Introduction

Modern methods of treating diseases of the musculoskeletal system place high demands on materials used in medicine, including those used for bone implants fabrication. These materials should have many properties (e.g. mechanical, physicochemical, corrosive), but most of all they should be biocompatible (biotolerant) and non-toxic1.

In recent years, biocompatible and biodegradable implants, such as polymer, ceramic, and metallic (e.g. magnesium-based alloys), which degrade slowly after implantation, have been extensively researched. This kind of biomaterial is more beneficial because it reduces the cost of healthcare and eliminates the next surgery to remove implants2,3,4. Among these biomaterials proposed for bone implants, calcium-based alloys are found. In the literature, there is not as much work on the use of Ca-based alloys with various additives, such as magnesium, zinc, ytterbium, boron, and gold in implantology. Therefore, it seems to be very interesting to present the behavior of such alloys in the human body environment, especially from a biological point of view.

It should be mentioned that calcium and most calcium-based alloying elements are present in the human body as micronutrients (e.g. zinc, boron) and macronutrients (e.g. calcium, magnesium). These are elements necessary for healthy functioning (also known as bioelements). Calcium is one of the minerals present in the body, necessary for its proper functioning. The total amount of calcium in an adult human body is approximately 1200 g, which is 1.5% of body weight. 99% of Ca is present in bones in the form of bound to apatite, the remainder is present in the ionized form in the intracellular and extracellular fluid. Furthermore, calcium is an inexpensive element with low density (1.55 g/cm3), which is essential for the application of short-term calcium-based alloys in orthopedic implants5. Biodegradable magnesium is highly biocompatible. It has mechanical properties similar to natural bones. Furthermore, Mg as an alloying element improves mechanical properties and corrosion resistance. This extremely important macroelement has a multidirectional effect on the human body5. Therefore, it has been used in the treatment and prevention of many diseases. Most of all, magnesium is essential for the proper structure of bone tissue. In addition to calcium, vitamin D, and phosphorus, it affects proper bone mineralization, protecting human against osteoporosis. Mg is responsible for the activity of osteoblasts and osteoclasts, i.e. cells that are involved in bone metabolism4.

The next element, zinc, is a catalyst for many reactions. It participates in the transformation of proteins, fats, and carbohydrates. Zn is necessary to maintain the stability of cell membranes. It has regulatory and structural functions6. In addition, zinc participates in energy transformations and is an element necessary to maintain proper body weight. In recent years, ytterbium, as a rare earth element, has been used as a promising alloy additive to modify microstructures and improve the mechanical properties of traditional magnesium alloys7. It has also been noted that ytterbium salts stimulate metabolism. Yu et al.8 stated that the addition of 4% Yb to the Mg66Zn30Ca4 alloy decreases the cytotoxicity in in vitro studies. Boron is a very important element that has a positive effect on immunity and has anti-inflammatory properties. It has a positive effect on bone health and prevention of arthritis, playing a potential role in bone formation9,10, calcium, vitamin D, magnesium, phosphorus, and fluoride metabolism (ensuring that these micronutrients are at the correct levels and that vitamins are better absorbed). These make boron useful in the prevention of osteoporosis and rheumatic diseases. Some works suggest that boron may be a natural anticancer agent11. Gold as an alloying additive also has anticancer properties12. The results show that Au compounds are more cytotoxic to cancer cells than to healthy ones13. In addition, gold compounds show healing properties. Due to reactions with skin lipids, gold nanoparticles are able to open the stratum corneum and penetrate through it14, and the degree of penetration depends on their physicochemical properties15.

The correct selection of the composition of a potential biomaterial is extremely important. Improper selection of a metal material or alloy used in an implantology can result in an allergic reaction in the body, known as metallosis16. It is caused by the passage of implant fragments into its environment (which is usually caused by abrasive wear on the implant surface), migration of corrosion products from the implant surface to the surrounding environment, or by the reaction of metal ions in contact with human body fluids. Metallosis can occur in several ways. There is quiet, sharp, and discreet metallosis.

