Abstract
New recommendations on evaluation of peritoneal membrane function suggest ruling out catheter dysfunction when evaluating patients with low ultrafiltration capacity. We introduce the use of a combination of parameters obtained from the cycler software PD Link with HomeChoicePro (Baxter International Inc., Illinois, United States) cyclers for predicting catheter dysfunction in automated peritoneal dialysis patients (APD). Out of 117 patients treated at the Medical University of Vienna between 2015 and 2021, we retrospectively identified all patients with verified catheter dysfunction (n = 14) and compared them to controls without clinical evidence of mechanical catheter problems and a recent X-ray confirming PD catheter tip in the rectovesical/rectouterine space (n = 19). All patients had a coiled single-cuff PD catheter, performed tidal PD, and received neutral pH bicarbonate/lactate-buffered PD fluids with low-glucose degradation products on APD. Icodextrin-containing PD fluids were used for daytime dwells. We retrieved cycler data for seven days each and tested parameters' predictive capability of catheter dysfunction. Total number of alarms/week > 7 as single predictive parameter of catheter dislocation identified 85.7% (sensitivity) of patients with dislocated catheter, whereas 31.6% (1-specificity) of control patients were false positive. A combination of parameters (number of alarms/week > 7, total drain time > 22 min, ultrafiltration of last fill < 150 mL) where at least two of three parameters appeared identified the same proportion of patients with catheter dislocation, but was more accurate in identifying controls (21.1% false positive). In contrast to yearly PET measurements, an easily applicable combination of daily cycler readout parameters, also available in new APD systems connected to remote monitoring platforms shows potential for diagnosis of catheter dysfunction during routine follow-up.
Similar content being viewed by others
Introduction
Successful treatment and technique survival on peritoneal dialysis (PD) depend on a long-term peritoneal access that provides effective peritoneal drainage of dialysis fluid for maintenance of an adequate fluid and salt homeostasis1. PD catheter tip migration is a major complication that usually requires catheter reposition or replacement and may eventually lead to technique failure2. Catheter dislocation may be associated with slow and incomplete drainage of dialysis fluid from the peritoneal cavity. It is important to note that while drain volume can be larger than fill volume even with malpositioned catheters3, drainage may be incomplete, leading to increased abdominal pressure and a higher risk of hernias and dialysate leaks4.
Low ultrafiltration (UF) capacity may be misleading in patients with PD catheter dysfunction as incomplete peritoneal drainage of dialysate can result in insufficient ultrafiltration and fluid overload. Therefore, the recent ISPD recommendations suggest that low UF capacity should be evaluated by considering mechanical drainage problems first, before suspecting membrane dysfunction5. Previous publications have suggested the association of slow drainage of PD fluid, poor net UF or constant alarms during cycler treatment with catheter dysfunction6,7. However, until now no standard procedure has been systematically evaluated on the usability to assess mechanical drainage problems in PD patients.
In contrast to isolated measurements of ultrafiltration capacity, for example during the peritoneal equilibration test (PET), automated peritoneal dialysis (APD) cycler readouts allow meticulous daily analysis of important treatment parameters. The aim of this proof-of-concept study was to analyze the usability of cycler parameters routinely used in our center for identifying APD patients with catheter malfunction.
Results
Out of 117 APD patients treated between 2015 and 2021, 14 patients with verified catheter dysfunction (12%) and 19 controls (16%) could be identified. There were no significant differences in demographic data (Table 1) between both groups. The median daily glucose load was 132.0 g/day [IQR (73.6)] in the controls, and 124.1 g/day [IQR (8.0)] in the cases (MWU test, P = 0.0471). Other PD prescription data were not statistically different between groups (Table 2). Eight out of the 14 patients with catheter dysfunction could continue PD treatment after conservative measures (reduction of tidal volume, laxatives and exercises in order to relocate catheter tip). Four patients needed catheter repositioning or replacement (laparoscopically) and 2 patients were transferred to hemodialysis since it was not possible to continue with PD.
Number of alarms
The mean number of cycler alarms was 19.3 ± 16.5/week in the cases and 5.2 ± 4.0/week in the controls, with a median of 13.5 alarms/week [IQR (16.8)] and 5.0 alarms/week [IQR (6.0)], MWU test P = 0.0002).
