Abstract
Advances in instrumentation and tracer materials are still required to enable sensitive, accurate, and localized in situ 3D temperature monitoring by magnetic particle imaging (MPI). We have developed a high-resolution magnetic particle imaging instrument and implemented a low-noise multi-harmonic lock-in detection method to observe and quantify temperature variations in iron oxide nanoparticle tracers using the harmonic ratio method for determining temperature. Using isolated harmonics for MPI and temperature imaging revealed an apparent dependence of imaging resolution on harmonic number. Thus, we present experimental and simulation studies to quantify the imaging resolution dependence on temperature and harmonic number, and directly validate the fundamental origin of MPI imaging resolution on harmonic number based on the concept of a harmonic point-spread-function.
Similar content being viewed by others
Introduction
Magnetic particle imaging (MPI) is a tracer-based imaging modality that is experiencing rapid growth towards biomedical and clinical applications1,2,3,4,5,6,7,8,9,10,11,12,13,14,15. In this work, our efforts are directed at establishing magnetic nanoparticles (MNP) as a traceable nano-thermometer standard for in situ accurate temperature sensing and imaging, relying on MPI as a core technology. Since MPI provides a non-invasive, image-guided approach towards heat-based theranostics, numerous authors have discussed potential prospects for MPI-based temperature sensing16,17,18,19,20 and hyperthermia21,22,23,24,25,26 to complement the capabilities of ultrasound, fluoroscope, computed tomography (CT) and magnetic resonance imaging (MRI)27,28. Achieving accurate MPI-based thermal imaging and targeted heat-based therapy is demanding and requires a systematic evaluation of the effects of technical noise and temperature on detection sensitivity and imaging resolution.
While most MPI studies have focused on measurements of in situ concentration distributions of superparamagnetic tracers, nanoparticle thermometry has also been a longstanding goal of MPI18,26. Two key approaches for nanoparticle thermometry have been proposed: multi-color, system matrix-based18,29,30 and multi-harmonic-based measurements16,19,20,31. The system matrix approach requires measurements of calibration system matrices to reconstruct both concentration and temperature, and a linear interpolation step for intermediate temperatures18. Generating a fine grid of temperature calibration matrices is not only time-consuming, but achieving high temperature spatial resolution in image reconstruction is difficult if the variation of the particle response with temperature is small32. Here, we will focus on the multi-harmonic technique, where the temperature can be determined by measurement of two harmonics and their ratios16. The harmonic approach requires only a single calibration matrix of particle response across magnetic field and temperature, which is used for image reconstruction of temperature spatial profiles if the excitation magnetic field is known. This is akin to determining the reference function of resistance-based thermometers, where the analog in our case would be the harmonic response function of the magnetic nanoparticle sensors. By avoiding the interpolation required by the system matrix approach, the harmonic approach in principle could achieve more accurate results with a higher thermal spatial resolution. Even though nanoparticle thermometry based on measurement of individual harmonics has been presented previously17,19,31,33,34,35,36 the effects of temperature and harmonic number on MPI imaging (sensitivity and spatial resolution) are not quantitatively understood. By both experiments and simulations, we investigate the impact of temperature and harmonic measurement on MPI, which will have important implications for image reconstruction whenever temperature is involved.
For MPI, the magnetization response of MNP is driven into saturation by a sinusoidal excitation field and most often detected by inductive coil sensors. Typical MPI systems record the magnetization response in the time domain with a digitizer. While this procedure has the advantage of detecting a broadband harmonic signal, often exceeding 1 MHz, the broad bandwidth carries along with it excess broadband noise. Broadband noise-matching is a possible solution, but the complex impedance of the inductive coil introduces frequency-dependence of noise that presents complications for bandwidths exceeding 1 MHz37,38,39,40,41,42. Towards sensitive temperature measurements with the multi-harmonic technique, we use an alternative and more general approach by implementing a parallel lock-in detection scheme in which many harmonics are measured simultaneously while still leveraging the narrowband detection and low-noise properties of lock-in amplifiers. In contrast to Goodwill et al.37, we are detecting many harmonics from the particle response in a narrowband fashion, thereby overcoming the resolution-bandwidth trade-off. It is a general goal of magnetic imaging to reach the body noise limit, which can potentially be achieved at a high frequency (1 MHz to 100 MHz) by detecting at a high harmonic number and reaching coil resistance (thermal) noise. Here, we present our multi-harmonic lock-in detection approach to compare with the traditional digitizer-based measurement, and provide an in-depth characterization of the harmonic and temperature dependence of MPI.
Results
MPI characterization
The MPI instrumentation (Fig. 1a, see Methods section for more detail) was designed for quantitative thermal imaging using magnetic nanoparticles. To assess imaging spatial resolution, we use a glass phantom with dimensions of 8 mm x 12 mm (diameter x height). The phantom has four channels (1 mm diameter, 8 mm height) where MNP samples can be introduced and sealed by epoxy. To demonstrate 3D capabilities, Fig. 1b presents 3D MPI images of this glass phantom in two rotated orientations, with four clearly resolved channels. For a more systematic evaluation of spatial resolution limit, we used a 3D printed acrylic phantom shown in Fig. 2. Here, the sets of four channels have different edge-to-edge hole spacing to evaluate the spatial resolution based on the Rayleigh criterion. The 2D image in Fig. 2 reveals that we can resolve 250 μm in the x-direction, the direction of the highest gradient field. Our results are competitive with the highest resolution MPI instruments reported9,43,44. These results, along with the measured point-spread-function (PSF45,46) of a commercial MNP, Vivotrax, indicate that the magnitude of the gradient field, G, is \(\approx 20\) T/m for the x-axis and \(\approx 10\) T/m for the y-axis. These results are consistent with our COMSOL simulations of the permanent magnet pair. We note that such a high gradient field is challenging to transfer to clinical settings. However, it has been shown47 that MPI using super-ferromagnetic (SFM) nanoparticles with their narrow PSF can achieve superior spatial resolution despite a lower gradient field. But for commercial nanoparticles, this work and previous MPI reports9,43,44 show that a high gradient field is required to achieve sub-millimeter native spatial resolution.
