Abstract
Needle artifacts pose a major limitation for MRI-guided interventions, as they impact the visually perceived needle size and needle-to-target-distance. The objective of this agar liver phantom study was to establish an experimental basis to understand and reduce needle artifact formation during MRI-guided abdominal interventions. Using a vendor-specific prototype fluoroscopic T1-weighted gradient echo sequence with real-time multiplanar acquisition at 1.5 T, the influence of 6 parameters (flip angle, bandwidth, matrix, slice thickness, read-out direction, intervention angle relative to B0) on artifact formation of 4 different coaxial MR-compatible coaxial needles (Nitinol, 16G–22G) was investigated. As one parameter was modified, the others remained constant. For each individual parameter variation, 2 independent and blinded readers rated artifact diameters at 2 predefined positions (15 mm distance from the perceived needle tip and at 50% of the needle length). Differences between the experimental subgroups were assessed by Bonferroni-corrected non-parametric tests. Correlations between continuous variables were expressed by the Bravais–Pearson coefficient and interrater reliability was quantified using the intraclass classification coefficient. Needle artifact size increased gradually with increasing flip angles (p = 0.002) as well as increasing intervention angles (p < 0.001). Artifact diameters differed significantly between the chosen matrix sizes (p = 0.002) while modifying bandwidth, readout direction, and slice thickness showed no significant differences. Interrater reliability was high (intraclass correlation coefficient 0.776–0.910). To minimize needle artifacts in MRI-guided abdominal interventions while maintaining optimal visibility of the coaxial needle, we suggest medium-range flip angles and low intervention angles relative to B0.
Introduction
Magnetic resonance imaging (MRI) demonstrates excellent characteristics for image guidance of interstitial interventional procedures, including the lack of ionizing radiation, high soft tissue contrast, and multiplanar needle guidance with simultaneous acquisition of axial, coronal, and sagittal image data sets in near real-time1,2,3,4,5,6,7,8. MRI-guided interventions are routinely performed in a broad range of organs, with a primary focus on body regions in which MRI is superior to computed tomography (CT) and ultrasound such as spine, prostate, and liver1,9,10,11,12,13. In particular, the combination of fast fluoroscopic T1-weighted gradient echo (GRE) sequences and hepatobiliary contrast agents has even enabled MRI-guided biopsy of small, CT-occult liver lesions with a diameter < 1 cm14.
To be MR-compatible and to assure a safe procedure for the patient and the performing radiologist, coaxial needles are made of alloys causing minimal interference with the outer magnetic field, e. g. nickel-titanium (Nickel Titanium Naval Ordnance Laboratory, Nitinol), titanium, glass fiber, or steel15. However, needle artifacts presenting as low intensity signal in the region of the coaxial needle cannot be completely excluded and hinder accurate visualization of the needle. In order to successfully perform MRI-guided interventions, a thorough understanding and knowledge of needle artifacts is crucial. The artifact size depends on orientation of the intervention device relative to the B0 field (intervention angle), the strength of the static magnetic field, diameter and alloy of the instruments, as well as on pulse sequence type and individual sequence parameters7,16,17,18. Despite being virtually inevitable, needle artifacts pose a major limitation for needle interventions, as they impact the visually perceived size of the needle and distance of the needle tip to the target during the procedure. Therefore, it is essential to limit needle artifacts to a minimum.
Technologic advances in magnet, protocol, coil, biopsy needle, and probe design have made MRI a clinically valuable image-guidance technique. Due to the recent, continuing innovations in targeting software, augmented reality, and compatible devices, it is essential to reassess technical and methodological fundamentals, such as susceptibility artifacts produced by metallic needles, in consideration of these advances. We here present a systematic investigation of artifact behavior of different commercially available MR-conditional Nitinol coaxial needles as a function of the intervention angle and sequence parameter variations in a vendor-specific prototype fluoroscopic T1-weighted GRE sequence with real-time multiplanar acquisition in a dedicated liver phantom at 1.5 Tesla (T). In addition, our study also adds to the field as not only selected but a broad range of clinically relevant parameters was evaluated under standardized and controlled conditions. The aim of the present study was to establish an experimental basis to understand and reduce artifact formation during clinical MRI-guided abdominal interventions in patients.
Results
Mean values and standard deviations of the artifact diameters of both readers are presented in Table 1.
