Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

Optimisation of T2 and T2* sequences in MRI for better quantification of iron on transfused dependent sickle cell patients

An Author Correction to this article was published on 26 August 2021

This article has been updated

Abstract

This work aimed to investigate the effect of different shim techniques, voxel sizes, and repetition time (TR) on using theT2 and T2* sequences to determine their optimum settings to investigate the quantification of iron in transfused dependent sickle cell patients. The effect of each of these parameters was investigated on phantoms of different Gadolinium (Gd) concentrations, on 10 volunteers and 25 patients using a1 5T MRI Philips scanner. No significant difference between the three shim techniques was noticed in either T2 or T2* sequence measurements. Pixel sizes of 1 × 1 and 2 × 2 mm provided optimum results for T2 measurements. At 1 × 1 mm pixel size the T2* measurements experienced less error in measurements than the size of 2.5 × 2.5 mm used in the literature. Even though the slice thickness variation did not provide any changes in T2 measurements, the 12 mm provided optimum T2* measurements. TR variation did not yield significant changes on either T2 or T2* measurements. These results indicate that both T2 and T2* sequences can be further improved by providing more reliable measurements and reducing acquisition time.

Introduction

Sickle cell anaemia (SCA) is a haemoglobinopathy caused from a single point mutation in the β-chain of human haemoglobin1 It causes rigid and sickle-shaped red blood cells, eventually ends up in anaemia, which leads to multiple complications, the most serious of which is the Overt stroke2.

In Africa, sickle cell disease is raising great health concerns, particularly in the western part of Sudan among the Baggara (Misseriya tribe). Archibald3 stated that the first incidence of sickle cell in Sudan was reported in 1926. The number of cases of this disease reached 30% among pastoral tribes in western Sudan. Up to 16% was found in immigrant tribes in the area4.

The prevalence of patients with SCA in Sub-Saharan Africa ranges between 5 and 40%5. Their life expectancy is < 20 years old6, and patients younger than 5 years are at high risk of death.

Blood transfusions in sickle cell patients could lead to iron overload particularly in the liver. Iron overload promotes the formation of toxic oxygen radicals that can cause cell damage. Therefore, the knowledge of iron levels in these patients is important to mitigate the health consequences of iron overload.

The levels of iron in blood-transfusion patients are often monitored using the serum ferritin method. This method known to be an unreliable marker for assessing iron in organs such as the heart and liver7.

The use of magnetic resonance imaging (MRI) techniques involving transverse relaxation times such as the T2 and T2* have shown promising results in estimating iron concentration in transfusion-dependent patients8,9. It is therefore, expected that the serum ferritin and invasive biopsy methods will be replaced by the MRI methods, however, there is limited data in the literature that systematically compared T2 and T2* sequences in sickle cell anemia (SCA)10. For the quantification of iron using MRI techniques, the literature highlighted that MR signal measurements could be affected by sequence protocol11,12,13,14. Nevertheless, many of these studies were limited to narrow investigations of parameter effects on T2 and T2* measurements. Researchers stated that T2 techniques are not affected by the size or shape of the imaging voxel15 or shimming techniques16, and commented that T2* methods were affected by voxel size17, shimming technique and TR.

The empirical effects of shimming techniques, range of voxel sizes, and TRs were not been sufficiently investigated. There is also a lack of understanding of how these changes in parameters affect T2 or T2* measurements. Reported T2 and T2* values in healthy subjects showed large variation between centres which could be explained due to the acquisition standardisation18. While the literature highlighted some guidelines for use of protocols for T2* measurements for cardiovascular image19, different acquisition approaches were implemented for T2 sequences and hence lead to variations in T2 values20. It was also suggested that changing T2* sequence parameters may particularly be significant at high liver concentrations9.

Gradient echo sequence (T2*), and Ferriscan (T2 sequences) have been used for evaluating iron in patients with iron overload, however, Ferriscan is not widely available in Africa.

The T2 sequence is currently a validated sequence for quantification of iron, on the other hand it is technically difficult to implement and requires additional costs and delay for data analysis21. The T2 sequence that is available in most scanners is the Fast Spin Echo Sequence or Turbo Spin Echo (TSE) sequence, and has been suggested to be used for iron quantification in transfusion-dependent patients22.

