Introduction

Glucose monitoring has a fundamental role in the prevention and management of diabetes, the fastest growing disease in the world1. Nearly 10% of the world’s population, living with diabetes, rely on periodic glucose monitoring to manage the disease. Traditional capillary blood glucose (CBG) meters measure the concentration of glucose from capillary blood samples obtained by the patient. Each measurement requires pricking a finger with a lancet to obtain a blood sample, a painful and inconvenient procedure.

Current trends lean towards continuous glucose monitoring (CGM). Novel devices based on interstitial fluid (ISF) include Freestyle Libre from Abbot2, Dexcom G6 from Dexcom3 (subcutaneous needle) and Eversense from Senseonics4 (subcutaneous sensor). These devices improve the quality of life for people living with diabetes5, and effectively, help patients to better manage the disease. For instance, persons with type-1 diabetes who use CGM devices are less prone to hypoglycemia6,7, stay longer within target glucose range8,9, and have reduced HbA1c levels6,7,8,9. Similar results have been observed in patients with type 2 diabetes10,11. Devices for CGM enable glucose trend tracking, which has sparked interest beyond diagnosed cases of diabetes. For example, CGM can motivate people with pre-diabetes to make improvements in life-style choices. It can also aid in optimizing the recovery and peak performance of athletes through nutrition management12,13.

While traditional glucose meters use capillary blood, most commercial CGM devices measure glucose from ISF, which can be accessed using minimally invasive or non-invasive approaches. Many compounds, such as glucose, are transported from the blood into the cells via ISF14,15. The glucose concentration in ISF strongly correlates with the blood glucose concentration16,17. This correlation makes ISF attractive for glucose monitoring. However, sampling of ISF non-invasively is challenging due to the barrier function of the skin. Thus, all approaches based on ISF that have reached commercial success are based on needles or microneedles that penetrate the skin and reach the ISF in the dermis.

Other approaches for non-invasive CGM under active research include sampling of interstitial fluid with reverse iontophoresis18,19,20 and ultrasound21,22, and detection of glucose through the skin using light23, or radio waves24. These approaches, however, are yet to demonstrate their capability of delivering the required accuracy and precision out of laboratory settings.

Both invasive and non-invasive methods for CGM in interstitial fluid face the common challenge of a lag time between the concentrations of glucose in ISF and in blood. The glucose concentration in ISF depends on the diffusion rate of glucose from capillaries to the interstitial fluid and the rate of glucose intake by cells25. Consequently, most of the time, the glucose concentration in ISF lags the concentration of glucose in blood26,27,28. The magnitude of this lag varies with time26 and within a person over time depending on their physical activity and blood glucose concentrations27.

The changing lag time increases the complexity of determining the accuracy (proximity to the true value) and precision (spread) of individual glucose measurements on ISF using blood glucose measurements as the reference. The accuracy of glucose monitoring methods is typically quantified using the Mean Absolute Relative Difference (MARD) and, with increasing popularity, also the Precision Absolute Relative difference (PARD)29. Error grids such as the Consensus (i.e., Parkes) Error Grid (CEG)30 visually represent the accuracy. These error grids display predicted vs. reference blood glucose concentrations and are divided into risk zones A-E that correspond to the severity of clinical outcome associated with a discrepancy between the predicted and reference glucose31.

In our previous work32, we introduced MHD as a novel physical mechanism for extraction of dermal interstitial fluid. Also, using an ex vivo model, we demonstrated a 13-fold increase in extraction rate of glucose when a 300 mT magnetic field is added32. In this paper we present for the first time the use of MHD for extraction of interstitial fluid in healthy volunteers and its application to glucose monitoring. Over the course of a glucose tolerance test, we obtain samples of interstitial fluid with MHD (needle-free) from the skin of healthy volunteers. Concurrently, we measure CBG with a reference glucose meter. We determine the glucose concentration in the ISF samples using a fluorometric assay. Then, we investigate the correlation between the two sets of measurements. To improve the reliability of the correlation analysis, we calibrate the ISF measurements using a least squares optimization algorithm to consider the temporal lag.

