Abstract
Null functions of an imaging system are functions in the object space that give exactly zero data. Hence, they represent the intrinsic limitations of the imaging system. Null functions exist in all digital imaging systems, because these systems map continuous objects to discrete data. However, the emergence of detectors that measure continuous data, e.g. particle-processing (PP) detectors, has the potential to eliminate null functions. PP detectors process signals produced by each particle and estimate particle attributes, which include two position coordinates and three components of momentum, as continuous variables. We consider Charged-Particle Emission Tomography (CPET), which relies on data collected by a PP detector to reconstruct the 3D distribution of a radioisotope that emits alpha or beta particles, and show empirically that the null functions are significantly reduced for alpha particles if ≥3 attributes are measured or for beta particles with five attributes measured.
Similar content being viewed by others
Introduction
The performance of an imaging system is determined by the many physical processes involved. Imaging is ultimately about transferring information about the object to the detector. The information is often carried by photons and charged particles. The physical processes involved in photon and particle transport include emission, absorption, scattering, propagation and detection1. Each process affects the performance of an imaging system in a complicated way, but these effects in combination can be characterized by null functions.
Null functions represent the intrinsic limitation of a linear imaging system. A null function [ref.2, p. 34–38] is any non-zero function in the object space that produces no data in the data space. In other words, null functions are invisible to an imaging system because, if added to an object, any difference in the collected data is only due to random noise in the imaging system. For any linear system, an object can be decomposed into two components, the measurement component and the null component. The measurement component of an object provides a perfect fit to the noise-free data. The null component of an object, which is also referred to as a null function, makes no contribution to the data and cannot be recovered even from noise-free data.
Null functions exist in all digital imaging systems3,4,5,6,7,8,9, because real-world objects are functions of continuous variables while a digital image consists of an array of pixel values or other forms of histograms. The mapping from continuous objects to discrete data, which can be represented by a continuous-to-discrete (CD) operator, causes an infinite number of objects to produce exactly the same data. Therefore, null functions, which are the differences of those objects, are inevitable for a CD system.
Recent work9 has shown that null functions can be reduced if continuous data are collected instead of discrete data for Single Photon Emission Computed Tomography (SPECT). The continuous data are collected with photon-processing detectors10,11,12,13,14,15,16 that apply a maximum-likelihood11,13,17,18,19,20 method to estimate the interaction position, deposited energy and other attributes of each photon-interaction event.
Our previous work21 extended the concept of photon-processing detectors to particle-processing detectors that detect alpha and beta particles. Two main applications of alpha and beta particles are studying the metabolisms of pharmaceuticals and targeted radionuclide therapy for cancer. In both applications, the cellular-level distribution of the charged-particle-emitting radionuclide is valuable information that can be acquired through imaging. The conventional imaging technique for charged particles is autoradiography22,23,24,25,26, which provides two-dimensional, ex vivo images of thin tissue slices. In that paper21, we presented a direct three-dimensional autoradiography technique, which was named as charged-particle emission tomography (CPET). CPET was enabled by particle-processing detectors (PPDs).
A particle-processing detector (PPD) detects individual particles, estimates information about each particle, and stores the estimated particle parameters as continuous variables in a list14,15,16,18,27. On a detector plane, the interaction of a charged particle with the detector can be described by up to five parameters including two variables describing the interaction position, the deposited energy and two direction cosines describing the propagation direction. We refer to each parameter that describes the interaction event as an attribute of the detected particle. Upon interacting with a detector, each charged particle forms a track and generates secondaries along the track. The secondaries, which are photons for a scintillation detector and electron-hole pairs if a semiconductor detector is used, may contain information about a subset of the five particle attributes. From the raw signals produced by the secondaries, particle attributes can be estimated rigorously with maximum-likelihood estimation11,13,17,18,19,20. A few examples of PPDs were introduced in the main text and appendix of Ding, et al.21.
In this paper, with the goal of understanding the benefits of using a PPD, we calculate the null functions of CPET when different combinations of attributes are measured. Starting from Monte Carlo simulations or theoretical calculations of charged-particle transport, we use the Singular Value Decomposition (SVD) to calculate the null functions corresponding to a test object. Our results show that the null functions are significantly reduced for alpha particles with q ≥ 3 or for beta particles with q = 5, where q is the number of attributes estimated for each particle.
This paper is organized as follows. Section 1 introduces the setup and physics of CPET. Section 2 presents a mathematical model of the imaging system, which serves as a foundation for Section 3 and 4. Section 3 introduces the mathematical concept of null function. Section 4 presents detailed methods of this study. The results are discussed in Section 5. Section 6 provides a summary and directions for future work.
