Reasons for the mechanical preparation of the root canal

The use of mechanical instrumentation is accepted as being one of the most important steps in root canal treatment.1 It is also recognised as being one of the more difficult practical challenges in operative dentistry. The main objectives of canal preparation include:

  • The removal of vital and necrotic pulpal tissue from the root canal space

  • The removal of microorganisms and their products from the canal space

  • The removal of infected dentine

  • To create a space that can be both irrigated and medicated

  • Allow predictable placement of a root canal filling material

  • Maintain sound root and coronal tissue to allow effective restoration and function of the tooth.

Mechanical shaping must be accompanied by the thorough irrigation of the resultant space, a concept often referred to as 'cleaning and shaping', although many operators suggest that 'shaping and cleaning' is a more accurate term where files shape and irrigants clean. In practice these two objectives can be achieved simultaneously as part of an overall chemo-mechanical protocol.

Schilder2 is often cited as being one of the most influential authors in describing the form of root canal preparation shapes and is credited with describing important design concepts that still hold true today:

  • A preparation that is continuously tapering with its largest diameter coronally and its narrowest diameter apically

  • The root canal preparation should encompass the shape and original anatomy of the natural canal

  • The original position of the apical foramen should be maintained

  • The foraminal size of the preparation should be kept as small as practical (Fig. 1).

    Figure 1: a Graphical representation of the ideal 'Schilder' canal preparation shape
    figure 1

    b Post-operative radiograph exhibiting a canal preparation with a continuously tapering shape and maintenance of original foraminal position

Unfortunately for clinicians the inherent complex anatomy of root canal systems3 including variations in length, curvature and diameter, as well as the presence of intricate canal anastomoses, provides a major challenge to the processes of canal shaping (Fig. 2).

Figure 2
figure 2

The complex microscopic anatomy of the root canal system (after Hess3)

Historical approaches to canal preparation

Although the preparation of root canals has been described in literature since the start of the eighteenth century4 no systematic preparation sequence of instrumentation appears to have been proposed till 1961.5 For many years following this, canal preparation was completed using a sequence of stainless steel files or reamers in a serial step-back or incremental technique.6,7 This technique consisted of initially negotiating the canal to working length followed by enlargement of the apical portion of the canal, with shaping of the body of the canal being completed by using successively larger files at progressively shorter lengths (Fig. 3). This regularly proved to be a difficult procedure, which often utilised sharp ended instruments in a push-pull filing motion. Disadvantages of this approach included considerable difficulty in achieving initial canal negotiation, premature blockage of canals with debris, inadequate irrigant solution penetration, and an increased incidence of canal ledging, transportation or perforation (Fig. 4).8 Often this technique proved to be time consuming and frustrating, especially where the original canal diameter was small, and the canal curvature more severe.

Figure 3
figure 3

Serial step-back procedure and file selection for preparation, and confirmation of deep canal shape

Figure 4: a Apical ledge formation is often created by the use of files with inappropriate tip size and lack of flexibility.
figure 4

Such blockages and ledges can be extremely difficult to negotiate

b Apical foramen transportation caused by over-instrumentation and length determination error

In response to the limitations of this technique an alternative step-down, or crown-down, approach to preparation was developed.9,10 In essence this technique completely reversed the step-back approach, advocating the preparation of the coronal aspect of the canal first followed by sequentially smaller files penetrating the canal until working length is achieved. In addition controlled rotational movements of hand files, such as the 'balanced-force' technique11 could be incorporated to minimise canal transportation. The crown-down approach offers several advantages over a step-back method:

  • Enhances irrigation penetration and maintains a reservoir of irrigant coronally through which instruments can pass

  • Minimises coronal flute binding of files, allowing enhanced control of the file tip and reducing procedural errors

  • Reduces the straightening of pre-curved files once in the canal

  • Enhances working length determination

  • Reduces the possibility for extrusion of debris into the periradicular tissues

  • Reduces the chances of inoculation of coronally placed pathogens into apical canal architecture

  • Enhances straight line access for instruments into the canal system.

The crown-down technique is now widely accepted as the preferred approach to canal preparation, irrespective of the type of shaping instrument utilised. A summary of a typical instrument sequence is outlined in Table 1.

Table 1 Summary of a typical instrument sequence

Classic approaches to canal filing have utilised ISO standardised 0.02 taper instruments by hand. These files have a diameter increase of 0.02 mm per millimeter from the file tip to the terminus of the 16 mm flute pattern (Fig. 5). The characteristic of these files to become inflexible in larger sizes and their tendency to produce relatively narrow preparations has been seen as a major disadvantage.12 These shortcomings have led clinicians and manufacturers to develop instruments that can complete canal shape more effectively and efficiently.

