Abstract
The generation of low-emittance electron beams has received significant interest in recent years. Driven by the requirements of X-ray-free electron lasers, the emittance of photocathode injectors has been reduced significantly, with corresponding increase in beam brightness. This has put increasingly stringent requirements on the instrumentation to measure the beam size. These requirements are even more stringent for novel accelerator developments, such as laser-driven accelerators based on dielectric structures or on a plasma. We present here the generation and measurement of a sub-micrometer electron beam, at a particle energy of 330 MeV, and a bunch charge below 1 pC. An electron-beam optics with a vertical β-function of a few millimeters has been setup. The beam is characterized through a wire scanner that employs a 1 μm wide metallic structure fabricated using the electron-beam lithography. The smallest (rms) transverse beam size presented here is <500 nm.
Similar content being viewed by others
Introduction
Nowadays, scientists are striving for electron beams with extremely small transverse sizes and emittances, driven by the high demand of X-ray-free electron lasers (XFELs) and novel accelerators development. In linear colliders, one of the main design goals is to achieve a small beam size at the interaction point to maximize the luminosity. The design of the International Linear Collider, for example, uses a nanometer-sized beam at the interaction point1,2. Similar requirements have to be met for advanced accelerator concepts, such as dielectric laser accelerators3,4 and plasma wakefield accelerators5,6,7. These schemes utilize micron scale transverse accelerating structures, defined by either the plasma or laser wavelength, and have been shown to achieve high accelerating gradients. In a dielectric laser accelerator, the electron beam is accelerated through a dielectric microstructure that requires beam transverse dimensions in the sub-μm scale8. In the field of plasma-based acceleration, reducing the transverse emittance down to the nanometer scale would open a path towards many different applications of this technology, such as colliders and light sources9. Concerning XFELs the reduced beam emittance would allow for a compact design with lower beam energy and shorter undulator length10,11,12. For future applications with ultralow emittances and beam sizes, special care may be required to avoid a deterioration of the beam quality due to space-charge effects: the so-called laminarity parameter13 should be kept smaller than one to ensure a negligible space-charge contribution.
The challenge of designing ultralow emittance accelerators brings the need for transverse beam size diagnostics with sub-micrometer resolution. However, the resolution limit of commonly used beam profile monitors hinders the possibility to measure sub-micrometer spot sizes14,15,16 (details are provided in ‘Discussion’).
We designed a novel wire scanner (WSC hereinafter) with resolution in the sub-micrometer range. In a WSC measurement, a thin metallic wire scans the beam transversally. The generated particle shower (loss signal) is detected downstream, enabling the reconstruction of the beam transverse profile (see ‘Methods’). The design of a WSC with improved resolution requires a reduction of the wire width. We developed a high-resolution profile monitor exploiting nanofabrication techniques to produce a 1 μm wide metallic stripe on a membrane by electron-beam lithography and electroplating. The proposed design ensures sub-micrometer resolution as well as enables the integration of a wire scanner on-a-chip. We present in this paper the generation and characterization of an electron beam with a 53 nm normalized emittance in the vertical plane, for a particle energy of 330 MeV and a bunch charge below 1 pC. This nanometer emittance has been attained at SwissFEL11,17, the X-ray-free electron laser at the Paul Scherrer Institute. We achieved sub-micrometer spot sizes setting up an electron-beam optics to meet a 2.61 mm β-function in the vertical plane. The presented acceleration setup, beam optics and wire scanner on-a-chip, have enabled the generation and measurement of beams with rms (root mean square) sizes <500 nm.
Results
Generation of sub-micrometer relativistic electron beams
Figure 1a shows the layout of the SwissFEL injector, where we generated and characterized the relativistic electron beams with rms transverse sizes below 500 nm.
In the SwissFEL injector, electron beams are emitted in an radio-frequency (RF) photoinjector gun18 and initially accelerated in an S-band RF booster linac (consisting of two sections: Booster 1 and 2) up to a maximum energy of 330 MeV19. In the last injector section, an experimental chamber20 (Fig. 1b) located downstream of a bunch compressor, offers the possibility of scanning the novel wire scanner on-a-chip (Fig. 1c) vertically through the beam to measure the vertical beam profile.
