Abstract
Laparoscopic TAPP/TEP approaches are well-established options for the cure of inguinal hernias. As in the open approach, mesh fixation and poor-quality biologic response represent controversial questions and are a source of concerns. Furthermore, hernia defect patency represents another problem which seems not well acknowledged among surgeons. These problems are considered the cause of frequent intra and postoperative complications. To overcome these concerns, recently a different concept of cure has emerged. Based on a newly developed dynamic responsive 3D scaffold named ProFlor, a permanent hernia defect obliteration has been finalized. Following its inherent centrifugal expansion due to its dynamic responsivity, this hernia device is positioned fixation free within the defect and induces a probiotic biological response allowing for the re-establishment of the degenerated inguinal barrier. A laparoscopic approach with the 3D scaffold has been tested on 71 patients to demonstrate its effectiveness in reducing intra and postoperative complications. The operated patients presented with bilateral and/or recurrent inguinal hernia. Overall, 122 hernia defects were obliterated with 119 dynamic responsive scaffolds. The procedures were carried out from January 2018 to January 2022 with a defined protocol and detailed procedural steps. The laparoscopic technique with the 3D hernia scaffold allowed for fixation free placement, permanent defect obliteration and dynamically induced regenerative effects. The technique proved effective in reducing intra and postoperative complications. In particular, early postoperative pain and discomfort significantly decreased. No chronic pain and no recurrences were reported during follow up. The results achieved with the described laparoscopic technique seem to embody an innovative concept for inguinal hernia repair. Fixation free, dynamic responsive, permanent defect obliteration, histologically proven regenerative effects are the distinctive features of this 3D scaffold. It seems to embody a more physiological and pathogenetically coherent concept of cure, thus improving treatment results of this widespread disease.
Similar content being viewed by others
Introduction
Many techniques and several types of implants are used for the repair of inguinal hernias. The concept of reinforcing the herniated groin by deploying a flat mesh, usually made of polypropylene, has demonstrated effectiveness in reducing the high recurrence rate of the pre-prosthetic era1. Nowadays, all implants used for hernia repair are flat, static, passive and do not cope with movements of the abdominal wall. Open anterior or posterior repair, laparoscopic transabdominal preperitoneal (TAPP) or totally extra-peritoneal (TEP) are two of the most popular herniorrhaphy methods, nevertheless the surgical community is still searching for a gold standard in the treatment of this disease2,3. In the past two decades, laparoscopic hernia repair has been more frequently reported as the method of choice, especially in the case of recurrence after open anterior repair or bilateral inguinal hernias4. However, laparoscopic TAPP and TEP present several challenges related to anatomical conditions, surgical technique and prosthetics5. For large defects, to avoid mesh invagination through the hernia opening under the push of the abdominal viscera, larger and heavier meshes with multiple fixation tools are used. This implies wider peritoneal dissection and increased risk of intra- and postoperative complications5,6. A consequence of broader tissue dissection and use of wider/harder meshes is an increase of fibrotic bridles and implant induced scar plaque in the delicate inguinal area that potentially increases postoperative pain and is considered harbinger of chronic pain syndrome7,8,9,10,11,12,13,14,15. Furthermore, the patency of the hernia defect after flat mesh deployment embodies another challenge whose relevance has not been adequately considered nor addressed. Leaving the inguinal hernia defect permanently open after mesh deployment is considered the most common cause of recurrence after laparoscopic repair16. To resolve these problems, a novel concept has recently been developed. It concerns the use of a 3D dynamic responsive scaffold that is introduced into the hernia opening achieving a permanent defect obliteration17,18,19. This 3D hernia device is made of polypropylene material like conventional hernia prosthetics but owns inherent springiness, is positioned fixation free and, moving together with the groin, induces a probiotic biologic response. Modulated by specific growth factors, the development of newly formed connective tissue, vessels, nerves and muscles within the 3D scaffold has been documented in a series of scientific studies19,20,21,22,23,24. These proprietary regenerative effects, likely result of its intrinsic dynamic responsivity, allow the categorization of this new inguinal hernia device not as a conventional implant but rather a regenerative scaffold. Already used in inguinal hernia repair with open approach, recently a specific laparoscopic TAPP approach has been finalized aimed to further reduce surgical trauma achieving permanent inguinal defect obliteration25,26,27. The present report highlights the outcomes of this novel laparoscopic technique with ProFlor in a cohort of patients suffering from inguinal hernia.
