Abstract
The development is reported of an ultra-rapid, point-of-care diagnostic device which harnesses surface acoustic wave (SAW) biochips, to detect HIV in a finger prick of blood within 10 seconds (sample-in-result-out). The disposable quartz biochip, based on microelectronic components found in every consumer smartphone, is extremely fast because no complex labelling, amplification or wash steps are needed. A pocket-sized control box reads out the SAW signal and displays results electronically. High analytical sensitivity and specificity are found with model and real patient blood samples. The findings presented here open up the potential of consumer electronics to cut lengthy test waiting times, giving patients on the spot access to potentially life-saving treatment and supporting more timely public health interventions to prevent disease transmission.
Similar content being viewed by others
Introduction
Ebola and Zika viruses offer a stark reminder that infectious diseases rank among the gravest threats to human health, and can spread rapidly and unpredictably. New infections will continue to emerge each year, and old enemies re-emerge, increasingly with acquired-drug resistance (e.g. gonorrhoea and HIV). Rapid diagnosis plays a crucial role in any outbreak situation, empowering patients to gain faster access to potentially life-saving treatment, and informing prevention strategies to protect the wider public. However, routine diagnostic tests based on enzyme linked immunosorbent assays (ELISAs) and polymerase chain reaction (PCR) are confined to centralized laboratories often requiring large, sophisticated, costly instrumentation and highly trained staff. Inherent delays occur between taking samples, conveying them to the laboratory, waiting for results to come back and subsequent follow up appointments1,2,3. This means that a patient often has to make multiple visits to a clinic in order to receive treatment, potentially over long distances. This delays prescribing of treatment with increased risk of suffering, mortality, and also incorrect prescription of antimicrobials.
Recent policy drivers aim to widen access to testing using so called ‘rapid’ point-of-care tests (POCT) but the performance and implementation of these tests still remain a challenge4. The most common tests based on lateral flow technology are still relatively slow, requiring a 10–20 minute waiting time for results5. This exceeds a typical doctor’s appointment (8–10 mins in the UK6) necessitating changes to patient pathways within a clinic with additional on-costs and staffing implications. It is also notoriously difficult to interpret a faint lateral flow test line by eye, particularly for non-experts (e.g. self-testers)1. Those tests that are currently available are insensitive to recent (acute) infections7 and lack the ability to automatically capture test results electronically, risking an incorrect reading, missed opportunities to link patients to care pathways and potential data loss for public health (e.g. during an Ebola outbreak)8. Alternatively, uneccesary treatment may be initiatied for those with false positive results9. An additional review of more recent diagnostics that have appeared in the literature (but not yet in the clinic) is given by Wang et al.10.
Here we report the development of an ASSURED (Affordable, Sensitive, Specific, User friendly, Rapid and Robust, Equipment Free and Delivered11) smartphone-connected SAW medical device, to detect biomarkers of HIV infection in a patient sample within seconds. The marriage between diagnostics and consumer electronics opens up the potential to dramatically widen access to testing and support convenient on the spot follow-up testing and care. With more mobile phone subscriptions than people on the planet (7.4 billion in 201612), their reach could help to bring tests outside of centralized hospital laboratories, to where they are most needed at the point-of-care, particularly in resource-limited settings across the developing world and even into the home. Smartphone usage is on the rise with an estimated 2 billion subscriptions worldwide and rapid growth in developing countries −34% of South Africans currently own a smartphone and forecasts predict 500 million smartphones in Sub-Saharan Africa by 202013. Beyond improving access, their battery life, processing power, display screen and inbuilt sensors could, in principle, be harnessed to significantly improve the performance of current POCTs and offer faster access to care, including follow up appointments in local clinics.
There have been a number of advances recently in the field of smartphone-connected diagnostics, both with in vivo and in vitro testing, using either the built-in smartphone sensors or their data transmission capabilities to link with existing diagnostic systems14. They include a smartphone-based system to monitor changes in pH in sweat or saliva, using disposable strips;15 colorimetric rapid quantification of vitamin D levels, using the built-in smartphone camera, paper strips and image analysis algorithms;16 as well as a more general smartphone-based reader for lateral flow immunochromatic assays17. Another recent development is the report of nanophotonic-devices which interface with smartphone cameras to read out current commercial lateral flow tests18, overcoming the need to visually read tests by eye. Microfluidic technologies such as the mChip and a ‘dongle’ recreate all the functions of an ELISA to diagnose HIV and syphilis within 15 minutes with nanoparticle signal enhancements and 6 wash steps19. By contrast, our approach requires no optics, microfluidics, analyte labelling, amplification or wash steps and instead exploits tiny microelectromechanical piezoelectric SAW sensors. SAW microelectronic filters, found within every smartphone, are thus transformed into biochips which can directly detect disease biomarkers in bodily fluids such as blood.
This platform technology can in principle be applied to a range of diseases and here we focus on HIV as an exemplar. This choice was driven by the compelling unmet human and economic needs which have triggered major policy drivers to widen access to HIV testing to hospital emergency services, doctors’ surgeries, community outreach centers20 and self-testing in the home21. The AIDS pandemic ranks among the most devastating infectious diseases in human history, infecting more than 78 million people and resulting in 39 million deaths22 with tremendous burden of illness in Sub-Saharan Africa. A significant proportion of those infected remain unaware of their infection – estimates range from 17% in UK23 to 55% in sub-Saharan Africa24. Late diagnosis is associated with a 10-fold increased risk of death23 and increased risk of unknowingly transmitting the infection25. Early diagnosis and access to antiretroviral treatment increases life-expectancy by 10 years26, reduce infant mortality by 76%27 and in pregnant women can reduce the risk of transmitting the virus to their babies to less than 1%28. According to the US Centre for Disease Control and prevention (CDC), every case of HIV that is prevented saves $380,000 in lifetime treatment costs28.