The beam size was brought down to sub-micrometer scale by minimizing the beam emittance at the photoinjector and the vertical β-function at the interaction point with the wire scanner (see ‘Methods’). The emittance was minimized by suitably reducing the laser spot size on the cathode and by operating the gun at a very low charge (below 1 pC) where space-charge effects are negligible. The measured vertical normalized emittance was 53 nm, with estimated errors of ~10%.
Once the gun is optimized for the smallest emittance, the achievement of a sub-micrometer size hinges on the minimization of the vertical β-function. For this purpose, we calculated and implemented in the accelerator a specifically suited optics. The computed vertical β-function at the wire scanner location was 2.61 mm.
Using an electron energy of 330 MeV and the presented vertical normalized emittance and β-function, we were able to generate a beam with an expected vertical rms size of 460 ± 20 nm at the wire scanner location.
Beam profile measurements with sub-micrometer resolution
The goal of measuring sub-micrometer beam sizes can not be achieved using conventional profile monitors such as screens and wire scanners, since their resolution is typically limited to the micrometer scale (see ‘Discussion’). Among them wire scanners are highly promising devices in terms of spatial resolution, which mainly depends on the wire width. In the conventional design, this dimension is constrained to a few micrometers21 (see ‘Methods’). To overcome this limitation, we exploited the possibilities offered by nanofabrication techniques to develop a novel WSC on-a-chip with improved sub-micrometer resolution. The device we present here consists of a silicon chip with a silicon nitride membrane at the center. On the membrane, two gold stripes are produced by electron-beam lithography and electroplating. They are characterized by different widths, namely w1 = 1 μm and w2 = 2 μm. Consequently, their rms geometrical resolutions are 0.3 and 0.6 μm, respectively (see ‘Methods’). A vertical scan through the beam is performed for each stripe independently to measure the beam profile. For comparison, we also scan through the beam a conventional cylindrical 5 μm tungsten wire whose rms geometrical resolution is 1.25 μm16.
In Fig. 2, we present measurements of the beam vertical profile obtained scanning each of the three wires through the beam and acquiring the corresponding loss signal by a beam loss monitor (BLM). For each wire, we show three sequential beam profile acquisitions. We evaluated the rms size of each of the three profiles. In Table 1, we report their mean values σrms, with their associated statistical errors. Only the σrms from the 1 μm gold stripe measurements is comparable with the expected beam dimension of 460 ± 20 nm. This behavior can be attributed to the resolution limits of the 5 μm tungsten wire and the 2 μm gold stripe. The measured traces are indeed the convolution of the beam profile with the wire cross-section. The cylindrical shape of the 5 μm wire is clearly visible in Fig. 2.
Nevertheless, the beam size can be extracted from the measured profiles fitting the data with the convolution of a function describing the wire shape and one describing the beam, which we assume to be Gaussian (details are provided in ‘Methods’). Figure 3 shows one of the three beam profiles measured by means of the conventional 5 μm W wire, the novel 2 and 1 μm gold stripes, and the corresponding fit curves. It also shows the reconstructed Gaussian profiles obtained by the described deconvolution. The profile measured by the 1 μm gold stripe and the corresponding reconstructed Gaussian profile are comparable. This confirms the goodness of our initial assumption of a Gaussian beam shape.
The beam vertical sizes σy derived from the fit procedure are summarized in Table 1. Each value is the average of the fit sigma over the three sequential beam profile measurements, with the corresponding standard deviation. The measured sizes are in good agreement with each other and with the expected beam dimension of 460 ± 20 nm.
Discussion
We have demonstrated the generation of sub-micrometer relativistic electron beams at SwissFEL and their characterization via wire scanner profile measurements.
This is a particularly promising result in view of advanced accelerator concepts experiments. A notable example is the development of an all-on-a-chip dielectric laser accelerator, pursued by the Accelerator on a Chip International Program22. A proof-of-principle experiment of high-gradient dielectric laser acceleration will be conducted at PSI using the SwissFEL high-brightness electron beam20,23. Generating electron beams with sub-micrometer spot size will be mandatory in order to achieve full transmission through the accelerating structure8.
