Table 3 Verification in practice.

From: Verification, analytical validation, and clinical validation (V3): the foundation of determining fit-for-purpose for Biometric Monitoring Technologies (BioMeTs)

Documentation you can expect Manufacturer should provide evidence of their BioMeT’s:
• Performance specifications for the integrated hardware
• Output data specifications
• Overview of software system tests
• Limitations to the verification testing
• e.g., specific known items that were not tested during verification
Clinical users’ questions answered by verification Is the performance of this BioMeT and each of its components sufficient to generate sample-level data of acceptable quality such that it can be used as an input to generate the processed data and downstream clinical measurement that I am interested in?