Abstract
We present a method for continuous, photon energy dependent micro Raman spectroscopy. A narrow excitation line is selected from a continuum laser by an acousto-optic tunable filter (AOTF) plus an additional monochromator (MC). Automation of laser, AOTF, MC and tunable long pass filters enables us to continuously scan the wavelength over the full visible range while synchronously acquiring Raman spectra over a photon energy range from 1.85 eV to 2.83 eV. We demonstrate the applicability of our method on a well-studied sample, reduced graphene oxide (rGO), where we measure the Raman scattering over the whole visual range and use the photon energy dependence of the D, G and GS band as verification for the method we present here. We complement this set of data with additional results from a Ti:sapphire laser source, covering the 1.75 to 1.41 eV range. From the full photon energy range of 1.41 to 2.83 eV, we noticed a small deviation from linearity for the dispersion of the D band.
Similar content being viewed by others
Introduction
Besides the typical photon energy (~ν4) dependence of the Raman signal1, many materials2,3,4 as well as nanomaterials, such as quantum dots5,6, carbon nanotubes7,8,9,10,11, semiconductors12, semiconducting nanowires13,14 and graphene15,16 show resonances at specific photon energy ranges accompanied by strong enhancement of the Raman signal by several orders of magnitude2,3,9,16,17,18,19. Accurate determination of these resonances requires light sources with large tunable energy ranges and narrow emission lines. Commonly used gas lasers (Ar+, Kr+, HeNe …) provide only a few distinct emission lines within the visible range not capable of covering photon energies in a quasi-continuous way. Although tunable sources like dye lasers, Ti:sapphire lasers and OPOs are available, they usually require realignment to change the wavelength over a large spectral range. Here we demonstrate a solution by applying a continuum laser source which can provide a continuous emission spectrum typically ranging from 3.1 eV (400 nm) to 0.517 eV (2400 nm). Selection of the desired excitation energy (2.83 to 1.85 eV) for the Raman measurement was here achieved by sending the laser continuum output through an AOTF and an additional monochromator (MC). The power demanding NIR range from 1.75 to 1.41 eV was covered with and additional Ti:sapphire laser which was sent through the monochromator as well. In combination with tunable long pass filters, this approach enables for recording Raman signals in the best case from approximately 200 cm−1 while continuously scanning the photon energy. To demonstrate our method, we used rGO, which provides a good Raman signal over the whole visible range20,21,22. The well-characterized Raman bands of rGO in the literature act as a good reference to validate our new method23,24. The reduced graphene oxide (rGO) was synthesized according to the method from Eigler et al. and recently used as an ink for printing semitransparent electrodes25,26.
Results and Discussion
Besides having sufficient intensity, a Raman spectroscopy light source should offer a spectral narrow emission line. Therefore, gas ion lasers are typically used since the underlying atomic transitions provide very sharp laser lines. Unfortunately, these gas lasers only provide fixed photon energies which prevent continuous scanning of the energy in Raman spectroscopy.
In recent years, continuum laser sources have become available and offer an alternative possibility delivering a continuous spectrum. Implementing AOTFs and MCs enables to select a specific photon energy. This approach delivers laser tuning in steps only limited by the MC (~0.1 nm for 600 grooves/mm grating). Due to typical light scatter levels (10−4–10−6) in commonly used MCs, suppression of the undesired photon energies might not be sufficient27. These residuals can be further suppressed by applying double MCs or simply sending the laser light twice through a single MC as performed in this work. The scanning of the excitation energy can easily be achieved by turning the MC grating which is usually equipped with a stepper motor. The continuous sweep of excitation energy requires also an adaptation of the long pass filters in case of monitoring low wavenumber Raman bands. This adaptation is accomplished by tunable long pass filters whose transmission depends on the incidence angle of the incoming light.