We sought to engineer the first prototype smartphone-connected SAW device to diagnose HIV with high sensitivity and specificity, and to harness the processing power of smartphones to speed up the delivery of results to make testing more convenient and avoid lengthy waiting times. We first optimized SAW biochips and capture coatings to detect model HIV antibodies and recombinant antigens (anti-p24 and p24 respectively). This work was then extended to a proof of concept assay testing real patient samples, using differential measurements with reference tests to achieve high specificity and sensitivity within seconds.
Results
SAW biosensor device and biochips for HIV
The principle of SAW was first described by Lord Rayleigh in 188529. Here we harness advances in microfabrication and the piezoelectric effect to generate a shear horizontal surface acoustic wave (SH-SAW) on a millimetre sized biochip. The wave generates a surface particle displacement perpendicular to the direction of the wave propagation30, and makes the biochip sensitive to reactions occurring on the surface. Our prototype devices and biochips are shown in Fig. 1(a–d) and described fully in the Experimental Section. In brief, our device comprises a disposable biochip, a pocket-sized control box reader and a mobile device (laptop or smartphone) to analyse, display and transmit results. Each biochip comprises a plane piezoelectric quartz crystal (36°Y-cut 90°X-propagation), with gold input and output interdigitated electrodes and a ‘sensing’ area in between. We tailored the sensing area with a thin gold film and a layer of capture proteins. If a biomarker of HIV is present in a finger prick of blood, it binds to the capture proteins on the surface of the biochip. This gives rise to a perturbation of mass and viscoelasticity which can be readily detected by the phase change of the SH-SAW - the difference in wave phase measured in degrees (°) between the input and the output electrodes (Δφ, Fig. 1e). The diagnosis of HIV is based on the detection of an immunological (anti-p24 antibody) and/or virological (p24 antigen) biomarker (Fig. 1f) which become detectable in blood at relatively defined time points post exposure31,32,33. Our laboratory prototype device comprises a small hand-held control box (measuring 14 × 10 × 4 cm) and up to four individual disposable SAW biochips (each measuring 25 × 7 × 2 mm) which can be connected in parallel for multiplexed analysis of biomarkers and control measurements. A more advanced development prototype device, containing all the same elements as the laboratory prototype, is shown in Fig. 1b, where the SAW biochip is mounted on a disposable cassette resembling a USB stick and results sent to a smartphone app either via a cable or via bluetooth. We note that all the measurements presented herein were performed using the laboratory prototype presented in Fig. 1a.
Key technical breakthroughs contributing to the performance of our ASSURED device include (i) Affordable - mass manufacture of SAW biochips using low cost quartz/gold materials instead of more common and expensive materials (e.g. lithium tantalate). Billions of SAW filters are manufactured each month for the mobile phone industry. The price per SAW filter is currently less than a couple of tens of cents each. It is expected that the total cost of goods will be economically viable due to the ability to mass manufacture since they are purely based on microelectronics, standard thin film deposition of gold, and bulk production of bio-reagents, therefore making the cost of goods compatible with single sample disposable tests. The detection circuit of the control box is expected to cost less than $30 and we are also looking at stand-alone disposable formats. The SAW biosensor presented here is also inherently low power and the control box can be charged by a smartphone. (ii) Sensitive – it operates at an optimized frequency of 251.5 MHz, 1–2 orders of magnitude higher than a typical quartz crystal microbalance benchtop analyzer (1–30 MHz). The SH-SAW is highly sensitive to changes in mass and mechanical properties of materials close to the biochip surface and minimizes losses to bulk modes30. (iii) Specific diagnosis – is achieved via a multiplex array of 4 biochips which allows differential measurements using reference biochips to reduce the risk of non-specific signals (e.g. temperature and viscosity) leading to false positive results. (iv) User friendly – results are electronically displayed on the user friendly app/software (iOS or Android) interface (e.g. ‘Reactive/Yes/No’), overcoming problems with interpreting and capturing lateral flow tests results. Results can be encrypted and securely sent to a local healthcare system to receive a follow up appointment. (v) Robust and Rapid – SAW biochips are small, rugged and very stable34. The patented micro fabricated glass-polymeric capping layer protects the interdigitated electrodes35 meaning we do not require any sophisticated microfluidics to be interfaced with the device, which can be prone to clogging. The biochips are insensitive to environmental lighting conditions, which can affect the interpretation of a lateral flow test and other optical assays36. Importantly, tests can be performed ultra-rapidly via direct detection methods overcoming the need for multiple steps, amplification and wash steps used in microfluidic devices. (vi) Equipment free – removes the need for costly instrumentation by using the smartphone battery supply, processing power and display screen. (vii) Delivered – in future this technology could leverage on established diagnostics and telecommunications supply chains.
