constraint of FPGA memory utilization. For the results presented here, each 100 nm × 100 nm map was split into 200 contours, yielding sub-nanometer resolution (an average of 0.5 nm between contour intersections). Each contour’s position from the horizontal difference channel on the QAPD was checked against each vertical difference channel. Every intersection’s 16-bit X and Y coordinates were joined into a single 32-bit number and stored in order from top left to bottom right (Fig. 2(e)). Contours without an intersection used a placeholder value to maintain the order of the intersections. The end result of this process was an ordered 40,000 member (200 contours from each difference channel) 1D array, henceforth referred to as a “map”, which was stored on the FPGA’s memory for quick access.
For live tracking, the real-time normalized difference channel values were converted to an index in FPGA memory and the corresponding X, Y coordinates were retrieved from memory. This method minimizes the calculations required in the FPGA and maximizes the rate at which the position of the fiducial can be determined, producing a new position value at a rate of 1 MHz - limited by the digitization rate. In the FPGA logic we combine this with digitization-level averaging (100-point average of non-overlapping points) to reduce stochastic noise, which results in a position tracking rate of 8.6 kHz.
Stabilizing fiducial position using feedback
The tracked fiducial position is used in the FPGA to calculate the deviation from a desired setpoint. Feedback commands to the sample stage are also generated on the FPGA and transmitted to the stage controller through a direct interface. By using an array of digital output lines from the FPGA, we control the position of the piezo sample stage, commanding movement at a rate up to the piezo controller’s servo update rate (20 kHz). This allows us to optionally maintain a desired position on the map by finding the displacement of the current tracked position from a desired position, and commanding the stage move to compensate for the measured displacement, forming a feedback loop. While commanding movement, we simultaneously poll the stage’s reported position. Completing these actions after a tracked position was calculated reduced the feedback rate to 5.9 kHz. The actual movement of the stage was slower due to the piezo’s physical stiffness limitation, with a resonant frequency of 340 Hz. Data on the tracked fiducial and stage position were periodically sent to the host computer through a set of buffers, decoupling the FPGA real-time processing from monitoring on the host computer.
Active stabilization of single fluorescent molecules
To demonstrate our stabilization technique for fluorescent single molecules, we immobilized fluorescent streptavidin (Cy3-Sta) proteins on a glass coverslip. We first cleaned coverslips in a piranha solution (75% H2SO4: 25% H2O2) and glued an AFM chip in place using marine epoxy. We then placed a 100 µL drop of 1 g/L biotinylated bovine serum albumin (BSA-biotin) near the fiducial. The BSA readily adsorbs to the surface of the glass coverslip, while the attached biotin serves as a strong binding point for Cy3-Sta. We allowed the BSA-biotin to adsorb to the surface for 30 mins before rinsing the area with buffer. Next, we incubated the coverslip with 50 pM Cy3-Sta for 15 min. This protocol ensures optically resolvable binding of individual Cy3-Sta. After incubation we rinsed the coverslip twice and imaged in buffer by raster scanning the sample.
Because the feedback demonstrations took place over hours, we ensured the laser focus was maintained by locking the relative distance between the sample and the objective with the CRISP autofocus system (Applied Scientific Instrumentation, MS-2000 controller). This system uses an infrared LED inserted into the optical pathway to determine the distance between the objective and the sample, correcting any displacement by changing the voltage on a z-axis piezo (Physik Instrumente, PD72Z1CAQ) mounted between the microscope nosepiece and the objective.
Results
We simultaneously obtained luminescence and backscatter images of the AFM tip (Fig. 1). The luminescence image of the AFM tip apex (Fig. 1(e)) was fit to a 2D Gaussian, which allowed us to accurately determine the apex position. Backscatter images of the pyramidal AFM tip, simultaneously generated by the QAPD and the APD (Fig. 1(c) and (d)), were similar and also permitted visualization of the tip apex. The tip apex determined from the backscattered images were within one standard deviation of a 2D Gaussian fit of the luminescence intensity, putting all three detection schemes in good agreement. For subsequent tracking and feedback, we used only the backscattered light from the fiducial.