The basic property of the material, which determines its suitability for biomedical applications, is biotolerance. Biocompatibility evaluation17, according to an international standard, requires in vitro tests on isolated cells or tissues, as well as in vivo tests in animals and preclinical trials18. In vitro tests are a way to initially determine the behavior of living cells in the presence of the tested biomaterial. One of the basic in vitro tests is a biological cytotoxicity of materials used in medicine. Toxic substances released by the material in contact with tissues or body fluids can damage the cell membrane, change the metabolic activity of cells, and damage the genetic material of the cell. Cytotoxicity is the evaluation of the effect of a potential biomaterial on cells observed under a microscope after a predetermined period of exposure or by the activity of enzymes proving the viability of the cells. However, it should be mentioned that the cytotoxicity test does not ensure the maintenance of all the conditions of the physiological environment, but it is a study evaluating the behavior of living cells in contact with the biomaterial. Each medical device or the materials used for its production, also intended for implantation, are compliant with biological tissues, cells, and body fluids. The types of methods for in vitro and in vivo tests related to the assessment of biological compliance of medical devices constitute the content of the applicable PN-EN ISO 10993-1:2010 standard (Biological assessment of medical devices—Part 1: Evaluation and testing in the risk management process). The selection of method depends on the use of the product, the time the implant spent in the human body, and the type of contact (including external, internal contact, intact skin, mucous membranes, contact with blood, bone tissue, tissue fluids, etc.). Taking into consideration the time of contact of the biomaterial with the body, this standard classifies medical devices into those that remain in the patient's body for no longer than 24 h, over 24 h, but no longer than 30 days, and which are in constant contact, i.e. over 30 days. Among the biocompatibility tests, depending on the duration of contact with the organism and the intended use of the biomaterial, the following can be distinguished19:

  •  − in vitro cytotoxicity,

  •  − sensitizing effect,

  •  − irritating effect or intracutaneous reactivity,

  •  − acute systemic toxicity,

  •  − subacute and subchronic toxicity,

  •  − genotoxicity,

  •  − post-implantation reaction and compatibility with blood.

Different compounds interact with cells in different ways, disrupting or not their life processes. One of the basic documents for the assessment of the toxicity of medical devices is the PN-EN ISO 10993-5:2009 standard. Biological evaluation of medical devices—Part 5: In vitro cytotoxicity tests. Cytotoxic activity can be determined on the basis of changes that occur in cells under the influence of the studied material in relation to the control sample20. The control sample is a cell culture carried out without biomaterial or with material in the generally known cytotoxic concentration range. When determining the viability of cells on the basis of the number of dead cells in contact with the biomaterial, this study may disqualify the biomaterial for further use. One of the most widely used assays to test cell viability is the MTT assay19. This test does not provide information on the type of toxins that occur. It determines the biological effects known as the cytotoxicity of materials of different composition21.

In this context, the aim of the work was to study the biological and corrosion behavior of six Ca–Mg–Zn–Yb alloys with various additions of boron and gold. In vitro cytotoxicity tests do not allow one to recreate all the conditions present in a living organism, but they are an excellent way to initially determine the behavior of living cells in the contact with tested potential biomaterial. Moreover, the authors wanted to verify which of the proposed alloy compositions is more cytocompatible and corrosion resistant.

Materials and methods

Materials characterisation

The investigations were carried out on Ca32Mg12Zn38Yb18-xBx alloys (x = 0, 1, 2, 3 at.%) and Ca32Mg12Zn38Yb18-2xBxAux alloys (x = 1, 2 at.%). Base alloys were produced by induction melting of elements with a purity of 99.9%. The alloys were remelted several times to obtain a homogeneous composition of ingots. All alloys were cast under an inert atmosphere. The casting chamber was purged with argon, but this did not provide enough air reduction in the chamber. Rapidly cooled samples in the form of plates with a length and width of 10 mm and a thickness of 1 mm were cast using copper mold casting.

The topography and roughness measurements of the samples were performed using the ZEISS LSM Exciter 5 confocal microscope (Zeiss, Oberkochen, Germany) with an observation system of 4 lasers of wavelength in the range of 405 to 633 nm. The device was equipped with a ZEN image acquisition and analysis system. The surfaces of the samples were mechanically polished with SiC papers from 500 to 2400 gradation before the observations by confocal microscope. Then, they were polished with a diamond suspension and cleaned with alcohol.