Detection of anti-p24 antibodies at clinically relevant levels
Anti-HIV antibodies are the most common target in POCT for HIV, the immunological response typically being detected several weeks after exposure to the virus. Here we initially focused on the detection of antibodies against the viral capsid p24, one of the most conserved and abundant proteins found in HIV. This is achieved using SAW biochips functionalised with recombinant p24 capture ligands (Fig. 2a) via dithiobis[succinimidyl propionate] (DSP) thiol linker chemistry, whereby the gold-sulphur bond drives the formation of a relatively well defined self-assembled monolayer and the terminal reactive ester group forms a covalent bond to the lysine residues on p24. We incubated a 20 µl drop of buffer for 2.5 minutes, to establish a stable baseline before loading the sample. The subsequent injection of 500 nM (75 µg/ml) anti-p24 antibodies triggered a rapid increase in the measured phase shift (Δφ) leading to a total phase shift of 46° after just five minutes (Fig. 2b). This phase shift indicates a change in mass and/or viscosity at the surface proximity associated with the specific binding of anti-p24 antibodies to p24 capture proteins on SAW biochips. We sought to carefully evaluate the analytical specificity of the SAW biochips, via negative control experiments. This involved injecting non-specific proteins - including a blank injection of buffer containing bovine serum albumin and then a non-specific antibody called anti-GBP5, with no known affinity to HIV p24, under exactly the same reaction conditions and concentration as the specific antibody, anti-p24. These results are shown in Fig. 2b - an injection of buffer with BSA shows no change in phase shift (orange trace Fig. 2b) and a sample containing 500 nM (75 µg/ml) of anti-GBP5 shows a negligible response (negative control, red trace Fig. 2b).
Next we systematically tested the relationship between the phase shift Δφ and the concentration of anti-p24 antibodies in solution at clinically relevant levels and found it scaled to a Langmuir adsorption isotherm, R2 = 0.994 (Fig. 2c). All measurements were performed in triplicate. At low antibody concentrations (5–20 nM, which corresponds to 0.75–3 µg/ml), the phase signal increases linearly from 4 ± 0.1 to 22 ± 0.6°. Above 20 nM (3 µg/ml), the signal begins to plateau, presumably when all the available binding sites on the surface are occupied. Testing three independent biochips, the lowest detected concentration (LDC) was found to be 2 nM (300ng/ml) anti-p24 antibodies in buffer within 5 minutes and the limit of detection (LOD) is 1.1 nM (165ng/ml), estimated using the function fitting the data presented in the insert of Fig. 2c, and defined as the concentration corresponding to a signal readout of the mean plus three times the standard deviation of the measurements taken for 0 nM samples. Both the LDC and LOD are well below the clinically relevant range of concentrations for anti-p24 antibodies, which is reported to be 82–1,900 µg/ml, or 0.55–12.67µM32. The small error bars reflect the high reproducibility of these assays between different biochips and functionalisation steps. These results compared favourably to those obtained using commercial lateral flow tests which we have found can detect 10–30 nM (1.5–4.5 µg/ml).
Testing times - ultra-fast diagnosis of HIV within seconds
Having established that SAW biosensors can detect antibodies at clinically relevant levels, we sought to test how quickly results could be delivered by analysis of the phase change in real-time. Reducing the time to deliver a test result is crucial for its utility within a typical primary healthcare setting, allowing treatment and care to be given on the spot, overcoming the need for multiple visits and reducing the risk that patients may not return to receive their results. The faster the test can be delivered the more time will be available for post-test counselling and care. The Foundation for Innovative Diagnostic (FIND)’s target product profile for a HIV self-test for use in the home underlines the need for results in under 5 minutes37.
Figure 2d shows the average phase shift (for three biochips) recorded shortly after sample injection, for the range of antibody concentrations tested. The results demonstrate that clinically relevant HIV antibody concentrations can be detected and distinguished just 10 seconds after injection of sample (sample-in-result-out) down to 50 nM (7.5 µg/ml). To our knowledge there are no tests commercially available or reported in the literature that are this rapid and exhibit a similar limit of detection; those currently on the market require a visual readout with the associated difficulties1.
Detection of recent HIV infections using recombinant p24 antigen detection
Building on the detection of HIV antibodies on SAW biochips, we next sought to investigate whether SAW biochips could detect a virological marker of infection, namely p24 antigen. This is a key biomarker of acute HIV infections and is used in combination with antibody detection in ‘gold-standard’ fourth generation HIV diagnostic assays in centralised laboratories. However, this protein is only present in human blood in minuscule levels, around a million-fold lower concentrations than its antibody counterpart32,38,39,40,41,42 and a billion times lower than human serum albumin, presenting a major challenge for POCT diagnostics. Fourth generation p24 POCT are beginning to emerge but their performance in the field has been suboptimal, and this has been attributed to a number of factors, including the low concentration of p24 antigen in patient samples4.
Here we present the first proof of concept for detection of p24 by developing SAW biochips functionalised with novel capture ligands engineered from anti-p24 llama antibodies. These novel capture ligands are llama VHH (the Variable region of the heavy chain of heavy chain-only antibodies), and are one-tenth the size of conventional antibodies43. Their small footprint means that VHH can access hidden clefts which are inaccessible to larger proteins and in principle they can be packed into dense arrays of capture ligands for diagnostic applications. They have remarkable temperature stability44, have shown strong affinity to p2445 and are therefore ideal candidates for capture ligands on SAW biochips. We immobilised the llama VHH on SAW biochips using DSP chemistry. Characterisation by X-ray photoelectron spectroscopy provides evidence of monolayer coverage, which we estimate to be around 0.11 VHH/nm2 (Figure S1 in Supplementary Information).
Figure 3a shows direct p24 detection on SAW biochips coated with anti-p24 llama VHH. A sample concentration of 4 nM (96 ng/ml) p24 gave rise to a phase change of only 1.2°, even after a 20 minute incubation. Since p24 is only one-sixth the mass of an antibody, a strategy of signal amplification was required through the creation of a larger immuno-sandwich, to achieve detection of p24, corresponding to a six-fold increase in mass per antigen. We tested five combinations of detection antibodies from commercial sources and the AIDS Reagent Program. All antibodies used in Fig. 3b are described in Table 1.
