Abstract
Bacterial culture and drug susceptibility testing are used to identify pathogen infections. Nevertheless, the process requires several days from collection to the identification of bacterial species and drug-resistance patterns. The digital PCR system is a rapidly developing quantitative detection technology widely applied to molecular diagnosis, including copy number variations, single nucleotide variant analysis, cancer biomarker discovery, and pathogen identification. This study aimed to use a droplet digital PCR system to identify bacteria in blood samples and explore its ability to identify pathogen in bacteremia. Then, we designed primers and probes of SWG-9 and COA gene for E. coli and S. aureus to identify in blood samples with the ddPCR system. The system had demonstrated extremely high detection accuracy in blood samples, and the detection rate of E. coli was 13.1–21.4%, and that of S. aureus was 50–88.3%. Finally, blood samples containing both E. coli and S. aureus were tested to evaluate further the accuracy and applicability of this method, indicating the detection rates range from 18.1% to 97%. The ddPCR system is highly promising as a qualitatively and quantitatively screening method for rapidly detecting pathogen.
Similar content being viewed by others
Introduction
Despite economic development, with health systems in various countries worldwide continuously improving, infectious diseases are still one of the main threats to human health. The 2018 Global Burden of Disease Study reported that 7.069 million people died from infectious diseases worldwide, accounting for 12.6% of the global deaths, and the mortality rate was 97/1000001. Infectious diseases remain a significant public health problem. Early diagnosis and rational application of antibiotics are essential for significance for the management of infectious diseases. At present, the clinical use of antibiotics depends on empirical treatment. Also, the over-use of antibiotics increases the resistance of bacteria and drug-induced diseases and consumes substantial health care resources2. Studies have reported that the annual cost of treatment for infectious diseases was about $1.5–2 billion3. One of the reasons was that the pathogenic microorganisms could not be quickly and accurately identified in the early stage of infections, leading to irrational use of antibiotics and other medical treatment measures.
The most common pathogenic bacteria in general surgery departments are Escherichia coli, Staphylococcus aureus, Pseudomonas aeruginosa, Klebsiella pneumoniae, and Acinetobacter baumannii4. Traditional bacterial detection and identification methods are based on bacterial culture and biochemical identification, and testing usually requires several days, even to one week. As a gold standard, the diagnosis of bacterial infectious diseases often relies on the specific detection of pathogenic bacteria. The use of antibiotics must be based on the bacterial culture results and drug sensitivity test results in clinical diagnosis and treatment. However, due to detection time, bacterial culture growth characteristics, non-pathogenic bacteria contamination, and improper material selection, the test results in actual work may generate false negatives or false positives. Incorrect identification results not only in delayed treatment, wasting medical resources, but in mistaken diagnosis and treatment. For these reasons, rapid, accurate identification of pathogenic microorganisms has become an intense focus of research.
In recent years, molecular biology techniques have been applied to detect pathogenic microorganisms. They have attracted attention due to their high specificity, lower time-requirement, and reduced incidence of cross-infection. These techniques include Enzyme-Linked Immunosorbent Assay2, Polymerase Chain Reaction5 and Gene microarray chips6. To date, Polymerase Chain Reaction technology has been widely used in many research fields because of its simple, intuitive, economical, and rapid detection characteristics7. Digital polymerase chain reaction (dPCR) was first proposed by Kenneth Kinzler and Bert Vogelstein in 1999. This is a PCR technology that truly achieves absolute quantification after qPCR technology8. Droplet digital PCR (ddPCR) is a new technology that enables absolute quantification of nucleic acids with high analytical sensitivity and precision. ddPCR splits PCR reagents into tens of thousands of nanoliter or picoliter partitions by a microfluidic chip so that each droplet contains 0 or 1 DNA template. After PCR amplification and fluorescence detection, the target nucleic acids are calculated from the number of positive and negative droplets by Poisson statistics9. Compared with traditional qPCR, digital PCR has higher sensitivity, specificity, and accuracy. It plays an essential role in many fields, including early diagnosis of tumor markers10, gene expression product analysis11, food safety testing12, pathogenic microorganism testing13, common genetic disease testing, and non-invasive prenatal diagnosis14.
In the present study, we designed primers and probes of SWG-9 and COA gene for E. coli and S. aureus. Then, ddPCR was applied to test the advantage and reliability for identifying E. coli and S. aureus in simulated bacteremia blood samples to provide the theoretical basis for subsequent clinical sample testing.