Materials and methods
Compliance with ethical standards
The research has been approved by the Ethics Commission of Policlinic Hospital of the University of Palermo, Italy—Approval ID: 10—November 11, 2020.
Human and animal right
The investigation was carried out in accordance with the Declaration of Helsinki for experiments involving humans. Informed consent was obtained for experimentation with human subjects.
Registration
ClinicalTrials.gov—Registration ID: NCT04718298—Registration date: 22/01/2021. https://clinicaltrials.gov/ct2/show/NCT04718298?term=Proflor&draw=2&rank=2.
Participants and tools
From January 2018 to January 2022, 71 individuals underwent laparoscopic hernia defect obliteration using the 3D dynamic responsive scaffold ProFlor type E. This 3D hernia device is made from low weight, large porous polypropylene, has a multilamellar cylindrical 3D structure with reinforced edges, and is available in two sizes: 25 and 40 mm in diameter, both types 15 mm in thickness. The center of the scaffold core is connected on one surface to an oval shaped flat mesh measuring 8 × 10 cm, which is intended to be deployed to cover the inguinal and femoral area counterfacing the peritoneal sheath (Fig. 1). The 3D core of the scaffold is compressible on both longitudinal and transversal planes. Endowed with inherent dynamic responsivity, it contracts and relaxes with the movements of the groin. Its proprietary centrifugal expansion allows fixation free positioning in the hernia defect for permanent obliteration.
Inclusion criteria for patients of the cohort undergoing inguinal hernia repair were:
-
(a)
Unilateral recurrent inguinal hernia previously operated with anterior open approach.
-
(b)
Bilateral primary inguinal hernia or recurrent unilateral + primary hernia on the contralateral groin.
-
(c)
Hernia defect diameter < 37 mm preoperatively assessed by US/CT scan.
-
(d)
Age between 18 and 85 years.
-
(e)
ASA score < IV.
-
(f)
BMI < 34.
Details of inclusion criteria and demographic data of the patients are included in Table 1. All patients of the cohort were male. Mean age was 54 (range 33–78) and mean BMI 27.8 (range 25–31).
Informed consent was obtained from all the participants involved in the experiment.
Preoperative clinical details
Preoperative defect measurement carried out with US and/or CT scans was considered mandatory for all patients (Fig. 2A,B). The aim of assessment of the hernia defect is to preoperatively determine the size of the implant core to be chosen for the procedure, e.g. a hernia opening > 37 mm was not contemplated and considered an exclusion criterion. Twenty-seven patients presented with bilateral inguinal hernia and among these 12 had double ipsilateral hernias. Eighteen patients suffered from recurrent inguinal hernia after anterior open approach. Three inguinal hernia patients, one of whom was operated for recurrence, had concomitant femoral hernia.
Surgical procedure
All surgeries were carried out in general anesthesia with TAPP technique. A protocol-based antibiotic prophylaxis with a single dose of a cephalosporin was administered one hour before operation. After establishing pneumoperitoneum, three trocars were placed at the umbilical level (12 mm large optic in the navel and two working ports over the right and left rectus muscle); all working ports were inserted under direct vision. Once the hernial defect was identified and the visceral protrusion pulled back to the peritoneal cavity, a dissection of parietal peritoneum to expose the inguinal backwall was made. Adhesiolysis of hernia sac followed. The dissected hernia sac was then returned to the abdominal cavity. After reducing the hernia, the ProFlor implant, rolled upon the flat part and compressed along the longitudinal axis, was introduced through the 12 mm trocar (Figs. 3A–D, 4A). Once delivered into the abdominal cavity, the springiness of the 3D scaffold allows a prompt regain of the original shape (Fig. 4B). Then, with simple maneuvers, the core of ProFlor E was inserted and positioned in the hernial defect with the edges facing the borders to fully obliterate the hernia opening (Figs. 4C, 5). Once the hernia device was firmly delivered into the defect, a simple test carried out by a repeated tentative of pulling out the flat part of the scaffold confirmed the permanence of the 3D core of the scaffold within the defect. Thanks to the centrifugal expansion of the multilamellar 3D core, no sutures or other fixation methods were needed to hold the ProFlor E in place, in any of the patients. At this stage, the flat mesh of the implant could be deployed to cover the inguinal and femoral area (Figs. 5, 6). Closure of the peritoneal flap finalized the procedure.