Results

Sampling of interstitial fluid using MHD

We developed a setup for ISF sampling, consisting of acrylic wells, electrodes, and block magnets (Fig. 1a-d). The position of the wells mimics the position of a fitness watch, and we positioned them on the flattest part of the dorsal side of the lower arm.

Figure 1
figure 1

Experimental set up and scheme. (a) Top-view of the acrylic wells placed on the volunteer’s lower arm. (b) A side-view of the acrylic wells. (c) The volunteer’s arm was placed between two neodymium magnets separated by an aluminum support. ( d) A schematic presentation of the set-up and the MHD extraction. Two Ag/AgCl disk electrodes inside the well drive a current through the skin (J). A magnetic field with direction (B) orthogonal to the current is applied to create Lorentz force (F) pointing towards the skin surface. The scheme is not drawn to scale. Figure Credits: Artic Frame Studio https://www.arcticframe.com.

MHD extraction during glucose tolerance tests

The performance of MHD extraction was evaluated by conducting a series of glucose tolerance tests. Figure 2a shows a schematic of the timeline for an individual glucose tolerance test.

Figure 2
figure 2

Glucose tolerance tests. ( a) Schematic presentation of the glucose tolerance test timeline (black dashed line). Volunteers entered the test after overnight fasting (9 h). 5 min MHD extractions were conducted with alternating current polarity. The red and blue bars indicate the direction of the applied electric current for each extraction. Capillary blood glucose measurements, represented by red blood drops and red dashed lines, were taken twice before sugar intake (green dashed line). 8 more capillary blood glucose readings were obtained after the glucose drink with 20 min intervals. The skin hydration and transepidermal water loss readings were recorded before and after the test. (b) Raw data from a representative experiment. These data are from the MHD cathode central well (grey diamonds), the control well (blue triangles), and reference blood glucose measurements (red circles). The first data point after warm-up is shown as a star. Error bars for the MHD and diffusion raw data are based on the standard error of the diffusion data recorded after the warm-up period. Error bars for the CBG concentrations are based on the standard error of our reference device. (c) All 10 experiments shown after linear interpolation of the reference CBG and after applying calibration. Each column consists of two experiments involving a single participant. Error estimates for the interstitial fluid glucose (ISFG) (not shown) are based on a combination of the standard error of the control well measurements and the standard error of our reference device and are in almost all cases ± 1–2 mM.

The extraction voltages drifted during the first ~ 100 min of each experiment (warm-up period). Thus, MHD samples extracted prior to 100 min were not used to investigate the correlation between ISFG and CBG concentrations.

We used an enzymatic assay to determine the amount of glucose in the extracted samples. Figure 2b). MHD and CBG glucose readings were taken from both the left (n = 5, Fig. 2c #1a-#5a) and the right (n = 5, Fig. 2c #1b-#5b) arms. The results from all 10 experiments after least squares-based calibration, as described in “Materials and Methods”, are presented in Fig. 2c. During the 300 min of sampling, we obtained a blood glucose peak for each volunteer. In most cases, the same individual exhibited a similar profile for the glucose tolerance test in the two experiment days. Volunteer #1, for instance, had a slow rise of CBG reading and only slight drop after the peak at 200 min before the experiment was concluded on the two experiment days (#1a and #1b).

Correlation between glucose in extracted ISF samples and blood glucose concentrations

The results from the glucose tolerance tests are shown in a CEG (Fig. 3). In total, the ten experiments yielded 88 measurement pairs. Of these measurement pairs, 84% fell within Zone A and 16% fell within Zone B. The corresponding MARD value is 12.9%. We found a correlation (R2 = 0.92) between the MHD extracted glucose concentrations and the reference blood glucose measurements. The resulting relationship between MHD glucose concentrations and reference CBG indicates the presence of a temporal lag between CBG and ISFG. Such a lag has also been observed by others17,27,28,33,34, and was thus expected. We accounted for this lag in our calibration (described in Materials and Methods).

