Introduction

In 2001, carbon nanotube film was used to fabricate a double-electrode discharge gas sensor in our lab1, which triggered an explosion of research on carbon nanotube based gas sensors1,2,3,4,5,6,7,8,9,10,11,12,13. Unfortunately, these gas sensors exhibited multi-valued sensitivity of breakdown voltage or discharge current in response to targeted gas component at 1 atm total operating pressure, therefore their applications are limited because a single-valued sensitivity14,15,16,17,18,19,20 to detection target is required for gas sensors. Additionally, the sensors in this configuration of double-electrode operated in self-sustaining discharge states, in which positive ions heavily bombard the cathode comprised of either nanotubes or a metal plate. In case of either a carbon nanotube cathode or a nanotube anode is used, current density can rise to a considerably large value around 100 A/m2, resulting in nanotube film damage21 and a short operating life1,2,3,4,5,6,7,8,9,10,11.

A tripolar electrode configuration has been reported7, however, it still showed multi-valued sensitivity to acetone and ethanol because the second electrode collected most positive ions, resulting in current passing the first electrode of nanotube anode larger than that passing through the third electrode (sampling electrode) by 2 orders and also causing quick damage of the nanotube electrode. Here we report a triple-electrode structure with single-valued sensitivity and longer electrode life. The structure used a carbon nanotube array as the first electrode, featuring two electric fields with directions opposite each other and is highly sensitive to various gases and capable of the detection directly from gas mixtures without separation in a wide range of the concentrations from 0 to 100%, as well as capable of monitoring temperature in a range of 17–120°C at the same time. In comparison to other existing sensors22,23,24,25,26,27,28, our design integrated gas and temperature detection into several tripolar electrode structures, featuring distinct interelectrode separations, capable of the operation at low power. Potentially, this could lead to a miniaturized gas ionization sensor with capability of multi-functional detections.

Results

A triple-electrode structure is designed for gas detection at 1 atm, which is comprised of a carbon nanotube array cathode, an extracting electrode and a collecting electrode (Fig. 1a and Supplementary Fig. S1). The multiwalled nanotube (MWNT) film (Fig. 1b) was grown by thermal chemical vapor deposition29,30 (TCVD) on one side of the Si cathode covered by a Ti/Ni/Au film. Nanotubes in the film are 50 nm in diameter and 5–6 μm in length. The distance between nanotubes is 50–100 nm. Ti/Ni/Au film was also sputtered on both sides of the extracting electrode (Fig. 1a, c) and the inner side of the collecting electrode for increasing initial discharge current. The sensor array was placed in a detection chamber with an electrical input and output interface and air was pumped out of the chamber to establish a low vacuum with a gas pressure of 5 kPa. The gas to be detected was then introduced in a controlled fashion into the chamber until pressure is increased to 1 atm. The sensor displays very fast response within around the order of 10−6 s31 in comparison to other sensors2,31,32 due to its non-self-sustaining discharge state.

Figure 1
figure 1

The triple-electrode sensor device.

(a) Exploded view of triple-electrode showing MWNT film used as the cathode. (b) SEM micrograph of a TCVD-grown, vertically aligned MWNT film. (c) Schematic diagram of the test system. Ue is higher than Uc, which offers two electric fields E1 and E2 in reversed field direction. (d) Calculated potential schematic and its effect on ion diffusion and migration. Positive ions, generated by discharge (α process) between the cathode and the extracting electrode, diffuse to the extracting electrode and then move to the collecting electrode due to acceleration by the electric field E2 and thereby form the positive ionic current Ic.

During operation, suitable d.c. voltages of Ue and Uc are applied between the cathode and the extracting electrode and between the cathode and the collecting electrode (Fig. 1c), respectively. Ue is higher than Uc, which generates two electric fields E1 and E2 in reversed field direction. The field near the nanotube tips is much higher than that in the other region. Consequently, ionization takes place near the tips in around 1 μm range (reference). Through collision and diffusion, significant amount of positive ions overcome the weak field near the extracting electrode and move into the region between the extracting and the collecting electrodes, then they are accelerated by E2 towards the collecting electrode (Fig. 1d and Supplementary Fig. S2A). Therefore bombardment of the positive ions on nanotubes is considerably reduced, which hindered occurrence of self-sustaining discharge, resulting in a significantly longer life-span of nanotube electrode and consequently a single-valued sensitivity.

