Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Original Communication
  • Published:

Accuracy of an eight-point tactile-electrode impedance method in the assessment of total body water

Abstract

Objective: To establish the accuracy of an eight-polar tactile-electrode impedance method in the assessment of total body water (TBW).

Design: Transversal study.

Setting: University department.

Subjects: Fifty healthy subjects (25 men and 25 women) with a mean (s.d.) age of 40 (12) y.

Methods: TBW measured by deuterium oxide dilution; resistance (R) of arms, trunk and legs measured at frequencies of 5, 50, 250 and 500 kHz with an eight-polar tactile-electrode impedance-meter (InBody 3.0, Biospace, Seoul, Korea).

Results: An algorithm for the prediction of TBW from the whole-body resistance index at 500 kHz (height 2/R500 where R is the sum of the segmental resistances of arms, trunk and legs) was developed in a randomly chosen subsample of 35 subjects. This algorithm had an adjusted coefficient of determination (r2adj) of 0.81 (P<0.0001) and a root mean square error (RMSE) of 3.6 l (9%). Cross-validation of the predictive algorithm in the remaining 15 subjects gave an r2adj of 0.87 (P<0.0001) and an RMSE of 3.0 l (8%). The precision of eight-polar BIA, determined by measuring R three times a day for five consecutive days in a fasting subject, was ≤2.8% for all segments and frequencies.

Conclusion: Eight-polar BIA is a precise method that offers accurate estimates of TBW in healthy subjects. This promising method should undergo further studies of precision and its accuracy in assessing extracellular water and appendicular body composition should be determined.

Sponsorship: Modena and Reggio Emilia University.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Figure 1
Figure 2

Similar content being viewed by others

References

  • Baumgartner RN, Chumlea WC, Roche AF . 1989 Estimation of body composition from bioelectric impedance of body segments Am. J. Clin. Nutr. 50: 221–226

    Article  CAS  Google Scholar 

  • Bedogni G, Polito C, Severi S, Strano CG, Manzieri AM, Alessio M, Iovene A, Battistini N . 1996 Altered body water distribution in subjects with juvenile rheumatoid arthritis and its effects on the measurement of water compartments from bioelectric impedance Eur. J. Clin. Nutr. 50: 335–339

    CAS  PubMed  Google Scholar 

  • Bedogni G, Bollea MR, Severi S, Trunfio O, Manzieri AM, Battistini N . 1997 The prediction of total body water and extracellular water from bioelectric impedance in obese children Eur. J. Clin. Nutr. 51: 129–133

    Article  CAS  Google Scholar 

  • Deurenberg P . 1994 International consensus conference on impedance in body composition Age Nutr. 5: 142–145

    Google Scholar 

  • Glantz SA, Slinker BK . 2001 Primer of Applied Regression and Analysis of Variance New York: McGraw-Hill

    Google Scholar 

  • Guo SS, Chumlea WC, Cockram DB . 1996 Use of statistical methods to estimate body composition Am. J. Clin. Nutr. 64: (Suppl): 428S–435S

    Article  CAS  Google Scholar 

  • Heitmann BL . 1994 Impedance: a valid method in the assessment of body composition? Eur. J. Clin. Nutr. 48: 228–240

    CAS  PubMed  Google Scholar 

  • Heymsfield SB, McManus CB, Smith J, Stevens V, Nixon DW . 1982 Anthropometric assessment of muscle mass: revised equations for calculating bone-free muscle area Am. J. Clin. Nutr. 36: 680–690

    Article  CAS  Google Scholar 

  • Kushner RF . 1992 Bioelectric impedance analysis: a review of principles and applications J. Am. Coll. Nutr. 11: 199–209

    CAS  Google Scholar 

  • Kushner RF, Schoeller DA, Fjeld CR, Danford L . 1992 Is the impedance index (Ht2/R) significant in predicting total body water? Am. J. Clin. Nutr. 56: 835–839

    Article  CAS  Google Scholar 

  • Lohman TG, Roche AF, Martorell R (eds). 1988 Anthropometric Standardization Reference Manual Champaign, IL: Human Kinetics Books

    Google Scholar 

  • Lukaski HC, Johnson PE . 1985 A simple inexpensive method of determining total body water using a tracer dose of deuterium oxide and infrared absorption of biological fluids Am. J. Clin. Nutr. 41: 363–370

    Article  CAS  Google Scholar 

  • Mazess RB . 1991 Do impedance changes reflect weight, not composition? Am. J. Clin. Nutr. 53: 178

    Article  CAS  Google Scholar 

  • Organ LW, Bradham B, Gore DT, Lozier SL . 1994 Segmental bioelectric impedance analysis: theory and application of a new technique J. Appl. Physiol. 77: 98–112

    Article  CAS  Google Scholar 

  • Schoeller DA . 1996 Hydrometry In: Human Body Composition ed. AF Roche, SB Heymsfield & TG Lohman, pp 25–44 Champaign, IL: Human Kinetics Books

    Google Scholar 

  • Wang ZM, Deurenberg P, Wei W et al. 1999 Hydration of fat-free body mass: review and critique of a classic body-composition constant Am. J. Clin. Nutr. 69: 833–841

    Article  CAS  Google Scholar 

  • Wotton MJ, Thomas BJ, Cornish BH, Ward LC . 2000 Comparison of whole body and segmental bioimpedance methodologies for estimating total body water Ann. NY Acad. Sci. 904: 181–186

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to G Bedogni.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Bedogni, G., Malavolti, M., Severi, S. et al. Accuracy of an eight-point tactile-electrode impedance method in the assessment of total body water. Eur J Clin Nutr 56, 1143–1148 (2002). https://doi.org/10.1038/sj.ejcn.1601466

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/sj.ejcn.1601466

Keywords

This article is cited by

Search

Quick links