In the present experiment, six different tunable filters are used to cover the full visible and parts of the NIR range. Figure 1A shows a schematic representation of the experimental setup. A continuum laser (SuperK EXTREME EXB-6 with SuperK SELECT AOTF wavelength selector) was integrated into a confocal microscope to cover the visible excitation range. The SuperK SELECT provides a laser line of about 3.5 nm full width at half maximum (see Fig. 1B dashed line). The light is sent twice through a Czerny-Turner-type MC (Acton Research SP300i) with a blazed grating of 600 grooves/mm and a focusing length of 0.3 m resulting in an adjustable laser line of 0.3 nm full width at half maximum and a minimum step size of about 0.12 nm (see Fig. 1B solid line). For measuring Raman spectra in the NIR range (1.75 to 1.41 eV) a Ti:sapphire laser is used and sent through the monochromator as well for cleaning up the laser line. The laser light is then reflected on a 30:70 beam splitter (XF122 Omega Optical) and focused onto the sample by an oil immersion objective (Olympus, UPLFLN 100× NA = 1.3). According to the laser output the intensity varies between 0.25 to 0.5 kW/cm2 in the visible range for the continuum laser and between 30 to 200 kW/cm2 in the NIR range for the Ti:sapphire laser (see supporting information). The Raman signal is collected by the same objective. Afterwards, the remaining laser light is blocked by a tunable long pass filter chosen according to the actual excitation range (VersaChrome Edge tunable long pass filters, 448 nm to 501 nm, 501 nm to 561 nm, 561 nm to 628 nm, 628 nm to 704 nm, 697 nm to 805 nm and 783 nm to 905 nm, Semrock) and mounted on a home-buildt rotational stage while the Raman signal is acquired with a spectrograph (Princeton Instruments SPEC-10:100B/LN_eXcelon CCD camera, SP 2356 spectrometer, 300 or 1200 grooves/mm). Wavelength calibration of all measurement ranges was performed with a neon spectral lamp (6032 Newport). This includes also correction of the tilting influence of the tunable filters. The silicon peak of the substrate at 520 cm−1 served as a reference point in all spectra since it is more accurate than the MC determined excitation energy. A 10:90 beam splitter provided the signal for laser power correction which was acquired with a photodiode power sensor (S120VC Thorlabs). Synchronization of laser, monochromator, tunable long pass filters, spectrograph and photometer was accomplished by self-written LabVIEW routines.
As compared to standard Raman instruments with fixed laser sources the clear advantage of our approach is the flexibility in choosing variable excitation energies over a broad range from theoretically 400 nm to 2400 nm. The light source can be tuned in steps of 0.12 nm as mentioned above. The most limiting factor for applications is the comparably low available excitation power (up to 0.5 kW/cm2, see Fig. S1 supporting information) which requires long exposure times of up to several 100 seconds. Measuring higher resolution Raman spectra (for example with 1200 grooves/mm gratings and above) will be even more time consuming, since the signal of a Raman band will be spread over more pixels. However, searching for resonances in materials like CNTs should be facilitated by the versatility of the proposed method. In addition, one must be aware that a tunable laser source always requires either a separate precise determination of the current photon energy or a reference Raman signal, which in our case is given by the silicon peak.
We demonstrate the feasibility of our Raman scanning approach on a sample of rGO synthesized according to the method from Eigler et al.25. rGO as well as graphite and graphene show strong Raman signals making these samples ideal for our demonstration. Particularly suitable are the D and G bands, which arise from bond stretching of pairs of sp2 atoms and the breathing modes of sp2 atoms21,22. The sample was prepared by oxidizing graphite flakes in concentrated sulfuric acid and permanganate under careful cooling. Subsequently, the resulting graphene oxide was spin coated onto a silicon wafer and reduced with a 1:1 mixture of trifluoroacetic acid and hydroiodic acid. The vapor generated the reduced graphene oxide26.