The structure enables a non-self-sustaining discharge near the nanotube tips at lower voltage Ue of less than 150 V. Non-self-sustaining discharge current I is mainly determined by gas partial pressure, electrode separation and gas temperature; and the gas partial pressure is proportional to gas concentration. Therefore, at a given electrode separation, I is direct measurements of gas concentration and temperature. We found that distances between electrodes had a distinct effect on the detection. By varying the distances between electrodes, we were able to identify different gas species in a gas mixture without component separation and measure temperature at the same time. Single-valued sensitivity to several gases and mixtures were obtained under electric field above a critical value by control of electrode separation and applied voltages on electrodes. Single-valued sensitivity to gas temperature was also observed. We also found that temperature has a great effect on gas detection sensitivity, the higher gas temperature, the higher sensitivity.

Discussion

For all our tests, gases probed were mixed with N2. The gases with first ionization energies higher than 10.3 eV are harder to ionize than nitrogen31, but they affect energy and charge transformation33 during gas discharge in a mixture with N2. Since the first ionization energies of the gases probed are greater than 10.3 eV, the ionization process will depend mainly on the ionization of N2 from its two metastable states, and with lower excitation energies33,34,35,36,37 of 6.2 and 8.4 eV, respectively. The first ionization coefficient32 of N2, α, reflects ionization ability of gas from collision with electrons and is determined by the applied electric field E and the partial pressure P of N2 (Supplementary Fig. S3),

where A and B are constants related to gas species and temperature. When the electric field rises up to a critical value, α increases with increasing N2 partial pressure from 0 to 1 atm, displaying single-valued α−P curves. As collecting ionic current Ic is a part of total discharge current I, it increases with α in a non-self-sustaining discharge state32,

where I0 is initial current, d is electrode separation between nanotube cathode and extracting electrode. When the concentration of gas probed is increased in a gas mixture with N2, it reduces N2 concentration at the same time, therefore single-valued decrease of collecting current with increasing concentration of the gas probed can be observed at given electric field up to a critical value.

Gas ionization properties of the triple-electrode structure were tested for new insights of the mechanism of the detection. First, we measured the current Ie flowing through the nanotube cathode in C2H2 at 80 V Ue with 200 μm electrode separations (between cathode and extracting electrode and between extracting electrode and collecting electrode, in general we kept two separations the same). We found that Ie = 2Ic (Supplementary Fig. S4). Current density je passing the nanotubes can be calculated by dividing Ie with the cross-sectional area of all nanotubes. To understand the effect of gas concentration and electric field on gas conductivity, we tested NO in a concentration range of 0–690 ppm at various extracting voltages of Ue and 120 μm electrode separations. Gas conductivity of NO, calculated by dividing je with the average electric field between the cathode and the extracting electrode38, was observed that it remained constant with rising average electric field and monotonically decreased with increasing NO concentration (Supplementary Fig. S5A). The results indicate a single-valued sensitivity mechanism of our structure for NO detection. Furthermore we evaluate the effect of gas temperature and electric field on gas ionization. Current density je was found exponentially dependent on gas temperature and linearly dependent on average electric field in pure N2 at 200 μm separations (Supplementary Fig. S6A, C), which follows the field-assisted thermal emission law38. It showed a potential of the structure for detecting gas temperature. To study the effect of gas concentration on electron emission ability of carbon nanotubes which greatly contributed to gas discharge current, we conducted tests using two sensors with 100 μm and 120 μm electrode separations at 80 V Ue with NO concentrations from 0 to 700 ppm and at gas temperatures from 30 to 80°C (Supplementary Fig. S7A, B). Schottky barrier was calculated according to Eq. 3. We observed that slightly barrier increasing with NO concentration (Supplementary Fig. S7C, D) could induce large change of collecting current. The result showed the effect of gas concentration on emission ability of nanotubes and also sensing potential of our structure.

where A denotes the Richardson constant, A = 1.20173 × 106 A·m−2·K−2, K denotes the Boltzmann constant, K = 1.38065 × 10−23 J·K−1 and denotes the Schottky barrier. In contrast, double-electrode device displayed ruleless change of conductivity with both average electric field and NO concentration (Supplementary Fig. S5B), current density was also higher and up to 102 A/m2 order (Supplementary Fig. S6D). This arises from the secondary electron emission induced by heavy bombardment of the positive ions on the nanotube cathode. Higher current density resulted in shorter life-span of nanotubes and multi-valued sensitivity of double-electrode structure2,3,4,5,6,7,8,9,10,11.