Using our tracking scheme, we tracked the 2D position of the tip apex in real-time at 8.6 kHz, and then turned on feedback and tracked the tip at 5.9 kHz (Fig. 3(a–c)). As Fig. 3(a) and (b) show, without feedback the tracked position varies as much as 55 nm in X and 35 nm in Y in only ten seconds. In contrast, with our feedback system active (Fig. 3(c)) the setpoint is maintained with sub-nanometer RMS along both axes, as shown in the projections (Fig. 3(d) and (g)). These histograms of the stabilized X and Y position monitored for ten seconds show a tight grouping around the zero setpoint (RMS = 0.48 nm and 0.85 nm, respectively). These values are consistent with the spacing between intersects during map generation and the fact that the relationship between tracked and stage coordinates is non-linear (see Fig. 2). Fourier transforms of the tracked data with feedback off and feedback on (Fig. 3(e) and (h)) demonstrate that fiducial motion with amplitudes exceeding ~0.2 nm (all in the lower frequency ranges) are effectively eliminated when feedback is enabled. To confirm our feedback method also corrects drift over longer time periods, we first tracked the AFM tip apex and used feedback to maintain the zero setpoint position for 4.5 minutes. We then turned feedback off and allowed the tip apex to drift for the remainder of the time (Fig. 3(f) and (i)). These data show a faithful handoff between stage movement and tracked position when feedback is toggled.
To correlate tracked position with stage location, we tracked the fiducial position while commanding the stage to step back and forth a set distance along each axis, with 17 ms wait between each position. First, we correlated tracked and stage sensor positions along the X axis. At each of 250 stage positions, 200 tracked and stage sensor positions were collected and averaged. The displacements between adjacent steps were recorded. This was done for 0 nm, 2 nm, 4 nm, 6 nm, 8 nm, 10 nm, and 20 nm step sizes, with the resulting displacements plotted in Fig. 4(a). As seen in the figure, the stage’s sensor displacement was tightly grouped for each step size while sample fluctuations caused the tracked displacement to vary a great deal. In this configuration, the mean of the tracked values is significantly shifted (0.15 nm) from the commanded 4 nm displacement due to the sample drift over the time required to collect the data (Fig. 4(b)). Next, we repeated the step tests for the Y axis while the X axis was held steady. For these tests, we collected and averaged 700 tracked and stage sensor positions for 125 movements with the same step distances (Fig. 4(c)). Along the Y axis the mean step-size is 0.86 nm greater than the commanded step-size of 4 nm (Fig. 4(d)).
Next, to demonstrate our method’s accuracy and precision, the feedback was turned on and the experiment was repeated for the X axis with the feedback set-point varied instead of the commanded stage position. As expected, these results showed a tight grouping of tracked displacement while the sensor readout was spread as the stage moved to compensate for sample fluctuations (Fig. 5(a)). By projecting the +4 nm data from Fig. 5(a) and fitting the histogram to a Gaussian, we obtained a measure of the feedback accuracy and precision. With a commanded 4 nm movement, the tracked step size was peaked at 4.015 ± 0.004 nm, a difference of just 15 ± 4 pm which corresponds to feedback accuracy (Fig. 5(b)). The standard deviation of the measured step sizes was 130 pm, which demonstrates the precision of this technique relative to the sample stage. When these experiments were repeated along the Y axis (Fig. 5(c)), the corresponding accuracy and precision are 10 ± 11 pm and 160 pm, respectively (Fig. 5(d)).
With the feedback method verified and benchmarked, we sought to stabilize single fluorescent biomolecules under physiologically relevant conditions. As described in the methods section, we immobilized optically well separated Cy3-Sta on our sample. We imaged fluorescent molecules near our fiducial and then allowed the system to drift for 100 minutes. We scanned the area a second time, and the resulting Cy3-Sta displacement can be seen in the overlaid image, with markers on the fluorophores’ centroid determined by a 2D Gaussian fit (Fig. 6(a)). By taking the difference of eight such centroids from a wider view, before and after drift, we determined the drift is 4.9 pixels in X and 3.02 pixels in Y (pixel size = 60 nm × 60 nm). Next, we imaged a new set of Cy3-Sta near the fiducial, after which feedback was enabled for 100 minutes before rescanning the area. The stage sensor values in Fig. 6(b) show the 2D drift correction during this time. The second set of overlaid images with fitted centroids are shown in Fig. 6(c). The average of eleven centroid displacements was only 0.4 pixels in X and 0.06 pixels in Y. It is important to note that since feedback was turned off during image acquisition, centroids measured in subsequent images were offset due to sample drift. During the 3 seconds it takes to image a single Cy3-Sta, a sample can drift up to 55 nm (Fig. 3(a) and (b)) which may account for the tiny centroid displacements measured in the presence of feedback (Fig. 6(c)).
