Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Original Article
  • Published:

Ultrasound-guided umbilical catheter insertion in neonates

Abstract

Objective:

Umbilical catheter placement is a routine neonatal emergency procedure that has large variability in technical methods. Commonly used methods are unable to accurately estimate insertion lengths, and X-rays cannot always identify malpositioned catheters. In clinical practice, the placement of umbilical lines takes time away from nursing during a critical transition period. Ultrasound is a safe and commonly used tool in the nursery for clinical management of sick neonates and has been shown to readily identify central catheter tip position. In this study, we sought to determine a more time-efficient and accurate means of umbilical catheter placement using bedside ultrasound.

Study Design:

This is a prospective, randomized, pilot trial of infants of any age or weight admitted to the neonatal intensive care unit who required umbilical catheter placement. Infants were excluded if they had congenital heart disease, abdominal wall defects or had a single umbilical artery. Catheters were placed using either the conventional method, with blinded evaluation of the catheters using ultrasound, or with ultrasound guidance, with input pertaining to catheter tip location. The number of X-rays required to confirm proper positioning, completion time points throughout the procedure and manipulations of the lines were recorded for both groups.

Result:

Ultrasound use decreased the time of line placement with an average saving of 64 min, as well as decreased the number of manipulations required and X-rays taken to place the catheters. The average X-ray time from request to viewing per X-ray was 40 min for all subjects.

Conclusion:

Ultrasound-guided umbilical catheter placement is a faster method to place catheters requiring fewer manipulations and X-rays when compared with conventional catheter placement.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Figure 1
Figure 2

Similar content being viewed by others

References

  1. Dunn PM . Localization of the umbilical catheter by post-mortem measurement. Arch Dis Child 1966; 41 (215): 69–75.

    Article  CAS  Google Scholar 

  2. Shukla H, Ferrara A . Rapid estimation of insertional length of umbilical catheters in newborns. Am J Dis Child 1986; 140 (8): 786–788.

    CAS  PubMed  Google Scholar 

  3. Ades A, Sable C, Cummings S, Cross R, Markle B, Martin G . Echocardiographic evaluation of umbilical venous catheter placement. J Perinatol 2003; 23 (1): 24–28.

    Article  Google Scholar 

  4. Greenberg M, Movahed H, Peterson B, Bejar R . Placement of umbilical venous catheters with use of bedside real-time ultrasonography. J Pediatr 1995; 126 (4): 633–635.

    Article  CAS  Google Scholar 

  5. George L, Waldman JD, Cohen ML, Segall ML, Kirkpatrick SE, Turner SW et al. Umbilical vascular catheters: localization by two-dimensional echocardio/aortography. Pediatr Cardiol 1982; 2 (3): 237–243.

    Article  CAS  Google Scholar 

  6. Lopriore E, Verheij GH, Walther FJ . Measurement of the ‘shoulder-umbilical’ distance for insertion of umbilical catheters in newborn babies: questionnaire study. Neonatology 2008; 94 (1): 35–37.

    Article  Google Scholar 

  7. Wright IM, Owers M, Wagner M . The umbilical arterial catheter: a formula for improved positioning in the very low birth weight infant. Pediatr Crit Care Med 2008; 9 (5): 498–501.

    Article  Google Scholar 

  8. Jeng JJ, Kao HA, Jiang CJ, Guo HR . [A study of insertional length of umbilical artery catheters in newborns]. Zhonghua Min Guo Xiao Er Ke Yi Xue Hui Za Zhi 1989; 30 (2): 100–104.

    CAS  PubMed  Google Scholar 

  9. Baker DH, Berdon WE, James LS . Proper localization of umbilical arterial and venous catheters by lateral roentgenograms. Pediatrics 1969; 43 (1): 34–39.

    CAS  PubMed  Google Scholar 

  10. Butt WW, Whyte H . Blood pressure monitoring in neonates: comparison of umbilical and peripheral artery catheter measurements. J Pediatr 1984; 105 (4): 630–632.

    Article  CAS  Google Scholar 

  11. Kitterman JA, Phibbs RH, Tooley WH . Aortic blood pressure in normal newborn infants during the first 12 h of life. Pediatrics 1969; 44 (6): 959–968.

    CAS  PubMed  Google Scholar 

  12. Oppenheimer DA, Carroll BA . Ultrasonic localization of neonatal umbilical catheters. Radiology 1982; 142 (3): 781–782.

    Article  CAS  Google Scholar 

  13. Oppenheimer DA, Carroll BA, Garth KE, Parker BR . Sonographic localization of neonatal umbilical catheters. Am J Roentgenol 1982; 138 (6): 1025–1032.

    Article  CAS  Google Scholar 

  14. Field T, Ignatoff E, Stringer S, Brennan J, Greenberg R, Widmayer S et al. Nonnutritive sucking during tube feedings: effects on preterm neonates in an intensive care unit. Pediatrics 1982; 70 (3): 381–384.