To study single-valued sensitivity, we chose hydrogen (H2) as the test gas and controlled electric field strength by varying voltage and distance between electrodes. To find the critical field needed for single-valued sensitivity, we kept the distances between the cathode and the extracting electrode and between the extracting and the colleting electrodes constant at 200 μm, fixed Uc constant at 10 V (0 V at the cathode and 10 V at the collecting electrode), changed Ue from 40 to 80 V and carried out hydrogen detection at 1 atm chamber pressure. When Ue increased, current characteristics changed from non-monotonic to monotonic at 80 V (Fig. 2a). Therefore, for H2, the threshold value of Ue was found to be around 80 V at 200 μm separations. We also found threshold voltages and electrode separations to obtain single-valued sensitivities for other gases and mixtures, including C2H2, O2, C2H4, SO2, SO2/NO and H2/C2H2/C2H4 mixtures.

Figure 2
figure 2

Effect of gas concentration on positive ionic current.

(a) Single-valued decrease of positive ionic current namely collecting current with increasing H2 concentration at 200 μm separations and 80 V of Ue (black). Multi-valued sensitivities to hydrogen at 200 μm separations with Ue as 40 V (purple) and 60 V (red), respectively. (b) Single-valued decrease of collecting current with increasing C2H2 concentration at 200 μm separations and 80 V of Ue (black). Single-valued decrease of collecting current with increasing O2 concentration at 90 V of Ue (red), at 200 μm separation between the cathode and the extracting electrode and 100 μm separation between the extracting and the collecting electrodes. (c) Sensitivity curves Nos. 1 and 2 of 200 μm separation sensors to C2H2 at 80 V of Ue, exhibiting our finding on the reproducible single-valued characteristic of the sensors in C2H2.

Single-valued sensitivities were found to be distinct to different gases. This is because different gases have distinct physical and chemical properties. A sensor with 200 μm electrode separations exhibited the sensitivities of −8 pA/ppm to 151 ppm (1 ppm = 1 μL/L) for H2 (Fig. 2a) and of −533 pA/ppm to 1 ppm for C2H2 (Fig. 2b) at 80 V of Ue. Fig. 2b also shows current vs. concentration change for O2. And the sensitivity was −327 pA/10000 ppm (−0.03 pA/ppm) to 6% O2 within a wide range of concentration from 0 to 100% (at 1 atm chamber pressure). Distinct sensitivities were also found for gases of C2H4, NO, NO2 and SO2 (Supplementary Fig. S8). Fig. 2c displays two adjacent sensitivity curves for C2H2, exhibiting our finding on the reproducible characteristic of the sensors in C2H2. Based on these results, we anticipated that our sensor might have the potential to detect a gas component in a mixture.

To evaluate the potential of our sensor for detecting gas components in a gas mixture, we tested the effect of the electrode distance on sensitivity of the sensors to mixed gases at 150 V of Ue (Fig. 3a). For these tests, we also kept the distances between the cathode and the extracting electrode and between the extracting and the collecting electrodes the same. For a sensor with 75 μm electrode separations, we measured output currents larger than 12 nA and observed the largest change of 4.8 nA corresponding to four gas mixtures; but for a sensor with 120 μm separations, the output currents were lower than 4.2 nA with a change of 1.4 nA corresponding to the same gas mixtures, displaying smaller sensitivity. Therefore, we can choose different electrode separations and make the sensors having various distinct sensitivities to different gas components, exploiting identification and detection of gas component in a mixture.

Figure 3
figure 3

Effect of electrode separation on ionic current and detection of gas mixture.

(a) Various separations are given as 75 μm, 80 μm, 100 μm and 120 μm, respectively. Case 1 denotes a concentration in gas mixture of SO2 (280 ppm)/NO (1115 ppm)/NO2 (195 ppm), 2 of SO2 (736 ppm)/NO (1115 ppm)/NO2 (195 ppm), 3 of SO2 (280 ppm)/NO (296 ppm)/NO2 (218 ppm) and 4 of SO2 (280 ppm)/NO (818 ppm)/NO2 (218 ppm). Four cases are all mixed with 208798 ppm O2 in N2 ambient with 150 V Ue. It shows that various separations enable different collecting currents. (b) Two sensors with different separations constitute a sensor array for detecting two components of SO2 and NO in a mixture at 100 V Ue, with one sensor with 100 μm separations for NO detection and the other with 150 μm separations for SO2 detection. (c) Collecting current of the NO sensor decreases with increasing NO concentration at different constant SO2 concentrations at 100 μm separations, showing interference from SO2 component. (d) Collecting current of the SO2 sensor decreases with increasing SO2 concentration at different constant NO concentrations at 150 μm separations, showing interference from NO component to SO2 detection.

