Introduction

In recent years, magnetometers have been extensively used in a wide range of applications, including biomedical, automotive industry, robotics, and nondestructive material testing1,2,3,4. In addition, the significant improvement in the performance of magnetic sensors, such as excellent sensitivity, high resolution, minimum detectable magnetic field, low power consumption, high stability, wide bandwidth, small size, low cost, and excellent linearity, has allowed them to be used in navigation (1 nT–600 μT), biomagnetic (100 fT–0.1 μT), and archeology (1 pT–700 µT) applications5. Some applications, such as inertial navigation system require high sensitivity and low noise6, whereas small size sensors are well-suited for probing microscopy7. In scanning Hall probe microscopy, miniature magnetic sensors can be integrated with temperature sensors for temperature compensation, thereby reaching a better trade-off between magnetic performance and spatial resolution1,7.

Various types of magnetic sensors have been developed, such as magnetoresistive and Hall-effect magnetometers, which have attracted attention due to their high sensitivity and reliability8,9. However, these magnetometers suffer many challenges, such as hysteresis effect and incompatibility of the material with standard manufacturing processes10. Micro-electromechanical Systems (MEMS) magnetometers with low power consumption and low cost have emerged as a promising alternative11,12,13. In addition, these magnetometers can be implemented on silicon without the need for any special magnetic materials13. Recently, considerable efforts have been focused on resonant Lorentz-force MEMS magnetometers14,15,16,17,18,19, which are based on measuring the resonance frequency of a microstructure using optical20,21, piezoresistive17,22, and capacitive techniques23,24.

The optical sensing technique needs a simple read-out circuitry and has good immunity to electromagnetic interference23. However, this method is bulky and cannot be easily integrated with the sensor for mass production. The piezoresistive method is a simple sensing technique that requires an easy fabrication process23. Nevertheless, the technique can have voltage offset and temperature dependence issues5. The capacitive sensing technique is the most used owing to its low noise and low power consumption25. However, it suffers from parasitic capacitances. Still it has low temperature dependence19.

Efforts have been made to improve the sensor performance, such as its sensitivity and resolution, by using different approaches, which can be classified into extrinsic and intrinsic techniques. The extrinsic technique involves the new modulation and control systems used to drive the MEMS Lorentz force sensors, either using open- or closed-loop12,26,27,28,29,30. The closed-loop aims to operate the sensor exactly at mechanical resonance26. Using the closed-loop control improves significantly the dynamical behaviour of the MEMS sensors, which enhances the measurement accuracy31. However, this approach needs a lock-in amplifier that increases the circuit complexity and complicates the integration of the control system with the MEMS device32. Compared to the closed-loop, the open-loop approach is compact with less complexity for the driving and sensing electronics in spite of the sensitivity to parameter uncertainties of the input signal32. On the other hand, the intrinsic technique can be attributed to the device itself using advanced MEMS capabilities23. Many research studies have been reported to amplify the Lorentz force and thus boost sensitivity using internal amplification17 or micro-leverage mechanism14. Moreover, several MEMS dynamical phenomena have been utilized to improve the MEMS sensor sensitivity19,33,34, such as veering and buckling.

In a previous work, we proposed a highly sensitive Lorentz-force magnetic micro-sensor based on an in-plane electrothermally heated initially curved micro-beam exhibiting the veering phenomenon between its first two symmetric vibration modes35. In a recent work, we presented a miniature highly sensitive wide-range resonant magnetic Lorentz-force micro-sensor based on measuring the resonance frequency of straight heated micro-beam operating around the buckling zone36. The resonance frequency is measured with optical sensing (laser) and the sensor is operated at ambient pressure. In this paper, we explore the performance of a similar sensor to detect experimentally low magnetic field using open loop control system (capacitive sensing). We show that the proposed sensor is capable of detecting a magnetic field of 100 μT. In addition, a Galerkin-based reduced-order model (ROM) is developed to understand and verify the micro-sensor performance, which can further help in the designing of ultra-high sensitivity magnetometer.

