Abstract
This pilot study uses a micro-optical coherence tomography (micro-OCT) system with ~1 μm axial resolution specifically to image the cornea and corneal scars in vivo. We used an established murine corneal scar model by irregular phototherapeutic keratectomy in ten C57BL/6 mice, with serial imaging using the micro-OCT and compared to anterior segment (AS-OCT) (RTvue, Optovue, Fremont, CA) before and after scar induction. Main outcome was agreement between the AS-OCT and micro-OCT using Bland-Altman plots (95% limits of agreement, LoA).We analysed 10 control eyes and 10 eyes with corneal scars and found that there was good agreement between AS-OCT and micro-OCT (P > 0.05) LOA: lower limit −14 µm (95% CI: −19 to −8.8 µm) upper limit 23 µm (95% CI: 18 to 28.5 µm) in terms of central corneal thickness. There was also good agreement between AS-OCT and micro-OCT in terms of corneal scar measurements (P > 0.5; correlation coefficient >0.99) LOA lower limit −2.1 µm (95% CI: −2.8 to −1.5 µm); upper limit 1.8 µm (95% CI: 1.1 to 2.4 µm). Our pilot study suggests that this novel in vivo micro-OCT imaging technique was able to measure central corneal thickness and scar thickness in agreement with current AS-OCT techniques.
Similar content being viewed by others
Introduction
Optical coherence tomography (OCT) uses low-coherence interferometry to enable non-contact, in vivo high-resolution imaging of the cornea1. However, most commercial anterior segment OCT (AS-OCT) systems are only able to achieve up to 5 μm axial resolution2 and thus, not able to differentiate layers within the cornea3. We have previously described a micro-OCT that uses broadband light to produce images of the cornea with up to 1–2 μm spatial resolution4,5,6. The micro-OCT was able to image corneal endothelial cells4,5 with a much higher resolution compared to a spectral domain OCT7.
Corneal scarring secondary to various insults may cause permanent visual impairment, where corneal transplantation may be the only way to restore vision8. As most corneal lesions or scars do not exist throughout the cornea, lamellar replacement of the diseased corneal layers9, may provide better tectonic outcomes10, or faster visual rehabilitation11,12,13. Thus, delineation of the cornea is important to allow for targeted treatment – while real-time AS-OCT guidance has now become possible in our surgical practice14. However, current AS-OCT systems have limited resolution, which are only able to detect gross changes to the cornea without detailed imaging of corneal layers, especially in the presence of corneal scars or intrastromal lesions.
Therefore, we performed a pilot study using a superluminescent diode array based micro-OCT system with ~1 μm axial resolution to detect diseased corneal layers, i.e. even in the presence of corneal scarring. Here, we compared this in vivo micro-OCT to a commercially available AS-OCT system in an established animal model to assess normal and scarred corneas.
Results
The analysis was done across all corneal samples (10 control and 10 injured). First, we compared normal corneal thickness measurements between established AS-OCT and micro-OCT to validate that the novel micro-OCT was comparable – Fig. 1. We found that the overall mean at baseline central corneal thickness (CCT) was 94.8 ± 4.0 µm using the AS-OCT and 91.4 ± 6.0 µm using the micro-OCT. There were no significant differences in mean CCT measured under AS-OCT and micro-OCT (Table 1). There was a good agreement between these two imaging techniques (P > 0.05) as the LOA were ranged from a lower limit of −14 µm (95% CI: −19 to −8.8 µm) to upper limit of 23 µm (95% CI: 18 to 28.5 µm) – Fig. 2.
Next, we compared the ability of the micro-OCT to measure the corneal scar thickness as compared to the AS-OCT. The mean scar thickness was 6.9 ± 3.3 µm measured by AS-OCT and 6.9 ± 3.0 µm by micro-OCT wherein no significant difference was observed between both the imaging modalities (Table 2). The central scar was developed to the greatest thickness at week 2 and subsequently reduced by week 4. There was a good agreement between these two imaging techniques (P > 0.5; correlation coefficient >0.99) with the LOA ranged from a lower limit of −2.1 µm (95% CI: −2.8 to −1.5 µm) to upper limit of 1.8 µm (95% CI: 1.1 to 2.4 µm) – Fig. 3.
