Abstract
Given that reduced graphene oxide (rGO)-based biosensors allow disposable and repeatable biomarker detection at the point of care, we developed a wafer-scale rGO patterning method with mass productivity, uniformity, and high resolution by conventional micro-electro-mechanical systems (MEMS) techniques. Various rGO patterns were demonstrated with dimensions ranging from 5 μm up to several hundred μm. Manufacture of these patterns was accomplished through the optimization of dry etching conditions. The axis-homogeneity and uniformity were also measured to verify the uniform patternability in 4-inch wafer with dry etching. Over 66.2% of uniform rGO patterns, which have deviation of resistance within range of ±10%, formed the entire wafer. We selected amyloid beta (Aβ) peptides in the plasma of APP/PS1 transgenic mice as a study model and measured the peptide level by resistance changes of highly uniform rGO biosensor arrays. Aβ is a pathological hallmark of Alzheimer’s disease and its plasma concentration is in the pg mL−1 range. The sensor detected the Aβ peptides with ultra-high sensitivity; the LOD was at levels as low as 100 fg mL−1. Our results provide biological evidences that this wafer-scale high-resolution patterning method can be used in rGO-based electrical diagnostic devices for detection of low-level protein biomarkers in biofluids.
Similar content being viewed by others
Introduction
Reduced graphene oxide (rGO) has garnered significant attention as a promising nanomaterial because of its scalable production at low cost, high portion of chemically active sites, and solution processability1,2,3,4,5,6,7,8. High electrical conductivity and a large number of reaction sites are significant benefits of rGO-based biosensors for applications in clinical diagnostics of pathological biomarkers9,10,11. Because of these benefits, rGO-based biosensors are considered highly useful in rapid, disposable, and repeatable diagnostics at the point of care12,13. However, most of rGO-based biosensors have low fabrication yield and poor reliability due to the lack of uniform rGO patterning method in large areas. Thus, it is essential to develop rGO patterns enabling massive production with reliable reproducibility for simple and compact diagnostic instruments detecting biomarkers in low concentration.
Some mass production patterning methods have already been introduced: laser writing on assembled graphene oxide (GO) layers at a charged surface by repeating coating and washing steps14 or by spin coating15 and shadow-mask-assisted rGO spray-patterning for accurate patterns in large areas16. However, a large volume of GO solution was required and the thickness of the patterns was controllable only at the micron-scale. The surface-energy-engineered stamping method for patterning of the uniform rGO thin films prepared by a meniscus-dragging deposition (MDD) technique was introduced in our previous research17,18; the patternability of small patterns and arrays with dimensions on the order of several μm was demonstrated in a controllable ultra-thin rGO layer with remarkably high accuracy. Although our previous method forms the rGO micropatterns efficiently, the stamping method has some limitations, such as the directional dependency of striping caused by the molds; this is one of the critical obstacles to reproducible alignment during subsequent process steps such as photolithography. Cost-effective, high-yield fabrication methods are needed to ensure the practicality of graphene-based biosensing applications.
In this study, we introduce an effective wafer-scale rGO patterning method for biosensing applications based on dry etching in conventional micro-electro-mechanical systems (MEMS) equipment. Various shapes and sizes of rGO micro-patterns have been demonstrated through the optimization of patterning in conjunction with dry etching. The axis-homogeneity, which is calculated by analysing the electrical resistance of rGO patterns oriented in different directions on the wafer, was measured to verify the improved uniformity of patterning with dry etching. To assess biological applications of our rGO biosensors for precise detection of protein biomarkers in low concentration, wafer scale fabricated array of rGO biosensor were utilized as shown in Fig. 1. We collected blood of APP/PS1 transgenic Alzheimer mice and measured the plasma level of human amyloid beta (Aβ) peptides. The Aβ peptide is a blood-brain-barrier-penetrating biomarker of Alzheimer’s disease19,20 and requires highly sensitive detection methods because of its low molecular weight (4.5 kDa)21 and picomolar concentration in plasma. The Aβ peptides in plasma were detected to show the reliable immunoreaction response of the rGO biosensor through change of resistance with wafer scale fabrication.