Micro-sensor working description and principle of operation

Figure 1a shows an optical image of the micro-sensor. It consists of a clamped–clamped straight micro-beam fabricated from SOIMUMPs process by MEMSCAP37. The fabrication process is based on highly doping 30 µm Si device layer of Silicon On-Insulator SOI substrate. As seen in Fig. 1a, the micro-beam is sandwiched between two electrodes to electrostatically induce in-plane vibration and to detect the output AC current. Figure 1b shows a cross-sectional view of a SOIMUMPs processed device similar to the studied one. The micro-beam has a 600 μm in length (L), 2 μm in width (h), and 30 μm in thickness (b). The gap (g) between one of the electrodes and the resonating micro-beam is 7 μm. Applying an electrothermal voltage (VTh) between the anchors of the micro-beam, generates an induced current (ITh), through Joule’s heating, which causes a compressive axial stress38. Increasing ITh leads to increase in the compressive axial load, and thus the resonance frequency (stiffness) of the micro-beam decreases. Increasing further ITh, the resonance frequency decreases until reaching a minimal value (very low stiffness). This is known as the buckling phenomenon39, see Fig. 1c. In addition, the micro-beam is electrostatically actuated with a static DC voltage to avoid the dip of the resonance frequency near buckling (bias the buckling), Fig. 1c.

Figure 1
figure 1

(a) Optical image of the in-plane clamped–clamped micro-beam with the geometrical parameters. (b) Cross-sectional view of the fabricated structure by SOIMUMPs. (c) Schematic illustrating the shift in the first resonance frequency of the micro-beam under input current (ITh), bias DC static voltage (VDC), and flux magnetic field (B).

In this work, the proposed MEMS magnetometer is based on the Lorentz force principle. Hence, to generate a constant magnetic flux B, the device is placed inside two identical circular magnetic coils, which is measured using a Gaussmeter with a commercial Hall probe. A schematic of the experimental setup and a photo of the experiment platform are shown in Fig. 2a,b, respectively. As seen in the Fig. 2a, to provide an AC harmonic voltage (VAC), the drive electrode is connected to the network analyzer output (Keysight N9913A). To sense the output AC motional current (IO) due to the micro-beam in-plane motion, the sense electrode is connected to the input of the network analyzer. In addition, we use a low-noise amplifier (LNA SR560) to amplify IO. Note that the existence of VDC creates an initial deflection, which enhances the output motional current. As we can see from Figs. 1a and 2a, with the presence of a magnetic field (B) in the -z-axis and with a DC current (ITh) in the + x-axis, Lorentz-force (FL) is generated normal to the micro-beam in the + y-axis. As shown in21,36, to achieve maximum sensitivity, the Lorentz force is chosen to be in the same direction of the initial deflection and buckling direction (+ y-axis).

Figure 2
figure 2

(a) Schematic of the experimental setup. (b) Photo of the experiment platform.

Dynamic analytical model

Assuming an Euler–Bernoulli nonlinear beam model, the structural in-plane behaviour of the micro-beam is governed by the following equation and boundary conditions:

$$\left\{\begin{array}{c}EI\frac{{\partial }^{4}w\left(x,t\right)}{\partial {x}^{4}}+\rho A\frac{{\partial }^{2}w\left(x,t\right)}{\partial {t}^{2}}+\tilde{c }\frac{\partial w\left(x,t\right)}{\partial t}=\left(\frac{EA}{2L}{\int }_{0}^{L}\left[{\left(\frac{\partial w}{\partial x}\right)}^{2}\right]dx-\frac{EA}{L}{\int }_{0}^{L}\alpha \left(T\left(x\right)-{T}_{amb}\right)dx\right)\frac{{\partial }^{2}w\left(x,t\right)}{\partial {x}^{2}}+\\ +\frac{{\varepsilon }_{0}b{V}_{DC}^{2}}{2{\left(g-w\left(x,t\right)\right)}^{2}}+B{I}_{Th}\\ w\left(x=0,t\right)=w\left(x=L,t\right)=0; \, \, \, \, \, \frac{\partial w\left(x=0,t\right)}{\partial x}=\frac{\partial w\left(x=L,t\right)}{\partial x}=0; \, \, \end{array}\right.;$$
(1)