Herein we thought to investigate the effect of different shim techniques, voxel sizes, and repetition times (TR) on the quantification of iron in transfusion-dependent sickle cell patients and examining optimum settings for appropriate assessment of patients.

The present study was built on the experimental hypothesis in changing settings and optimizing parameters that may lead to improving the quality of the images produced, which leads to the appropriate evaluation transfusion dependent patients.

Materials and method

The study was conducted at Dar-Alateeba Hospital, a private hospital located in Khartoum State, Sudan. The measurements involved the use of a 1.5 T whole-body Philips-MRI scanner. The scanner was equipped with high performance gradients with a maximum strength of 40 mT/m and a maximum slew rate of 200 T/m/s. A four-element torso phased array coil was used for all the scans. The research work explored the best settings that can be used in patients’ measurement via investigations conducted first on a phantom, on healthy volunteers and finally on patients as described in the following subsections.

Phantom study and calibration

Phantom was used in the study to simulate the relaxation properties of tissues with different concentrations of iron. It consisted of a Plexiglas holder containing eight plastic bottles, each with a volume of 17 ml. Seven bottles contain gadolinium (Gd) at concentration of 0.5, 1, 2, 3, 4, 5 and 20%. One of the bottles contained pure water for reference. The effect of the paramagnetic gadolinium on the signal intensity measurements using T1, T2 and T2* sequences has been previously investigated in number of studies23,24,25. Researchers reported that the increase in gadolinium concentrations led to a significant decrease on the T2 and T2* relaxation26,27.

The phantom was initially scanned with a balance fast field echo (FFE) sequence to localise the best plane for analysis. Following this a Turbo Spin Echo (or T2 sequences) and a Gradient Echo (or T2* sequence) were run on the phantom first for calibration using the following parameters: shim = default, pixel size = 2 × 2 mm, slice thickness = 10 mm, TE/ES = 4.4/4.4 and TR = 400 ms, which gives an acquisition time of 27.2 s. For T2* sequence the settings were: shim = default, pixel size = 2 × 2 mm, slice thickness = 10 mm, TE/ES = 2.5/1.1 ms, and TR = 200 ms and this gives an acquisition time of 7.8 s. The number of echoes for each of the T2 and T2* sequences were 20 and 15 respectively. The scan for each of the two calibration protocols was repeated 3 times. Their mean and standard deviation (SD) of the repeat measurements were obtained.

Following this, optimization protocols were carried out for both the T2 and T2* sequences and it included the effects of the calibration parameters as well as the effects of the remaining: the shim techniques, four pixel sizes, two slice thicknesses and four TR values. For each of the imaging parameters, the scan is repeated three times on which also the mean and SD were calculated for the two sequences.

Volunteers and patients

The results obtained from the phantom study were tested on 10 volunteers and 25 patients. The volunteers were healthy, with no known previous diseases. Initially, serum ferritin measurements were performed for all patients and volunteers. The studied subjects were scanned for the following parameters: Shim default, auto and volume for both T2 and T2* sequence. Pixel sizes were 2 × 2, 2.5 × 2.5, 3.28 × 3.28 and 4 × 4 mm for T2 and 1 × 1, 2 × 2, 2.5 × 2.5 and 3.28 × 3.28 mm for T2*. TR values were 245, 330, 400, and 500 ms for T2, and 100,160, 200, and 230 ms for T2* sequence. Slice thickness of 8, 10, and 12 mm was also investigated on the phantom for both T2 and T2* sequences.

The patients’ study was conducted at the optimum settings obtained from both the phantom and volunteers’ studies using both T2 and T2* sequences. These were then compared to the calibration/standard settings, which was implemented before the incremental changes. The settings for the optimised sequence for T2 were: shim = auto, pixel size = 2 × 2 mm, slice thickness = 10 mm, TE/ES = 4.4/4.4 ms, and TR = 245 ms, with an acquisition time = 16.9 s. The optimised settings for the T2* sequence were as follow: shim = default, pixel size = 1 × 1 mm, slice thickness = 12 mm, TE/ES = 3.7/1.7 ms and TR = 100 ms-with acquisition time = 5.8 s.