The micro-OCT successfully identified the area of corneal scarring as confirmed by corresponding in vivo confocal microscopy imaging (Fig. 4) and histology (Fig. 5). Hematoxylin and eosin histochemistry in scarred corneas at three weeks confirmed corresponding scarring with increased inflammatory cell infiltration in the corneal stroma, and distinct epithelial disorganization. We also used the micro-OCT to identify the areas of active scarring, confirmed by the presence of myofibroblasts and fibroblasts using immunostaining for alpha smooth muscle actin (α-SMA) and fibronectin – Fig. 5. The scar detected by the micro-OCT corresponded to the areas of α-SMA staining and strong fibronectin expression compared to the control, normal corneas.
Discussion
Our study using a novel in vivo micro-OCT imaging technique suggested that the measurement of central corneal thickness and scar thickness were generally of good agreement with AS-OCT images, with at least 95% of measurements and differences within the 95% prediction intervals and 95% LOA, respectively. We found that the images acquired were able to discern the five basic corneal layers. When imaging the irr-PTK injured eyes, the scar tissue was also detected prominently. For example, the mean central corneal (80.2 ± 11.28 μm) and scar thickness (9.9 ± 2.28 μm) in mice cornea determined by the micro-OCT showed good agreement with immunofluorescent histology measurements of (82 ± 5.5 μm) CCT and (9.2 ± 3.5 μm) scar obtained at Week 2 follow-up. The resolution of distinction of the corneal layers in the processed micro-OCT acquired images, are comparable to the histology results, suggesting the potential for OCT to produce high resolution images of specific corneal layers non-invasively. When comparing the micro-OCT images with the AS-OCT imaging modality, the former performs better than AS-OCT in prominently displaying the scar distribution and intensity.
However, it is important to note the limitations of the in vivo micro-OCT used in this pilot study. One of the main problems encountered was the widespread presence of motion artefacts. The reason behind this could be attributed to the heart beat rate (310–840 beats per minute) of the mice used herein, which was much higher than the image acquisition speed of 60 frames per second. In view of this, future outlook for this area of study may involve improving the image acquisition speed. Moreover, the image acquisition for the micro-OCT is longer (estimated 1–2 minutes per image) compared to IVCM (15–30 seconds) and AS-OCT (5 seconds). Nonetheless, we obtained promising results, where this non-invasive imaging technique could produce images with quality comparable to that of ex vivo histology, and with more detail as compared to the current AS-OCT. Further improvements in the image processing technique and image acquisition speed, may allow better results for human eyes15,16.
The recent advances in OCT technology now allows, not only higher resolution imaging of the cornea17, but also additional information such as en face reconstruction for the cornea and ocular surface18, or even delineating vascular flow within the cornea19,20,21. High-resolution OCT is currently able to image corneal layers i.e. epithelium, Bowman’s layer, stroma, Descemet membrane, endothelium - and even the ocular surface tear film22. However, we have described a micro-OCT system that is able to provide more details similar to in vivo microscopy imaging, which may have a potential role in corneal pathologies that specifically require more information such as surgical planning in corneal dystrophies23, atypical corneal infections requiring microscopic detail of the pathogen24, or assessing the cornea after refractive surgery where high definition stromal layer assessment is required25,26.
In summary, despite the recognized limitations of this in vivo micro-OCT system, we provide promising early results from an animal study that suggests high resolution images with histology like detail are obtained in both normal and scarred corneas. Technical improvements such as developing a swept source micro-OCT system with higher A-line speed (up to 4 MHz)27, or introduction of a line scan camera with full-range OCT imaging28, may improve the image quality even further. If successful, the increased amount of detail in corneal structure analysis could have clinical applications in assessing disease within corneal layers or planning for selective corneal replacement. Further studies for human corneal imaging use in vivo would be required to establish this promising micro-OCT system in the future.