Follow up protocol
Patients were discharged within 24 h of surgery with a prescription for an analgesic prophylaxis based on oral administration of 1000 mg paracetamol 3 times per day until the third postoperative day. Follow up was carried out starting after 1 week, 4 weeks, 3 months, 6 months and each following year postoperatively. The protocol included the use of the visual analogue scale (VAS) to assess pain, and US control of the operated groin to demonstrate the permanence of the 3D scaffold in the defect. To highlight the quality of tissue incorporation within the 3D hernia device through the assessment of signal intensity, an MRI of the lower pelvis was programmed between 6 and 12 months, post-operation.
Results
The described 3D scaffold was employed to obliterate inguinal hernia defects fixation free in all 71 patients of the cohort. Despite no preoperative selection being made, from the preliminary US/CT imaging all patients met the inclusion criteria showing defect dimensions < 37 mm. This allowed the use of both 3D scaffold types with the 25- and 40-mm sized 3D core. Concerning hernia types, 33 patients of the cohort had unilateral recurrent hernias, two of whom with concomitant femoral hernia (forgotten hernia). 34 individuals presented with bilateral primary hernias, 6 others with double ipsilateral hernia in one groin and two additional patients with a double ipsilateral hernia in both groins (Figs. 6 A,B, 7). Three patients with bilateral hernias also had a concomitant femoral hernia. A total of 122 hernia defects were obliterated with 119 scaffolds since three double ipsilateral defects could be obliterated with just one 40 mm 3D scaffold that simply displaced laterally the divisor septum between the defects. With regard to location, 79 defects were in the lateral fossa, 24 in the medial fossa, 8 in the supravesical fossa, 6 were combined involving two neighboring fossae and an additional five in the femoral ring. Concerning the types of 3D scaffold used, 53 units were 25 mm in size and 66 units sized 40 mm in diameter. All details of the intraoperative findings, as well as the scaffolds utilized, are highlighted in Table 2. The mean duration of the operative procedure in unilateral (all recurrent) hernias was around 55 min (range 45–65 min.), in bilateral hernia, with single defect each side, the procedure lasted approximately 65 min (range 55–75 min) while in the cases with additional ipsilateral hernia mean time length was 75 min (range 65–85 min.). The time needed for obliterating the three reported femoral hernias was approximately ten additional minutes for each scaffold deployment. Overall, no intraoperative adverse events occurred. All patients were discharged the day after the operation. In the postoperative period, only 3 seromas (4.2%) were reported, and these were successfully managed conservatively. No hematomas, sepsis or other common adverse events occurred. During long-term follow-up, US scans showed no dislocation or migration of the 3D scaffold, which remained firmly in place fully obliterating the defect. The return to daily activities was nearly immediate, starting already from 43th to 5th postoperative day. No sense of foreign body, discomfort, or chronic pain was reported among the operated patients, even long term. In a follow up, from 36 to 6 months, no recurrence occurred in any patients of the cohort. All data of the postoperative outcomes are detailed in Table 3. Postoperative pain, assessed with the VAS scale, showed very low-pain scores starting from the first post-operative day. Regardless of unilateral recurrent or bilateral inguinal hernia repair with the 3D dynamic responsive scaffold, four days postop. all patients were pain-free (Fig. 8). Between 6 and 12 months postop. all individuals underwent MRI control of the lower pelvis to document the permanence in situ of the 3D scaffold obliterating the defect, as well as the quality of the newly developed tissue into the 3D scaffold by means of comparative signal intensity (Fig. 9A,B).