Figure 3
figure 3

Correlation of glucose concentrations in extracted ISF and capillary blood samples. CEG showing 88 measurement pairs from 10 experiments. These data are shown after applying the basic least squares calibration algorithm. Percentages within each zone are as follows: Zone A: 84.0%, Zone B: 16.0%, Zones C-E: 0%. The corresponding MARD is 12.9%.

Effect of extraction on the skin

To investigate potential effects of MHD extraction on the skin, we took photographs of the skin (Fig. 4a) and measured the transepidermal water loss (TEWL) before and after the glucose tolerance test. Thus, we examined the effect of multiple extractions. The skin photographs were visually analysed by an experienced dermatologist. We observed slight redness under extraction wells (distal and central wells) immediately after the experiment for all participants. This redness disappeared after a few hours. In some cases, the adhesive caused slight temporary redness on the participant’s arm. Figure 4b shows that the TEWL rate changed due to the extraction experiment on the volar forearm locations of the two active wells and the control well. Figure S3a, shows the TEWL measured at the extraction sites before and after the extraction experiment. The results demonstrate a higher TEWL at the more distal location on the forearm, which is consistent with the reported values and demonstrate a known trend for higher TEWL on a more distal part of the forearm35. Although the TEWL rate increased in measurements done 15 min after the test, a comparable increase was also observed for the control well, implying that this observation mostly resulted from the skin occlusion, a known cause of a transient increase in TEWL. Occlusion is occasionally used as a skin barrier stress test in combination with TEWL measurements36. Furthermore, we observed increased epidermal and dermal water content after the extraction experiments (Fig. 4c), with a greater increase at the MHD active well locations (distal and central) compared to the control well location (proximal). The p-test of epidermal and dermal water content in MHD wells compared to the control well resulted in statistically significant difference of p = 0.007 and p = 0.002, respectively.

Figure 4
figure 4

Effect of extraction on the skin. (a) Photographs of the skin MHD well sites before and after the glucose tolerance test that includes 30 extractions lasting 5 min. The extraction well outline was marked with a red makeup pen before the experiment. (b) ΔTEWL and (c) Δ skin water content for each well from all 10 experiments. Δ values were obtained by measuring the skin parameters before and after the experiment and by calculating the difference between these values (Δ = after – before). p values were obtained via t-test, with p ≤ 0.05 considered significant.

Discussion

While ongoing research efforts target non-invasive solutions to CGM, such solutions have so far been elusive. Over the last few years, CGM devices based on subcutaneous needles have developed rapidly to meet high standards for accuracy and safety1. These technologies benefit a rapidly growing number of people with diabetes, in particular, insulin users37. However, the adoption of CGM devices by people with- or at risk of getting diabetes remains low38. To solve this problem, further development of existing and new technologies for CGM should lead towards reducing costs, improving usability, and minimizing environmental impact. A clear path towards improvement on these 3 areas is the development of a non-invasive CGM solution. For instance, a solution based on extraction of interstitial fluid with MHD can be implemented in the form of a reusable wearable device combined with disposable low-cost screen-printed biosensors.

In this study we present needle-free ISF glucose sampling using MHD in healthy volunteers. We performed 10 glucose tolerance tests on 5 healthy volunteers while sampling CBG concentrations and ISFG concentrations (Fig. 2). The resulting data and calibration yielded a MARD of 12.9% and a PARD of 13.1%. These values are within the range 10–15% typically reported by commercial CGM devices39. Furthermore, commercial devices leverage sophisticated calibration algorithms to improve their accuracy. While being outside the scope of this investigation, a calibration algorithm that considers environmental settings, MHD parameters, and previous measurements (glucose trends), could be developed to improve the accuracy of the ISFG measurements.