Electrochemical measurements

The corrosion resistance was assessed on the basis of electrochemical tests. Measurements were carried out in PWE (NaCl—5.75 g·cm−3, KCl—0.38 g·cm−3, CaCl2·6H2O—0.394 g·cm−3, MgCl2·6H2O—0.2 g·cm−3, CH3COONa—4.62 g·cm−3, Na3C6H5O7·2H2O—0.9 g·cm−3) and Ringer’s solution (8.6 g/dm3 NaCl, 0.3 g/dm3 KCl, and 0.48 g/dm3 CaCl2–6H2O) at 37 °C using an Autolab 302 N potentiostat (Metrohm AG, Herisau, Switzerland). The conditions of the experiment were similar to the natural environment inside the organism. The potentiostat was equipped with a cell containing the reference electrode (saturated calomel electrode) and the counter electrode (platinum rod). The CaMgZnYbBAu alloys in the form of plates were tested with 3600 s of open circuit potential (EOCP). Potentiodynamic curves with Tafel’s extrapolation were recorded in a potential range from EOCP − 250 mV to EOCP + 250 mV, a scan rate was 1 mV· s−1. The corrosion potential (Ecorr) and corrosion current density (jcorr) were determined by Tafel’s extrapolation using cathodic and anodic branches of the polarization curves.

Corrosion products analysis

After 48 h of immersion in Ringer’s and PWE solutions, the surfaces of the corroded samples were observed with corrosion products using scanning electron microscope (SEM, Thornwood, New York, USA), which was equipped with an energy dispersive X-ray spectrometer (EDS) detector. Fourier transform infrared (FTIR) spectroscopy was also used to analyze the corrosion products. FTIR spectra were recorded at room temperature for the Ca32Mg12Zn38Yb15B3 and Ca32Mg12Zn38Yb14B2Au2 alloys using a Nicolet 6700/8700 FTIR spectrometer (Thermo Fisher Scientific, Waltham, USA). Before the measurements, the corrosion products were collected from the surface of the immersed samples and mixed together with dry KBr. The samples were performed in transmission mode in a mid infrared range of 4000–400 cm−1.

Cytotoxicity tests

Human osteosarcoma cells (U2-OS) derived from the ATCC+ HTB-96™ collection were used to evaluate the cytocompatibility of the experimental alloys. Cells were cultured using McCoy’s 5a Medium Modified medium, supplemented with 10% Fetal Bovine Serum (FBS), and 100 U ml−1 penicillin and 100 µg·ml−1 streptomycin. The cells were stored in an atmosphere of 37 °C saturated with water vapor and enriched with 5% CO2. The medium was changed every 2–3 days and cells were passed at 90% confluence. Biocompatibility was performed by indirect contact. Samples in the form of plates for cytotoxicity assessment were 1 mm thick. They were sterilized in an autoclave at 121 °C for 21 min at a steam pressure of 1.5 bar.

According to ISO 10993-5:1999, the extraction medium was prepared using serum-free culture medium with a surface area to extraction medium ratio of 1 cm2/ml in an atmosphere saturated with steam with water supplemented with 5% CO2 at 37 °C for 24 and 72 h. The extract was stored at 4 °C prior to the cytotoxicity test. Pure extract (100%) diluted with culture medium 1:1, 1:2, 1:4, and 1:8 were used for the tests.

Cells were incubated in 96-well culture plates at a density of 5 × 103 cells/100 µl medium in each well and incubated for 48 h to allow cell adhesion and log growth to be achieved. After this time, the medium was replaced with 100 μl of extraction medium and incubated for 24 and 48 h. After this time, the supernatant was decanted, and 100 µl of MTT reagent dissolved in medium at a final concentration of 1 mg/ml was added to each well. The samples were incubated with the MTT reagent for 4 h at 37 °C. The medium was then gently decanted and 200 µl of DMSO was added to dissolve the resulting formazan crystals. The spectrophotometric absorbance of the samples was measured with a microplate reader (BioTek ELx800) at 570 nm with a reference wavelength of 630 nm. The schematic illustration of the cytotoxicity test provided in this work is presented in Fig. 1. (1, 2) The extraction medium was prepared using a ratio of alloy culture medium with a surface area to extraction medium ratio of 1 cm2/ml for 24 h (2) and 72 h (2’). (3) After the alloy extraction period the sample was removed and the medium was centrifuged (4). (5) In O2-OS cells the indirect extract was added and the cells were incubated for 24 h or 48 h. (6) After the incubation period 24 h (6) or 48 h (6’) the supernatant was decanted and 100 µl of MTT reagent dissolved in medium was added to each well. Microplate was incubated for 4 h in a humidified atmosphere, such as + 37 °C, 5% CO2. The medium was then gently decanted and 200 µl of DMSO was added to dissolve the crystals. (7) A 96 well plate after the MTT assay. Metabolism of MTT to a formazan salt by viable cells, as shown in a 96-well plate. The decrease in the number of viable and metabolically active cells results in the decreasing intensity of the purple color observed.