Discussion
Laparoscopic inguinal hernia repair is gaining increasing popularity among surgeons. However, this surgical procedure suffers from the same limitations as the open approach. As in conventional open prosthetic herniorrhaphy, uncontrolled foreign body response and mesh fixation are the main concerns28,29,30,31,32,33,34. The laparoscopic use of very broad meshes, often made of heavy-weight texture, logically intensifies the foreign body response induced by the biocompatible material causing an uncontrolled fibrotic apposition upon and around the flat mesh35,36. Postoperatively, the effects of excessive fibrotic ingrowth can lead to discomfort during movements and/or increased risks of groin nerve entrapment resulting in chronic pain. Concerning fixation, apart from expensive, short acting fibrin glue or the seemingly toxic cyanoacrylate adhesives37,38, in laparoscopic herniorrhaphy (both TEP and TAPP) mesh fastening is commonly carried out with metallic or resorbable tacks. This implies an evident negative impact on the abdominal wall physiology, as the screwed tacks, sharply piercing the delicate muscular surround, logically hinder groin kinetics39,40. Postoperatively, the pressure of the abdominal content puts the mesh under stress. If the hernia opening is larger than 2 cm, visceral pressure pushes the implant towards this weak point with possible detachment of the fixation tools. This may lead to mobilization and invagination of the mesh into the patent defect and is considered the main cause of recurrence in laparoscopic inguinal hernia repair6. If detached, tacks may tear the musculature causing bleeding and increased postoperative pain. Unfastened tacks falling in the preperitoneal space can migrate into the abdominal cavity leading sometimes to serious adverse events such as intestinal perforation41,42. Nevertheless, by critically analyzing the literature, it appears evident that, despite its significance, the issue of defect patency does not seem to be taken into adequate consideration. Years of scientific evidence have confirmed that the larger the defect the more recurrences occur, explaining why patency of the hernial opening continues to represent an unresolved and disappointingly undebated dilemma. To resolve the problem, some surgeons envisage that choosing larger and harder meshes plus increasing tack fixation could be the solution. However, this empirical solution does not appear to have met the needs. Therefore, it would seem evident that hernia defect patency is a crucial issue still waiting for a solution, as in clinical and experimental literature there are no reports that deal with the basic question: how can permanent obliteration of the hernia defect in laparoscopic groin hernia repair be achieved? In recent years, a different line of thought concerning the treatment of inguinal hernia has emerged based on updated evidence in physiopathology and the degenerative genesis of inguinal protrusions43,44,45,46,47,48,49,50. Following these new findings, a newly designed 3D device with dynamic responsivity has been developed for inguinal hernia repair. Made of the same polypropylene material as conventional meshes and owing an intrinsic centrifugal expansion, this inguinal hernia device, named ProFlor, is intended to be inserted fixation free into the defect thus achieving permanent obliteration. Recent literature has demonstrated that the dynamic compliance of this 3D hernia scaffold allows for a totally different biological response achieving development of newly formed connective tissue, vessels, nerves and muscle bundles, to full maturation20,21,22,23. The pathogenetically and physiologically coherent therapy model embodied by this 3D hernia device is diametrically opposed to the concept of reinforcing the groin with fibrotic mesh incorporation, which has been widespread for more than half a century. Conversely, the regenerative features of ProFlor mean that this 3D hernia device cannot be categorized as a conventional hernia prosthetics but rather as a regenerative scaffold. The above-depicted characteristics, undoubtedly uncommon for a hernia device, appear to demonstrate its effectiveness in the results achieved during the clinical study here presented. It should also be highlighted that on observing the size of the 3D scaffold, especially the 40 mm sized type used for laparoscopic hernia repair, it may be hard to believe that it can be introduced through a trocar channel into the abdominal cavity. Nevertheless, both clinical practice and specific physical characteristics of this hernia device make it possible to accomplish this crucial procedural step. As the 3D dynamic responsive scaffold used for laparoscopy is easily compressible, it can be enfolded by rolling the flat portion (Fig. 3A). With this configuration, it is possible to deliver both the 25 mm and the 40 mm types through a 12 mm channel by means of forceps (Fig. 3B–D). The laparoscopic procedure with this 3D hernia device is shorter and safer as, unlike the conventional laparoscopic approach, there is no need for a wide dissection of the inguinal backwall to deploy the meshes, usually 15 × 10 cm, to reduce the risks of invagination and/or recurrence51. On the contrary, achieving permanent defect obliteration renders wide dissection and deployment of large meshes superfluous, thus eliminating risks of injuries and reducing time needed. The 8 × 10 cm large, oval shaped, preperitoneal connected flat part of the 3D scaffold amply suffices for complete prophylactic coverage of the posterior inguinal area (Fig. 6), which surface ranges between 4.7 and 5.5 cm in length and 3–6.5 cm in height52. In addition, the low weight and very large porous texture of ProFlor’s flat portion, are far away from the mass and thickness of the meshes usually used for laparoscopic hernia repair also, like in double ipsilateral hernia may occur, in case of placement of two 3D devices in same groin.