The different CEG zones (Fig. 3) represent different risk levels. Zone A represents the lowest risk. Measurement datapoints falling within the Zone A impose no risk of adverse clinical outcome for the patient. The risk increases through Zones B, C, D, and E. Zone E is associated with the most severe risk for the patients.

The emergence of solutions for CGM based on the use of interstitial fluid has motivated research on the lag time between blood glucose and glucose in ISF34. However, no clear consensus on this topic has yet been reached. The microcirculation in the skin is a complex dynamic phenomenon that causes inter- and intra-subject variations in the lag time26,28,40. Furthermore, different ISF reservoirs and different sampling mechanisms can lead to different results. While the clinical relevance of the ISFG has yet to be fully understood, current approaches for CGM in the interstitial fluid using needles have proven to be valuable for people with both type 1 and type 2 diabetes. Our results are consistent with the literature, showing a temporal lag between CBG concentrations and ISFG (Fig. S5). Our results also indicate different lag times among different individuals. For this reason, we calibrated our measurements on ISF for individual experiments (Supplementary Information). Hence, the specific lag times we present should be considered as parameters in an empirical model – rather than direct quantitative and absolute measurements of the time lags. Finally, we note that continuous ISFG and CBG measurements would be better suited to lag quantification. However, the experiment was not designed to target the lag times as that quantification was not a primary goal of this work. Thus, the temporal resolution of our data imposes limitations on the lag assessment and the lag estimates should be treated accordingly. Our future research on a system integrating MHD extraction with electrochemical biosensors will enable a more thorough investigation of the lag time and discussion of its clinical implications.

The determination of glucose concentration in interstitial fluid was done by using a standard fluorometric assay. In this way, the glucose concentration in ISF and blood were measured by two independent methods, in order to minimize the uncertainty related to electrochemical biosensors commonly used in glucometers. In future work, the ISF glucose will be determined by an electrochemical biosensor, which will allow miniaturization of the device. We have demonstrated the integration of MHD and electrochemical sensors for glucose monitoring in a previous work 41. As evident from that work, the use of electrochemical sensors overcomes the need for sample collection because the samples can be measured at the extraction site. As the electrochemical sensors are thin (e.g., 200 μm), small magnets can be arranged in close proximity to the skin to generate the desired B-field at the extraction site. For instance, a Neodymium, N52, nickel-plated magnet 5 × 5 × 3 mm (Q-05–05-03-N52N, Supermagnete, Germany) provides a similar magnetic field strength (300 mT) near its surface.

The use of electrochemical sensors also allows reducing the sample dilution. In this study, the interstitial fluid samples were collected and diluted into 200 µL of PBS. Extraction times of 5 min allowed us to collect sufficient volumes of interstitial fluid to enable the quantification of glucose using standard fluorometric assays. Smaller sample volumes and consequently shorter extraction times should be needed when the sample dilution is lower. The extraction time limits the frequency of the measurements. Hence, shorter extraction times are desired. Most CGM devices measure glucose levels at time intervals (e.g., up to every 5 min) rather than continuously. A consequent reduced accuracy during periods of rapidly changing glucose levels has been documented42. However, CGM offers the advantage of facilitating frequent glucose measurements for prolonged periods. This gives patients and physicians reliable information about the health condition. In particular, time in range (TIR), enabled by CGM, has demonstrated big value in diabetes diagnosis and treatment43,44,45.