Figure 1
figure 1

Schematic ilustration of cytotoxicity test according with ISO 10993-5:1999.

Results and discussion

The resorbable CaMgZnYbBAu alloys in a form of plates had an amorphous structure (Ca32Mg12Zn38Yb18, Ca32Mg12Zn38Yb16B2), an amorphous structure with some crystalline phases, such as CaZn, MgZn (Ca32Mg12Zn38Yb17B1, Ca32Mg12Zn38Yb15B3) and an amorphous structure with CaZn, MgZn and CaZn2 crystalline phases (Ca32Mg12Zn38Yb16B1Au, Ca32Mg12Zn38Yb14B2Au2). All identified phases were embedded in the amorphous structure22,23,24. The structure studies of the alloys were completed by the surface topography analysis. The surface topography of the alloys studied was observed using a confocal microscope (Fig. 2). Measurements were carried out in an area of 3.6 × 105 µm2 (600 × 600 µm). The surfaces of the Ca32Mg12Zn38Yb18-xBx alloys (x = 0, 1, 2, 3 at.%) and Ca32Mg12Zn38Yb18-2xBxAux alloys (x = 1, 2 at.%) were similar in their topography, and had a granular structure. This was in agreement with small changes in the structure of these alloys. It is well known that the surface roughness values of alloys influence the corrosion resistance and cell viability. Any scratches on the metal surface may increase stress, i.e., reduce crack resistance. This leads to stress corrosion cracking or hydrogen embrittlement, which finally causes premature failure of biodegradable implants during operation25,26. Many works, especially on Mg-based alloys, show that initial surface roughness influences the intensity of corrosion behavior27,28. Additionally, for orthopedic implants that come into contact with blood, it is important to minimize surface roughness.

Figure 2
figure 2

Confocal images of the Ca32Mg12Zn38Yb18-xBx (x = 0, 1, 2, 3 at.%) and Ca32Mg12Zn38Yb18-2xBxAux (x = 1, 2 at.%) alloys in a form of plate: (a) Ca32Mg12Zn38Yb18;; (b) Ca32Mg12Zn38Yb17B1;; (c) Ca32Mg12Zn38Yb16B2; (d) Ca32Mg12Zn38Yb15B3; (e) Ca32Mg12Zn38Yb16B1Au; (f) Ca32Mg12Zn38Yb14B2Au2.

In addition, a surface roughness analysis was also performed. The surface roughness parameters (roughness average—Ra and root mean square—RS) of the Ca-based alloys were determined and listed in Table 1. The analysis showed that the roughness values for all of the alloys studied were similar. However, the lowest roughness values were obtained for the Ca32Mg12Zn38Yb15B3 and Ca32Mg12Zn38Yb18 alloys. These may suggest that these alloys were more corrosion resistant, because the lower the roughness values, the higher the corrosion resistance29,30. It should also be noted that corrosion resistance is dependent not only on the surface roughness but also on the ability of the alloy to form a protective passive film on its surface30. Mitchell et al.27 studied the effect of surface roughness on the corrosion behavior of AZ31 Mg alloys for 672 h of immersion in a solution of 3.5 wt.% NaCl. The results have shown that the corrosion rate and the formation of the oxide layer (MgO) on the surface of the AZ31 alloy were proportional to the average roughness value (Ra) of the samples. The surface of the roughened sample had pores and microcracks that were potential places for the corrosion initiation. Moreover, in this sample, the dissolution of the metals was greater, resulting in mass reduction. In the studies performed by Walter et al.28 the passivation behavior of the AZ91 alloy with different surface roughness was presented. The authors suggested that the surface roughness of the Mg alloy played a critical role in the corrosion resistance in the 3.5% NaCl environment. They confirmed that the sample with a smoother surface has shown a greater tendency to passivation and pitting resistance than the sample with a higher surface roughness.