We found that the largest phase change was measured for a combination of llama VHH capture ligand and the NIH-3537 detection antibody (labelled immune-complex 3 in Fig. 3b). Figure 3b shows that this combination generated a 5.5° phase shift in response to 4 nM (96ng/ml) p24 within 20 minutes. This corresponds to a five-fold increase in signal generated by the larger mass of the immuno-sandwich complex although additional factors such as surface viscoelastic changes may also contribute to this enhancement. To test the specificity of this signal we ran a control experiments including a sample with no p24 and injecting non-p24-targeting detection antibodies (Fig. 3c). The phase shift measured was significantly smaller than when detecting a p24/anti-p24 immuno-sandwich, indicating that the immuno-sandwich complex gave specific detection of p24.
Next we titrated the concentration of p24 while keeping the anti-p24 antibody fixed and in excess (200 nM − 30 µg/ml). Testing three independent biochips per concentration, a linear relationship was observed in the range 2–40 nM (48–960ng/ml) p24 (Fig. 3d-e) followed by plateauing of the signal - closely fitting a Langmuir adsorption isotherm model (Fig. 3d solid black line fit, R 2 = 0.96). This demonstrates the ability of our biosensor to detect HIV p24 proteins down to low nanomolar concentrations in the presence of other background proteins at one million fold higher concentrations (2% w/v or 20 mg/ml BSA). The LDC of p24 was found to be 2 nM (48ng/ml), however, this remains above the clinically relevant concentrations of p24 measured using 4th generation HIV assay laboratory systems (1–1000 pg/ml or 0.04 to 41.7 pM48). For example, the bioMérieux VIDAS measures 3–400 pg/ml (0.13–16.7pM) of p24 (manufacturer’s information). A benchmarking study using a quartz crystal microbalance was used to corroborate the p24-VHH assay (Figure S2 in Supplementary Information) and suggest that with identical surface chemistries, the SAW biosensor offers a 20-fold signal enhancement.
Testing clinical patient plasma samples for HIV
Our results show that SAW biosensors have the ability to rapidly detect biomarkers of HIV with high reproducibility down to low nanomolar (ng/ml) concentrations using model samples in buffer with 2% w/v BSA. As a next step towards the proof of concept of their clinical utility, we tested their performance with anonymised patient samples in partnership with University College London Hospital (UCLH). The challenge of testing patient samples is that blood plasma contains a complex background of serum albumins, globulins, fibrinogen, glucose, clotting factors, hormones, and electrolytes, which are typically present at a million-fold higher concentrations than HIV biomarkers. The high protein concentration of plasma, typically 70 mg/ml, also has a significantly higher viscosity than buffer (1.5 to 1.8 times more viscous49). Therefore, it is essential to differentiate the contribution of the specific biomarker binding to the surface from non-specific background contributions e.g. sample viscosity, to avoid the risk of false positives and false negatives.
To achieve this, we developed a multiplexed biochip assay involving the detection of HIV anti-p24 antibodies on ‘test’ biochips (coated with p24) and reference biochips (coated with an anti-adsorptive protein, non-animal protein, which has no affinity for either of the HIV biomarkers tested) to account for non-specific signals such as viscosity and temperature (Fig. 4a). The biochips were incubated in buffer for 30 seconds to record a stable baseline. Two patient blood plasma samples were then tested: one HIV positive sample and one negative for anti-HIV antibodies. The phase shifts of both the reference and the test biochips are plotted against time in Fig. 4b. Upon addition of the HIV-positive patient samples, the phase shift of the test biochip (solid line) rapidly increases within 10 seconds to 56° while the reference biochip (dashed line) increases to 20° and effectively plateaus at this level. We remove the non-specific signals that can arise from known differences in viscosity and temperature between buffer and human plasma by taking a differential measurement (test minus reference for each sample). The differential signal shown in Fig. 4c, shows the specific binding of anti-p24 antibodies to p24-coated test chips with a 36° within just 10 seconds (red trace Fig. 4c). By contrast, the differential signal from a healthy volunteer HIV-negative sample gave only a 1° phase shift after 10 seconds.
Discussion
Our findings show how SAW-based diagnostic devices can be used to rapidly detect biomarkers of HIV infection in model and patient samples with high sensitivity. The microelectromechanical device, comprising a disposable biochip and a small control box which in this study was powered by a laptop, benefits from direct detection of biomarkers on a disposable quartz biochip without the need for any labelling, amplification or wash steps.
We are not aware of any test on the market that can produce such a rapid result. Our findings demonstrate that these SAW devices have the ability to meet the 5 minute time-to-result requirement of an ‘ideal’ HIV self-test outlined by FIND37. This proof of concept study is focused on HIV but our platform technology can be applied to a range of different infectious diseases and other non-communicable diseases. The speed of the tests will help to support patients and front-line health-workers by widening access to testing outside of hospital settings, and cut waiting times to give results on the spot. This reduces the need for multiple visits to a clinic to receive a diagnostic result, increases the likelihood that the patient will receive the correct treatment - reducing dependence on presumptive treatments - and reduces the risk that the patient’s health will deteriorate before a correct diagnosis is made. The WHO strongly recommends early testing of infants as a key to initiate treatment faster50. Early diagnosis is also known to significantly reduce the risk of onwards transmission by people unaware of their infection25 and could be used in a variety of point-of-care settings - from an emergency room to a local doctor’s office, pharmacist, community outreach setting, a rural clinic in a developing country and in the home.