We chose two electrode separations for fabricating two sensors (Fig. 3b) and conducted tests for detecting two components of SO2 and NO in a mixture with N2 without component separation. One sensor with 100 μm separations detected NO in a 0–1128 ppm range and the other with 150 μm separations detected SO2 in a 0–738 ppm range. Collecting currents as single-valued functions of two-component concentrations were measured when Ue was increased to 100 V (Fig. 3c, d). Fig. 3c shows that the collecting current of the NO sensor decreased from 0.71 to 0.47 nA when maintaining constant NO concentration and increasing SO2 concentration from 141 to 738 ppm, showing interference from SO2 component. Similarly, Fig. 3d shows interference from NO component to the SO2 detection.

Equations (1) and (2) show that collecting current Ic is an exponential function of electrode separation d and component partial pressure32. When detection chamber volume is given, partial pressure is proportional to component concentration. Therefore the collecting current Ic in Fig. 3c and d is determined mainly by three factors of electrode separation, NO concentration and SO2 concentration , which can be described as follows:

where d1 = 100 μm, d2 = 150 μm, I1 and I2 are collecting current of the 100 μm and 150 μm separation sensors, respectively. The functions of f0, f1 and f2 in Eq. (4) describe the effect of d1, and on I1, respectively; g0, g1 and g2 in Eq. (5) describe the effect of d2, and on I2, respectively. a0, a1 and a2 are the coefficients of f0, f1 and f2, respectively; b0, b1 and b2 are the coefficients of g0, g1 and g2, respectively. From the equations (4) and (5), values of and can be solved. Therefore, by careful study of gas component interaction mechanism with electrodes and design of the sensor configuration, sensors capable for multi-component detection may be designed and concentrations of the two components could be obtained through detection followed by data processing according to the equations (4) and (5). Fig. 4 shows that this technique is also valid for direct detection of H2, C2H2 and C2H4 in a mixture by using three sensors with different electrode separations (Fig. 4). To C2H2, detection was sensitive down to 1 ppm. The results suggested that the sensor array with different electrode separations has potential for use in trace gas and gas component detections.

Figure 4
figure 4

(a) Detection of a H2/C2H2/C2H4 mixture by using three sensors at 70°C, 150 V Ue and 1 V Uc. (b) the first sensor with 75 μm separations for C2H2 test, showing detection ability of the sensor down to 1 ppm trace gas, (c) the second with 100 μm separations for H2 test and (d) the third with 120 μm separations for C2H4 test. The sensitivities of the three sensors are obviously different and collecting current monotonically decreases with increasing H2, C2H2 and C2H4 concentrations in the preset concentration ranges.

Two sensors with different electrode separations were fabricated to simultaneously detect 0–700 ppm NO and gas temperature at 80 V of Ue, respectively (Fig. 5a, b and Supplementary Fig. S7A, B). Tests were conducted while NO concentration increased from 0 to 700 ppm at various given temperature values in a range of 30–80°C. The sensor with 100 μm electrode separation detected NO in a range of 0–700 ppm and exhibits the highest sensitivity of −23 pA/ppm to 335 ppm NO (Supplementary Table S1). The other sensor with 120 μm separation detected gas temperature in a range of 30–80°C and exhibits the highest sensitivity of 1 nA/°C to 70°C (Supplementary Table S2). Fig. 5a and b also show the effect of gas temperature on NO detection and that of NO concentration on temperature detection, respectively. The highest cross sensitivity of the NO sensor to temperature is 2.0 × 10−2/°C (Supplementary Table S1), two orders higher than that of the temperature sensor to NO, −3.1 × 10−4 ppm−1 (Supplementary Table S2), indicating a stronger effect of temperature on NO detection than that of NO on temperature detection.

Figure 5
figure 5

Simultaneous detection of NO and gas temperature.