Table 1 Surface roughness parameters for the Ca32Mg12Zn38Yb18-xBx (x = 0, 1, 2, 3 at.%) and Ca32Mg12Zn38Yb18-2xBxAux (x = 1, 2 at.%) alloys.

Electrochemical tests in PWE solution were presented in both the open circuit potential (EOCP) and the potentiodynamic curves (Fig. 3). Changes in the EOCP potential were measured to evaluate protective properties of the alloys Ca32Mg12Zn38Yb18−xBx (x = 0, 1, 2, 3 at.%) and Ca32Mg12Zn38Yb18−2xBxAux (x = 1, 2 at.%). The curves determined for a stationary potential as a function of immersion time indicate that CaMgZnYbBAu alloys in plate form were active in PWE solution (Fig. 3a). Furthermore, the alloys studied were characterized by significant fluctuations. It can also be observed that the Ca32Mg12Zn38Yb14B2Au2 and Ca32Mg12Zn38Yb16B2 alloys were more stable in PWE solution compared to the others. The EOCP potentials were similar for all CaMgZnYbBAu alloys. They were in the range of − 1.3 V to − 1.2 V. Furthermore, the differences in EOCP potential between alloys with 1 and 2 at.% Au were not greater than 50 mV. The same situation was observed for the alloys with and without boron content. Polarization curves for alloys with 1 and 2 at.% of B were located at higher current values above 100 μA/cm2 (Fig. 3b). This may suggest that the Ca32Mg12Zn38Yb17B1 and Ca32Mg12Zn38Yb16B2 alloys exhibit lower corrosion resistance. The extrapolation of the polarization curves using the Tafel’s method allowed us to determine the corrosion current density (jcorr) and polarization resistance (Rp) of the CaMgZnYbBAu alloys (Table 2).

Figure 3
figure 3

Changes of open-circuit potential with time (a) and polarization curves (b) of CaMgZnYbBAu alloys in a form of plates in PWE solution at 37 °C.

Table 2 Comparison of electrochemical properties for Ca32Mg12Zn38Yb18-xBx (x = 0, 1, 2, 3 at.%) and Ca32Mg12Zn38Yb18-2xBxAux (x = 1, 2 at.%) alloys in Ringer’s solution and PWE fluid at 37 °C.

The lowest value of corrosion current density was observed in PWE fluid, equal to 10.6 µA·cm−2, was observed for the Ca-based alloy with 2 at.% Au. The slightly higher values of jcorr (equal to 18 and 29 µA·cm−2) were obtained for alloys with 1 at.% Au and 3 at.% B. These alloys (Ca32Mg12Zn38Yb15B3, Ca32Mg12Zn38Yb16B1Au1, Ca32Mg12Zn38Yb14B2Au2) were characterized by higher polarization resistance that indicate better corrosion resistance compared to alloys B0, B1 and B2. The low jcorr and high Rp of the Ca32Mg12Zn38Yb15B3 alloy is consistent with the small roughness values.

For comparison, the changes in EOCP with time and polarization curves for all alloys in Ringer’s solution were presented in Fig. 4. The lowest values of corrosion current density in Ringer’s solution were observed for alloys with 1 and 2 at.% addition of Au22,23,24.

Figure 4
figure 4

Changes of open-circuit potential with time (a) and polarization curves (b) of CaMgZnYbBAu alloys in a form of plates in Ringer’s solution at 37 °C22,23,24.

The surface roughness of implants is assumed to be an important factor in the osseointegration of these materials31. This parameter is responsible for the degradation of metallic materials32,33. Studies on the impact of surface roughness on the corrosive behavior of magnesium alloy (AZ91) in a chloride-containing environment at room temperature showed that the pitting corrosion of the alloy critical depends on the roughness of the surface28.

The smoother surface of Ca alloy with 3 at.% of B improved the corrosion resistance in the chloride environment. Therefore, it is very important to design alloys with optimal surface roughness to improve their corrosion properties34,35. Reddy et al.34 studied the effect of surface roughness on the corrosion behavior in 3.5 wt.% NaCl of two AZ31 and AZ80 alloys. The researchers stated that the higher corrosion resistance of AZ80 compared to AZ31 was due to lower surface roughness. Hence, the lower surface roughness of the samples improves surface integrity in such a way as to increase corrosion resistance36. Similar results were obtained in terms of the correlation between surface roughness and corrosion resistance of biomedical materials for stainless steel and titanium alloys37,38,39.