Figure 4 shows the effect of the MHD extraction on skin. The data demonstrated that the skin barrier function was not significantly affected by the extraction when comparing to skin under the control well. We observed an increase in TEWL under each well, including the control (Fig. 4b). Its presence in the control well implies that this increase mostly results from the skin occlusion. The temporary dysfunction of the skin barrier after occlusion may result from increased water content in stratum corneum in the absence of evaporation, leading to temporary disrupted structure of the skin barrier’s lipid multilayer46. Related to this disruption, we observed increased epidermal and dermal water content after the extraction experiments (Fig. 4c). This increase was greater in the MHD active sensor locations compared to the control well. One may hypothesize that any active means to percolate fluid flow through the skin towards its surface likely leads to a transient increase in trapped fluid within the skin. The participants were queried about the presence of adverse skin reactions (e.g., redness, rash, itching) on the test day, the next day, one week after the test day, and four weeks after the test day. Of the 5 participants, one reported erythema and absence of other symptoms on the day of the experiment. Another participant reported erythema persisting the day of the experiment and to the next day but, more importantly, none of the participants had any evidence of rash after one week or after four weeks post experiments. Overall, the volunteers considered the experience pleasant. They reported slight tingling sensation in the beginning of the experiment. The sensation was usually present only in the first few seconds of each extraction. Furthermore, in most cases the sensation usually disappeared within the first hours of the experiment. These reports are consistent with the skin property and skin reaction data indicating that the extraction itself is not causing damage to the skin.

Outlook

We present the first ever study in humans of the safety and effectiveness of MHD to extract dermal ISF to enable needle-free monitoring of glucose. Our results show a high correlation between the concentration of glucose in blood and the concentration of glucose in samples of ISF extracted with MHD. Furthermore, no evidence of long-lasting effects on the skin were observed. Our results indicate a high potential value of MHD as a tool to enable needle-free glucose monitoring. Beyond glucose monitoring, sampling of ISF with MHD has the potential to enable monitoring of other analytes also present in ISF.

Materials and methods

Materials

D-Glucose (#G8270-100G) was purchased from VWR, 10 mM PBS powder pouches (#P3813), Glucose oxidase (GOx, #G7141-1MU), Horseradish peroxidase (HRP, #P8250-25KU) from Sigma Aldrich, and AmplexRed (#A12222) from ThermoFisher. All the chemicals were used without additional purification. 75 mM glucose drinks (GlucoTest) were purchased from Suomen diabeteskauppa (Finnish Diabetes Shop, www.diabeteskauppa.fi).

Extraction of interstitial fluid

Extraction involved applying 300 µA/cm2 current density through the skin using a custom-made current source (block diagram in Fig. S1), two Ag/AgCl disk electrodes (r = 2 mm, Warner Instruments) and two neodymium magnets (70 × 70 × 30 mm3, Goliath, Supermagnete, Germany). An aluminium spacer maintained 8 cm distance between the magnets. The extracted ISF was collected into circular acrylic wells (fluid contact area to the skin: 0.5 cm2) designed for this experiment and ordered from Protohouse (Salo, Finland). These wells were attached to the volunteer’s wrist using double-sided tape (9474LE, 3 M) and filled with 200 µL of PBS. Two wells were used for extraction (distal and central), whereas the third well (proximal) was used as a control well (passive diffusion alone) (Fig. 1a). The control well was used for two purposes. First, to determine the background noise in the sample analysis. Second, as a reference for the skin damage assessment. The distal well was placed 2 cm from the wrist bone (ulna head). The central and proximal wells were placed further along the arm, with 5 mm outer wall-to-wall distance from each other. If the volunteer had any skin damage, such as scars or rash, the wells were placed outside of affected areas. The wells and the skin were rinsed twice with PBS. Rinsing was followed by filling the wells with 200 µL of PBS and then starting the extraction. Two Ag/AgCl disk electrodes (r = 2 mm, E202, Warner Instruments) inserted into the wells ensured a low-impedance electrical connection with the solution and prevented pH changes due to the electric current. The electrodes were positioned carefully into the PBS solution while avoiding touching the skin. Both the current intensity and the voltage between the extraction wells was monitored using a DMM (34465A, Keysight).