Results
Assay design and identification of bacterial strains using the ddPCR system
To examine the method's dynamic range and detection limit, serial dilutions of E. coli and S. aureus DNAs were subjected to ddPCR. Bacterial DNA was quantified using Qubit 3.0, and the estimated copy number was calculated. Then, we diluted the bacterial DNA templates in different concentrations.
The ddPCR-based bacterial strains detection workflow mainly consisted of five steps: sample preparation, preparation of reaction mixture, droplet generation, PCR amplification, and fluorescence detection and data analysis. For strains of bacteria, and the number of bacterial nucleic acids can be detected simultaneously. For NC, all experiments performed in this study showed no positive signal in either the FAM or the VIC channel, indicating no contamination in the ddPCR system. To examine the dynamic range of the method, serial dilutions of E. coli or S. aureus were quantified by ddPCR, and the results of three replicates were analyzed (Fig. 1, S3; Tables 1, 2, S11, S12).
The ddPCR system identifies various E. coli ATCC25922 nucleic acid (Fig. 1A, corresponding to A05 to H05). The ddPCR signals were linear over the range from 0 to 105 copies/μL with R2 = 0.9995 for all sites. The actual copy number of bacteria from A05 to E05 measured using the ddPCR system (Fig. 1A Table 1). The ddPCR system identified various S. aureus ATCC25923 nucleic acid (Fig. 1B, corresponding to A01–H01), and the measured number of bacterial nucleic acid templates were very close to the theoretical value. The ddPCR signals were linear over the range from 0 to 104 copies/μL with R2 = 1 for all sites. The actual copy number of bacteria nucleic acid templates from A01 to C01 measured using the ddPCR system (Fig. 1 B, Table 2).
We compared the lowest detection limit of ddPCR and real-time quantitative PCR in different numbers of bacterial nucleic acid templates. When the number of nucleic acid templates is less than 100, the real-time quantitative PCR system cannot produce a useful amplification curve (Figure S1, S2; Table S8–S10).
The specificity of ddPCR for simultaneously identified two bacterial strains
To validate the specificity of this method, two groups of ddPCR amplification experiments, including E. coli nucleic acid and S. aureus nucleic acid, were performed in triplicates for each detection site. In the two sets of experiments, the actual DNA copy numbers of ATCC25922 and ATCC25923 were very close to the theoretical values, and there was almost no interference between them. The detection rate of bacteria was as high as 80% or even 90% (Fig. 2, Tables 3 and 4).
Bacteria count
We added E. coli and S. aureus to 50 ml of Luria–Bertani bacterial culture medium and placed it in an orbital shaker at 37 °C overnight. We then diluted the bacterial solutions 106 times and performed a rolling ball scratching board. We calculated the number of colonies in the culture medium and found that the number of colonies of E. coli ATCC25922 was 107, and the number of colonies of S. aureus ATCC25923 was 70 (Fig. 3). Conversion is made according to the bacterial solution's dilution factor. The concentration of the E. coli ATCC25922 bacterial solution was 1.25 × 109 cells/ml, and the concentration of S. aureus ATCC25923 bacterial solution was 6.7 × 108 cells/ml.
Detection of simulated bacteria using the ddPCR system
We used the ddPCR system to identify bacterial nucleic acid extracted from the simulated bacteremia blood samples. We counted the number of bacteria by counting the number of nucleic acid templates. The number of bacteria detected using the ddPCR system and the theoretical number of bacteria are listed in Tables 5 and 6. We took the average number of bacteria detected in the three groups and calculated the logarithm of the average number of bacteria to perform fitting analysis (Fig. 4, ATCC25922: R2 = 0.9998, ATCC25923: R2 = 0.9996). The E. coli ATCC25922 nucleic acid detection rate in simulated bacteremia blood samples was 13.1–21.4%, and the S. aureus ATCC25923 nucleic acid detection rate in simulated bacteremia blood samples was 50–88.3%.
We used the ddPCR system to measure bacterial nucleic acid extracted from the simulated bacteremia blood samples contaminated by E. coli and S. aureus simultaneously (Table 7 and 8). We took the logarithm of the measured number of bacteria to make a fitting curve (Fig. 5, ATCC25922: R2 = 0.9532, ATCC25923: R2 = 0.9958). The bacterial nucleic acid detection rate of E. coli ATCC25922 was 18.1–77%, and that of S. aureus ATCC25923 was 44.9–97%.