Raman spectra for visible excitation light were recorded for 150 s in the range from 2.83 eV to 2.20 eV and for 200 s in the range from 2.20 eV to 1.85 eV. The spectra are shown in Fig. 2. Additional Raman spectra for NIR excitation from 1.75 eV to 1.41 eV can be found in Fig. S2 in the supporting information. The assignment of the bands was done based on the nomenclature by Heller et al.28. The shoulder of the G peak, becoming more pronounced at excitation energies below 2.5 eV, is historically and more commonly referred to as D’, but we will use in this manuscript GS instead22,29,30,31. Besides the pronounced peak of the silicon substrate at 520 cm−1 the D and G band are present at about 1370 cm−1 and 1590 cm−1 for the highest photon energy16,22. At this energy, the D feature has an amplitude about twice as much as the G feature. In addition, the 2D and D + GS are visible at 2695 cm−1 and 2935 cm−1, respectively. Depending on the transmission profile of the tunable long pass filters that we used, a maximum transmission window of approximately 200 cm−1 to 3700 cm−1 was achieved for specific photon energies, while a minimum transmission range of 500 cm−1 to 2300 cm−1 was achieved over the whole 2.83 eV - 1.85 eV energy range. The latter can be seen for example as some of the higher wavenumber 2D and D + GS bands are cut-off at specific energies.
As previously reported, the position of the disorder-induced D band shows a photon energy dependence with a shift of about 50 cm−1/eV21,22,32,33,34. We applied Lorentzian fits to the D and G peak features to determine the peak positions and compare them to the literature. The peak positions are shown in Fig. 3A as a function of photon energy. In agreement with previous studies on carbon materials, the D band features a shift from 1374 ± 5 cm−1 at 2.83 eV to 1297 ± 5 cm−1 at 1.41 eV21,32,33,34. Based on a linear fit to the data, we obtained a shift of (51.5 ± 0.5) cm−1/eV, in agreement with the literature33. However, precise inspection of the D-related data in Fig. 3A show a systematic deviaton from a linear energy dependency which can be best fitted by a parabolic function. A similar trend can be seen in the data from Pócsik et al. which, however, has not been further discussed32. In addition, the G band is not expected to show a significant dispersion as was demonstrated for microcrystalline graphite and in the framework of Kramers-Heisenberg-Dirac theory based calculations28,32. Nevertheless, as shown here in Fig. 3A and in other studies, the G feature seems to shift with decreasing photon energy from 1585 ± 5 cm−1 at 2.83 eV to about 1600 ± 5 cm−1 at 1.85 eV35,36. We interpret this apparent shift as a consequence of the overlap between the G band and the defect induced GS band with increased amplitudes of GS at lower photon energies28,37,38,39,40,41. This causes an apparent shift of the Raman frequencies when fitting the spectra with only one Lorentzian peak. This is also consistent with findings from King et al.36. To verify this, we acquired spectra with higher resolution (spectrometer grating with 1200 grooves/mm) and by using more intense gas lasers (75–125 kW/cm2, approximately 100 times higher intensity than the CL). Three representative spectra for 2.54 eV (blue), 2.28 eV (green) and 1.96 eV (red) are shown in the range between 1500 cm−1 and 1700 cm−1 in Fig. 3B). From the spectra, it is obvious that G and GS are overlapping, which results in an apparent photon energy dependent shift when fitting with only one Lorentzian peak35,38.
Conclusion
We introduced a novel approach of continuous scanning Raman spectroscopy based on a continuum laser source which enables Raman spectroscopy over the whole visible (and essentially also NIR) range. Applying our method to chemically reduced graphene oxide25,26, we acquired consecutive Raman spectra excited over the full visible range (438–672 nm) and parts of the NIR range (710–880 nm). Analysis of Raman spectra by fitting the D and G band with single Lorentzian functions revealed good literature agreement with the observed energy dispersions of both bands. The shift of about (51.5 ± 0.5) cm−1/eV for the D band is in good agreement with literature. However, a so far not so often discussed minor deviation from a linear dispersion is observed. Furthermore, we can attribute the apparent shift of the G band to an overlap of G and GS 21,32. For the presented method, the spectral resolution with respect to the laser linewidth and step size, is determined by the utilized diffraction grating of the monochromator and can be adapted to the specific material under investigation. Finally, more powerful continuum laser sources than the one we used, will result in a reduced integration time per spectrum. Given the application of Raman spectroscopy in all fields of materials sciences, we believe that our approach can add versatility and easy access to photon energy dependent Raman spectra.