Flip angle
Artifact size increased gradually with increasing FA for each needle diameter (16G: 12.1–16.1 mm, 18G: 11.9—15.4 mm, 20G: 6.7–9.8 mm, 22G: 5.8–8.5 mm; p = 0.002). Pairwise comparison did not show any significant difference between stepwise increased parameters. However, multiple testing revealed significant differences between 9 pairs of FA. For all 4 needles, significant and strong positive correlations between FA and artifact diameter (r between 0.816 and 0.965; p < 0.01) (Fig. 1) were observed. In addition to the central hypointense needle artifact, a hyperintense peripheral rim was observed at FA > 45°, which was included in the measurements.
Scan series 1. Mean values (of both readers) of the measured artifact diameters (a) of the 16G, 18G, 20G, 22G needles with different FA of the T1-weighted GRE sequence. Note the positive correlation of artifact diameter and FA (p ≤ 0.05) for 16G (r = 0.859), 18G (r = 0.816), 20G (r = 0.965), and 22G (r = 0.950) MR-compatible needles, in (b–k) the visual correlate is shown. Note the hyperintense rim around the central hypointense needle artifact occurring with FA > 45° representing as an alteration of the image contrast due to the needle. FA = flip angle, G = gauge.
Bandwidth
Modifying the receiver BW, we found no significant difference in artifact diameter for any of the different needles (16G: 12.2–12.7 mm, 18G: 10.8–12.0 mm, 20G: 8.4–9.0 mm, 22G: 5.6–6.7 mm; p = 0.304) and also no significant correlation (− 0.70 < r < 0.62, p > 0.19) (Fig. 2) between the artifact diameter and the BW.
Scan series 2. Mean values (of both readers) of the measured artifact diameters (a) of the 16G, 18G, 20G, 22G needles with different BWs of the T1-weighted GRE sequence. There was no significant correlation of artifact diameter with BW for 16G (r = 0.300), 18G (r = − 0.100), 20G (r = − 0.600), and 22G (r = − 0.700) MR-compatible needle, in (b–f) the visual correlate is shown. BW = bandwidth, Hz = Hertz, G = gauge.
Matrix
Artifact diameters differed significantly between the 4 matrix sizes for each needle diameter (16G: 10.9–12.7 mm, 18G: 9.5–11.6 mm, 20G: 6.8–8.6 mm, 22G: 5.8–7.9 mm; p = 0.002); however, pairwise comparisons did not demonstrate any significant differences excluding between the matrix sizes 96 × 96 and 192 × 192 (p = 0.004) as well as between 96 × 96 and 256 × 256 (p = 0.006). A significant negative correlation was found between the mean artifact diameter of both readers and the matrix for the 16G needle (r = − 0.993, p = 0.007). For the other needle sizes, there was no significant correlation (Fig. 3) between the artifact diameter and the matrix. In addition, increasing differentiation between the display of the “actual” needle and the surrounding artifact at larger matrices could be observed.
Scan series 3. Mean values (of both readers) of the measured artifact diameters (a) of the 16G, 18G, 20G, 22G needles with different matrix sizes of the T1-weighted GRE sequence. There was a negative correlation (p < 0.005) of artifact diameter with matrix for 16G (r = − 0.993) needles, but no significant correlation for 18G (r = − 0.905), 20G (r = − 0.611), and 22G (r = − 0.549) needles. In (b–e) the visual correlate is shown. Note the increasing differentiation between actual needle and artifact at larger matrices. G = gauge.
Slice thickness
Modifying slice thickness, we found no significant difference in artifact diameter for each needle size (16G: 11.8–12.3 mm, 18G: 10.2–11.3 mm, 20G: 6.5–7.4 mm, 20G: 5.2–5.5 mm; p = 0.191) and no significant correlation between artifact diameter and slice thickness for any of the needles either (− 0.787 < r < 0.662; p > 0.21).
Readout direction
For the 2 varying readout directions (right > left, anterior > posterior), no significant differences in artifact diameter were found according to Wilcoxon signed-rank test (p = 0.623) with generally similar artifact diameters for both read-out directions (Table 1).
Intervention angle
Artifact size increased gradually with increasing intervention angles for each needle diameter (16G: 5.8–15.2 mm, 18G: 6.3–13.3 mm, 20G: 5.0–9.3 mm, 22G: 3.9–7.6 mm; p < 0.001). Pairwise comparisons did not show any significance between the stepwise changes, exemplarily from 0° to 15° (p = 1.000) or from 60° to 75° (p = 1.000). Multiple testing revealed significant differences of the artifact diameter for 7 pairs of intervention angles. For all 4 needles, we observed a significant positive correlation (r between 0.960 and 0.980; p ≤ 0.001) with strong effect strength, which is shown in Fig. 4. In addition, a ball-like tip artifact was observed at small intervention angles (0° and 15°).
