Abstract
This work explores the possibility of depth profiling of inorganic materials with Megaelectron Volt Secondary Ion Mass Spectrometry using low energy primary ions (LE MeV SIMS), specifically 555 keV Cu2+, while etching the surface with 1 keV Ar+ ions. This is demonstrated on a dual-layer sample consisting of 50 nm Cr layer deposited on 150 nm In2O5Sn (ITO) glass. These materials proved to have sufficient secondary ion yield in previous studies using copper ions with energies of several hundred keV. LE MeV SIMS and keV SIMS depth profiles of Cr-ITO dual-layer are compared and corroborated by atomic force microscopy (AFM) and time-of-flight elastic recoil detection analysis (TOF-ERDA). The results show the potential of LE MeV SIMS depth profiling of inorganic multilayer systems in accelerator facilities equipped with MeV SIMS setup and a fairly simple sputtering source.
Similar content being viewed by others
Introduction
Secondary Ion Mass Spectrometry using MeV ions (MeV SIMS)1 is a fairly new Ion Beam Analysis (IBA) technique that is being increasingly used for the analysis and imaging of organic materials in various fields, such as forensics (fingerprints2 and inks3,4,5), cultural heritage (paints6), biology (plants7 and tissues8), etc. Conventional keV SIMS, on the other hand, is a well-established technique used mainly in the analysis and depth profiling of inorganic materials, with the most popular application in the semiconductor industry, i.e. studies of dopant profiles9,10, diffusion, corrosion11, contamination12, etc. It is also convenient for analyzing biomolecules, but with decreased ionization efficiency compared to MeV SIMS1 and keV cluster SIMS13,14. Techniques developed in order to enhance the ionization efficiency involve the use of high energy gas cluster ion beams (GCIBs) such as C60+ and Arx (x = 500–5000)13,14 and recently water cluster ions (H2O)n+ (n = 1−10 000)15,16, which have proven superior over former.
KeV SIMS is a surface-sensitive technique (ions are detected from a few uppermost monolayers) and can be extended to depth profile analysis by introducing ion sputtering. A variety of applications in SIMS depth profiling arise from high sensitivity to inorganic species and excellent depth resolution17,18. Recently, depth profiling of organic films has proven promising by employing either cluster ion beams or low energy Cs+ ions for sputtering, due to reduced surface degradation19. Depth resolution, a quantitative measure of the depth range, is by convention the sputtered depth measured between 84 and 16% of the maximum yield at an ideally sharp interface between two media20. Several different parameters contribute to the profile broadening that originate from instrumental factors, ion beam-sample interactions, and sample characteristics. These include surface roughening caused by sputtering21, atomic mixing in the collision cascade22, information depth of the technique23, non-uniform etching ion beam density in the analyzed area24, differential sputtering due to crystalline/amorphous regions25, etc. For thicker metallic layers, sputtering-induced roughness is generally the dominating contribution to depth resolution26. The best resolution in SIMS (below 5 nm) is achieved with low sputtering ion beam energy27 (below 1 keV) and high incidence angle, in combination with sample rotation in order to minimize sputtering-induced roughness28. At this point, atomic mixing and information depth are of increasing importance for depth resolution26.
On the other hand, modification of the original element distribution due to matrix effects poses a significant limitation in quantitative SIMS depth profiling. This phenomenon arises from the secondary ion yield dependence on the surrounding chemical state (of the matrix). The matrix effect depends largely on experimental conditions, namely the nature of the primary ion, the incident angle, the detected species, and the energy of secondary ions29. In data analysis, this is usually tackled with the use of reference materials containing similar matrix10,30,31. Many studies have investigated the parameters involved in matrix effects and possible methods for reducing them32,33.