Our findings also highlight the importance of reference sensors and differential measurements. Herein we used individual chips to acquire test and differential measurements but research is underway to engineer a multichannel biochip where up to 4 reagents can be tested simultaneously with a single sample and deployed in a larger pilot clinical study to assess the clinical sensitivity and specificity and time to result of our technology in real world settings (Gray, E. R. et al. Manuscript in preparation. (2017)). We show that the platform technology can also be tailored to detect p24, via an immuno-sandwich complex. Further work to understand the fundamental origin of the SAW signal is underway51. These developments aim to enable detection of low level of p24 associated with early infections, where a person may unknowingly be 26 times more infectious than during later asymptomatic infection52. Indeed, a recent study in North America has shown that early infection accounted for approximately half of all onward transmissions53.
Despite these promising results, there are a number of limitations of our current work which we hope to address in future. (i) To date, all the laboratory prototype devices, schematised in Fig. 1a, still require manual pipetting in two steps (buffer stabilisation followed by the addition of sample. The goal is to have a single sample addition step, using the smartphone connected prototype device shown in Fig. 1b. (ii) The biochips used in this work were functionalised manually and this could lead to variations between biochips. In the future, a semi-automated protocol (such as ink-jet printing) will be adopted for biochip functionalisation. (iii) The laboratory prototype has two sets of two different individual biochips (see Fig. 1a) which is necessary to acquire differential measurements in order to compensate for non-specific signals (due to temperature or sample viscosity). However, this set up is bulky and variations between biochips could lead to potential errors. The next generation of biochips have an in-situ reference channel coated with a non-specific capture ligand, overcoming the need for multiple chips and reducing the volume of sample required. SAW biochips featuring multiple delay lines have been designed, and work is underway to demonstrate their ability to be used for referencing and multiplexed detection of various biomarkers. Examples include co-infections such as HIV and Mycobacterium tuberculosis or syphilis, using a single SAW biochip. (iv) Our current study uses frozen plasma samples and the next step is to use fresh whole blood samples. (v) The analysis of two clinical samples here gives proof of concept. A much larger number of clinical samples are needed to determine clinical sensitivity and specificity. (vi) Future work will investigate the ability of the device to detect different strains of HIV, for example, testing diversity panels representing viruses found worldwide, with controls for other blood-borne viruses such as Hepatitis. (vii) Here we have shown that a SAW biosensor has the required sensitivity needed to detect HIV antibodies in a HIV positive patient sample, however, the detection p24 was not yet within the clinically relevant range. More work is needed to improve the sensitivity to reach pg/ml levels, and subsequently test for p24 in patient samples.
To close, our findings and underlying concepts open up a new field of mass manufacturable, ultra-fast smartphone-enabled consumer diagnostics. True health and economic benefits will be realised when secure on-line patient pathways54 are built to link patients to care in developed and developing countries to empower millions of people at risk of HIV, in much the same way that glucose-tests are empowering people with diabetes to manage their own health at home. In future, geo-located information from mobile devices could also be used to support more timely public health responses to emerging infections such as Ebola or Zika viruses.
Methods
The laboratory prototype SAW biosensor (Fig. 1a) has three components. Firstly, a pocket-sized control box (14 × 10 × 4 cm3). Secondly, the disposable SAW biochip (25 × 7× 2 mm3). Up to four biochips can be connected in parallel to the control box, allowing for multiplex detection of different biomarkers and control measurements. The control box sends and receives an analogue signal to and from the SAW biochips, and converts it to a digital signal. The third component is the smartphone (or laptop) and associated software, which receives the digital signal from the control box, analyses, displays and then can wirelessly transmit results to healthcare databases.
The SH-SAW biochip was designed using a quartz crystal (36°Y-cut 90°X-propagation). The gold interdigitated transducers (IDT) were evaporated onto the crystal and consisted of 80 finger pairs with an aperture of 1mm, exciting a SH-SAW with a wavelength of 20 µm at a frequency of 251.5 MHz. A 2 nm thin film of titanium was evaporated in between the IDTs, followed by the evaporation of a 90 nm thin film of gold to form the sensing area. The IDTs are protected from liquids by a glass lid and epoxy walls, which are constructed using a photolithography technique as described by Kogai et al.35.
The gold surface of the SAW biochips was cleaned by incubation in a 2% (volume/volume) solution of Hellmanex III (Hellma Analytics, UK) for 20 minutes, then thoroughly washed with deionised water. It was then functionalised with the relevant capture protein using two successive 30 minute incubations: a first incubation in a solution of Dithiobis[succinimidyl propionate] (DSP - Thermo Scientific Pierce) dissolved in dimethyl sulfoxide (DMSO, Fischer Scientific, UK) at a concentration of 4 mg/ml, followed by incubation in a solution of the capture protein dispersed in phosphate buffer saline (PBS, Sigma-Aldrich, pH 7.4) at 100 µg/ml. The gold surface was washed with DMSO, then PBS in between the two incubations, and with PBS after the second incubation to wash away any unbound protein. On contact with the gold surface, the disulfide bond of DSP cleaves to yield two identical thiols55 with an amine-reactive N-hydroxysuccinimide (NHS) ester end, which then reacts during the second incubation with any primary amine available on the capture protein to form a stable amide bond56.
Following the functionalisation, the surface was blocked using two successive incubations. First for 15 minutes in a solution of tris buffer saline-tween buffer (TBS-T, pH 7.6 with 0.05% Tween 20, CalbioChem, UK) to block the unreacted DSP molecules via a similar reaction with a primary amine. Secondly in a solution of bovine serum albumin (BSA 2% w/v, Sigma-Aldrich) in TBS-T to reduce unspecific binding during detection of the target protein. The surface was then left incubated in a solution of BSA in TBS-T.
