Abstract
Mass spectrometry-based targeted proteomics allows objective protein quantitation of clinical biomarkers from a single section of formalin-fixed, paraffin-embedded (FFPE) tumor tissue biopsies. We combined high-field asymmetric waveform ion mobility spectrometry (FAIMS) and parallel reaction monitoring (PRM) to increase assay sensitivity. The modular nature of the FAIMS source allowed direct comparison of the performance of FAIMS-PRM to PRM. Limits of quantitation were determined by spiking synthetic peptides into a human spleen matrix. In addition, 20 clinical samples were analyzed using FAIMS-PRM and the quantitation of HER2 was compared with that obtained with the Ventana immunohistochemistry assay. FAIMS-PRM improved the overall signal-to-noise ratio over that from PRM and increased assay sensitivity in FFPE tissue analysis for four (HER2, EGFR, cMET, and KRAS) of five proteins of clinical interest. FAIMS-PRM enabled sensitive quantitation of basal HER2 expression in breast cancer samples classified as HER2 negative by immunohistochemistry. Furthermore, we determined the degree of FAIMS-dependent background reduction and showed that this correlated with an improved lower limit of quantitation with FAIMS. FAIMS-PRM is anticipated to benefit clinical trials in which multiple biomarker questions must be addressed and the availability of tumor biopsy samples is limited.
Similar content being viewed by others
Introduction
In targeted therapeutics, expression levels of primary drug targets are often associated with patients’ clinical responses to treatment and thus serve as predictive biomarkers. Precise measurement of such biomarkers plays a crucial role in selecting patients most likely to benefit from therapy1. The U.S. Food and Drug Administration has approved companion diagnostic (CDx) assays to select eligible patients for treatment on the basis of molecular characteristics of the tumor2. To date, CDx assays have been approved for just five protein targets, namely, ALK, EGFR, HER2, KIT, and PD-L1, and all of these protein assays are based on immunohistochemistry (IHC)3. For example, the HercepTest was the first protein-based CDx assay that identified HER2-positive cancer patients eligible for trastuzumab treatment4. Currently, an IHC stain intensity of 3 + , as annotated by a pathologist, is considered HER2 positive. A HER2 IHC of 2 + is also considered HER2 positive if an additional fluorescence in situ hybridization assay (FISH) proves that the gene is amplified. Although clinical studies have demonstrated a strong correlation between IHC-based HER2 expression level and trastuzumab sensitivity, a series of mixed results from recent clinical trials5 and ambiguous quantitation of the antibody-based HercepTest necessitates more objective and quantitative protein assays6. Furthermore, recent advances in antibody–drug conjugate (ADC) therapies that conjugate a small-molecule drug, such as the topoisomerase I inhibitor deruxtecan, to trastuzumab demonstrate that such ADCs can efficiently target tumor cells with low HER2 expression by leveraging the conjugated warheads in heterogeneous tumors7. This new focus on tumors with low HER2 expression highlights an important need for more sensitive and quantitative protein assays.
In addition to patient stratification, quantitative analysis of protein biomarkers in biopsy samples is commonly applied to assess the pharmacodynamic effects of drugs under clinical investigation. This includes monitoring the abundance of therapeutic target proteins as well as downstream efficacy markers upon drug treatment8,9. IHC is also a standard method for pharmacodynamic assessments in clinical trials.
Mass spectrometry (MS)-based targeted proteomics with selected reaction monitoring (SRM) has emerged as a promising technology for protein quantitation to overcome the limitations of IHC. In reports of clinical trials using SRM-based HER2 measurements, clinical cutoff values of HER2 expression level, which is predictive of response to trastuzumab, were determined by SRM readouts6,10. The SRM method, using a triple-quadrupole mass spectrometer, is known for its high sensitivity, which is achieved by fixing the mass analyzer to transmit only selected ions. However, sensitivity can be hampered by high non-specific background signal that interferes with the signals of targeted ions. This is a very common problem in analyzing high-complexity clinical samples and has been addressed by using parallel reaction monitoring (PRM) to leverage high-resolving power mass analyzers, allowing a flexible choice of fragment ions post-acquisition11,12,13,14. With greater signal-to-noise ratios than SRM, PRM is advantageous when an assay suffers from high levels of non-specific background signal, as is often the case for clinical analyses involving formalin-fixed, paraffin-embedded (FFPE) tumor biopsies15,16.