The characterization of sub-micrometer beams requires the development of diagnostics with spatial resolution of the same order of magnitude. Several different techniques are commonly used to measure the beam transverse profile. Screens are the most used two-dimensional beam profile monitors. In the case of the SwissFEL YAG:Ce screens the spatial resolution is 8 μm with a smallest measured beam size of 15 μm15. A higher spatial resolution has been achieved at the UCLA Pegasus laboratory utilizing a 20 μm YAG:Ce crystal with an in-vacuum infinity-corrected microscope objective coupled to a CCD camera. This set-up allowed measurements of beam sizes down to 5 μm and transverse emittances of 5 nm × 10 nm24. The spatial resolution of optical transition radiation screens (OTRs) is only limited by the camera sensor and optics14. In the Accelerator Test Facility 2 at KEK, a vertical beam size of 750 nm has been measured by means of OTR monitors25,26. Nevertheless, OTRs are limited by the emission of coherent optical transition radiation from compressed bunches, which prevents their usage in FELs27. Laser wire scanners are one-dimensional profile monitors that require a laser beam focused to a diffraction-limited waist. The minimum achievable laser waist and therefore measurable beam size is limited by the laser wavelength. At the SLAC Final Focus Test Beam, scientists have measured an electron beam vertical size of 70 nm by means of a laser beam split and folded onto itself to produce an interference fringe pattern2. The minimum achievable laser wavelength coupled with space constraints and costs represents the main limitation of such profile monitors14.
In this work, we focused on wire scanners, since they are the most promising profile monitors in terms of spatial resolution. The latter is limited by the encoder readout, the wire diameter and vibrations. In SwissFEL, a wire-geometry dominated resolution of 1.25 μm was reached with a 5 μm tungsten wire. This was achieved thanks to a motion system provided of an encoder with resolution of 0.1 μm, and to measured wire-vibrations largely below the geometrical resolution16,21. Therefore, a natural way to push the geometrical resolution of a WSC towards the sub-micrometer scale was by decreasing the wire diameter. However, the strength of a wire reduces with its diameter. Thus, the conventional manufacturing technique of stretching a wire onto a wire fork limits its width to a few micrometers14. Nanofabrication has been used to develop wire scanners with a resolution on the micrometer scale28. We were able to overcome this limitation and fabricate wires with a width down to 1 μm employing innovative nanofabrication techniques. In principle, a minimum width of 0.2 μm could be reached using the same technique. It should be noted though that reducing the stripe width implies a decrease of the signal-to-noise ratio. The reduction of the signal-to-noise ratio with a smaller stripe width could be compensated by increasing the stripe thickness.
It should also be considered that during a measurement, the electron beam transfers energy to the stripe, which increases its temperature. For the beam parameters in the present experiment, this is not a problem. The stripes were inspected with a scanning electron microscope after the beam tests, and no damage was found. Additionally, we modeled the temperature rise, and this model supports the finding: the maximum temperature for beam parameters as discussed in the paper is 458 K. For a higher bunch charge and beam energy, the temperature rise needs to be examined.
Methods
Accelerator setup
Our goal was the generation of electron beams with rms vertical size below one micrometer at the interaction point with the wire scanner. The rms transverse size of the beam at position s along the accelerator is
where β(s) is the β-function of the magnetic lattice at position s, εn is the normalized emittance and γ is the electron Lorentz factor. The beam size can be minimized by reducing the normalized emittance, increasing the beam energy, and reducing the β-function at the interaction point. The final energy E of the SwissFEL injector is 330 MeV, so we focused our efforts in minimizing the emittance and the β-function at the WSC location.
In RF photoinjectors, the emittance is determined by three different contributions: intrinsic or thermal emittance, space charge, and RF fields29. To minimize the space-charge contribution to the emittance increase, we operated at a very low charge, below 1 pC. The RF gun was set to its maximum accelerating gradient, while the gun phase was empirically adjusted to minimize the beam energy spread. In these conditions, the emittance was practically determined by the intrinsic or thermal contribution, which is proportional to the laser beam size on the cathode. We therefore reduced the laser iris to the smallest possible diameter, corresponding to an rms laser spot size on the cathode of about 80 μm. The electron-beam emittance was measured in both transverse planes downstream of the bunch compressor using the symmetric single-quadrupole-scan technique30. The measured normalized emittances are εn,x = 42 nm and εn,y = 53 nm in the horizontal and vertical planes, respectively, with estimated uncertainties of ~10%. The different transverse emittances derive from a slight asymmetry of the laser spot on the cathode.