Scan series 6. Mean values (of both readers) of the measured artifact diameters (a) of the 16G, 18G, 20G, 22G needles with different intervention angles of the T1-weighted GRE sequence. Note the positive correlation of artifact diameter with intervention angle (p ≤ 0.05) for 16G (r = 0.980), 18G (r = 0.966), 20G (r = 0.960), and 22G (r = 0.972) MR-compatible needles; in (b–h) the visual correlate is shown. The upper row provides demonstration purposes of the used angulation. The lower row shows the sequence images used for artifact measurements. Note the ball-like tip artifact occurring at small intervention angle (0° and 15°). IA = intervention angle, G = gauge.
Needle diameter
The 4 needle diameters (16–22G) showed significant differences (p > 0.001) in artifact diameter (Fig. 5), which also proved to be significant (p < 0.05) for each pairwise comparison, except between 16 and 18G (p = 0.064).
Boxplot of artifact size depending on needle diameter. Medians (of both readers) of all measured artifact diameters of all modified parameters separated in 16G, 18G, 20G, 22G needles. Kruskal–Wallis test demonstrated a significant difference of artifact diameter (p < 0.001) for the 4 sizes of MR-compatible needles.
Interrater reliability
For all 4 needles, the intraclass correlation coefficients were between 0.776 and 0.910 (p < 0.001) indicating excellent interrater reliability (Table 2).
Discussion
In this phantom study, we investigated the influence of different sequence parameters of a vendor-specific prototype fluoroscopic T1-weighted GRE sequence with near real-time multiplanar acquisition and of the intervention angle on artifact behavior of commercially available, MR-compatible coaxial needles at 1.5 T. Needle artifacts in MRI are caused by several physical processes, but in particular by inhomogeneities of the static magnetic field (B0) experienced by the nuclei. The needle-induced B0 inhomogeneity stems from the geometric characteristics and the magnetic susceptibility of the object being imaged. These static field errors cause distortions of the spatial geometry, as well as intra-voxel dephasing. The latter is caused by local disturbances of B0 within a voxel resulting in severe intensity losses due to uncompensated intravoxel phase dispersion (T2* effect), which is particularly relevant for GRE sequences18,19. Other causes for needle artifacts are radio-frequency effects such as B1 enhancement20. We aimed to investigate these processes (summarized by the term “needle artifacts” in this study) in whole regarding their relevance for MR-guided abdominal procedures depending on easily modifiable sequence parameters and intervention angle. Our study adds to the field as it provides a comprehensive, systematic basis for future investigation and potential reduction of needle artifacts during clinical MR-guided abdominal interventions.
Increasing artifact diameters were observed with increasing FA. The hyperintense peripheral rim around the central hypointense needle artifact occurring at high FA (> 45°) was included in the measurement. It extends the area of potential misinterpretation of needle position and should be considered when choosing a high FA for T1-weighted GRE sequences. Another workgroup already investigated needle artifacts by modifying the FA in T1-weighted GRE sequences14. In line with our results, they did not report any relevant change of the artifact diameter in a stepwise comparison of different artifact diameters for FA < 45°. Only when comparing FA 45° to 55° and 55° to 65°, a significant increase in artifact diameter was demonstrated, while including the hyperintense zone in the measurement. Overall, the results are in accordance with our observation that optimal artifact diameters are found with FA < 45°14. Another aspect to consider is the signal-to-noise ratio (SNR) decreasing with increasing FA above the Ernst angle (αE) in GRE sequences21. It is known that the MR signal is maximized at αE, which can be calculated from the equation αE = arccos (exp(− TR/T1)). Using the presented real-time, T1-weighted GRE sequence at 1.5 T with a short TR of 5.3 ms, this would result in an αE of 10.7°, calculated for healthy liver parenchyma showing T1 relaxation time of ~ 300 ms after 10–20 min of Gd-EOB-DTPA administration22,23. In the case of cirrhotic or functionally impaired liver parenchyma, longer T1 and thus an even lower αE would have to be expected. In general, the αE is low in combination with the short TR used in our study, so that high as well as medium FAs may not lead to optimal SNR.
The modification of the BW showed no significant influence on the artifact diameter though BW has been mentioned in the literature as one of the crucial modification factors to minimize needle artifacts24. The theory behind this is, that by a metallic object of a given size and susceptibility, the Larmor frequencies of the hydrogen protons are altered to a certain amount25. Reducing the Hz/pixel ratio increases the number of pixels affected by the variance in frequencies and thus increases the size of the susceptibility artifact. Overall, it can be concluded that for MR-guided interventions with T1-weighted GRE sequences, BW variations did not show any significant effect on needle artifact size in the investigated range, which might have been too small to reveal significant differences. However, the BW range was chosen according to standard BW used in clinical routine.