In theory, additional reduction of unwanted signals may further improve assay performance. Although sample fractionation or enrichment strategies can be applied to reduce complexity and non-specific background signals17,18, they also decrease the overall throughput and, more importantly, are not feasible when there is limited availability of clinical samples. Ion mobility spectrometry (IMS) can reduce complexity in the gas phase without having to fractionate samples in advance and has been applied to peptide quantitation by SRM19,20,21,22.
High-field asymmetric waveform IMS (FAIMS) is a commercially available technology that is distinct from other IMS techniques. In FAIMS, ions are transported by a flow of carrier gas through an asymmetrical electric field applied perpendicularly to the direction of travel23. Selection of a FAIMS DC compensation voltage (CV) allows continuous transmission of the target ion population, in contrast to the temporal separation of ions that characterizes other IMS approaches. This feature may be ideal for a targeted proteomics workflow, in which the focus is on a selected list of molecules. In this context, the addition of FAIMS may further increase the performance of PRM. To date, the application of FAIMS on an Orbitrap instrument for PRM has been described in a single publication, without extensive characterization of the benefit of FAIMS24. Here we present a systematic evaluation of a FAIMS-PRM method combining ion mobility and a high-resolution Orbitrap mass analyzer to achieve high-sensitivity quantitation of critical biomarkers in oncology clinical studies.
Materials and methods
Synthetic peptide preparation
All peptides were synthesized as light and heavy pairs by 21st Century Biochemicals (Marlborough, MA). Heavy peptides were labeled with C-terminal R [13C615N4] or K [13C615N2] with > 99% isotopic enrichment. For cysteines, carbamidomethylated cysteines were used. Synthesized peptides were further purified by reverse-phase HPLC, using a YMC Hydrosphere 20 × 150 mm C18 column, packed with 5 micron beads with 120 Å pore size, to attain > 95% purity, and net peptide content was verified by amino acid analysis.
Clinical sample preparation
FFPE tumor biopsy samples from breast cancer patients, collected at the Moscow Cancer Research Center with informed consent under Institutional Review Board approval (Russion Oncological Research Centre; Protocol/Study number PG-ONC 2003/1) by certified medical pathologists, were purchased from ProteoGenex (Inglewood, CA). All use of these samples was in accordance with local guidelines and regulations. Three tissue sections per sample were generated for hematoxylin and eosin (H&E) staining, HER2 IHC, and laser microdissection (LMD).
Image analysis and LMD
Pathology evaluation for laser-microdissected tumor epithelium was conducted by the study pathologist on H&E-stained slides, which were digitally imaged with an Aperio ScanScope AT scanner (Leica Microsystems, Wetzlar, Germany). HALO AI (Indica Labs, Albuquerque, NM) was used to classify and annotate the slides to guide LMD of tumor epithelium (Supplemental Materials and Methods).
Sample processing for LC–MS analysis
Microdissected tissue samples collected from the slides were dissolved in 0.1% RapiGest, incubated at 95 °C for 90 min, alkylated with chloroacetamide at 37 °C, prior to overnight trypsin digestion (Supplemental Materials and Methods).
LC–MS data acquisition
Desalted samples (1.2 µg) were combined with synthetic isotope-labelled peptides (6 fmol). Five-sixths of this mixture (1 µg of total peptide and 5 fmol of each synthetic peptide) was loaded onto EvoTip trapping columns before separation with the EvoSep One nanoLC system (EvoSep, Odense, Denmark) coupled to an Orbitrap Fusion Lumos mass spectrometer with a FAIMS-PRO interface (Thermo Fisher). Peptides were eluted over a 44-min gradient, from 7 to 30% acetonitrile (on-column), at a flow rate of 500 nL/min (Supplemental Materials and Methods). The FAIMS-PRM experiment employed higher-energy collisional dissociation (HCD) fragmentation with an isolation window of 0.7 mass-to-charge ratio (m/z), a target automatic gain control of 1E6 ions, and a maximum injection time of 100 ms. Tandem MS (MS/MS) scans were acquired in centroid mode with the Orbitrap detector, using 30 K resolution at 200 m/z (unless otherwise stated it the text). FAIMS was operated at the standard resolution, with no additional FAIMS gas.
Data analysis
PRM data was analyzed with Skyline (University of Washington, Seattle, WA)25, using high-selectivity extraction. Fragment ions with interference were identified by manual analysis, comparing coelution and fragment ion ratios between endogenous and reference peptides. Any fragment ions showing interference were flagged and omitted from use in quantitation (Supplemental Materials and Methods).
