Introduction

Interactions between high energy and high intensity lasers with matter produce particle flux and electromagnetic radiation over a wide range of energy1,2. The generation of transient fields of very high intensity in the radiofrequency-microwave regime has been observed for femtosecond to nanosecond laser pulses with 1011–1020 W/cm2 intensity, on both conductive and dielectric targets3,4,5,6,7,8,9,10,11,12,13,14,15,16,17,18,19,20,21. These fields have a bandwidth of several GHz and last for hundreds of nanoseconds. They often cause saturation and damage to the electronic equipment inside and near the experimental chamber, but on the other hand can also become an effective diagnostic tool. For these reasons, the absolute characterization of EMPs is of great importance and nowadays a very hot topic for present and future plants for laser-plasma acceleration (PETAL22, ELI23,24,…) and for inertial-confinement-fusion (NIF6,12,13, LMJ22,…). Once the sources will be well understood, laser-plasma interaction could be used also as powerful generator of transient RF-microwave fields, volumetrically distributed with respect to the point of emission, for many applications as interaction with materials and devices, EMP-radiation-hardening and electromagnetic compatibility studies.

EMPs have been previously characterized with conductive probes3,4,5,6,7,8,9,10,11,12,13,14,15,16,17,18,19,20,21, but only a few works specifically address the absolute measurements of the associated electric fields, which are accomplished by D-dot and B-dot probes (sensitive to and respectively)25,26. This is mainly due to the difficulties in performing reliable measurements with suitable signal-to-noise ratio (SNR). The conductive structure of these probes usually determines low measurement accuracy in near-field characterization27. This is a major problem in experimental chambers for laser-plasma-interaction (LPI) experiments, normally filled with metallic equipment and sets a limit on the probe minimum distance from the interaction locus. Conductive probes cannot be put in direct view of the plasma, because parasitic currents can be induced on them by ionizing radiation emitted from LPI (charged particles and UVs-Xrays by photoelectric effect) and can propagate to the oscilloscope heavily affecting the E-field measurement. EMPs may further induce spurious currents on the external conductor of the coaxial cables connecting probes to oscilloscope. Moreover, time integration is required to get electric fields from measurements; this greatly amplifies the low-frequency noise (where important part of the signal is) and requires suitable filtering6. measurements are even worst, since plane-wave approximation (inappropriate inside a conductive chamber) is used. In this work we describe the first time-resolved direct measurements, using the linear electro-optic (Pockels) effect in crystals28, of single vectorial components of EMP electric fields due to laser-plasma interactions on nanosecond regime, at laser intensities typical of inertial-confinement-fusion. This allowed us to perform, for the first time, measurements with the detector in direct view of the plasma and rather close to it.

Dielectric electro-optic (EO) probes have allowed reliable measurement of transient electric fields29,30,31,32,33,34,35, with many advantages versus classical conductive probes, in terms of field vectorial selectivity (one component measurement rejecting the orthogonal ones), bandwidth, dynamics, temporal and spatial resolution, low dimensions and invasiveness32,33,34,36,37. Indeed, their effective permittivity may induce a local perturbation on fields measured in vacuum. This remains much lower and localized than for conductive probes32,33,34 and with offline calibration can be estimated for de-embedding. The dielectric structure insures no parasitic currents even in these harsh environments. Their intrinsic low sensitivity is generally not detrimental when dealing with high intensity field measurements, as for EMPs. So far, they have been used in air, water and for atmospheric-pressure plasma jet sources32,34,38,39,40. Here we report about their application in vacuum and specifically in LPI.