Multi-harmonic detection for MPI
A voltage-based measurement system using an inductive coil (IC) sensor with a pre-amp has SNR given by37,40:
The quantity in the numerator is the product of the coil sensitivity, \(\varepsilon\), and voltage induced by the MNP’s modulated magnetization, dM/dt. The denominator denotes the noise sources from the sensor, an IC sensor with real (\(R_L\)) and complex impedance (\(Z_L\)), and the pre-amp at temperature T. The pre-amp has a voltage and current noise spectral density (NSD, \(V/\sqrt{Hz}\)), \(\bar{e}_n\) and \(\bar{i}_n\), respectively. Since \(\Delta f\) is the detection bandwidth, the best SNR is obtained by minimizing the bandwidth. This ideal scenario is, however, at odds with the MPI technique in that it is desirable to retain all information-carrying harmonics by using a high sampling rate (2 MSa/s for 1 MHz). Reducing the sampling rate (detection bandwidth) could improve SNR, but at the expense of spatial resolution45. To overcome the incompatibility of noise-matching and the bandwidth requirement, we implemented lock-in detection with parallel multi-harmonic capabilities that can improve SNR by permitting detection at arbitrarily high frequencies limited only by the lock-in amplifier bandwidth.
Relying on measurement of only a few harmonics has been proposed for in situ temperature measurement16,20, and also as a general modality for imaging13,42. Isolated harmonics can be sensitively measured with lock-in detection, where the noise bandwidth (1 Hz to 100 Hz range) is adjustable based on the desired acquisition speed and sensitivity. For detection of a single harmonic, the equivalent noise bandwidth (\(\Delta f_{LI}\)) is set by the acquisition time, \(\tau\), which is typically 10 ms in our experiment. For a first order filter, \(\Delta f_{LI} = 1/4\tau\), which gives a \(\Delta f_{LI} \approx 25\) Hz. The noise bandwidth of the Zurich lock-in (Zurich, HF2LI, 50 MHz) scales as \(\Delta f_{LI} \propto n^{-5/4}\), where n is an integer denoting the filter order number. Increasing the filter order is a simple way to lower the \(\Delta f_{LI}\) without increasing \(\tau\). For the equivalent measurement of harmonic amplitudes with a digitizer, the noise bandwidth is related to the acquisition time \(\tau\) by \(\Delta f_{D} = F_s/N = 1/\tau\), where \(F_s\) is the sampling rate and N is the sample length.
To compare the noise performance of multi-harmonic lock-in detection with the digitizer for MPI, we acquired images of the Vivotrax glass phantom using both techniques. Figure 3 show the 1D image slice across two channels of the glass phantom. The four panels show a comparison of the 3rd to 9th harmonic signals measured by the lock-in and digitizer. Six harmonics were recorded simultaneously in the lock-in case, but only four are presented for clarity. Here, the magnetization data was recorded at each spatial position using the same measurement time of \(\tau\) = 10 ms for each instrument. This acquisition time corresponds to noise bandwidths of 100 Hz and 25 Hz for the digitizer and lock-in, respectively, for each harmonic. From Eq. (1), the corresponding SNR enhancement is \(\sqrt{100/25}=2\). Combined with the lower noise floor of the lock-in relative to the digitizer, the improved SNR (signal peak divided by the RMS noise of the baseline) with the lock-in is most clearly observed in the higher harmonic data of Fig. 3. In general, the lock-in amplifier is not the limiting factor for MPI imaging speed since the integration time can be arbitrarily set as short as 1 \(\mu s\). The fundamental limitation in our imaging speed is the slow raster scan mechanisms (mechanical translation or electromagnetic shifting), which is slower than the Lissajous scanning at \(\approx\) 25 kHz in each axis. In principle, the lock-in amplifier can be implemented with the Lissajous scanning, as long as the lock-in data acquisition and Lissajous scanning are synchronized.
Temperature dependent MPI
Temperature affects the equilibrium magnetization and dynamics of MNP, which are encoded on the MPI signal. For MNP systems that display magnetization dynamics at experimentally-resolvable timescales, we have shown the effect of temperature on distinct Néel and Brownian mechanisms48. There have been numerous other reported studies on the temperature dependence of suspended magnetic nanoparticles16,17,18,19,20,26,31,33,34,35,36,49,50 that focused on determining the magnetization response sensitivity to temperature. Those that rely on harmonic measurements16,20,31,33,36, as opposed to the system matrix approach, to measure temperature typically select low harmonics (2nd to 5th) that display the largest signal to improve measurement sensitivity. However, these reports did not provide an assessment of how harmonic measurements impact imaging spatial resolution, a key component for thermal imaging performance. Here, to investigate the impact of temperature and harmonic-based measurements on MPI imaging resolution, we used our high resolution MPI instrument with the imaging phantom in Fig. 1 to characterize these dependencies.