The protocol for testing all samples, including patient plasma samples, involved incubating 20 ul TBS-T buffer on the surface of functionalised SAW biochips for a period of 30 seconds up to 2.5 minutes to establish a stable baseline signal. The buffer was then removed from the surface of the sensor, and replaced with a 20 µl drop of sample, using a micropipette. No further sample handling was needed until the end of the measurement.
Differential measurements were taken by running a test biochip and a reference biochip in parallel: Δφ(differential) = Δφ(test) − Δφ(reference). The reference biochip was functionalised with a protein with no affinity for the analyte (in most cases non-animal protein, G-Biosciences).
The antibodies used were from the following sources: BC1071 (Aalto Bio Reagents, UK), NIH-3537 (AIDS Reagent Program, USA)46, NBS500–473 (Novus Biologicals, UK)47, Capricorn HIV 1/2 (Capricorn Products, USA), C65489M (Meridian Life Science, USA)47, ab89284 (Abcam, UK). The recombinant p24 was obtained from Aalto Bioreagents, UK and is based on GenBank MI5654.
The XPS measurements were conducted at the NEXUS Lab, Newcastle University, UK. The SAW biochips were cleaned using the same protocol as described above. The data were fitted and analysed using CasaXPS software.
The QCM sensors (QSX 301, Biolin Scientific, UK) were cleaned and functionalised with DSP offline, using the same protocol as the one used for the SAW biochips, then inserted into the system (Q-sense E4, Biolin Scientific, UK) to be functionalised with the relevant protein.
The HIV-positive sample was anonymized and had clearance for discard from the UCLH diagnostic laboratory. They were collected as part of the ICONIC project, approved by the Ethical Committee NRES Committee London - Surrey Borders HRA, Research Ethics Committee (REC) London Centre Study title: InfeCtion respONse through vIrus genomiCs (ICONIC) REC reference: 13/LO/1303 HIV viral load was over 5,000 c/mL. Aside from the viral load, no further eligibility criteria were imposed. No clinical measurement of anti-HIV antibody quantitative levels is made by the hospital. The HIV-negative sample was taken after informed consent was obtained from a healthy volunteer from UCL staff. All samples were collected in Ethylenediaminetetraacetic acid BD vacutainers, centrifuged at 3000xg for 20 minutes to separate fractions and plasma, aliquoted and stored at −80 °C until use. All methods were performed in accordance with relevant guidelines and regulations.
The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.
Ethics statement
The University College London Hospital Research Clinical Microbiology Department reviewed and exempted the HIV-positive samples used in this study from ethics review because it was an assay development, and waived the need for consent due to the fact the patient material used was fully anonymised. HIV-negative samples were obtained from UCL staff and students who gave full informed consent. The study was reviewed by UCL Ethics Board and given study number 6109/001.
Disclaimer
This article/paper/report presents independent research funded by the National Institute for Health Research (NIHR). The views expressed are those of the author(s) and not necessarily those of the NHS, the NIHR or the UK Department of Health.
References
Grusky, O., Roberts, K. J. & Swanson, A.-N. Failure to return for HIV test results: a pilot study of three community testing sites. J. Int. Assoc. Physicians AIDS Care (Chic). 6, 47–55 (2007).
Gray, R. H. et al. Limitations of rapid HIV-1 tests during screening for trials in Uganda: diagnostic test accuracy study. BMJ 335, 188 (2007).
Pilcher, C. D. et al. Performance of rapid point-of-care and laboratory tests for acute and established HIV infection in San Francisco. PLoS One 8, e80629 (2013).
Wolpaw, B. J. et al. The failure of routine rapid HIV testing: a case study of improving low sensitivity in the field. BMC Health Serv. Res. 10, 73 (2010).
WHO. Rapid HIV tests: Guidelines for use in HIV testing and counselling services in ressource-constrained settings. WHO (World Health Organization, 2004).
NHS. GP appointments - The NHS in England - NHS Choices. Available at: http://www.nhs.uk/NHSEngland/AboutNHSservices/doctors/Pages/gp-appointments.aspx. (Accessed: 24th February 2016) (2016).
Patel, P. et al. Rapid HIV screening: missed opportunities for HIV diagnosis and prevention. J. Clin. Virol. 54, 42–7 (2012).
Desai, D., Wu, G. & Zaman, M. H. Tackling HIV through robust diagnostics in the developing world: current status and future opportunities. Lab Chip 11, 194–211 (2011).
Eaton, J. W., Johnson, C. C. & Gregson, S. The Cost of Not Retesting: Human Immunodeficiency Virus Misdiagnosis in the Antiretroviral Therapy ‘Test-and-Offer’ Era. Clin. Infect. Dis. doi:https://doi.org/10.1093/cid/cix341(2017)
Wang, S. et al. Advances in addressing technical challenges of point-of-care diagnostics in resource-limited settings. Expert Rev. Mol. Diagn. 16, 449–459 (2016).
Mabey, D., Peeling, R. W., Ustianowski, A. & Perkins, M. D. Diagnostics for the developing world. Nat. Rev. Microbiol. 2, 231–40 (2004).
Carson, S. et al. Ericsson Mobility Report. 4–7 (2015). Available at: http://www.ericsson.com/mobility-report. (Accessed: 29th January 2016)
GSM Association. The Mobile Economy Sub-Saharan Africa 2015 (2015).
Xu, X. et al. Advances in Smartphone-Based Point-of-Care Diagnostics. Proc. IEEE 103, 236–247 (2015).
Oncescu, V., O’Dell, D. & Erickson, D. Smartphone based health accessory for colorimetric detection of biomarkers in sweat and saliva. Lab Chip 13, 3232 (2013).