Methodology

Electro-optic probe

Figure 1 shows the EO-probe structure32,41. Measurements of external electric fields are performed by detecting the change of polarization state, induced by electro-optic effect, of a continuous-wave laser probing beam having λp = 1550 nm and circular polarization (obtained with the λ/4 plate), propagating in a <111>-cut m Bi12SiO20 (BSO) crystal of 5 mm length. This is the polarization state modulation technique29,30,31,32,33,3442,43. BSO is intrinsically isotropic, becoming here birefringent because of the component of the external field, orthogonal to the laser wave-vector which is parallel to the <111> direction of the crystal. Actually, in our probe has been slightly misaligned with this direction, so that light coming from the first fiber can be collected by the second, after total reflection on the dielectric Bragg mirror. This double-passage increases the sensitivity and simplifies the probe mounting close to plasma. Elliptical polarization on the output probe beam is caused by the generated anisotropy; information on electric field is contained on: (1) induced dephasing between the two linearly-polarized components of the elliptical polarization , (2) orientation of eigendielectric axes with respect to the optical axis: ξ± = π (3 ± 1)/4 − αE/2, being αE the angle between and that axis31,42,43. In particular, it is 43, where r41 and n0 are the Electro-Optical coefficient of BSO crystal and its refractive index in absence of applied external E-field, respectively. At λp wavelength n0 = 2.405 and r41 has been measured for known external electric fields in the (10 Hz–2 GHz) range, as reported in ref. 44. As a brief summary of these measurements, at low frequencies (below 100 kHz) it resulted r41 = 5.4 ± 0.4 pm/V and at higher ones (above 10 MHz) r41 = 4.7 ± 0.2 pm/V, with more complex behaviour in the (100 kHz–10 MHz) range. Thus, by using a wave polarization analyzer - reported in detail in refs 42 and 43 −Δθ and ξ± can be measured and from them the two components of can be simultaneously determined29,30,31,32,33,3442,43. The laser probe beam has a double passage in the crystal and the electric-field is then averaged over a τ = 80 ps time frame (flying time of photons through 10 mm of BSO crystal). From this, it is possible to achieve the probe cutoff frequency29,32: fc = 0.443/τ = 5.5 GHz.

Figure 1
figure 1

Scheme of the EO-probe.

KapteosTM built a custom version of the EOP-P2R02-BS050 probe to adapt it to vacuum conditions used in the experimental chamber of ABC facility45,46. A polarization maintaining (PM) fiber is joined to the set of lens +λ/4 plate, followed by crystal and mirror (Fig. 1). An Alumina sleeve (30 mm length and 4 mm diameter) contains the whole structure. The custom probe was then enclosed in a 3 mm thick Teflon shield, having 4 mm and 10 mm internal and external diameters, respectively and protecting it from direct X-ray radiation coming from plasma. Shield permittivity εr-Teflon ~ 2.1 leads to partial impedance matching of the probe (effective permittivity εr-eff ~ 931,32,33,34) with surrounding vacuum, increasing the sensitivity. The overall configuration and the small bandwidth, centered at 1550 nm, of the detection system (tailored infrared fast photodiodes, PM fibers and optics) ensures effective rejection of main laser (λ0 = 1054 nm) stray light and of 400–700 nm light emission from scintillation of BSO by X-rays47,48,49. All components are dielectric and non-magnetic, with ultra-low loss tangent; the probe is insensitive to magnetic fields up to more than 3 T36, three orders higher than those expected in this experiment in the probe position. Optical fiber connection allowed the electronics to be placed suitably far from experiment, eliminating their free-space-coupling with EMP. The sensitivity of complete setup was preserved by using two tailored vacuum optical-feedthroughs with minimum insertion loss. The unit eoSense HF-2A-09L provided laser probe generation, monitored probe sensitivity along time, executed demodulation and determination of the EX’ and EY’ components of Electro-optic effect occurs on femtosecond time-scales, leading to intrinsic bandwidths exceeding 10 THz, with fmin in the kilohertz range32. Indeed, system bandwidths are limited by round-trip time of laser through the crystal (fc) and electronics frequency cutoffs. To improve the SNR, dedicated low-noise-amplifiers were used before the 3.5 GHz Lecroy-735Zi oscilloscope and this determined fmax ~ 0.5 GHz.