Setup and physics of Charged-Particle Emission Tomography
Radioisotopes that emit charged particles, such as alpha and beta particles, are widely used to label molecules or cells of interest in pharmacology23,24,25,26 and targeted radiation therapy28,29,30,31. Imaging of tissue samples containing such radioisotopes provides information about the distribution of a pharmaceutical or radiation dose. In the conventional charged-particle imaging technique, which is autoradiography, if cellular-level resolution is desired, one has to slice the tissue samples to 4- to 10-μm thick. This leads to a large number of tissue samples needed to be imaged and limits autoradiography to ex vivo imaging.
A direct three-dimensional imaging technique for radioisotopes that emit alpha or beta particles, Charged-Particle Emission Tomography (CPET), has been presented before21,32,33. CPET would allow imaging of thick tissue sections with high spatial resolution, which reduces the number of tissue samples needed for imaging. CPET also has the potential to be used for in vivo imaging. CPET is enabled by particle-processing detectors (PPDs)21. A PPD detects single particles and measures a subset of particle attributes, such as position, direction, and energy, for each detected particle.
In CPET21, the objects imaged are molecules or cells that have been labeled with charged-particle-emitting radioisotopes in tissue; a planar PPD is placed in contact with or in close proximity to one side of a sample tissue, as illustrated in Fig. 1. Beta Emission Tomography (BET) and Alpha Emission Tomography (αET, pronounced as AlphET) are two special cases of CPET. In each case, the detected-particle attributes are fed into an algorithm to reconstruct the 3D distribution of the radioisotopes. There is no motion of the detector or the object, making CPET a case of extreme limited-angle tomography34.
BET is an imaging technique for fast electrons, which include beta particles, conversion electrons and Auger electrons. Beta particles have broad, continuous energy spectra, while conversion electrons and Auger electrons usually have energy spectra with several sharp peaks. While traversing media, fast electrons deflect easily due to their low mass, and therefore fast electrons tend to have tortuous paths35. Figure 1(a) shows the simulated 3D beta particle tracks projected on a 2D plane, where the particles are emitted from a 18F-point source.
αET produces images of radioisotopes that emit alpha particles21,36. Alpha particles have discrete energy spectra with highly monoenergetic emission lines associated with particular nuclear transitions. In low-atomic-number materials such as water or tissue, alpha particles travel in nearly straight lines because Rutherford scattering rarely happens35. As alpha particles travel through matter, they interact with the molecules in the material through Coulomb interactions and lose energy continuously. The energy loss of an alpha particle in a material is a known function of its path length. At the end of its path, an alpha particle loses all of its energy and is stopped in the material. Figure 1(b) shows the simulated 3D alpha particle tracks projected on a 2D plane, where the particles originate from a 239Pu-point source.
Due to the short range of alpha particles and significant scattering of beta particles in matter, CPET is designed differently from conventional emission tomography techniques such as Positron Emission Tomography (PET) and Single-Photon Emission Computed Tomography (SPECT), both of which collect 2D projections of the 3D object from multiple angles37. In contrast, CPET achieves tomographic reconstruction from a single projection data. Another fundamental difference between CPET and the conventional emission tomography (PET and SPECT) is that charged particles cannot penetrate the body of a patient unless the source is close to the skin. Potential applications of CPET in medicine include 3D imaging of superficial lesions, endoscopic tomography, surgical probes and 3D molecular imaging of a small animal with a window chamber.
Mathematical Model of CPET
Rigorous analysis of CPET requires a mathematical model, which we derive step by step in this section. For an example on how to apply this mathematical model, see Section 4. In CPET, the object is the spatial density of radioactive decays per unit time. We denote the object as a scalar-valued function f (R), where R = (x, y, z) is a point within the object. We assume that the object is square-integrable and supported within a region S f in \({{\mathbb{R}}}^{3}\). Therefore, the object function f(R), which is also referred to as f in the following discussion, is a vector in the Hilbert space \({{\mathbb{L}}}_{2}({{\bf{S}}}_{f})\).
