Craniosynostosis is characterised by the premature fusion of cranial sutures and can be associated with intracranial hypertension (IH), which can damage the brain and vision if left unaddressed [1]. The gold standard method for measuring intracranial pressure (ICP) is invasive ICP monitoring, requiring hospital admission, general anaesthesia and risks such as bleeding and infection [2]. Optical coherence tomography (OCT) is a non-invasive imaging technique providing ultra-high resolution cross-sectional images of the optic nerve within seconds. OCT has demonstrated promise as a non-invasive surrogate marker for IH, but standard table-mounted OCT devices are difficult to use in infants and young children [3]. Here, we assessed the feasibility of a portable OCT device (Fig. 1) in children with craniosynostosis.
This was a multi-centre feasibility study. Fifty children aged under 18 years with a clinical and/or genetic diagnosis of craniosynostosis were included. A portable OCT system was used (Spectralis Flex, Heidelberg Engineering, Heidelberg, Germany). Success rates were recorded for the bilateral and unilateral acquisition of optic nerve head (ONH) images of analysable quality. This was defined as an ONH image where the edges of the disc margins and the cup profile, including its lowermost point, were clearly visualised. This study was approved by Oxford University Hospitals Audit Department (registration ID 5956).
Of the fifty children included in this study, 47 were imaged at the Oxford Craniofacial Unit in outpatient clinics, while three were imaged at Great Ormond Street Hospital (GOSH) on the recovery ward during invasive ICP monitoring. 28 patients were male (56%), 22 were female (44%), 12 had syndromic craniosynostosis (24%), 38 had non-syndromic craniosynostosis (76%). Median age was 47 months (range: 2–173; IQR: 11–80). All 50 patients were awake during imaging, and all OCT images were taken with an assistant operating the computer. No patients were dilated. Bilateral OCT imaging was successful in 39 (78%) and unilateral OCT imaging successful in 44 (88%). The reason for non-acquisition was poor patient compliance in all cases.
To the best of our knowledge, this is the first study to demonstrate that portable OCT imaging using the Spectralis Flex system is feasible in children with craniosynostosis. The success rates for bilateral and unilateral ONH imaging were similar to those recently obtained at Great Ormond Street Hospital (GOSH), using a handheld OCT device [4], and higher than that reported by Patel et al. [5] in a large normative study of children without eye disease (n = 351). This study carried the limitations expected of a feasibility study. Only three patients at GOSH underwent OCT imaging using the Spectralis Flex, because the handheld device (Envisu C2300, Leica Microsystems, Wetzlar, Germany) and table-mounted OCT (Spectralis) are predominantly used in these patients at GOSH––interestingly, all three patients had successful bilateral OCT scans during invasive ICP monitoring. The feasibility of OCT during ICP monitoring permits further prospective work to determine the optimal OCT parameters to maximise sensitivity and specificity for IH in craniosynostosis, and to inform clinical guidelines.
Data availability
The datasets generated and/or analysed during the current study are available from the corresponding author on reasonable request.
References
Wilkie AOM, Johnson D, Wall SA. Clinical genetics of craniosynostosis. Curr Opin pediatrics. 2017;29:622–8.
Tamburrini G, Caldarelli M, Massimi L, Santini P, Di Rocco C. Intracranial pressure monitoring in children with single suture and complex craniosynostosis: a review. Childs Nerv Syst. 2005;21:913–21.
Rufai SR, Hisaund M, Jeelani NUO, McLean RJ. Detection of intracranial hypertension in children using optical coherence tomography: a systematic review. BMJ Open. 2021;11:e046935.
Rufai SR, Bowman R, Bunce C, Panteli V, McLean RJ, Teli S, et al. Feasibility and repeatability of handheld optical coherence tomography in children with craniosynostosis. Transl Vis Sci Technol. 2021;10:24.
Patel A, Purohit R, Lee H, Sheth V, Maconachie G, Papageorgiou E, et al. Optic nerve head development in healthy infants and children using handheld spectral-domain optical coherence tomography. Ophthalmology 2016;123:2147–57.
Funding
Supported by a grant from the Medical Research Council (MRC; MR/N004566/1). SRR is funded by a National Institute for Health Research (NIHR) Doctoral Fellowship Award ID: NIHR300155. This paper presents independent research funded by the NIHR and MRC. The views expressed are those of the authors and not necessarily those of the MRC, National Health Service, NIHR, or the Department of Health and Social Care.
Author information
Authors and Affiliations
Contributions
Concept and design: all authors; acquisition, analysis, or interpretation of data: all authors; drafting of the manuscript: SRR, RP; critical revision of the manuscript for important intellectual content: all authors; obtained funding: SRR; supervision: NUOJ, DJ, TPL.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Purohit, R., Rufai, S.R., Patel, C.K. et al. Feasibility of a portable optical coherence tomography system in children with craniosynostosis. Eye 37, 576–577 (2023). https://doi.org/10.1038/s41433-022-02205-0
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/s41433-022-02205-0