Discussion
The specific genes of bacterial strains were selected according to the GenBank database. Nandakafle and Brozel first discovered the SWG-9 UIDA gene in 2015. The majority of E. coli strains carry the SWG-9 gene, which contains 566 bases and encodes a protein product containing 188 amino acids. So we designed a primer and MGB probe for this gene to detect E. coli. Also, Most S. aureus carries the Staphylococcus coagulase gene, which can encode protein products of 647 amino acids, and this gene plays a vital role in the pathogenic process of S. aureus15. Therefore, we selected the conserved sequence of this gene in the experiment to design PCR primers and probes to identify strains.
Furthermore, in this study, the sensitivity and accuracy of the ddPCR system have obvious advantages compared with real-time fluorescent quantitative PCR. The ddPCR system retains high sensitivity and accuracy when simultaneously detecting two pathogenic microorganisms. Also, the ddPCR system performs absolute quantification without the need for a standard curve. The detection rate of simulated bacteremia blood samples using the ddPCR system is lower than that of purified bacteria nucleic acid samples, mainly because of the bacterial nucleic acid extraction rate in simulated bacteremia blood samples. When the bacterial nucleic acid is extracted, human blood, EDTA in the anticoagulation tube, whether the cell walls are easily digested and cleaved by enzymes, and the elution and enrichment of nucleic acid, various operating processes and factors have a negative impact during the bacterial nucleic acid extraction. Even so, the rate of extracting bacterial nucleic acid from bacterial blood samples by the ddPCR system was between 15 and 80%.
We found that when the concentration of bacteria in the blood was greater than ten cells/ml, the ddPCR system detected bacteria stably. At present, clinical laboratories determine whether bacteremia is present based on the blood culture instruments reporting positive, and the reporting time is about 6–18 h. Generally speaking, it takes about three days from the collection of blood samples to the completion of bacterial culture and antibiotic sensitivity testing. Suppose we design unique probes and primers for specific bacterial and drug-resistance genes. In that case, the ddPCR system can identify bacterial species and drug-resistance genes simultaneously, with detection times less than 4 h. Therefore, our research suggests that the ddPCR technical platform is superior for the identification of pathogenic microorganisms. The ddPCR system accurately and effectively detect pathogenic microorganisms, allowing for more targeted antibiotic therapy and minimizing misuse of antibiotics, drug toxicity, and drug resistance, all of which could reduce economic burdens on patients.
Currently, the coronavirus (COVID-19) epidemic is spreading all over the world. The pathogenic microorganism of this pneumonia has been identified as a new coronavirus (Severe Acute Respiratory Syndrome Coronavirus 2, SARS-CoV-2). In acute respiratory infections, reverse transcription-polymerase chain reaction is usually used to detect pathogenic viruses from respiratory secretions in nucleic acid testing. It is the "gold standard" for the diagnosis of related cases16. However, the RT-PCR test result can also produce false negatives17. The reasons are the time of specimen collection, the quality or type of specimen, the transportation of the specimen, the ability of the virus to mutate, and the low sensitivity of RT-PCR and polymerase chain reaction inhibition18,19,20. Therefore, an efficient, fast, and accurate laboratory diagnosis technology is the key to ensuring early diagnosis, timely treatment, and preventing the development of the epidemic. Based on its advantages, digital PCR provides a new, efficient, and accurate detection method for pathogenic microorganisms such as viruses and bacteria detection.
Methods
Bacterial strains, experimental reagents, and instruments
The bacterial strains used were provided by the Laboratory of Peking University People's Hospital (Table S1), and the reagents and instruments used were provided by the laboratory of Tsinghua University (Tables S2 and S3).
Extraction of bacterial nucleic acid
The bacterial strains were propagated and cultured in a 50 ml Luria–Bertani bacterial culture medium. A total of four strains of bacteria were used in the study, S. aureus ATCC25923, S. aureus ATCC29213, E. coli ATCC25922, and a strain of E. coli named E. coli-clin extracted from a patient. According to the methods suggested in the manufacturer's instructions, nucleic acids were extracted from the four bacteria using the QIAamp cador Pathogen Mini Kit (QIAGEN, Germany). We used Nanodrop to measure the extracted nucleic acid concentrations and stored samples at –20 °C.