At first glance, ProFlor’s procedural concept might resemble the plug-and-patch repair technique. Nevertheless, there are significant dissimilarities between these two models of treatment. Firstly, conventional plugs are static, not dynamic and, to avoid migration, must be fixated to the hernia opening. Further, plugs are inserted into the hernia opening mimicking defect obliteration, which, however results ineffective since within a few weeks the plug becomes incorporated by fibrotic fibers and shrinks up to 70%. These adverse phenomena do not occur in ProFlor that, positioned fixation free to permanently obliterate the defect, moves in tune with the inguinal backwall. Lastly, ProFlor does not shrink because its intrinsic dynamic responsivity induces a unique biologic response with development of newly formed vessels, nerves and muscle fibers characterizing this 3D device as a regenerative scaffold20,21,22,23,24. These proprietary features of ProFlor embody a substantial difference to patch and plug technique.
Analyses of the intra and postop. Outcomes show that no adverse events of note occurred in the described patient cohort. Reduced intraoperative trauma allowed all patients to be discharged within 24 h with no or negligible pain. Discomfort or pain intensity was sufficiently low for all individuals to return to the daily activity within 5 days. Except for three seromas, no major or minor complications, no chronic pain or recurrence postoperatively were reported. These positive results may be the consequence of a series of factors. Among these, the inherent dynamic behavior of the 3D scaffold is crucial, as it allows centrifugal expansion into the defect assuring permanent obliteration of the hernial opening without need for fixation. No fixation and dynamic compliance to groin movements are essential for a significant reduction of postoperative pain and avoid risk of developing chronic pain. US and MR imaging carried out during follow up demonstrated the permanence in situ of all dynamic scaffolds also in the long term. By fully and permanently occupying the hernia defect, the 3D hernia device impedes visceral re-protrusion through the inguinal barrier; hence no reported recurrences. Moreover, within a few months, the regenerative feature of the 3D scaffold leads to complete re-establishment of the inguinal barrier formerly wasted by hernia disease. The dynamically induced probiotic response of ProFlor in humans is clearly documented by the assessment of the signal intensity in MR imaging53. This, as seen in Fig. 9A,B, allows characterization of the enhanced quality of ingrown tissue in the 3D scaffold.
There are, however, some limitations which should be addressed in future studies. Firstly, even if statistically significant, the patient cohort is limited. This was mainly due to the negative impact of the recent COVID pandemic on elective surgery schedules. Furthermore, the study design also fully considered the recommendation of the EHS guidelines that strongly suggest laparoscopic inguinal hernia repair for recurrent hernias after open approach and for the treatment of bilateral inguinal hernias54,55. Therefore, patients with unilateral inguinal hernia were excluded from the study: these patients underwent open anterior hernia repair with the 3D scaffold in local anesthesia in the outpatient care system. Also, defects wider than 37 mm were excluded from the investigation. Whilst no preoperative selection was carried out, none of the preliminary US or CT assessed dimensions of the hernia defects was wider than 35 mm in the described cohort of patients. It must also be stated that more than 90% of hernia defects overall encountered in our clinical practice are less than 35 mm. For future cases of defects greater than 37 mm the planned solution involves delivering two 3D hernia devices connected together as previously documented in open ProFlor technique through a 12 mm channel24. Lastly, we are well aware that to be fully validated these results should be compared with those obtained with other laparoscopic techniques for inguinal hernia repair. However, the present report merely represents a preliminary study, which outcomes serve as a basic dataset for a comparative study with the TAPP technique carried out with conventional flat meshes, which is still ongoing.
Conclusion
The described laparoscopic technique and the clinical results achieved with the 3D dynamic responsive and regenerative hernia scaffold seems to embody an innovative concept for the cure of inguinal hernias. Dynamic responsivity, fixation free deployment, permanent defect obliteration and regenerative behavior, are totally new features developed to overcome unphysiological and pathogenetically discrepant concepts of a therapy that was established more than half a century ago. In accordance with current international guidelines56,57, TAPP hernia repair technique with the 3D hernia scaffold ProFlor likely represents a concrete and promising alternative for decisively improving the laparoscopic treatment results of this widespread disease.
