Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Original Article
  • Published:

Accuracy of a New Low-flow Sidestream Capnography Technology in Newborns: A Pilot Study

Abstract

OBJECTIVE: To evaluate the accuracy of a new low-flow sidestream capnography technology and analyze components of the capnogram in mechanically ventilated newborns with and without pulmonary disease.

METHODS: Twenty patients were prospectively identified. Eligible infants were mechanically ventilated and had an indwelling arterial catheter. Two groups were identified: newborns who were receiving mechanical ventilation for pulmonary diseases, and newborns who were receiving postoperative mechanical ventilation for nonpulmonary conditions. End-tidal CO2 (PetCO2) was measured for 1-minute pre- and post-arterial blood sampling, and PetCO2 and PaCO2 were compared for each patient. Eight quantitative waveform parameters were also measured on all patients.

RESULTS: Newborns in the pulmonary group (n=13) (persistent pulmonary hypertension of the newborn/meconium aspiration syndrome, respiratory distress syndrome, pneumonia) and newborns in the control group (n=7) were matched for birth weight, gestational age, and postnatal age. PetCO2–PaCO2 Gradient values were higher in the pulmonary group (7.4±3.3 mm Hg) than controls (3.4±2.4 mm Hg). Four waveform parameters (ascending slope, alveolar angle, alpha angle, descending angle) were identified, which independently differentiated patients with pulmonary disease from controls.

CONCLUSIONS: Low-flow capnography with Microstream technology accurately measured alveolar CO2 in newborns without pulmonary disease, as demonstrated by normal PetCO2–PaCO2 gradients. The measured PetCO2–PaCO2 gradient, as expected, was significantly higher in newborns with pulmonary disease. We also identified four quantitative waveform parameters that may be useful in differentiating between mechanically ventilated newborn patients with and without lung disease.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Figure 1
Figure 2

Similar content being viewed by others

References

  1. Lee KS, Khoshnood B, Wall SN, Chang YP, Hsieh HL, Singh JK . Trend in mortality from respiratory distress syndrome in the United States, 1970–1995 J Pediatr 1999 134: 434–40

    Article  CAS  Google Scholar 

  2. Rodriguez RJ, Martin RJ . Exogenous surfactant therapy in newborns Respir Care Clin North Am 1999 5: 595–616

    CAS  Google Scholar 

  3. Wessel DL, Adatia I, VanMarter LJ et al. Improved oxygenation in a randomized trial of inhaled nitric oxide for persistent pulmonary hypertension of the newborn Pediatrics 1997 100: E7

    Article  CAS  Google Scholar 

  4. Cornfield DN, Maynard RC, deRognier RA, Guiang SF, Barbato JE, Milla CE . Randomized, controlled trial of low-dose inhaled nitric oxide for the treatment of term and near-term infants with respiratory failure and pulmonary hypertension Pediatrics 1999 104: 1089–94

    Article  CAS  Google Scholar 

  5. Plavka R, Kopecky P, Sebron V, Svihovec P, Zlatohlavkova B, Janus V . A prospective, randomized comparison of conventional mechanical ventilation and very early high frequency oscillatory ventilation in extremely premature newborns with respiratory distress syndrome Intensive Care Med 1999 25: 68–75

    Article  CAS  Google Scholar 

  6. American Society of Anesthesiologists Policy Statements on Practice Parameters: Standards for Basic Anesthetic Monitoring 1999 http://www.asahq.org/Standards/02.html#2

  7. Badgwell JM, Heavner JE, May WS et al. End-tidal PCO2 monitoring in infants and children ventilated with either a partial rebreathing or a non-rebreathing circuit Anesthesiology 1987 66: 405–10

    Article  CAS  Google Scholar 

  8. Kirpalani H, Kechagias S, Lerman J . Technical and clinical aspects of capnography in neonates J Med Eng Tech 1991 15: 154–61

    Article  CAS  Google Scholar 

  9. Gravenstein N . Capnometry in infants should not be done at lower sampling flow rates J Clin Monit 1989 5: 63–4

    Article  CAS  Google Scholar 

  10. Sasse FJ . Can we trust end-tidal carbon dioxide measurements in infants? J Clin Monit 1985 1: 147–8

    Article  CAS  Google Scholar 

  11. Gravenstein JS, Paulus DA, Hayes TJ editors. Capnography in clinical practice Stoneham, MA: Butterworth Publishers 1989 43–9