Ion injection numbers were extracted from raw files as RawOvFtT, using rawDiag26.
HER2 IHC assay
FFPE tissue sections were stained with the Ventana Pathway anti-Her2/neu (4B5) assay according to the manufacturer’s instructions on a Ventana Benchmark Ultra staining device (Ventana Medical, Oro Valley, AZ).
Results
Integration of FAIMS into the PRM method
In this study, we selected 10 peptides representing HER2, EGFR, ER, cMET, and KRAS, which are primary targets of oncology drug development and actionable biomarkers in the clinic5,8,9,27. To determine the optimal FAIMS CV that transmits the maximum ion flux, a range of CVs from 0 to – 100 V was scanned per precursor, with direct infusion of synthetic peptides. For example, the optimal transmission of the EGFR IPLENLQIIR [M + 2H]2 + precursor was between – 62 and – 58 V, with a half-maximal intensity of – 72 to – 52 V (Fig. 1A). The optimal CV value determined per precursor was integrated into the final scheduled PRM method. This allowed the FAIMS unit to apply the optimal CV values for the targets at the scheduled acquisition time of the individual peptide, according to the PRM method. The final FAIMS-PRM method for these peptides is summarized in Table 1. FAIMS CVs for the 10 peptides ranged from – 28 to – 58 V. For each target the optimal collision-induced dissociation energy was also determined (Table 1). Additional optimization of the precursor isolation width and MS/MS resolution was carried out. Precursor isolation widths of 0.4, 0.7 and 1.6 m/z were compared: 0.7 m/z was selected as the preferred balance between reducing interference, shown in Supplementary Fig. S1A for the HER2 ELVSEFSR peptide, and maintaining signal of the target peptide (Supp. Fig. S1B). MS/MS resolutions of 30 K (64 ms transient length), 60 K (128 ms) and 120 K (256 ms) were compared: the additional scan time required for higher resolution data acquisition did not appear to give sufficient reduction in interference to be warranted. Higher resolution acquisition also resulted in lower fragment ion signal (Supp. Fig. S2), and therefore the 30 K resolution was preferred. For the four peptides shown in Supplementary Fig. S2, the difference in summed area between the 30 K resolution and 120 K resolution ranges from 1 to 29%, with the median percentage drop being 22%. The lower signal with higher resolution acquisition may be related to the longer measurement time in the Orbitrap analyzer, allowing signal decay to occur, for instance due to collisions with background gas28.
Performance of FAIMS-PRM
To assess the impact of FAIMS integration into the PRM method, the limits of quantitation (LOQs) of the PRM assays obtained with and without FAIMS were determined in trypsin-digested, formalin-fixed human spleen. Because the target proteins were present in the spleen matrix, a reverse-dilution approach was chosen, with spiking-in of a constant 5 fmol of light peptide and a 10-point dilution series of heavy peptide ranging from 100 fmol to 5 amol. Blank (light peptides plus matrix) and double-blank (matrix only) samples were included in this set.
An LOQ curve was acquired on two separate occasions, with and without FAIMS, for a total of four LOQ curves and 192 raw files. The quantitative range, spanning from lower to upper limits of quantitation (LLOQ to ULOQ), was defined as the points on the curve with CV values of < 20%, mean accuracy within 80–120%, and mean fragment ion ratio within 70–130% of the coeluting reference fragment ratio (Supplemental Fig. 3). Table 2 shows the lowest LLOQ with and without FAIMS from the replicate curves. The full list of LOQs, including ULOQ and range, is shown in Supplemental Table 1. Of 11 precursors, 7 showed a threefold reduction in LLOQ with the addition of FAIMs. For all precursors, the PRM experiments without FAIMS had a median LLOQ of 137 amol/µg; with FAIMS, this was reduced to 46 amol/µg. In parallel, we calculated the limits of quantitation for additional fragment ions (71 in total; ranging from four to ten per precursor), and also for summing all fragment ions signals for each peptide (Supplementary Tables S2 and S3). Summing all fragment ions gave the same result as picking the best fragment ion: the same seven precursors had the lowest LLOQ with FAIMS, while the other four precursors were tied. For the complete set of fragment ions, 37 had the lowest LLOQ with FAIMS, 30 were tied and 4 had the lowest LLOQ without FAIMS. The majority of the ties (22 of 30) were for the same four precursors which did not have an overall improved LLOQ.