Data availability
The datasets used and analyzed during the study are available from the corresponding author upon reasonable request.
References
Kingsnorth, A. & LeBlanc, K. Hernias: Inguinal and incisional. Lancet 362, 1561–1571 (2003).
Bracale, U. et al. Which is the best laparoscopic approach for inguinal hernia repair: TEP or TAPP? A systematic review of the literature with a network meta-analysis. Surg. Endoscopy 26(12), 3355–3366 (2012).
McCormack, K. et al. Laparoscopic techniques versus open techniques for inguinal hernia repair. Cochr. Database Syst. Rev. https://doi.org/10.1002/14651858 (2003).
Köckerling, F. et al. TEP versus TAPP: Comparison of the perioperative outcome in 17,587 patients with a primary unilateral inguinal hernia. Surg. Endosc. 29(12), 3750–3760 (2015).
de Beaux, A. C. Abdominal hernias in core topics in general and emergency surgery: E-Book: A companion to specialist surgical practice. Edisi Ke-6th. London: Elsevier Health Sciences pp. 81–93 (2018).
Schmedt, C. G., Sauerland, S. & Bittner, R. Comparison of endoscopic procedure vs Lichtenstein and other open mesh techniques for inguinal hernia repair: A meta-analysis of randomized controlled trials. Surg. Endosc. 19, 188–189 (2005).
Amid, P. K. Causes, prevention, and surgical treatment of postherniorrhaphy neuropathic inguinodynia: Triple neurectomy with proximal end implantation. Hernia 8, 343–349 (2004).
Bay-Nielsen, M., Perkins, F. M. & Kehlet, H. Pain and functional impairment 1 year after inguinal herniorrhaphy: A nationwide questionnaire study. Am. J. Surg. 233, 1–7 (2001).
Bay-Nielsen, M., Nilsson, E., Nordin, P. & Kehlet, H. Chronic pain after open mesh and sutured repair of indirect inguinal hernia in young males. Br. J. Surg. 91, 1372–1376 (2004).
Kehlet, H. & Bay-Nielsen, M. Nationwide quality improvement of groin hernia repair from the Danish Hernia Database of 87,840 patients from 1998 to 2005. Hernia 12, 1–7 (2008).
Nienhuijs, S. et al. Chronic pain after mesh repair of inguinal hernia: A systematic review. Am. J. Surg. 194, 394–400 (2007).
Loos, M. J., Roumen, R. M. & Scheltinga, M. R. Classifying postherniorrhaphy pain syndromes following elective inguinal hernia repair. World J. Surg. 31, 1760–1765 (2007).
Aasvang, E. K. et al. Predictive risk factors for persistent postherniotomy pain. Anesthesiology 112(4), 957–969 (2010).
Burgmans, J. P. et al. 2010) Long-term results of a randomized double-blinded prospective trial of a lightweight (Ultrapro) versus a heavyweight mesh (Prolene) in laparoscopic total extraperitoneal inguinal hernia repair (TULP-trial. Ann. Surg. 263(5), 862–866 (2016).
Mitura, K., Garnysz, K. & Michałek, I. Long-term follow-up of a randomized controlled trial of Lichtenstein repair vs the Valenti technique for inguinal hernia. Hernia 23, 547–554 (2019).
Andresen, K. & Rosenberg, J. Management of chronic pain after hernia repair. J. Pain Res. 11, 675–681 (2018).
Mayer, F. et al. When is mesh fixation in TAPP-repair of primary inguinal hernia repair necessary? The register-based analysis of 11,230 cases. Surg. Endosc. 30(10), 4363–4371 (2016).
Amato, G., Agrusa, A. & Romano, G. Fixation-free inguinal hernia repair using a dynamic self-retaining implant. Surg. Technol. Int. 22, 107 (2012).
Amato, G. et al. Dynamic inguinal hernia repair with a 3d fixation-free and motion-compliant implant: A clinical study. Surg. Technol. Int. 24, 155–165 (2014).
Amato, G., Lo Monte, A. I., Cassata, D. G., Romano, G. & Bussani, R. A new prosthetic implant for inguinal hernia repair: Its features in a porcine experimental model. Artif. Organs 35(8), E181–E190 (2011).
