Abstract
Phenolic compounds are pollutants of major concern, and effective monitoring is essential to reduce exposure. Electrochemical sensors offer rapid and accurate detection of phenols but suffer from two main shortcomings preventing their widespread use: electrode fouling and signal interference from co-existing isomers. Here we demonstrate a potential solution based on environmentally friendly and biocompatible carbon nanomaterials to detect monophenols (phenol and cresol) and biphenols (hydroquinone and catechol). Electrode fouling is tackled in two ways: by introducing electrochemically resistant nanodiamond electrodes and by developing single-use nanocarbon electrodes. We provide a comprehensive analysis of the electrochemical performance of three distinct carbon materials (graphene, nanodiamond and nanocarbon). Nanocarbon exhibits the lowest detection limit below 10−8 M, and one order of magnitude higher sensitivity than the other carbon nanomaterials. We detect co-existing phenol isomers with nanocarbon electrodes and apply it in river water and green tea samples, which may pave the way towards low-cost industrial scale monitoring of phenolic compounds.
Introduction
Phenolic compounds, a class of chemicals containing one or more hydroxyl groups attached to an aromatic ring, are ubiquitous in the environment. They occur naturally in plants and form an essential part of the human diet including wine and fruits1. Also of artificial origin, these compounds are used in the manufacturing industry and are found in wastewaters, exhaust gases and as by-products of chemical engineering processes2,3. Their occurrence in waterbodies is undesirable, as they can cause severe adverse effects on animals and humans above a specific dose, resulting in severe kidney, liver and cardiovascular diseases4,5. At least 11 phenolic compounds are listed by the US Environmental Protection Agency and the European Union as priority pollutants6,7. Rapid and accurate monitoring of phenolic compounds in solution would be beneficial to quantifying and reducing the exposure to these contaminants8.
Detection of phenolic compounds is possible employing chromatography9, spectrophotometry10 and capillary electrophoresis11, reaching detection limits in the micromolar range. Using the redox reaction of hydroxyl groups on the aromatic ring, electrochemical methods have complemented the previous schemes, offering a quick, reliable and low-cost assay of phenolic compounds in the aqueous phase12,13. Biphenols, like hydroquinone (HQ) and 1,2-dihydroxybenzene (catechol), can be directly oxidised at electrodes at low over-potential13. However, monophenol detection requires the use of enzymes, such as tyrosinase, lactase and polyphenol oxidase8,14. These enzymes oxidise monophenols into o-biphenols which are subsequently oxidised to o-quinones, thereby facilitating their detection.
Electrochemical detection of phenolic compounds faces two challenges that hinder their use at industrial scales: First, reaction by-products lead to fouling of electrodes, which limits their accuracy and lifespan. Secondly, similar chemical structures of co-existing phenol isomers, such as HQ and catechol, result in overlapping redox peaks that makes simultaneous detection of these species challenging.
We show that carbon materials can offer solutions to these challenges. They are abundant, highly biocompatible and more importantly, span a wide range of physical and chemical properties, which highly depend on the carbon–carbon molecular orbital hybridisation15,16. Carbon electrode materials have impacted a range of disciplines from electrochemical sensing17,18, to energy storage and conversion15,19. Graphene-like carbon materials have been intensively studied as electrochemical sensors for phenolic compound detection. Notable examples are graphene and carbon nanotubes20,21,22. However diamond-like materials and amorphous carbon23, which interact with phenols differently due to their molecular bonding, have been largely overlooked.
In this work, we demonstrate that both nanodiamond and amorphous nanocarbon offer a facile solution to the aforementioned challenges due to their consistently high sensitivities, prevention from passivation and fouling and superior biocompatibility to enzyme modification24,25,26. We provide a comprehensive analysis of their electrochemical performance and compare them to graphene-modified electrodes for detection of mono and biphenols. The wide range of electrochemical methods, including voltammetry and amperometry allow the comparison of three distinct carbon materials systematically and provide sensing performance benchmarks. Further, we demonstrate the simultaneous detection of co-existing phenolic isomers using nanocarbon electrodes, and apply these methods to detect phenolics in real Thames river water samples and green tea.