Design of primers and probes
The specific genes of bacterial strains were selected according to the GenBank database. The specific genes of E. coli and S. aureus were SWG-9 and coagulase COA gene, respectively. Primers and probes were designed for conserved regions in SWG-9 and coagulase COA gene sequences (Table S4 and S5). We used Oligo 7.0 and Primer Express 3.0 software to design the primers and probes for the nucleic acid sequence containing the site to be detected. The sequences were delivered to Thermo Fisher Scientific (CN) to synthesize the primers and probes. The 5 'end of the E. coli probe was labeled with a 6-FAM fluorescent molecule, the 5′ end of the S. aureus probe was labeled with VIC fluorescent molecule.
Workflow of the ddPCR system
According to the manufacturer's instructions, we carried out ddPCR using the QX200 ddPCR system (Bio-Rad, CA). The QX200 ddPCR system (Bio-Rad, CA) consists of four steps: (1) preparation of the reaction mixture using a ddPCR amplification volume 20 µL (Table S6); (2) droplet generation; (3) ddPCR amplification (Table S7); and (4) droplet reading and analysis of results using QuantaSoft 3.0 software.
Detection of bacterial strains using a real-time quantitative PCR system
The real-time quantitative PCR system with the MGB probe system was used to quantify the gradient-diluted nucleic acid of E. coli standard strain (ATCC25922) to determine the sensitivity range of the two real-time quantitative PCR. The template dilution gradient (copies/20 µL) of ATCC25922 was 106, 105, 104, 103, 102, 101, 1, and 0. Each group of experiments was repeated three times in parallel.
Detection of bacterial strains by ddPCR system
We used the ddPCR system to quantify the gradient-diluted nucleic acid of E. coli (ATCC25922) to determine the sensitivity range of the ddPCR system. The template dilution gradient (copies/20 μl) of ATCC25922 were 105, 104, 103, 102, 25, 6, and 0. Each group of experiments was repeated three times in parallel.
The ddPCR system simultaneously detects two bacterial strains
Various numbers of DNA templates, specific primers, and probes (ATCC25922: 5′-FAM/3′MGB, ATCC25923: 5′-VIC/3′MGB) of the two bacterial strains were added to a ddPCR reaction system (Group A, bacteria counts of E. coli: bacteria counts of S. aureus = 10:1, Group B, bacteria counts of E. coli: bacteria counts of S. aureus = 1:10). We counted DNA template numbers of E. coli and S. aureus and calculated the detection rate of the two bacteria.
Detection of blood samples of a simulated bacteremia blood sample using the ddPCR system
The blood samples were obtained from the Department of Gastroenterological Surgery, Peking University People’s Hospital. All of the blood sample donors included in this study provided their informed consent. The study was approved by the Human Ethics Committee of the Peking University People’s Hospital (China). The blood collection tube was pretreated with EDTA anticoagulation. Three simulations of bacteremia blood samples were prepared; Group A: 100 μL of the serially-diluted suspensions of E. coli (ATCC25922) were injected into 400 μL of blood treated with EDTA. Group B: 100 μL of the serially-diluted suspension of the S. aureus strain (ATCC25923) was injected into the 400 μL blood treated with EDTA. Group C: 100 µL of the ATCC25922 and ATCC25923 suspensions were diluted into gradients, then added to the blood samples with 400 µL EDTA. Extraction of nucleic acid from bacterial blood specimens was performed as follows: nucleic acid from bacterial blood specimens was extracted using a bacterial nucleic acid kit (Qiagen, Germany). We used 50 µl of solution buffer to elute DNA from the column and stored it at 80 °C. The extracted nucleic acid was detected using the ddPCR system. Repeat the test three times in parallel.
Statistical analysis
The detection results were statistically analyzed using SPSS 22.0, GraphPad Prism 5, and OriginPro 2016 software, and the linear dynamic range of the method was measured by R2 value.
Ethics approval and consent to participate
The present study was approved by the Human Ethics Committee of the Peking University People’s Hospital (China) and written informed consent was obtained from all subjects. Methods were carried out in “accordance” with the approved guidelines.
References
Roth, G. A. et al. Global, regional, and national age-sex-specific mortality for 282 causes of death in 195 countries and territories, 1980–2017: a systematic analysis for the Global Burden of Disease Study 2017. The Lancet 392(10159), 1736–1788 (2018).
Alekseeva, L. P. et al. Comparative Assessment of Dot-Immunoassay and Immunochromatography Methods for Detection of 01 Serogroup of Vibrio Cholerae. Zh. Mikrobiol. Epidemiol. Immunobiol. 6, 88–93 (2010).