Amato, G. et al. Biologic response of inguinal hernia prosthetics: A comparative study of conventional static meshes versus 3D dynamic implants. Artif. Organs 39(1), E10 (2015).
Amato, G. et al. Neomyogenesis in 3D dynamic responsive prosthesis for inguinal hernia repair. Artif. Organs. 42(12), 1216–1223 (2018).
Amato, G. et al. Neo-nervegenesis in 3D dynamic responsive implant for inguinal hernia repair: Qualitative study. Int. J. Surg. 76, 114–119 (2020).
Amato, G. et al. Enhanced angiogenesis in the 3D dynamic responsive implant for inguinal hernia repair ProFlor®. Artif. Organs. 00, 1–10 (2021).
Amato, G. et al. Fixation free inguinal hernia repair with the 3D dynamic responsive prosthesis ProFlor: Features, procedural steps and long-term results. Int. J. Surg. Open 21, 34–44 (2019).
Amato, G., Romano, G., Calò, P. G., Di Buono, G. & Agrusa, A. First-in-man permanent laparoscopic fixation free obliteration of inguinal hernia defect with the 3D dynamic responsive implant ProFlor-E®: Case report. Int. J. Surg. Case Rep. 77S(Suppl), S2–S7 (2020).
Amato, G. et al. Inguinal hernia: Defect obliteration with the 3D dynamic regenerative scaffold Proflor™. Surg. Technol. Int. 4, 38 (2021).
Klinge, U., Klosterhalfen, B., Muller, M. & Schumpelick, V. Foreign body reaction to meshes used for the repair of abdominal wall hernias. Eur. J. Surg. 65, 665–673 (1999).
Weyhe, D. et al. Experimental comparison of monofile light and heavy polypropylene meshes: Less weight does not mean less biological response. World J. Surg. 30(8), 1586–1591 (2006).
O’Dwyer, P. J. et al. Randomized clinical trial assessing impact of a lightweight or heavyweight on chronic pain after inguinal hernia repair. Br. J. Surg. 92, 166–170 (2005).
Iftikhar, N. & Kerawala, A. Quality of life after inguinal hernia repair. Pol. Przegl Chir. 93(3), 1–5. https://doi.org/10.5604/01.3001.0014.8218 (2021).
Kebabci, E., Ozturk, S. & Unver, M. Outcomes of endoscopic totally extraperitoneal (TEP) repair of clinically occult inguinal hernia diagnosed with ultrasonography. Pol. Przegl Chir. 93(4), 11–14 (2021).
Mitura, K., Garnysz, K., Wyrzykowska, D. & Michałek, I. The change in groin pain perception after transabdominal preperitoneal inguinal hernia repair with glue fixation: A prospective trial of a single surgeon’s experience. Surg. Endosc. 32, 4284–4289. https://doi.org/10.1007/s00464-018-6178-0 (2018).
Mitura, K. & Kozieł, S. The influence of different sterilization types on mosquito net mesh characteristics in groin hernia repair. Hernia 22(3), 483–490. https://doi.org/10.1007/s10029-018-1756-6 (2018).
Bjurstrom, M. F., Nicol, A. L., Amid, P. K. & Chen, D. C. Pain control following inguinal herniorrhaphy: Current perspectives. J. Pain Res. 7, 277–290 (2014).
Charalambous, M. P. & Charalambous, C. P. Incidence of chronic groin pain following open mesh inguinal hernia repair, and effect of elective division of the ilioinguinal nerve: Meta-analysis of randomized controlled trials. Hernia 22, 401–409 (2018).
Katkhouda, N. et al. Use of fibrin sealant for prosthetic mesh fixation in laparoscopic extraperitoneal inguinal hernia repair. Ann. Surg. 233, 18–25 (2001).
Pascual, G. et al. Cytotoxicity of cyanoacrylate-based tissue adhesives and short term preclinical in vivo biocompatibility in abdominal hernia repair. PLoS ONE 11(6), e0157920 (2016).
Shah, N. et al. Mesh fixation at laparoscopic inguinal hernia repair: A meta- analysis comparing tissue glue and tack fixation. World J. Surg. 38, 2558–2570 (2014).
Reynvoet, E. & Berrevoet, F. Pros and cons of tacking in laparoscopic hernia repair. Surg. Technol. Int. 25, 136–140 (2014).