The corrosion resistance of orthopedic implants depends on the alloy additives. Calcium as a base metal increases strength and hardness. The influence of Ca on the microstructure of the alloy and electrochemical behavior can also be seen4. This element is important in the context of the kinetics of implant degradation because it increases their resorption. Magnesium in Ca-based alloys has mechanical properties similar to those of human bones, but it is characterized by a high corrosion rate4. This disadvantage of Mg can be reduced by the addition of Zn, Yb, B, and Au metals to Ca-based alloys. The addition of Zn causes grain refinement and the formation of secondary phases, which affects the mechanical and corrosive properties of Mg alloys, because zinc improves the corrosion potential of the biomedical material40. Moreover, Zn transforms impurities such as iron, copper, and nickel into harmless intermetallic compounds, thus reducing their corrosion-promoting effects7. The physical and chemical properties of Yb as a rare-earth element are similar to those of calcium. The authors40 studied the corrosion behavior in simulated body fluid (SBF) of amorphous Mg–Zn–Yb–Ag alloys with different contents of ytterbium. They stated that the addition of 4 at.% of Yb significantly improved the corrosion resistance compared to the other alloys tested. In the work24 there was observed that ytterbium has a positive effect on the corrosion resistance of Ca–Mg–Zn alloys in chloride-rich Ringer solution. In addition, the authors in24 suggested that the addition of boron and gold to a Ca-based alloys slowed the corrosion rate (due to creating a barrier among Ca, B, and Cl ions and forming a thick layer of corrosion products) and decreased the evolution of H2.

Furthermore, Hernandez-Rodriguez et al.41 tested the effect of different boron content (0.06, 0.25, 0.5 and 1 wt.%) in a CoCrMo biocompatible alloy on corrosion resistance in phosphate buffered saline (PBS). They observed an improvement in corrosion resistance with increased B content. Similar electrochemical corrosion results were achieved during immersion in Ringer’s solution22,23,24. The good anticorrosive properties of such alloys were also resulted from the addition of gold as an alloying element42 or as nanoparticles located in an applied coating43,44. Lee et al.42 studied the corrosion resistance of Ti-xAu alloys (x = 5, 10, 15, 20 and 40 wt.%) in 0.9% NaCl solution at 37 °C. They stated that the Ti-20Au alloy exhibited the lowest Ecorr value, equal to − 278.33 mV, but the lowest corrosion current density was observed for the Ti-5Au alloy. Thin gold films (with a thickness of about 50 and 100 nm) which have been applied to Ni–Cr–Mo alloys using magnetron sputtering technique were tested by Wadullah et al.44. The corrosion behavior of the samples was studied in an artificial saliva solution at 37 °C. The authors observed lower values for corrosion current densities and higher values for corrosion potentials for alloys coated with gold.

The biocompatibility of the CaMgZnYbBAu alloys was studied using the MTT assay, which is an indicator of cell viability, proliferation, and cytotoxicity (see Fig. 5). It is known that osteoblastic cell adhesion, their growth, and proliferation were correlated with surface roughness. Moreover, the interactions between osteoblasts and orthopedic biomaterials depend on the development of bone-implant interfaces. Because of that, osteoblastic adhesion is very important for the first bone-biomaterial interaction45,46.

Figure 5
figure 5

Cytotoxicity of U2-OS cells cultured in extraction mediums after 24 and 48 h of incubation for alloys: (a) Ca32Mg12Zn38Yb18; (b) Ca32Mg12Zn38Yb17B1; (c) Ca32Mg12Zn38Yb16B2; (d) Ca32Mg12Zn38Yb15B3; (e) Ca32Mg12Zn38Yb16B1Au; (f) Ca32Mg12Zn38Yb14B2Au2. Data are the mean ± SD of three independent experiments. *p < 0.05 and **p < 0.01 versus control.