    Google Scholar 

  12. Schieber RA, Namnoum A, Sugden A et al. Accuracy of expiratory carbon dioxide measurements using the coaxial and circle breathing circuits in small subjects J Clin Monit 1985 1: 149–55

    Article  CAS  Google Scholar 

  13. Swedlow DB . Capnometry and capnography: the anesthesia disaster early warning system Semin Anesth 1986 3: 194–205

    Google Scholar 

  14. You B, Peslin R, Duvivier C, Dang Vu V, Grilliat JP . Expiratory capnography in asthma: evaluation of various shape indices Eur Respir J 1994 7: 318–23

    Article  CAS  Google Scholar 

  15. Yaron M, Padyk P, Hutsinpiller M et al. Utility of the expiratory capnogram in the assessment of bronchospasm Ann Emerg Med 1996 28: 403–7

    Article  CAS  Google Scholar 

  16. Colman Y, Krauss B . Microstream capnography technology: a new approach to an old problem J Clin Monit 1999 15: 403–9

    Article  CAS  Google Scholar 

  17. Smalhout B, Kalenda Z . An Atlas of Capnography vol 1: 2nd ed Kerckebosch-Zeist, The Netherlands 1981

    Google Scholar 

  18. Zar JH . Biostatistical Analysis 3rd ed Englewood Cliffs, NJ: Prentice-Hall International 1996 474–80

    Google Scholar 

  19. Miller RG . Simultaneous statistical inference 2nd ed New York: Springer-Verlag 1981 90–3

    Book  Google Scholar 

  20. Hosmer DW, Lemeshow S . Applied Logistic Regression 2nd ed New York: Wiley 2000 47–69

    Book  Google Scholar 

  21. Epstein MF, Cohen AR, Feldman HA, Raemer DB . Estimation of PaCO2 by two non-invasive methods in the critically ill newborn infant J Pediatr 1985 106: 282–6

    Article  CAS  Google Scholar 

  22. McEvedy BA, McLeod ME, Kirpalani H, Volgyesi GH, Lerman J . End-tidal carbon dioxide measurements in critically ill neonates: a comparison of side-stream and mainstream capnometers Can J Anaesth 1990 37: 322–6

    Article  CAS  Google Scholar 

  23. Rich GF, Sconzo JM . Continuous end-tidal sampling within the proximal endotracheal tube estimates arterial CO2 tension in infants Can J Anaesth 1991 38: 201–3

    Article  CAS  Google Scholar 

  24. Rozycki HJ, Sysyn GD, Marshal MK, Malloy R, Wiswell TE . Mainstream end-tidal carbon dioxide monitoring in the neonatal intensive care unit Pediatrics 1998 101: 648–53

    Article  CAS  Google Scholar 

  25. Badgwell JM, McLeod ME, Lerman J, Creighton RE . End-tidal PCO2 measurements sampled at the distal and proximal ends of the endotracheal tube in infants and children Anesth Analg 1987 66: 959–64

    Article  CAS  Google Scholar 

  26. Hopper AO, Nystrom GA, Deming DD, Brown WR, Peabody JL . Infrared end-tidal CO2 measurement does not accurately predict arterial CO2 values or end-tidal to arterial PCO2 gradients in rabbits with lung injury Pediatr Pulmonol 1994 17: 189–96

    Article  CAS  Google Scholar 

  27. Bloom BT et al. Comparison of Infasurf (calf lung surfactant extract) to Survanta (Beractant) in the treatment and prevention of respiratory distress syndrome Pediatrics 1997 100: 31–8

    Article  CAS  Google Scholar 

  28. Hudak ML et al. A multicenter randomized masked comparison trial of synthetic surfactant versus calf lung surfactant extract in the prevention of neonatal respiratory distress syndrome Pediatrics 1997 100: 39–50

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Hagerty, J., Kleinman, M., Zurakowski, D. et al. Accuracy of a New Low-flow Sidestream Capnography Technology in Newborns: A Pilot Study. J Perinatol 22, 219–225 (2002). https://doi.org/10.1038/sj.jp.7210672

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/sj.jp.7210672

This article is cited by

Search

Quick links