Results and Discussion
The major steps of the rGO patterning method are accomplished with conventional photolithography and dry etching processes to form accurate patterns with high yield at the wafer scale. First of all, the rGO thin film layer was deposited on the Si/SiO2 wafer. The rGO film was produced by the reduction of a GO layer that had been uniformly deposited by the MDD method17,18. The detailed process steps of the rGO deposition are presented in Fig. S1 of the Supporting Information section. The robust and uniform rGO film layer formed with MMD methods was treated the same as any other layer that is generally used in semiconductor fabrication and MEMS techniques. The sheet resistance of the deposited rGO layer across a 4-inch wafer was measured to demonstrate the high uniformity of the film formed by the MDD method; the results are summarized in Fig. S2(b). After the deposition of the rGO film, a photoresist (PR) layer was deposited on the rGO layer by spin coating as shown in Fig. S3. The photolithography, including exposure to ultra-violet (UV) light to form patterns on the PR layer, was performed with a commercial aligner. The patterned PR on the rGO layer was used as a sacrificial layer for selective etching of the rGO layer. The etching process was implemented with reactive ion etching (RIE), which is a kind of dry etching, in an oxygen atmosphere. In this way, the patterns were successfully transferred from the PR to the rGO layer on the SiO2 substrate.
Various high-resolution micro-patterns were formed on the rGO to verify the transfer of the PR patterns to the rGO, as shown in Fig. 2. The optical images of Fig. 2a,b are the PR patterns of the logo ‘KIST’ and an array of rectangular patterns, respectively. Figure 2a′ show the optical images of the rGO patterns after the etching process. The rGO patterns are indistinguishable from the PR patterns. The PR functioned very effectively as a sacrificial-layer to transfer its own patterns to the rGO layer with high resolution and reproducibility.
The smallest width in the final rGO patterns in Fig. 2a′ was 5 μm as indicated by the white arrows. The square patterns of rGO in Fig. 2b′ were also obtained from the PR patterns through the etching process. The feature length of the single square pattern and the pitch size between the patterns are approximately 10 μm in the case of both the PR and the rGO as shown in Fig. 2b,b′, respectively. Other shapes such as stars, circles, the ‘CAU’ logo, and a pattern of parallel lines were also produced to show the patternability of various shapes with conventional dry etching. The array of 50-μm-wide single stars and the circular patterns with a radius of 25 μm were properly formed as shown in the optical images of Fig. 2c,d, respectively. The centre-to-centre distance between features was 100 μm in the arrays of squares and circles. In Fig. 2f, the ‘CAU’ logo was also reproduced within a 1.4 mm × 0.8 mm area, which is a relatively large area, with detailed expressions of the rough edges in each character. Patterns comprising 20-μm-wide lines with 20-μm spacing were also demonstrated as shown in Fig. 2h. As shown in the atomic force microscopy (AFM) image of Fig. 2e, narrow line patterns with 5-μm width and spacing were also fabricated and measured to verify the resolution of the patterning method. The thickness profile measured by AFM from the image of Fig. 2e is shown in Fig. 2g. The thickness of the rGO line patterns is about 8.7 nm, which is the expected value, based on the results of previously reported MDD studies17.
The resistance of fabricated rGO sensor devices was measured and analysed to show the uniform formation of rGO sensor arrays through patterning of rGO by dry etching across the entire 4-inch wafer. The detailed fabrication process is shown in Fig. S3. The variation of resistance in the rGO sensors as a function of position on the wafer was demonstrated as shown in Fig. 3a,b. The fabricated devices on the wafer and the detailed layout of the wafer are shown in Fig. S4. The average resistance (18.3 kΩ) of 24 rGO devices, each of which contains 24 individual arms (resulting in a total of 576 arms) was measured and set as a reference for the entire wafer. Each arm has 50-μm-wide and 100-μm-long rGO patterns as shown in Fig. S4(d). The deviation of resistance in each section from the average value was measured and calculated as a percentage.