(a) Collecting current Ic dependence on NO concentration at 60°C, 70°C and 80°C, respectively, showing a greater effect of gas temperature T on NO detection, at 100 μm electrode separations and 80 V Ue. (b) Collecting current Ic dependence on gas temperature T at 119 μL/L, 335 μL/L and 700 μL/L of NO, respectively, showing a slight effect of NO on gas temperature detection, at 120 μm electrode separations and 80 V Ue.

Analysis on simultaneously detecting NO and gas temperature follows: Since collecting current Ic is an exponential function of electrode separation d, component partial pressure P, gas temperature T and electric field intensity E (Eqs. (1) and (2)), we can obtain an equation for Ic in dependent of d, and T at given voltages applied on three electrodes. In this case, the collecting current Ic is determined by the three factors of electrode separation d, NO concentration and gas temperature T, which can be described as follows:

where d1 = 100 μm, d2 = 120 μm, I1 and I2 are collecting current of the sensors with 100 μm and 120 μm electrode separation, respectively. The functions of f0, f1 and f2 in Eq. (6) is dependent of d1, and T, respectively; g0, g1 and g2 in Eq. (7) is dependent of d2, and T, respectively. a0, a1 and a2 in Eq. (6) are the coefficients of f0, f1 and f2, respectively; b0, b1 and b2 in Eq. (7) are the coefficients of g0, g1 and g2, respectively. These coefficients can be determined experimentally (Fig. 5a, b and Supplementary Fig. S7A, B). The values of and T can be obtained by solving Eqs. (6) and (7). Similarly, a proper data processing of the data shown on Fig. 5a and b could be conducted, resulting in detection of NO concentration and gas temperature at the same time. Moreover, three sensors with various distinct electrode separations were fabricated and used to simultaneously detect a SO2–NO mixture and gas temperature (Fig. 6). The sensor with a 100 μm electrode separation showed a highest sensitivity of −7.7 nA/ppm to 298 ppm NO (Supplementary Table S3), the sensor with a 75 μm electrode separation showed a highest sensitivity of −3.8 nA/ppm to 280 ppm SO2 (Supplementary Table S4) and the sensor with a 120 μm electrode separation showed a highest sensitivity of 83.2 nA/°C to 120°C (Supplementary Table S5). The highest cross sensitivity to temperature is one order higher than that to NO and SO2 (Supplementary Tables S3–5), respectively, indicating a greater effect of temperature on NO and SO2 detection than that of NO or SO2 concentration on temperature detection. A proper data processing of the data presented in Fig. 6 could also be conducted and could simultaneously detect the two components and gas temperature. Sensitivity of the sensor to gas temperature in pure N2 was also tested (Fig. 6d) and displayed a highest sensitivity value of 14.9 μA/°C to 100°C (Supplementary Table S6), considerably higher than that to gas concentration by three orders and higher than that of existing temperature measurement technology by one order26. The results further suggested that our sensor with distinct gaps can work as gas sensor and temperature sensor at the same time, leading to the development of integrated sensing devices with capability for temperature monitoring and concentration detection at the same time. Additionally, our sensor is superior to the existing temperature measurement technology26,27,28.

Figure 6
figure 6

Simultaneous detection of a NO−SO2 mixture and gas temperature by using three sensors at 150 V Ue and 10 V Uc and temperature sensitivity test.

(a) The first sensor with 75 μm separations for SO2 test, (b) the second with 100 μm separations for NO test and (c) the third with 120 μm separations for gas temperature test. The sensitivities of the three sensors are distinctly different. Collecting currents I1 and I2 of the sensors in (a) and (b) monotonically decrease with increasing SO2 and NO concentration, respectively; nevertheless I3 in (c) increases with gas temperature in the preset range, showing a considerably higher sensitivity of the sensor to gas temperature. (d) Temperature sensitivity of our sensor at 200 μm separation between the cathode and the extracting electrode, 220 μm separation between the extracting and the collecting electrodes and at 100 V Ue in pure N2.