MeV SIMS is also surface-sensitive and can in theory be extended to depth profiling. The fundamental difference between keV and MeV SIMS lies in the interaction mechanism of the primary ions with the material. While keV SIMS operates with energies of a few tens of keV through direct energy transfer to the secondary ions via nuclear stopping, MeV SIMS uses heavy energetic ions of a few tens of MeV that interact through electronic stopping with the target’s electronic system. It was shown in our previous work34 that lowering the primary ion beam energy from a standardly used few MeV to a few hundred keV enhances the efficiency of secondary ion detection for some inorganic species. Their detection was successful at all three primary ion beam energies used (200 keV Cu2+, 440 keV Cu2+, and standardly used 5 MeV Si4+), exhibiting the expected behavior of secondary ion yield with respect to the primary ion velocity: their yields decreasing with increasing velocity, i.e., decreasing nuclear stopping in the inorganic material, which is a driving force in collisional sputtering. This energy mode was named LE (Low Energy) MeV SIMS. In the present work, LE MeV SIMS depth profiling was explored based on previously confirmed good sensitivity to several inorganic species, in order to compare the achieved depth resolution against keV SIMS, as well as to observe the magnitude of eventual matrix effects, given that specific primary ion beam conditions are introduced in LE MeV SIMS. Up until now, the authors have found no record of an attempt of a dual-beam MeV SIMS depth profiling of inorganic matter in the literature.
This first demonstration of MeV SIMS depth profiling of an inorganic target at low primary ion beam energy opens up possibilities in both fundamental understanding of the impact of the primary ion characteristics on secondary ion yield of inorganic species, as well as expanding the application of MeV SIMS and other types of instruments, such as ion implanters, to perform mass spectrometry of inorganic species within their technical limits. This also presents a novelty for Ion Beam Analysis (IBA) laboratories in terms of the possibility to expand (and possibly improve) the set of existing techniques capable of depth profiling.
Experimental and methods
A dual-layer Cr-ITO sample was prepared by magnetron sputtering of roughly 50 nm Cr on top of 150 nm ITO (In2O5Sn), deposited on a soda-lime glass substrate. The sample was first characterized by Time-of-Flight Elastic Recoil Detection Analysis (TOF-ERDA) to determine layers’ atomic content and layer thicknesses. TOF-ERDA measurement was performed by 23 MeV 127I7+ ions with 20° incidence angle toward the sample surface. TOF-ERDA spectrometer was positioned at an angle of 37.5° toward the beam direction. More details about the used TOF-ERDA spectrometer can be found in the Ref.35,36.
The sample was also examined by atomic force microscopy (AFM) in tapping mode at several different places over areas of 2 × 2 μm2 to 10 × 10 μm2 before and after sputtering, in order to obtain RMS roughness.
KeV TOF SIMS depth profiling was performed on the dual-layer Cr-ITO sample on a TOF.SIMS 5 instrument produced by ION TOF, Germany, at Jožef Stefan Institute in Ljubljana, Slovenia, in a dual-beam mode using 30 keV Bi+ analysis ion beam with ion current of 2 pA and 1 keV Ar+ etching ion beam with ion current of 127 nA. The analysis was performed over an area of 100 × 100 μm2, while etching was done over 400 × 400 μm2. The base pressure in the chamber was 5·10–9 mbar. Hydrogen was introduced in the analysis chamber at pressure 7·10–7 mbar to reduce matrix effects32. Mass spectra were obtained in the positive secondary ion mode.
LE MeV SIMS depth profiling was performed on the dual-layer Cr-ITO sample on an in-house MeV TOF SIMS setup described elsewhere37. The chosen analysis beam was 555 keV Cu2+, having currents in the range of 1–5 fA in pulsed primary ion mode. The analysis covered an area of 300 × 300 μm2, under vacuum pressure of 10–6–10–7 mbar. Sputtering was carried out using PREVAC Ion Source IS 40C1, operating with 1 keV Ar+ ion beam and emission current of 10 mA. The ion source was mounted at an angle of 45° with respect to the surface normal. The sputtering beam spot size was approximately 1 × 1 cm2, as provided by the manufacturer. During ion sputtering, the sample holder was scanned in two directions in an attempt to homogenize sputtering, thus creating a sputtering area in the shape of a parallelogram with an unspecified area. Ar pressure was ranging from 6·10–5 to 3·10–4 mbar and the unsuppressed ion current measured on the target was ranging from 6 to 12 μA (changed between cycles). Depth profiling was conducted in a dual-beam mode, in cycles consisting of SIMS analysis of the sample surface by 555 keV Cu2+ and sputtering by 1 keV Ar+ for 10–20 min. Mass spectra were obtained in a positive secondary ion mode. While generating depth profiles, peak areas of selected secondary ions were normalized to the primary ion current and measurement time.