In the case of the EGFR IPLENLQIIR peptide, the lowest LLOQ obtained by PRM analysis without FAIMS was 137 amol (Table 2). This value was improved to 46 amol with FAIMS-PRM. MS/MS spectra of the 46-amol IPLENLQIIR peptide spiked into the spleen matrix and analyzed by FAIMS-PRM vs. PRM are shown in Fig. 1C and D. At this low level, the PRM MS/MS data showed a higher total ion count (3.9 E5), including interfering background ions, but FAIMS-PRM MS/MS demonstrated a cleaner spectrum, with a fourfold-lower total ion count (9.9 E4) and higher relative intensities of the fragment ions derived from the IPLENLQIIR precursor.
Extracted ion chromatograms (XICs) of fragment ions from the 46-amol IPLENLQIIR peptide analyzed by PRM vs. FAIMS-PRM are shown in Fig. 1E and F. Interferences detected in PRM, even with high-resolution (30 K) Orbitrap detection (Fig. 1F), were removed when PRM was acquired with FAIMS (Fig. 1E).
We hypothesized that the peptides for which FAIMS reduced the LLOQ in this study were more likely to elute in a region of high matrix complexity with higher levels of non-specific background signal. These peptides would benefit from FAIMS to a greater extent after the background ions were filtered out. The format of our LOQ curve allows us to accurately quantify non-specific background, as this constitutes the only signal in the heavy channel for the LOQ blank injections, with the appropriate elution window defined by the light synthetic peptide. The average number of non-specific background ions injected for each targeted m/z window are shown in Fig. 2A. Background ions for all targeted m/z windows decreased with FAIMS (Fig. 2B). However, the seven peptides showing a reduction in LLOQ with FAIMS integration demonstrated a significantly greater decrease in background ions with FAIMS. This was shown by a two-sided t test comparing percent background reduction for peptides having a lower LLOQ with peptides having an unchanged LLOQ (P < 0.001; mean reduction of 46%; 95% confidence interval 32.1 to 60.6%).
The impact of FAIMS on total ions injected across the LOQ curve is shown in Supplemental Fig. S4 for the EGFR IPLENLQIIR [M + 2H]2+ precursor. The contribution of the spiked heavy peptide was negligible up to 137 amol. From 412 amol, the median ion injection number increased with additional heavy peptide with FAIMS. Without FAIMS, the median ion injection number only noticeably increased from addition of 3.7 fmol heavy peptide, due to the increased background.
Application of FAIMS-PRM in clinical samples
To demonstrate the performance of our FAIMS-PRM assay on clinical samples, we applied the method to 20 FFPE tumor biopsies from breast cancer patients. Tumor epithelial cells identified by AI-based image analysis were collected by LMD for precise measurement of tumoral concentrations. The overall workflow is described in Fig. 3A.
All targeted proteins were successfully quantified from the 20 tumor biopsy samples (Supplemental Table S4, Supplemental Fig. S5. Examples of FAIMS-PRM–based quantitation for EGFR and HER2 are shown in Fig. 3B and C. EGFR quantitation from two samples with levels of 244 and 62 amol/µg is shown in Fig. 3B. Even at the 62-amol level, the low level of non-specific background signal surrounding the peaks of interest is apparent. Quantitation of HER2 for two samples with levels of 5718 and 210 amol/µg is shown in Fig. 3C. HER2 levels in these samples were quantified using the two best-performing peptides, ELVSEFSR and SGGGDLTLGLEPSEEEAPR, which both showed improved LLOQs with FAIMS integration. The HER2 quantitation results obtained independently from these two peptides were highly correlated (R2 = 0.991) (Fig. 4A).
We also performed IHC analysis for these 20 samples, using the Ventana Her2 (4B5) assay. The pathologist’s scoring results for HER2 expression based on this IHC assay were compared with the tumoral concentrations acquired by FAIMS-PRM. Although the PRM readouts correlated well with the IHC scores, in agreement with earlier publications, the MS quantitation highlighted a wide range of tumoral concentrations within the same IHC scores (Fig. 4B). The outliers within an IHC class may identify therapeutic opportunities. E.g. within the eight samples classified as 1 + by IHC, two samples have a HER2 concentration close to the median concentration of the HER2 2 + samples. These patients may be suitable candidates for a HER2-targeted therapy, with a positive FISH test.