Using Vivotrax samples in the glass phantom, we studied the imaging spatial resolution dependence on harmonic number and temperature. 1-D scans across the two channels of the phantom are presented in Fig. 4 for both the 7th and 9th harmonics. Only a modest change in the full-width-half-maximum (FWHM) of the peaks was observed as a function of temperature. A more quantitative assessment of temperature on spatial resolution can be ascertained from PSF data (Fig. 5). In comparison with measurements in Fig. 4, the PSF in Fig. 5 includes all harmonics and therefore exhibits significantly higher SNR. By fitting the PSF curves to a Voigt function, the temperature dependence of the FWHM was obtained (Fig. 5(b)). A modest slope of (0.041±0.006) mT/K is consistent with a small temperature dependence for data presented in Fig. 4. Finally, Fig. 5c shows the PSF for FFP scans in the positive and negative directions for Vivotrax at 298 K. The high symmetry of the PSF for both directions suggests a small relaxation effect at our measurement conditions51.
For thermal imaging applications using magnetic nanoparticles, measurement of harmonic ratios, rather than individual harmonic amplitudes, has been proposed to remove the concentration dependence16. We recorded 2D images using individual harmonics and also their ratios, and the data is presented in Fig. 6. In the case of the 5th/3rd and 7th/3rd harmonic ratios, the higher harmonic dominates the spatial resolution of the ratio image, which mostly resembles the image of the higher harmonic number. In general, images obtained with either individual harmonics or harmonic ratios clearly show a harmonic dependence of spatial resolution, and suggests a corresponding PSF dependence on harmonic number. The concept of a harmonic PSF is crucial for this imaging modality, and will be discussed in detail in the following sections.
Harmonic dependence of spatial resolution
Most MPI instruments typically utilize all harmonics for image reconstruction, which obscures the impact of individual harmonics on spatial resolution. The effect of harmonics on the reconstructed image has been introduced previously2, especially in the context of feedthrough (1st harmonic) suppression52. More importantly, these studies reveal a direct mapping of harmonic frequency space to spatial frequency in the image space for sinusoidal excitation. Our work extends upon these analyses by experimentally determining the impact of individual harmonics on image reconstruction, a new concept denoted as the harmonic PSF.
To demonstrate the existence of a harmonic PSF, we acquired MPI images reconstructed using individual harmonic amplitudes by means of (1) the spectrum of the digitized time signal and (2) the lock-in amplifier demodulated signal. Using both approaches, the apparent dependence of resolution on harmonic number was observed for two different MNP samples. Here, a glass phantom with four channels was used as a sample holder for 70 nm Synomag (Fig. 7) and for Vivotrax (Fig. 8). These two widely used MNP commercial systems have significant differences in their magnetic properties but both exhibit improved spatial resolution with increasing harmonic number. Although the crystallite diameters for Vivotrax and Synomag are both about 5 nm, Vivotrax shows non-uniform aggregation, but Synomag is a “nanoflower” with well-defined cluster geometry53. They exhibit differences in their MPS signatures, AC susceptibility, and PSF widths. Figure 7a shows a strong resolution dependence on harmonic number, as the four wells become more spatially resolved with increasing harmonic number. To quantitatively analyze the resolution dependence, an acrylic phantom with two sample wells spaced by 0.5 mm (edge-to-edge) was machined and imaged in 1D (Fig. 7b). The spatial resolution was estimated by applying the Houston criterion54 to the fitted FWHM of the measured profile, and the result is plotted in Fig. 7c. The spatial resolution displays an exponential improvement with increasing harmonic number and saturates at higher harmonics. A similar analysis was done for Vivotrax tracers (Fig. 8) with an observed exponential-like dependence of resolution on harmonic number.
Harmonic PSF theory and experiment
To derive the fundamental origin of the harmonic PSF, or the spatial variation of the PSF with respect to harmonic number, we start by describing the 1D MPI signal of a spatial distribution \(\rho (x)\) of identical particles by:
where \(s_{part}(x,t)\) is the signal from a single particle, given by:45,52
This is the voltage response for the case of an inductive receive coil where \(m_{part}\left( H_\textit{Eff}(x,t)\right)\) is the particle magnetic moment component in the direction of the applied field and \(\varepsilon\) is the coil sensitivity. In the case of periodic excitation with an effective magnetic field \(H_{\textit{Eff}}(x,t)\), Eq. (3) can be decomposed as a Fourier series:
where \(S_n(x)\) is the spatially-dependent Fourier component and \(\omega _o=2\pi f_o\) is the angular frequency of the drive field.
The effective field \(H_\textit{Eff}\) is a combination of a spatially varying gradient field \(H_S(x)\), with gradient G, and a time-varying excitation field \(H_D(t)\), with frequency \(f_o=1/T\) and amplitude A, such that \(H_{\textit{Eff}}(x,t)=H_S(x)-H_D(t)=Gx-A\cos (\omega _o t)\). The Fourier coefficients are then given by:
where \(m'_{part}(H_{\textit{Eff}})\) is the derivative of \(m_{part}\) with respect to its argument. Along the drive trajectory, we can use the relation \(t(H_D)=\mp \frac{1}{\omega _o}\arccos \left( \frac{H_D}{A}\right)\) to parameterize by field in the integral of Eq. (5). Arccos is only defined on the range \([0,\pi ]\), so the first half period can be identified with negative values for time and the second half with positive values. Following the development by Rahmer:2
If we set the condition that
then the limits of integration in Eq. (6) can be set to infinity. Recalling that \(H_S(x)=Gx\), we arrive at the following form to describe the spatial dependence of the Fourier coefficients, \(S_n(x)\):
Equation (7), which closely resembles the Chebyshev polynomials that are truncated for \(|x|>A/G\), is significant because these polynomials form a complete orthogonal basis which can describe any MPI harmonic image2,52.