Total drain time and drain volume at the end of the cycler session
Total drain volumes at the end of cycler treatment were not statistically different between groups, with a median of 2080 mL [IQR (384)] in the cases and 2065 mL [IQR (1062)] in the controls (MWU test P = 0.7329). However, the total drain time at the end of the cycler session was significantly larger in the cases than in the controls (mean 29.0 ± 11.3 min vs 20.2 ± 6.8 min, P = 0.0012) and a median of 24.3 min [IQR (14.3)] vs 17.7 min [IQR (5.3)].
Net UF of last fill (daytime dwell)
All patients received icodextrin-containing PD fluid for the daytime dwell, except for one patient in the control group who performed nightly intermittent PD without last fill. The mean net UF of the cases was statistically different and more negative than in the controls (− 12.7 ± 153.3 mL vs 206.3 ± 228.4 mL, P = 0.0047). The median net UF of last fill was − 10.0 mL [IQR (259.9)] in the cases vs 227.6 mL [IQR (328.9)] in the controls.
Number of days with negative UF of last fill per week
The number of days with negative UF of last fill (drain volume lower than fill volume) tended to be higher in the cases than in the controls (mean 3.3 ± 2.6 days/week vs 1.6 ± 2.1 days, MWU test P = 0.0749). The median for the cases was 3.0 days/week [IQR (6.0)] vs 1.0 day/week [IQR (2.5)], respectively.
Net UF and gcUF during cycler treatment (without UF of last fill)
The mean net UF per cycler treatment in controls was 393.5 ± 445.8 mL, whereas the cases had a lower mean net UF of 102.2 ± 372.2 mL per treatment (median 242.1 mL [IQR (607.6)] vs 160.5 mL [IQR (826.1)], P = 0.1320). Nevertheless, the mean glucose-corrected UF (gcUF) was not statistically different between groups, 1.5 ± 2.9 mL/g/day in the controls vs. 0.06 ± 2.3 mL/g/day in the cases (median 1 mL/g/day [IQR (4.3)] vs 0.42 mL/g/day [IQR (3.3)], P = 0.2258).
Supplementary Table S1 provides a summary of all analyzed parameters.
Defining cut-off values for prediction
The parameters with the highest area under the curve (AUC) were number of total alarms (0.87, P = 0.0004), total drain time (0.82, P = 0.0017), and mean net UF of last fill (0.79, P = 0.0056). We decided to include number of days with negative UF of last fill (0.68, P = 0.0839) because there was a statistical trend towards significance. Furthermore, in our experience, it is easy to report in clinical practice for both, patients and physicians. The smallest AUC was identified with the mean gcUF (0.63; P = 0.2155). Therefore, this parameter was not included in further analysis.
For these four selected parameters the cut-off values for predicting catheter dislocation based on ROC curves were > 7 alarms/week (specificity: 85%, sensitivity: 79%), > 22 min drain time at end of the cycler session (specificity: 79%, sensitivity: 89%), < 150 mL mean net UF of last fill (specificity: 93%, sensitivity: 66%), and > 2 days per week with negative UF of last fill (specificity: 71%, sensitivity: 78%) (Fig. 1, Supplementary Table S2).
Table 3 summarizes the simple and stepwise logistic regression of the parameters for predicting catheter dislocation. Each additional alarm per week increased the odds of catheter dislocation by a factor of 1.298 (95% CI 1.062; 1.586, P = 0.011) leading to an increase by 40.6 for 14 alarms (mean difference between cases and controls). Each additional minute of total drain time increased the odds of catheter dislocation by a factor of 1.125 (95% CI 1.015; 1.247, P = 0.025). A 1 ml (ml) decrease of mean UF of last fill increased the odds for catheter dislocation by a factor of 1.006 (95% CI 0.989; 0.999, P = 0.013). The mean difference between cases and controls was approximately 200 mL leading to an increase of the odds by 3.33 for a catheter dislocation. Furthermore, an increase of days with negative UF of last fill tended to be associated with catheter dislocation (P = 0.063). In the stepwise logistic regression analysis (including only the four parameters) the total number of alarms improved the fit of the model for the prediction of catheter dislocation compared to the null model. DeLong test showed a significant difference between the two AUCs (P<0.001).
After adding the standard deviation of the residuals (SD-Res) of the net UF of the last fill as well as the SD-Res of the gcUF as measures of variability of the net UF to the existing four covariates into the stepwise regression model and using BIC as criterion for selection, all covariates except total alarms were eliminated, indicating that SD-Res does not improve the prediction model.
One patient with catheter dysfunction developed a peritonitis during the second part of the study period. We repeated analysis of the ROC curves and stepwise logistic regression analysis after excluding this patient, and the results were consistent.