Results
Electrochemical characterisation of carbon electrodes
In this study we selected three types of carbon materials based on their carbon–carbon hybridisation, i.e. graphene (mainly sp2), amorphous carbon (a mixture of sp2 and sp3) and boron-doped nanodiamond (mainly sp3)24,25,26. Conversion into carbon nanoparticles was achieved by microbead-assisted ultrasonic disintegration (Fig. 1a, b), yielding particle sizes of 4 to 8 nm, as measured by dynamic light scattering. These particles were drop-coated onto glassy carbon electrodes (GCE) (Fig. 1c) and were used to detect biphenols (Fig. 1d) and monophenols with the addition of the enzyme tyrosinase (Fig. 1e).
Scheme of electrode preparation for phenolic compound detection. a Microbead-assisted ultrasonic disintegration of carbon materials suspended in water for 12 h. b Carbon nanomaterials after ultrasonication (size 4–8 nm). c Drop-coating of resulting suspension of nanocarbon onto a polished glassy carbon electrode (GCE). d Direct use of a modified electrode for biphenol detection, e.g. hydroquinone oxidises to quinone. e Additional modification of electrode with the enzyme tyrosinase for monophenol detection, e.g. phenol is first converted into hydroquinone by the enzyme (cartoon adapted from PDB ID: 3NM838) and then oxidises into quinone at the electrode
We first characterised the modified carbon electrodes by cyclic voltammetry, i.e. graphene-GCE, nanodiamond-GCE and nanocarbon-GCE. Experiments were performed in 0.1 M phosphate buffer and 0.2 mM HQ. The cyclic voltammograms (CV) in Fig. 2a showed that nanocarbon-GCE exhibited eight times higher current density than GCE. CVs were obtained at nanocarbon-GCE with increasing scan rates (Fig. 2c), showing a linear relationship between peak current and scan rate. This linearity indicates that the main contribution to the measured current is coming from adsorbed HQ at the surface, rather than by diffusion from the bulk27. To test the adsorption of HQ at nanocarbon-GCE, we dipped the electrode into 2 mM HQ for periods up to 600 s and then performed CV in 0.1 M phosphate buffer (Fig. 2d). The observed cumulative redox peaks after successive dipping in HQ confirm that the species adsorbs at the electrode surface. Analogous experiments were carried out at GCE, nanodiamond-GCE and graphene-GCE. For these cases, the peak current was found to depend linearly on the square root of the scan rate (Supplementary Figure 1), indicative of a diffusion-controlled process, as described by the Cottrell equation27.
Electrochemical characterisation with cyclic voltammetry of modified carbon electrodes. a Cyclic voltammograms (CV) of modified carbon electrodes in 0.1 M phosphate buffer and 0.2 mM HQ: glassy carbon electrode (GCE) in black, nanodiamond-GCE in orange, graphene-GCE in blue and nanocarbon-GCE in red dashed lines. Nanocarbon-GCE exhibited eight times higher current density than GCE; b magnified CV from a of GCE, nanodiamond and graphene-modified GCE. c CV of nanocarbon-GCE at increasing scan rates immersed in the same solution as a. The inset figure shows a linear relationship between the peak current and the scan rate (from 20 to 200 mV s−1), indicative of adsorption of HQ at the surface; d CV in 0.1 M phosphate buffer after exposing nanocarbon-GCE to a solution of 2 mM HQ for the increasing dipping times. e The plot indicates peak currents achieved in d at shown dipping times; f CV in 0.2 mM cresol of modified GCE with nanocarbon-GCE without tyrosinase (red), tyr-GCE without nanocarbon (blue) and tyr-nanocarbon-GCE (black). A significant redox peak was only observed for tyr-nanocarbon-GCE (black)
Nanocarbon-modified electrodes showed enzyme biocompatibility. The detection of phenol was facilitated by the presence of the enzyme tyrosinase, which converted monophenol molecules into biphenol (Fig. 1e). We performed CV in cresol solutions and observed redox peaks when nanocarbon-GCE was modified with tyrosinase. However, in the absence of either nanocarbon or tyrosinase, no significant peak was detected (Fig. 2f). Enzyme modification was also possible for graphene-GCE (Supplementary Figure 2a), while nanodiamond-GCE did not show similar properties (Supplementary Figure 2b), which we attribute to its sp3 hybridisation.