Concerning the optics, the strengths of five quadrupole magnets between the bunch compressor and the chamber were calculated to minimize the vertical β-function at the interaction point, while keeping the horizontal β-function to a reasonable value. This optimization was performed using the code elegant31. Figure 4 shows the evolution of the β-functions from the bunch compressor to the interaction point for the calculated optics. During the experiment, the optics at the exit of the bunch compressor were iteratively matched to their design values using several quadrupole magnets upstream of the bunch compressor. Then, the five quadrupole magnets were set to the pre-calculated settings. Finally, the last quadrupole magnet upstream of the chamber was empirically adjusted to minimize the beam size at the wire scanner. This adjustment compensated any possible errors in the incoming beam optics at the bunch compressor, the electron-beam energy, the quadrupole field, and the exact position of the wire scanner.
The computed β-functions at the chamber location are βx = 0.273 m in the horizontal plane and βy = 2.61 mm in the vertical one. According to Eq. (1), the rms vertical beam size is 460 ± 20 nm for the considered experimental parameters E, εn,y, and βy (see Table 2).
Sub-micrometer resolution wire scanner on-a-chip
Figure 5 shows the sketch of a sub-μm resolution WSC on-a-chip (a) and the fabricated devices mounted on sample holders (b). Each WSC on-a-chip consists of a 6 × 6 mm2 Si chip having a 2 × 2 mm2 silicon nitride (Si3N4) membrane at its center. On the membrane two metal stripes of different widths w1 = 1 μm and w2 = 2 μm, are electroplated. The separation between the two stripes is 0.67 mm. This distance has been chosen such that each stripe can be separately scanned through the beam. The stripe material selected is gold. This heavy metal provides a strong scattering cross-section for a good signal-to-noise ratio of the loss signal, as well as sufficient resistance against heating from the beam interaction.
The implemented fabrication procedure is an application of the nanofabrication technique described in refs.32,33 The prototype fabrication starts with the evaporation of a Cr–Au–Cr metal layer sequence (5–20–5 nm) on a 250 nm thick Si3N4 membrane. Then a 3.2 μm thick layer of poly-methyl-methacrylate (PMMA) resist is spin-coated (Fig. 6a). The two Cr layers ensure better adhesion between Au and the membrane as well as the resist32. In a second step, applying electron-beam lithography we wrote two parallel stripes into the resist, exposing it to a 100 keV electron beam at 5 nA beam current and a dose of 1900 μC cm−2. The electron beam changes the chemical and physical properties of the PMMA resist, making it soluble in the developer solution. In particular, the exposed regions were developed by immersing the resist in a solution of isopropanol and water (7:3 by volume) and then rinsing the sample in de-ionized water and blow-drying it in a N2 gas jet. The development time was 45 s, and as a result the exposed parts of the PMMA have been completely removed (Fig. 6b). After the development, we etched the sample by Cl2/CO2 plasma for 45 s, to remove the top Cr layer and reveal the underlying Au sheet. The developed resist trenches were subsequently filled with Au by electroplating in a cyanide-based plating bath at a plating current density of 2.5 mA cm−2 for 14 min. After the electroplating, the PMMA resist was completely removed by ashing in an oxygen plasma for 15 min. In this way, only the Au stripes remained on top of the membrane (Fig. 6c). The manufactured samples were inspected and characterized with a scanning electron microscope (SEM). Figure 7 shows SEM images of a 2 and a 1 μm wide Au stripe on the Si3N4 membrane. The measured stripe thickness is 3 μm.
Detection setup
Figure 8a shows a schematic drawing of the experimental chamber in the SwissFEL injector, employed to perform the wire scanner measurements. The chamber is equipped with an in-vacuum manipulator operated by a 2-phase stepper motor via a feed-through. Four sample holders are mounted on the manipulator where four different samples can be settled. The stepper motor translates vertically the manipulator so that a sample can be inserted into the SwissFEL vacuum chamber along this direction and brought to the interaction point with the electron beam. By using a load-lock pre-chamber, the samples can be installed on the sample holders without breaking the vacuum employing a manually controllable manipulator20.