Results and discussion
Dual-layer Cr-ITO sample was first evaluated with TOF-ERDA depth profiling. The resulting depth profile of the main species constituting the sample is shown in Fig. 1. Analysis of TOF-ERDA spectra was performed by software Potku38. Despite the fact that depth resolution in ERDA deteriorates with depth due to multiple scattering, one can obtain the thickness of each layer by integrating the depth profile, or directly by measuring the distance between half of the maximum of the elemental curve at the layer surface (which marks the beginning of the layer) and at the interface (which marks the end of the layer). Cr and ITO layer thickness is calculated to be 44 ± 3 nm and 154 ± 9 nm, respectively (atomic density of Cr and ITO was considered in conversion from atoms/cm2 to nm). It should be noted that because of deteriorating depth resolution (which is roughly 22 nm at the Cr-ITO interface), ERDA has limited capability to resolve eventual structure in the depth profile at the layers’ interface. Since ERDA is a quantitative technique, elemental concentrations are provided in atomic percentages.
KeV SIMS profiles of selected species with respect to the sputter ion dose density (ions/cm2) are shown in Fig. 2. Cr-ITO interface is clearly resolved, and layers’ compositions are mostly uniform. Another ToF SIMS measurement (not shown here) using a 2 keV Cs+ etching beam observed the presence of increased CrO- at the surface of the Cr layer and interface with ITO. As a result, partial oxidation of Cr at the Cr-ITO interface causes a matrix effect by enhancing secondary ion yield of Cr+, as shown in Fig. 2. Many other oxides were detected as well, but with much lower efficiency.
LE MeV SIMS profiles of detected secondary ions from inorganic species are shown in Fig. 3. The x-axis is expressed in “quasi-dose” (cumulative sputtering ion current x sputtering time, per sputter cycle) since the sputter ion dose density could not be calculated due to the inability to precisely define the sputtering area. The main constituents of both layers and substrate (Cr+, In+, Sn+, Si+) are detected as positive ions, together with their isotopes with expected abundancies. In the case of Cr and In, the depth profile represents the sum of normalized peak areas of secondary ions of all detected isotopes. The profiles demonstrate significant chemical sensitivity to inorganic secondary ions. No oxides were detected in both positive and negative secondary ion mode, possibly due to lower efficiency to eject oxides with 555 keV Cu2+ primary ion beam, and/or due to the presence of a higher amount of background in the mass spectra while operating in the low energy mode. Also, there is generally a significantly lower total secondary ion yield in negative than in positive secondary ion mode. No cluster secondary ions were detected in positive secondary ion mode.
The satisfactory quality of LE MeV SIMS depth profile is reflected in the depth resolution at the Cr-ITO interface that is comparable to that in keV SIMS: the estimated difference in depth between 16 and 84% of the plateau is roughly 10 nm and 11 nm, for keV SIMS (Cr2+) and LE MeV SIMS (Cr+), respectively. This estimation was done by directly converting the x-axes to thickness (in nm) according to ERDA profiles. However, at the end of the dual-layer, the depth resolution in LE MeV SIMS worsens compared to keV SIMS, presumably due to Ar ion gun sputtering conditions during LE MeV SIMS analyses, which could potentially be improved. Another significant observation in LE MeV SIMS depth profile is the reduced matrix effect on Cr+ secondary ion yield at the Cr-ITO interface where keV SIMS indicated the presence of oxidized Cr. There is no sharp increase in secondary ion yield at the surface and at the end of the Cr layer as it is observed in the keV SIMS analyses for the Cr+ signal, instead the profile is rather steady. However, it is known that monoatomic metallic ions are more prone to matrix effect than cluster ions, hence, in keV SIMS, such layers are usually displayed through metallic dimers or trimers.
AFM of the pre-sputtered sample was performed in several segments. For virgin surface (Cr), an average RMS of 3.6 ± 0.2 nm was obtained. Virgin ITO surface before deposition of Cr layer revealed an average RMS of 3.8 ± 0.1 nm. Because the glass plate came with an already deposited ITO layer, the roughness of the virgin substrate (soda-lime glass) was measured from the substrate backside with an average RMS of 4.6 ± 0.5 nm. In the case of LE MEV SIMS, the crater in the post-sputtered sample revealed an average RMS of 8.1 ± 2.4 nm, whereas for keV SIMS, an average RMS of 9.1 ± 0.4 nm was obtained for the crater. The depth resolution degradation at the end of the ITO layer in LE MeV SIMS compared to keV SIMS is obviously not a consequence of roughness since the crater RMS is comparable to that of keV SIMS. However, it could be attributable to inhomogeneous Ar+ sputtering-induced non-horizontal crater bottom since Ar+ beam was not focused as in keV SIMS, and the analysis beam covered a fairly large area of 300 × 300 μm2.