Figure 5: Graphical representation of an ISO size 15 file.
figure 5

D0 is 0.15 mm followed by 0.02 mm increases in diameter till D16 is reached (0.47 mm)

Nickel-titanium (NiTi) preparation systems

Since the mid 1990s the introduction of root canal shaping instruments manufactured from nickel-titanium alloys have dramatically influenced preparation techniques. There are now dozens of different product types on the market, a range of choice that can lead to bewilderment for the dentist using such instruments for the first time. The metallurgic properties of NiTi alloys allows instruments to be manufactured with enhanced flexibility and shape memory compared to stainless steel products. This has obvious advantages when preparing canals with curvatures, as the potential for canal straightening is reduced. Another major advantage of NiTi instruments is the ability to make files with increased taper while retaining flexibility. Tapers ranging from a two-fold increase (0.04 taper) to a six-fold increase (0.12 taper) are now readily available (Fig. 6). Instruments like these have, until recently, been predominantly for use by hand, or driven through a 360 degree rotation by specialised speed/torque controlled electric motors. This provides the opportunity for canal preparation with regular and greater tapers, using fewer instruments often in less time (Fig. 7).13

Figure 6: Greater taper hand files.
figure 6

These files have increased taper rates: White - 0.06, yellow - 0.08, red - 0.10, blue - 0.12. Flexibility is maintained by NiTi alloy composition

Figure 7: The original ProTaper file series.
figure 7

These variable taper designs allowed a potentially reduced number of files for canal preparation

NiTi rotary systems have developed a strong following in clinical practice and their use in general dental practice is becoming well established.14 While these instruments are no doubt extremely useful adjuncts to canal preparation, they alone cannot address every intricate variation in canal shape and the additional use of hand files must be considered essential to achieve thorough preparation. Following manufacturer's recommendations for use is a prerequisite for the avoidance of procedural errors such as file breakage. The following guidelines for rotary NiTi file usage apply to all manufacturers' systems:

  • Proper straight line access should initially be achieved

  • Rotary NiTi files are for canal enlargement of the canal, not negotiation

  • NiTi files should follow a hand file created glide-path

  • Use in a crown-down fashion

  • The use of an appropriate speed/torque control motor is highly recommended

  • Exert minimal apical pressure during canal instrumentation

  • Lubricate/irrigate frequently

  • Use each NiTi file within the canal for three to four seconds before withdrawing, cleaning and inspecting

  • Regard NiTi files as single use items.

The last point in this list is particularly relevant to rotary NiTi files due to the increased levels of torsional and flexural stresses on the instrument.15 As a result of such stresses, rotary NiTi files should be discarded after each tooth preparation to ensure quality control of endodontic file efficiency. In the United Kingdom all endodontic files and reamers are now considered as single use.16

A substantial amount of research has been carried out on the shaping and cleaning efficacy, preparation time, safety, and straightening effects of rotary NiTi systems.17 Overall conclusions from these studies state that:

  • The use of rotary NiTi instruments in curved root canal systems results in a reduction of canal straightening and better centred preparations

  • NiTi instruments alone are not capable of preparing all the canal walls

  • As long as manufacturers' guidelines are followed the use of NiTi instruments can be considered safe.

Practitioners should therefore be fully aware of the need for appropriate instrument training and familiarisation with different systems. The importance of thorough concurrent canal irrigation cannot be overstated. When the above factors are taken into consideration canal preparations can be achieved quickly and effectively, while maintaining the mechanical objectives of preparation shape previously noted (Fig. 8).

Figure 8: Pre- and post-operative radiographs of a mandibular molar tooth prepared using the ProTaper system to finishing file F2.
figure 8

The mechanical objectives of canal shape have been observed

Rotary NiTi instrument designs

The detailed designs of the numerous rotary NiTi systems available are outside the scope of this article. However, there are basically two groups of design differences in regard to cross sectional shape: files with radial landed features and non-landed types. (Fig. 9). As a generalisation landed files allow a slower, but slightly more predictable preparation to their ability to stay centred in a canal. Non-landed instruments have a more effective cutting action, and usually allow a more rapid preparation technique. Non-landed files require a higher level of operator care and experience in order to avoid preparation errors, especially in relation to working length determination.

Figure 9
figure 9

Cross sectional SEM views of (a) a landed rotary NiTi file (Profile) and (b) a non-landed type (ProTaper)

Measuring working length

The question of where to finish the apical extent of a root canal preparation has divided the endodontic community for a number of years. Most endodontic practitioners will aim to prepare to the apical constriction of the canal, in essence where the pulpal tissues become peri-apical tissues. Locating this position, however, can be demanding as it cannot be identified radiographically, it is highly variable in relation to the radiographic apex of a root, and disease processes may alter or destroy it.18 In practice a combination of measurement techniques, including radiography, electronic apex location, tactile awareness, and the use of paper point measuring techniques often allows a more consistent method of measurement than a single technique alone. Modern electronic apex locators are now so consistently accurate that they form an invaluable aid to endodontic treatment.