Discussion
The data presented here demonstrate a feedback method that maintains fiducial location with sub-nanometer accuracy and precision while using low intensity laser light. This configuration is ideal for single molecule fluorescence measurements. Because our resolution is currently limited by the resolution of the map, the effective resolution of our technique can be easily increased by either increasing the number of contours used to find intersections or decreasing the map scan area. Ultimately, the FPGA onboard memory limits how many intersections can be stored, but an optimal map can be determined by balancing desired travel range with intersect density.
To implement our method in AFM applications, feedback in three dimensions is necessary. It is important to point out that our method is not fundamentally limited to two dimensions. We can extend our method to stabilize a sample in 3D, for instance, by utilizing a set of maps at different heights along the optical axis.
While we tracked the position of the fiducial at a rate of 8.6 kHz, our tracking speed was limited by the fact that we used an average of 100 consecutive digitized points to reduce noise in the tracked position. At the possible expense of a lowered signal to noise ratio, reducing the number of averages or eliminating them altogether could greatly increase the tracking speed. The amount of averaging can be tuned to optimize the method for different applications.
We anticipate that the feedback technology presented here has broad applications in stabilizing instruments that combine single molecule AFM and fluorescence-based approaches. An existing version of this instrument, termed ‘single-molecule cut-and-paste’12, has been used to construct biological nanostructures by using a specially functionalized AFM tip to move molecules from a reservoir to a target area, one molecule at a time. Our feedback method will allow the AFM tip to repeatedly address the same location several times, potentially constructing more complex structures. By addressing specific locations or fiducials, the placement of molecules would also be specific, eliminating the need to ‘blindly’ pick up and place molecules over a large area.
Our method also has broad applications in stabilizing conventional fluorescence microscopes. In these microscopes, long term molecular dynamics are visualized by taking repeated fluorescence images, while utilizing feedback methods to stabilize the microscope so that the molecules do not drift out of view. Unlike other ultrastable platforms which employ a high laser intensity that would photobleach the sample, our method uses lower light intensity. We have demonstrated our method can stabilize individual fluorescent molecules without photobleaching them.
In addition to stabilizing microscope platforms, our method has broad applications in tracking the motion of single molecules in vitro and in live-cells. As many biological processes rely on sub-millisecond motions of single molecules, the high tracking rate presented here would allow these movements to be monitored with high temporal resolution. For instance, protein movement has previously been tracked by covalently attaching gold nanoparticles (GNPs) to a protein of interest and monitoring backscattered light on a QPD27. While this method boasts an impressive 40 kHz tracking rate with 1.5 nanometer resolution, the tracking is not conducted in “real-time” since the position data cannot be retrieved until the experiment is completed. In contrast, our tracking method allows for sub-nanometer tracking resolution and real-time position information. Furthermore, our low laser intensity operation would allow this sort of tracking to be done in conjunction with fluorescence microscopy techniques.
References
Gordon, M. P., Ha, T. & Selvin, P. R. Single-molecule high-resolution imaging with photobleaching. Proc. Natl. Acad. Sci. USA 101, 6462–6465 (2004).
Churchman, L. S., Okten, Z., Rock, R. S., Dawson, J. F. & Spudich, J. A. Single molecule high-resolution colocalization of Cy3 and Cy5 attached to macromolecules measures intramolecular distances through time. Proc. Natl. Acad. Sci. USA 102, 1419–1423 (2005).
Qu, X., Wu, D., Mets, L. & Scherer, N. F. Nanometer-localized multiple single-molecule fluorescence microscopy. Proc. Natl. Acad. Sci. USA 101, 11298–11303 (2004).
Thompson, R. E., Larson, D. R. & Webb, W. W. Precise nanometer localization analysis for individual fluorescent probes. Biophys. J. 82, 2775–2783 (2002).
Yildiz, A. et al. Myosin V walks hand-over-hand: single fluorophore imaging with 1.5-nm localization. Science 300, 2061–2065 (2003).
Butt, H. J., Cappella, B. & Kappl, M. Force measurements with the atomic force microscope: Technique, interpretation and applications. Surf. Sci. Rep. 59, 1–152 (2005).
Hinterdorfer, P. & Dufrene, Y. F. Detection and localization of single molecular recognition events using atomic force microscopy. Nat. Methods 3, 347–355 (2006).
Hugel, T. & Seitz, M. The study of molecular interactions by AFM force spectroscopy. Macromol. Rapid Comm. 22, 989–1016 (2001).
Manibog, K., Yen, C. & Sivasankar, S. Measuring Force-Induced Dissociation Kinetics of Protein Complexes Using Single-Molecule Atomic Force Microscopy. Methods Enzymol. 582, 297–320 (2017).
Gumpp, H. et al. Triggering Enzymatic Activity with Force. Nano Lett. 9, 3290–3295 (2009).