Another difference between keV and LE MeV SIMS profiles concerns the ratio of the widths of Cr and ITO layer profiles. The width of each profile depends on the sputter yield of that element, given the energy and species of the etching ion beam (atoms/ion). Since profiles from both techniques were generated with the same type of etching beam (1 keV Ar+), one would expect a linear dependence between their x-axes. However, the vacuum conditions during sputtering were notably different (7·10–7 mbar for keV SIMS, and 6·10–5–3·10–4 mbar for LE MeV SIMS). There is evidence in the literature that the sputtering yield may be affected by background pressure39. Other than that, the authors have found no other explanation for this discrepancy.
Conclusions
Obtained LE MeV SIMS depth profiles of a dual-layer Cr-ITO sample demonstrate significant chemical sensitivity to inorganic secondary ions, as well as satisfactory depth resolution comparable to that of keV SIMS performed on the same type of sample using the same type of etching beam (1 keV Ar+). However, at the end of the dual-layer, depth resolution in LE MeV SIMS worsens compared to keV SIMS, but this is probably due to Ar sputtering conditions, which could potentially be improved by focusing the beam and defining a more uniform beam rastering. A notable revelation was a sign of majorly reduced matrix effect on Cr + secondary ion at the partially oxidized locations in Cr layer (surface and interface with ITO), compared to keV SIMS. This phenomenon is worth further exploring systematically.
One should note that LE MeV SIMS depth profiling is not as straightforward as on the commercial keV SIMS instruments, providing a significantly fewer number of points per sputter cycle, which can be time-consuming. All things considered, all of the observed limitations in LE MeV SIMS profiles seem to be a consequence of the sputtering rather than the analysis conditions. Overall, this work shows the benefit for other IBA laboratories that possess MeV SIMS instrument in expanding its application to inorganic samples by lowering the energy of the primary ion beam, thus gaining multiple orders of magnitude higher efficiencies of inorganic ions and obtaining depth profiles of intermediate-thickness samples with satisfactory depth resolution. Moreover, this can be realized using a relatively low-cost sputtering source.
In theory, this also paves the way for MeV SIMS depth profiling of hybrid organic/inorganic samples such as OLED screens or hybrid solar cells, at least in terms of the ability to detect secondary ions of both organic and inorganic species simultaneously in the low energy mode.
References
Nakata, Y. et al. Matrix-free high-resolution imaging mass spectrometry with high-energy ion projectiles. J. Mass Spectrom. 44, 128–136 (2009).
Bailey, M. J. et al. Depth profiling of fingerprint and ink signals by SIMS and MeV SIMS. Nucl. Instrum. Methods Phys. Res., Sect. B 268, 1929–1932 (2010).
Moore, K. L. et al. Determination of deposition order of toners, inkjet inks, and blue ballpoint pen combining Mev-secondary ion mass spectrometry and particle induced x-ray emission. Anal. Chem. 91, 12997–13005 (2019).
Malloy, M. C., Bogdanović Radović, I., Siketić, Z. & Jakšić, M. Determination of deposition order of blue ballpoint pen lines by MeV secondary ion mass spectrometry. Forensic Chem. 7, 75–80 (2018).
Barac, M. et al. Comparison of optical techniques and MeV SIMS in determining deposition order between optically distinguishable and indistinguishable inks from different writing tools. Forensic Sci. Int. 331, 111136 (2022).
Radović, I. B. et al. Identification and imaging of modern paints using secondary ion mass spectrometry with MeV ions. Nucl. Instrum. Methods Phys. Res., Sect. B 406, 296–301 (2017).
Jenčič, B. et al. Molecular imaging of cannabis leaf tissue with MeV-SIMS method. Nucl. Instrum. Methods Phys. Res., Sect. B 371, 205–210 (2016).