where \(w\left( {x,t} \right)\) denotes the in-plane (along y-direction) displacement in space x and time t, ρ is mass density, E is the Young’s modulus, A = bh and I = bh3/12 are the cross-sectional area and second moment of inertia, \(\tilde{c}\) represents the overall viscous damping coefficient, α is the thermal expansion coefficient, \({T}_{amb}\) is the ambient temperature, and \({\varepsilon }_{0}\) is the dielectric constant of air. The space-dependent temperature profile \(T\left( x \right)\) can be obtained through solving the steady-state heat transfer equation as detailed in40.

The resultant equations and boundary conditions system, Eq. (1), are numerically discretized using a Galerkin’s modal based expansion technique41. Accordingly, the micro-beam deflection is expanded as follows:

$$w\left(x,t\right)={\sum }_{i}{\psi }_{i}\left(x\right){u}_{i}\left(t\right); \, (1\le i\le N)$$
(2)

where \(\psi_{i} \left( x \right)\) represents the mode-shape of an un-forced beam and \(u_{i} \left( t \right)\) is the corresponding unknown time-dependent modal coordinate amplitude. Next, Eq. (2) is substituted into Eq. (1). After that, the resultant outcome is multiplied by the mode-shape functions \(\psi_{j} \, (1 \le j \le N)\) and the subsequent N-equations are then integrated from 0–L. This produces N Ordinary-Differential Equations (ODEs) in \(u_{j} \left( t \right) \, (1 \le k \le N)\) forming a Reduced-Order-Model (ROM). A convergence analysis was carried out and it was found that five modes (N = 5) is sufficient for convergence. The derived ROM can be used to calculate the micro-resonator natural frequencies. Toward this, the static deflection of the beam is first calculated. For this, we evaluate the stationary deflection by setting all time dependent terms in the ROM equal to zero. Then the modal amplitudes \(u_{i} \left( t \right)\) are replaced by unknown constant coefficients ak. This leads to a system of nonlinear algebraic equations that can be solved numerically using Newton–Raphson algorithm. Finally, the eigenvalue problem of the micro-resonator is solved by perturbing the response around the calculated static position and considering only the linear part of the ROM. Further detail of the eigenvalue problem can be found in42.

Results and discussion

To perform the measurements, the magnetometer is placed inside a vacuum chamber with a pressure of 6 mT. The DC voltage VDC and the electrostatic driving rms amplitude VAC are set to 30 V and 9.988 mVrms (− 27 dBm). Figure 3a shows the experimental and analytical results of the variation of the fundamental resonance frequency of the micro-beam while changing the electrothermal current ITh. Good agreement is shown among the results. It can be seen that the resonance frequency decreases with the increase of ITh until it reaches the buckling zone (ITh = 0.91 mA). Within this range, the resonance frequency varies from 52 to 35 kHz, confirming that the presence of VDC causes perturbation around the buckling zone (imperfect buckling). Thus, the system shows a midpoint displacement of 0.25 μm (initial curvature), see the inset of Fig. 1a. Hence, operating around this zone, the micro-sensor is very sensitive to any stiffness change (for instance from the magnetic distributed forces). After the imperfect buckling zone, the resonance frequency increases due to the induced curvature of the micro-beam.

Figure 3
figure 3

(a) Experimental and analytical results of the resonance frequency variation of the micro-beam with the electrothermal current ITh. (b) The simulated mode shape for VDC = 30 V and for various of ITh. The inset shows the analytical result of the static deflection of the micro-beam with VDC = 30 V and ITh = 0 mA.

To further investigate the dynamic response of the system with ITh, we plot the mode shapes of the micro-beam for VDC = 30 V and for various ITh, Fig. 3b. It is clear that, before the buckling point, there is no effect of the compressive axial load on the mode shapes.