To understand the high peak current observed at nanocarbon-GCE (Fig. 2a), we calculated the heterogeneous rate constant of hydroquinone using the Laviron equation28,29,30. This equation relates the heterogeneous rate constant k with the variations of the peak potentials as a function of the scan rate (full calculations are given in Supplementary Figure 3). We report a transfer coefficient and number of transferred electrons of α = 0.46 and n = 1.7, in agreement with previous results31,32,33,34. Using these parameters in the Laviron equation yields a rate constant of k = 2.06 cm s−1, which is significantly higher than the value at graphene/chitosan-GCE (0.0147 cm s−1)31, CILE (0.0629 cm s−1)33 and carbon fibres (0.0219 cm s−1 in water and 0.167 cm s−1 in 1 M H2SO4)34.
Electrochemical sensing of phenolic compounds
It is possible to detect biphenols (HQ and catechol) directly at carbon electrodes. Here we test the performance of GCE, nanodiamond-GCE, graphene-GCE and nanocarbon-GCE as biphenols sensors by three electrochemical methods: cyclic voltammetry (CV), square-wave voltammetry (SWV) and amperometry. Figure 3 shows representative results obtained using nanocarbon-GCE for HQ detection. When the HQ concentration raised, the corresponding peak current of the redox reaction scaled linearly, as measured both in CV (Fig. 3a and Supplementary Figure 4a) and SWV (Fig. 3b and Supplementary Figure 4b). To detect low concentrations of HQ at nanocarbon-GCE, we performed amperometry by successive additions of HQ, which resulted in sharp current jumps (Fig. 3c). Current jumps of 15 ± 1 nA were measured for additions of 0.2 μM of HQ. Analogous experiments using nanodiamond-GCE, together with GCE and graphene-GCE as comparison were carried out (Supplementary Figure 5). Measured current changes as a function of HQ concentrations are plotted in Fig. 3d for four electrodes. The sensitivity of each electrode is given by the slope of linear fits in Fig. 3d, among which nanocarbon-GCE stands out as the electrode with the highest sensitivity. Similar measurements for detection of catechol are given in Supplementary Figure 6.
HQ detection at nanocarbon-modified electrode. a CV at increasing concentration of HQ as indicated. b Square-wave voltammetric measurement of HQ at increasing concentrations, as indicated. c Amperometric detection of successive addition of HQ at indicated concentrations with applied potential of +0.2 V. Plot shows the first 600 s of the full experiment in the inset figure. d Results from amperometric experiments for four carbon-modified electrodes, i.e. GCE, nanodiamond-GCE, graphene-GCE and nanocarbon-GCE. The plot shows the current increase against added HQ concentration. Nanocarbon-GCE exhibits highest sensitivity, as indicated by the slope
Monophenols (phenol and cresol) were detected after additional modification of nanocarbon-GCE with the enzyme tyrosinase following the scheme in Fig. 1e. At 0 V we measured current jumps by successive addition of phenol (Fig. 4a, b) and cresol (Fig. 4c, d). The detection limits of four phenolic compounds for all measured carbon electrodes, together with previously reported ones are listed in Table 1. Among them, nanocarbon-GCE exhibited the lowest detection limit, reaching values below 100 nM.
Amperometric detection of monophenols with nanocarbon-GCE. a Current change upon addition of phenol at nanocarbon-GCE modified with tyrosinase with applied potential of 0 V. b Current jumps as a function of phenol concentration extracted from amperogram a). c Analogous amperogram performed with addition of cresol. d Current jumps as a function of cresol concentration extracted from amperogram c)
Low surface fouling at nanodiamond-glassy carbon electrodes
Fouling of electrodes is a problem for phenolic sensors35. The by-products from phenolic compound oxidation react with the electrode, thereby inactivating it. It is well-known that boron-doped diamond benefits from high chemical resistance22. To test whether nanodiamond-GCE resists fouling when used to detect phenolics, we performed cyclic voltammetry in HQ for consecutive cycles. After ten scan cycles (Fig. 5a in red) the voltammogram did not change significantly. In contrast, a dramatic drop of redox peak current was observed at a graphene electrode (Fig. 5b) indicating surface fouling. This comparison shows the excellent inertness of nanodiamond against phenolic compounds. We observed that peak currents at nanodiamond-GCE scale linearly with HQ concentration up to 100 μM (Fig. 5c). This shows that nanodiamond-GCE have a wider detection range. In comparison, graphene-GCE depart from linearity at concentrations above 10 μM (Fig. 5d), as well as nanocarbon-GCE (Supplementary Figure 4). We attribute fouling of graphene and nanocarbon electrodes to their sp2 carbon–carbon hybridisation, which facilitates the adsorption and reaction of phenolics oxidation by-products.
