Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Article
  • Published:

Umbilical venous catheter insertion depth estimation using birth weight versus surface measurement formula: a randomized controlled trial

Abstract

Objective

To determine if umbilical venous catheter (UVC) insertion depth estimated by surface measurement (SM) results in optimal catheter tip position on ultrasound as compared with formula using birth weight (BW).

Methods

In this randomized controlled trial, eligible infants were randomized to UVC insertion depth estimated by SM or BW method. We compared proportion of optimum UVC position on ultrasound read by neonatologist masked with group assignment.

Results

UVC was inserted to estimated depth in 164 of 200 enroled infants. There was no difference in the proportion of correctly positioned UVCs between the groups (SM 33/82 (40.2%) vs BW 27/82 (32.9%), p = 0.33). Among BW < 1000 g, SM method had higher correctly positioned UVC (43.7% vs 22.5%, p = 0.07).

Conclusion

There was no difference in the rate of optimally positioned UVC tip between the two methods for estimating UVC insertion depth. However, SM method results in more optimal positioning of UVC tip among BW < 1000 g infants.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Fig. 1

Similar content being viewed by others

References

  1. Hermansen MC, Hermansen MG. Intravascular catheter complications in the neonatal intensive care unit. Clin Perinatol. 2005;32:141–56.

    Article  Google Scholar 

  2. Anderson J, Leonard D, Braner D, Lai S, Tegtmeyer K. Videos in clinical medicine. Umbilical vascular catheterization in neonates. N Engl J Med. 2008;359:e18.

    Article  Google Scholar 

  3. Hogan MJ. Neonatal vascular catheters and their complications. Radiol Clin North Am. 1999;37:1109–25.

  4. Dunn PM. Localization of the umbilical catheter by post-mortem measurement. Arch Dis Child. 1966;41:69–75.

    Article  CAS  Google Scholar 

  5. Shukla H, Ferrara A. Rapid estimation of insertional length of umbilical catheters in newborns. Am J Dis Child. 1986;140:786–8.

    CAS  Google Scholar 

  6. Vali P, Fleming SE, Kim JH. Determination of umbilical catheter placement using anatomic landmarks. Neonatology. 2010;98:381–6.

    Article  Google Scholar 

  7. Verheij GH, te Pas AB, Smits-Wintjens VE, Sramek A, Walther FJ, Lopriore E. Revised formula to determine the insertion length of umbilical vein catheters. Eur J Pedia. 2013;172:1011–5.

    Article  Google Scholar 

  8. Gupta AO, Peesay MR, Ramasethu J. Simple measurements to place umbilical catheters using surface anatomy. J Perinatol. 2015;35:476–80.

    Article  CAS  Google Scholar 

  9. Gomella TL, Cunningham MD, Eyal FG. Venous access: umbilical vein catheterization. In: Gomella TL, Cunningham MD, Eyal FG, eds. Neonatology: management, procedures, on-call problems, diseases, and drugs. 7th ed. New York, USA: McGraw-Hill Global Education Holdings, LLC; 2013. p. 317.

    Google Scholar 

  10. Kieran EA, Laffan EE, O’Donnell CP. Estimating umbilical catheter insertion depth in newborns using weight or body measurement: a randomised trial. Arch Dis Child Fetal Neonatal Ed. 2016;101:F10–5.

    Article  Google Scholar 

  11. Verheij GH, Te Pas AB, Witlox RS, Smits-Wintjens VE, Walther FJ, Lopriore E. Poor accuracy of methods currently used to determine umbilical catheter insertion length. Int J Pedia. 2010;2010:873167.

    Google Scholar 

  12. Lean WL, Dawson JA, Davis PG, Theda C, Thio M. Accuracy of five formulae to determine the insertion length of umbilical venous catheters. Arch Dis Child Fetal Neonatal Ed. 2019;104:F165–F169.

    Article  Google Scholar 

  13. Greenberg M, Movahed H, Peterson B, Bejar R. Placement of umbilical venous catheters with use of bedside real-time ultrasonography. J Pedia. 1995;126:633–5.

    Article  CAS  Google Scholar 

  14. Ades A, Sable C, Cummings S, et al. Echocardiographic evaluation of umbilical venous catheter placement. J Perinatol. 2003;23:24–8.

    Article  Google Scholar 

  15. Simanovsky N, Ofek-Shlomai N, Rozovsky K, Ergaz-Shaltiel Z, Hiller N, Bar-Oz B. Umbilical venous catheter position: evaluation by ultrasound. Eur Radio. 2011;21:1882–6.

    Article  Google Scholar 

  16. Michel F, Brevaut-Malaty V, Pasquali R, Thomachot L, Vialet R, Hassid S, et al. Comparison of ultrasound and X-ray in determining the position of umbilical venous catheters. Resuscitation. 2012;83:705–9.

    Article  Google Scholar 

  17. Pulickal AS, Charlagorla PK, Tume SC, Chhabra M, Narula P, Nadroo AM. Superiority of targeted neonatal echocardiography for umbilical venous catheter tip localization accuracy of a clinician performance model. J Perinatol. 2013;33:950–3.

    Article  CAS  Google Scholar 

  18. Harabor A, Soraisham A. Rates of intracardiac umbilical venous catheter placement in neonates. J Ultrasound Med. 2014;33:1557–61.

    Article  Google Scholar 

  19. Franta J, Harabor A, Soraisham AS. Ultrasound assessment of umbilical venous catheter migration in preterm infants: a prospective study. Arch Dis Child Fetal Neonatal Ed. 2017;102:F251–F5.

    Article  Google Scholar 

  20. Sharma D, Farahbakhsh N, Tabatabaii SA. Role of ultrasound for central catheter tip localization in neonates: a review of the current evidence. J Matern Fetal Neonatal Med. 2018;15:1–9.

    Google Scholar 

Download references

Acknowledgements

We would like to thank the participants and their parents, the nurses who provided care and assisted with procedures, and the clinicians who assisted in recruitment.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Amuchou Singh Soraisham.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Additional information

Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Trial registration number: This study was registered with clinical trials register (NCT02939690 (https://www.clinicaltrials.gov).

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Sheta, A., Kamaluddeen, M. & Soraisham, A.S. Umbilical venous catheter insertion depth estimation using birth weight versus surface measurement formula: a randomized controlled trial. J Perinatol 40, 567–572 (2020). https://doi.org/10.1038/s41372-019-0456-0

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/s41372-019-0456-0

This article is cited by

Search

Quick links