For paediatric patients not able to breath-hold, the sequence was implemented during free breathing using respiratory triggering.

All images acquired were analyzed by plotting a circular region of interest (ROI) using Matlab code (version 2015b). For the phantom study, the ROI was chosen around each of the eight bottles. As for the volunteers and patients, regions away from the vascular structure were chosen to avoid the contribution of artefacts from the blood. The pixel intensities of each of the selected region were averaged together and this region is propagated to other images with longer echo times (TE). The signals were then plotted versus their corresponding TE values using Levenberg–Marquardt algorithm of non-linear curve fitting. Data from T2 and T2* were fitted using a simple exponential decay formula28 (Eq. 1):

$${\text{S}} = {\text{S}}_{0} \exp ^{{ - {\text{TE}}/{\text{T}}2^{*} }}$$
(1)

where S is the final signal intensity, S0 is the initial signal and TE is the echo time.

All data points at noise level were removed using the truncation method as they are currently widely used29 and proven to provide more accurate measurements than the offset and baseline subtraction methods9.

All Statistical analysis was performed using EXCEL spreadsheets. To evaluate the differences between the two data sets, a paired Student’s t-test was calculated. A p value less than 0.05 was considered to be significant. Image analysis were carried out using a personally written programming code (Matlab 2015b).

Ethical approval was obtained from the Health Research Council-National Research Ethics Committee (Ministry of Health, Sudan). The committee was established to review ethical issues in health research proposals submitted for implementation, in the country, and review its results in accordance with regulations that adopt international guidelines. Written informed consent was obtained from each individual for carrying out the studies.

Results and discussion

Results

The effects of tested parameters were studied in terms of three performance indices: (i) The exponential decay curves of both T2 and T2*; (ii) the magnitude measurements of T2 or T2* values using Eq. 1 and (iii) the correlation coefficient R2, which determines the associated strength between two variables. All results are presented in in Figs. 1, 2, 3, 4 and 5 and Tables 1 and 2.

Figure 1
figure1

Decay curves of the 2% Gd concentration for T2 and T2* sequences, respectively, using; the three shims (A, B), Five different pixel sizes (C, D), three slice thickness (E, F) and five TR values (G, H).

Figure 2
figure2

Decay curves of T2 and T2* for a healthy volunteer using: the three shim techniques (A, B), pixel sizes of 2 × 2 to 4 × 4 mm for T2 (C), pixel sizes of 1 × 1 to 3.28 × 3.28 mm for T2* (D), slice thicknesses of 8, 10, 12 mm (E, F), TR values 245–500 ms for T2 (G) and TR values 100–230 ms for T2* (H).

Figure 3
figure3

Decay curve of a sickle cell patient: (A) using auto shim, 2 × 2 mm pixel size, 10 mm slice thickness, and TR = 245 ms for T2 sequence ; (B) at default shim, 1 × 1 mm pixel size, 12 mm slice thickness, and TR = 100 ms for T2* sequence.

Figure 4
figure4

Mean magnitude measurements (with SD) of the 25 patients using the conventional T2 and T2* sequences versus the new optimised sequences.

Figure 5
figure5

MRI images of the same patient using: (A) standard T2*, (B) optimised T2*, (C) standard T2 and (D) optimised T2 sequences.

Table 1 Phantom mean measurements ± standard deviation of T2 and T2* for different shim techniques.
Table 2 Phantom mean measurements ± standard deviation for different pixel size using T2 and T2*.

Phantom and volunteers results

The results of the phantom and volunteers measurements, are presented in Tables 1 and 2 and in Figs. 1 and 2. The figures show plots of the 2% Gd concentration versus echo time (TE) for three shims, five different pixel sizes, three slice thickness and five TR values. The tables show mean magnitude measurements of Gd concentrations 1–20% and of 10 volunteers respectively, for both T2 and T2*, using the tested parameters.