The electro-optic probe was offline calibrated using a Transverse ElectroMagnetic (TEM) cell50,51. This is a closed waveguide having 50 Ω characteristic impedance, with core basically constituted by parallel conductive plates where a TEM electromagnetic wave is propagating. A uniform electric field was thus applied to the electro-optic probe placed in the centre of the cell for calibration purposes. The field strength is simply given by the potential difference between the plates divided by the distance separating them and in our case anyway previously calibrated. The modulus of the scattering parameter S11 (reflection coefficient) of the TEM cell was verified to be lower than −20 dB when used with a 50 Ω termination load, with or without the presence of the EO-probe inside. It means that the traveling-wave regime was insured when calibrating the electro-optic probe and local wave-reflections due to the BSO/Alumina/Teflon/air interfaces did not modify remarkably the electric field at small distances from the probe. This guaranteed a correct and precise calibration procedure of the EO-probe up to the cutoff frequency of the TEM cell (6 GHz), providing accurate data on orientation of the crystal axes, measurement dynamics (>120 dB Hz0.5), intrinsic sensitivity (<20 kV/m for single shot pulses), vectorial selectivity (>40 dB) and spatial resolution (<5 mm), whereas the field sign was still undetermined. The calibrated probe is thus able to measure external electric fields with ±30% accuracy, much better than for classical conductive probes. The system is intrinsically stable for temperature variation and this was verified during the LPI experiments.

Results

Experimental measurements

Experiments have been performed with ABC laser, a Nd:phosphate-glass nanosecond facility45,46. One circularly polarized beam of 20–30 J, with FWHM ~ 3 ns, fundamental wavelength λ0 and 10-5 contrast, was focused by a F/1 lens up to 50 μm diameter, leading to ~(0.3–0.5) PW/cm2 intensity, for normal incidence on a thick (1710–1790 mm) Al target with ~40 mm2 plain surface (Fig. 2). For each shot the LPI was monitored by a large number of diagnostics45,46. Thermal ion emission from plasma was measured by Time-of-Flight (TOF) detection with a set of faraday-cups; a particle contribution with Eion/A ~ 1 keV (A = atomic number) is shown in Fig. 3a at φ = 53° from the target normal. As evident, strong coupling with EMP oscillations conceals detection of particles with Eion/A ≥ 4 keV. The problem is overcome with TOF monocrystalline diamond detectors52,53, fabricated by Università di Tor Vergata with microstrip surface-interdigital connections and high hardness to EMPs. In Fig. 3b signal measured by a diamond placed at φ = 65° is shown in the electron-energy domain. The huge peak on the right is due to X-rays; on its left falling edge a clear indication of fast electrons is present, with ~26 keV peak energy and ~40% FWHM. Part of this hot electron population is due to resonant absorption1,54,55, because of highly focused intensity (even higher in some regions because of self-focusing), large focusing angle (F/1 lens) and laser circular polarization. The remaining population is due to two-plasmon-decay instability1,54,56, also generating the 2λ0/3 = 703 nm harmonic54,56. In Fig. 3d the visible light spectrum detected by the Ocean Optics HR4000 spectrometer is shown, with the clear indication of components at λ0, 2λ0/3 and λ0/2 (527 nm). Hot electrons escaping from plasma create an electrostatic Debye sheath which accelerates ions2. On the same diamond signal the trace of ~20 keV fast protons (hydrogen is always present on target surfaces as impurity2) was in fact detected (Fig. 3c), together with the 1 keV thermal peak measured by the faraday-cup.

Figure 2
figure 2

Scheme of the experiment in the two configurations represented by the shots #1590 and #1597.

Figure 3
figure 3

Measurement of particles and of visible light spectrum.

(a) Faraday-cup at 53°; diamond detector at 65°: (b) electron and (c) ion energy domain; (d) Optical spectrometer. A = atomic mass number.