The two permanent magnets created a magnetic field along the lateral axis of the volunteer’s arm (Fig. 1c). At the extraction site, the measured magnetic field was 300 mT. The sampling of ISF with MHD was induced by driving an electric current (300 µA/cm2) through the electrodes, the electrolyte in the extraction wells, and the skin in the magnetic field (Fig. 1d and Fig. S1). The current (J) was driven orthogonally to the magnetic field (B) to create a Lorenz force (F) pointing from the dermis towards the skin surface, thus inducing magnetohydrodynamic extraction of ISF. The first extraction was always conducted with the distal well working as the cathode. Each extraction was followed by a 5 min collection time, where samples from each well were transferred into microtubes. The wells were then rinsed with PBS and fresh PBS added for the subsequent extraction. The polarity of the current and the direction of the magnetic field were swapped for each subsequent extraction. Reference blood glucose concentrations of the volunteer were measured using a glucometer (Contour XT, Ascensia Diabetes Care) following the manufacturer’s instructions. The CBG samples were taken from the arm which was not used for MHD extraction. This meant the right arm for the first five experiments and the left arm for the last five.

Glucose tolerance test

All participants for the study were volunteers who had submitted informed consent in writing before the tests. The experimental methodology complies with the clinical pilot study plan reviewed and approved by the Helsinki and Uusimaa Hospital District Ethical Committee II. This study is a clinical medical device study according to The Finnish Act on Medical Devices and Equipment (Laki Terveydenhuollon Laitteista ja Tarvikkeista 629/2010) and was performed in accordance with the guidelines within. This study was performed in accordance with the Declaration of Helsinki. The volunteer entered the test after fasting for 9 h. The test started with a first CBG measurement followed by an extraction of ISF. After 120 min, a second CBG measurement was taken. Then, the volunteer consumed a glucose drink (75 g of glucose, GlucoTest). Afterwards, 8 more CBG measurements were taken, one every 20 min (total of 10 during the test). The MHD extractions and CBG measurements were performed simultaneously over the duration of the experiment. MHD and CBG glucose readings were taken from both the left (n = 5) and the right (n = 5) arms. In total, 10 tests were performed on five different volunteers (2 males and 3 females, age range 28–43 years). Each volunteer participated twice in different days. First experiment was conducted from the left arm and the second from the right arm.

Skin parameter measurements

Skin was washed with soap (Neutral ®), rinsed with water, and dried with a paper towel both before and after the extraction experiment. After the wash, TEWL and skin moisture were measured at each extraction site and at the diffusion control site. The time between washing the arm and skin measurements was kept constant (15 min). Measurements were performed using VapoMeter47,48, and MoistureMeterD49 Compact (Delfin Technologies, Finland) according to the manufacturer’s instructions. Each location was measured three times, and an average of the three measurements was used in the analysis.

Determination of glucose concentration in the extracted ISF samples

The glucose concentration in the extracted ISF samples was measured using a GOx/HRP/AmplexRed assay. First, 20 µL of standard glucose solutions and each extracted sample were individually pipetted in a 384-well plate in duplicate (black, Corning). Next, we added 10 µL of 250 µM AmplexRed reagent. After adding the reagent, the plate was transferred to a plate reader (Varioskan LUX, Thermo Fisher), which automatically injected 10 µL of GOx/HRP solution to each well to initiate a cascade of reaction: \({C}_{6}{H}_{12}{O}_{6}+{O}_{2}\stackrel{GOx}{\to }{C}_{6}{H}_{10}{O}_{6}+{H}_{2}{O}_{2}; {H}_{2}{O}_{2}+ \mathrm{AmplexRed}\stackrel{HRP}{\to } resorfurin+ {H}_{2}O\)

Resorufin emits a fluorescent signal at 585 nm upon its excitation with λ = 571 nm light50. The reaction was allowed to progress for 8 min, and the end point value was used to quantify the glucose concentration in each sample by using a standard calibration curve obtained from the standard glucose samples.