Patients results

The results of serum ferritin for the volunteers ranged between 5.5 and 86.8 ng/ml. For the patients, the serum measurements ranged between 179 and 2000 ng/ml. T2 decay curves of standard (TR 400 ms) versus optimised (TR 245 ms) of a sickle cell patient with iron overload are shown in Fig. 3a. Figure 3b shows the decay curves of the standard (TR = 200 ms) versus optimised (TR = 100 ms) for T2* for the same patient. The magnitude measurements of the 25 patients using the conventional T2 and T2* sequences versus the new optimised sequences are shown in Fig. 4.

The mean magnitudes of the standard and optimised sequences for T2 were 41.72 ± 11.82 ms and 41.10 ± 11.31 ms, and for T2* they were 5.11 ± 3.39 and 5.25 ± 3.49 ms respectively.

Figure 5A,B present typical images of a patient acquired using standard and optimised T2*, while Fig. 5C,D obtained using standard and optimised T2, respectively.

Discussion

It was observed that all decay curves of the phantom studies appeared with offset at the first two data points of the curve. The T2* curves experience slight oscillation. The presence of field inhomogeneities probably caused this oscillation. In Fig. 1 the three shim techniques of T2 and T2* sequences appeared overlapping, indicating that there is no significant variation between their measurements. Increasing the pixel size (in- resolution) caused an increase in the bandwidth (the scanner was not set for a fixed Hz/pixel value). Nevertheless, with the increase in bandwidth, the signal-to-noise ratio was reduced simultaneously. The oscillation of data points around the curve was noticed to increase with higher Gd concentrations. Slice thickness had a minor effect on the T2 measures, while the signal intensity was noticed to increase with slice thickness for T2*. TR was observed to have a minor effect on both sequence measurements.

Results for volunteers, as shown in Fig. 2, indicate that T2 and T2* were loosely fit to its associated decay curves. Factors such as blood, the anatomy of the organ may have contributed to poor fit. However, T2 curves are bound together, indicating that shim has no significant effect on the measurements. Concurrently, T2* data points exhibited higher oscillation than those of the T2, which is attributed to the presence of field inhomogeneities. Such small oscillation can likely have profound effects on estimating long T2* values30. The scattering of data points were noticed to be specifically higher at pixel sizes of 3.28 × 3.28 and 4 × 4 mm, especially for T2* measurements. The trend of the decay curves for the different slice thickness for T2 appeared the same. However, the signal intensity measurements for T2* were noticed to be the highest at 12 mm.

Tables 1, 2, 3 and 4 represent the magnitude measurements (with SD) of Gd concentrations of 1–20% for the different shim techniques, pixel sizes, slice thicknesses, and TR values for both T2 and T2* measurements. Generally, it can be concluded that there is no significant differences in measurements for the three shim techniques. Statistical test showed correlations coefficients (R2) ranging from 0.98 to 0.99 for the three shim techniques. However, the 1% Gd concentration experienced different T2* measurements since it recorded the highest magnitude measurement with the lowest (SD). On the other hand, the increase in pixel sizes resulted in a steady increase in magnitude measurements. The 20% concentration recorded the lowest R2 and highest SD, especially at smaller pixel sizes. This is probably because low T2 measurements cannot be measured accurately due to the limitation of the short echo time and echo spacing used (5.9/5.9 ms). For T2* measurements, the 1 × 1 mm recorded the lowest magnitude measurements with low SD for the whole range of Gd concentrations. Measurements fluctuations can be observed at 4–20% Gd concentrations and the magnitude values increased by nearly twice as much when using pixel sizes 3.28 × 3.28 mm or 4 × 4 mm compared to the 1 × 1 mm pixel. R2 values ranged between 0.514 and 0.997 for all Gd concentrations using these pixel sizes, with the optimum obtained at 1 × 1 mm.

Table 3 Phantom mean measurements ± standard deviation of different slice thickness for T2 and T2*.
Table 4 Phantom mean measurements ± standard deviation of different TR using T2 and T2* sequences.