A numerical calculation of the analytical form of Eq. (8) is shown in Fig. 9 for a nanoparticle with a Langevin response and PSF FWHM of 13 mT, typical of commercial Vivotrax samples. In the Langevin case, the magnetic moment component in the direction of the applied field is:
Here, \(\mu _0\) is the vacuum permeability, m the magnitude of the particle magnetic moment. The moment is the saturation magnetization, \(M_s\), multiplied by the volume of the magnetic core, and \(\mathcal {L}\) is the Langevin function. In Fig. 9a, the PSF is approximated as a Lorentzian function45. Figure 9b shows the components of the Chebyshev polynomials corresponding to the particle harmonics with the condition set by Eq. (7). The final result calculated using Eq. (8) is presented in Fig. 9c and shows the spatial variation of the PSF for the 1st to 7th harmonics. Finally, the explicit temperature dependence of each individual harmonic PSF can be accounted for by using the analytical form for the derivative of the Langevin function45 through the argument \(\xi\) from Eq. 9:
A numerical calculation of the temperature-dependence for each harmonic PSF using Eq. (10) is displayed in Fig. 10. Here, a small relative change in the shape of the PSF from 250 K to 300 K is consistent with the measured data for the 7th and 9th harmonic in Fig. 4. In the scenario of (1) a low amplitude excitation magnetic field (\(< 10\) mT)46 and (2) slow (\(\approx\) 10 Hz) electromagnetic shifting or mechanical translation for image rastering, the Langevin model provides a reasonable approximation for the magnetization response because relaxation effects are minor, as evident in the minimal asymmetry observed in the negative and positive scanning of the PSF (Fig. 5c). In our PSF measurement and MPI image acquisition, the relatively smaller amplitude, but higher frequency excitation field traverses a smaller portion of the M vs H curve compared to the 8-fold larger bias field, which dominates the FFP shifting and the shape of the PSF.
According to the previous analysis, the MPI signal can be represented by a Fourier series, whose components define the harmonic PSF. To compare with the predicted form of these harmonics, we experimentally measure the individual harmonic PSF components directly by applying a slow varying (10 Hz) bias field (triangle waveform) at 80 mT peak-to-peak superimposed on the excitation field, similar to the MPS-type procedure by Croft et al.46. Different from previous PSF reconstruction procedures that include all harmonics, we resolved the individual harmonic PSFs by two distinct digital processing methods: (1) short-time Fourier transform (STFT) of the digitized time response signal and (2) time-swept lock-in detection.
For method (1) using the time signal f(t) from the digitizer, the STFT is implemented by first binning the digitized data in defined Hamming window segments w(t). At each time shift t, a new window \(w(t'-t)\) is applied and the FT is calculated of \(f(t')w(t'-t)\):
This procedure produces a spectrogram, or a plot of the Fourier spectrum at each time segment window. With the bias field turned on, the harmonic spectrum will only appear when the bias field amplitude is near zero, that is when the MNP are not fully saturated. For method (2), the harmonic PSF is obtained with the lock-in by implementing a time-domain acquisition of the output at the specified demodulation frequency (\(1f_0\), \(3f_0\), \(5f_0\), etc.) to resolve the time-swept PSF at each harmonic number.
Figure 11 is a step-by-step graphical layout of the STFT procedure with detailed descriptions in the figure caption. Harmonic PSF data taken with the digitizer (method 1, STFT) and lock-in amplifier (method 2) are plotted in Fig. 12a,b. The general shape of the PSF obtained using the two independent approaches display qualitative agreement for each harmonic, and provides experimental justification for the origin of the harmonic dependence of PSF. By inspection, it can be seen that the harmonic PSF displays a decreasing FWHM (higher spatial resolution) with increasing harmonic number, and the FWHM also converges to an asymptotic value. The insets show the fitted FWHM using a Lorentzian function for each harmonic PSF. This trend explains our imaging results for nanoparticle samples in Figs. 7 and 8, and provides experimental confirmation of the fundamental harmonic frequency to spatial frequency correspondence general to MPI, analogous to the k-space to image-space transformation in MRI55.
Simulations of harmonic PSF experiments
To obtain additional validation and insight into the origin of the experimentally observed spatial resolution dependence on harmonic number, we performed the following direct simulation study, taking into account the experimental parameters of our MPS-type harmonic PSF measurements. The magnetic moment of a single magnetic nanoparticle is modeled using the Langevin function, and the PSF is calculated as the time-derivative of the Langevin function45. Finally, the harmonic dependence of the PSF is calculated numerically for both the STFT and lock-in methods using simulated bias and excitation fields, and including information such as the STFT window width and lock-in integration time constant. Measured and simulated harmonic PSFs presented in Figs. 12a,b and c,d, respectively, show that the PSF FWHM, defined by the broad envelope at each harmonic, is narrower with increasing harmonic number, with nearly an exponential-like improvement in resolution56. The general shape of the harmonic PSF shows qualitative agreement with the calculated PSF shown in Fig. 9. More importantly, the simulations provide intuition for the origin of the narrower PSF for higher harmonics: as the FFP location moves further from the particle location, the signal amplitude of the higher harmonics decreases faster than that of the lower harmonics. In other words, as the FFP position moves away from the particle location, the magnetization response becomes more linear.