Here, the buckling bifurcation will be used to sense the out-of-plane magnetic field in the low ranges. The Lorentz-force alters the micro-beam stiffness, which increases its resonance frequency. This frequency shift forms the basis of the proposed micro-sensor. Figure 4a shows the variation of the normalized resonance frequency (Δf/f0) variation versus input magnetic field (B) for ITh = 0.81 mA (around buckling point) where Δf is the frequency shift, which is defined as (ff0). f and f0 are the resonance frequency of the micro-beam with and without B. In Fig. 4a, the sensitivity can be evaluated as the slop of the variation Δf/f0 with B. As seen in Fig. 4a, the device capable of sense a wide range of B from 100 μT to 4 mT. Also, it can be seen that the sensor has high sensitivity (S) of 11.6 T−1. Further, we check the linearity of the proposed micro-sensor. Figure 4a demonstrates that the micro-sensor has high linearity less than 0.5% in the entire magnetic field range. Next, we plot the results for ITh = 0.67 mA, inset of Fig. 4a. The result shows a measured S of 10.8/T with 2% of non-linearity. This value is lower than the value near the buckling point 0.81 mA. The analytical results of the variation Δf/f0 with B while changing ITh are shown in Fig. 4b. The results agree with the experimental data, showing a high sensitivity of 11.92 T−1 at the buckling point (ITh = 0.91 mA).

Figure 4
figure 4

(a) Experimental results of the normalized resonance frequency variation versus input magnetic field for ITh = 0.81 mA. The inset shows the normalized resonance frequency shift versus input magnetic field for ITh = 0.67 mA. The slope of the plot presents the sensitivity of the device (S). (b) Analytical results of the variation of the normalized resonance frequency shift of the micro-beam for different ITh values. (c) Experimental and analytical results of S versus ITh.

To further analyze the effect of the operating current on the sensitivity, Fig. 4c illustrates the analytical and experimental results of S with ITh before the buckling zone. The theoretical results match well with the experimental data. As expected, S increases with the increase of ITh. At 0.91 mA, the sensor achieves a sensitivity that is 14 times higher than the sensitivity of 0.2 mA. The results confirm well the concept that by operating at the buckling point, the Lorentz-force magnetometer is extremely sensitive to out-of-plane magnetic fields. According to the results presented in Fig. 4b,c, the optimized sensitivity of the device can be achieved by carefully choosing the operating current ITh to reach the bifurcation point.

Next, we discuss the total power consumption of the proposed magnetic sensor. It results from the combination of the electrostatic drive power of the resonator (in order of nW) and the electrothermal power43 (in order of mW). Thus, the overall power consumption is dominated by the electrothermal voltage. It is estimated based on resistive heating of the resonator (VTh2/R) where R is the micro-beam resistance. For 600 µm micro-beam, R is found to be 1.49 KΩ and with ITh of 0.91 mA, the magnetometer power consumption (P) is around 1.2 mW. Moreover, ITh can be decreased to decrease the power dissipation by Joule’s effect. This will be on the expense of decreasing the sensitivity, i.e., SITh. It is important to note that there are many ways to improve the micro-sensor performance, including designing devices with low R, which reduces the dissipation and enhances the sensitivity.

Note here that the current detection method using capacitive sensing may suffer from parasitic capacitances and other sources of noise. Moreover, this sensing method can have an impact on the resolution of measurements. The measured quality factor Q of the sensor is found to be around 200. Thus, for the noise analysis of the system, we assume only the thermo-mechanical white noise (Q > 100 for frequency-modulated sensors) since the system shows a low noise voltage related to the LNA instrumentation amplifier around 4 nV.Hz−1/244. Thus, the evaluated thermo-mechanical noise (Brownian limited resolution) of the proposed magnetic sensor can be expressed as15,45:

$$Noise=\frac{\sqrt{4{k}_{B}Tb}}{L{I}_{Th}}$$
(3)

where kB is Boltzmann’s constant, T is absolute temperature in Kelvin, and b is the damping coefficient (b = 0.0025). The theoretical Brownian noise is found to be 13.6 µT.Hz−1/2 for a 0.81 mA input current. Clearly, a longer micro-beam can decrease the noise limit. The Brownian limited resolution may be decreased further by increasing the power or by lowering the damping of the system. Hence, the micro-sensor design can be adjusted to improve the noise performance.