For the various matrix sizes, a significant difference in artifact diameter was observed. In addition, a negative correlation between artifact diameter and selected matrix was found for the 16G needle. The other needles also showed smaller artifact diameters at higher matrix sizes, but the effect was not strong enough to be significant for the matrix sizes evaluated in this work. An increased matrix not only reduces the artifact diameter but also optimizes the image quality due to decreased voxel volume producing MR images that may show more small details. The matrix determines the spatial resolution and is therefore a quality feature of the acquired image data26. The improved differentiation of the actual needle and artifact shown in our scan series allows more exact biopsy of target lesions. These advantages of a smaller artifact diameter and a higher spatial resolution are offset by an increase in acquisition time, which is the crucial point that the matrix cannot be set as high as possible14.
As the matrix yields an influence on the voxel size, it is also impacted by slice thickness. If the slice thickness decreases, which is equivalent to a reduction of the voxel size, the field inhomogeneity within each individual voxel decreases. This results in lower artifact dependence and generates smaller needle artifacts26,27. In the present study, we could not observe any significant differences between the selected slice thicknesses with regard to artifact size. However, it must be noted that the differences in the extent of the change of the artifact diameter may have been too small.
The comparison of the readout directions (right > left, anterior > posterior) at an intervention angle of 45° revealed no significant differences with regard to artifact diameters. Our results are in accordance with previous studies. Lewin et al. compared parallel and perpendicular readout directions at the intervention angle of 90° and found no significant differences for GRE sequences; this is in contrast to spin echo (SE) and turbo spin echo (TSE) sequences, in which artifacts were more pronounced when the readout direction was perpendicular to the needle shafts28. Similarly, Frahm et al. reported no provable effect swapping the phase-encoding and frequency-encoding axes on GRE sequences at the intervention angle of either 0° or 90°, respectively26.
The intervention angle is closely related to the size of the resulting artifact18,29,30. In accordance, our results also showed the positive correlation of artifact size with increasing intervention angle relative to the B0 field. Using low susceptibility materials is advantageous as these may be used at higher angles. Frahm et al. investigated the relationship between magnetic field strength and intervention angle26. They demonstrated that at lower field strengths (0.2 vs. 1.5 T), the needle artifact growths less with increasing angle. At high field strengths, the artifact correlates closely with the increasing angle relative to the B0 field26. In addition, a ball-like tip artifact at the tip of the needle was observed at the low intervention angles (0°–10°). As described by the working group of Liu et al., the magnetic field is most strongly influenced in the area of the needle tip31. This is particularly noticeable in needles with lower magnetic susceptibility, such as carbon fiber or titanium, compared to chromium, cobalt, or nickel, respectively. This three-dimensional tip artifact is significantly larger than the rather small artifact along the needle body at small angles and extends in all directions14,32. As already shown in previous works14,29, the influence of the needle diameter can be observed in each individual parameter tested and also in the evaluation of all measurements performed: Larger needles show increased artifact sizes, but the most significant difference was observed between 18 and 20G.
With regard to clinical practice, we recommend the lowest reasonably achievable intervention angles relative to the B0 field, even if repositioning of the patient becomes necessary. Furthermore, we suggest medium-range FA (30°–45°) to minimize the hypointense needle artifact and to avoid the hyperintense rim in the artifact periphery, which exclusively occurred at high FA in our scan series.