Discussion
Throughout the evolution of targeted proteomics, as in all other analytical methodologies, improved sensitivity has been achieved through limiting noise and increasing signal. The original SRM method continuously transmits ions only within both the precursor and fragment m/z windows, thereby limiting noise from other species and increasing signal 29. In practice, multiple transitions per peptide must be monitored to increase confidence in identification and to mitigate potential interference from variable matrix components. However, each additional SRM transition reduces the transmission time per transition and therefore the overall sensitivity of the method. In the PRM method, fragment ion selection is carried out after acquisition from the complete MS/MS data set, and more stringent separation of signal from non-specific background is possible when relying on the high-resolution data. To further increase sensitivity, and if sufficient sample is available (typically at least 10 × more than for direct analysis), front-end fractionation or immunoprecipitation can be added into the workflow17,18. For clinical samples with limited material, incorporation of a gas-phase ion mobility precursor selection step is an alternative approach to enhance sensitivity. We integrated a FAIMS step into the scheduled PRM method to determine whether this integration would improve the quantitation performance for oncology targets in a clinical setting.
The benefit of FAIMS with the EGFR peptide IPLENLQIIR is shown in Fig. 1. This peptide showed a dramatic reduction in non-specific background after addition of FAIMS. The effect was evident in the MSMS spectra (Fig. 1C vs. D) and within the target fragment extracted ion chromatograms (Fig. 1E vs. F). Lowering the level of non-specific background peptides proximal to the target precursor ion led to detection of more fragment ions with more consistent scan-to-scan signal and consequently higher-quality quantitation.
In the PRM approach with an Orbitrap analyzer, sensitivity can be modulated by ion injection time; longer injection times of up to 500 ms have been shown to improve signal-to-noise ratio by collecting more ions in the C-trap30. As ion injection time is increased, however, throughput is decreased. In our study, we employed a maximum ion injection time of 100 ms per precursor, allowing up to 15 co-eluting peptide pairs to be quantified with six points across a 20-s chromatographic peak width. Dynamic scheduling, based on either elution markers or reference peptide detection, may allow more time to be spent per peptide, further boosting sensitivity12,31. We note that our 100 ms injection time permits the acquisition of higher resolution data (50 K versus 30 K), without a significant time penalty. This has been implemented successfully in later version of the method.
The addition of an ion mobility step to an SRM method has been described previously19,20,21. One report describes a method combining IMS with quadrupole time-of-flight fragment ion detection for the quantitation of host cell proteins in protein biopharmaceutical products22; in this case the IMS was integral to the mass spectrometer, prohibiting a comparison of the same method with and without ion mobility. The FAIMS unit employed in our work has been previously deployed in global proteomic studies, with either label-free or stable-isotope labeling quantitation; to our knowledge, however, its performance in targeted proteomics has not been assessed32,33,34,35,36. Our analysis allowed a direct assessment of the contribution of FAIMS to the performance of the PRM method. Figure 2A shows the variable level of background ions detected within the different precursor windows, quantified as ions per scan, in a human tissue sample. The target of 1E6 ions usually was not reached within the 100-ms fill time. Without FAIMS, the median background level varied by 24-fold, from the highest level of 5.8 E5 ions per scan for the ELVSEFSR (2 +) region to the lowest level of 2.4 E4 for the FVVIQNEDLGPASPLDSTFYR (2 +) region. With FAIMS, the median background ions per scan was reduced for all 11 precursors, and the degree of reduction ranged from 13 to 86%. The targets with a FAIMS-dependent reduction in LLOQ also showed the greatest FAIMS-dependent reduction in background signal, with a reduction of at least 48.5% (Fig. 2B). We measured the background ions, with and without FAIMS, for the heavy isolation window in a human spleen matrix; levels of background for the light window and with other complex matrices will differ slightly, but we expect the same trend of reduced background with the addition of FAIMS.
We selected five protein targets, HER2, EGFR, cMET, KRAS, and ER, as real-world examples of clinically relevant biomarkers requiring quantitation from small amounts of clinical material—in this case a single 10-µm-thick FFPE section. The tumor areas selected for LMD ranged from 4.6 to 14.1 mm2, highlighting the low sample requirements for this sensitive approach. Expression levels for most of these targets varied widely (Supplemental Fig. S5). For example, HER2 levels in our samples spanned a 30-fold range, from 200 to 6000 amol/µg (Fig. 4). FAIMS-PRM increased the quantitation sensitivity of the clinically relevant biomarker HER2, reducing the LLOQ from 137 to 46 amol/µg. This compares favorably with reported LLOQs from SRM assays of 125 amol/µg37, 5 fmol/μg38, 155 amol/μg39, and 83 amol/µg40, although these assays employed different matrices and LLOQ calculation methods.