The charged particles produced by the object propagate through tissue, and some of them enter the planar detector and produce signals. Each detected particle can be characterized by an attribute vector (r d , s, E), where r d = (x d , y d ) is the 2D position of the particle at the detector plane, s = (s x , s y ) is the direction cosines representing the propagation direction, and E is the residual energy. A statistical ensemble of particles is described by a phase-space distribution function, which is referred to as the radiance in the imaging literature1,2. The radiance of the charged-particles at the entrance plane of the detector is related to the object through
where \({\mathscr{P}}\) is a propagation operator describing the interaction between the charged particles and tissue. The kernel of \({\mathscr{P}}\) is pr (r d , s, E|R)S(R), where pr(r d , s, E|R) is the probability density of a detected particle with true attribute vector (r d , s, E) given that the event originated with an emission at R; and S(R) is the probability that an emission at R will be detected and result in a list entry. The function S(R) is often referred to as the sensitivity function. Operator \({\mathscr{P}}\), which includes the physical processes of scattering and energy loss, can be derived analytically or approximated by Monte Carlo simulation.
A PPD collects signals from each detected particle and estimates a subset of the five attributes (x d , y d , s x , s y , E) to some uncertainty. The estimated attribute vector, which is a vector in the q-D attribute space (q = 2, 3, 4, 5), is denoted as \(\hat{{\bf{A}}}\). If a total of J particles are detected, the output of a PPD is a list of J attribute vectors. We will denote this list as \({\mathscr{A}}=\{{\hat{{\bf{A}}}}_{1},{\hat{{\bf{A}}}}_{2},\ldots ,{\hat{{\bf{A}}}}_{J}\}\), where \(\hat{{\bf{A}}}{j}\) is the estimated attribute vector for the j th detected event. Because the radioactive decays are independent events, the following Poisson point process [ref.2, p. 649–669] can be used to describe the list-mode data15,27,38,39
where δ(...) denotes the q-D Dirac delta function. For a given object and a fixed exposure time τ, the mean of \(u(\hat{{\bf{A}}})\) over an ensemble of attribute lists (more specifically, over the number of events J and each attribute vector \({\hat{{\bf{A}}}}_{j}\)) is27,32
where \({\rm{prf}}(\hat{{\bf{A}}}|{\bf{R}})\) is the point response function with the point source located at R.
By measuring the attribute vector for each detected particle, the detector maps the phase-space distribution function L 0(r d , s, E) to data \(\bar{u}(\hat{{\bf{A}}}|{\bf{f}})\). This detection process can be described by a measurement operator \( {\mathcal M} \) as
The imaging system can be described by combining Equation (1) and Equation (4):
where \( {\mathcal L} = {\mathcal M} {\mathscr{P}}\) is the system operator, and \({\bf{u}}=\bar{u}(\hat{{\bf{A}}}|{\bf{f}})\) is a vector in the Hilbert space \({{\mathbb{L}}}_{2}({{\mathbb{R}}}^{q})\). Thus, \( {\mathcal L} \) is a linear operator that maps a vector f in \({{\mathbb{L}}}_{2}({{\bf{S}}}_{f})\) (object space) to a vector u in \({{\mathbb{L}}}_{2}({{\mathbb{R}}}^{q})\) (image space).
As shown in Equation (3), the kernel function of \( {\mathcal L} \) is \({\rm{prf}}(\hat{{\bf{A}}}|{\bf{R}})\). This kernel function is given by
where pr(A|R)S(R) is obtained by marginalizing pr(r d , s,E|R)S(R) over the particle attributes that are not estimated by the detector; and pr\((\hat{{\bf{A}}}|{\bf{A}})\) is a probability density function of the estimates \(\hat{{\bf{A}}}\) when the true underlying attribute is A. The estimation uncertainty originates from noise in the detector.
SVD and null functions
Null function analysis is based on Singular Value Decomposition (SVD). In this section, we introduce SVD and show how to use it to calculate null functions. In addition, we discuss the symmetries of CPET and how to apply symmetries to simplify the SVD analysis.