We acknowledge several limitations to the present study. First, imaging was performed under standardized laboratory conditions, as the phantom was fixed, motionless, and provided a high background signal intensity that facilitated the depiction of signal voids. In vivo, lower artifact contrast must be expected, as well as organ movement due to diaphragmal excursions during breathing. In addition, the clinical employment of higher FA may be limited by high specific absorption rates (SAR) and strong signal saturation, potentially hindering visibility of the target structures. Second, the evaluation of the artifact diameter was performed by manual measurements. Automatic artifact measurements would minimize the potential impact of the reader. However, high interreader agreement was observed in the present study. Third, we only investigated a single alloy (Nitinol). Different alloys may yield different artifact properties and should be investigated in further studies. Fourth, the phantom study was performed at a single field strength (1.5 T). For other organ regions, such as the prostate gland, higher field strengths (3 T) will need to be investigated. Fifth, when modifying the parameters in our evaluation, we tried to avoid changing another parameter at same time in order to not have any additional confounding variables. However, BW was not modified separately but only coupled to TR due to the fact that the minimum TR was systematically chosen in our experimental setting. One consequence of that choice could actually be misleading as the SNR increases with increasing TR. In addition, in a static phantom, images do not suffer from additional motion artifact with increasing image acquisition time (with increasing TR). Sixth, low artifact size in some instances does not necessarily mean that the “true” position of the needle is better known, as it is extremely difficult to be sure about the exact needle position from the real-time MRI visualization. Further studies are needed here, e.g., using a coordinate/navigator system to determine and verify the actual needle position and particularly the exact position of the needle tip in longitudinal direction along the needle axis. However, several issues remain to be resolved here, such as integration of, e.g., a coordinate system into the agar phantom and minimizing errors of measurement using the 2D T1 GRE sequence, taking into consideration that the distance between the grid and the needle will never be zero. Yamada et al. recently made an attempt in this direction, as they applied real-time US imaging fused with reformatted static MR images and coordinate registration for needle guidance during MR-guided percutaneous tumor ablations, with a mean targeting error of 1.6 ± 0.6 mm33. However, to date there is a lack of similar studies using real-time MR sequences. Seventh, we performed 2-dimensional (D) artifact measurements. However, it is acknowledged that in 3D image-guided interventions, needle artifact volume may be a relevant additional parameter. Though, depending on the sequence acquisition, such an assessment may be not adequately feasible, and several recent studies were also conducted in 2D34,35. Nevertheless, artifact volume and therefore possible differences to our measurement method should be investigated in future studies. Eighth, individually iterated the optimal parameters in our scan series, the combination of these settings and their interaction with regard to artifact formation needs to be investigated in further studies and in vivo.
Conclusion
Low intervention angles relative to the B0 field and medium-range FAs minimized the needle artifacts while maintaining optimal visualization of the intervention needle in this phantom study at 1.5 T.
Methods
This phantom study did not require institutional review board approval.
Image acquisition
MRI was performed on a closed-bore clinical 1.5 T MRI scanner (Magnetom Aera, Siemens Healthineers, Erlangen, Germany) with a short open bore design (system length cover-to-cover 145 cm, bore diameter 70 cm). The gradient system had a maximum gradient strength of 45 mT/m and a slew rate of 200 T/m/s. The MRI protocol was based on a prototype real-time fluoroscopic T1-weighted GRE sequence (Siemens WIP package ASP 1075G “Needle AutoAlign”, Siemens Healthineers, Erlangen, Germany). This prototype package allows for visual real-time update and interactive graphical modification of the slice geometry during imaging. A 6-channel body coil (Siemens Healthineers, Erlangen, Germany) with a weight of 1.4 kg (322 mm × 533 mm × 70 mm) was used as receive coil.
Agar phantom
The dedicated liver phantom was created by mixing 5 L of water with 100 g sodium chloride and 100 g agar–agar. The suspension was poured into a polyvinyl chloride (PVC) container and the subsequent curing process occupied 24 h at room temperature (21 °C). The agar phantom had a viscosity and signal intensity similar to that of the human liver parenchyma, as previously described36,37. The course of the needle could easily be followed from the outside due to the corresponding transparency. The repeated production of the phantom was not necessary as the complete scan series could be performed using the same model, assuring comparability of the results. An artifact-free wooden needle holder device with predetermined small openings was used for the exact and parallel positioning of the intervention needles in the B0 field and held the needles in a stable position. The experimental setup is shown in Fig. 6.
Experimental setup. Agar phantom with needle holder device (a): The phantom was composed of water, agar–agar, and sodium chloride and cured in a polyvinyl chloride container. Coaxial MR-compatible needles with different diameters (b) were introduced in a needle holder device firmly attached to the polyvinyl chloride container at a 45° angle. To evaluate the individual intervention angles to the B0 field, the phantom was tilted towards the B0 field for the intervention angles 30°, 15° and 0° and in the opposite direction for the angles 60°, 75° and 90°.
MR-compatible intervention needles
We investigated 4 different commercially available MR-compatible coaxial Nitinol needles (ITP Innovative Tomography Products GmbH, Bochum, Germany) sized 16G (outer diameter 1.60 mm, length 140 mm, trocar cut), 18G (outer diameter 1.25 mm, length 150 mm, standardized facet cut), 20G (outer diameter 0.90 mm, length 150 mm, standardized facet cut), and 22G (outer diameter 0.70 mm, length 100 mm, standardized facet cut).