Sun, B. Nosocomial Infection in China: Management Status and Solutions. Am. J. Infect. Control. 44, 851–852 (2016).
Shan, K. et al. Evaluation of resistance by clinically pathogenic bacteria to antimicrobials and common disinfectants in Beijing, China. . J. Nippon Med. Sch. 85, 302–308 (2018).
Blais, B. W. & Martinez-Perez, A. A simple PCR-based macroarray system for detection of multiple gene markers in the identification of priority enterohemorrhagic Escherichia Coli. J Food Prot. 74, 365–372 (2011).
Herbert, M. A. et al. Genetic islands of streptococcus agalactiae strains NEM316 and 2603VR and their presence in other group B streptococcal strains. BMC Microbiol. 5, 31 (2005).
Saiki, R. K. et al. Enzymatic amplification of beta-globin genomic sequences and restriction site analysis for diagnosis of sickle cell anemia. Science 230, 1350–1354 (1985).
Vogelstein, B. & Kinzler, K. W. Digital PCR. Proc Natl Acad Sci U S A. 96, 9236–9241 (1999).
Pinheiro, L. B. et al. Evaluation of a droplet digital polymerase chain reaction format for DNA copy number quantification. Anal. Chem. 84, 1003–1011 (2012).
Oehler, V. G. et al. Absolute quantitative detection of ABL tyrosine kinase domain point mutations in chronic myeloid leukemia using a novel nanofluidic platform and mutation-specific PCR. Leukemia 23, 396–399 (2009).
White, A. K. et al. High-throughput microfluidic single-cell RT-qPCR. Proc Natl Acad Sci U S A. 108, 13999–14004 (2011).
Corbisier, P., Bhat, S., Partis, L., Xie, V. R. & Emslie, K. R. Absolute quantification of genetically modified MON810 maize (Zea Mays L.) by digital polymerase chain reaction. Anal. Bioanal. Chem. 396, 2143–2150 (2010).
Strain, M. C. et al. Highly precise measurement of HIV DNA by droplet digital PCR. PLoS ONE 8, e55943 (2013).
Lee, S. Y. et al. A new approach of digital PCR system for non-invasive prenatal screening of trisomy 21. Clin. Chim. Acta. 476, 75–80 (2018).
McAdow, M., Missiakas, D. M. & Schneewind, O. Staphylococcus aureus secretes coagulase and Von Willebrand factor binding protein to modify the coagulation cascade and establish host infections. J. Innate Immun. 4, 141–148 (2012).
Xu, Y. et al. Significance of serology testing to assist timely diagnosis of SARS-CoV-2 infections: implication from a family cluster. Emerg Microbes Infect. 9, 924–927 (2020).
Sethuraman, N., Jeremiah, S. S. & Ryo, A. Interpreting diagnostic tests for SARS-CoV-2. JAMA 323, 2249–2251 (2020).
Kucirka, L. M., Lauer, S. A., Laeyendecker, O., Boon, D. & Lessler, J. Variation in false-negative rate of reverse transcriptase polymerase chain reaction-based SARS-CoV-2 tests by time since exposure. Ann. Intern. Med. 173, 262–267 (2020).
Caruana, G. et al. Diagnostic strategies for SARS-CoV-2 infection and interpretation of microbiological results. Clin Microbiol Infect. 1, 1 (2020).
Schrader, C., Schielke, A., Ellerbroek, L. & Johne, R. PCR inhibitors—occurrence, properties and removal. J. Appl. Microbiol. 113, 1014–1026 (2012).
Acknowledgements
We thank members of our laboratory for helpful discussion and Professor Guo Yong of Tsinghua University, Co, for the support of bioinformatics analysis.
Author information
Authors and Affiliations
Contributions
Z.L., K.J., T.Z., F.W., Z.C., J.X., and Z.N. conceived and designed the study. T.Z. and Z.N. wrote the manuscript. Z.L. and K.J. reviewed and edited the manuscript. All authors reviewed the manuscript.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Zhang, T., Niu, Z., Wu, F. et al. Qualitative and quantitative detection of surgical pathogenic microorganisms Escherichia coli and Staphylococcus aureus based on ddPCR system. Sci Rep 11, 8771 (2021). https://doi.org/10.1038/s41598-021-87824-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-87824-5
This article is cited by
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.