Predictive power for combination of parameters
To facilitate the utilization of parameters of the model in clinical practice, models with the highest AUCs were selected for parameter combinations: total alarms/week, drain time at the end of the cycler session, UF of last fill (or, alternatively, days with negative UF of last fill). Total number of alarms/week > 7 as single predictive parameter identified 85.7% of the cases, 31.6% of controls were false positive. The presence of at least two of three parameters (number of alarms/week > 7, total drain time > 22 min, UF last fill < 150 mL) identified the same proportion of cases compared with number of alarms alone (85.7%), but it was more accurate in identifying control patients (21.1% vs 31.6% false positive). Stricter definitions, such as fulfilling all three criteria, reduced the predictive power for diagnosing catheter dislocation, but also decreased the number of false positive controls (Table 4).
Discussion
We demonstrated that a protocol assessing several cycler readout parameters might be a feasible, easy-to-perform tool to adequately screen APD patients for catheter dysfunction.
In this proof-of-concept study, we confirmed the clinically observed impression that the parameters under investigation serve as predictors for diagnosis of catheter dysfunction. This is the first controlled analysis evaluating parameters obtained from the APD cycler software for the diagnosis of catheter dysfunction in two well-defined patient groups. The selected cases and controls represent typical APD patients with radiological confirmation of PD catheter position and clinical assessment of catheter performance.
Patients with a malfunctioning catheter may clinically present with slow drainage of PD fluid, prolonged drain times, poor net UF6,7 or constant alarms during cycler treatment8, resulting in impairment of quality of life. Early diagnosis prevents the possibility of volume overload and complications associated with increased intraperitoneal volume that subsequently may lead to technique failure9.
Recent ISPD recommendations suggest that mechanical problems should be excluded prior to membrane function testing in patients with insufficient UF capacity. However, there exists no standard protocol on how to diagnose catheter dysfunction5.
One of the requirements for good catheter performance is an adequate position of the PD catheter tip. The ISPD guidelines on PD access recommend using the upper border of the pubic symphysis as a reference point for the ideal location of the catheter tip deep in the pelvic area7. During or after catheter implantation, the position of the catheter tip might be confirmed with a plain abdominal radiological examination, but this is not routinely performed in all centers10.
Single diagnostic tests performed in yearly intervals (e.g. ultrafiltration capacity during the PET) may not be suitable for diagnosing mechanical problems. Net ultrafiltration is more variable than peritoneal small solute transport. Availability of daily treatment data extracted from the APD cycler card software allows a better clinical evaluation of mechanical drainage problems in routine practice.
In our study, in the stepwise logistic regression analysis, only number of alarms/week was a statistically significant predictor of catheter dislocation. As the Pearson and Spearman correlation coefficients between number of total alarms and the other metric variables revealed values above 0.5 (indicating moderate to strong correlations), none of the other variables could contribute anything in addition. However, considering the limited patient number this analysis does not refute the importance of other parameters for diagnosing mechanical drainage problems.
A combination of at least two of three parameters which had significant ROC curves and an AUC of > 0.75 (number of alarms, total drain time, net ultrafiltration of last fill) could identify a similar number of patients with catheter dislocation compared to using the number of alarms alone (85.7%). However, when using a combination of parameters, the percentage of false positive control decreased (Table 4).
The results of this study are strengthened by the accordance of the cut-off values derived from the ROC curves with conclusions of other authors in the recent literature. We determined that a cut-off value of > 7 alarms/week is a good predictor of catheter dysfunction. According to Neri et al.11, the use of tidal PD optimized catheter flow function, resulting in the presence of < 1 alarm per day (corresponding to < 7 alarms per week). Other authors also suggest that outflow failure prompts alarms in APD patients treated with HomeChoicePro8. Moreover, the cut-off value for mean drain time of > 22 min for predicting catheter dysfunction found in our study is similar to the conclusion of other authors who suggested that a drainage time of less than 20 min is a marker of good catheter performance3,12. A sub-optimal catheter position results in poor hydraulic function that may cause outflow failure and can prolong drain time at the end of the cycler session10,13. We demonstrated that low or negative UF of last fill during the daytime dwell was evident in patients with catheter dislocation. Our selected cut-off value of UF for predicting catheter dysfunction of 150 mL per daytime dwell is consistent with data reported by Plum et al.14 who found a mean ultrafiltration volume of last fill of 278 ± 43 mL/day in APD patients treated with icodextrin during the long dwell, which was comparable with the mean net UF of last fill of our control group (206.3 ± 228.4 mL).