Methods
This study was approved by the Institutional Animal Care and Use Committee of Singapore Health Services and followed the guidelines in the use of Animals in Ophthalmic and Vision Research. We used a murine corneal scar model by irregular phototherapeutic keratectomy as previously described13. Ten C57BL/6 mice were anaesthetized by an intraperitoneal injection of ketamine hydrochloride (20 mg/kg body weight) and xylazine hydrochloride (2 mg/kg body weight), with only one eye treated under topical anesthesia (topical 1% lignocaine hydrochloride). Briefly, a standard 2 mm diameter central corneal wound was made using #64 Beaver blade and denuded of epithelium. Irregular PTK (irrPTK) was then performed to the ablation zone by firing 105 laser pulses (ablation depth ~10 μm) with an excimer laser. A fine mesh screen was positioned in the path of laser after firing 50% of the pulses performed with PTK. Discomfort was relieved by subcutaneous injection of Buprenorphine, 0.05 mg/kg body weight, twice daily for 1 day. Tobramycin ointment was applied twice daily for 2 days to prevent infection. Both injured and fellow control eyes were imaged under micro-OCT, slit lamp biomicroscopy, in vivo confocal microscopy and standard OCT (described in detail below) at day 3 before and 7, 14, 21, 28 days after scar induction. The mice were sacrificed at the 5th week.
Micro-Optical Coherence Tomography Imaging
The micro-OCT essentially uses light scattered elastically from within tissue in three dimensions and uses the electric field amplitude and the optical delay of light returned from the sample to measure the depth of the tissue. In this study, the micro-OCT system was modified from that described previously29, with significant improvements to axial and lateral resolution. The high-bandwidth (~165 nm centered at 830) and short coherence length light, is obtained by use of a superluminescent diode source (T-850, Superlum, Ireland), allowing axial resolution of 2.5 μm in air. The system includes an interferometer while the reference and sample arms intersect at the beam-splitter. A lateral resolution of 2.5 μm (10X objective lens) and 1.3 μm (20X objective lens) is achieved. Effective power incident on sample was kept less than 2 mW. The galvanometer scanning motors were controlled using customizable software, in correspondence with the spectral-data from a scan camera (E2V, AViiVA EM4). The line rate (1024 per frame), frame rate (60 per second) and the scan geometry (3.5 mm × 3.5 mm) were used. Three dimensional images (3.5 mm × 3.5 mm × 1 mm, W × H × D) were obtained with an acquisition time of less than 3 minutes per 3D image.
Optical Coherence Tomography Image Enhancement and Data collection
Current image quality from standard OCT is greatly hampered by the presence of shadow artifacts and poor tissue visibility in the deeper layers - due to signal attenuation, whereby signal strength diminishes as a function of tissue depth. Moreover, the animal model for this experiment, mice, have a very high heart rate of 310–840 beats per minute, which being much higher than the frame rate results in noise and distortion in the captured image. Filters were applied to remove the background noise and edge distortions from the images using MATLAB R2017a (The MathWorks, Inc., Natick, Massachusetts, United States). Further, the captured micro-OCT images were improved using compensation algorithms to compensate for light attenuation and contrast enhancement with post-processing software15, used to remove the shadow artefacts from dense structures and improve the visibility of the deep tissue layers, as previously described – Fig. 116. The post-processed images were then used to obtain information on the corneal, epithelial and the scar thickness (in case of the injured samples). All measurements were done using ImageJ and each measurement was computed as the average of 10 measurements taken across the sample.