Peach, G. & Tan, L. C. Small bowel obstruction and perforation due to a displaced spiral tacker: A rare complication of laparoscopic inguinal hernia repair. Hernia 12(3), 303–305 (2008).
Fitzgerald, H. L., Orenstein, S. B. & Novitsky, Y. W. Small bowel obstruction owing to displaced spiral tack after laparoscopic TAPP inguinal hernia repair. Surg. Laparosc. Endosc. Percutaneous Tech. 20(3), e132–e135 (2010).
Haltmeier, T. & Groebli, Y. Small bowel lesion due to spiral tacks after laparoscopic intraperitoneal onlay mesh repair for incisional hernia. Int. J. Surg. Case Rep. 4(3), 283–285 (2013).
Amato, G. et al. Histological findings of the internal inguinal ring in patients having indirect inguinal hernia. Hernia 13, 259–262 (2009).
Amato, G. et al. Nerve degeneration in inguinal hernia specimens. Hernia 15, 53–58 (2011).
Amato, G. et al. Damage to the vascular structures in inguinal hernia specimens. Hernia 16, 63–67 (2012).
Amato, G. et al. Muscle degeneration in inguinal hernia specimens. Hernia 16, 327–331 (2012).
Amato, G. et al. Histological findings in direct inguinal hernia. Hernia 17, 757–763 (2013).
Amato, G. et al. Combined inguinal hernia in the elderly: Portraying the progression of hernia disease. Int. J. Surg. 33(Suppl 1), S20–S29 (2016).
Amato, G. et al. The septum inguinalis: A clue to hernia genesis? J. Invest. Surg. 31, 1–9 (2018).
Amato, G. et al. Inguinal hernia: The destiny of the inferior epigastric vessels and the pathogenesis of the disease. Surg. Technol. Int. 18, 36 (2020).
Rudiman, R., Winata, A. A. New laparoscopic surgery in inguinal hernia repair. In Hernia Surgery September LicenseCC BY 3.0, https://doi.org/10.5772/intechopen.89028. (2019).
Rabe, R., Yacapin, C. P., Buckley, B. S. & Faylona, J. M. Repeated in vivo inguinal measurements to estimate a single optimal mesh size for inguinal herniorrhaphy. BMC Surg. 2(12), 19 (2012).
Amato, G. et al. A regenerative scaffold for inguinal hernia repair: MR imaging and histological cross evidence: Qualitative study. Int. J. Surg. 96, 106170 (2021).
Simons, M. P. et al. European Hernia Society guidelines on the treatment of inguinal hernia in adult patients. Hernia 13(4), 343–403 (2009).
HerniaSurge Group. International guidelines for groin hernia management. Hernia 22, 1–165 (2018).
Bittner, R. et al. Guidelines for laparoscopic treatment of ventral and incisional abdominal wall hernias: International Endohernia Society (IEHS). Surg. Endosc. 28, 2e29 (2014).
Funding
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Author information
Authors and Affiliations
Contributions
G.A.: Conceptualization, writing the manuscript. A.A.: Investigation, data Curation. P.G.C.: Methodology, supervision. G.D.B.: Investigation, resources. S.B.: Conceptualization, review. A.C.: Project administration, supervision. G.Z.: methodology, G.R.: review and editing. All authors have read the final manuscript and approved its submission for publication.
Corresponding author
Ethics declarations
Competing interests
Dr. Giuseppe Amato is the developer and patent owner of the device being discussed in the article, despite in the recent past he had financial/commercial relationships, such connections are not present at the moment. Drs. Antonino Agrusa, Pietro Giorgio Calò, Giuseppe Di Buono, Salvatore Buscemi, Adriana Cordova, Guido Zanghì and Giorgio Romano have no conflict of interest or financial ties to disclose. Overall, except those disclosed by Dr. Giuseppe Amato, there are no other relationships or activities to declare that could have potentially influenced the outcomes of the submitted work.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Amato, G., Agrusa, A., Calò, P.G. et al. Fixation free laparoscopic obliteration of inguinal hernia defects with the 3D dynamic responsive scaffold ProFlor. Sci Rep 12, 18971 (2022). https://doi.org/10.1038/s41598-022-23128-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-022-23128-6
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.