Singular Value Decomposition
SVD analysis of a linear system involves the system operator and its adjoint operator [ref.2, p. 17]. The adjoint operator of \( {\mathcal L} \), denoted as \({ {\mathcal L} }^{\dagger }\), maps a function in the attribute space to a function in the object space. The adjoint operator is defined by
where 〈.,.〉 is the inner product in Hilbert space \({{\mathbb{L}}}_{2}({{\bf{S}}}_{f})\) or \({{\mathbb{L}}}_{2}({{\mathbb{R}}}^{q})\). The adjoint operator \( {\mathcal L} \) † acting on a point process \(u(A)={\sum }_{j\mathrm{=1}}^{J}\delta (\hat{{\bf{A}}}-{\hat{{\bf{A}}}}_{j})\) is a function of R and it is given by
SVD is performed by solving the eigenequation generated by two operators \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\) and \(\, {\mathcal L} {\mathcal L} {}^{\dagger }\). The operators \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\) and \(\, {\mathcal L} {\mathcal L} {}^{\dagger }\) are Hermitian and non-negative definite, which means the operators have real and non-negative eigenvalues. Furthermore, the two operators have the same eigenvalue spectra2. The operator \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\) maps a function from the object space back to the object space; while \(\, {\mathcal L} {\mathcal L} {}^{\dagger }\) maps a function from image space back to the image space. In tomography literature, \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\) is often referred to as the projection/backprojection operator. The eigenfunctions of \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\) and \(\, {\mathcal L} {\mathcal L} {}^{\dagger }\) are in the object space and the image space, respectively.
We will denote the eigenfunction of \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\) as w n (x, y, z), the eigenfunction of \(\, {\mathcal L} {\mathcal L} {}^{\dagger }\) as v n (A), and the corresponding eigenvalue as μ n , where n can be an integer, a continuous variable, or a combination of both. When an eigenfunction in the object space is propagated through the imaging system, the resulted function is the corresponding eigenfunction in image space. This can be written in abstract form as
where \({{\bf{w}}}_{n}={w}_{n}(x,y,z)\) is the n th eigenfunction in object space, \({{\bf{v}}}_{n}={{\bf{v}}}_{n}(\hat{{\bf{A}}})\) is the n th eigenfunction in image space, and μ n is the n th eigenvalue of both \(\, {\mathcal L} {\mathcal L} {}^{\dagger }\) and \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\). The value \(\sqrt{{\mu }_{n}}\) is referred to as the n th singular value of \( {\mathcal L} \).
Calculation of null functions
For any linear system, an object can be decomposed into two orthogonal components, the measurement component and the null component. In this section, we introduce how to calculate null functions, which are represented by the null components of any objects2,9.
When n is discrete, we can use the eigenfunction in object space, w n , and the eigenvalues to calculate the measurement component of any object function f as
where R is the rank2 of \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\) and the summation is over n. If n is a continuous variable, the summation turns into an integration.
The null component, which does not contribute to the data, can be calculated by
The measurement component and null component lie in different subspaces of object space [2, p~34–44]. One can prove that the measurement component and the null component of the same object are orthogonal. In mathematical terms, the inner product of the two components is
Singular-value decomposition of objects into their measurement and null components is a powerful technique to investigate the intrinsic limitations of imaging systems4,5,6,7,8,9. SVD analysis of an imaging system that maps a function in 3D object space to q-D data space can be complex due to the high dimensions. However, this complexity is reduced by the symmetry of CPET.
Symmetries
The symmetry of CPET is contained in the propagation operator \({\mathscr{P}}\). The propagation operator has translational and rotational symmetries under two conditions: (1) the detector is large compared to the size of the object and the range of the charged particles, and (2) the medium where radioactive sources are located is uniform. Under the assumption that the measurement operator \( {\mathcal M} \) does not break these symmetries, the system operator \( {\mathcal L} \) has translational and rotational symmetries, which can be used to simplify the characterization of the imaging system.
In CPET, the kernel function of the projection/backprojection operator, \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\), is
where the point response function prf(A|R) has been defined in Equation (6).
As discussed in Section 7.2.10 in Barrett and Myers (2004)2, the translational symmetry on the xy-plane simplifies the kernel of \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\):
where r = (x, y), R = (x, y, z) and z max is either the thickness of the tissue sample or the maximum-tissue thickness a particle can penetrate. Due to the lateral-shift invariance, the 2D-Fourier-basis functions, exp (−2πi ρ ⋅ r), are eigenfunctions associated with the transverse direction, and the 2D spatial frequency ρ serve as an index for the eigenvalue equations.
The eigenvalue equation of \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\) turns into
where K(ρ, z, z′) is the 2D-Fourier transform of k(r − r′, z, z′) with respect to vector r − r′; W j (ρ, z)exp(2πi ρ ⋅ r) is an eigenfunction with indices ρ and j; μ ρ,j is the corresponding eigenvalue; ρ is a continuous index and associated with the transverse directions; the index j is connected to the longitudinal direction.