Recently, a new anti-HER2 ADC, trastuzumab-deruxtecan (Enhertu; AstraZeneca) showed antitumor activity not only in patients with HER2-positive cancer (HER2 3+ and HER2 2+/FISH+) but also in low-HER2–expressing tumors (HER2 1+ and HER2 2+) that were previously categorized as HER2 negative and not eligible for anti-HER2 monoclonal antibody treatment7. Importantly, as shown in Fig. 4B, HER2 in breast cancer samples still exhibited a range of expression levels in these low-expression samples, and all of them were above the LLOQ of our FAIMS-PRM HER2 assay. Previous studies have shown similar general agreement between targeted MS and IHC HER2 measurements, while also identifying a subset of samples with HER2 levels outside the expected range based on the IHC classification37,38. Although the tumoral HER2 concentrations used in the box plots are based on the ELVSEFSR peptide, a high correlation (R2 = 0.991) was obtained between the assays with two different peptides (ELVSEFSR and SGGGDLTLGLEPSEEEAPR), consistent with the expected quantitation accuracy of our clinical proteomics workflow and FAIMS-PRM assay.
The future development of treatments intended to benefit patients with low HER2 expression will place an increased emphasis on lower-level HER2 quantitation accuracy and robustness. Likewise, the FAIMS-PRM assays for other potential ADC targets such as EGFR and cMET, for which LLOQs were improved (Table 2), will aid investigations of therapeutic benefits for tumors with low target expression. In our study, the addition of FAIMS to the PRM method reduced the LLOQ for most of the peptides evaluated. This increased sensitivity was obtained without additional analysis time or sample material. The clinical proteomics workflow for FFPE tissue samples presented here is anticipated to be of great utility for clinical trials in which multiplexing of protein targets allows proof-of-mechanism and pharmacodynamic questions to be addressed.
Abbreviations
- ADC:
-
Antibody-drug conjugate
- AI:
-
Artificial intelligence
- CDx:
-
Companion diagnostic
- CV:
-
Compensation voltage
- FAIMS:
-
High-field asymmetric waveform ion mobility spectrometry
- FFPE:
-
Formalin-fixed: paraffin-embedded
- H&E:
-
Hematoxylin and eosin
- HCD:
-
Higher-energy collisional dissociation
- IHC:
-
Immunohistochemistry
- IMS:
-
Ion mobility spectrometry
- LC:
-
Liquid chromatography
- LLOQ:
-
Lower limit of quantitation
- LMS:
-
Laser microdissection
- LOQ:
-
Limit of quantitation
- MS:
-
Mass spectrometry
- MS/MS:
-
Tandem mass spectrometry
- PRM:
-
Parallel reaction monitoring
- SRM:
-
Selected reaction monitoring
- ULOQ:
-
Upper limit of quantitation
- XIC:
-
Extracted ion chromatogram
References
Vargas, A. J. & Harris, C. C. Biomarker development in the precision medicine era: Lung cancer as a case study. Nat. Rev. Cancer 16(8), 525–537 (2016).
U.S. Food and Drug Administration List of cleared or approved companion diagnostic devices (in vitro and imaging tools). https://www.fda.gov/media/119249/download (March 24, 2021),
Scheerens, H. et al. Current status of companion and complementary diagnostics: Strategic considerations for development and launch. Clin. Transl. Sci. 10(2), 84–92 (2017).
Vogel, C. L. et al. Efficacy and safety of trastuzumab as a single agent in first-line treatment of HER2-overexpressing metastatic breast cancer. J. Clin. Oncol. 20(3), 719–726 (2002).
Bang, Y. J. et al. Trastuzumab in combination with chemotherapy versus chemotherapy alone for treatment of HER2-positive advanced gastric or gastro-oesophageal junction cancer (ToGA): A phase 3, open-label, randomised controlled trial. Lancet 376(9742), 687–697 (2010).
An, E. et al. Quantitative proteomic analysis of HER2 expression in the selection of gastric cancer patients for trastuzumab treatment. Ann. Oncol. 28(1), 110–115 (2017).