Resistance to electrode fouling of nanodiamond. a Plot shows first scan cycle at nanodiamond-GCE in 0.2 mM HQ (black) and after 10 scan cycles (red). b Analogous experiment as a performed with a graphene-GCE for comparison. c Peak currents at nanodiamond-GCE plotted against HQ concentration. d Results for an analogous experiment with a graphene-GCE
Simultaneous detection of phenolics
In addition to electrode fouling, interference from co-existing isomers disrupts the electrochemical signal due to their similar chemical structure and overlapping redox peaks. Here we used nanocarbon-GCE to resolve the peaks from HQ and catechol, two common phenolic contaminants that coexist in environmental waters13. We performed a series of experiments via SWV (Fig. 6). First, we added equal amounts of HQ and catechol successively, reaching equal final concentrations. We observed that two distinguishable oxidation peaks increased simultaneously with a significant separation of about 56 mV. When the catechol concentration was successively increased, but not HQ, only the catechol peak exhibited higher peak currents (Fig. 6b). Similarly, when catechol levels remained constant, but HQ was added, only the HQ peak showed an increment in current (Fig. 6c). Plotting peak currents against final concentrations of each species yielded a linear relationship shown in Fig. 6d. The achieved detection limits are below 0.02 µM. A comparison with other reported sensors is listed in Table 2.
Simultaneous detection of HQ and catechol at nanocarbon-GCE. a Square-wave voltammogram in 0.1 M phosphate buffer and successive addition of equal amounts of HQ and catechol reaching indicated final concentrations; b only addition of catechol from 8 to 13 µM while maintaining a fixed concentration of HQ (8 µM). c only addition of HQ from 8 to 13 µM while maintaining a fixed concentration of catechol (13 µM). d Plot of peak currents against concentration of HQ (black) and catechol (red), demonstrating the simultaneous detection of both species. The data points are extracted from a–c as indicated
Nanocarbon as reproducible single-use electrodes
We tested the suitability of nanocarbon as a single-use disposable electrode by performing voltammetry at increasing concentrations of catechol (Fig. 7a) and HQ (Fig. 7b). Each cycle was performed at a freshly prepared nanocarbon-GCE and discarded afterwards. When plotting peak currents against concentrations a linear relationship was measured (slope 1.88 ± 0.03 A/M and, R2 = 99.7 %), indicative of an accurate and highly reproducible approach.
Detection of HQ and catechol with single-use nanocarbon electrode. Square-wave voltammograms in 0.1 M phosphate buffer and increasing concentrations of catechol (a) and increasing concentrations of HQ (b). Each cycle has been performed at a freshly prepared nanocarbon-GCE. Inset plots show the peak current against concentration of each species
We used single-use nanocarbon-GCE to detect HQ and catechol in Thames river water samples. Known concentrations of HQ and catechol were added to river water samples (Fig. 8a in red) and SWV was used at nanocarbon-GCE to measure their concentration (Fig. 8a in blue). Found concentrations were extrapolated using the linear relationship shown in Fig. 7. Each measurement was repeated ten times. The average recovery rate was 102% for HQ and 101% for catechol. This experiment demonstrates that nanocarbon-GCE can be successfully applied in river water samples, without significant interference from additional co-existing compounds.
Phenolic compound detection in river water and green tea using a nanocarbon-GCE. a Detection of phenolic compounds in Thames river water after addition of known concentrations of HQ and catechol (in red). SWV method was used at nanocarbon-GCE and measured concentrations (in blue) were extrapolated using the linear relationship shown in Fig. 7. Each measurement was repeated ten times (error bars indicate standard deviations). b Voltammograms in 0.1 M phosphate buffer (background signal in black), green tea sample (blue) and after addition of 10 µM HQ and catechol to the green tea sample (red)
It is known that green tea contains high levels of various phenolic compounds, giving it its antioxidant properties1. We applied nanocarbon-GCE to detect HQ and catechol levels in green tea. We first performed SWV in PBS buffer (Fig. 8b, in black). We then boiled green tea in PBS buffer following the protocol in Methods section. The resulting peaks are shown in Fig. 8b (in blue). To test whether these peaks corresponded to HQ and catechol, we added equal amounts of each and obtained peaks at similar potentials (Fig. 8b in red). As shown in Fig. 6d, we assume a linear relationship between phenolic compound concentration and its detected peak current in this concentration range. Therefore, we extrapolated the HQ and catechol levels in the green tea sample to be 48 μM g−1 and 64 μM g−1, respectively, which is of the same magnitude as reported in the literature36,37.