We also simulated the temperature dependence of the full PSF to validate the observations in Fig. 5. Note that due to the explicit temperature dependence of \(\xi\) in Eq. (9), the magnetization of a single particle shows a corresponding temperature dependence, even if the saturation magnetization was set to be constant. Figure 13a shows the simulated temperature dependence of the PSF over the range 250 K to 300 K, in which only a modest change in the PSF FWHM with temperature (0.043 mT/K) was observed, in good agreement with the Vivotrax data from Fig. 5. Despite its generality, the Langevin function does not account for relaxation effects that may impact the magnetization response in MPI51,57,58. To account for relaxation, the Langevin response in Eq. (9) was modified to include a relaxation kernel51,59:
Here, \(\tau\) is the effective relaxation time constant and u(t) is the Heaviside function. Despite the small impact of relaxation effects on our measurements, both Brownian and Néel relaxation time constants are affected by temperature58,60 and may introduce additional changes to the harmonic PSF. To quantify the impact of temperature, we used the magnetization model with relaxation (Eq. (12)) and computed the harmonic PSF with varying relaxation time constants (\(\tau\) = 100 ns to 100 \(\mu s\)) expected within our measurement temperature range (Supplementary Information, SI). Relaxation has a negligible impact on the harmonic PSF FWHM (Fig. 13b,c), but reduces the harmonic PSF’s amplitude at increasing harmonic number (SI). The latter is consistent with other reports of relaxation effects decreasing the signal amplitude46,61.
Discussion
Towards the goal of advancing temperature measurement and thermal imaging with magnetic nanoparticles, this work comprised experimental and theoretical studies on the harmonic and temperature dependence of MPI and analyses of their fundamental origin. To capture small temperature variations of the particle response with improved sensitivity, we implemented a multi-harmonic lock-in detection scheme well-tailored for the harmonic analysis approach to temperature determination. Analysis of individual harmonics in MPI revealed an apparent dependence of imaging resolution with harmonic number—the harmonic PSF. For the harmonic PSF, we have shown agreement between the analytical form, numerical simulations of experiments, and experimental results using both STFT and lock-in methods. Harmonic PSF width decreases as harmonic number increases, consistent with theory, simulations, and imaging spatial resolution in our MPI scanner measurements. Our experimental validation of the harmonic PSF establishes a direct connection between the imaging spatial resolution and harmonic measurement for MPI, which is required for understanding and predicting thermal imaging resolution using the harmonic measurement approach. This analysis of the harmonic PSF is the first step towards understanding the thermal spatial resolution. A more direct and systematic characterization of spatial resolution with temperature variations, or a temperature PSF using isolated harmonics and harmonic ratios, requires (1) a thermal gradient phantom and (2) extension of the harmonic PSF theory to the “temperature PSF” to validate measurements of a thermal gradient. This would also allow for quantification of temperature measurement resolution due to image reconstruction artifacts, e.g. from gridding errors due to particles experiencing non-zero magnetic field in the vicinity of the FFP. These studies are currently underway and beyond the scope of this work. Finally, this work also validates the correspondence of parallel multi-harmonic lock-in and digitizer acquisition for MPI imaging and thermometry, with the former displaying superior noise performance and being straightforward to implement. Further improvements in detection sensitivity will require developments of both detection and lock-in electronics that are noise-matched to the receive chain and operate at high frequencies.
Methods
Similar to our temperature-tunable AC magnetometers48,62, the MNP sample holder is machined out of thermally conductive Shapal ceramic. The temperature of the sample is tuned by water-cooling with a recirculating chiller, and the temperature is measured by two 100 ohm platinum resistance thermometers. An additional water-cooling line is used to control the temperature of the excitation solenoid coil to ensure thermal stability over long acquisition times. A CAD drawing of the MPI instrument is presented in Fig. 1a. The selection field (gradient field: 20 T/m in x-axis, 10 T/m in y-xis and z-axis) is generated from a pair of permanent magnets (diameter = 50 mm, height = 50 mm, N52 grade NdFeB) separated by \(\approx\) 30 mm to 40 mm (bore size) to produce the field-free-point (FFP). The FFP is fixed in our MPI imager, and spatial scanning is achieved by mechanical translation of a 3D stage in which the sample holder and excitation coils are mounted. This image acquisition scheme relies on spatial mechanical step-scanning at a maximum rate of \(\approx 5\) Hz, which makes it more similar to the x-space method12,13,52,63 that uses low frequency (\(\approx\) 10 Hz) shift coils rather than the Lissajous pattern-based electromagnetic shifting with three orthogonal fields near 25 kHz1,7,39.
The transmit chain is a resonant circuit at the excitation frequency \({f}_0\) = 31.75 kHz with a second order low-pass filter to attenuate (50 dB) the higher frequency feedthrough signals. Magnetization response from AC excitation (10 \(\normalsize mT_p\)) is recorded with an inductive receive coil, wound as a first-order gradiometer in series with an additional \(\approx 30\) dB notch filter at the excitation frequency. This output is directly connected to a SR560 pre-amplifier, whose output is connected in series to an impedance matched, cascaded notch filter stage with \(\approx 60\) dB of attenuation at \({f}_0\) and \(\approx 40\) dB at \(2 {f}_0\). The filtered magnetization response is directed to either a digitizer (GaGe, CSE161G2, 16 bit, 1 GSa/s) or lock-in amplifier (Zurich, HF2LI, 50 MHz) to measure the harmonic amplitudes for image reconstruction. For the digitizer, the harmonic amplitudes are determined from the Fourier transform of the time domain signal. The multi-frequency capability of the HF2LI lock-in amplifier allows the recording of six independent harmonics simultaneously in a parallel fashion.