As we notice above, the thermo-mechanical noise is dependent on Q, and its magnitude can be affected by the input current. Hence, we measured the Q change versus ITh for VDC = 30 V and 40 V, see Fig. 5a. The results show an increase of Q as increasing ITh or VDC. It is clearly noticeable that the sensor has the highest Q at the buckling zone. Thus, operating around the buckling zone, the proposed magnetometer exhibits high sensitivity, good linearity, and also high quality factor.

Figure 5
figure 5

(a) Measured quality factor (Q) versus ITh for VDC = 30 V and 40 V. (b) Simulated results of the temperature dependence of frequency (TCF) of micro-resonator.

Also, high-performance magnetic sensors should have high sensitivity over a wide temperature range. To investigate the effect of the temperature on the sensor, we simulate the temperature dependence of the resonance frequency from 0 to 45 °C, see Fig. 5b. As can be seen in Fig. 5b, the results show a temperature coefficient of frequency (TCF) of − 0.0055/°C. For the temperature compensation, we can use a highly n-doped silicon when the temperature coefficient of the electrical stiffness becomes large enough to compensate or supersede the effect of the temperature coefficient of the mechanical stiffness46. However, at high doping, the increase of electrical stiffness may increase the operating point to higher bias currents.

Further, analytical simulations are performed to predict the device performance. Figure 6 illustrates the effect of VDC on the power consumption and sensitivity. Figure 6a shows the variation of the resonance frequency while varying ITh for different values of VDC. The inset depicts a comparison between analytical and experimental results for VDC = 30 and VDC = 25 V.

Figure 6
figure 6

(a) Analytical results of the resonance frequency variation of the micro-beam with ITh for various values of VDC. The inset shows the analytical and experimental variation of the resonance frequency for different ITh for VDC = 30 V and 25 V. (b) Results of the normalized resonance frequency variation versus B for different bias VDC. (c) Results of the variation of S versus VDC.

As seen in the figure, as VDC decreases, the corresponding resonance frequency drops until reaching a value of 17 kHz. In addition, its corresponding operating point increases to 1.05 mA. For low VDC value (@ 10 V), the frequency response has the sharpest curve at the perturbed buckling zone (low stiffness), which can be extremely sensitive to any output forces. Figure 3b shows the results of Δf/f0 variation versus B for different bias VDC.

As can be seen in Fig. 6b, the increase in Δf/f0 follows a linear function with B for different values VDC. For further study of the effect of VDC on the sensor sensitivity, we plot S as increasing VDC, Fig. 6c. The results show that as the actuation VDC increases, the S decreases as a quadratic polynomial. Subsequently, first S drops rapidly, and then slowly decreases, as VDC increases. Decreasing VDC from 30 to 10 V, the sensitivity can be further improved by 920%. Indeed, these results demonstrate that the sensitivity is highly dependent on the applied VDC. In the proposed sensor with high airgap, adding high value of VDC is necessary to improve the motional current on the sensing electrode. Accordingly, future efforts should minimize electrostatic driving VDC with ultra-small air-gap size to make the output signal large enough to be sensed.