    CAS  Google Scholar 

  15. Tsui BC, Richards GJ, Van Aerde J . Umbilical vein catheterization under electrocardiogram guidance. Paediatr Anaesth 2005; 15 (4): 297–300.

    Article  Google Scholar 

  16. Biban P, Cavalli C, Santuz P, Soffiati M, Rugolotto S, Zangardi T . [Positioning of umbilical vein catheter with ECG-guided technique: randomized study]. Acta Biomed Ateneo Parmense 2000; 71 (Suppl 1): 647–650.

    PubMed  Google Scholar 

  17. Raval NC, Gonzalez E, Bhat AM, Pearlman SA, Stefano JL . Umbilical venous catheters: evaluation of radiographs to determine position and associated complications of malpositioned umbilical venous catheters. Am J Perinatol 1995; 12 (3): 201–204.

    Article  CAS  Google Scholar 

  18. Kuschel CA . National Women's Newborn Service's Clinical Guideline—Umbilical Catheterisation (sic) (New Zealand) http://www.adhb.govt.nz/newborn/Guidelines/VascularCatheters/UmbilicalCatheters.htm.

  19. Donadieu J, Zeghnoun A, Roudier C, Maccia C, Pirard P, Andre C et al. Cumulative effective doses delivered by radiographs to preterm infants in a neonatal intensive care unit. Pediatrics 2006; 117 (3): 882–888.

    Article  Google Scholar 

  20. Sutton PM, Arthur RJ, Taylor C, Stringer MD . Ionising radiation from diagnostic X-rays in very low birthweight babies. Arch Dis Child Fetal Neonatal Ed 1998; 78 (3): F227–F229.

    Article  CAS  Google Scholar 

  21. Ono K, Akahane K, Aota T, Hada M, Takano Y, Kai M et al. Neonatal doses from X ray examinations by birth weight in a neonatal intensive care unit. Radiat Prot Dosimetry 2003; 103 (2): 155–162.

    Article  CAS  Google Scholar 

  22. Wilson-Costello D, Rao PS, Morrison S, Hack M . Radiation exposure from diagnostic radiographs in extremely low birth weight infants. Pediatrics 1996; 97 (3): 369–374.

    CAS  PubMed  Google Scholar 

  23. Corbett SW, Andrews HG, Baker EM, Jones WG . ED evaluation of the pediatric trauma patient by ultrasonography. Am J Emerg Med 2000; 18 (3): 244–249.

    Article  CAS  Google Scholar 

  24. Froehlich CD, Rigby MR, Rosenberg ES, Li R, Roerig PL, Easley KA et al. Ultrasound-guided central venous catheter placement decreases complications and decreases placement attempts compared with the landmark technique in patients in a pediatric intensive care unit. Crit Care Med 2009; 37 (3): 1090–1096.

    Article  Google Scholar 

  25. Kluckow M, Seri I, Evans N . Echocardiography and the neonatologist. Pediatr Cardiol 2008; 29 (6): 1043–1047.

    Article  CAS  Google Scholar 

  26. Lanoix R, Leak LV, Gaeta T, Gernsheimer JR . A preliminary evaluation of emergency ultrasound in the setting of an emergency medicine training program. Am J Emerg Med 2000; 18 (1): 41–45.

    Article  CAS  Google Scholar 

  27. Lynch RE, Lungo JB, Loftis LL, Ream RS, Gale GB . A procedure for placing pediatric femoral venous catheter tips near the right atrium. Pediatr Emerg Care 2002; 18 (2): 130–132.

    Article  Google Scholar 

  28. Manasia AR, Nagaraj HM, Kodali RB, Croft LB, Oropello JM, Kohli-Seth R et al. Feasibility and potential clinical utility of goal-directed transthoracic echocardiography performed by noncardiologist intensivists using a small hand-carried device (SonoHeart) in critically ill patients. J Cardiothorac Vasc Anesth 2005; 19 (2): 155–159.

    Article  Google Scholar 

Download references

Acknowledgements

We thank all the healthcare professionals who helped with this study and all the parents of the participating study infants. This study was supported by a research grant from General Electric Healthcare. This paper was presented in poster presentation format at the Pediatric Academic Societies Annual Meeting in Baltimore, 5 May 2009.

Author contributions

All coauthors have read, approve of and concur with the submitted manuscript. All individuals designated as authors who made substantial contributions to the conception and design, execution, or analysis and interpretation of data, were involved in the drafting of the manuscript, and approved the manuscript as submitted.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to J H Kim.

Ethics declarations

Competing interests

The authors do not have commercial affiliation or consultant roles that could be construed as a conflict of interest. None of the original material contained in the manuscript has been previously published nor is currently under review for publication elsewhere.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Fleming, S., Kim, J. Ultrasound-guided umbilical catheter insertion in neonates. J Perinatol 31, 344–349 (2011). https://doi.org/10.1038/jp.2010.128

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/jp.2010.128

Keywords

This article is cited by

Search

Quick links