The slice thickness study on the 1% and 2% Gd using T2* sequence demonstrated higher magnitude measurements at 8 mm, while the 12 mm thickness showed low SD for all the concentrations with R2 ranging 0.975–0.997at p < 0.01. Moreover, TR values had small effects on T2 measurements while 1–2% Gd concentration demonstrated a slight increase in measurements with TR. It was also observed that TR has a minor effect on T2* measurements with the lowest fluctuations of measurements found for concentrations of 3–20% with R2 range between 0.98 and 0.996.

Tables 5 and 6 show mean ± SD of the result of magnitude measurements calculated on volunteers for both T2 and T2* sequences. Average T2 measurements on normal volunteers' livers using the T2 sequence in this study was 44 ± 7 ms comparable with other similar studies31,32,33. We also observed that the average magnitude measurement using the default shim was less than the measurements recorded with auto and volume shim techniques by approximately 0.1 and 0.2%, respectively. The default shim recorded high SD values, while the low fluctuation of measurements was recorded using the auto-shim technique. The overall T2* measurements recorded from the volunteers fell between 30 and 45 ms (average 36 ± 5 ms) comparable with literature34. The T2 magnitude measurements were presented with a slight increase for pixel sizes larger than 2.5 × 2.5 mm.

Table 5 Volunteers measurement of magnitude ± standard deviation for T2 and T2* sequences.
Table 6 Volunteers measurement of magnitude ± standard deviation of the ten volunteers using different TR values for both T2 and T2* sequences.

The T2* magnitude measurements at different pixel sizes showed similarity to the magnitude measurement obtained for the 1% Gd concentration. The lowest measurement was recorded at 1 × 1 mm. Larger pixel sizes provided T2* measurements of up-to 20% than that of the 1 × 1 mm. It is also observed that the SD values became higher with pixel size (reaching up to 50% for pixel size of 4 × 4 mm).

In comparison with the literature, our study found that the increase of pixel size resulted in an increase in the sensitivity of the magnitude measurements35. Larger pixel sizes have caused the measurements to become less reliable. Similar to the T2 measurements the utilization of higher receiver bandwidth in this study incorporated more noise on the acquired images, and subsequently caused scattering in signal intensity measurements. This led to wide fluctuations of measurements especially in the case of high pixel sizes of 3.28 × 3.28 mm and 4 × 4 mm (R2 ~ 0.614). The T2 sequence results showed no significant differences between the magnitude measurements for the slice thicknesses of 8, 10, and 12 mm. As for T2*, the 12 mm thickness was associated with the highest magnitude measurements and lowest fluctuations of the measurements.

Mean magnitude measurements of T2 for the range of studied TR witnessed no significant difference in their values. T2* measurements at TR values of 100-160 ms, recorded low mean magnitude of measurements (by ~ 30%) compare to those at 200-230 ms. Long T2* components, like those found in healthy volunteers, require long TR values for accurate measurements. It is observed that long TRs of 200–230 ms provides consistent results (R2 range 0.95–0.99).

In general, there was a slight difference in magnitude measurements between the conventional and optimised T2 sequences for patients studies (Figs. 3, 4). The conventional sequence provided a higher measurement by 1.5% and higher SD by 4% than its corresponding optimised sequence. The correlation coefficients obtained from the conventional sequence were 0.978–0.994, whilst for the optimised sequence the range was 0.979–0.994 (p < 0.01). We also observed similar image qualities were obtained for the patients using the T2 and T2* optimised sequences as compared to their corresponding standard sequences (Fig. 5).

Thus, the results indicate that the standard T2 sequence could easily replace the optimised sequence without compromising sickle cell patients' measured data with iron overload. Similarly results obtained for T2* sequences were in close agreement with this conclusion (with uncertainty of ~ 3%). The correlation coefficient for the standard sequence ranged between 0.942–0.99, while for the optimised (0.945–1.0) at p-value < 0.01.

Our studies on the effects of these parameters were for the first time investigated on the livers of transfusion-dependent paediatric patients using both T2 and T2* sequence. Our investigations showed a room for further improvement in the T2* sequence by adopting the optimised sequence instead of the conventional sequence. Another advantage of this optimisation is that the acquisition time for the T2 sequence was reduced by almost 38% and for T2* by 25%. We also noticed that the measurements' reliability has dramatically improved.