Modi, S. et al. Antitumor activity and safety of trastuzumab deruxtecan in patients with HER2-low–expressing advanced breast cancer: Results from a phase Ib study. J. Clin. Oncol. 38(17), 1887–1896 (2020).
Nardone, A. et al. The oral selective oestrogen receptor degrader (SERD) AZD9496 is comparable to fulvestrant in antagonising ER and circumventing endocrine resistance. Br. J. Cancer 120(3), 331–339 (2019).
Janes, M. R. et al. Targeting KRAS mutant cancers with a covalent G12C-specific inhibitor. Cell 172(3), 578-589.e17 (2018).
Nuciforo, P. et al. High HER2 protein levels correlate with increased survival in breast cancer patients treated with anti-HER2 therapy. Mol. Oncol. 10(1), 138–147 (2016).
Gallien, S., Bourmaud, A., Kim, S. Y. & Domon, B. Technical considerations for large-scale parallel reaction monitoring analysis. J. Proteomics 100, 147–159 (2014).
Gallien, S., Kim, S. Y. & Domon, B. Large-scale targeted proteomics using internal standard triggered-parallel reaction monitoring (IS-PRM). Mol. Cell Proteomics 14(6), 1630–1644 (2015).
Gallien, S. et al. Targeted proteomic quantification on quadrupole-Orbitrap mass spectrometer. Mol. Cell Proteomics 11(12), 1709–1723 (2012).
Cifani, P. & Kentsis, A. High sensitivity quantitative proteomics using automated multidimensional nano-flow chromatography and accumulated ion monitoring on quadrupole-Orbitrap-linear ion trap mass spectrometer. Mol. Cell Proteomics 16(11), 2006–2016 (2017).
Kim, Y. J., Gallien, S., van Oostrum, J. & Domon, B. Targeted proteomics strategy applied to biomarker evaluation. Proteomics Clin. Appl. 7(11–12), 739–747 (2013).
Peterson, A. C., Russell, J. D., Bailey, D. J., Westphall, M. S. & Coon, J. J. Parallel reaction monitoring for high resolution and high mass accuracy quantitative, targeted proteomics. Mol. Cell Proteomics 11(11), 1475–1488 (2012).
Morales-Betanzos, C. A. et al. Quantitative mass spectrometry analysis of PD-L1 protein expression, N-glycosylation and expression stoichiometry with PD-1 and PD-L2 in human melanoma. Mol. Cell Proteomics 16(10), 1705–1717 (2017).
Do, M. et al. Clinical application of multiple reaction monitoring-mass spectrometry to human epidermal growth factor receptor 2 measurements as a potential diagnostic tool for breast cancer therapy. Clin. Chem. 66(10), 1339–1348 (2020).
Klaassen, T., Szwandt, S., Kapron, J. T. & Roemer, A. Validated quantitation method for a peptide in rat serum using liquid chromatography/high-field asymmetric waveform ion mobility spectrometry. Rapid Commun. Mass Spectrom. 23(15), 2301–2306 (2009).
Xia, Y.-Q., Wu, S. T. & Jemal, M. LC-FAIMS-MS/MS for quantification of a peptide in plasma and evaluation of FAIMS global selectivity from plasma components. Anal. Chem. 80(18), 7137–7143 (2008).
Norman, R. L., Singh, R., Langridge, J. I., Ng, L. L. & Jones, D. J. L. The measurement of KRAS G12 mutants using multiplexed selected reaction monitoring and ion mobility mass spectrometry. Rapid Commun. Mass Spectrom. 34(S4), e8657 (2020).
Doneanu, C. et al. An HS-MRM assay for the quantification of host-cell proteins in protein biopharmaceuticals by liquid chromatography ion mobility QTOF mass spectrometry. J. Vis. Exp. 134, e55325 (2018).
Cooper, H. J. To what extent is FAIMS beneficial in the analysis of proteins?. J. Am. Soc. Mass Spectrom. 27(4), 566–577 (2016).
Renuse, S. et al. A mass spectrometry-based targeted assay for detection of SARS-CoV-2 antigen from clinical specimens. EBioMedicine 69, 103465 (2021).
MacLean, B. et al. Skyline: An open source document editor for creating and analyzing targeted proteomics experiments. Bioinformatics 26(7), 966–968 (2010).