In the chamber, we installed the sub-μm resolution WSC on-a-chip as well as a reference WSC consisting of a 5 μm tungsten wire. The wire in this case is mounted using the conventional technique of stretching it onto the sample holder and fixing it between two pins (Fig. 5b). Figure 8b shows the prototypes installed in the SwissFEL injector beamline.
During a WSC beam profile measurement, the wire is scanned through the beam along the vertical direction at a constant velocity. The wire position is acquired by an absolute optical encoder (0.1 μm resolution). When the electron beam intercepts the wire a shower of primary scattered electrons and secondary emitted e+ e− γ is produced. The shower intensity is proportional to the fraction of the beam intercepted by the wire and represents the signal of interest (loss signal). The loss signal is detected outside of the vacuum chamber by a beam loss monitor (BLM) installed 4.5 m downstream of the experimental chamber. SwissFEL BLMs consist of a fast scintillator fiber (Saint Gobain BCF-20) wrapped around the beam pipe. This scintillator fiber is connected via a plastic optical fiber to a photomultiplier tube, which has remotely adjustable gain in the range [5 × 103, 4 × 106]. The signal is finally sent to an ADC unit for digitization and processing34. SwissFEL’s BLM acquisition and processing system enables resolving two bunches with a temporal separation on the nanosecond scale, being specifically designed for resolving two bunches emitted at the SwissFEL injector with 28 ns spacing between them. The BLM gain was set to 9 × 104. This value guaranteed a suitable signal quality and amplification during the beam profile measurements with the three wires.
For every wire position along the beam, we measured the loss signal for one second at a beam repetition rate of 10 Hz. Every point in the beam profiles presented in Figs. 2, 3 is the average of these acquired values.
Data processing
Every infinitesimal element of the wire width induces on the beam, a loss that is proportional to the product of the wire thickness and the beam distribution35. Under the assumption of a Gaussian beam shape, the measured loss signal is given by
where y is the wire position along the measurement direction and t(·) is a function describing the wire thickness shape. α, γ are the Gaussian function amplitude and centroid, respectively; Δ is an offset and σ is the Gaussian function standard deviation. The data were fitted through this function, with free parameters Δ, α, σ, γ.
The conventional 5 μm tungsten wire is cylindrical, hence to describe the wire shape we choose the function
where D = 5 μm is the wire diameter.
Instead, the 2 and 1 μm gold stripes are rectangular so the proper function to describe their shape is the step function
where tw = 3 μm is the stripe thickness and w is the stripe width.
Since the tungsten wire radius and the widths of the gold stripes are known from measurements, it is not necessary to consider them free parameters in the fit.
The results summarized in Table 1 and Fig. 3 support the conclusion that by applying this analysis we can reconstruct the beam size from the profiles acquired with all three wires. It should be noted though that this analysis is only valid for Gaussian beams. Conversely, the 1 μm stripe can be used to accurately measure the profile of beams with transverse size of 400–500 nm and directly extrapolate the beam dimension, without the need of an analysis demanding a-priori assumptions on the beam shape. Since the resolution limit is lower than the beam dimension the effects of the convolution in the measured profile are negligible, whereas small details would be washed out by the convolution with a larger wire. The σy and σrms of the profile measured with the 1 μm stripe are indeed in good agreement, with an error below 9% (see Table 1).
Wire scanner geometrical resolution
The geometrical resolution r of a wire is defined by its rms size14,16. The square rms size of a wire whose thickness is described by the function t(·) is
where
A wire that is round in cross-section has an rms size equal to D/436, since its thickness is described by the function t(u) in Eq. (3). As a result, the 5 μm W wire geometrical resolution is equal to r = 1.25 μm. For a rectangular stripe, we have r = \(w{\mathrm{/}}\sqrt {12}\) (cf. Eq. (4)). Therefore, the geometrical resolutions of the 1 and 2 μm Au stripes are equal to 0.3 and 0.6 μm, respectively.
Data availability
The data that support the findings of this study are available from the corresponding author upon request.