Our study has certain limitations. We acknowledge that we are reporting a retrospective, single-center study of patients on APD, which may limit statistical power and generalizability of our results. For example, albeit not statistically significant, assessing the number of treatment days with negative UF of last fill may be more practicable than analyzing cut-off values of ultrafiltration with icodextrin. Although we included all APD patients treated in our center between 2015 and 2021 who were suitable for inclusion in either groups, the sample size is still small and exclusion of 29% of eligible patients with missing data may have limited the validity of the stepwise logistic regression analysis and the representativeness of the results. Larger, prospective studies are required to confirm our clinical observations.
In conclusion, a timely intervention in patients with catheter dysfunction may prevent adverse events with potential negative impact on technique survival9. In this context, the use of total number of alarms/week as well as a combination of parameters derived from the APD cycler card management software with the selected cut-off values are good predictors for assessing catheter function. Our findings may be especially interesting because these parameters are also available in new APD cyclers with devices connected to remote monitoring platforms, which are rapidly gaining importance and availability9,15.
Material and methods
Study design
This retrospective case–control, proof-of-concept study included patients with end-stage chronic kidney disease treated with HomeChoicePro APD system (Baxter International Inc., Illinois, United States) at the Division of Nephrology and Dialysis, Medical University of Vienna.
To validate parameters related to mechanical drainage problems we studied two well-characterized patient groups using data retrieved from the APD cycler card management software PD Link (Baxter International Inc., Illinois, United States). In order to minimize bias, retrospective analysis of the PD catheter position based on posteroanterior and/or lateral abdomen X-ray, if available, and review of clinical records to assess catheter drainage function and patients’ demographic data was performed by two investigators. This study was performed in line with the principles of the Declaration of Helsinki and involves retrospectively collected data. Informed consents were collected from all our patients before inclusion in our database and biobank. Approval was granted by the Local Ethics Committee of the Medical University of Vienna (study protocol EK 2035/2015).
Patient selection
Patients included in this study received regular APD between January 2015 and March 2021 at the Medical University of Vienna. All patients were treated with neutral pH bicarbonate/lactate-buffered PD solutions with low concentration of glucose degradation products (GDP) (Physioneal 40, Baxter Healthcare) and icodextrin-containing PD fluids (Extraneal, Baxter Healthcare) for the daytime dwell (except for one patient in the control group who performed nightly intermittent PD without last fill). All patients had a coiled single-cuff PD catheter and performed tidal peritoneal dialysis. Patients with incomplete data (no abdominal X-ray or data on catheter performance available) were withdrawn. Patients with a history of large open abdominal or pelvic surgery, ascites, malignant pelvic neoplasia or death within 15 days after APD start, were excluded (Fig. 2).
Cases
All patients with clinical evidence of drainage problems that required a diagnostic and/or therapeutic intervention (repeated abdominal X-ray to verify catheter position, use of laxatives, catheter repositioning or replacement) and with confirmed catheter dislocation in an abdominal X-ray (verifying catheter tip dislocation outside the deep pelvic area) were selected as cases.
Controls
Patients with an uncomplicated course of treatment without clinical evidence of mechanical problems and a recent routine X-ray confirming position of the PD catheter tip in the recto-vesical/ recto-uterine space were selected as controls. Patients with a computed tomography (CT) peritoneography or an explorative laparoscopy/adhesiolysis after start of PD were excluded from the control group since requirement of these diagnostic or therapeutic measures do not represent an uncomplicated course of treatment.
Parameters
Based on previous publications8,9,11,15,16 and on our own clinical expertise, daily cycler readout parameters are routinely used at our center to detect mechanical drainage problems early during routine follow-up. Number of alarms, drain time at the end of cycler session, net UF from last fill, number of days with negative UF of last fill and net UF from cycler treatment (net UF and glucose-corrected UF [gcUF]) were analyzed in the present study.
The observation period for both groups consisted of seven days of continuous APD treatment. Among cases, these parameters were analyzed in the week immediately before the diagnostic imaging studies confirming catheter tip migration. The time-period for the analysis of the control group was seven days selected randomly depending on data availability, within a pre-defined time window between 1 and 3 months after the start of PD. This time window was chosen considering the appearance of changes of peritoneal membrane function parameters early after PD start that may influence ultrafiltration5.