Jenčič, B. et al. MeV-SIMS TOF imaging of organic tissue with continuous primary beam. J. Am. Soc. Mass Spectrom. 30, 1801–1812 (2019).
Wei, X., Zhao, L., Wang, J., Zeng, Y. & Li, J. Characterization of nitride-based LED materials and devices using TOF-SIMS: Characterization of LED materials and devices using TOF-SIMS. Surf. Interface Anal. 46, 299–302 (2014).
Gong, B. & Marjo, C. E. Quantitative ToF-SIMS depth profiling of a multi-phased III-V semiconductor matrix via the analysis of secondary cluster ions: Quantitative TOF-SIMS depth profiling of III-V matrix. Surf. Interface Anal. 48, 422–427 (2016).
Wang, L. et al. Study of the surface oxides and corrosion behavior of an equiatomic CoCrFeMnNi high entropy alloy by XPS and ToF-SIMS. Corros. Sci. 167, 108507 (2020).
Mowat, I., Lindley, P. & McCaig, L. A correlation of TOF-SIMS and TXRF for the analysis of trace metal contamination on silicon and gallium arsenide. Appl. Surf. Sci. 203–204, 495–499 (2003).
Rabbani, S., Barber, A. M., Fletcher, J. S., Lockyer, N. P. & Vickerman, J. C. TOF-SIMS with argon gas cluster ion beams: A comparison with C60+. Anal. Chem. 83, 3793–3800 (2011).
Angerer, T. B., Blenkinsopp, P. & Fletcher, J. S. High energy gas cluster ions for organic and biological analysis by time-of-flight secondary ion mass spectrometry. Int. J. Mass Spectrom. 377, 591–598 (2015).
Sheraz née Rabbani, S. et al. Prospect of increasing secondary ion yields in ToF-SIMS using water cluster primary ion beams: Examining new primary ion sources for ToF-SIMS. Surf. Interface Anal. 46, 51–53 (2014).
Sheraz née Rabbani, S., Barber, A., Fletcher, J. S., Lockyer, N. P. & Vickerman, J. C. Enhancing secondary ion yields in time of flight-secondary ion mass spectrometry using water cluster primary beams. Anal. Chem. 85, 5654–5658 (2013).
Baryshev, S. V. et al. High-resolution secondary ion mass spectrometry depth profiling of nanolayers: Depth profiling of nanolayers by high-resolution SIMS. Rapid Commun. Mass Spectrom. 26, 2224–2230 (2012).
Chakraborty, P. Ultra-high depth resolution SIMS for the interface analysis of complex low-dimensional structures. Nucl. Instrum. Methods Phys. Res., Sect. B 266, 1858–1865 (2008).
Mouhib, T. et al. Molecular depth profiling of organic photovoltaic heterojunction layers by ToF-SIMS: comparative evaluation of three sputtering beams. Analyst 138, 6801 (2013).
ASTM. Standard Terminology Relating to Surface Analysis. American Society for Testing and Materials, Committee E‐42 on Surface Analysis, 4 (1992).
Yan, X. L., Duvenhage, M. M., Wang, J. Y., Swart, H. C. & Terblans, J. J. Evaluation of sputtering induced surface roughness development of Ni/Cu multilayers thin films by Time-of-Flight Secondary Ion Mass Spectrometry depth profiling with different energies O2+ ion bombardment. Thin Solid Films 669, 188–197 (2019).
Andersen, H. H. The depth resolution of sputter profiling. Appl. Phys. 18, 131–140 (1979).
Hofmann, S. & Schubert, J. Determination and application of the depth resolution function in sputter profiling with secondary ion mass spectroscopy and Auger electron spectroscopy. J. Vac. Sci. Technol., A: Vac., Surf. Films 16, 1096–1102 (1998).
Magee, C. W. & Honig, R. E. Depth profiling by SIMS?depth resolution, dynamic range and sensitivity. Surf. Interface Anal. 4, 35–41 (1982).
Graham, D. J., Wagner, M. S. & Castner, D. G. Information from complexity: Challenges of TOF-SIMS data interpretation. Appl. Surf. Sci. 252, 6860–6868 (2006).
Satori, K., Haga, Y., Minatoya, R., Aoki, M. & Kajiwara, K. Factors causing deterioration of depth resolution in Auger electron spectroscopy depth profiling of multilayered systems. J. Vac. Sci. Technol., A: Vac., Surf. Films 15, 478–484 (1997).