Scan series
The phantom was positioned in the isocenter of the XZ plane using the light visor of the magnetic resonance tomograph. The influence of the following parameters on artifact formation was investigated: flip angle, receiver bandwidth, matrix, slice thickness, read-out direction, and intervention angle (i. e. the needle angle relative to the B0 field). As one parameter was modified, the others remained constant in the following predefined setting: The matrix was fixed to 128 × 128 voxels, which was a compromise between acquisition time and spatial resolution. Echo time (TE) and repetition time (TR), which yield an influence on acquisition time, were set to a minimum (TE 2.35 ms, TR 5.3 ms) as fixed parameters, resulting in an acquisition time of 0.68 s per plane. The field of view (FOV) was uniformly set to 300 × 300 mm2. Image acquisition was performed at a 45° angle relative to the B0 field as fixed intervention angle. The predefined setting was 15° for the flip angle (FA), 500 Hz/pixel for the bandwidth (BW), and 10 mm for the slice thickness, respectively. The fixed read-out direction was anterior to posterior. A detailed overview of these fixed parameters of predefined setting is provided in Table 3.
Starting with these default settings, each of the parameters mentioned above was modified, as described in detail in Table 4, resulting in acquisition times of 0.68–1.53 s per plane. For each parameter modification, the T1-weighted real-time sequence was performed in the same manner. The needle holder was attached to the phantom at a 45° (default setting) angle to the orientation of the B0 field extending longitudinally into the bore. To evaluate the individual intervention angles to the B0 field, the whole phantom was tilted towards the B0 field for the intervention angles 30°, 15°, and 0°, as well as in the opposite direction for the angles 60°, 75°, and 90°, respectively. Prior to the start of the fluoroscopic T1-weighted GRE sequence, the correct angle (accepted deviation of 0–2°) between the needles and the B0 field and the positioning in the isocenter of the MR imager was verified by test sequences.
Artifact diameter measurement
For image acquisition and evaluation of the artifact diameter, the software syngo Studio VB36E (Siemens Healthineers, Erlangen, Germany) was used. To ensure comparability of the individual artifact diameter, artifact diameter was measured in standardized planes at 2 defined positions: in 15 mm distance from the perceived needle tip and at 50% of the total needle length. The artifact diameter was determined for each of the 4 different coaxial MR-compatible needles for every modification of the scan series. In total, we implemented 2 measurements per needle artifact, which were performed by 2 independent and blinded readers (with 2 and 6 years of diagnostic MR imaging experience) for each modification of the evaluated parameters.
Statistical analysis
Statistical analysis was performed using dedicated statistics software (SPSS version 26, SPSS Inc., Chicago, IL). For descriptive statistics, the numerical values are presented as mean values plus standard deviation at 95% confidence intervals. To evaluate differences between the modified sequence parameters in the related samples, we used the Wilcoxon signed-rank test in case of 2 values of the modified parameter and in case of > 2 values the Friedman test including post hoc testing and Bonferroni multiple testing correction. Furthermore, we evaluated possible positive or negative correlations between the values of the modified parameters and the size of artifact diameter. For this purpose, we calculated the correlation coefficient Bravais–Pearson, which was tested for significance on both sides. In order to assess the significance of the results, the effect strengths r of the correlation coefficient Bravais–Pearson was additionally presented using Cohen's classification (r = 0.10 weak effect, r = 0.30 medium effect, r = 0.50 strong effect). To determine differences between the 4 diameters of MR-compatible intervention needles as unrelated samples, we used the Mann–Whitney U test. To measure the interrater reliability between the 2 blinded readers, we calculated the intraclass correlation coefficient (ICC). For all tests, p < 0.05 was considered significant.
Data availability
All relevant data are provided in the manuscript.
References
Winkelmann, M. T. et al. MR-guided high-power microwave ablation in hepatic malignancies: Initial results in clinical routine. Cardiovasc. Intervent. Radiol. 43, 1631–1638. https://doi.org/10.1007/s00270-020-02584-5 (2020).
Elfatairy, K. K., Filson, C. P., Sanda, M. G., Osunkoya, A. O. & Nour, S. G. In-bore MRI-guided prostate biopsies in patients with prior positive transrectal us-guided biopsy results: Pathologic outcomes and predictors of missed cancers. Radiol. Imaging Cancer 2, e190078. https://doi.org/10.1148/rycan.2020190078 (2020).
Winkelmann, M. T. et al. MRI-guided percutaneous thermoablation in combination with hepatic resection as parenchyma-sparing approach in patients with primary and secondary hepatic malignancies: Single center long-term experience. Cancer Imaging 20, 37. https://doi.org/10.1186/s40644-020-00316-z (2020).
Weiss, J. et al. Feasibility, efficacy, and safety of percutaneous MR-guided ablation of small ( mm) hepatic malignancies. J. Magn. Reson. Imaging 49, 374–381. https://doi.org/10.1002/jmri.26252 (2019).