Discussion
To tackle the challenges for the large-scale use of electrochemical sensors of phenolic compounds, we developed two promising electrodes based on carbon nanomaterials, i.e. nanodiamond electrode and nanocarbon electrode. On the one hand, nanodiamond, with its unique carbon bonding hybridisation, exhibited excellent resistance to fouling, which led to a wider detection range and longer lifetime, compared with graphene-modified electrodes.
On the other hand, we converted the abundant, cheap and biodegradable amorphous carbon into 5 nm size nanocarbon material by bead-assisted ultrasonication disintegration. Biphenols (HQ and catechol) adsorbed strongly onto the surface of nanocarbon-modified electrodes and exhibited redox peak currents one order of magnitude higher than other carbon nanomaterials. We also demonstrated that this nanocarbon material is biocompatible with the enzyme tyrosinase, thereby enabling the detection of monophenols (phenol and cresol). The high sensitivity and low detection limits make nanocarbon materials very efficient sensors to detect phenolic compounds. The strong adsorption properties also enables nanocarbon electrodes to resolve the signal from co-existing isomers (HQ and catechol), therefore making it an attractive option for simultaneous detection of phenolic compounds.
The protocol to prepare this nanocarbon electrode is easy to operate, safe, stable and highly reproducible, which facilitates the development of single-use disposable sensors using this material. We tested the single-use nanocarbon electrodes in real samples, i.e. river water and green tea, and accurately detected levels of HQ and catechol. This successful measurement in real contexts with a facile and biodegradable system paves the way for low-cost industrial scale monitoring of phenolic compounds.
Methods
Materials and reagents
The commercial Monarch 430 amorphous carbon (Cabot Corporation, CB M 430) was purchased from James M Brown Ltd (Staffordshire, UK). Graphene nano-platelets of 3 μm diameter and less than 10 nm thickness were acquired from Angstrom materials. The nanodiamond powder (ND-105/G4) was purchased from Yorkshire Bioscience Ltd. Other chemicals were obtained from Sigma-Aldrich unless otherwise stated.
Preparation of modified electrodes by microbead-assisted ultrasonic disintegration
Glassy carbon electrodes were cleaned before use by polishing with alumina powder, which was then removed by ultrasonication in water for 3 min. A total of 5 g milling beads were mixed with 1 mg powder of carbon nanomaterials and 5 mL Milli-Q water. The mixture was sonicated for 12 h, and then centrifuged (3 × 5 min, 10,000 rpm). Ten microlitre carbon materials suspensions were drop-coated onto a glassy carbon electrode with a surface area of 0.0707 cm2, and subsequently dried in the fume hood. Tyrosinase-modified electrodes used to detect phenol and cresol were prepared by dropping 10 μL 1 mg mL−1 tyrosinase solution onto the electrode surface and dried for 1 h (Tyr/nanocarbon-GCE). The protocol is illustrated in Fig. 1.
Electrochemical measurements
Electrochemical measurements were performed at room temperature (20 ± 2 ℃) with μ-AUTOLAB III potentiostat (Eco-Chemie, Netherlands), running GPES software (Version 4.9). A three-electrode system was used consisting of a working electrode, a Pt wire as counter electrode and an Ag/AgCl reference electrode. Electrochemical experiments were carried out in 0.1 M phosphate buffer solution (pH 7.4) at room temperature. All solutions and subsequent dilutions were prepared with Milli-Q water ( > 18 MΩ cm). The SWV used frequency of 10 Hz, sensitivity of 1 mV and amplitude of 10 mV.