References
Bernath, P. F. Spectra of atoms and molecules. (Oxford university press, 2015).
Behringer, J. In Raman spectroscopy 168–223 (Springer, 1967).
Johnson, B. B. & Peticolas, W. L. The resonant Raman effect. Annu. Rev. Phys. Chem. 27, 465–521 (1976).
Spiro, T. G. & Strekas, T. C. Resonance Raman spectra of heme proteins. Effects of oxidation and spin state. J. Am. Chem. Soc. 96, 338–345 (1974).
Cheng, H.-M., Lin, K.-F., Hsu, H.-C. & Hsieh, W.-F. Size dependence of photoluminescence and resonant Raman scattering from ZnO quantum dots. Appl. Phys. Lett. 88, 261909 (2006).
Dzhagan, V. et al. Resonant Raman scattering study of CdSe nanocrystals passivated with CdS and ZnS. Nanotechnology 18, 285701 (2007).
Dresselhaus, M. S., Dresselhaus, G., Jorio, A., Souza Filho, A. G. & Saito, R. Raman spectroscopy on isolated single wall carbon nanotubes. Carbon 40, 2043–2061 (2002).
Dresselhaus, M. S., Dresselhaus, G., Saito, R. & Jorio, A. Raman spectroscopy of carbon nanotubes. Phys. Rep. 409, 47–99 (2005).
Fantini, C. et al. Optical Transition Energies for Carbon Nanotubes from Resonant Raman Spectroscopy: Environment and Temperature Effects. Phys. Rev. Lett. 93, 147406 (2004).
Pimenta, M. et al. Resonant Raman effect in single-wall carbon nanotubes. J. Mater. Res. 13, 2396–2404 (1998).
Telg, H., Maultzsch, J., Reich, S., Hennrich, F. & Thomsen, C. Chirality distribution and transition energies of carbon nanotubes. Phys. Rev. Lett. 93, 177401 (2004).
Scott, J., Leite, R. & Damen, T. Resonant Raman effect in semiconductors. Phys. Rev. 188, 1285 (1969).
Tanta, R. et al. Raman spectroscopy and electrical properties of InAs nanowires with local oxidation enabled by substrate micro-trenches and laser irradiation. Appl. Phys. Lett. 107, 243101 (2015).
Zardo, I. et al. E1(A) Electronic Band Gap in Wurtzite InAs Nanowires Studied by Resonant Raman Scattering. Nano Lett. 13, 3011–3016 (2013).
Chen, Z. et al. Lateral Fusion of Chemical Vapor Deposited N = 5 Armchair Graphene Nanoribbons. J. Am. Chem. Soc. 139, 9483–9486 (2017).
Malard, L. M., Pimenta, M. A., Dresselhaus, G. & Dresselhaus, M. S. Raman spectroscopy in graphene. Phys. Rep. 473, 51–87 (2009).
Efremov, E. V., Ariese, F. & Gooijer, C. Achievements in resonance Raman spectroscopy: Review of a technique with a distinct analytical chemistry potential. Anal. Chim. Acta. 606, 119–134 (2008).
Malard, L. M. et al. Resonance Raman study of polyynes encapsulated in single-wall carbon nanotubes. Phys. Rev. B 76, 233412 (2007).
Parker, F. S. Applications of infrared, Raman, and resonance Raman spectroscopy in biochemistry. (Springer Science & Business Media, 1983).
Mathur, R. B., Singh, B. P. & Pande, S. Carbon Nanomaterials: Synthesis, Structure, Properties and Applications. (CRC Press, 2016).
Ferrari, A. & Robertson, J. Resonant Raman spectroscopy of disordered, amorphous, and diamondlike carbon. Phys. Rev. B 64, 075414 (2001).
Ferrari, A. C. Raman spectroscopy of graphene and graphite: Disorder, electron–phonon coupling, doping and nonadiabatic effects. Solid State Commun. 143, 47–57 (2007).
Pei, S., Zhao, J., Du, J., Ren, W. & Cheng, H.-M. Direct reduction of graphene oxide films into highly conductive and flexible graphene films by hydrohalic acids. Carbon 48, 4466–4474 (2010).