This study has some limitations to generalize its outcomes for a number of reasons: (i) the number of samples is relatively small (10 volunteers + 25 patients). (ii) To get the optimum pixel size and keep the Hz/pixel fixed, this requires a change in both the field of view (FOV) and the matrix size. This type of modification affects the routine of work in the hospital so we had to keep the FOV and the size of the matrix constant throughout the study period.

Conclusion

The aim of this study was to investigate the effect of different shim techniques, voxel sizes, and repetition time (TR) on T2 and T2* sequences and to determine optimum settings for the evaluation of the amount of iron in the livers of transfusion dependent sickle cell patients. Phantom with different concentrations of Gd (1–20%), and ten volunteers were used to define these settings, which were then investigated on 25 patients. The optimized protocols resulted in a reduction of acquisition time by 38% and 25% for T2 and T2*, respectively. In addition, the quality of images were not compromised as compared to the conventional sequences. The outcomes of this study could be particularly beneficial for young adults who cannot tolerate MRI scans for long periods of time. The research results indicated that T2 and T2* sequences can be further refined to provide more reliable measurements and on transfusion dependent-sickle cell patients.

Change history

References

  1. 1.

    Marotta, C. A., Wilson, J. T., Forget, B. G. & Weissman, S. M. Human 13-globin messenger RNA 111: nucleotide sequences derived from complementary DNA. J. Biol. Chem. 25, 5040–5053 (1977).

    Article  Google Scholar 

  2. 2.

    Kirkham, F. J. Therapy insight: stroke risk and its management in patients with sickle cell disease. Nat. Clin. Pract. Neurol. 3, 264–278 (2007).

    CAS  Article  PubMed Central  Google Scholar 

  3. 3.

    Archibald, R. G. A case of sickle cell anemia in the Sudan. Tran. R. Soc. Trop. Med. Hyg. 19, 389–393 (1926).

    Article  Google Scholar 

  4. 4.

    Ahmed, H. A. & Baker, E. A. Sickling in the Sudan; result of surveys in Blue Nile Province. East Afr. Med. J. 6, 395–399 (1986).

    Google Scholar 

  5. 5.

    Daak, A. A. et al. Effect of Omega-3 (n-3) fatty acid supplementation in patients with sickle cell anemia: randomized, double-blind, placebo-controlled trail. Am. J. Clin. Nurt. 97, 37–44 (2013).

    CAS  Article  Google Scholar 

  6. 6.

    Makani, J. et al. Mortality in sickle anemia in Africa: a prospective cohort study in Tanzania. PLoS ONE 6, e14699 (2011).

    ADS  CAS  Article  PubMed Central  Google Scholar 

  7. 7.

    Bayav, M., Ozbulbul, N. I. & Bor, O. Assessment of cardiac and liver iron overload by magnetic. J. Int. Med. Res. 48(7), 1–6 (2020).

    Google Scholar 

  8. 8.

    Padeniya, P. et al. Comparison of liver MRI R2(FerriScan) VS liver MRI T2* as a measure of body iron load in a cohort of beta thalassaemia major patients. Orphanet. J. Rare Dis. 15, 26 (2020).

    Article  PubMed Central  Google Scholar 

  9. 9.

    Henninger, B., Alustiza, J., Garbowski, M. & Gandon, M. Practical guide on the quantification of hepatic iron with MRI. Eur. Radiol. 30, 383–439 (2020).

    Article  PubMed Central  Google Scholar 

  10. 10.

    Wood, J. C. et al. Liver iron concentration measurements by MRI in chronically transfused children with Sickle Cell Anemia: baseline results from the TWiTCH Trail. Am. J. Hematol. 90, 806–810 (2015).

    CAS  Article  PubMed Central  Google Scholar 

  11. 11.

    Zaman, A. et al. Robust myocardial T2 and T2* mapping at 3T. J. CMR 41, 1013–1020 (2015).

    Google Scholar 

  12. 12.

    Hancu, I., Govenkar, A., Lenkinski, R. E. & Lee, S. K. On shimming approaches in 3T breast MRI. MRM 69, 862–867 (2013).

    Article  Google Scholar 

  13. 13.