The deviations along the y-axis, as shown by the red bars in Fig. 3a, were: −2.5 ± 2.0%, 9.4 ± 4.2%, 0.7 ± 1.4%, −4.5 ± 3.8%, and 0.7 ± 2.9% at sections (1), (2), (3), (4), and (5), respectively. The average values are derived from over 96 or 144 arms of rGO-based sensors in each section of the wafer. Resistance uniformity was better than 5%, except in section (2). Furthermore, the variation within each section is also quite low. The trend in deviations along the x-axis was similar to the y-axis, which shows that the dry etching method can achieve axis-homogeneous patterning. The deviation of resistance along the x-axis, as shown by the red bars in Fig. 3b, were: −13.8 ± 0.5%, −1.9 ± 0.1%, 3.3 ± 0.1%, 5.1 ± 0.1%, 6.9 ± 0.1%, and 0.4 ± 1.7% at sections (a), (b), (c), (d), (e), and (f), respectively. Most sections have around 5% of deviation from the average resistance, and the minimum variation in each section was 0.1%; this shows that the dry etching method can produce very uniform patterns throughout the entire wafer. The uniformity that was achieved here was higher than that reported by another group22, using a different method; and also higher than in our own previous study that utilized a polydimethylsiloxane (PDMS) stamping method13. A comparison of the uniformity and axis-homogeneity achieved by dry etching and by PDMS stamping is presented in Fig. S5.
The distribution of the deviations from the mean value of resistance was also analysed, as shown in Fig. 3c. The number of devices that had specific values of deviation was counted and calculated to display the percentages. The histogram shows that 66.2% of the rGO arms formed by dry etching were found within the deviation range of ±10% [range −20~−15%:2.7%, −15~−10%:5.7%, −10~−5%:10.5%, −5~0%:14.9%, 0~5%:22.7%, 5~10%:18.1%, 10~15%:9.2%, and 15~20%:4.0%, red bars of Fig. 3c].
To verify the uniformity of the pattern formation across the entire wafer, the thickness of the films produced by the dry etching method was measured as a function of the gas composition, the RF power of the plasma, and the etching time, as shown in Fig. 3d–f. The gases oxygen (O2), argon (Ar), and nitrogen (N2) were utilized at 300 W of RF power for dry etching by RIE to keep the etching times at reasonable levels. As shown in Fig. 3d, the O2 gas has a relatively linear etching profile that is sufficient to control etching thickness by controlling the etching time, whereas the other gases do not display this feature. To confirm the dependence of etching rate on RF power, the etched thickness in an O2 atmosphere after 30 s etching time was also measured, as shown in Fig. 3e. The etched thickness increased with increasing RF power in the range from 100 W to 200 W. Above 200 W of RF power, the etched thickness was not proportional to the power and then saturated, becoming almost independent of power. The etching time in an O2 atmosphere was also varied at two different fixed power levels to optimize the etching conditions. The etched thickness at 100 W and 300 W of RF power, which are, respectively, the lowest and highest values studied, was measured as a function of the etching time as shown in Fig. 3f. The etched thickness was linearly proportional to the etching time (from 10 to 60 s) with an etch rate of 0.43 nm/s at 100 W RF power. The etch rate of 300 W RF power was 2.3 nm/s; this is a higher value than the etch rate at 100 W. Thus, the controllable conditions of RF power and etching time in an O2 atmosphere could be adjusted to produce the desired thickness of target materials. The properties of the dry-etched rGO thin films were further characterized by their Raman and XPS spectra, as shown in Figs S6–S8.
Finally, the detection of Aβ peptide was demonstrated to show the reliable characteristics of the rGO-based sensors that had been fabricated on 4-inch wafers. The reactions of Aβ antibody and Aβ were detected through changes of resistance in the rGO sensor. The change of resistance is defined by Equation (1),
where ∆R, R0, and Ra are the absolute values of resistance changes, the initial resistance of the rGO, and the resistance after immobilization of antibodies, respectively.
As shown in in Fig. 4a, the antibody for Aβ (6E10) was immobilized at carboxyl reaction sites of rGO by coupling with Ethyl-(dimethylaminopropyl) carbodiimide (EDC) and N-Hydroxysuccinimide (NHS). After the immobilization, the Aβ40 was reacted with the immobilized antibody on the rGO surface as an immunoassay reaction. The device used for this demonstration is shown in Fig. 4b. To verify the reproducibility of the stable immobilization of antibodies on the rGO surface, the measurement of resistance changes was repeated with 13 devices which were randomly selected from the same wafer as shown in Fig. 4c. The average value change in resistance [the red line in Fig. 4c] for 13 devices was 10.4 ± 1.4%. The values of each bar were acquired from a single device with 24 resistance-sensitive arms. Therefore, the data in Fig. 4c represent a total of 312 resistance measurements, which demonstrate the excellent uniformity of the response of the rGO sensors (from the same wafer) to immobilization of Aβ antibodies.