As we mention above, the device sensitivity is defined as the fractional change in frequency per unit of magnetic field density, which can be expressed as14,15,30:

$$S=\frac{\partial \left(\frac{\Delta f}{{f}_{0}}\right)}{\partial B}=\frac{{I}_{Th}L}{2k}$$
(4)

where k is the micro-beam’s stiffness. As can be seen from Eq. (4), S can be further enhanced by modifying the physical dimensions of the device more carefully or taking advantages from more complaint micro-beams. On the other hand, as implied by Eq. (4), increasing L or reducing h increases S. Hence, it responds more to the Lorentz-force for the same given magnetic flux. This explains that the device performance depends strongly on the geometrical parameter of the micro-beam. In the following, we investigate the influence of the micro-beam length on the resonance frequency and sensitivity. Based on the analytical model, we increase L from 700 μm to 900 μm and compute the resulting resonance frequencies. Figure 7 displays the results as tuning ITh and VDC for micro-beams with different lengths L. This figure shows that the initial natural frequency at ITh = 0 mA decreases as L increases. In addition, we observe that at ITh = 0 mA, the frequency responses increase with VDC for micro-beam lengths 800 μm and 900 μm, respectively. This is due to the mid-plane stretching that is dominate over electrostatic force for a longer micro-beam and large capacitive gap. As noted from the figure, increasing L shifts the position of the buckling point from high to low values of ITh. This can be explained from the increase of the micro-beam’s resistance and the decrease of its stiffness. Hence, these results can be useful to reduce the energy consumption. Increasing the length to 900 μm can reduce the power consumption by 94% (P = 0.07 mW). Operating around the buckling point, Fig. 8 presents the analytical results of the variation of S as a function of VDC and for different lengths L. As shown, increasing L results in the decrease of the sensor sensitivity. S decreases to the minimum of 2.4 T−1 for L of 900 μm and VDC of 30 V. We conclude here that to improve the sensor sensitivity, the micro-beam length L and the input current ITh should be increased (SITh L). However, in both cases there is the cost of increased power dissipation by Joule’s effect. It also shows that the operating current of the magnetic sensor and VDC can be tuned to achieve higher sensitivity.

Figure 7
figure 7

The simulated variation of the resonance frequency versus ITh for various values of VDC and different micro-beam’s length L (a) L = 700 μm, (b) L = 800 μm, and (c) L = 900 μm. The insets show the corresponding S versus voltage VDC.

Figure 8
figure 8

Sensitivity variation with VDC for different micro-beam length L.

These results can be used as guidelines to choose the geometric parameters to improve the sensor performance. The power consumption can be reduced by a factor of 2 if a 900 μm long beam is used. Moreover, the device sensitivity is expected to improve after reducing VDC. More analytical studies can be conducted to investigate the effects of the length-to-thickness (L/h) ratio on the energy dissipation and sensitivity. One should mention that the proposed sensor design is flexible and can be optimized to measure ultra-low biomagnetic fields’ applications.

Table. 1 shows the performance comparisons of the proposed micro-sensor with some of the reported Lorentz-force-based magnetometers. For a fair comparison, only MEMS magnetometers with frequency-modulation (FM) output, which the proposed sensor belongs to, have been included. It can be seen from Table 1 that the proposed miniature magnetic sensor has higher sensitivity compared with the state of the art. Moreover, the high sensitivity, low-power consumption, and good linearity, make the sensor promising for wide range of applications.

Table 1 Summary of the sensor’s performance of this work compared with some of the reported frequency modulation (FM) Lorentz-force-based magnetic micro-sensors.

Conclusions

In this paper, we proposed an out-of-plane Lorentz-force magnetic micro-sensor by exploiting the buckling point bifurcation of a straight micro-beam. The concept is based on the frequency shift measured at different magnetic field strengths. We showed that the sensitivity of the proposed magnetic sensor is very high. In addition, we demonstrated analytically an optimal sensor design whose sensitivity can be further enhanced to around 920%. Also, the proposed sensor shows good linearity. These features, combined with low-power consumption, simplicity of fabrication, low-cost, and scalability, make the sensor promising for variety of applications. We demonstrated experimentally the capability of the magnetometer to sense a low magnetic field around 100 μT using open loop control system. Future research work will include a closed loop control system to improve the noise performance. It was found that the performance of the proposed magnetometer depends strongly on the micro-beam length, operating point, and the electrostatic driving voltage. One should mention that the micro-sensor design is flexible and can be optimized to measure ultra-low biomagnetic fields’ applications.