    Mulder, M. J., Keuken, M. C., Bazin, P. L., Alkemade, A. & Forstmann, B. U. Size and shape matter: the impact of voxel geometry on the identification of small nuclei. PLoS ONE https://doi.org/10.1371/journal.pone.0215382 (2019).

    Article  PubMed  PubMed Central  Google Scholar 

  14. 14.

    Pavitt, H. L. et al. The effect of reducing repetition time TR on the measurement of liver R2 for the purpose of measuring liver iron concentration. MRM 65, 1346–1351 (2011).

    CAS  Article  Google Scholar 

  15. 15.

    Wood, J. C. et al. Cardiac iron determines cardiac T2*, T2 and T1 in the gerbil model of iron cardiomyopathy. Circulation 112(535), 543 (2005).

    Google Scholar 

  16. 16.

    He, T. et al. Multi-center transferability of a breath-hold T2 technique for myocardial iron assessment. J. Cardiovasc. Magn. Reson. 10, 11 (2008).

    CAS  Article  PubMed Central  Google Scholar 

  17. 17.

    Westwood, M. A. et al. Interscanner reproducibility of cardiovascular magnetic resonance T2* measurements of tissue iron in thalassemia. J. Magn. Reson. Imaging 18, 616–620 (2003).

    Article  Google Scholar 

  18. 18.

    Snel, G. H. et al. Cardiovascular Magnetic Resonance native quantitative values for T2* and T2 cardiomypathies and heart transplantations: a systematic review and meta-analysis. JCMR 22, 44 (2020).

    Google Scholar 

  19. 19.

    Messroghli, J. et al. Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: a consensus statement by the Society for Cardiovascular Magnetic Resonance (SCMR) endorsed by the European Association for Cardiovascular Imaging (EACVI). JCMR 19, 75 (2017).

    Google Scholar 

  20. 20.

    Schwitter, J. Comparison of three different cardiac T2-mapping techniques at 1.5 tesla. Biomed. J. Sci. Tech. Res. 3(2), 3143–3150 (2018).

    Google Scholar 

  21. 21.

    Labranche, R. et al. Liver iron quantification with MR imaging: a primer for radiologists. Radiographics 38, 392–412 (2018).

    Article  PubMed Central  Google Scholar 

  22. 22.

    Sammet, C. L. et al. Measurement and correction of stimulated echo contamination in T2-based iron quantification. Magn. Reson. Imaging 31, 664–668 (2013).

    CAS  Article  PubMed Central  Google Scholar 

  23. 23.

    Rohrer, M., Bauer, H., Mintorovitch, J., Requardt, M. & Weinmann, H. J. Comparison of magnetic properties of MRI contrast media solutions at different magnetic field strengths. Invest. Radiol. 40, 715–724 (2005).

    Article  PubMed Central  Google Scholar 

  24. 24.

    Kanematsu, M. et al. Thick-section half-Fourier rapid acquisition with relaxation enhancement MR cholangiopancreatography: effects of iv administration of gadolinium chelate. AJR Am. J. Roentgenol. 178, 755–761 (2002).

    Article  PubMed Central  Google Scholar 

  25. 25.

    Kuperman, V. Y. & Alley, M. T. Differentiation between the effects of T1 and T2* shortening in contrast-enhanced MRI of the breast. J. Magn. Reson. Imaging 9, 172–176 (1999).

    CAS  Article  PubMed Central  Google Scholar 

  26. 26.

    May, D. A. DJ Pennington (2000) Effect of gadolinium concentration on renal signal intensity: an in vitro study with a saline bag model. Radiology 216, 232–236 (2000).

    CAS  Article  PubMed Central  Google Scholar 

  27. 27.

    Lee, M. J. et al. The T2-shortening effect of gadolinium and the optimal conditions for maximizing the CNR for evaluating the biliary system: a phantom study. Korean J. Radiol. 12(3), 364 (2011).

    Google Scholar 

  28. 28.

    Donald, W. M. From Picture to Proton, 2nd edn, 2007, ISBN-13: 978-0521683845.

  29. 29.

    Alam, M. H. et al. Validation of T2* in-line analysis for tissue iron quantification. JCMR 18, 23 (2016).