Trachsel, C. et al. rawDiag: An R package supporting rational LC–MS method optimization for bottom-up proteomics. J. Proteome Res. 17(8), 2908–2914 (2018).
Sellmann, C. et al. Balancing selectivity and efficacy of bispecific epidermal growth factor receptor (EGFR) × c-MET antibodies and antibody-drug conjugates. J. Biol. Chem. 291(48), 25106–25119 (2016).
Perry, R. H., Cooks, R. G. & Noll, R. J. Orbitrap mass spectrometry: Instrumentation, ion motion and applications. Mass Spectrom. Rev. 27(6), 661–699 (2008).
Picotti, P. & Aebersold, R. Selected reaction monitoring–based proteomics: Workflows, potential, pitfalls and future directions. Nat. Methods 9, 555–566 (2012).
Nguyen, C. D. L. et al. sensitive and simple targeted proteomics approach to quantify transcription factor and membrane proteins of the unfolded protein response pathway in glioblastoma cells. Sci. Rep. 9(1), 8836 (2019).
Remes, P. M., Yip, P. & MacCoss, M. J. Highly multiplex targeted proteomics enabled by real-time chromatographic alignment. Anal. Chem. 92(17), 11809–11817 (2020).
Pfammatter, S. et al. A novel differential ion mobility device expands the depth of proteome coverage and the sensitivity of multiplex proteomic measurements. Mol. Cell. Proteomics 17(10), 2051–2067 (2018).
Bekker-Jensen, D. B., del Val, A. M., Steigerwald, S., Rüther, P., Fort, K., Arrey, T. N., Harder, A., Makarov, A., Olsen, J. V. A compact quadrupole-Orbitrap mass spectrometer with FAIMS interface improves proteome coverage in short LC gradients. Preprint at http://biorxiv.org/content/early/2019/11/30/860643.abstract.
Hebert, A. S. et al. Comprehensive single-shot proteomics with FAIMS on a hybrid Orbitrap mass spectrometer. Anal. Chem. 90(15), 9529–9537 (2018).
Pfammatter, S., Bonneil, E., McManus, F. P. & Thibault, P. Accurate quantitative proteomic analyses using metabolic labeling and high field asymmetric waveform ion mobility spectrometry (FAIMS). J. Proteome. Res. 18(5), 2129–2138 (2019).
Schweppe, D. K. et al. Characterization and optimization of multiplexed quantitative analyses using high-field asymmetric-waveform ion mobility mass spectrometry. Anal. Chem. 91(6), 4010–4016 (2019).
Hembrough, T. et al. Application of selected reaction monitoring for multiplex quantification of clinically validated biomarkers in formalin-fixed, paraffin-embedded tumor tissue. J. Mol. Diagn. 15(4), 454–465 (2013).
Guerin, M. et al. Development of parallel reaction monitoring (PRM)-based quantitative proteomics applied to HER2-Positive breast cancer. Oncotarget 9(73), 33762–33777 (2018).
Steiner, C. et al. Quantification of HER2 by targeted mass spectrometry in formalin-fixed paraffin-embedded (FFPE) breast cancer tissues. Mol. Cell Proteomics 14(10), 2786–2799 (2015).
Kennedy, J. J. et al. Demonstrating the feasibility of large-scale development of standardized assays to quantify human proteins. Nat Methods 11, 149 (2013).
Acknowledgements
We thank Deborah Schuman for proof-reading the manuscript.
Funding
This study was funded by AstraZeneca.
Author information
Authors and Affiliations
Contributions
All authors confirmed they have contributed to the intellectual content of this paper and have met the following 4 requirements: (a) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of data; (b) drafting or revising the article for intellectual content; (c) final approval of the published article; and (d) agreement to be accountable for all aspects of the article thus ensuring that questions related to the accuracy or integrity of any part of the article are appropriately investigated and resolved.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Sweet, S., Chain, D., Yu, W. et al. The addition of FAIMS increases targeted proteomics sensitivity from FFPE tumor biopsies. Sci Rep 12, 13876 (2022). https://doi.org/10.1038/s41598-022-16358-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-022-16358-1
This article is cited by
-
Recent developments in mass-spectrometry-based targeted proteomics of clinical cancer biomarkers
Clinical Proteomics (2024)
-
Mass spectrometry quantifies target engagement for a KRASG12C inhibitor in FFPE tumor tissue
Clinical Proteomics (2023)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.