He, Y., Lu, M., Cao, J. & Lu, H. P. Manipulating protein conformations by single-molecule AFM-FRET nanoscopy. ACS Nano 6, 1221–1229 (2012).
Kufer, S. K., Puchner, E. M., Gumpp, H., Liedl, T. & Gaub, H. E. Single-molecule cut-and-paste surface assembly. Science 319, 594–596 (2008).
Strackharn, M., Stahl, S. W., Puchner, E. M. & Gaub, H. E. Functional Assembly of Aptamer Binding Sites by Single-Molecule Cut-and-Paste. Nano Lett. 12, 2425–2428 (2012).
Strackharn, M., Pippig, D. A., Meyer, P., Stahl, S. W. & Gaub, H. E. Nanoscale Arrangement of Proteins by Single-Molecule Cut-and-Paste. J. Am. Chem. Soc. 134, 15193–15196 (2012).
Puchner, E. M., Kufer, S. K., Strackharn, M., Stahl, S. W. & Gaub, H. E. Nanoparticle Self-Assembly on a DNA-Scaffold Written by Single-Molecule Cut-and-Paste. Nano Lett. 8, 3692–3695 (2008).
Eigler, D. M. & Schweizer, E. K. Positioning single atoms with a scanning tunneling microscope. Nature 344, 524–526 (1990).
Stroscio, J. A. & Eigler, D. M. Atomic and molecular manipulation with the scanning tunneling microscope. Science 254, 1319–1326 (1991).
King, G. M., Carter, A. R., Churnside, A. B., Eberle, L. S. & Perkins, T. T. Ultrastable Atomic Force Microscopy: Atomic-Scale Stability and Registration in Ambient Conditions. Nano Lett. 9, 1451–1456 (2009).
Sigdel, K. P., Grayer, J. S. & King, G. M. Three-Dimensional Atomic Force Microscopy: Interaction Force Vector by Direct Observation of Tip Trajectory. Nano Lett. 13, 5106–5111 (2013).
Churnside, A. B. & Perkins, T. T. Ultrastable atomic force microscopy: Improved force and positional stability. FEBS Lett. 588, 3621–3630 (2014).
Youn, Y., Ishitsuka, Y., Jin, C. & Selvin, P. R. Thermal nanoimprint lithography for drift correction in super-resolution fluorescence microscopy. Opt. Express 26, 1670–1680 (2018).
Lee, S. H. et al. Using fixed fiduciary markers for stage drift correction. Opt. Express 20, 12177–12183 (2012).
Walder, R., Paik, D. H., Bull, M. S., Sauer, C. & Perkins, T. T. Ultrastable measurement platform: sub-nm drift over hours in 3D at room temperature. Opt. Express 23, 16554–16564 (2015).
Li, H., Yen, C.-F. & Sivasankar, S. Fluorescence Axial Localization with Nanometer Accuracy and Precision. Nano Lett. 12, 3731–3735 (2012).
Poolton, N. & Cros, Y. Luminescence excitation and bleaching involving dangling bonds in silicon nitride. J. Phys. I 1, 1335–1345 (1991).
Aydinli, A., Serpengüzel, A. & Vardar, D. Visible photoluminescence from low temperature deposited hydrogenated amorphous silicon nitride. Solid State Commun. 98, 273–277 (1996).
Nan, X., Sims, P. A. & Xie, X. S. Organelle Tracking in a Living Cell with Microsecond Time Resolution and Nanometer Spatial Precision. ChemPhysChem 9, 707–712 (2008).
Acknowledgements
Research reported in this publication was supported in part by the National Institute of General Medical Sciences of the National Institutes of Health (R01GM121885) and by a Signature Research Initiative Award from the College of Liberal Arts and Science at Iowa State University. The content is solely the responsibility of the authors and does not necessarily represent the official views of the funding agencies.
Author information
Authors and Affiliations
Contributions
P.D.S., J.G.L., and S.S. designed the study. P.D.S. and B.H.R. performed experiments and analyzed data. P.D.S., J.G.L., and S.S. wrote the manuscript. All authors edited and approved the final manuscript.
Corresponding author
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Schmidt, P.D., Reichert, B.H., Lajoie, J.G. et al. Method for high frequency tracking and sub-nm sample stabilization in single molecule fluorescence microscopy. Sci Rep 8, 13912 (2018). https://doi.org/10.1038/s41598-018-32012-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-018-32012-1
Keywords
This article is cited by
-
An alternative framework for fluorescence correlation spectroscopy
Nature Communications (2019)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.