Data availability
The datasets used and/or analysed during the current study available from the corresponding author on reasonable request.
References
Gleich, B. & Weizenecker, J. Tomographic imaging using the nonlinear response of magnetic particles. Nature 435, 1214–1217. https://doi.org/10.1038/nature03808 (2005).
Rahmer, J., Weizenecker, J., Gleich, B. & Borgert, J. Signal encoding in magnetic particle imaging: Properties of the system function. BMC Med. Imaging 9, 4. https://doi.org/10.1186/1471-2342-9-4 (2009).
Goodwill, P. W., Konkle, J. J., Saritas, E. U., Zheng, B. & Conolly, S. M. Projection X-space magnetic particle imaging patrick. IEEE Trans. Med. Imaging 31, 1076–1085. https://doi.org/10.1038/jid.2014.371 (2012).
Saritas, E. et al. Magnetic particle imaging (MPI) for NMR and MRI researchers. J. Magn. Reson. 23, 1–7. https://doi.org/10.1038/jid.2014.371 (2013).
Vogel, P. et al. Traveling wave magnetic particle imaging. IEEE Trans. Med. Imaging 33, 400–407. https://doi.org/10.1109/TMI.2013.2285472 (2014).
Panagiotopoulos, N. et al. Magnetic particle imaging: Current developments and future directions. Int. J. Nanomed. 10, 3097–3114 (2015).
Gräfe, K., von Gladiss, A., Bringout, G., Ahlborg, M. & Buzug, T. M. 2d images recorded with a single-sided magnetic particle imaging scanner. IEEE Trans. Med. Imaging 35, 1056–1065. https://doi.org/10.1109/TMI.2015.2507187 (2016).
Knopp, T., Gdaniec, N. & Möddel, M. Magnetic particle imaging: From proof of principle to preclinical applications. Phys. Med. Biol. 62, R124. https://doi.org/10.1088/1361-6560/aa6c99 (2017).
Vogel, P. et al. Micro-traveling wave magnetic particle imaging - sub-millimeter resolution with optimized tracer LS-008. IEEE Trans. Magn.https://doi.org/10.1109/TMAG.2019.2924198 (2019).
Talebloo, N., Gudi, M., Robertson, N. & Wang, P. Magnetic particle imaging: Current applications in biomedical research. J. Magn. Reson. Imaging 51, 1659–1668. https://doi.org/10.1002/jmri.26875 (2020).
Han, X. et al. The applications of magnetic particle imaging: From cell to body. Diagnostics (Basel) 10, 800 (2020).
Kurt, S., Muslu, Y. & Saritas, E. U. Partial FOV center imaging (PCI): A robust X-space image reconstruction for magnetic particle imaging. IEEE Trans. Med. Imaging 39, 3441–3450 (2020).
Mason, E. E. et al. Concept for using magnetic particle imaging for intraoperative margin analysis in breast-conserving surgery. Sci. Rep. 11, 13456. https://doi.org/10.1038/s41598-021-92644-8 (2021).
Tay, Z. W. et al. Magnetic particle imaging: An emerging modality with prospects in diagnosis, targeting and therapy of cancer. Cancers 13, 5285 (2021).
Chandrasekharan, P. et al. Non-radioactive and sensitive tracking of neutrophils towards inflammation using antibody functionalized magnetic particle imaging tracers. Nanotheranostics 5, 240–255 (2021).
Weaver, J. B., Rauwerdink, A. M. & Hansen, E. W. Magnetic nanoparticle temperature estimation. Med. Phys. 36, 1822–1829. https://doi.org/10.1118/1.3106342 (2009).
Garaio, E., Collantes, J. M., Garcia, J. A., Plazaola, F. & Sandre, O. Harmonic phases of the nanoparticle magnetization: An intrinsic temperature probe. Appl. Phys. Lett.https://doi.org/10.1063/1.4931457 (2015).
Stehning, C. & Gleich, B. R. J. Simultaneous magnetic particle imaging (MPI) and temperature mapping using multi-color MPI. Int. J. Magn. Part. Imaging 2, 1–6 (2016).
Wells, J., Paysen, H., Kosch, O., Trahms, L. & Wiekhorst, F. Temperature dependence in magnetic particle imaging. AIP Adv.https://doi.org/10.1063/1.5004506 (2018).
Zhong, J., Schilling, M. & Ludwig, F. Excitation frequency dependence of temperature resolution in magnetic nanoparticle temperature imaging with a scanning magnetic particle spectrometer. J. Magn. Magn. Mater. 471, 340–345. https://doi.org/10.1016/j.jmmm.2018.09.112 (2019).
Franke, J. et al. System characterization of a highly integrated preclinical hybrid MPI-MRI scanner. IEEE Trans. Med. Imaging 35, 1993–2004. https://doi.org/10.1109/TMI.2016.2542041 (2016).
Bauer, L. M., Situ, S. F., Griswold, M. A. & Samia, A. C. S. High-performance iron oxide nanoparticles for magnetic particle imaging - guided hyperthermia (HMPI). Nanoscale 8, 12162–12169. https://doi.org/10.1039/C6NR01877G (2016).