Let us consider a fixed ρ for now. The function K(ρ, z, z′), which can also be denoted as K ρ (z, z′), is considered as a function of two variables z and z′. For a fixed ρ, Equation (15) is an eigenvalue equation. The compactness of the linear operator defined by Equation (15) determines if the index j is discrete or continuous. An operator is compact if its kernel satisfies the Hilbert-Schmidt condition2, which in the case of Equation (15) assumes the form
When K(ρ, z, z′) is bounded and z max is finite, the operator defined by the integration has a discrete spectrum and the index j is discrete. As a side note, the Hilbert-Schmidt condition is satisfied only in the z direction, but not on the xy plane. All CPET setups discussed in the following section have bounded K(ρ, z, z′) for a fixed ρ, except for αET with a detector that measures 5D data (x d , y d , s x , s y , E).
In numerical calculations, the 1D integration described in Equation (15) can be approximated with a matrix multiplication. If N different depths z are sampled when K(ρ, z, z′) is calculated, Equation (15) turns into an eigenvalue problem for an N × N matrix at each ρ.
If a system also has rotational symmetry with respect to z-axis, the kernel K(ρ, z, z′) reduces to K(ρ, z, z′), where ρ is the magnitude of the 2D spatial frequency ρ. As a result, for two frequencies, ρ 1 and ρ 2, the 1D eigen-analysis problem described in Equation (15) is the same if ρ 1 = ρ 2.
Methods
We use SVD analysis to compare 2D-, 3D-, 4D- and 5D-detectors for both BET and αET. The number of dimensions simply refers to the number of attributes measured by the detector. More specifically, the four combinations of attributes to be measured are:
-
(x d , y d ),
-
(x d , y d, E),
-
(x d , y d, s x , s y ),
-
(x d , y d, s x , s y , E),
where (x d , y d , s x , s y , E) have been introduced in Section 2. For each setup, K(ρ, z, z′) is calculated and an eigenvalue problem is solved at each ρ.
The function describing the interaction between charged particles and tissue, pr(A|R)S(R), is approximated with Monte Carlo simulation results for beta particles and solved analytically for alpha particles.
The estimation error of the detector, which is described by pr(Â|A), can be modeled asymptotically as a multivariate Gaussian distribution function with the inverse covariance matrix equals to Fisher Information Matrix40,41, if Maximum-Likelihood (ML) estimation is used13,17,18,19,41. The asymptotic properties of ML estimation are satisfied in the limit of a large number of secondaries (photoelectrons for scintillation detectors or electron-hole pairs for semiconductor detectors). However, in the present treatment, in order to explore the upper limit on how well the imaging technique can perform, we effectively “turn off” the uncertainties due to the estimation process in the detector. Mathematically speaking, we assume \(\mathrm{pr}(\hat{{\bf{A}}}|{\bf{A}})=\delta (\hat{{\bf{A}}}-{\bf{A}})\).
Monte Carlo simulations for BET systems
In this section, we consider an 18F source, because it is widely used in Positron Emission Tomography. Two other sources, an 131I source and a 400-keV-monoenergetic electron source, are considered in Appendix B.
We use the Monte Carlo simulation toolkit Geant442,43 to simulate the passage of beta particles through tissue. Even though Monte Carlo simulation has intrinsic randomness, the uncertainties can be reduced by simulating a large number of particles.
For each type of detector, a point source is simulated at depth z in a layer of 100-μm tissue (approximated by water), where z belongs to {5, 10, ..., 100} μm. For each depth, the emission and propagation of 107 beta particles are simulated. A detector of size 1024 × 1024 μm 2 is placed on the top surface of the tissue layer and measures q attributes from each particle that enters the detector. The output of each simulation is a list of particle attributes. To estimate \(\mathrm{pr}(\hat{{\bf{A}}}|{\bf{R}})\), we bin the list-mode data into histograms. The dimensions of the bins are defined as: Δx = Δy = 4 μm, Δθ = 2°, Δϕ = 2° and ΔE = 20 keV. The bin size is determined so that finer sampling rate does not produce significantly different results.
Theoretical calculations for αET systems
We approximate the tracks of alpha particles in tissue as straight lines and derive the point spread function of αET analytically. The derivations are provided in Appendix C. We consider a 239Pu source, which emits alpha particles with energies 5.15 MeV (73 %), 5.14 MeV (15 %) and 5.11 MeV (10 %). We use a monoenergetic source that emits 5.15-MeV alpha particles as an approximation. The range of a 5.15-MeV-alpha particle in tissue (approximated by water) is 39.4 μm.