References
Behnke, T. et al. The international linear collider technical design report-volume 1: executive summary. Preprint at arXiv:1306.6327 (2013).
Balakin, V. et al. Focusing of submicron beams for TeV-scale e+ e− linear colliders. Phys. Rev. Lett. 74, 2479 (1995).
England, R. J. et al. Dielectric laser accelerators. Rev. Mod. Phys. 86, 1337 (2014).
Peralta, E. A. et al. Demonstration of electron acceleration in a laser-driven dielectric microstructure. Nature 503, 91 (2013).
Blumenfeld, I. et al. Energy doubling of 42 GeV electrons in a metre-scale plasma wakefield accelerator. Nature 445, 741–744 (2007).
Sears, C. M. S. et al. Emittance and divergence of laser wakefield accelerated electrons. Phys. Rev. Spec. Top. Accel. Beams 13, 092803 (2010).
Weingartner, R. et al. Ultralow emittance electron beams from a laser-wakefield accelerator. Phys. Rev. Spec. Top. Accel. Beams 15, 111302 (2012).
Wootton, K. P., McNeur, J. & Leedle, K. J. Dielectric laser accelerators: designs, experiments, and applications. Rev. Accel. Sci. Technol. 9, 105–126 (2016).
Xu, X. L. et al. Low emittance electron beam generation from a laser wakefield accelerator using two laser pulses with different wavelengths. Phys. Rev. Spec. Top. Accel. Beams 17, 061301 (2014).
Rosenzweig, J. B. et al. Next generation high brightness electron beams from ultra-high field cryogenic radiofrequency photocathode sources. Preprint at arXiv:1603.01657 (2016).
Milne, C. J. et al. SwissFEL: the Swiss X-ray free electron laser. Appl. Sci. 7, 720 (2017).
Prat, E. et al. Emittance measurements and minimization at the SwissFEL injector test facility. Phys. Rev. Spec. Top. Accel. Beams 17, 104401 (2014).
Chao, A. W., Mess, K. H., Tigner, M. & Zimmermann, F. Handbook of Accelerator Physics and Engineering (3rd Printing) (World Scientific, Singapore, 1999).
Tenenbaum, P. & Shintake, T. Measurement of small electron-beam spots. Annu. Rev. Nucl. Part Sci. 49, 125–162 (1999).
Ischebeck, R., Prat, E., Thominet, V. & Ozkan-Loch, C. Transverse profile imager for ultrabright electron beams. Phys. Rev. Spec. Top. Accel. Beams 18, 082802 (2015).
Orlandi, G. L. et al. Design and experimental tests of free electron laser wire scanners. Phys. Rev. Accel. Beams 19, 092802 (2016).
Ganter, R. SwissFEL-conceptual design report. Tech. Rep. Paul Scherrer Institute (PSI) (2010).
Raguin, J. Y., Bopp, M., Citterio, A. & Scherer, A. The SwissFEL RF Gun: RF design and thermal analysis. In Proc. of International Linear Accelerator Conference (LINAC12): Tel Aviv, Israel, September 9–14, 2012 495–497 (2013).
Raguin, J. Y. The Swiss FEL S-band accelerating structure: RF design. In Proc. of International Linear Accelerator Conference (LINAC12): Tel Aviv, Israel, September 9–14 498–500 (2013).
Ferrari, E. et al. The ACHIP experimental chambers at the Paul Scherrer Institut. Nuclear Inst. Methods Phys. Res. A (2018).
Orlandi, G. L. et al. First experimental results of the commissioning of the SwissFEL wire-scanners. In Proc. of Int. Beam Instrumentation Conf. (IBIC’17), Grand Rapids, Michigan, USA, August 20–24, 2017 (JACOW, Geneva, Switzerland, 2017).
Wootton, K. et al. Towards a fully integrated accelerator on a chip: dielectric laser acceleration (DLA) from the source to relativistic electrons. In Proc. of International Particle Accelerator Conference (IPAC’17), Copenhagen, Denmark, 14–19 May, 2017, International Particle Accelerator Conference No. 8 2520–2525 (JACoW, Geneva, Switzerland, 2017).