Hensley, D. et al. Combining magnetic particle imaging and magnetic fluid hyperthermia in a theranostic platform. Phys. Med. Biol. 62, 3483–3500. https://doi.org/10.1088/1361-6560/aa5601 (2017).
Tay, Z. W. et al. Magnetic particle imaging-guided heating in vivo using gradient fields for arbitrary localization of magnetic hyperthermia therapy. ACS Nano 12, 3699–3713. https://doi.org/10.1021/acsnano.8b00893 (2018).
Chandrasekharan, P. et al. Using magnetic particle imaging systems to localize and guide magnetic hyperthermia treatment: Tracers, hardware, and future medical applications. Theranostics 10, 2965–2981. https://doi.org/10.7150/thno.40858 (2020).
Salamon, J. et al. Visualization of spatial and temporal temperature distributions with magnetic particle imaging for liver tumor ablation therapy. Sci. Rep. 10, 7480. https://doi.org/10.1038/s41598-020-64280-1 (2020).
Weidensteiner, C. et al. Real-time MR temperature mapping of rabbit liver in vivo during thermal ablation. Magn. Reson. Med. 50, 322–330. https://doi.org/10.1002/mrm.10521 (2003).
Perälä, J. et al. MRI-guided laser ablation of neuroendocrine tumor hepatic metastases. Acta Radiol. Short Rep. 3, 2047981613499753 (2014).
Rahmer, J., Halkola, A., Gleich, B., Schmale, I. & Borgert, J. First experimental evidence of the feasibility of multi-color magnetic particle imaging. Phys. Med. Biol. 60, 1775–1791. https://doi.org/10.1088/0031-9155/60/5/1775 (2015).
Pantke, D., Holle, N., Mogarkar, A., Straub, M. & Schulz, V. Multifrequency magnetic particle imaging enabled by a combined passive and active drive field feed-through compensation approach. Med. Phys. 46, 4077–4086. https://doi.org/10.1002/mp.13650 (2019).
Rauwerdink, A. M., Hansen, E. W. & Weaver, J. B. Nanoparticle temperature estimation in combined ac and dc magnetic fields. Phys. Med. Biol. 54, L51. https://doi.org/10.1088/0031-9155/54/19/L01 (2009).
Quelhas, K. N., Henn, M.-A., Farias, R. C., Tew, W. L. & Woods, S. I. Parallel 3d temperature image reconstruction using multi-color magnetic particle imaging (mpi). AIP Conference Proceedings (2023, In Press).
Zhong, J., Dieckhoff, J., Schilling, M. & Ludwig, F. Influence of static magnetic field strength on the temperature resolution of a magnetic nanoparticle thermometer. J. Appl. Phys.https://doi.org/10.1063/1.4964696 (2016).
Draack, S., Viereck, T., Kuhlmann, C., Schilling, M. & Ludwig, F. Temperature-dependent MPS measurements. Int. J. Magn. Part. Imaging 3(1), 3–6. https://doi.org/10.18416/ijmpi.2017.1703018 (2017).
Draack, S. et al. Determination of dominating relaxation mechanisms from temperature-dependent Magnetic Particle Spectroscopy measurements. J. Magn. Magn. Mater. 474, 570–573. https://doi.org/10.1016/j.jmmm.2018.11.023 (2019).
Janssen, K.-J., Zhong, J., Viereck, T., Schilling, M. & Ludwig, F. Quantitative temperature visualization with single harmonic-based magnetic particle imaging. J. Magn. Magn. Mater. 563, 169915. https://doi.org/10.1016/j.jmmm.2022.169915 (2022).
Goodwill, P. W., Scott, G. C., Stang, P. P. & Conolly, S. M. Narrowband magnetic particle imaging. IEEE Trans. Med. Imaging 28, 1231–1237. https://doi.org/10.1109/TMI.2009.2013849 (2009).
Graeser, M., Knopp, T., Grüttner, M., Sattel, T. F. & Buzug, T. M. Analog receive signal processing for magnetic particle imaging. Med. Phys. 40, 1–9. https://doi.org/10.1118/1.4794482 (2013).
Weizenecker, J., Gleich, B., Rahmer, J., Dahnke, H. & Borgert, J. Three-dimensional real-time in vivo magnetic particle imaging. Phys. Med. Biol. 54, L1–L10. https://doi.org/10.1088/0031-9155/54/5/L01 (2009).
Zheng, B. et al. Optimal broadband noise matching to inductive sensors: Application to magnetic particle imaging. IEEE Trans. Biomed. Circuits Syst. 11, 1041–1052. https://doi.org/10.1109/TBCAS.2017.2712566 (2017).
Malhotra, A., Schwegmann, H., Schumacher, J., Chen, X. & Buzug, T. M. Fully differential low noise amplifier for MPI/MPS. Int. J. Magn. Part. Imaging 6, 1–3. https://doi.org/10.18416/IJMPI.2020.2009006 (2020).
Mason, E. E. Magnetic Particle Imaging for Intraoperative Breast Cancer Margin Assessment and Functional Brain Imaging. Ph.D. thesis, Massachusetts Institute of Technology (2020).
Arami, H. et al. Tomographic magnetic particle imaging of cancer targeted nanoparticles. Nanoscale 9, 18723–18730 (2017).
Trisnanto, S. B. & Takemura, Y. Modulating relaxation responses of magnetic nanotracers for submillimeter imaging. Appl. Phys. Lett. 115, 123101. https://doi.org/10.1063/1.5102081 (2019).