Barkhausen, J. et al. White paper: Interventional MRI: Current status and potential for development considering economic perspectives, part 1: General application. Rofo 189, 611–623. https://doi.org/10.1055/s-0043-110011 (2017).
Fischbach, F. et al. Assessment of fast dynamic imaging and the use of Gd-EOB-DTPA for MR-guided liver interventions. J. Magn. Reson. Imaging 34, 874–879. https://doi.org/10.1002/jmri.22691 (2011).
Jungmann, P. M., Agten, C. A., Pfirrmann, C. W. & Sutter, R. Advances in MRI around metal. J. Magn. Reson. Imaging 46, 972–991. https://doi.org/10.1002/jmri.25708 (2017).
Bock, M. et al. Interventional magnetic resonance imaging: An alternative to image guidance with ionising radiation. Radiat. Prot. Dosimetry 117, 74–78. https://doi.org/10.1093/rpd/nci731 (2005).
Tempany, C., Straus, S., Hata, N. & Haker, S. MR-guided prostate interventions. J. Magn. Reson. Imaging 27, 356–367. https://doi.org/10.1002/jmri.21259 (2008).
Fischbach, F. et al. MR-guided freehand biopsy of liver lesions with fast continuous imaging using a 1.0-T open MRI scanner: Experience in 50 patients. Cardiovasc. Intervent. Radiol. 34, 188–192. https://doi.org/10.1007/s00270-010-9836-8 (2011).
Fritz, J. et al. Freehand real-time MRI-guided lumbar spinal injection procedures at 1.5 T: Feasibility, accuracy, and safety. AJR Am. J. Roentgenol. 192, W161-167. https://doi.org/10.2214/AJR.08.1569 (2009).
Weiss, C. R., Nour, S. G. & Lewin, J. S. MR-guided biopsy: A review of current techniques and applications. J. Magn. Reson. Imaging 27, 311–325. https://doi.org/10.1002/jmri.21270 (2008).
Dianat, S. S., Carter, H. B. & Macura, K. J. Magnetic resonance-guided prostate biopsy. Magn. Reson. Imaging Clin. N. Am. 23, 621–631. https://doi.org/10.1016/j.mric.2015.05.005 (2015).
Bauch, S. Evaluation MR kompatibler Nadeln und interaktiver Sequenzen zur interventionellen Bildgebung an einem offenen 1.0 Tesla MR- Tomographen (Panorama-HFO), Otto von Guericke University Library, Magdeburg, Germany, (2014).
Penzkofer, T., Peykan, N., Schmidt, K., Krombach, G. & Kuhl, C. K. How MRI compatible is “MRI compatible”? A systematic comparison of artifacts caused by biopsy needles at 3.0 and 1.5 T. Cardiovasc. Intervent. Radiol. 36, 1646–1657. https://doi.org/10.1007/s00270-013-0616-0 (2013).
Thomas, C. et al. In vitro assessment of needle artifacts with an interactive three-dimensional MR fluoroscopy system. J. Vasc. Interv. Radiol. 21, 375–380. https://doi.org/10.1016/j.jvir.2009.11.012 (2010).
Ludeke, K. M., Roschmann, P. & Tischler, R. Susceptibility artefacts in NMR imaging. Magn. Reson. Imaging 3, 329–343 (1985).
Ladd, M. E. et al. Biopsy needle susceptibility artifacts. Magn. Reson. Med. 36, 646–651 (1996).
Bakker, C. J., Bhagwandien, R., Moerland, M. A. & Fuderer, M. Susceptibility artifacts in 2DFT spin-echo and gradient-echo imaging: The cylinder model revisited. Magn. Reson. Imaging 11, 539–548. https://doi.org/10.1016/0730-725x(93)90473-q (1993).
Graf, H., Lauer, U. A., Berger, A. & Schick, F. RF artifacts caused by metallic implants or instruments which get more prominent at 3 T: An in vitro study. Magn. Reson. Imaging 23, 493–499. https://doi.org/10.1016/j.mri.2004.12.009 (2005).
Ernst, R. R. & Anderson, W. A. Application of Fourier transform spectroscopy to magnetic resonance. Rev. Sci. Instrum. 37, 93. https://doi.org/10.1063/1.1719961 (1966).
Kim, J. E., Kim, H. O., Bae, K., Choi, D. S. & Nickel, D. T1 mapping for liver function evaluation in gadoxetic acid-enhanced MR imaging: Comparison of look-locker inversion recovery and B(1) inhomogeneity-corrected variable flip angle method. Eur. Radiol. 29, 3584–3594. https://doi.org/10.1007/s00330-018-5947-4 (2019).