Our study overcame a key barrier of double-electrode and tripolar electrode ionic gas sensors, that is multi-valued sensitivity2,3,4,5,6,7,8,9,10,11 to gas concentration at 1 atm operating pressure. Through design of the triple-electrode gas sensor and careful control of the distances from the extracting electrode to the cathode and the collecting electrode as well as the field direction and the voltages between them, we were able to significantly reduce ion bombardment on the carbon nanotube cathode, hence increase field strength at lower voltage near the tips without quick tip damaging, which leads to a single-valued sensitivity in a wide range. We have demonstrated multi-gas detection as well as simultaneously gas and temperature detection capability and that our design is superior to existing commercial gas and temperature sensors22,23,24,25,26,27,28. With further development, we believe this work could lead to miniaturized, multi-functional sensing devices.

Methods

Sensor fabrication

The triple-electrode sensor using carbon nanotube array was fabricated by using Microelectro Mechanical Systems (MEMS) technology. The silicon slices (500 μm thick, 27 mm in length, 8 mm in width) were cut as the substrates of cathode, extracting electrode and collecting electrode, respectively. They were processed through photolithography and dry etching, resulting in the cathode with two round holes, the extracting electrode with one round hole and the collecting electrode with a 200 μm deep square blind hole. 50 nm thick Ti film, 400 nm thick Ni film and 125 nm thick Au film were deposited in sequence by magnetron sputtering on both sides of extracting electrode (Fig. 1a, c) and the inner side of cathode and collecting electrode. Then the three electrodes were rapidly annealed (Supplementary Fig. S1A–C) in a low vacuum of 3.0 Pa at 450°C for 50 seconds to alloy the Ti/Ni/Au film. On the annealed Ti/Ni/Au film of cathode, vertically aligned multiwalled nanotube (MWNT) array (Supplementary Fig. S1D) was grown by thermal chemical vapor deposition (TCVD) method. Polyester film was cut as the insulating strips between electrodes with 2-mm width and various given thickness which corresponds to various electrode separation of the device. In the final step golden wires were bonded on the edge of electrodes and the three electrodes were bonded together with insulating glue and finally, the sensor device (Supplementary Fig. S1E) was fabricated.

Gas mixture preparation and temperature control system

An automatic gas mixing system was consisted of an operation platform based on a computer, a gas mixing chamber and a gas and temperature detection chamber. There were five gas cylinders containing high purity gas with concentration accuracy of 2%. One of them supplied the environment gas N2 and the others supplied the measured gas. Five mass flow controllers (MFC, Line Tech M3030V with 1% accuracy) were used to continuously regulate gas flux for preparing mixed gas concentrations. The full scales of the five MFCs were two with 50 mL/min, one with 100 mL/min, one with 500 mL/min and one with 1000 mL/min. 2.5% accuracy of the gas concentration in two/three/four component mixtures was obtained through calculation on the basis of 1% accuracy of MFC and 2% accuracy of source gas. After well mixed in a gas mixing chamber, gas mixture flowed into a sealable stainless steel detection chamber with a pressure meter to detect gas pressure. A temperature control device was used to control gas temperature and the temperature can be increased to the set value within 1 minute. A vacuum pump was equipped for pumping gas out of the detection chamber and quick gas exchange.

Electrical test system

A PXI bus test system was constructed to provide high stable powers for the sensors and to precisely measure the collecting currents from pA to nA. The test system based on PXI bus was consisted of a 14-Slot Chassis (NI PXI-1044), a controller module (NI PXI-ExpressCard8360), five digital multimeter modules (NI PXI-4071) and three power modules (NI PXI-4132). The three power modules provided voltages needed for the electrodes (Fig. 1c). Two of the power modules were connected in series to provide 0–150 V voltage for extracting electrodes of the sensors and the other power module provided 0–100 V voltage for collecting electrodes. The digital multimeter modules with shielded cables are capable of acquiring the collecting currents from pA to nA for five sensors at the same time. All acquired data were recorded to a computer through a LabVIEW program linked to MXI interface.

Sensor measurements

The sensor array was placed in a detection chamber and wires bonded to electrodes were extended outside through an interface. Sensors with distinct gaps in the array simultaneously detected different physical parameters such as concentration of different gas species and gas temperature. Measurements were conducted with increasing gas concentration in a preset range at various temperatures from 17 to 120°C. Before the measurements, temperature was controlled at a set value and then air was pumped out of the chamber to establish a low vacuum with a gas pressure of 5 kPa, later the measured gases and the environment gas N2 were introduced in a controlled fashion until chamber pressure is increased to 1 atm. One minute after voltages were applied on three electrodes, collecting currents were measured and recorded to the computer through the MXI interface at various gas concentration and temperature conditions.