The colorimetric MTT assay is characterized by high sensitivity and therefore it can be used to study a large number of different samples21. The amount of MTT assay is proportional to the number of metabolically active cell cultures. It should be mentioned that cells have limited ability to proliferate. They can divide about 50 times with a culture medium that is frequently changed. Samples with absorption values less than 50% of division activity are considered toxic21. In our studies, the biocompatible effect of the samples was studied by indirect contact19. The extracts of the samples (in a form of plates) after 24 and 72 h were added to the cells. The time for cell incubation was 24 and 48 h. In addition, the content of the extracts was 100, 50, 25 and 12.5%, adequately. The results of the cytotoxicity tests showed that the Ca32Mg12Zn38Yb15B3 alloy was characterized by the highest cell viability (Fig. 5d). Cell viability greater than 100% was visible both for 24 and 48 h of incubation and for 24 and 74 h of extraction. Slightly lower cell viability was observed for the Ca-based alloy with 2 at.% Au (Fig. 5f). The cytotoxicity results were in agreement with the corrosion resistance of the Ca32Mg12Zn38Yb15B3 and Ca32Mg12Zn38Yb14B2Au2 alloys. It can also be stated that the alloy without boron is toxic and cannot be used as a potential biomaterial. Hakki et al.10 studied boron effect of the addition on the cell proliferation. The impact of B concentrations on cell viability was evaluated at 24 and 48 h using an MTT assay. The authors stated that boron plays an important role in bone metabolism11. The effect of Au was commented on in47 in which its influence on Fe-based alloy properties was investigated. The authors47 assessed that the addition of 5 wt.% Au in Fe-based alloys is an alternative to bioresorbable materials due to mechanical properties and biocompatibility understood, among others, as a cell vialibity and with the appropriate degradation rate.

Approximately a hundred data are related to the biocompatibility of bioresorbable Mg-based alloys48,49,50. Article50 presents data on mitochondrial metabolic activity of cells exposed, among others, to pure Mg alloy for 24 h. Analysis using MTT and LIVE/DEAD assays indicated metabolic acticity over 80% that was summerized as high cell viability50. In another work51, cytotoxicity studies showed that Mg–1.2Ca–1Zn alloy extraction medium resulted in higher cell viability than composition with Bi addition. The authors51 referred to the study52, in which the in vitro cytotoxicity of Y, Nd, Dy, Pr, Gd, La, Ce, Eu, Li, and Zr was analyzed. According to work52, the cytotoxicity of Ce and La is strongly associated with ionic radii, therefore, the biocompatibility of alloys with the addition of Bi was lower compared to the Mg-1.2Ca-1Zn alloy51. In this work, the cytotoxicity of Ca-based alloys could also be related with the differences between the ionic radii of B and Au. In the work53 there are also results of cytotoxicity studies on bioresorbable Zn-Mg alloys. According to the plot of cell viability measured through the MTT test presented in53, the viability of cells after 24 h in the Zn–1Mg alloy was slightly below 100%. After 72 h, the value increased above 100%53. Moreover, in paper49 the Ca-P coatings on the Mg-Zn-Zr alloy were investigated. According to cytotoxicity studies, after three days of incubation, the Ca-P coating had a favorable effect on biocompatibility compared to the uncoated alloy49.

After 48 h of immersion in Ringer’s solution and PWE fluid at 37 °C, the surface morphologies of the samples with corrosion products were microscopically observed. The analysis aimed to examine alloys with a lower corrosion rate and a higher cell viability, such as Ca32Mg12Zn38Yb15B3 and Ca32Mg12Zn38Yb14B2Au2. The SEM images are presented in Fig. 6. Microcracks were observed in all alloy samples, due to dehydration during the drying of the samples. In addition, the surfaces of the alloys were covered with corrosion products, and some corrosion products were also observed to fall off of the surface. It can also be seen that the entire surface of the samples with 3 at.% of boron and with 2 at.% of Au after immersion in PWE was not covered with corrosion products (Fig. 6e,g). Furthermore, the shapes of the corrosion products were different depending on the type of corrosive solution (Ringer’s or PWE solution). The samples exposed to Ringer’s solution were covered with cuboid-shaped and needle-shaped corrosion products compared to the samples exposed to PWE fluid, where flaky-shaped corrosion products can be seen.

Figure 6
figure 6

SEM images of samples’ surfaces with corrosion products after 48 h of immersion in Ringer’s solution at 37 °C for: (a,b) Ca32Mg12Zn38Yb15B3; (c,d) Ca32Mg12Zn38Yb14B2Au2, and after immersion in PWE fluid: (e,f) Ca32Mg12Zn38Yb15B3; (g,h) Ca32Mg12Zn38Yb14B2Au2.