We performed two series of measurements, both with EO-probe in direct view of the target and at 85 mm distance. For the first series the probe was mounted on the xy plane, as in configuration indicated for shot #1590 in Fig. 2, with φp = 70°, longitudinal axis parallel to in cylindrical coordinates and Bragg mirror toward the target. In the whole campaign we were able to measure the component of but not the one, because of technical reasons: lower sensitivity to this component and increased background noise of the associated channel amplifier. In particular, it was (Fig. 2) , being γ = 81° ± 3°. In Fig. 4a,b the measured E1X′ component is shown for shot #1590 of this first series, in two different time scales; the axis origin was chosen at the beginning of the first intense peak. The signal appears to be constituted by two spectral components, as also shown in Fig. 5. The main contribution lasts for the total signal duration (~1 μs) and concerns frequencies below ~50 MHz. The second affects only the first 300 ns and contains higher frequencies. Maxima higher than 100 kV/m are present during the first 250 ns. Background noise is visible for t < 0, leading to very good SNRs for each shot. In the whole campaign it was not possible to determine the absolute field phase (i.e. sign). For this #1590 shot, a first high ‘positive’ peak is present (FWHM = 6.7 ns) and then a ‘negative’ large peak at ~40 ns. Thus, a rather sharp one is at ~80 ns, where the field reaches its maximum value: From plane-wave approximation and on the hypothesis of isotropic emission with respect to target, the overall EMP energy is thus estimated ~0.46 J, ~1.8% of the laser one.

Figure 4
figure 4

Measured EX′ field component for shot #1590 (a,b, in two different time scales) and #1597 (c).

Figure 5
figure 5

Time and frequency domain of signals of EO-probe and SWB antenna for shot #1590.

A superwideband (SWB) antenna was also used at ~48 cm from the target14,15,16,57, in a region of the chamber close to the surface and well protected against direct plasma radiation by thick conductive objects. Results in time and frequency domain are compared in Fig. 5 with those from EO-probe. The SWB signal duration is lower, whereas its main spectral contributions near ~130 MHz and ~200 MHz find correspondence in the EO signal. The high-pass behaviour of the antenna was expected because it is sensitive to the time derivative of the electric field and because of its 50 Ω impedance matching, worst at the smallest frequencies57. Moreover, the antenna is placed in a section of the vacuum chamber having transversal dimension much lower than the chamber radius. So, only the EMP components at higher frequency can reasonably penetrate that small space. The antenna is also close enough to the chamber surface, where quartz windows and many vacuum radiofrequency-feedthroughs are present. As observed in references15,16, these are source of coupling of EMPs to the outside of the vacuum chamber. This energy leakage causes local electromagnetic modes excited in this section of the chamber to have low Q values and then related fall-time behaviour faster than for modes excited in the chamber core15,16. We could suppose that the spectral content of the EO signal for f < 50 MHz, not revealed by the SWB antenna, might be due to direct plasma radiation (X-rays and particles), exciting these components only in the chamber core.

In the second series of measurements, probe was rotated of 90° clockwise with respect to the z axis (shot #1597 in Fig. 2), so . In Fig. 4c the related E2X′ component is shown. Presence and time-duration of the two spectral contributions observed for shot #1590 are confirmed here, too. There is still a high sharp ‘positive’ peak (correspondent to that of shot #1590) having FWHM = 5.4 ns, leading to Then, the field slowly decreases, changing sign at ~30 ns. There are some large oscillations, decreasing below 100 keV/m after ~80 ns from the main peak. The time evolution for t > 300 ns is similar to shot #1590 and then not reported. In this case the EMP energy is ~0.54 J, ~2.2% of laser one.