Furthermore, daily glucose load was calculated as total grams of glucose in the fresh PD fluid infused each day. Daily glucose corrected ultrafiltration (gcUF) was calculated as UF in mL divided by glucose load in grams. APD prescription was not modified during the observation period.
Statistical analysis
Statistical analysis was performed using IBM SPSS Statistics (IBM Corp., Armonk, NY, USA), R (GNU General Public License17) and GraphPad Software (California, USA). Results were expressed as relative frequencies for categorical variables, mean with standard deviation (SD) for continuous variables, and median with interquartile range (IQR) for skewed distributions. Comparison of categorical variables was performed with Fisher’s exact test. Continuous variables were analyzed with Mann–Whitney U-test (MWU). The receiver-operating curves (ROC) for each parameter were computed from the observed proportion utilizing Clopper method18, without correction for multiple comparisons of area calculation. Selected cut-off values for the parameters were identified via means of the highest sensitivity and specificity. Stepwise logistic regression (catheter dislocation versus control as dependent variable) was performed to evaluate the contribution of several covariates to anticipation of catheter dislocation using “stepwise” in the R-package “RcmdrMisc”19 with BIC (Bayesian Information Criterion) as criterion for selection. To account for the variability of net UF of the last fill as well as the gcUF as a potential predictors for mechanical problems, we determined the expected values for both variables via linear regression for each patient for all seven days according to the method stated in a publication by Yang et al.20 on hemoglobin variability in 2007. Residuals of net UF in the last fill as well as the gcUF were calculated per patient day and summarized as standard deviation of the residuals (SD-Res) per patient. P values of < 0.05 were considered to be statistically significant.
Change history
10 May 2022
A Correction to this paper has been published: https://doi.org/10.1038/s41598-022-12074-y
References
Brown, E. A. et al. International Society for Peritoneal Dialysis practice recommendations: Prescribing high-quality goal-directed peritoneal dialysis. Perit. Dial. Int. 40, 244–253. https://doi.org/10.1177/0896860819895364 (2020).
Béchade, C. et al. Early failure in patients starting peritoneal dialysis: A competing risks approach. Nephrol. Dial. Transplant. 29, 2127–2135. https://doi.org/10.1093/ndt/gft055 (2014).
Palomar, R. et al. The position of the peritoneal dialysis catheter is not essential for a correct performance. Clin. Nephrol. 70, 554–557. https://doi.org/10.5414/cnp70554 (2008).
Castellanos, L. B. et al. Clinical relevance of intraperitoneal pressure in peritoneal dialysis patients. Perit. Dial. Int. 37, 562–567. https://doi.org/10.3747/pdi.2016.00267 (2017).
Morelle, J. et al. ISPD recommendations for the evaluation of peritoneal membrane dysfunction in adults: Classification, measurement, interpretation and rationale for intervention. Perit. Dial. Int. 10, 352–372. https://doi.org/10.1177/0896860820982218 (2021).
Crabtree, J. H. Selected best demonstrated practices in peritoneal dialysis access. Kidney Int. Suppl. 103, 27–37. https://doi.org/10.1038/sj.ki.5001913 (2006).
Crabtree, J. H. et al. Creating and maintaining optimal peritoneal dialysis access in the adult patient: 2019 update. Perit. Dial. Int. 39, 414–436. https://doi.org/10.3747/pdi.2018.00232 (2019).
Chaudhry, R. I. & Golper, T. A. Automated cyclers used in peritoneal dialysis: Technical aspects for the clinician. Med. Devices (Auckl). 23, 95–102. https://doi.org/10.2147/MDER.S51189 (2015).
Drepper, V. J., Martin, P. Y., Chopard, C. S. & Sloand, J. A. Remote patient management in automated peritoneal dialysis: A promising new tool. Perit. Dial. Int. 38, 76–78. https://doi.org/10.3747/pdi.2017.00054 (2018).
Stegmayr, B. G. et al. Few outflow problems with a self-locating catheter for peritoneal dialysis: A randomized trial. Medicine 94, e2083. https://doi.org/10.1097/MD.0000000000002083 (2015).
Neri, L., Viglino, G., Cappelletti, A. & Gandolfo, C. Evaluation of drainage times and alarms with various automated peritoneal dialysis modalities. Adv. Perit. Dial. 17, 72–74 (2001).
Daugirdas, J. T. et al. (eds) Handbook of Dialysis 458–459 (Wolters Kluwer Health, Philadelphia, 2015).