Pei, S. & Cheng, H.-M. The reduction of graphene oxide. Carbon 50, 3210–3228 (2012).
Eigler, S. et al. Wet Chemical Synthesis of Graphene. Adv. Mater. 25, 3583–3587 (2013).
Overgaard, M. H. et al. Highly Conductive Semitransparent Graphene Circuits Screen‐Printed from Water‐Based Graphene Oxide Ink. Adv. Mater. Tech. 2, 1700011 (2017).
Sharpe, M. & Irish, D. Stray light in diffraction grating monochromators. Opt. Acta 25, 861–893 (1978).
Heller, E. J. et al. Theory of graphene raman scattering. ACS Nano 10, 2803–2818 (2016).
Cancado, L., Pimenta, M., Neves, B., Dantas, M. & Jorio, A. Influence of the atomic structure on the Raman spectra of graphite edges. Phys. Rev. Lett. 93, 247401 (2004).
Ferralis, N. Probing mechanical properties of graphene with Raman spectroscopy. Journal of materials science 45, 5135–5149 (2010).
Mohiuddin, T. et al. Uniaxial strain in graphene by Raman spectroscopy: G peak splitting, Grüneisen parameters, and sample orientation. Phys. Rev. B 79, 205433 (2009).
Pócsik, I., Hundhausen, M., Koós, M. & Ley, L. Origin of the D peak in the Raman spectrum of microcrystalline graphite. J. Non-Cryst. Solids 227–230, Part 2, 1083–1086 (1998).
Vidano, R., Fischbach, D., Willis, L. & Loehr, T. Observation of Raman band shifting with excitation wavelength for carbons and graphites. Solid State Commun. 39, 341–344 (1981).
Matthews, M., Pimenta, M., Dresselhaus, G., Dresselhaus, M. & Endo, M. Origin of dispersive effects of the Raman D band in carbon materials. Phys. Rev. B 59, R6585 (1999).
Cançado, L. G. et al. Quantifying defects in graphene via Raman spectroscopy at different excitation energies. Nano Lett. 11, 3190–3196 (2011).
King, A. A. et al. A New Raman Metric for the Characterisation of Graphene oxide and its Derivatives. Sci. Rep.-UK 6 (2016).
Cancado, L. et al. General equation for the determination of the crystallite size L a of nanographite by Raman spectroscopy. Appl. Phys. Lett. 88, 163106 (2006).
Claramunt, S. et al. The importance of interbands on the interpretation of the Raman spectrum of graphene oxide. J. Phys. Chem. C 119, 10123–10129 (2015).
Kaniyoor, A. & Ramaprabhu, S. A Raman spectroscopic investigation of graphite oxide derived graphene. Aip. Adv. 2, 032183 (2012).
Kudin, K. N. et al. Raman spectra of graphite oxide and functionalized graphene sheets. Nano Lett. 8, 36–41 (2008).
Pimenta, M. et al. Studying disorder in graphite-based systems by Raman spectroscopy. Phys. Chem. Chem. Phys. 9, 1276–1290 (2007).
Acknowledgements
We gratefully acknowledge financial support from the “Center for Synthetic Biology” at Copenhagen University funded by the UNIK research initiative of the Danish Ministry of Science, Technology and Innovation (Grant 09-065274), bioSYNergy, University of Copenhagen’s Excellence Programme for Interdisciplinary Research, the Villum Foundation (Project number VKR023115), the Carlsberg Foundation (CF14-0388), the Danish Council of Independent Research (Project number DFF-7014-00027).
Author information
Authors and Affiliations
Contributions
S.K. performed the experiment and analysis. S.K. and T.V. wrote the paper. M.H.O. fabricated the sample. T.V. supervised the analysis and experiment.
Corresponding author
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Krause, S., Overgaard, M.H. & Vosch, T. Photon Energy Dependent Micro-Raman Spectroscopy with a Continuum Laser Source. Sci Rep 8, 11621 (2018). https://doi.org/10.1038/s41598-018-29921-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-018-29921-6