Yoon, J. H. et al. Quantitative assessment of liver function by using gadoxetic acid-enhanced MRI: Hepatocyte uptake ratio. Radiology 290, 125–133. https://doi.org/10.1148/radiol.2018180753 (2019).
Hargreaves, B. A. et al. Metal-induced artifacts in MRI. AJR Am. J. Roentgenol. 197, 547–555. https://doi.org/10.2214/AJR.11.7364 (2011).
Talbot, B. S. & Weinberg, E. P. MR imaging with metal-suppression sequences for evaluation of total joint arthroplasty. Radiographics 36, 209–225. https://doi.org/10.1148/rg.2016150075 (2016).
Frahm, C., Gehl, H.-B., Melchert, U. H. & Weiss, H.-D. Visualization of magnetic resonance-compatible needles at 1.5 and 0.2 Tesla. Cardiovasc. Interv. Radiol. 19(5), 335–340. https://doi.org/10.1007/BF02570186 (1996).
Muller-Bierl, B., Graf, H., Lauer, U., Steidle, G. & Schick, F. Numerical modeling of needle tip artifacts in MR gradient echo imaging. Med. Phys. 31, 579–587. https://doi.org/10.1118/1.1640971 (2004).
Lewin, J. S. et al. Needle localization in MR-guided biopsy and aspiration: Effects of field strength, sequence design, and magnetic field orientation. AJR Am. J. Roentgenol. 166, 1337–1345. https://doi.org/10.2214/ajr.166.6.8633445 (1996).
Thomas, C. et al. Carbon fibre and nitinol needles for MRI-guided interventions: First in vitro and in vivo application. Eur. J. Radiol. 79, 353–358. https://doi.org/10.1016/j.ejrad.2010.07.007 (2011).
Wachowicz, K., Thomas, S. D. & Fallone, B. G. Characterization of the susceptibility artifact around a prostate brachytherapy seed in MRI. Med. Phys. 33, 4459–4467. https://doi.org/10.1118/1.2364052 (2006).
Liu, H., Hall, W. A., Martin, A. J. & Truwit, C. L. Biopsy needle tip artifact in MR-guided neurosurgery. J. Magn. Reson. Imaging 13(1), 16–22. https://doi.org/10.1002/1522-2586(200101)13:1%3c16::aid-jmri1003%3e3.0.co;2-b (2001).
Reichenbach, J. R., Wurdinger, S., Pfleiderer, S. O. R. & Kaiser, W. A. Comparison of artifacts produced from carbon fiber and titanium alloy needles at 1.5 T MR imaging. J. Magn. Reson. Imaging 11(1), 69–74. https://doi.org/10.1002/(sici)1522-2586(200001)11:1%3c69::aid-jmri11%3e3.0.co;2-8 (2000).
Yamada, A. et al. Magnetic resonance and ultrasound image-guided navigation system using a needle manipulator. Med. Phys. 47, 850–858. https://doi.org/10.1002/mp.13958 (2020).
Singh, S., Torrealdea, F. & Bandula, S. MRI guided intervention: Evaluation of MR conditional Biopsy and Ablation Needle Tip Artifacts at 3T using a balanced fast field echo sequence. J. Vasc. Interv. Radiol. https://doi.org/10.1016/j.jvir.2021.03.536 (2021).
Okamoto, S. et al. Needle artifact characteristics and insertion accuracy using a 1.2T open MRI scanner: A phantom study. Diagn. Interv. Imaging. https://doi.org/10.1016/j.diii.2020.12.007 (2021).
Hue, Y. K. et al. Magnetic resonance mediated radiofrequency ablation. IEEE Trans. Med. Imaging 37, 417–427. https://doi.org/10.1109/tmi.2017.2753739 (2018).
Farina, L., Nissenbaum, Y., Cavagnaro, M. & Goldberg, S. N. Tissue shrinkage in microwave thermal ablation: Comparison of three commercial devices. Int. J. Hyperthermia 34, 382–391. https://doi.org/10.1080/02656736.2017.1362115 (2018).
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Contributions
V.F.S., O.D., M.S., M.A., and P.M.K. conceived the experiments. V.F.S. and P.M.K. conducted the experiments. V.F.S. and F.A. analyzed the data. V.F.S. and P.M.K. jointly wrote the manuscript. All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Schmidt, V.F., Arnone, F., Dietrich, O. et al. Artifact reduction of coaxial needles in magnetic resonance imaging-guided abdominal interventions at 1.5 T: a phantom study. Sci Rep 11, 22963 (2021). https://doi.org/10.1038/s41598-021-02434-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-02434-5
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.