After the uniform immobilization of antibodies, the relative resistance changes (∆R/Ra, %) resulting from the reaction of Aβ40 with its antibody was plotted as a function of the Aβ40 concentration to characterize the sensor as shown in Fig. 4d,e. The resistances were calculated with Ohm’s law from measurement of current according to the applied voltages (from 2.0 V to 3.0 V) as shown in Fig. 4d. The absolute values of ∆R/Ra in Fig. 4e–g were derived by replacing R0 with Ra and Ra with RAβ in equation (1), where RAβ is the absolute resistance after the reaction with Aβ40.
The resistance change was proportional to the Aβ40 concentration in the range of 100 fg mL−1 to 100 pg mL−1 as shown in the black line of Fig. 4e. The limit of detection (LOD) of Aβ is approximately 100 fg mL−1, which shows that the rGO sensor has ultra-high sensitivity. The test of selectivity was also tried with the antibody of Prostate-specific antigen (PSA) as shown by the red line of Fig. 4e. The PSA antibody was also immobilized at rGO surface with the EDC and NHS coupling. After the immobilization of the PSA antibody, Aβ40 was also applied to react with the PSA antibody. The resistance changes had no discernible dependence on the concentration of Aβ40 in the case of the PSA antibody. Each point of Fig. 4e was acquired from 12 arms of rGO, resulting in consumption of 120 arms for the experiments; 60 arms were used for each of the conditions. To show the more practical usage of wafer scale-fabrication for biosensing applications, the response of the Aβ40 reaction with randomly selected devices in the same wafer was also measured as shown in Fig. 4f. The devices were from positions (3, c) and (5, b) which are relatively close to the middle and edge parts of the wafer, respectively. The percentage of ∆R/Ra was measured with both devices, and the values tracked each other very closely as the concentration of Aβ40 increased in the range from 100 fg mL−1 to 100 pg mL−1. The ∆R/Ra was measured by sequentially applying Aβ40 from lower to higher concentration. Next, we measured levels of human Aβ peptides in mouse plasma samples to show the reliable working of the rGO biosensor with real samples as shown in Fig. 4g. We collected blood plasma of 7-month-old male transgenic (TG, n = 4) mice, with human Aβ peptides, as an Alzheimer patient model; their age-matched male wild type (WT, n = 3) mice were used as a control sample. The number of consumed devices per individual mouse was 1, resulting in utilizing 7 devices and 168 arms. The plasma of each mouse was reacted with immobilized antibody the same as in the previous recombinant Aβ reaction. Each group of mice has a different value of resistance changes and the test discriminated with high significant value (p = 0.0452) as shown in Fig. 4g.
Conclusion
In summary, the effective wafer-scale rGO patterning was accomplished by adapting the highly reliable and reproducible dry etching techniques that are generally used in MEMS or semiconductor fabrication method for biosensors. The gold electrode was also aligned and formed on the rGO patterning with conventional photolithography and lift-off, thanks to the well-formed rGO layer created by the MDD method. Most existing rGO-based devices have an rGO layer only at the outermost layer after forming the other structures, such as electrodes. These results also show that our method can be applicable in various fields that need subsequent fabrication steps after forming the rGO patterns. The resistance changes by the reaction of Aβ40 and Aβ antibody in rGO sensors, which were fabricated at wafer-scale, were also reproducible because of the reliable rGO patterning method with ultra-high sensitivity (LOD ≈ 100 fg mL−1). This level of performance is sufficient to permit the utilization of the sensor in the diagnosis of Alzheimer’s disease19. We hope that widespread usage of graphene-based biomolecule detectors will become feasible with our fabrication method. For that, we would also try to acquire more intense reactions by treatment of the rGO and then be able to detect biomolecules from actual clinical samples in biofluids such as human blood and urine which is the one of the most convenient media for diagnosis.
Methods
Deposition of rGO thin films on 4-inch wafers
Detailed information is described in the Supporting Information section in Fig. S1.