Limitations of rotary NiTi instrumentation

If correct procedure is followed, many root canals can be effectively shaped using rotary NiTi files. A 'golden rule' is to explore the canal system using small, flexible hand files, usually sizes 8, 10 and 15, before inserting a rotary file to the same length. This 'scouting' of the canal can provide valuable tactile information on the safety of the preparation case. Specifically, canals that exhibit sharp curvatures and re-curvatures, division, or convergence with abrupt change of direction, can provide challenging environments for rotating NiTi files due to increases in torsional and flexural stresses (Fig. 10). In these situations it is often prudent to complete preparation using pre-curved fine stainless steel hand files.

Figure 10: Radiographic sequence showing a canal system with a distinct re-curvature.
figure 10

Preparation of such shapes can be a challenge for rotary NiTi files, and in this case the apical re-curve portion was prepared using hand instruments

The 'single file' concept

Many rotary NiTi systems require several files to achieve their preferred preparation shape. Along with concerns over file separation, decontamination and cost issues have driven manufacturers to try and reduce the number of NiTi instruments necessary to create an ideal shape. Some rotary NiTi systems are now available that have a greatly reduced number of files for example, Twisted Files® (Sybron Endo, Orange, CA), and others that utilise a single rotary instrument (One Shape®, Micro Mega®, Bescanson, France). Both these systems retain a full 360 degree rotational motion, and TF files utilise an altered state NiTi alloy that is purported to be more resistant to distortion and fracture.19 More recently the use of NiTi files in a reciprocating motion have been commercially developed that claim to allow canal shaping using only a single file (Wave One™, Dentsply Maillefer, Ballaigues, Switzerland, and Reciproc®, VDW®, Munich, Germany). This concept was first reported by Yared20 where a Protaper F2 finishing file was driven in a reciprocating fashion rather than through a conventional 360 degree rotation. The clockwise (144 degree) and anti-clockwise (72 degree) motions were controlled via an electronic motor, and the root canal was only pre-negotiated by an ISO 0.02 taper size 8 file. Such an approach was claimed to be effective at shaping and maintaining canal curvature. Additionally, advantages in reducing potential instrument cross contamination and associated cost reductions were claimed.

These types of instruments have now been launched commercially, with a small number of preparation files to simplify the clinician's choice between small, medium and large canal sizes (Fig. 11). Both these reciprocating systems are made from a modified NiTi alloy (M-wire) and claim to be able to prepare adequate canal shape with a minimised risk of file fracture due to the reciprocating nature of the file movement. Reciproc® is recommended for use without the need to create a 'glide-path' before its introduction into the canal, a claim that appears to have some experimental merit in roots with little or moderate canal curvature.21

Figure 11: WaveOne file system.
figure 11

Only three files are in this series: Small (yellow), primary (red) and large (blue)

Another study compared the shaping and cleaning potential of these new designs compared to 'conventional' rotary NiTi systems.22 Both reciprocating systems maintained root canal curvature and were considered safe. The shaping time for the reciprocating files was naturally shorter due to the reduced number of files used, but it was noted that irrigation time of the canal should not be compromised by this reduced mechanical preparation time. Varying results regarding the amount and position of debris produced by reciprocating files has also been assessed23 and further studies are merited to investigate the performance of this type of file.

There is no doubt that the introduction of reciprocating files has generated a high level of interest, especially in the general dental practitioner population. Their use in selected situations can produce consistently good results (Fig. 12), and the reduced number of files needed to prepare adequate shapes has obvious financial and time saving potential. The need to purchase a specialised motor capable of driving the reciprocating file correctly does initially offset some of these cost savings, but many of these dedicated units also allow the use of more conventional rotary systems, allowing the clinician to apply hybrid approaches to canal preparation using a combination of instruments. The authors of this article remain convinced that being proficient in the use of a number of different endodontic instruments allows a much more satisfying ability to address and treat a wider range of canal shapes.

Figure 12
figure 12

Pre- and post-operative radiographs of a mandibular molar prepared using a single WaveOne primary file after 'glide-path' management

Conclusions

The shaping of root canals is often considered by clinicians as a complex and challenging practical procedure. Historically there have been multiple approaches recommended to make it a more efficient and easier process, some of which have appeared to be contradictory to one another. The evolution of a crown-down approach and the introduction of advances in material design have revolutionised the clinical practice of many endodontists and general practitioners. The use of such mechanical method alone, however, has not impacted on the clinical success of endodontic procedures. It is essential that the mechanical preparation of the root canal system is backed up by a sound biologic knowledge of the endodontic disease process, and the application of complimentary anti-bacterial therapies to ultimately improve the clinician's ability to maintain the natural dentition. In this respect, the importance of thorough and effective canal irrigation cannot be over emphasised. Irrigation and its microbiological explanation are covered in another article within this endodontic issue.