Lee, S., Oncescu, V., Mancuso, M., Mehta, S. & Erickson, D. A smartphone platform for the quantification of vitamin D levels. Lab Chip 14, 1437–42 (2014).
Mudanyali, O. et al. Integrated rapid-diagnostic-test reader platform on a cellphone. Lab Chip 12, 2678 (2012).
Wei, Q. et al. Fluorescent imaging of single nanoparticles and viruses on a smart phone. ACS Nano 7, 9147–55 (2013).
Laksanasopin, T. et al. A smartphone dongle for diagnosis of infectious diseases at the point of care. Sci. Transl. Med. 7, 273re1 (2015).
House of Lords. House of Lords - Select Committee on HIV and AIDs in the United Kingdom - First Report. Available at: http://www.publications.parliament.uk/pa/ld201012/ldselect/ldaids/188/18802.htm. (Accessed: 28th January 2016) (2011).
UK, D. of H. Modernisation of HIV rules to better protect public - Department of Health (UK). (2013).
UNAIDS. UNAIDS 2014 - Fact sheet. Available at: http://www.unaids.org/en/resources/campaigns/2014/2014gapreport/factsheet. (Accessed: 29th January 2016) (2014).
Public Health England. HIV in the UK – Situation Report 2015 - Incidence, prevalence and prevention. Public Health England Public Health England (2015).
UNAIDS. The Gap Report | UNAIDS 2014. Available at: http://www.unaids.org/en/resources/documents/2014/20140716_UNAIDS_gap_report. (Accessed: 9th May 2016) (2014).
Cohen, M. & S., M. D. et al. Prevention of HIV-1 Infection with Early Antiretroviral Therapy. N. Engl. J. Med. 365, 493–505 (2011).
Bor, J., Herbst, A. J., Newell, M.-L. & Bärnighausen, T. Increases in adult life expectancy in rural South Africa: valuing the scale-up of HIV treatment. Science 339, 961–5 (2013).
Violari, A. et al. Early Antiretroviral Therapy and Mortality among HIV-Infected Infants. N. Engl. J. Med. 359, 2233–2244 (2008).
HIV Cost-effectiveness | Guidance | Program Resources | HIV/AIDS | CDC. Available at: http://www.cdc.gov/hiv/programresources/guidance/costeffectiveness/index.html. (Accessed: 9th May 2016) (2016).
Rayleigh, L. On Waves Propagated along the Plane Surface of an Elastic Solid. Proc. London Math. Soc. s1–17, 4–11 (1885).
Krieger, R. E. & Auld, B. A. Acoustic fields and waves in solids. (R.E. Krieger, 1990).
Tomaras, G. D. & Haynes, B. F. HIV-1-specific antibody responses during acute and chronic HIV-1 infection. Curr. Opin. HIV AIDS 4, 373–379 (2009).
Binley, J. M. et al. Differential regulation of the antibody responses to Gag and Env proteins of human immunodeficiency virus type 1. J. Virol. 71, 2799–809 (1997).
Liao, H.-X. et al. Initial antibodies binding to HIV-1 gp41 in acutely infected subjects are polyreactive and highly mutated. J. Exp. Med. 208, 2237–49 (2011).
Rocha-Gaso, M.-I., March-Iborra, C., Montoya-Baides, Á. & Arnau-Vives, A. Surface Generated Acoustic Wave Biosensors for the Detection of Pathogens: A Review. Sensors 9, 5740–5769 (2009).
Kogai, T., Yoshimura, N., Mori, T. & Yatsuda, H. Liquid-Phase Shear Horizontal Surface Acoustic Wave Immunosensor. Jpn. J. Appl. Phys. 49, 07HD15 (2010).
O′Farrell, B. Lateral Flow Technology for Field-Based Applications—Basics and Advanced Developments. Top. Companion Anim. Med. 30, 139–147 (2015).
Peck, R. B., Lim, J. M, Wellhausen, JD, Lee, AM V. M. Target Product Profile: HIV Self-Test Version 4.1: A White Paper on the Evaluation of Current HIV Rapid Tests and Development of Core Specifications for Next-Generation HIV Tests - PATH (2014).
Brown, A. E. et al. Quantitative relationship of circulating p24 antigen with human immunodeficiency virus (HIV) RNA and specific antibody in HIV-infected subjects receiving antiretroviral therapy. The RV43 Study Group. J. Infect. Dis. 172, 1091–5 (1995).
Böni, J. et al. Simple monitoring of antiretroviral therapy with a signal-amplification-boosted HIV-1 p24 antigen assay with heat-denatured plasma. AIDS 11, F47–F52 (1997).
Erikstrup, C. et al. p24 as a predictor of mortality in a cohort of HIV-1-infected adults in rural Africa. J. Acquir. Immune Defic. Syndr. 48, 345–349 (2008).
Spacek, L. A. et al. Diagnostic accuracy of ultrasensitive heat-denatured HIV-1 p24 antigen in non-B subtypes in Kampala, Uganda. Int. J. STD AIDS 22, 310–4 (2011).
Yates, N. L. et al. HIV-1 gp41 envelope IgA is frequently elicited after transmission but has an initial short response half-life. Mucosal Immunol. 6, 692–703 (2013).
Hamers-Casterman, C. et al. Naturally occurring antibodies devoid of light chains. Nature 363, 446–8 (1993).
Gorlani, A. et al. Llama antibody fragments have good potential for application as HIV type 1 topical microbicides. AIDS Res. Hum. Retroviruses 28, 198–205 (2012).