Patterning of rGO
The PR (GXR601, AZ electronic materials) was deposited by spin coating at 3000 rpm for 30 s on the rGO film layer that had been formed by the MDD method to obtain approximately 1 μm of thickness. Photolithography was performed, using a commercial aligner (MA6 Mask aligner, Karl Suss) with 15 mW mm−2 of UV light exposure for forming the sacrificial pattern of PR. The coated substrate was pre-baked at 100 °C for 90 s before developing with developer (AZ® 300 MIF, AZ electronic materials). Hard-baking at 115 °C for 60 s was also utilized to ensure a robust sacrificial layer during the etching process. RIE (Advanced Vacuum & STS, VISION 320 RIE) was applied in an O2 atmosphere with the RF power and time optimized according to the expected etched thickness to produce the rGO patterns on the SiO2 substrate. After the etching, the remaining PR was removed with acetone, and the substrate was sequentially washed with methanol, isopropyl alcohol, and deionized (DI) water. The etched thickness was measured with an AFM (Park systems, XE-100, Korea).
Fabrication of rGO biosensor
Detailed information is introduced in the Supporting Information section in Fig. S3.
Electrical measurements
The sheet resistance of the rGO films was measured by the four-point-probe method with a Keithley 2400. The resistance changes were recorded with a semiconductor analyzer (Agilent, B1500A) and probe station (Unitek corp., AP-8000, Korea).
Antibody immobilizations and Aβ40 reaction
Both 6E10 monoclonal antibody (Covance, USA, 1 mg mL−1) for Aβ and PSA antibody (Fitzgerald, USA, 1 mg mL−1) were diluted to a concentration of 10 μg mL−1 with 0.1X phosphate-buffered saline (PBS) solution. After the dilution, each antibody was immobilized at the rGO surface with 2 mM of EDC and 8 mM of NHS for 2 h. Washing for removing residual materials was accomplished with 0.1X PBS and DI water. The recombinant Aβ40, which has 1–40 amino acids was diluted in 10 mM PBS buffer to get a target concentration from 100 fg mL−1 to 1 ng mL−1. The diluted recombinant Aβ40 was dropped on the antibody-immobilized rGO surface and reacted for 20 min. After the reaction, the surface was washed with PBS buffer and DI water.
Preparation of transgenic mice
Double APP/PS1 transgenic mice (TG, strain name: B6C3-Tg (APPswe, PSEN1De9) 85Dbo/J) and their WT mice (B6C3F1) were purchased from Jackson Laboratory (Bar Harbor, Maine, USA). Expression of both genes, APP and PSEN1, was confirmed before the experiment began via a PCR instrument from Bio-Rad (S1000 Thermal-Cycler) using the standard PCR condition from Jackson Laboratory, the PCR-remix provided by Cosmo-Genetech (G-taq PCR premix kit, CMT-6002), and DNA from mice tails. The subject male mice (TG and WT; n = 4 and n = 3, respectively) were 7 months of age at the time of blood collection, and they were housed in a facility under controlled temperature, with an alternating 12 h light-dark cycle and access to food and water ad libitum. All animal experiments were carried out in accordance with the National Institutes of Health guide for the care and use of laboratory animals (NIH Publications No. 8023, revised 1978). The animal studies were approved by the Institutional Animal Care and Use Committee of Korea Institute of Science and Technology.
Additional Information
How to cite this article: Kim, J. et al. Wafer-scale high-resolution patterning of reduced graphene oxide films for detection of low concentration biomarkers in plasma. Sci. Rep. 6, 31276; doi: 10.1038/srep31276 (2016).
References
Lee, C. et al. Measurement of the elastic properties and intrinsic strength of monolayer graphene. Science 321, 385–358 (2008).
De, S. & Coleman, J. N. Are There Fundamental Limitations on the Sheet Resistance and Transmittance of Thin Graphene Films? ACS Nano 4, 2713–2720 (2010).
Bae, S. et al. Roll-to-roll production of 30-inch graphene films for transparent electrodes. Nat. Nanotechnol. 5, 574–578 (2010).
Wang, Y. et al. Large area, continuous, few-layered graphene as anodes in organic photovoltaic devices. Appl. Phys. Lett. 95, 063302 (2009).
Han, M. Y. et al. Energy Band-Gap Engineering of Graphene Nanoribbons. Phy. Rev. Lett. 98, 206805 (2007).