A detector of size 160 × 160 μm 2 is placed at the top surface of the tissue layer and measures q attributes from each particle that enters the detector from the tissue. A detailed discussion of αET can be found in Ding et al.21. For each type of detector, a point source is considered at depth z in a layer of 40- μm tissue, where z can take any value in {1, 3,..., 39} μm. When q takes values in {2, 3, 4}, K(ρ, z, z′) is evaluated numerically for (z, z′) on a 20 × 20 grid. When a 5D detector is used (q = 5), the eigenvalue spectrum of the system can be calculated analytically. However, in order to compare to the numerical results for q = {2, 3, 4}, we consider the 5D αET on the same discrete depth z used for the other detectors, z = {1, 3, ...39} μm.
Results
In order to explore the intrinsic limitations in CPET, we calculate measurement and null components of a test object. We design an object f that contains several cylinders uniformly filled with radioactive substances. The size of the object scales to the tissue thickness of the system being studied (100 μm for BET and 40 μm for αET). An illustration of the object, which shows the relative size, is provided in Fig. 2. For actual size of the objects in BET and αET, see the cross-section views shown in column one of Figs 3 and 4, respectively.
The eigenvalue spectra and the eigenfunctions of \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\) are calculated for each scenario. The eigenvalues of \(\,{ {\mathcal L} }^{\dagger }\, {\mathcal L} \,\) are scaled to the maximum eigenvalue and presented in the fourth column of Figs 3 and 4 for BET and αET, respectively. In the plots, the values λ n (ρ) = μ n (ρ)/μ max are plotted as functions of spatial frequency ρ, where μ max is the maximum eigenvalue. At each frequency, 20 eigenvalues are plotted with different colors.
The measurement component f meas is calculated according to Equation 10, where the summation is for all j and ρ where λ j,ρ > λ min . The threshold λ min is chosen so that the eigenfunctions contributed to the measurement component are eigenfunctions with eigenvalues significantly above zero. As shown in the eigenvalue plots, λ j,ρ are noisy (due to the numerical errors in the calculation of K(ρ, z, z′)) when λ j,ρ < 10−6. Therefore, we set λ min to 10−6. The results are shown for BET and αET in Figs 3 and 4, respectively. Both figures present the object in the first column, the measurement component of the object in the second column, the null component in the third column. The four types of detectors that measure four different combinations of data are shown in rows one to four, respectively. All plots are xz-projection views of the 3D volume.
BET
As shown from Row 1 to Row 4 in Fig. 3, when the number of measured attributes q increases from 2 to 5, the null component of the object reduces and the measurement component approaches the true object gradually. When all five attributes, (x d , y d , s x , s y , E), are measured from each detected particle, the null space of BET is almost completely eliminated. This means, despite the broad spectra of beta decays, it is necessary to measure the residual energy of the particles to achieve good depth resolution. For BET21, a preferred detector should have the ability to measure position, direction and energy of each detected particle.
αET
The figures in the third column of Fig. 4 show that the null functions of αET essentially disappear when the detector measures any of the following combinations of attributes (1) position and energy, (2) position and direction, and (3) position, direction and energy.
Norm of the null functions
In order to quantitatively characterize the effect of measuring additional attributes on null functions, we calculate the scaled \({{\mathbb{L}}}_{2}\) norm of null functions. The \({{\mathbb{L}}}_{2}\) norm of f is defined as [ref.2, p.4]
We define the scaled norm of null functions as
where the subscript s represents scaled, and f null is the null component of the object f.
The scaled norm of null functions as a function of the number of attributes measured by the detector is plotted for both BET and αET in Fig. 5. The plots show that as the number of attributes (q) measured by the detector increases, the scaled norm of the null functions decreases. The scaled norm of the null function goes to zero for αET with q ≤ 3 and for BET with q = 5.
Conclusions and future work
We have demonstrated the potential of particle-processing (PP) detectors in reducing null functions of Charged-Particle Emission Tomography (CPET). By collecting data with a PP detector, CPET is able to reconstruct the 3D distribution of a radioactive tracer that emits alpha or beta particles. Furthermore, this tomographic reconstruction is based on single-projection data. CPET is enabled by PP detectors that detect single particles and measure direction and energy attributes in addition to position for each particle.
Null functions are calculated for CPET for four different combinations of particle attributes. Our results show that, as the number of attributes measured increases, the null functions reduce for both alpha and beta particles. The direction attributes and the energy attribute, which are often ignored, contain important information for 3D reconstruction. The null functions of CPET are significantly reduced (1) for alpha particles when the direction or energy is measured along with the position for each particle and (2) for beta particles when the position, direction and energy are all measured for each detected particle.