Gray, E. R. et al. Unravelling the Molecular Basis of High Affinity Nanobodies against HIV p24: In Vitro Functional, Structural, and in Silico Insights. ACS Infect. Dis. 3(7), 479–491 (2017).
Chesebro, B., Wehrly, K., Nishio, J. & Perryman, S. Macrophage-tropic human immunodeficiency virus isolates from different patients exhibit unusual V3 envelope sequence homogeneity in comparison with T-cell-tropic isolates: definition of critical amino acids involved in cell tropism. J. Virol. 66, 6547–54 (1992).
Wong, E. Y. & Hewlett, I. K. HIV diagnostics: challenges and opportunities. HIV Ther. 4, 399–412 (2010).
Duiculescu, D. C., Geffin, R. B., Scott, G. B. & Scott, W. A. Clinical and immunological correlates of immune-complex-dissociated HIV-1 p24 antigen in HIV-1-infected children. J. Acquir. Immune Defic. Syndr. 7, 807–15 (1994).
Harkness, J. & Whittington, R. B. The viscosity of human blood plasma: its change in disease and on the exhibition of drugs. Rheol. Acta 10, 55–60 (1971).
WHO. WHO | Consolidated guidelines on the use of antiretroviral drugs for treating and preventing HIV infection. WHO (World Health Organization, 2016).
Brookes, J., Bufacchi, R., Kondoh, J., Duffy, D. M. & McKendry, R. A. Determining biosensing modes in SH-SAW device using 3D finite element analysis. Sensors Actuators, B Chem. 234, 412–419 (2016).
Hollingsworth, T. D., Anderson, R. M. & Fraser, C. HIV-1 transmission, by stage of infection. J. Infect. Dis. 198, 687–93 (2008).
Brenner, B. G. et al. High Rates of Forward Transmission Events after Acute / Early HIV-1 Infection. J. Infect. Dis. 195, 951–959 (2007).
Aicken, C. et al. Online Clinical Management Pathways for Chlamydia Treatment: Enriching Formative Evaluation of a Complex E-Health Intervention. BMJ Open 5, (2015).
Abruña, H. D. Dithiobissuccinimidyl propionate as an anchor for assembling peroxidases at electrodes surfaces and its application in a H2O2 biosensor. Anal. Chem. 71, 5530–5537 (1999).
Lomant, A. J. & Fairbanks, G. Chemical probes of extended biological structures: synthesis and properties of the cleavable protein cross-linking reagent [35S]dithiobis(succinimidyl propionate). J. Mol. Biol. 104, 243–61 (1976).
Busch, M. P. & Satten, G. A. Time course of viremia and antibody seroconversion following human immunodeficiency virus exposure. Am. J. Med. 102, 117–124 (1997).
BUSCH, M. P. et al. Time course of detection of viral and serologic markers preceding human immunodeficiency virus type 1 seroconversion: implications for screening of blood and tissue donors. Transfusion 35, 91–97 (1995).
Acknowledgements
This work was funded by grants from the NIHR i4i program (II-LA-1111–20004) and EPSRC i-sense Early Warning Sensing Systems in Infectious Disease (EP/K031953/1). VT’s studentship was funded via the UCL Impact award in collaboration with OJ-Bio and RM was supported by the Royal Society Wolfson Merit Fellowship. EN receives funding from the UCLH/UCL NIHR Biomedical Research Centre. We thank OJ-Bio Ltd and Professor Jun Kondoh (Shizuoka University, Japan) for useful discussions, Molly Stevens’s laboratory at Imperial College London for QCM measurements and the Nexus Laboratory (Newcastle University, UK) for the XPS measurements. The p24 VHH was kindly provided by QVQ Holding b.v. The Netherlands.
Author information
Authors and Affiliations
Contributions
R.A.M. and D.A. jointly initiated the collaboration and together with D.P., V.E., R.W., and H.Y. designed the research. V.T. and E.G. carried out all experiments. E.N. contributed samples and clinical input, D.A., H.Y., R.W. and E.N. contributed key reagents, prototypes, biochips and patient samples. C.V. generated the llama V.H.H. V.T. and R.A.M. wrote the manuscript, with revisions from all co-authors.
Corresponding author
Ethics declarations
Competing Interests
The following are declared: D.A., V.L., and H.Y. work for O.J.-Bio which produces the SAW prototype. C.V. is employed by a commercial company, Q.V.Q. B.V. This does not alter their adherence to policies on data sharing and materials. A studentship to V.T. was part-funded by O.J.-Bio. All other authors declare no competing interests exist.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Turbé, V., Gray, E.R., Lawson, V.E. et al. Towards an ultra-rapid smartphone- connected test for infectious diseases. Sci Rep 7, 11971 (2017). https://doi.org/10.1038/s41598-017-11887-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-017-11887-6
This article is cited by
-
An aptamer-based shear horizontal surface acoustic wave biosensor with a CVD-grown single-layered graphene film for high-sensitivity detection of a label-free endotoxin
Microsystems & Nanoengineering (2020)
-
A Prototype Antibody-based Biosensor for Measurement of Salivary MMP-8 in Periodontitis using Surface Acoustic Wave Technology
Scientific Reports (2019)
-
Taking connected mobile-health diagnostics of infectious diseases to the field
Nature (2019)
-
Highly sensitive immunodiagnostics at the point of care employing alternative recognition elements and smartphones: hype, trend, or revolution?
Analytical and Bioanalytical Chemistry (2019)
-
Ultra-rapid, sensitive and specific digital diagnosis of HIV with a dual-channel SAW biosensor in a pilot clinical study
npj Digital Medicine (2018)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.