Hofmann, S. Compositional depth profiling by sputtering. Prog. Surf. Sci. 36, 35–87 (1991).
Zalar, A. Improved depth resolution by sample rotation during Auger electron spectroscopy depth profiling. Thin Solid Films 124, 223–230 (1985).
Saha, B. & Chakraborty, P. MCsn+-SIMS: An innovative approach for direct compositional analysis of materials without standards. Energy Procedia 41, 80–109 (2013).
Wilson, R. G. & Novak, S. W. Systematics of secondary-ion-mass spectrometry relative sensitivity factors versus electron affinity and ionization potential for a variety of matrices determined from implanted standards of more than 70 elements. J. Appl. Phys. 69, 466–474 (1991).
Zanderigo, F., Ferrari, S., Queirolo, G., Pello, C. & Borgini, M. Quantitative TOF-SIMS analysis of metal contamination on silicon wafers. Mater. Sci. Eng., B 73, 173–177 (2000).
Ekar, J., Panjan, P., Drev, S. & Kovač, J. ToF-SIMS depth profiling of metal, metal oxide, and alloy multilayers in atmospheres of H2, C2H2, CO, and O2. J. Am. Soc. Mass Spectrom. 33, 31–44 (2022).
Priebe, A., Xie, T., Bürki, G., Pethö, L. & Michler, J. The matrix effect in TOF-SIMS analysis of two-element inorganic thin films. J. Anal. At. Spectrom. 35, 1156–1166 (2020).
Barac, M., Brajković, M., Bogdanović Radović, I., Kovač, J. & Siketić, Z. MeV TOF SIMS analysis of hybrid organic/inorganic compounds in the low energy region. J. Am. Soc. Mass Spectrom. 32, 825–831 (2021).
Siketić, Z., Radović, I. B. & Jakšić, M. Development of a time-of-flight spectrometer at the Ruder Bošković Institute in Zagreb. Nucl. Instrum. Methods Phys. Res., Sect. B 266, 1328–1332 (2008).
Siketić, Z., Radović, I. B. & Jakšić, M. Quantitative analysis of hydrogen in thin films using time-of-flight elastic recoil detection analysis. Thin Solid Films 518, 2617–2622 (2010).
Tadić, T. et al. Development of a TOF SIMS setup at the Zagreb heavy ion microbeam facility. Nucl. Instrum. Methods Phys. Res., Sect. B 332, 234–237 (2014).
Arstila, K. et al. Potku – New analysis software for heavy ion elastic recoil detection analysis. Nucl. Instrum. Methods Phys. Res., Sect. B 331, 34–41 (2014).
Husinsky, W., Betz, G., Girgis, I., Viehböck, F. & Bay, H. L. Velocity distributions and sputtering yields of chromium atoms under argon, oxygen and carbon ion bombardment. J. Nucl. Mater. 128–129, 577–582 (1984).
Acknowledgements
Authors acknowledge support by the RADIATE project under the Grant Agreement 824096 from the EU Research and Innovation program HORIZON 2020 and by the European Regional Development Fund for the ‘Center of Excellence for Advanced Materials and Sensing Devices’ (Grant No. KK.01.1.1.01.0001), Ruđer Bošković Institute, Zagreb, Croatia. M. Brajković acknowledges support by the Croatian Science Foundation (CSF) project "Young Researchers’ Career Development Project—Training of Doctoral Students" co-financed by the European Union, Operational Program “Efficient Human Resources 2014-2020”.
Author information
Authors and Affiliations
Contributions
M.B. planned and optimized the experimental setup and performed LE MeV SIMS measurements together with Z.S. and M.B. M.B. prepared the manuscript text, as well as interpreted the results together with Z.S., J.K., and I.B.R. KeV SIMS depth profile measurements performed by J.E. and J.K. TOF-ERDA measurements and analysis performed by Z.S. AFM measurements performed by I.Š.R. All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Barac, M., Brajković, M., Siketić, Z. et al. Depth profiling of Cr-ITO dual-layer sample with secondary ion mass spectrometry using MeV ions in the low energy region. Sci Rep 12, 11611 (2022). https://doi.org/10.1038/s41598-022-16042-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-022-16042-4
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.