Goodwill, P. W. & Conolly, S. M. The x-space formulation of the magnetic particle imaging process: 1-d signal, resolution, bandwidth, snr, sar, and magnetostimulation. IEEE Trans. Med. Imaging 29, 1851–1859. https://doi.org/10.1109/TMI.2010.2052284 (2010).
Croft, L. R. et al. Low drive field amplitude for improved image resolution in magnetic particle imaging. Med. Phys. 43, 424–435. https://doi.org/10.1118/1.4938097 (2016).
Tay, Z. W. et al. Superferromagnetic nanoparticles enable order-of-magnitude resolution & sensitivity gain in magnetic particle imaging. Small Methods 5, 2100796. https://doi.org/10.1002/smtd.202100796 (2021).
Bui, T. Q. et al. Advanced characterization of magnetization dynamics in iron oxide magnetic nanoparticle tracers. Appl. Phys. Lett. 120, 012407. https://doi.org/10.1063/5.0077016 (2022).
Hankiewicz, J. H., Celinski, Z., Stupic, K. F., Anderson, N. R. & Camley, R. E. Ferromagnetic particles as magnetic resonance imaging temperature sensors. Nat. Commun. 7, 12415 (2016).
Abel, F. M. et al. Thermosensitivity through exchange coupling in ferrimagnetic/antiferromagnetic nano-objects for magnetic-based thermometry. ACS Appl. Mater. Interfaces. 15, 13439–13448. https://doi.org/10.1021/acsami.2c19673 (2023).
Croft, L. R., Goodwill, P. W. & Conolly, S. M. Relaxation in x-space magnetic particle imaging. IEEE Trans. Med. Imaging 31, 2335–2342. https://doi.org/10.1109/TMI.2012.2217979 (2012).
Lu, K., Goodwill, P. W., Saritas, E. U., Zheng, B. & Conolly, S. M. Linearity and shift invariance for quantitative magnetic particle imaging. IEEE Trans. Med. Imaging 32, 1565–1575. https://doi.org/10.1109/TMI.2013.2257177 (2013).
Rietberg, M. T. et al. Modelling of dynamic behaviour in magnetic nanoparticles. Nanomaterialshttps://doi.org/10.3390/nano11123396 (2021).
Houston, W. V. A compound interferometer for fine structure work. Phys. Rev. 29, 478–484. https://doi.org/10.1103/PhysRev.29.478 (1927).
Sodickson, A. & Cory, D. G. A generalized k-space formalism for treating the spatial aspects of a variety of nmr experiments. Prog. Nucl. Magn. Reson. Spectrosc. 33, 77–108. https://doi.org/10.1016/S0079-6565(98)00021-1 (1998).
Henn, M., Bui, T. & Woods, S. Investigating the harmonic dependence of MPI resolution. Int. J. Magn. Part. Imaginghttps://doi.org/10.18416/IJMPI.2023.2303059 (2023).
Tay, Z. W., Hensley, D. W., Vreeland, E. C., Zheng, B. & Conolly, S. M. The relaxation wall: Experimental limits to improving MPI spatial resolution by increasing nanoparticle core size. Biomed. Phys. Eng. Express 3, 035003 (2017).
Arsalani, S. et al. Temperature dependent magnetorelaxometry of magnetic nanoparticle ensembles. Phys. Med. Biol. 68, 175017. https://doi.org/10.1088/1361-6560/acec28 (2023).
Henn, M.-A., Quelhas, K. N., Bui, T. Q. & Woods, S. I. Improving model-based MPI image reconstructions: Baseline recovery, receive coil sensitivity, relaxation and uncertainty estimation. Int. J. Magn. Part. Imaging 2022(8), 1–9. https://doi.org/10.18416/IJMPI.2022.2208001 (2022).
Deissler, R. J., Wu, Y. & Martens, M. A. Dependence of brownian and néel relaxation times on magnetic field strength. Med. Phys. 41, 012301. https://doi.org/10.1118/1.4837216 (2014).
Tay, Z. W. et al. Pulsed excitation in magnetic particle imaging. IEEE Trans. Med. Imaging 38, 2389–2399. https://doi.org/10.1109/TMI.2019.2898202 (2019).
Bui, T. Q., Tew, W. L. & Woods, S. I. Ac magnetometry with active stabilization and harmonic suppression for magnetic nanoparticle spectroscopy and thermometry. J. Appl. Phys. 128, 224901. https://doi.org/10.1063/5.0031451 (2020).
Goodwill, P. W. et al. X-space MPI: Magnetic nanoparticles for safe medical imaging. Adv. Mater. 24, 3870–3877 (2012).
Acknowledgements
We acknowledge Jo Wu for providing the glass imaging phantom. We thank Angela Hight Walker, Cindi Dennis, Michael Donahue, Adam Biacchi, Frank Abel, Klaus Quelhas, and Eli Mattingly for helpful discussions. We acknowledge financial support from the NIST Innovation in Measurement Science (IMS) program.
Reference is made to commercial products to adequately specify the experimental procedures involved. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology, nor does it imply that these products are the best for the purpose specified.
Author information
Authors and Affiliations
Contributions
T.Q.B. and S.I.W. conceived the experiment(s); T.Q.B. conducted the experiments; M.A.C. and W.L.T. provided the imaging phantoms and equipment. T.Q.B., M.A.H., S.I.W. analyzed the results; All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Bui, T.Q., Henn, MA., Tew, W.L. et al. Harmonic dependence of thermal magnetic particle imaging. Sci Rep 13, 15762 (2023). https://doi.org/10.1038/s41598-023-42620-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-023-42620-1
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.