In this paper, we studied the upper limits of the system performance by considering ideal detectors with no estimation uncertainty. For future work, the effect of estimation uncertainty on the null functions can be considered for a specific setup. Another avenue for future work includes studying charged-particle transport by solving the Boltzmann transport equation, which describes the temporal change of the radiance due to the physical processes of emission, scattering and energy loss. A numerical or analytical solution of the Boltzmann transport equation might provide more insights on beta emission tomography.
References
Caucci, L., Myers, K. J. & Barrett, H. H. Radiance and photon noise: imaging in geometrical optics, physical optics, quantum optics and radiology. Optical Engineering 55, 013102–013102 (2016).
Barrett, H. H. & Myers, K. J. Foundations of Image Science (John Wiley & Sons, Hoboken, NJ, 2004).
Barrett, H. H., Aarsvold, J. N. & Roney, T. J. Null functions and eigenfunctions: tools for the analysis of imaging systems. Progress in clinical and biological research 363, 211–226 (1990).
Culver, J., Ntziachristos, V., Holboke, M. & Yodh, A. Optimization of optode arrangements for diffuse optical tomography: A singular-value analysis. Optics Letters 26, 701–703 (2001).
Burvall, A., Barrett, H. H., Dainty, C. & Myers, K. J. Singular-value decomposition for through-focus imaging systems. Journal of the Optical Society of America A 23, 2440–2448 (2006).
Kyprianou, I. S., Badano, A., Gallas, B. D. & Myers, K. J. Singular value description of a digital radiographic detector: Theory and measurements. Medical Physics 35, 4744–4756 (2008).
Burvall, A., Barrett, H. H., Myers, K. J. & Dainty, C. Singular-value decomposition of a tomosynthesis system. Optics Express 18, 20699–20711 (2010).
Roumeliotis, M. et al. Analysis of a photoacoustic imaging system by the crosstalk matrix and singular value decomposition. Optics Express 18, 11406–11417 (2010).
Jha, A. K. et al. Singular value decomposition for photon-processing nuclear imaging systems and applications for reconstruction and computing null functions. Physics in medicine and biology 60, 7359 (2015).
Furenlid, L. R. et al. FastSPECT II: a second-generation high-resolution dynamic SPECT imager. IEEE Transactions on Nuclear Science 51, 631–635 (2004).
Moore, S., Barrett, H. & Furenlid, L. ModPET-a compact PET system employing modular gamma cameras, maximum-likelihood event-parameter estimation, and list-mode ML-EM reconstruction. Journal of Nuclear Medicine 53, 491–491 (2012).
Miller, B. W. et al. The iQID camera: An ionizing-radiation quantum imaging detector. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 767, 146–152 (2014).
Barrett, H. H. et al. Maximum-likelihood methods for processing signals from gamma-ray detectors. IEEE transactions on nuclear science 56, 725–735 (2009).
Caucci, L. et al. Image science with photon-processing detectors. Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE 1–7 (2013).
Parra, L. & Barrett, H. H. List-mode likelihood: EM algorithm and image quality estimation demonstrated on 2-D PET. IEEE transactions on medical imaging 17, 228–235 (1998).
Snyder, D. L. & Politte, D. G. Image reconstruction from list-mode data in an emission tomography system having time-of-flight measurements. IEEE Transactions on Nuclear Science 30, 1843–1849 (1983).
Hesterman, J. Y., Caucci, L., Kupinski, M. A., Barrett, H. H. & Furenlid, L. R. Maximum-likelihood estimation with a contracting-grid search algorithm. IEEE transactions on nuclear science 57, 1077–1084 (2010).
Caucci, L., Furenlid, L. R. & Barrett, H. H. Maximum likelihood event estimation and list-mode image reconstruction on GPU hardware. 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC) 4072–4076 (2009).
Barrett, H. H., Dainty, C. & Lara, D. Maximum-likelihood methods in wavefront sensing: stochastic models and likelihood functions. Journal of the Optical Society of America A 24, 391–414 (2007).
Chen, J. & Barrett, H. H. Likelihood window, energy window, and bayesian window for scatter rejection in gamma cameras. IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) 3, 1414–1416 (1993).
Ding, Y., Caucci, L. & Barrett, H. H. Charged-particle emission tomography. Medical Physics 44, 2478–2489 (2017).
Boyd, G. A. Autoradiography in biology and medicine (Academic Press, New York, 1955).