Samples from the control well were subtracted from each corresponding MHD cathode well sample. Subtracting the control sample readings serves to remove the contribution of glucose due to passive diffusion from the total glucose in our samples. The distal and central well MHD cathode samples for each sample pair were averaged. Next, least squares fits to the control-subtracted MHD cathode samples vs. CBG concentrations were made for each individual experiment. The MHD control-subtracted cathode samples for each experiment were then divided by their respective slopes from the least squares fits. This approach enables a robust analysis of correlation with minimal data processing. The corresponding Consensus Error Grid is shown in Fig. 3. It is important to note that this calibration is done retrospectively, using all the data from each individual experiment, as opposed to in real-time. We acknowledge the importance of a real-time algorithm. For these data, a real-time algorithm could rely on two CBG reference points: one at the start, and then another near the glucose peak. These points, along with their corresponding ISFG measurements would yield the initial ISFG, as well as scale the ISFG to match the CBG range. However, the lag complicates this calibration. Consequently, alternative calibration methods and improved understanding of the lag via additional clinical studies are desirable. With this in mind, the data presented here are suboptimal for the development of such an algorithm.

The temporal lag between the CBG and ISFG concentrations was considered prior to the calibration. A detailed description of this adjustment and the lag characterization is given in the Supplementary Information (Fig. S5).

The resulting predicted MHD glucose concentrations and CBG reference data were then averaged for each sample over all experiments (i.e. Sample n from all experiments 1–10 were averaged, repeated for each Sample until Sample N). There was one exception for the first experiment, which used 10 min instead of 5 min extractions for the first 4 extractions during the initial warm-up period. Thus, the warm-up period for that experiment was longer than for the other experiments. Consequently, there were two fewer samples in this first experiment compared to the other experiments.

Data processing

We collected data from the MHD glucose concentrations, CBG reference measurements, skin moisture measurements, and extraction voltages. This pre-processing was performed in Python using standard data-handling libraries (Numpy51, SciPy52, Pandas53, Statsmodels54, and Plotly55 for display).

The MHD glucose and reference CBG concentrations were measured at different times (see experimental protocol in Fig. 2a). Thus, to compare the two, the reference CBG concentrations were matched in time to the MHD glucose concentrations via linear interpolation. (i.e. a line is fit between two CBG measurements. The point on that line used as a reference CBG is the point that corresponds to the MHD sample time that is being compared. The initial relative MHD and CBG measurement timelines are consistent with the example shown in Fig. 2b and the temporally aligned MHD and CBG glucose concentrations are shown in Fig. 2c).

The times assumed for each MHD glucose concentration sample in the analysis were their collection times (within 1 min of the current being shut off after each extraction). The final MHD concentrations used in our analysis were the mean concentrations when combining the distal and central wells (and thus the sample times for the two wells were averaged for the final analysis). Note that the MHD glucose concentrations were diluted in 200 µL of PBS – thus they are dissimilar from the ISFG concentration, but rather correlate with it. Our calibration compensates for this dilution.

Extraction voltages (Fig. S4) were used to determine the warm-up time by assessing how they drifted on the y-axis over time.

To assess covariance, we performed an ANOVA56 and Tukey’s Honestly Significant Difference test57. Our models included arm (categorical variable, left or right), the five individual participants (categorical variable, grouped as two experiment days per participant), the ten individual experiments explored individually, regardless of repeat participants (categorical variable), the standard error of the control well (over each experiment duration), and time (continuous variable, in seconds).

We set p < 0.05 as the criteria to reject the null hypothesis that the predicted ISFG is equivalent across the different categories in our model. The ANOVA demonstrated that the right and left arms produce equivalent results. When adding participant as a model parameter, the null hypothesis was rejected. To identify the reason, we performed a Tukey’s HSD test and found that participant 2 was an exceptional case. A subsequent Tukey’s HSD test of participant 2, revealed that the experiment #2b (see Fig. 2) was exceptional. This exception is most likely due to an outlier in the ISFG between 150 and 200 min (Fig. 2c). Finally, the ANOVA indicates some change in correlation between predicted ISFG and reference CBG over time (p = 0.008). Results from this analysis are summarized in Table 1.

Table 1 Model included in the ANOVA analysis. Asterisks indicate parameters having p < 0.05, resulting in rejection of the null hypothesis.