Size characterisation by dynamic light scattering
Dynamic light scattering (Viscotek, Malvern) was used to measure the hydrodynamic diameter of the carbon nanomaterials. The average size of graphene was about 80 nm initially and it reduced to 56 nm after sonication. Nanocarbon powder had initial sizes of 50–70 nm and decreased to 5–8 nm after sonication. The size of nanodiamonds was measured to be 4–5 nm.
River water and tea sample preparation
A total of 0.1 M phosphate buffer was prepared using river water collected from the Thames River (Oxford). Tea samples were prepared by adding 2.5 g green tea (Twinings) into 50 mL 80 °C Milli-Q water for 5 min. Tea leaves were then removed and the sample was filtered using a 11 μm pore-size filter (Whatman, Grade 1).
Data and code availability
The data that support the findings of this study are available from the corresponding author upon reasonable request.
References
Crozier, A., Jaganath, I. B. & Clifford, M. N. Dietary phenolics: chemistry, bioavailability and effects on health. Nat. Prod. Rep. 26, 1001–1043 (2009).
Davı̀, M. L. & Gnudi, F. Phenolic compounds in surface water. Water Res. 33, 3213–3219 (1999).
Kahru, A., Maloverjan, A., Sillak, H. & Põllumaa, L. The toxicity and fate of phenolic pollutants in the contaminated soils associated with the oil-shale industry. Environ. Sci. Pollut. Res. 9, 27–33 (2002).
Agency for Toxic Substances and Disease Registry. Toxicological Profile for Phenol (U.S. Department of Health and Human Services, Public Health Service, Atlanta, GA, 2008).
Pandey, K. B. & Rizvi, S. I. Plant polyphenols as dietary antioxidants in human health and disease. Oxid. Med. Cell Longev. 2, 270–278 (2009).
U.S. Environmental Protection Agency. 40 CFR Appendix A to Part 423 -126 Priority Pollutants (2013). https://www.epa.gov/eg/toxic-and-priority-pollutants-under-clean-water-act
Liu, Y., Su, G., Zhang, B., Jiang, G. & Yan, B. Nanoparticle-based strategies for detection and remediation of environmental pollutants. Analyst 136, 872–877 (2011).
Della Pelle, F. & Compagnone, D. Nanomaterial-based sensing and biosensing of phenolic compounds and related antioxidant capacity in food. Sensors 18, 462 (2018).
Pérez-Ramírez, I. F., Reynoso-Camacho, Ra, Saura-Calixto, F. & Pérez-Jiménez, J. Comprehensive characterization of extractable and nonextractable phenolic compounds by high-performance liquid chromatography–electrospray ionization–quadrupole time-of-flight of a grape/pomegranate pomace dietary supplement. J. Agric. Food Chem. 66, 661–673 (2018).
Escarpa, A. & Gonzalez, M. C. Approach to the content of total extractable phenolic compounds from different food samples by comparison of chromatographic and spectrophotometric methods. Anal. Chim. Acta 427, 119–127 (2001).
Ehala, S., Vaher, M. & Kaljurand, M. Characterization of phenolic profiles of Northern European berries by capillary electrophoresis and determination of their antioxidant activity. J. Agric. Food Chem. 53, 6484–6490 (2005).
Harborne, J. B. Biochemistry of Phenolic Compounds. (Academic Press., London, 1964).
Ghadimi, H., Ab Ghani, S. & Amiri, I. Electrochemistry of Dihydroxybenzene Compounds: Carbon Based Electrodes and Their Uses in Synthesis and Sensors. (Elsevier, New York City, NY, 2017).
Kochana, J., Gala, A., Parczewski, A. & Adamski, J. Titania sol-gel-derived tyrosinase-based amperometric biosensor for determination of phenolic compounds in water samples. Exam. Interf. Eff. Anal. Bioanal. Chem. 391, 1275–1281 (2008).
Mao, X., Rutledge, G. C. & Hatton, T. A. Nanocarbon-based electrochemical systems for sensing, electrocatalysis, and energy storage. Nano Today 9, 405–432 (2014).
Zhang, W. et al. Recent development of carbon electrode materials and their bioanalytical and environmental applications. Chem. Soc. Rev. 45, 715–752 (2016).
Yang, N., Swain, G. M. & Jiang, X. Nanocarbon electrochemistry and electroanalysis: current status and future perspectives. Electroanalysis 28, 27–34 (2016).