Prat, E. et al. Outline of a dielectric laser acceleration experiment at SwissFEL. Nucl. Instrum. Methods Phys. Res. Sect. A 865, 87–90 (2017).
Maxson, J. et al. Direct measurement of sub-10 fs relativistic electron beams with ultralow emittance. Phys. Rev. Lett. 118, 154802 (2017).
Bolzon, B. et al. Very high resolution optical transition radiation imaging system: comparison between simulation and experiment. Phys. Rev. Spec. Top. Accel. Beams 18, 082803 (2015).
Kruchinin, K. et al. Sub-micrometer transverse beam size diagnostics using optical transition radiation. J. Phys. 517, 012011 (2014).
Akre, R. et al. Commissioning the linac coherent light source injector. Phys. Rev. Spec. Top. Accel. beams 11, 030703 (2008).
Veronese, M. et al. A nanofabricated wirescanner with free standing wires: design, fabrication and experimental results. Nucl. Instrum. Methods Phys. Res. Sect. A 891, 32–36 (2018).
Fraser, J. S., Sheffield, R. L., Gray, E. & Rodenz, G. W. High-brightness photoemitter injector for electron accelerators. IEEE Trans. Nucl. Sci. 32, 1791–1793 (1985).
Prat, E. Symmetric single-quadrupole-magnet scan method to measure the 2D transverse beam parameters. Nucl. Instrum. Methods Phys. Res. Sect. A 743, 103–108 (2014).
Borland, M. Elegant: a flexible SDDS-compliant code for accelerator simulation. Technical Report (Argonne National Lab., IL, 2000).
Gorelick, S., Guzenko, V. A., Vila-Comamala, J. & David, C. Direct e-beam writing of dense and high aspect ratio nanostructures in thick layers of PMMA for electroplating. Nanotechnology 21, 295303 (2010).
Gorelick, S., Vila-Comamala, J., Guzenko, V. A. & David, C. High aspect ratio nanostructuring by high energy electrons and electroplating. Microelectron. Eng. 88, 2259–2262 (2011).
Ozkan-Loch, C. et al. System Integration of SwissFEL Beam Loss Monitors. In Proc. of International Beam Instrumentation Conference (IBIC2015), Melbourne, Australia, 13–17 September 2015 4, 170–174 (JACoW, Geneva, Switzerland, 2016).
Fernow, R. Introduction to Experimental Particle Physics (Cambridge University Press, Cambridge, UK, 1986).
Field, C. The wire scanner system of the final focus test beam at SLAC. Nucl. Instrum. Methods Phys. Res. Sect. A 360, 467–475 (1995).
Acknowledgements
We would like to express their sincere thanks to Thomas Schietinger for his careful proof-reading of the manuscript and to Beat Rippstein for his help during the mounting of the wire scanner on-a-chip. We are furthermore grateful to the SwissFEL vacuum, operation, and laser groups for their support. This research is supported by the Gordon and Betty Moore Foundation through Grant GBMF4744 to Stanford.
Author information
Authors and Affiliations
Contributions
M.B., S.B., C.D., R.I., and G.L.O. designed the wire scanner on-a-chip. V.G. nanofabricated the devices and with S.B. characterized them. S.B., R.I., and G.L.O. designed the experimental set-up. E.P. calculated the optics and minimized the beam size at the wire scanner location during the experiment. E.F. and E.P. setup and transported the electron beam through the accelerator. S.B., E.F., and R.I. installed the wire scanners. R.I. and E.P. wrote the data acquisition codes. S.B., E.F., R.I., G.L.O., C.O.-L., and E.P. performed the beam profile measurements. S.B., R.I., and G.L.O. performed the data analysis. S.B. wrote the manuscript, which was enriched by all the co-authors suggestions.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Borrelli, S., Orlandi, G.L., Bednarzik, M. et al. Generation and measurement of sub-micrometer relativistic electron beams. Commun Phys 1, 52 (2018). https://doi.org/10.1038/s42005-018-0048-x
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s42005-018-0048-x
This article is cited by
-
Generalized superradiance for producing broadband coherent radiation with transversely modulated arbitrarily diluted bunches
Nature Physics (2021)
-
Ultrafast Relativistic Electron Nanoprobes
Communications Physics (2019)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.