Anterior segment evaluation and histology
We performed serial slit-lamp microscopy (FS-3V Zoom Photo Slit Lamp, Nikon, Tokyo, Japan) and AS-OCT (RTvue, Optovue, Fremont, CA) to compare corneal measurements between the micro-OCT and AS-OCT. In vivo confocal microscopy (Heidelberg Engineering GmbH, Heidelberg, Germany) was also performed for corneal scars in vivo. After sacrifice, the murine corneas were extracted and fixed in 10% formaldehyde (Sigma Aldrich, St. Louis, MO, USA) for 60 min and PBS washes. They were dehydrated and processed for paraffin embedding, with sections (5 μm thick), haematoxylin and eosin (H&E) staining - using Mayer’s hematoxylin (Sigma Aldrich, USA), then differentiated with acid alcohol, extensively washed before stain with Eosin solution (Sigma Aldrich, USA) for another 2 minutes, washed, and dehydrated. The section was mounted with Permount reagent (Fisher) and viewed under light microscopy (Nikon C2 confocal microscope). Immunofluorescence staining for scar tissue markers was performed using Mouse On Mouse (MOM) Immunodetection kit (Vector Laboratories Inc., Burlingame, CA, US) with mouse monoclonal antibody against alpha smooth muscle actin (α-SMA) (DAKO M0851; 1:200 dilution) and mouse monoclonal antibody against fibronectin (FN) (Millipore MAB1940; 1:200 dilution; Temecula CA, US). After staining, the samples were mounted with Vectashield containing DAPI (Vector Laboratories) and viewed under fluorescence microscopy (ZEISS Axio Observer.Z1, Oberkochen, Germany).
Statistical analysis
Mean differences in corneal and scar measurements (mean ± standard deviation, SD) between the two machines were assessed using a paired t-test, after tests for normality. Agreements between the AS-OCT and µ-OCT machines were also described using the Bland-Altman, with 95% limits of agreement (LoA). Bland Altman analysis was performed with MedCalc Version 17.1. Intra-Class Correlation Coefficients were computed for inter-machine reliability of data. All other statistical tests were performed with IBM Corp. Released 2011 IBM SPSS Statistics for Windows, Version 20.0. Armonk, NY: IBM Corp.; Statistical significance was considered p < 0.05.
Data availability
The data that support the findings of this study are available from the corresponding author.
References
Ang, M. et al. Anterior segment optical coherence tomography study of the cornea and anterior segment in adult ethnic South Asian Indian eyes. Invest Ophthalmol Vis Sci 53, 120–125 (2012).
Ang, M. et al. Anterior segment optical coherence tomography. Prog Retin Eye Res. pii: S1350-9462(17)30085-X, https://doi.org/10.1016/j.preteyeres.2018.04.002. [Epub ahead of print] Review. PMID:29635068 (2018).
Girard, M. J., Strouthidis, N. G., Ethier, C. R. & Mari, J. M. Shadow removal and contrast enhancement in optical coherence tomography images of the human optic nerve head. Invest Ophthalmol Vis Sci 52, 7738–7748 (2011).
Liu, L. et al. Method for quantitative study of airway functional microanatomy using micro-optical coherence tomography. Plos One 8, e54473 (2013).
Liu, L. et al. Imaging the subcellular structure of human coronary atherosclerosis using micro-optical coherence tomography. Nature Medicine 17, 1010–1014 (2011).
Liu, L. et al. An autoregulatory mechanism governing mucociliary transport is sensitive to mucus load. American journal of respiratory cell and molecular biology 51, 485–493 (2014).
Liu, X., Chen, S., Cui, D., Yu, X. & Liu, L. Spectral estimation optical coherence tomography for axial super-resolution. Optics Express 23, 26521–26532 (2015).
Tan, D. T., Dart, J. K., Holland, E. J. & Kinoshita, S. Corneal transplantation. Lancet 379, 1749–1761 (2012).
Ang, M., Mehta, J. S., Arundhati, A. & Tan, D. T. Anterior lamellar keratoplasty over penetrating keratoplasty for optical, therapeutic, and tectonic indications: a case series. Am J Ophthalmol 147, 697–702 e692 (2009).
Ang, M., Mehta, J. S., Sng, C. C., Htoon, H. M. & Tan, D. T. Indications, outcomes, and risk factors for failure in tectonic keratoplasty. Ophthalmology 119, 1311–1319 (2012).
Tan, D., Ang, M., Arundhati, A. & Khor, W. B. Development of Selective Lamellar Keratoplasty within an Asian Corneal Transplant Program: The Singapore Corneal Transplant Study (An American Ophthalmological Society Thesis). Trans Am Ophthalmol Soc 113, T10 (2015).
Ang, M. & Mehta, J. S. Deep anterior lamellar keratoplasty as an alternative to penetrating keratoplasty. Ophthalmology 118, 2306–2307, author reply 2307 (2011).