Pandey, S. K. et al. Nanocarbon‐based electrochemical detection of heavy metals. Electroanalysis 28, 2472–2488 (2016).
Ji, Y., Huang, L., Hu, J., Streb, C. & Song, Y.-F. Polyoxometalate-functionalized nanocarbon materials for energy conversion, energy storage and sensor systems. Energy Environ. Sci. 8, 776–789 (2015).
Tsai, Y.-C. & Chiu, C.-C. Amperometric biosensors based on multiwalled carbon nanotube-Nafion-tyrosinase nanobiocomposites for the determination of phenolic compounds. Sens. Actuators B Chem. 125, 10–16 (2007).
Balasubramanian, K. & Burghard, M. Biosensors based on carbon nanotubes. Anal. Bioanal. Chem. 385, 452–468 (2006).
Vlamidis, Y., Gualandi, I. & Tonelli, D. Amperometric biosensors based on reduced GO and MWCNTs composite for polyphenols detection in fruit juices. J. Electroanal. Chem. 799, 285–292 (2017).
Dehsheikh, H. G., Ghasemi-Kahrizsangi, S. & Karamian, E. Addition impact of nano-carbon black on the performance of MgO.CaO compounds. Ceram. Int. 44, 5524–5527 (2018).
Koelmans, A. A. et al. Black carbon: the reverse of its dark side. Chemosphere 63, 365–377 (2006).
Boehm, H. P. Some aspects of the surface-chemistry of carbon-black and other carbons. Carbon N. Y. 32, 759–769 (1994).
Mochalin, V. N., Shenderova, O., Ho, D. & Gogotsi, Y. The properties and applications of nanodiamonds. Nat. Nanotechnol. 7, 11–23 (2012).
Bard, A. J., Faulkner, L. R., Leddy, J. & Zoski, C. G. Electrochemical Methods: Fundamentals and Applications. 2 (Wiley, New York, 1980).
Laviron, E. General expression of the linear potential sweep voltammogram in the case of diffusionless electrochemical systems. J. Electroanal. Chem. Interfacial Electrochem. 101, 19–28 (1979).
Laviron, E. Adsorption, autoinhibition and autocatalysis in polarography and in linear potential sweep voltammetry. J. Electroanal. Chem. Interfacial Electrochem. 52, 355–393 (1974).
Senthil Kumar, K. & Natarajan, P. Electrochemical behavior of two and one electron redox systems adsorbed on to micro-and mesoporous silicate materials: Influence of the channels and the cationic environment of the host materials. Mater. Chem. Phys. 117, 365–372 (2009).
Yin, H. S. et al. Electrochemical behavior of catechol, resorcinol and hydroquinone at graphene-chitosan composite film modified glassy carbon electrode and their simultaneous determination in water samples. Electrochim. Acta 56, 2748–2753 (2011).
Dąbrowski, A., Podkościelny, P., Hubicki, Z. & Barczak, M. Adsorption of phenolic compounds by activated carbon—a critical review. Chemosphere 58, 1049–1070 (2005).
Zhang, Y. & Zheng, J. B. Comparative investigation on electrochemical behavior of hydroquinone at carbon ionic liquid electrode, ionic liquid modified carbon paste electrode and carbon paste electrode. Electrochim. Acta 52, 7210–7216 (2007).
Ayranci, E. & Duman, O. Adsorption behaviors of some phenolic compounds onto high specific area activated carbon cloth. J. Hazard. Mater. 124, 125–132 (2005).
Wang, L. et al. Covalent modification of glassy carbon electrode with aspartic acid for simultaneous determination of hydroquinone and catechol. Ann. Chim. 97, 395–404 (2007).
Thangaraj, R., Manjula, N. & Kumar, A. S. Rapid simultaneous electrochemical sensing of tea polyphenols. Anal. Methods 4, 2922–2928 (2012).
Lin, H., Gan, T. & Wu, K. Sensitive and rapid determination of catechol in tea samples using mesoporous Al-doped silica modified electrode. Food Chem. 113, 701–704 (2009).
Sendovski, M., Kanteev, M., Ben-Yosef, V. S., Adir, N. & Fishman, A. First structures of an active bacterial tyrosinase reveal copper plasticity. J. Mol. Biol. 405, 227–237 (2011).