Ronco, C. et al. Remote patient management of peritoneal dialysis during COVID-19 pandemic. Perit. Dial. Int. 40, 363–367. https://doi.org/10.1177/0896860820927697 (2020).
Plum, J. et al. Efficacy and safety of a 7.5% icodextrin peritoneal dialysis solution in patients treated with automated peritoneal dialysis. Am. J. Kidney Dis. 39, 862–871. https://doi.org/10.1053/ajkd.2002.32009 (2002).
Stachowska-Pietka, J. et al. Water removal during automated peritoneal dialysis assessed by remote patient monitoring and modelling of peritoneal tissue hydration. Sci. Rep. 11, 15589. https://doi.org/10.1038/s41598-021-95001-x (2021).
Qayyum, A., Yang, L. & Fan, S. L. Optimizing peritoneal dialysis catheter placement by lateral abdomen X-ray. Perit. Dial. Int. 35, 760–762. https://doi.org/10.3747/pdi.2014.00263 (2015).
R Core Team. R: A language and environment for statistical computing. R Foundation for Statistical Computing. https://www.R-project.org/ (2020).
Clopper, C. J. & Pearson, E. S. The use of confidence or fiducial limits illustrated in the case of the binomial. Biometrika 26, 404–413. https://doi.org/10.1093/biomet/26.4.404 (1934).
Fox, J. RcmdrMisc: R Commander Miscellaneous Functions. R package version 2.7–1. https://CRAN.R-project.org/package=RcmdrMisc (2020).
Yang, W. et al. Hemoglobin variability and mortality in ESRD. J. Am. Soc. Nephrol. 18, 3164–3170. https://doi.org/10.1681/ASN.2007010058 (2007).
Acknowledgements
This work is part of “Identification and Management of Patients at Risk – Outcome and Vascular Events in Peritoneal Dialysis” (IMPROVE-PD), a project that receives funding from the European Union’s Horizon 2020 research and innovation program under the Marie Skłodowska-Curie Actions, grant agreement No 812699. The financial support by the Austrian Federal Ministry of Science, Research and Economy and the National Foundation for Research, Technology and Development is gratefully acknowledged. We sincerely thank Julia Andrejevic, Helene Rosenberg and Filip Zubic for their assistance with data collection.
Author information
Authors and Affiliations
Contributions
K.O.F and A.V. developed concept and design of the study, had full access to all data and take responsibility for the integrity and accuracy of data and subsequent analysis. Data collection was performed by K.O.F and M.U. K.O.F and A.V. conducted analysis and drafted the initial manuscript. K.O.F., F. E., C.A. and A.V. contributed to interpretation of data. Statistical analysis was done by L.K. and F.K. Critical revision of the manuscript for important intellectual content was done by all authors. All authors approved the final manuscript as submitted.
Corresponding author
Ethics declarations
Competing interests
A.V. has received honoraria and travel grants from Baxter and Fresenius and consulting fees from Baxter unrelated to this study. C.A. is cofounder of Zytoprotec GmbH, a spin-off of the Medical University Vienna that holds the patent “Carbohydrate-based peritoneal dialysis fluid comprising glutamine residue” (International Publication Number: WO 2008/106702 A1). M.U. and K.K. are former employees, K.K. is a consultant of Zytoprotec GmbH. K.O.F. is PhD candidate at the Medical University of Vienna and also employee of Baxter Healthcare GmbH, funded as early-stage researcher by the IMPROVE-PD (812699)/Horizon 2020 project. Baxter Healthcare did neither participate in the development of the protocol and study design, nor in the analysis of data or elaboration of the manuscript. All other authors have no conflicts of interest to declare.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
The original online version of this Article was revised: The original version of this Article contained an error in the Results section under the subheading ‘Defining cut‑off values for prediction’ where, “The parameters with the highest area under the curve (AUC) were number of total alarms (0.87, P = 0.0004), total drain time (0.82, P = 0.0017), and mean net UF of last fill (0.79, P = 0.0078.” now reads: “The parameters with the highest area under the curve (AUC) were number of total alarms (0.87, P = 0.0004), total drain time (0.82, P = 0.0017), and mean net UF of last fill (0.79, P = 0.0056).”
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Oviedo Flores, K., Kaltenegger, L., Eibensteiner, F. et al. Assessing mechanical catheter dysfunction in automated tidal peritoneal dialysis using cycler software: a case control, proof-of-concept study. Sci Rep 12, 5657 (2022). https://doi.org/10.1038/s41598-022-09462-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-022-09462-9
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.