Particle-in-cell simulations

To have better insight on possible origins of signals observed we performed simplified particle-in-cell (PIC) simulations of the experiment, by CST Particle Studio solver. Space-charge effects were considered, together with secondary electron emission from Teflon58 and superficial charge deposition on surfaces. Considering Fig. 2, the simulated box dimensions were 150 mm, 400 mm and 200 mm along x, y and z axes, respectively. Absorbing boundary conditions (free-space) were set on each box surface and adaptive meshing refinement was performed for a total number of about 254000 mesh cells. Minimum and maximum mesh step size were 0.85 mm and 5.15 mm respectively and a time step of 2.2 ps was used. The target surface was source of conical particle flows, uniform within their angle of emission φt to the target normal. We considered a Gaussian-shaped electron bunch with σ = 3 ns, 26 keV of peak energy and 40% energy spread, as estimated from diamond measurements (Fig. 3b). An equal and synchronized bunch of protons was added, modeling the fast-ion component. In Fig. 6a,b we compare, in time domain, EO-probe measurements with results of simulations for φt = 60°. The first peak of measured E1X′ can be effectively associated with the simulated fast-electron peak, whereas simulated fast ions can be associated with the following measurement decrease, having minimum at ~40 ns. Later oscillations might be associated with quasi-neutral thermal components, not considered in these preliminary simple calculations. In Fig. 6c,d the same comparison is shown for the related Fourier spectra. The simulated configuration is very simple and does not allow to get a remarkable fitting with experimental data; however, calculations indicate that the low-frequency components of measured spectra are compatible with fields generated by particle flows. Furthermore, this might confirm that the classical high-frequency components, also showed in the spectrum of the SWB-antenna signal (see Fig. 5), are likely cavity modes excited within the experimental chamber4,15,16.

Figure 6
figure 6

Measurement of E1X′ in shot #1590 with related simulations for φt = 60°: (a) time domain and (c) Fourier spectrum. The same for E2X′ in shot #1597: (b) time domain and (d) Fourier spectrum.

Future and more accurate modelings of the experiment have to consider photoionization due to X-rays from plasma, generating a cloud of cold electrons around the external surface of the Teflon. This is expected to create a pulsed electric field, rather synchronous with the peak due to fast electrons. Effects due to charge implantation on Teflon have to be taken carefully into account, too. These phenomena help to explain the absence of the first peak of measured E2X′ in these preliminary simulations and the higher intensity of measured E1X′ peak. In Fig. 7 we show simulations of E1−2X′ versus φt angle; when φt > 70° particles hit the probe. E1X′ (~Eφ) decreases with φt and this is due to anisotropic particle emission from target with respect to the probe. Indeed, Eφ became negligible in simulations with the EO-probe on the symmetry axis of the particle flow. On the other hand, E2X′ (~Er) is not particularly affected by this (Fig. 7b).

Figure 7
figure 7

Simulations for E1X′ (a) and E2X′ (b) for different φt.

Conclusions

We have described the first direct electric-field measurements of EMPs generated by laser-target interaction in ICF regime, performed by dielectric EO-probes directly viewing the plasma. A maximum electric field of 261 kV/m was measured, about two orders of magnitude higher than previous measurements in nanosecond lasers with much higher energy (LULI, OMEGA and NIF6,8,10,12,13,18,20,21) and comparable with maximum values reached with picosecond lasers18,20,21.

The analysis of measurements and preliminary simulations indicates that EMP signals are compatible with the emission of TOF-detected charged particles10. It also shows that EMP should be affected by anisotropic particle emission from target, X-ray photoionization and charge implantation on surfaces directly exposed to plasma. In experiments with a femtosecond laser of ~100 mJ (with moderate plasma ionizing-radiation) the main EMP contribution resulted from neutralization-currents flowing through the target holder18,21. In nanosecond facilities high intensity neutralization-currents were measured59, but it has been shown that their contribution to EMP is small18,21. Figure 5 suggests that here this could be associated with the main spectral components ~130 MHz and ~200 MHz detected by both EO-probe and SWB antenna.

The performed measurements allowed to get important information on laser-plasma interaction. Future experiments of this type and tailored numerical studies will be important for the better understanding of EMP sources in different regimes, key point of future plants for laser-plasma acceleration and for inertial-confinement-fusion, as well as for the use of EMPs as effective plasma diagnostics. This also opens the significant application of laser-plasma interaction as powerful source of transient RF-microwave fields of high intensity, volumetrically distributed with respect to the point of emission, for interaction with materials and devices, EMP-radiation-hardening and electromagnetic compatibility studies. Moreover the demonstration that electro-optic effect can be an extremely effective method for detecting electric fields in laser-plasma context, leads to a great potential for characterization of the intrinsic LPI and of the generated Terahertz radiation.