Ang, M., Mohamed-Noriega, K., Mehta, J. S. & Tan, D. Deep anterior lamellar keratoplasty: surgical techniques, challenges, and management of intraoperative complications. Int Ophthalmol Clin 53, 47–58 (2013).
De Benito-Llopis, L., Mehta, J. S., Angunawela, R. I., Ang, M. & Tan, D. T. Intraoperative anterior segment optical coherence tomography: a novel assessment tool during deep anterior lamellar keratoplasty. Am J Ophthalmol 157, 334–341 e333 (2014).
Girard, M. J. et al. Enhancement of Corneal Visibility in Optical Coherence Tomography Images Using Corneal Adaptive Compensation. Transl Vis Sci Technol 4, 3 (2015).
Chung, C. W. et al. Enhancement of Corneal Visibility in Optical Coherence Tomography Images with Corneal Opacification. Transl Vis Sci Technol 5, 3 (2016).
Ang, M. et al. Optical coherence tomography angiography: a review of current and future clinical applications. Graefes Arch Clin Exp Ophthalmol 256, 237–245 (2018).
Ang, M. et al. En face optical coherence tomography angiography for corneal neovascularisation. Br J Ophthalmol 100, 616–621 (2016).
Ang, M. et al. Optical Coherence Tomography Angiography for Anterior Segment Vasculature Imaging. Ophthalmology 122, 1740–1747 (2015).
Ang, M. et al. Optical coherence tomography angiography and indocyanine green angiography for corneal vascularisation. Br J Ophthalmol 100, 1557–1563 (2016).
Cai, Y., A. D. Barrio, J. L., Wilkins, M. R. & Ang, M. Serial optical coherence tomography angiography for corneal vascularization. Graefes Arch Clin Exp Ophthalmol 255, 135–139 (2017).
Moutsouris, K. et al. Optical coherence tomography, Scheimpflug imaging, and slit-lamp biomicroscopy in the early detection of graft detachment after Descemet membrane endothelial keratoplasty. Cornea 30, 1369–1375 (2011).
Siebelmann, S. et al. Anterior segment optical coherence tomography for the diagnosis of corneal dystrophies according to the IC3D classification. Surv Ophthalmol (2017).
Ang, M., Mehta, J. S., Mantoo, S. & Tan, D. Deep anterior lamellar keratoplasty to treat microsporidial stromal keratitis. Cornea 28, 832–835 (2009).
Rosas Salaroli, C. H., Li, Y. & Huang, D. High-resolution optical coherence tomography visualization of LASIK flap displacement. J Cataract Refract Surg 35, 1640–1642 (2009).
Ang, M. et al. Refractive lenticule extraction: transition and comparison of 3 surgical techniques. J Cataract Refract Surg 40, 1415–1424 (2014).
Fechtig, D. J., Schmoll, T., Grajciar, B., Drexler, W. & Leitgeb, R. A. Line-field parallel swept source interferometric imaging at up to 1 MHz. Opt. Lett. 39, 5333–5336 (2014).
Sarunic, M., Choma, M. A., Yang, C. & Izatt, J. A. Instantaneous complex conjugate resolved spectral domain and swept-source OCT using 3 × 3 fiber couplers. Optics Express 13, 957–967 (2005).
Ang, M. et al. Evaluation of a Micro-Optical Coherence Tomography for the Corneal Endothelium in an Animal Model. Sci Rep 6, 29769 (2016).
Acknowledgements
Singapore National Medical Research Council (NMRC/CNIG/1128/2015), and National Research Foundation Singapore (NRF-CRP13-2014-05).
Author information
Authors and Affiliations
Contributions
All authors (1) made substantial contributions to conception and design, acquisition of data, or analysis and interpretation of data; (2) drafting the article or revising it critically for important intellectual content; and (3) final approval of the version to be published.
Corresponding author
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ang, M., Devarajan, K., Das, S. et al. Novel application of In Vivo Micro-Optical Coherence Tomography to assess Cornea scarring in an Animal Model. Sci Rep 8, 11483 (2018). https://doi.org/10.1038/s41598-018-29761-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-018-29761-4
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.