In addition, EDS analysis and FTIR tests were performed for the corrosion products of the alloys and solutions mentioned above (Figs. 7 and 8). The results of the EDS analysis showed that Zn, Yb, and O are the dominant constituent elements of the particles. There were also low-intensity peaks from Mg, Ca, C, and Cl. The lower intensity of chloride ions in corrosion products for both Ca-based alloys may suggest that these alloys were corrosion resistant.

Figure 7
figure 7

EDS analysis for corrosion products after 48 h of immersion in Ringer’s solution at 37 °C for: (a) Ca32Mg12Zn38Yb15B3; (b) Ca32Mg12Zn38Yb14B2Au2, and after immersion in PWE fluid for: (c) Ca32Mg12Zn38Yb15B3; (d) Ca32Mg12Zn38Yb14B2Au2.

Figure 8
figure 8

FTIR spectra of Ca32Mg12Zn38Yb15B3 and Ca32Mg12Zn38Yb14B2Au2 samples after immersion test in Ringer’s and PWE solutions with marked characteristic vibrations.

The obtained FTIR spectra confirmed the presence of carbonates on the surface of the studied plates after 7 days of the immersion test in both PWE and Ringer’s solutions. The peak observed at 1645 cm−1 for the B3 sample (Ca32Mg12Zn38Yb15B3) kept in Ringer’s solution is associated with the δ O–H vibration of water54, while the highly visible broad peak around 854 cm−1 (from the Zn–O vibration) confirmed, that the corrosion products presented on the surface are probably related to the ZnCO355,56. The characteristic of carbonates υ1 symmetric, υ2 asymmetric and υ4 symmetric vibrations were observed at 1082, 874, 854 and 713 cm−154. Interestingly, the υ3 asymmetric vibrations of CO32− appeared at different wave numbers for analyzed samples. For example, sample B3 after immersion in the Ringer’s solution was characterized by very broad peaks at 1575, 1485 and 1405 cm−1, while the Ca32Mg12Zn38Yb14B2Au2 sample immersed in the same solution showed the υ3 asymmetric vibrations observed at 1490 and 1402 cm−1. This phenomenon is related to differences in the complex crystal structure of the resulting corrosion products. The same findings were described years ago by Lee and Condrate57. The carbonate ion lies in a lower symmetry than that of unperturbed carbonate. Although only two bands can be observed for free carbonate ion, for the carbonates (such as zincum carbonate), large splitting can be observed. Accordingly, the structure and chemical composition of the formed carbonates are different for all samples and depend on both the chemical composition of the alloys and the solution.

Conclusions

The purpose of the work was to study the biocompatible effect and corrosion behavior of the Ca32Mg12Zn38Yb18−xBx (x = 0, 1, 2, 3 at.%) and Ca32Mg12Zn38Yb18−2xBxAux (x = 1, 2 at.%) alloys. The results of the investigations allowed us to draw the following conclusions:

  • The Au addition of 1 and 2 at.% in CaMgZnYbB alloys and 3 at.% of B in CaMgZnYb improved corrosion resistance in PWE solution compared to Ca32Mg12Zn38Yb18 because of lower values of jcorr and higher Rp. The B addition of 1 and 2% negatively affected the corrosion properties of the Ca32Mg12Zn38Yb18 alloy in PWE solution.

  • The surface roughness analysis has shown that the lowest roughness values (Ra and Rs equal to 1.09 and 1.39 µm, respectively) were obtained for the alloy with 3 at.% of B. According to the results of the electrochemical measurements, the B addition of 3 at.% was more corrosion resistant compared to the addition of 2 and 1 at.% of B. The results of cytotoxicity tests using the MTT assay also showed that the alloy with 3 at.% of B was characterized by the highest cell viability, over 100% for both 24 and 48 h of incubation.

  • The FTIR results indicated the presence mainly of carbonates with different structure and chemical composition (due to different alloy and solution compositions), on the surfaces of both the studied Ca32Mg12Zn38Yb15B3 and Ca32Mg12Zn38Yb14B2Au2 samples studied immersed in Ringer’s solution and PWE fluid.