Cao, L., Sun, H., Li, J. & Lu, L. An enhanced electrochemical platform based on graphene-polyoxometalate nanomaterials for sensitive determination of diphenolic compounds. Anal. Methods 3, 1587–1594 (2011).
Fan, L. F., Wu, X. Q., Guo, M. D. & Gao, Y. T. Cobalt hydroxide film deposited on glassy carbon electrode for electrocatalytic oxidation of hydroquinone. Electrochim. Acta 52, 3654–3659 (2007).
Kong, Y., Ma, M. J., Mao, H. H., Ma, J. Q. & Yao, C. A novel amperometric sensor based on intercalated montmorillonite modified carbon paste electrode for hydroquinone determination. Anal. Methods 4, 748–752 (2012).
Wang, L. et al. Simultaneous determination of catechol and hydroquinone based on poly (diallyldimethylammonium chloride) functionalized graphene-modified glassy carbon electrode. J. Solid State Electrochem. 16, 1323–1331 (2012).
Ren, J., Kang, T.-F., Xue, R., Ge, C.-N. & Cheng, S.-Y. Biosensor based on a glassy carbon electrode modified with tyrosinase immmobilized on multiwalled carbon nanotubes. Microchim. Acta 174, 303–309 (2011).
Hashemnia, S., Khayatzadeh, S. & Hashemnia, M. Electrochemical detection of phenolic compounds using composite film of multiwall carbon nanotube/surfactant/tyrosinase on a carbon paste electrode. J. Solid State Electrochem. 16, 473–479 (2012).
Zhou, Y. & Zhi, J. Development of an amperometric biosensor based on covalent immobilization of tyrosinase on a boron-doped diamond electrode. Electrochem. Commun. 8, 1811–1816 (2006).
Shan, D., Zhu, M., Han, E., Xue, H. & Cosnier, S. Calcium carbonate nanoparticles: a host matrix for the construction of highly sensitive amperometric phenol biosensor. Biosens. & Bioelectron. 23, 648–654 (2007).
Huang, K., Yu, S., Wang, L., Gan, T. & Li, M. Simultaneous determination of hydroquinone, catechol and resorcinol based on graphene/cobalt hexacyanoferrate modified glassy carbon electrode. Acta Chim. Sin. 70, 735–740 (2012).
Deng, K., Li, C., Li, X. & Xu, G. Simultaneous electrochemical determination of hydroquinone and catechol at MWNTs and Cobalt(II) tetrakisphenylporphyrin modified electrode. Anal. Lett. 45, 883–893 (2012).
Guo, Q. H. et al. Simultaneous determination of catechol and hydroquinone using electrospun carbon nanofibers modified electrode. Sens. Actuator B Chem. 163, 179–185 (2012).
Huo, Z. et al. Sensitive simultaneous determination of catechol and hydroquinone using a gold electrode modified with carbon nanofibers and gold nanoparticles. Microchim. Acta 173, 119–125 (2011).
Li, M. et al. Sensitive and facile determination of catechol and hydroquinone simultaneously under coexistence of resorcinol with a Zn/Al layered double hydroxide film modified glassy carbon electrode. Electroanalysis 21, 1521–1526 (2009).
Ding, Y. P., Liu, W. L., Wu, Q. S. & Wang, X. G. Direct simultaneous determination of dihydroxybenzene isomers at C-nanotube-modified electrodes by derivative voltammetry. J. Electroanal. Chem. 575, 275–280 (2005).
Ahammad, A. J. S., Sarker, S., Rahman, M. A. & Lee, J. J. Simultaneous determination of hydroquinone and catechol at an activated glassy carbon electrode. Electroanalysis 22, 694–700 (2010).
Acknowledgements
LJ acknowledges the financial support from the program of China Scholarships Council (8260100311020).
Author information
Authors and Affiliations
Contributions
L.J. performed the experiments. L.J. and I.S. analysed the results. I.S. wrote the manuscript with the support from L.J. L.J., I.S. and J.F. discussed the results and commented on the final manuscript.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Jiang, L., Santiago, I. & Foord, J. Nanocarbon and nanodiamond for high performance phenolics sensing. Commun Chem 1, 43 (2018). https://doi.org/10.1038/s42004-018-0045-8
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s42004-018-0045-8
This article is cited by
-
A review of BioFET’s basic principles and materials for biomedical applications
Biomedical Engineering Letters (2021)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.