Stankovich, S. et al. Graphene-based composite materials. Nature 442, 282–286 (2006).
Chen, W. et al. Preparation of graphene by the rapid and mild thermal reduction of graphene oxide induced by microwaves. Carbon 48, 1146–1152 (2010).
Wang, X., Zhi, L. & Müllen, K. Transparent, Conductive Graphene Electrodes for Dye-Sensitized Solar Cells. Nano Lett. 8, 323–327 (2008).
Myung, S. et al. Graphene-Encapsulated Nanoparticle-Based Biosensor for the Selective Detection of Cancer Biomarkers. Adv. Materials 23, 2221–2225 (2011).
Huang, Y. et al. Nanoelectronic biosensors based on CVD grown graphene. Nanoscale 2, 1485–1488 (2010).
Mao, S. et al. Specific Protein Detection Using Thermally Reduced Graphene Oxide Sheet Decorated with Gold Nanoparticle-Antibody Conjugates. Adv. Materials 22, 3521–3526 (2010).
Fang, L. et al. A single-use, disposable iridium-modified electrochemical biosensor for fructosyl valine for the glycoslated hemoglobin detection. Sensors and Actuators B: Chemical 137, 235–238 (2009).
Chen, P. et al. A high selective disposable biosensor based on screen-printed technique with two working electrodes for eliminating interference signals. Sensors and Actuators B: Chemical 183, 589–593 (2013).
Zhou, Y. et al. Microstructuring of graphene oxide nanosheets using direct laser writing. Adv. Materials 21, 67–71 (2009).
Zhou, Y. et al. Direct imprinting of microcircuits on graphene oxides film by femtosecond laser reduction. Nano Today 5, 15–20 (2010).
Zhou, M. et al. Controlled Synthesis of Large-Area and Patterned Electrochemically Reduced Graphene Oxide Films. Chem. Eur. J. 15, 6116–6120 (2009).
Ko, Y. U. et al. Microlitre scale solution processing for controlled, rapid fabrication of chemically derived graphene thin films. J. Mater. Chem. 22, 3606–3613 (2012).
N. H. Kim et al. Surface Energy Engineered, High-Resolution Micropatterning of Solution-Processed Reduced Graphene Oxide Thin Films. Adv. Materials 25, 894–898 (2013).
Kim, H. Y. et al. EPPS rescues hippocampus-dependent cognitive deficits in APP/PS1 mice by disaggregation of amyloid-β oligomers and plaques. Nat. Comm. 6, 8997 (2015).
Schmechel, D. E. et al. Increased amyloid beta-peptide deposition in cerebral cortex as a consequence of apolipoprotein E genotype in late-onset Alzheimer disease. Proc. Natl. Acad. Sci. 90, 9649–9653 (1993).
Shoji, M. et al. Production of the Alzheimer amyloid beta protein by normal proteolytic processing. Science 258, 126–129 (1992).
Kurkina, T. et al. Self-Assembled Electrical Biodetector Based on Reduced Graphene Oxide. ACS Nano. 6, 5514–5520 (2012).
Acknowledgements
The authors are grateful for financial support from the KIST institutional Program (2V04870 & 2E26180) and Korea Health Industry Development Institute (KHIDI, HI14C3319). S.M. Lee was supported by the Chung-Ang University Freshman Academic Record Excellent Scholarship Grants in 2015.
Author information
Authors and Affiliations
Contributions
J.K., M.-S.C., S.M.L., D.J., B.C.L., J.H.L., S.T.C. and K.S.H. conceived the idea and Y.S.K. served the biological advice. J.K., M.-S.C. and D.J. performed most of the experiments, contributed to experimental design, data analysis and writing. S.T.C. and K.S.H. designed and supervised the study, analyzed the data, and wrote the manuscript. All authors discussed the results and commented on the manuscript.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing financial interests.
Supplementary information
Rights and permissions
This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/
About this article
Cite this article
Kim, J., Chae, MS., Lee, S. et al. Wafer-scale high-resolution patterning of reduced graphene oxide films for detection of low concentration biomarkers in plasma. Sci Rep 6, 31276 (2016). https://doi.org/10.1038/srep31276
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/srep31276
This article is cited by
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.