Solon, E. G., Schweitzer, A., Stoeckli, M. & Prideaux, B. Autoradiography, MALDI-MS, and SIMS-MS imaging in pharmaceutical discovery and development. The AAPS (American Association of Pharmaceutical Scientists) journal 12, 11–26 (2010).
Solon, E. G. Autoradiography: high-resolution molecular imaging in pharmaceutical discovery and development. Expert opinion on drug discovery 2, 503–514 (2007).
Johnström, P., Bird, J. L. & Davenport, A. P. Quantitative phosphor imaging autoradiography of radioligands for positron emission tomography, 205–220 (Humana Press, Totowa, NJ, 2012).
Unnerstall, J. R., Niehoff, D. L., Kuhar, M. J. & Palacios, J. M. Quantitative receptor autoradiography using [3H]Ultrofilm: application to multiple benzodiazepine receptors. Journal of neuroscience methods 6, 59–73 (1982).
Caucci, L. & Barrett, H. H. Objective assessment of image quality. V. Photon-counting detectors and list-mode data. Journal of the Optical Society of America A 29, 1003–1016 (2012).
Hricak, H., Adelstein, S. J., Conti, P. S. et al. Advancing nuclear medicine through innovation. Committee on State of the Science of Nuclear Medicine, National Research Council (2007).
O’Donoghue, J. A., Bardies, M. & Wheldon, T. E. Relationships between tumor size and curability for uniformly targeted therapy with beta-emitting radionuclides. Journal of Nuclear Medicine 36, 1902–1909 (1995).
Price, E. W. & Orvig, C. Matching chelators to radiometals for radiopharmaceuticals. Chemical Society Reviews 43, 260–290 (2014).
Miao, Y., Owen, N. K., Fisher, D. R., Hoffman, T. J. & Quinn, T. P. Therapeutic efficacy of a 188Re-labeled α-melanocyte–stimulating hormone peptide analog in murine and human melanoma-bearing mouse models. Journal of Nuclear Medicine 46, 121–129 (2005).
Ding, Y. Charged-Particle Emission Tomography. PhD dissertation, The University of Arizona (2016).
Barrett, H. H. et al. Beta and alpha emission tomography for three-dimensional autoradiography. https://www.google.com/patents/WO2015163961A2?cl=en WIPO patent application PCT/US2015/014,223 (2015).
Davison, M. E. The ill-conditioned nature of the limited angle tomography problem. SIAM Journal on Applied Mathematics 43, 428–448 (1983).
Knoll, G. F. Radiation detection and measurement (John Wiley & Sons, Hoboken, NJ, 2010).
Ding, Y., Caucci, L. & Barrett, H. αET: Alpha Emission Tomography. IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) 1–3 (2014).
Rowland, D. J. & Cherry, S. R. Small-animal preclinical nuclear medicine instrumentation and methodology. Seminars in Nuclear Medicine 38, 209–222 (2008).
Lehovich, A. List-mode SPECT reconstruction using empirical likelihood. PhD dissertation, The University of Arizona (2005).
Jha, A. K. Retrieving information from scattered photons in medical imaging. PhD dissertation, The University of Arizona (2013).
Fisher, R. A. Theory of statistical estimation. Mathematical Proceedings of the Cambridge Philosophical Society 22, 700–725 (1925).
Abbey, C. K., Clarkson, E., Barrett, H. H., Müller, S. P. & Rybicki, F. J. A method for approximating the density of maximum-likelihood and maximum a posteriori estimates under a gaussian noise model. Medical image analysis 2, 395–403 (1998).
Agostinelli, S., Allison, J. & Amako, K. et al. Geant4 a simulation toolkit. Nuclear instruments and methods in physics research section A: Accelerators, Spectrometers, Detectors and Associated Equipment 506, 250–303 (2003).
Allison, J., Amako, K. & Apostolakis, J. et al. Geant4 developments and applications. IEEE Transactions on Nuclear Science 53, 270–278 (2006).
Acknowledgements
This research was supported by NIH grants 5 R37 EB000803 and 5 P41 EB002035.
Author information
Authors and Affiliations
Contributions
Y.D. designed and carried out the study. L.C. contributed to the conception and execution of the study. H.H.B. developed the theory and supervised the research. All authors contributed in writing the paper.
Corresponding author
Ethics declarations
Competing Interests
The authors declare that they have no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ding, Y., Caucci, L. & Barrett, H.H. Null functions in three-dimensional imaging of alpha and beta particles. Sci Rep 7, 15807 (2017). https://doi.org/10.1038/s41598-017-16111-z
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-017-16111-z
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.