    Article  Google Scholar 

  30. 30.

    Ghurge, N. R., Enriquez, C. M., Coates, D. T., Nelson, M. D. & Wood, J. C. Improved R2* measurements in myocardial iron overload. J. Magn. Reson. Imaging 23, 9–16 (2006).

    Article  Google Scholar 

  31. 31.

    Abe, Y., Yamashita, Y., Tang, Y., Namimoto, T. & Takahashi, M. Calculation of T2 relaxation time from ultrafast single shot sequences for differentiation of liver tumors: comparison of echo-planar, HASTE, and spin-echo sequences. Radiat. Med. 18, 7–14 (2000).

    CAS  PubMed  PubMed Central  Google Scholar 

  32. 32.

    Foley, W. D. et al. Contrast optimization for the detection of focal hepatic lesions by MR imaging at 1.5 T. AJR Am. J. Roentgenol. 149, 1155–1160 (1987).

    CAS  Article  PubMed Central  Google Scholar 

  33. 33.

    Bernardino, M. E., Chaloupka, J. C., Malko, J. A., Chezmar, J. L. & Nelson, R. C. Are hepatic and muscle T2 values different at 0.5 and 1.5 Tesla?. Magn. Reson. Imaging 7, 363–367 (1989).

    CAS  Article  PubMed Central  Google Scholar 

  34. 34.

    Obrzut, M. et al. Value of liver iron concentration in healthy volunteers assessed by MRI. Sci. Rep. 10, 17887 (2020).

    ADS  CAS  Article  PubMed Central  Google Scholar 

  35. 35.

    Chavhan, G. B., Babyn, P. S., Thomas, B., Shroff, M. M. & Haacke, E. M. Principles, techniques, and applications of T2*-based MR imaging and its special applications. Radiographics 29, 1433–1449 (2009).

    Article  PubMed Central  Google Scholar 

Download references

Acknowledgements

The authors appreciate the help of Al Ateeba Hospital staff and specifically to Mr. Ahmed Abdelfatah for his time and dedication and for arranging the scanning time at their busy hospital. The authors gratefully acknowledge The Deanship of Scientific Research (DSR) at King Abdulaziz University, Jeddah DSR for technical and financial support under grant no. RG-12-135-41.

Author information

Affiliations

Authors

Contributions

In this work, the principal investigator A.A. was encouraged by the research supervisor I.S. to investigate the possibility of employing magnetic resonance imaging technology for the required measurements in the research problem that was discussed in details with the I.S., A.A. made all arrangements for collecting data. With the assistance of both A.B. and H.M. all work were done, including lab, patient checks, and arranging times for sample collection in the hospital. The collected data was analysed by A.A. and I.S. and then A.A. prepared the draft that include writing the main text and drawing the figures and graphs, reviewed and edited by I.S. The final reviews were done by A.S. with assistance and support from F.A.

Corresponding authors

Correspondence to Azza Ahmed or Isam Salih.

Ethics declarations

Competing interests

The authors declare no competing interests.

Additional information

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

The original online version of this Article was revised: The original version of this Article contained an error in Affiliation 5, which was incorrectly given as ‘Physics Department, College of Science, Imam Mohamed Ibn Saud Islamic University (IMSIU), P. O. Box 11642, Riyadh, Saudi Arabia’. The correct affiliation is ‘Physics Department, College of Science, Imam Mohammad Ibn Saud Islamic University (IMSIU), P. O. Box 11642, Riyadh, Saudi Arabia’.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and Permissions

About this article

Verify currency and authenticity via CrossMark

Cite this article

Ahmed, A., Baldo, A., Sulieman, A. et al. Optimisation of T2 and T2* sequences in MRI for better quantification of iron on transfused dependent sickle cell patients. Sci Rep 11, 8513 (2021). https://doi.org/10.1038/s41598-021-88116-8

Download citation

Comments

By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.

Search

Quick links

Nature Briefing

Sign up for the Nature Briefing newsletter — what matters in science, free to your inbox daily.

Get the most important science stories of the day, free in your inbox. Sign up for Nature Briefing